Entry  Fri May 26 21:53:20 2017, Koji, General, Configuration, Trans RF PD setup 
    Reply  Tue Jun 6 00:49:48 2017, Koji, General, Configuration, Trans RF PD setup 
       Reply  Tue Jun 6 22:00:36 2017, Koji, General, Configuration, Trans RF PD setup 
Message ID: 280     Entry time: Tue Jun 6 22:00:36 2017     In reply to: 279
Author: Koji 
Type: General 
Category: Configuration 
Subject: Trans RF PD setup 

- Replaced the PZT with the one used from the beginning. This must be PZT #21. After the replacement, the spot positions look very good. I even went up. So I decided this is the configuration to proceed to the gluing. The CM1 mirror has the HR arrow at the top.

- The input beam was realigned w.r.t. the OMC.

- Tried to use the IR viewer with the new rechargable battery brought from the 40m. But the view still didn't work. The possibility is a) the viewer is broken b) the battery is empty.

- Tried to use the stainless clean regulartor for the UHP N2. The outlet has a short tube with a different diameter. The O.D. of the old tube is 6.3mm, while the new one is 9.5mm. If I insert the thinner tube in the new tube, it approximately fits. But I don't believe this is the way...