Message ID: 180     Entry time: Mon Mar 3 02:46:21 2014
Author: Koji 
Type: General 
Category: Characterization 
Subject: Spot positions scanned 

Spot positions on CM1 and CM2 scanned according to the recipes provided by the previous entry.

The best result obtained was:

Transmission from FM2: 32.7mW
Incident on BS1: 34.4mW

Reflection (Unlocked): 5.99V
Reflection (Locked): 104mV
Reflection (Dark): -7.5mV

to accomodate the spot on BS1 it had to be about a mm moved from the template.

This gives us:
- Portion of the TEM00 carrier: R = 1-(104+7.5)/(5990+7.5) = 0.981
- Raw transmission: 32.7/34.4 = 0.950
- TEM00 transmission 0.950/R = 0.969
- Excluding the transmission of BS1: 0.969/0.9926 = 0.976
  => loss per mirror ~40ppm

 