Message ID: 355     Entry time: Thu Apr 25 15:05:19 2019
Author: Joe 
Type: Optics 
Category: Characterization 
Subject: Looking at PZT HOM spacing dependance and thinking about workflow 

[koji, joe]

The template or glass breadboard was wobbling, and we noticed that the caivty alignment became worse/better when it was pressed down. We saw that it was the glass breadboard, so it was fixed into the transport fixture more securely. Now its alignement didn't change when it was pressed down. We took a pzt mirror out and replaced it, the alignment din't change much so that was good. We set up posts to hold the pzt wires.

We noticed that the bottom of the mirrors were dirty, so we cleaned them, and once we were happy with the newton rings, we aligned the cavity

Took a photo of CM2, the spot is maybe 1 beam diameter vertically and horizontally from the centre, and quite a bright spot could be seen. The same problem with CM1. We thought it would be good to see a measurement of higher order mode spacing dependence on PZT DC voltage rather than doing the full characterisation since the alignment seems to change quite a lot when ever we do anything, and this cavity arrangement probably isn't very good anyway (can see scattering on both curved mirrors with the IR camera). 

did measurements of FSR, = 2.64835MHz

did HOM spacing for 0,75,150V on CM1 in pitch and yaw.

we want to come up with a work flow for how to do these measurements, and make automate parts of the analysis?

 