Message ID: 205     Entry time: Thu Jul 10 23:22:28 2014
Author: Koji 
Type: Optics 
Category: Characterization 
Subject: I1OMC QPD 

QPD#              QPD1       QPD2
Housing#          #006       #007
Diode#            #50        #51
Shim              1.25mm 03  1.25mm 02   (1.25mm = D1201467-10)

-------------------------------------
Power Incident    123.1-13.0 uW  124.5-8.0 uW
Sum Out            77.0 mV   82.5 mV
Vertical Out      -24.0 mV  - 8.8 mV
Horizontal Out      4.2 mV    9.0 mV
SEG1              -11.6 mV  -16.0 mV
SEG2              -12.6 mV  -18.0 mV
SEG3              -25.2 mV  -24.4 mV
SEG4              -21.4 mV  -21.4 mV
-------------------------------------
Spot position X   -21   um  -19   um  (positive = more power on SEG1 and SEG4)
Spot position Y   +102  um  +47   um  (positive = more power on SEG3 and SEG4)
-------------------------------------

Responsivity[A/W] 0.70      0.71
Q.E.              0.82      0.83
-------------------------------------

Arrangement of the segments
View from the beam
/ 2 | 1 X
|---+---|
\ 3 | 4 /

---------------

I(w,x,y) = Exp[-2 (x^2 + y^2)/w^2]/(Pi w^2/2)

(SEG_A+SEG_B-SEG_C-SEG_D)/(SEG_A+SEG_B+SEG_C+SEG_D) = Erf[sqrt(2) d/w]

d: distance of the spot from the center
w: beam width