Message ID: 184     Entry time: Wed May 14 02:15:15 2014
Author: Koji 
Type: Optics 
Category: Characterization 
Subject: FSR/TSM adjustment of the OMC cavity  

1. FSR was adjusted and measured with "the golden arches" technique again.

FSR = 264.8412 MHz +/- 1400Hz => Lcav = 1.13197 m. (nominal 1.132m)

2. Transverse mode spacings for the vertical and horizontal modes were measured.

TMS/FSR = 0.218144 (V) / 0.219748 (H)

This is almost perfect!

The 19th-order lower sideband hit the resonance. Next step is to glue some of the flat mirrors.