40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  Cryo Lab eLog, Page 2 of 62  Not logged in ELOG logo
ID Date Author Type Category Subject
  3089   Mon Feb 6 10:45:47 2023 aaronUpdatePSOMABake out of the PSOMA Chamber.

After an extended bakeout, the PSOMA chamber pressure is now consistently ~1e-7 torr.

  3088   Fri Feb 3 14:27:58 2023 JCUpdatePSOMABake out of the PSOMA Chamber.

Baking the PSOMA Chamber.

Preperation of Baking out the Chamber
To bake out the chamber, we need the following materials:
* Variac - This controls the voltage that supplied to the heater tape
* Heater Tapes - We used 6 heater tapes we uses individual outlets.
* Fiberglass Electrical Tape - We use [McMaster 7574A13](https://www.mcmaster.com/7574A13/)
* Powerstrip - This connect to the Variac and has as many outlets and you have cables

I started off by arranging the heater tape along the lid’s surface. I used to cover the lid and another 3 to wrap around the body of the chamber. When wrapping the heater tape, I had to be aware to keep the tape from interacting with itself and other routings. To press the tape nicely against the chamber, I used the Fiberglass Electrical Tape. The chamber with heater tape is shown in Attachment #1.



After wrapping the chamber nicely, I covered the chamber with aluminum foil. I didn’t really care if I punctured holes in the foil, this is mainly just to help the chamber maintain it’s heat. 

Next, we turned on the Variac (with the heater tapes plugged into the Powerstrip, and power strip plugged into the Variac) and set this to 50V. The oven is now being baked

  3087   Fri Feb 3 13:30:31 2023 aaronDailyProgressControl Systemtwo new slow loops

I restarted cymac1 and ran rtreset, and somehow the right combination of nds/standalone_edc/daqd were restarted such that historical data are available.

I found some settings for the N laser temperature PID that avoid Acromag glitches (usually occurs when the nominal output value changes by a small amount near zero); the same settings were in use for the S temperature PID. I also used an SR560 to lowpass the N TEC control signal at 10 Hz, 24 dB/oct rolloff. Both laser current control signals are now zero-mean. This avoids the PLL saturating as the S laser temperature drifts to follow the cavity, and extends the time that PLL and PDH can be simultaneously locked to many minutes (maybe hours, we'll see).

To lock the relative phase of resonant pump and probe, I used a function generator to AM at 10 kHz the 300 MHz tone before N laser's EOM. When the pump and probe are out-of-phase (as desired), the 10 kHz tone is nulled on BEAT PD (it's entirely PM wrt the pump). So, I fed BEAT MON to an SR830, and fed the lock-in's output to the external mod port of the Marconi (to FM the 300 MHz tone). I found that the following filters and gains worked well (though probably will want to integrate the lock-in for longer so I can inject 10s Hz signals to the amplifier):

  • lock-in time constant 30 ms, 24 dB rolloff
  • 1 V sensitivity for lock in (max)
  • 100 kHz / V ext mod coefficient on Marconi
  • I adjusted the lock-in phase to maximize loop gain, ended up at -17 deg

I'll need to transfer the lock-in controller to CDS system or use another SR560 to sum the low-frequency drift control with a 40-400 Hz signal.

Eh, on further inspection those lock-in and FM settings had way too much loop gain for the low-frequency drift. When I tried injecting a ~100s Hz signal that should have appeared as an AM peak in BEAT MON, I was unable. I was just introducing enough noise such that the 10 kHz tones were being washed out; this makes sense, 100 kHz/V FM deviation * mVs residual noise = 100 Hz residual noise injected as FM. I reduced the gain but needs more work.

  3086   Fri Feb 3 11:26:11 2023 aaronDailyProgressControl SystemLaser hysterisis curve and noise changed after power outage

After the power outage, I was able to lock the cavity with PDH, but unable to lock the N laser to the S laser with PLL. At the same TEC and current settings that I was using before, the width of the beat note is now O(3 MHz). I found that increasing the TEC setpoint (decreasing laser temperature) for both lasers to the next hysterisis curve resulted in a much cleaner beat note O(100 kHz). This is closer to what I was seeing before the power outage.

Attachments

  1. TEC settings before the outage (lower end of ~1 kOhm range). Laser power on BEAT MON is ~253 uW.
  2. Beat note at the previous TEC settings
  3. New TEC settings (lower end of ~1 kOhm range). Laser power on BEAT MON is ~409 mV, so I'll need to adjust output HWP to avoid saturating REFL.
  4. Beat note at new TEC settings
Attachment 1: FA0BD187-DCD6-475F-BF90-053718CDBA24.jpeg
FA0BD187-DCD6-475F-BF90-053718CDBA24.jpeg
Attachment 2: 10DCC765-960E-4341-94FE-A40D5BA4802A.jpeg
10DCC765-960E-4341-94FE-A40D5BA4802A.jpeg
Attachment 3: E530ACF7-AA55-44E4-9F46-D313676E9FC3.jpeg
E530ACF7-AA55-44E4-9F46-D313676E9FC3.jpeg
Attachment 4: 012A9978-8208-4676-A8D3-EDA48E747DEB.jpeg
012A9978-8208-4676-A8D3-EDA48E747DEB.jpeg
  3084   Thu Feb 2 18:13:32 2023 aaronDailyProgressLab WorkPSOMA experiment plans for upcoming week

JC baked the aluminum breadboard in the large oven at 40m over the last 2.5 days. Today we retrived it from the oven, wrapped in UHV foil and mylar sheeting, and returned it to the cryo lab.

JC used the fancy high temperature tape to secure our heater strips to the chamber, wrapped the chamber in kitchen-grade aluminum foil, and is baking it out at 85 C.

  3083   Thu Feb 2 17:46:54 2023 aaronDailyProgressControl Systemsignal injection

I tuned up the PDH servo and it does seem to have helped the PLL remain locked for longer (residual 40 Hz in PDH is smaller, so the low-frequency dynamic range of the PLL is expanded a bit).

Played a bit with out FM/PM/AM on the Marconi's RF signal sent to the HP spectrum analyzer. The 0-order RF is nulled when PMing with 2.40 and 5.52 radians amplitude, consistent with the value of the 0-order Bessel J function. I also found a ZP-3+ (couldn't find a ZP-3MH, but the 3+ at least goes to DC and is level <17) mixer to use for amplitude modulating. The ZP-3+ allows up to 40 mA at the IF port, so I'm limiting my AM driving tone to <1 V peak. Applying 0.5 Vdc and a 0.01 Vpp tone at 10 kHz from the function generator seems safe, and I can just barely see second order AM sidebands on the spectrum analyzer. Applying both FM and AM of course generates intermodulation products between the two driving frequencies, which may not be desirable but I'll leave that aside for now.

I was able to lock both PDH and PLL, drive the N laser EOM at the PLL offset frequency to produce a resonant probe, AM the probe at 10 kHz, tune the PLL phase to adjust ensure the 10 kHz does not appear in BEAT MON spectrum (indicating that the resonant pump and probe are 90 deg out of phase, as desired), and finally FM the Marconi tone to produce the 'signal' tone entering the cavity.

I was planning to set up the lock-in and controller to maintain the PLL/Marconi phase such that the 10 kHz tone is purely PM at BEAT MON... but during the physics colloquium Caltech (and Pasadena) had yet another power outage, so I had to recover the lab instead. It seems some combination of daqd, standalone_edc, and nds needs to be restarted, because ndscope once again has no history. (CW: restarting the rts-nds service restored access to the history. The other intervention that was needed was to press the start button on the particle counter to restart its data collection.)

  3082   Wed Feb 1 12:07:15 2023 aaronDailyProgressControl Systemsignal injection

After some effort tuning loops, I lock both PDH and PLL and am set up for signal injection.

The PLL is drifting out of range and losing lock in < minute, so I'm going to wrap another integrator around N laser's current control to feed to temperature. This will keep the PLL control signal zero-mean.

To indicate that the PLL is locked, I'm using the Q quadrature out of the PLL mixer (assuming the quadrature used for locking is I). When the loop is locked, Q is maximized. I had some trouble closing this loop because the N laser driver HF mon I'm sending to cymac1 via DB9 has a large DC offset (unlike the identical channel for the S laser driver, which is zero-mean for zero-mean control signals). Not sure why this is, but there wasn't a simple combination of N laser current driver mons that provided a zero-mean control readback... very annoying, I'll have to use the BNC output instead which will prevent me from simultaneously measuring the high frequency control noise spectrum. I just manually nulled the offset, hopefully it doesn't drift.

In the end, this didn't work very well. The acromag DAC sends occassional glitches that cause the PLL to lose lock, which we also saw with the S temperature loop.

It could be that quieting the PDH loop will help the PLL maintain lock. TRANS MON and REFL MON are noisier than usual.

  3081   Tue Jan 31 15:57:37 2023 aaronDailyProgressControl Systemback to locking

Even with loop-shaping in Moku:Pro, passing the control signal through a unity-gain Moku:Go buffer resulted in oscillation at 10s of kHz.

I at first suspected this is due to extra input voltage noise on Moku:Go's input ports. The Moku:Pro does some smart blending of signals between its low-frequency and high-frequency ADC, which suppresses input voltage noise below ~10 kHz. Attached is a voltage spectral density for

  1. Top plot, live trace is Moku:Pro with nothing connected to its input port sent through +60 dB proportional gain filter to its output port, then to SR785 (approx input voltage noise of Moku:Pro * 1000)
  2. Top plot, reference trace is Moku:Go with nothing connected to its input port sent through +60 dB proportional gain filter to its output port, then to SR785 (approx input voltage noise of Moku:Go * 1000)
  3. Bottom plot, live trace is Moku:Pro with nothing connected to its input port sent through -60 dB proportional gain filter to its output port, then to SR785 (approx output voltage noise of Moku:Pro)
  4. Bottom plot, reference trace is Moku:Go with nothing connected to its input port sent through -60 dB proportional gain filter to its output port, then to SR785 (approx output voltage noise of Moku:Go)

The input voltage noise of Moku:Pro is about 4x less than that of Moku:Go between 100 Hz and ~10 kHz, mostly due to the ADC blending (similar input voltage noise curve available on liquid instruments' webpage about ADC blending).

If the loop instability were due to input voltage noise on Moku:Go, I'd expect to be able to mitigate this by boosting the signal in Moku:Pro and reducing gain in Moku:Go... but this did not significantly change the residual noise or oscillation.

Now I suspect the ADC/DAC handoff on a second device is just introducing too much loop gain. I think our phase margin was only at best 45-90 degrees with maybe 65 kHz UGF (can dig through the elog). A delay of ~4 us would completely eliminate the phase margin at the UGF. I did find adding some derivative stabilized the loop a bit, but couldn't completely eliminate oscillation. Next I'll check the delay when passing a signal through a Moku:Go buffer.

Attachment 1: 7C313A5E-BEF5-44CD-BF56-0B8E9077D91B.jpeg
7C313A5E-BEF5-44CD-BF56-0B8E9077D91B.jpeg
  3080   Tue Jan 31 11:46:00 2023 aaronDailyProgressControl Systemback to locking

We want to inject signal both at the input of our amplifier and at the PDH (and maybe PLL) error points to null the phase modulation in cavity reflection and avoid unnecessarily driving the pump frequency (S laser current).

Previously, we used 2x lock-in instruments and 2x PID controllers in our Moku:Pro to achieve lock. However, this configuration does not let us modulate the PDH error point.

Yesterday, I tried using 2x lock-in instruments in Moku:Pro to generate PDH and PLL error signals, then routing the signals to a Moku:Go for loop shaping and feedforward. Unfortunately, I wasn't able to lock either the PDH or PLL when passing the error signal between Moku:Pro and Moku:Go. I got both loops locking independently with the error signal shaped internal to the Moku:Pro, but the same combination of filters split between Moku:Pro and Moku:Go could not acquire lock. It could be that some combination of DC offset drift between the Moku boxes and input ADC noise is to blame, but I wasn't able to resolve the issue by shifting around where gain is applied between Pro and Go.

Today I'll try handling loop-shaping in Moku:Pro, and just do feedforward with the Moku:Go. This way, the low frequency signal is boosted above 1/f noise in the ADC.

  3079   Mon Jan 30 16:59:27 2023 shrutiNotesPSOMAfields and loops and quadratures
Attachment 1: FieldPropagationShort.pdf
FieldPropagationShort.pdf FieldPropagationShort.pdf FieldPropagationShort.pdf FieldPropagationShort.pdf FieldPropagationShort.pdf FieldPropagationShort.pdf
  3078   Fri Jan 27 17:11:25 2023 aaronDailyProgressLab WorkPSOMA experiment plans for upcoming week

[JC, aaron]

We retrieved four wooden blocks from the 40m to place under the vacuum chamber during bake out.

We then vented the chamber with dry nitrogen. Because we don't have a second up-to-air valve in the system, we loosened the KF flange on valve 1 to avoid overpressuring the chamber. While the chamber was open, we flowed nitrogen at 2.5 psi at the regulator. We also cleared out the area around the chamber, and disconnected the vacuum gauge from power.

We used the hoist to lift the chamber body and move it away from the bottom plate. We placed the bottom plate on wooden blocks wrapped in UHV foil, and removed the styrofoam and remaining wrapping that had been holding the plates.

We wiped the bottom o-ring and plate with 100% IPA. We photographed a large indentation in the bottom o-ring, which we suspect could be responsible for our leak. We also tried to remove the o-ring from the groove using teflon forceps applied at the notch in the o-ring groove, but were unable to do so. The o-ring still has vacuum grease inside the groove.

We lowered the chamber body back onto the bottom plate. We reattached the KF flange on valve 1 and gave it a new o-ring. After waiting ~20 minutes, we recalibrated the vacuum gauge's 'atmosphere' point (as it was reading only 100 torr or so at atmospher). After being disconnected from power for about an hour, then being allowed to equilibrate and calibrate at atmosphere, the gauge is once again giving 'lower noise' readings. Good to know.

The pumpdown is proceeding very slowly, and I'm restarting the roughing pump several times as it times out (as usual). The turbo kicked on after about an hour of pumping, so I changed the roughing cycle timer to 90 minutes to avoid having to repeatedly restart roughing. I suspect a leak, once the turbo spun up fully the vacuum pressure exponentially decayed to >10 utorr.

Attachment 1: Screenshot_from_2023-01-27_18-47-06.png
Screenshot_from_2023-01-27_18-47-06.png
  3077   Fri Jan 27 13:22:35 2023 aaronDailyProgressLab WorkPSOMA experiment plans for upcoming week

I did a rate-of-rise test on the vacuum now that we've swapped CF flanges. The vacuum pressure did stabilize at a level below 1e-4 torr this time, and squinting at the leak rate plot you can make out a downward trend... but I didn't even add a trend line because the gauge has become much noisier since I last ran this test. I don't know why the gauge is so noisy, but it has up to 300+% deviations upwards in pressure that last for less than a minute before going back to a baseline under 1e-4 torr (even with the pump valved off, so there shouldn't be any gas sinks in the system).

Since the rate-of-rise test is promising, I think we should bake out the chamber over the weekend instead of venting to clean the o-rings.  We need to vent and open the chamber anyway to put wooden blocks under the chamber for baking, so we'll just wipe the bottom o-ring today.

Attachment 1: pressure.pdf
pressure.pdf
Attachment 2: leak.pdf
leak.pdf
  3076   Thu Jan 26 17:54:35 2023 aaronDailyProgressLab WorkPSOMA experiment plans for upcoming week

Shruti and I finished replacing gaskets and bolts on the remaining 3 CF flanges. The subsequent pumpdown is still proceeding slowly enough that I expect a leak (despite not yet performing a rate-of-rise test). The next step is to clean and reseat the large viton o-rings on the top and bottom of the chamber.

 

  3075   Wed Jan 25 15:05:43 2023 aaronLab InfrastructureLab Workdry nitrogen arrived in lab

I received compressed dry nitrogen this afternoon. I connnected the regulator we had been using, and valved off the line going to the cryo cavs experiment (there was already a valve at the cryo cavs table). Despite the low-pressure-side gauge on the regulator having a 100 psi range, the regulator can only supply ~10 psi. This is awkward because I spec'ed the calibrated leak to fill at ~30 psi; it will still leak, but the rate will be lower. We could decide to swap regulators if it's taking too long to vent through the leak (I've seen the regulator on our He line supply 100 psi).

  3074   Wed Jan 25 11:51:47 2023 aaronDailyProgressLab WorkPSOMA experiment plans for upcoming week

After Maty gave us a tutorial on CF flanges, JC and I

  • vented the PSOMA chamber
  • removed 3x CF35 flanges
  • began installing new gaskets before realizing we didn't have the right socket size (1/4" worked finger tight but no further)
  • Covered remaining open gaskets and blanks in UHV foil, then broke for lunch so JC can retrieve supplies from 40m

One of the CF flanges we removed had at least one bolt that galled well before it was fully seated in the plate nut; others had minor galling on some bolts. We put all the steel bolts, washers, and plate nuts in a bag labelled "BAD," and can either throw out the entire batch or just the ones with noticeable galling.

In the afternoon, we installed 3 out of 4 CF35 flanges and 2 out of 4 CF125 flanges. We'll finish tightening the remaining 3 flanges tomorrow.

  3073   Wed Jan 25 10:34:19 2023 JcDailyProgressLab WorkPSOMA experiment plans for upcoming week

Aaron, Madeline, and I met up in CRYO to discuss what would be the best move to check the PSOMA chamber. Attachment 1 has the plans for the upcoming week and what we plan to do.

Attachment 1: CC43B4CE-A4E8-4760-B3E6-32EA827D6D2A.jpeg
CC43B4CE-A4E8-4760-B3E6-32EA827D6D2A.jpeg
  3072   Tue Jan 24 11:44:46 2023 shrutiDailyProgressPSOMAhow much AM do we expect?

Here is a notebook that takes into account the change in the interference condition and phase-lock phase offset to estimate how much AM and PM we expect.

(More details coming soon...)

  3071   Mon Jan 23 15:50:18 2023 aaronUpdateVacuumnitrogen line connected

We were supposed to receive dry, compressed N2 in cryo lab today... but it never arrived. I've asked JC to follow up with airgas (or whoever was supposed to deliver).

In the meantime, I found fittings to adapt the 1/4" VCR male flange on our calibrated leak to the 3/8" tubing that carries nitrogen to the PSOMA experiment (1/4" male VCR -> 1/4" female VCR coupling body -> 1/4" VCR to swagelok tubing adapter -> 1/4" gas line filter -> 1/4"-to-3/8" right angle connector). See photo for assembly. I borrowed the VCR connectors from Hutzler lab, and identified swagelok SS-4-WVCR-6-400-SC11 as the adapter we want to purchase.

Attachment 1: 266FD66A-1E3D-49FF-9CD5-3952619B8F5F.jpeg
266FD66A-1E3D-49FF-9CD5-3952619B8F5F.jpeg
  3070   Sun Jan 22 21:23:09 2023 aaronNotesPSOMAproposed next measurements for PSOMA cavity

True

Quote:

I don't think it's necessary to have to do the whole matrix calculation again for breaking up blocks with no cross-coupling because each block is just a product of sub-blocks... 

 

 

  3069   Fri Jan 20 07:41:23 2023 Not aaronNotesPSOMAproposed next measurements for PSOMA cavity

I don't think it's necessary to have to do the whole matrix calculation again for breaking up blocks with no cross-coupling because each block is just a product of sub-blocks... 

 

  3068   Thu Jan 19 15:28:19 2023 aaronNotesPSOMAproposed next measurements for PSOMA cavity

Additional notes on measuring k

To make this proposal more explicit... since it's easy to measure v_1, v_2, and the error signals in loop without additional calibration, we could estimate k by exciting at \gamma_1 and measuring

k=\frac{v_2}{v_1}*\frac{1-A_2B_2}{A_2B_1}=\frac{f_2}{f_1}*\frac{1-A_2B_2}{A_2B_1}

I'm assuming we inject the excitation in such a way that v_1,v_2 can be identified with their respective nodes 2,5. This seems like a reasonable measurement between ~1 kHz and ~100 kHz, but might be impossible at lower frequency near the cantilever resonance where it's hard to measure the loop transfer functions. I included the equivalent estimate based on f_2/f_1, but in practice we would measure the ratio of error signals and would need to add a node inside of B_i to derive the appropriate transfer matrix.

Including the error points in the transfer matrix

We could also excite at \gamma_2 or at the error points. Breaking B_i into the plant P_i and controller C_i and labelling the new nodes 7,8 corresponding to error signals e_1, e_2, we have a new open loop transfer matrix

\textsl{T}=\begin{pmatrix} 0 & 0 & 0 & 0 & 0 & 0 & P_1 & 0\\ 0 & 0 & A_1 & 0 & X & 0 & 0 & 0\\ 1 & 0 & 0 & -1 & 0 & 0 & 0 & 0\\ 0 & 0 & 0 & 0 & 0 & 0 & 0 & P_2 \\ 0 & 0 & 0 & 0 & 0 & A_2 & 0 & 0\\ k & 0 & 0 & 1 & 0 & 0 & 0 & 0\\ 0 & C_1 & 0 & 0 & 0 & 0 & 0 & 0 \\ 0 & 0 & 0 & 0 & C_2 & 0 & 0 & 0 \end{pmatrix}

The matrix \vec{S}=(1-T)^{-1} is identical to the matrix above in the 6x6 upper-left quadrant, except with all references to B_i\to C_iP_i. The rest of the matrix is

\vec{S}=\begin{pmatrix} *&*&*&*&*&*& P_1-A_2C_2P_1P_2 & A_2C_1P_1P_2X-A_1C_1P_1P_2 \\ *&*&*&*&*&*& P_1A_1-(1+k)P_1A_1A_2C_2P_2 +A_2kX & A_2P_2X-A_1P_2 \\ *&*&*&*&*&*& P_1-(1+k)A_2C_2P_1P_2 & (1+k)A_2C_1P_1P_2X - P_2 \\ *&*&*&*&*&*& kA_2C_2P_1P_2 & P_2-A_1C_1P_1P_2-kXA_2C_1P_1P_2 \\ *&*&*&*&*&*& kA_2P_1 & (1+k) A_1A_2C_1P_1P_2-A_2P_2 \\ *&*&*&*&*&*& k P_1 & P_2 - (1+k)A_1C_2P_1P_2 \\ A_1C_1-(1+k)A_1A_2C_1C_2P_2+kA_2C_1X & C_1-A_2C_1C_2P_2 & A_1C_1-A_1A_2C_1C_2P_2 & A_2C_1X-A_1C_1 & C_1X-A_1C_1C_2P_2 & A_2C_1X-A_1A_2C_1C_2P_2 & 1-A_2C_2P_2 & A_2C_1P_2X-A_1C_1P_2 \\ kA_2C_2 & kA_2C_1C_2P_1 & kA_1A_2C_1C_2P_1 & A_2C_2-(1+k)A_1A_2C_1C_2P_1 & C_2-A_1C_1C_2P_1 & A_2C_2-A_1A_2C_1C_2P_1 & kA_2C_2P_1 & 1-A_1C_1P_1-kXA_2C_1P_1 \end{pmatrix}

 

It's not easy to FM the laser at low frequency using an EOM (requires phase modulation proportional to 1/f), so let's assume we can't directly excite the laser frequency either before or after the beamsplitter. Instead, we can only sum into the control and error signals at 2,5,7,8. Therefore, we only have access to columns 2, 5, 7, and 8 of \vec{S}. Likewise, we can directly measure only the control and error signals either before or after the injection point. This limits us to accessing rows 2, 5, 7, and 8 of \vec{S}.

(Note that we could consider measuring at node 3 using the driver's current mon, but since this is related to the control signal by a trivial transfer function I'm leaving it out).

This gives us a list of 4x4+1=17 transfer functions in the full system... the extra 1 is due to always being able to measure \frac{1}{1-H} with the transfer function measured just after the injection point. The transfer functions in the table below are normalized by \frac{\mathrm{excitation}}{1-H}, and one could measure any ratio of these transfer functions just as easily (assuming both are coherent at the relevant frequencies).

Injection point measure 2 measure 5 measure 7 measure 8    
2 1-A_2C_2P_2 kA_2C_1P_1 C_1-A_2C_1C_2P_2 kC_1C_2A_2P_1    
5 X-A_1C_2P_2 1-A_1C_1P_1 C_1X-A_1C_1C_2P_2 C_2-A_1C_1C_2P_1    
7 A_1P_1-(1+k)A_1A_2C_2P_1P_2 + kA_2X kA_2P_1 1-A_2C_2P_2 kA_2C_2P_1    
8 A_2P_2X-A_1P_2 A_2P_2-A_1A_2C_1P_1P_2(1+k) A_2C_1P_2X-A_1C_1P_2 1-A_1C_1P_1-kA_2C_1P_1X    

I'm having some trouble discerning which of these transfer functions are coherent at what frequencies, and I think we should just start trying them out in-lab. 


**Note: I reproduced \vec{S} in the 6x6 case, and I think there are a few minor typos above. Since there are several, someone should check again for final resolution, but there isn't a pattern to the differences so I suspect simple typos in the unwieldy matrix

  • Row 3 column 1 should read 1-A_2B_2(1+k)
  • Row 3 column 2 should read B_1-(1+k)A_1B_1B_2
  • Row 2 column 4 should read A_2X-A_1
  • Row 3 column 4 should read -1+(1+k)XA_2B_1
  • Row 6 column 4 should read 1-(1+k)A_1B_1

notes on signal injection

 

 

  3067   Wed Jan 18 15:09:57 2023 aaronDailyProgressVacuumleak checking

The PSOMA chamber reached ~9e-5 torr overnight. I'm trying to identify and correct the leaks in the PSOMA chamber. Using the rate-of-rise test, I measured an O(1e-5 torr*L/s) source of gas in the chamber, based on ~1000 L chamber volume. The test seems to indicate an actual leak, because the rate of rise is steady or increasing after ~1 hour with the pump valved off. Attachment 1 shows the rate-of-rise test and subsequent pumpdown. Attachment 2 shows the leak rate during the test.

After replacing the CF flange I suspected of being leaky based on methanol spritz, the vacuum quickly pumped down to ~20e-5 torr. I'm leaving the system heated to ~80 C overnight to bake out water.

Attachment 1: pressure.pdf
pressure.pdf
Attachment 2: leak.pdf
leak.pdf
  3066   Tue Jan 17 17:26:30 2023 aaronDailyProgressVacuumvent and pumpdown to vacuum chamber

I replaced the gasket of the CF flange with broken bolt.

After a very slow roughing cycle (~1 hour with roughing pump only), the vacuum pressure reached a sufficiently low level that the turbo kicked on. Ultimately the vacuum pressure reached 25e-6 torr. (attachment 1 is the pumpdown curve)

The rate-of-rise test (valve off the pump only and watch the rate of pressure increase) suggests the presence of a real leak, not simply outgassing, because the rate of pressure increase remains approximately constant over time. The test is visible in the attachment where the pressure rises after bottoming out, then falls again to ~25 utorr. I stopped the test somewhat early to avoid the vacuum gauge switching over to Pirani readout, and to avoid needing to spin down the turbo before pumping down again.

I sprayed methanol on all flanges and wasn't able to identify a single limiting leak. However, there are some candidate CF flanges:

  • between the 4-way cross and valve 3
  • On the Cf35 flange 135 degrees clockwise from the hinge.

I'm leaving the system pumping overnight to see if it reaches a lower ultimate pressure.

Attachment 1: Screenshot_from_2023-01-17_17-31-02.png
Screenshot_from_2023-01-17_17-31-02.png
  3065   Tue Jan 17 11:47:42 2023 aaronNotesPSOMAproposed next measurements for PSOMA cavity

Here are some comments on the proposed measurements. Intended to let me understand fully, no fault assumed.

A. Loop coupling by modulation: k

  • Currently, we can't lock both loops with X=0 (or maintain lock while X is turned down to 0). This might be possible once we're in vacuum, which would be my suggestion for attempting this measurement.
  • I think measuring 6\to 1 and 3\to 4 requires the following additional calibrations, based on my understanding that 6 is the North laser frequency, 3 is the South laser frequency, 1 is a weighted combination of frequencies 'seen' by the PDH plant, and 4 is the difference frequency between the N and S lasers.
    • Transfer function from AF drive voltage to N laser frequency (3)
    • Transfer function from AF drive voltage to S laser frequency (6)
    • Transfer function from f_2 to PLL error signal (which is inside block B_2)
    • Transfer function from f_1 to PDH error signal (which is inside block B_1). To avoid circular reasoning, I think this transfer function is identical whether measured with the PLL on or off, so we could eliminate the term in f_1 involving k

C. Full Loop Gain, H

Requires same calibration transfer functions as used in (A) to calibrate (6) and (1). I don't understand why the modulation at the N EOM should be converted to voltage at current driver input. Doesn't 6 correspond to frequency of the N laser, so we want to refer to N laser frequency?

D. Signal Transfer Function

What are E_S and E_N?

Since the current driver is a sub-block of A_i, wouldn't the equations instead apply to transfer functions when monitoring the control signal for either PLL or PDH? I may be misunderstanding whether 6 and 3 appear before or after the EOM.

 


Just as an aside, the only alternative I know to measuring the various loop transfer functions above is to pick off the beam just before the cavity and directly measure amplitude and phase fluctuations on the pickoff using BHD (or whatever). I'm not sure where we'd get the clean (unaffected by unwanted loop couplings) LO for such a pickoff measurement.

Since it seems pretty straightforward to measure all of the blocks independently, maybe the following computationally complicated measurement would actually be experimentally simpler:

  1. Measure the individual blocks as in (B)
  2. Measure transfer function from signal injection point to all available mon ports (minimally both PDH and PLL error and control points). This can be done in one shot of a swept sine.
  3. Fit k using the full matrix, which should be overdetermined.

We could do some meta-analysis of how much information each monitor point contributes (as a function of frequency), and eventually simplify the measurement. But since we wouldn't need to worry about calibrating and comparing multiple excitation points, this might minimize calibration uncertainty.

This could also be expanded to directly measure the optomechanical transfer function, as in the other log entries in this thread.

  Draft   Tue Jan 17 10:58:52 2023 shrutiNotesPSOMAMeasuring amplifier gain and noise figure

 

1. Measuring IO relations

1. Injecting AM and PM

The phase offset in the PLL sets both the signal quadrature (relative to the pump) and also the interference condition between the two lasers in the current setup. We always apply phase modulation at the north EOM...

Assume, \phi_{offset}=0, is when the modulation is AM wrt the pump.

 

2. Measuring AM and PM

- By measuring the DC port of the 1811 REFL PD directly, the AM can be measured (I think?)

- PM can be measured in the PDH error signal

 

2. Estimating PSOMA gain

A. As ratios derived from the IO relations:

\begin{pmatrix} b_1[\Omega] \\ b_2[\Omega] \end{pmatrix} = \exp(-i 2 \eta_{cav}) \begin{pmatrix} 1 & 0 \\ - K[\Omega] & 1 \end{pmatrix} \begin{pmatrix} a_1[\Omega]\\ a_2[\Omega] \end{pmatrix}

1) The ratio of PM measured in the PDH error signal when AM is injected to PM measured when PM is injected

[ Considering the change in power with the change in interference condition ]

3. Spurious couplings

1) Residual amplitude modulation

In principle, when AM is injected, and gain=0, this signal should not be detectable in the PDH error signal unless there is some residual amplitude modulation in the north EOM.

 

 

  3063   Tue Jan 17 09:22:18 2023 shrutiNotesPSOMAproposed next measurements for PSOMA cavity

1. Description

The two loops with feedforward that currently make PSOMA controls using two lasers are shown in Attachment 1.

The south laser is the pump and is locked to the cavity using PDH locking. The north laser is the probe/signal simulator and is locked to the south laser with a phase-locked loop. In practice, the beat between the two lasers is locked to a local oscillator at f_{offset} (150 or 300 MHz was used) and an additional modulation at f_{offset} is applied, one of whose sidebands is resonant with the cavity. The signal is an audiofrequency sideband of the cavity resonance. There is an additional feedforward X from the PDH to the PLL to allow the phaselock to follow the PDH loop when the cavity is very noisy.

The two lasers are mixed on a 50/50 BS and the PLL uses the beat in the AS port and symmetric port with (mostly) constructively interfering light is sent to the cavity.  Since the frequency noise measured around the cavity sees the north laser/path frequency noise only when the additional modulation is applied, the two loops' error signals depend on the north and south paths as:

\begin{pmatrix} f_1 \\ f_2 \end{pmatrix} = \begin{pmatrix} 1 & k\\ -1 & 1 \end{pmatrix} \begin{pmatrix} f_1' \\ f_2' \end{pmatrix}

where f_1' is the frequency noise of the south laser and f_2'  is that of the north laser.

M = \begin{pmatrix} 1 & k\\ -1 & 1 \end{pmatrix} is shown in the bottom-center of Attachments 1 and 2.

f_1 is the error signal seen by the PDH loop (in frequency units). For the PLL, the error signal is proportional to \Delta\phi = 2\pi\frac{f_2' - f_1'}{s} = \frac{2 \pi f_2}{s} in the Laplace domain.

\\ f_1' = A_1 v_1\\ v_1 = B_1 f_1\\ f_2' = A_2 v_2\\ v_2 = B_2 f_2

Explicitly,

f_1' = f_{south} - f_{cavity} \\ f_2' = f_{north} - f_{south} - f_{offset}

Also, v_1, v_2 are the control signals of the PDH and PLL loops respectively.

2. Loop calculations

For the nodes as numbered in Attachment 2, the open-loop transfer matrix is

\mathcal{T} = \begin{pmatrix} 0 & B_1 & 0 & 0 & 0 & 0 \\ 0 & 0 & A_1 & 0 & X & 0 \\ 1 & 0 & 0 & -1 & 0 & 0 \\ 0 & 0 & 0 & 0 & B_2 & 0 \\ 0 & 0 & 0 & 0 & 0 & A_2 \\ k & 0 & 0 & 1 & 0 & 0 \\ \end{pmatrix}

[convention: row index-> column index for both S and T]

In closed loop, the full set of relations are \mathcal{S} = \left( 1- \mathcal{T} \right)^{-1}

\mathcal{S} = \left(\frac{1}{1-k X A_2 B_1 + (1+k) A_1 A_2 B_1 B_2 - A_1 B_1 - A_2 B_2} \right) \\ \times \begin{pmatrix} 1-A_2 B_2 & B_1-A_2 B_1 B_2 & A_1 B_1-A_1 A_2 B_1 B_2 & X A_2 B_1-A_1 B_1 & -A_1 B_1 B_2+X B_1 & X A_2 B_1-A_1 A_2 B_1 B_2 \\ A_1+kX A_2 - (1+k)A_1A_2B_2 & 1-A_2 B_2 & A_1-A_1 A_2 B_2 & A_1-X A_2 & -A_1 B_2+X & -A_1 A_2 B_2+X A_2 \\ 1 - (1+k)A_1A_2 & B_1 -(1+k)A_1A_2B_2 & -k X A_2 B_1-A_2 B_2+1 & -1 +(1+k)XA_1B_2& X B_1-B_2 & X A_2 B_1-A_2 B_2 \\ kA_2 B_2 & kA_2 B_1 B_2 & kA_1 A_2 B_1 B_2 & -k X A_2 B_1-A_1 B_1+1 & B_2-A_1 B_1 B_2 & A_2 B_2-A_1 A_2 B_1 B_2 \\ kA_2 & kA_2 B_1 & kA_1 A_2 B_1 & A_2 - (1+k)A_1A_2B_1 & 1-A_1 B_1 & A_2-A_1 A_2 B_1 \\ k & k B_1 & k A_1 B_1 & 1 - (1+k)A_2B_2 & kX B_1+B_2 - (1+k)A_1B_1B_2 & 1-A_1 B_1 \\ \end{pmatrix}

 

3. Measurements for calibration

A. Loop coupling by modulation: k

With the feedforward turned off, X=0, while both loops are on, this can be measured  as  \frac{6\rightarrow 1}{3 \rightarrow 4}

Experimentally, this would mean measuring two transfer functions, the first by modulating the north EOM (the secondary AF sideband) and measuring the PDH error signal, and the second, by modulating the south EOM and measuring the PLL error signal in frequency units.

B. Individual blocks

X is just the constant factor (=0.5) specified in the Moku cross-coupling matrix, but can be easily measured between the PDH control signal and the North laser HF MON (by dividing out the voltage->current conversion of the north current driver).

Since both loops can be individually locked at the same servo settings, the A's and B's can be measured separately in each loop.

C. Full loop gain, H

The full/overall loop-suppression = 1 / (1 - H), where

H = k X A_2 B_1 - (1+k) A_1 A_2 B_1 B_2 + A_1 B_1 + A_2 B_2.

This factor can be measured directly (with or without feedforward) , once k is known.

H(s) = 1 - \frac{k}{(6 \rightarrow 1)(s)}

where 6->1 is the measured by modulating the north EOM (but converting it to voltage units at the current driver input) and measuring the PDH error signal.

D. Signal transfer functions

When a signal, \eta(s), is injected as the secondary audio sideband in the north EOM and both laser driver current monitors are measured, the following transfer functions are obtained:

For the south monitor,

\frac{I_{S}}{\eta}(s) = \left( \frac{E_N}{E_S}\right) \left( \frac{k A_1 B_1}{1 - H}\right) = \frac{k A_1(s) B_1(s)}{1 - H(s)}

E_S = 1 since the south EOM does not change affect the frequency between the south laser and the cavity. If \eta(s) is calibrated to units of frequency of the north laser, E_N = 1.

For the north monitor,

\frac{I_{N}}{\eta}(s) = \left( \frac{1}{L_N (s)}\right) \left( \frac{1 - A_1(s) B_1(s)}{1 - H(s)}\right)

L_N is the transfer function of the north laser in units of Hz/A.

 

Before pointing out all faults and reasons why these measurements are not immediately possible in the lab, please consider
1. These are what directly follow from the calculations as what occurred to me as the easiest way to measure those quantities theoretically when I couldn’t think of a more practical method. For e.g., X=0 cannot be possible while acquiring lock, but maybe it could be slowly decreased when the setup is stabilized. Please let me know if you do think of a better way to measure k.
2. Of course, from where we actually end up injecting the signal there may be an additional calibration of the drivers, etc that may be necessary which I can easily add. But I have taken into account the units whenever I thought was necessary.
 

 

Attachment 1: Controls_physical.pdf
Controls_physical.pdf
Attachment 2: Controls_block.pdf
Controls_block.pdf
  3062   Fri Jan 13 16:22:43 2023 aaronDailyProgressVacuumvent and pumpdown to vacuum chamber

Spritzing methanol around the CF flanges securing valves 2 and 3 reduced the vacuum pressure by ~3 ntorr (10%). I'll keep an eye on these flanges as we test the rest of the chamber.

I proceeded to vent the chamber, roughly following the steps to reach chamber isolated, vent ready, and vented in the elog above... with the exception that valve 2 remained closed at all times and I abridged steps related to venting with dry nitrogen (I'm just using room air at 1 atm).

Venting with room air

  1. Close valve 1.
  2. turn off the pump
    • Since the chamber is isolated close to atm by valve 2, I tried to open the chamber to inspect its o-ring. I unlatched the clamps on top of the chamber, but some residual vacuum inside the chamber prevented the lid from opening. However, after the vacuum gauge reached ~1e-5 torr, the lid swung open rapidly on its own. Apparently the hydraulic hinge pulls with sufficient force to open the chamber on its own, and a hand is required to prevent the lid from opening too quickly. It is easy to close and open the chamber, but going forward the chamber should not be left closed but unclamped. To avoid dust settling, I closed and clamped the chamber lid. The o-ring looks clean.
    • With valves 1, 2, and 3 closed, the pressure at the gauge increases ~1utorr/min. Since the chamber has a much larger volume than the hose around the gauge, this might actually be acceptable. If the chamber reaches ~10s mtorr in O(day), we'll be leak checking.
  3. Open valve 3
  4. Open valve 1
  5. Open up-to-air valve
  6. Open valve 2

I opened the chamber lid and wiped the o-ring and mating surface with a dry lint-free cloth. The o-ring already has some vacuum grease applied. Then, I pumped down the chamber.

Pumpdown

  1. Close valve 3 and up-to-air valve. Valves 1 and 2 are open.
  2. Turn on pumping station
    • The vacuum pressure stayed close to atmosphere, indicating a screaming leak (almost just an open port, except there isn't one).
    • When I closed valve 2, the pumpdown continued as normal, so it's a problem on one of the chamber flanges.

I turned off the pump and re-tightened the CF flanges. I had been tightening the flanges on the chamber to a torque specification, but this time I tightened until the metal surfaces of the flange completely touched.

On one of the CF35 flanges, the bolt broke off in the nut desite a visible gap present between the CF surfaces. Maybe I need to use a lubricant with these bolts? Since I now need to garb up and completely open at least one flange, I'm going to push this leak hunt to next time.

 

Attachment 1 shows the pressure during the above. The linear plot focuses on the leak rate with valves 1, 2, and 3 closed.

 

Attachment 1: Screenshot_from_2023-01-13_17-17-25.png
Screenshot_from_2023-01-13_17-17-25.png
  3061   Fri Jan 13 14:47:56 2023 aaronLab InfrastructureGeneraloptimal height of PSOMA optics table

[shruti, aaron]

Shruti and I checked out the ergonomics of the PSOMA cryostat and table. We determined that we should acquire new legs for the optics table that are 8.25-8.5" shorter than the current legs. The distance from the floor to the bottom of the rail around the enclosure is 28.25", and the table is 12" thick. The current legs are 22.25", and have a protrusion for the floating attachment to the table which ends ~24" from the ground.

We want the new legs to attach to the bottom of the optics table (28.25"-12") = 16.25" from the ground. Up to ~1" longer legs would be fine.

This would put the top surface of the optics table ~28.25" from the ground. This height is comfortable for both Shruti and I to work on from a standing position with no stool or bending over, and it would be comfortable to reach over the 19.5" tall cryostat especially with assistance from a stool or short platform.

  3060   Thu Jan 12 17:30:04 2023 aaronDailyProgressVacuumPSOMA vacuum closed, pumping down (hose only)

The PSOMA vacuum chamber is closed. The attachments describe the current system.

  1. A sketch of the system is in attachment 1
  2. Attachment 2 shows valve 2, valve 3, the calibrated leak, and the up-to-air valve
  3. Attachment 3 is the pumpdown curve from this afternoon. I'm pumping overnight only up to the vacuum gauge (so valve 1 is open but valves 2 and 3 are closed). I tried to grab the pumpdown curve from dataviewer, but somehow the data aren't being saved to frames? I can hold however much data in the ndscope buffer, but not retrieve dataviewer reports no data found (also "no data output" and "bad file descriptor"). Anyway, when I left after around 1 hour of pumping the pressure is ~0.25 utorr.
    • Update: Chris restarted a process and restored our access (elog pending). After ~24 hours of pumping, the pressure between the pump and closed valves 2/3 (so up to but not including the chamber) is <3e-8 torr! Pumpdown in attachment 3.

It would be helpful if someone could review and discuss these proposed procedures for operating this vacuum system.

Vacuum procedures

Pumpdown

  1. Start with valves 1 and 2 open and 3 closed, so the chamber is open to the pump but the venting path is closed.
  2. Turn on pumping station. Call this state at vacuum

Isolating the vacuum chamber

  1. Start "at vacuum"
  2. Close valve 1
  3. Turn off the pump. Call this state chamber isolated

We could consider adding a second gauge on the opposite side of the chamber, in which case valve 2 could be closed here instead of 1. This would allow continuous monitoring of both the chamber pressure and pressure in the line between the pump and chamber.

Preparing for vent

  1. Start at chamber isolated with the pump speed at 0 rpm
  2. Close valve 2
  3. Open the up-to-air valve
  4. Flush dry nitrogen into the venting path through the calibrated leak by supplying 200 kPa from the gas cylinder
  5. Close the up-to-air valve
  6. Open valve 3. The leak is now filling the line between closed valves 1 and 3 with nitrogen, with the pressure in this line monitored by the gauge.
  7. Once the vacuum gauge reading is greater than the pressure in the rough vacuum line, turn off the supply of nitrogen at the gas cylinder.
  8. Open valve 1.
  9. Turn on the pumping station.
  10. Once the pressure equilibrates (pumping rate is equal to leak rate with no nitrogen supplied from the cylinder), the system is vent ready

Venting

  1. Start vent ready
  2. Open valve 2
  3. Increase pressure in the dry nitrogen line to 200 kPa
  4. Close valve 1. The system is now venting at the leak rate, and the vacuum gauge is monitoring the chamber pressure.
  5. Turn off the pumping station
  6. Once the system approaches atmospheric pressure, expose the up-to-air valve to atmosphere. The system is vented
    • Note that to expose the system only to dry nitrogen, I'll need to add an additional relief valve. The calibrated leak cannot flow enough nitrogen into 1 atm.

I should make a state diagram of these processes.

Attachment 1: Screen_Shot_2023-01-12_at_18.39.22.png
Screen_Shot_2023-01-12_at_18.39.22.png
Attachment 2: 2AC90352-B687-4431-88DA-C56DF2B3DD22.jpeg
2AC90352-B687-4431-88DA-C56DF2B3DD22.jpeg
Attachment 3: Screenshot_from_2023-01-13_14-57-24.png
Screenshot_from_2023-01-13_14-57-24.png
  3059   Thu Jan 12 16:29:10 2023 aaronLab InfrastructureDAQnds errors

I'm having trouble with nds...

  • I can caget the expected value of slow channels defined in epics (eg X1:OMA-ERC_PRESSURE_UTORR reads ~atmosphere), but nds (ndscope, dataviewer) reports these channels as reading 0
  • nds (specifically ndscope) and epics agree on the value of slow channels defined by rtcds such as X1:OMA-ERC_REFL_MON_OUTPUT
  • nds (specifically ndscope) reports the expected value of fast channels defined by rtcds such as X1:OMA-ERC_REFL_MON_OUT, but also reports a "Low level daq error occurred [1]: Unspecified error". Epics (as expected) can't find these fast channels.

I tried restarting the epics services on cominaux, restarting rtcds using the rtreset script on cymac1, and rebooting all of cominaux, cymac1, and spirou workstation. systemctl on cominaux reports the usually services are running on that machine, and I don't know what to do based on the status screens on sitemap (attached).

Attachment 1: Screenshot_from_2023-01-12_16-28-16.png
Screenshot_from_2023-01-12_16-28-16.png
  3058   Thu Jan 12 11:51:19 2023 KojiLab InfrastructureGeneralHow to move the large engine hoist through the narrow door

How to move the large engine hoist through the narrow door

See http://nodus.ligo.caltech.edu:8080/Mariner/122

  3057   Wed Jan 11 16:52:13 2023 aaronDailyProgressVacuumclosed some CF flanges

I continued closing the PSOMA chamber, and will post a sketch of the vacuum system comorrow when all the bolts are tightened.

  3056   Tue Jan 10 16:29:03 2023 aaronDailyProgressPSOMAproposed next measurements for PSOMA cavity

tldr: need some input on measuring the fields before the cavity

We saw some promising signals, but need to say definitively what is happening in our amplifier cavity. We want to observe the dynamics of our plant (cavity), and preferably while minimally resorting to detailed modeling of the coupled PDH and PLL loops. So what measurements should we make? I'm writing this up following discussion with Chris to help clarify for myself.

Ignoring intracavity noises, the fields reflected from the cavity are

\begin{pmatrix}b_1\\b_2\end{pmatrix} = e^{2i\eta}\begin{pmatrix} 1 & 0 \\ \kappa & 1\end{pmatrix} \begin{pmatrix} a_1\\a_2\end{pmatrix}

Where the quadratures are amplitude (1) and phase (2), \eta is a frequency dependent phase due to the cavity pole, and \kappa is the ponderomotive coupling. To full characterize the optomechanical transfer function, we want to measure the transfer function from excitation to each of a_1, a_2, b_1, b_2. I think measuring all of these transfer functions using the same excitation point lets us characterize the cavity transfer function no matter how complicated our controller is (even if the controller itself involves many coupled loops), because the above equation must hold in steady state. We could even measure them simultaneously with a single excitation.

The DC MON on REFL is sensitive to b_1, while the cavity pole lets us measure b_2  after demodulating at the PDH sideband frequency (the usual PDH technique). Likewise, the BEAT PD is sensitive around DC to a_1, and is sensitive to a_\mathrm{probe,2} by demodulating at the PDH sideband frequency. The last claim is justified because the acoustic signal sidebands around the probe are not (directly) present on the pump laser or its RF sidebands.

Formally, the Beat PD sees a slightly different field than the cavity, but the transfer function measurements might be mappable between a and \tilde{a}. As long as a_\mathrm{probe,2}\approx a_2 and either the pump or probe dominates in the amplitude quadrature, measuring phase fluctuations at BEAT PD is equivalent to measuring the fields just before the cavity.

\begin{pmatrix} a_1\\a_2\end{pmatrix} = \begin{pmatrix}a_\mathrm{pump,1} \\ a_\mathrm{pump,2}\end{pmatrix} + \begin{pmatrix}a_\mathrm{probe,1} \\ a_\mathrm{probe,2}\end{pmatrix}

\begin{pmatrix} \tilde{a}_1\\\tilde{a}_2\end{pmatrix} = \begin{pmatrix} a_\mathrm{pump,1}\\a_\mathrm{pump,2}\end{pmatrix} - \begin{pmatrix}a_\mathrm{probe,1}\\a_\mathrm{probe,2} \end{pmatrix}

However, in practice I'm not sure we can monitor only BEAT PD. The PDH loop will inject any signal in b_2 back to a_2 by driving the pump frequency. Therefore, we probably cannot approximate a_\mathrm{probe,2}\approx a_2, especially in the limit of amplitude-quadrature signal injection (meaning direct coupling of signal to a_\mathrm{probe,2} is small) with high ponderomotive gain (meaning b_2 can be much larger than a_2); furthermore, the PLL drives \tilde{a}_2\to 0, so we would expect a_\mathrm{pump,2}\approx a_\mathrm{probe,2}. In other words, we might be able to use the BEAT PD to estimate fields entering the cavity, but only if we measure some transfer functions of the coupled PDH-PLL loop.

We may need to pick off some light just before the cavity instead (easy enough), or fully characterize the coupled control loops (probably worth it to avoid loss in an eventual sub-SQL demonstration). We also would need to measure the phase of the pickoff using BHD, since we want to know the contributions to a_2 of both the pump and the probe and suspect that the pump contribution may dominate. I'm not sure where to pick off a BHD LO -- both pump and probe current drivers are controllers in either the PDH or PLL, so neither laser has a clean carrier free from any injected signal.

Alternatively, can we excite below the PLL UGF then assume a_\mathrm{pump,2}\approx a_\mathrm{probe,2} and simply measure a_\mathrm{probe,2} at the pickoff before the cavity and demodulating at PDH sideband frequency?

next steps

To summarize a few possible approaches

  1. Keep the same measurement setup (attachment 1), and assume a_\mathrm{pump,2}\approx a_\mathrm{probe,2} and a_\mathrm{probe,1}>>a_\mathrm{pump,1} to map measurements at BEAT PD to fields entering the cavity
  2. Pick off just before the cavity and measure using a single RF PD. This lifts the assumption that a_\mathrm{probe,1}>>a_\mathrm{pump,1} by directly measuring a_1, but still requires a_\mathrm{pump,2}\approx a_\mathrm{probe,2} due to the incomplete measure of only the probe component of a_2
  3. Pick off just before the cavity and measure phase with a BHD (or another method?). This would provide a direct measurement of both a_1 and a_2. I'm not sure where we'd get a clean LO for this measurement.

In either (1) or (2) above, we might lift the assumptions about pump and probe fields by making appropriate (and unkown to me) measurements of the PDH-PLL transfer function.

Attachment 1: 39CA7615-6C69-4816-81C7-D2CB92A27AEA.jpeg
39CA7615-6C69-4816-81C7-D2CB92A27AEA.jpeg
  3055   Thu Dec 22 14:30:08 2022 aaronDailyProgressVacuumclosed some CF flanges

I closed the 6.25" CF flanges on the PSOMA chamber with blanks, though I didn't completely seal the copper gaskets (I left two bolts finger tight on each flange).

On three of the flanges, I used Kapton tape to secure the outer edge of the copper gasket to the flange so I could easily install the blank without dropping the gasket. I wasn't satisfied with the result since I suspect at least some of flanges have a small piece of Kapton tape between the gasket and the groove. The tape is well outside the knife edge, but I'm concerned it will prevent the copper from fully forming into the groove. I'm going to remove and reseat these flanges.

I also started re-shelving some optics from the cryo-cantilevers experiment to make room for the chamber on the PSOMA table.

Attachments

  1. vacuum workstation
  2. The flanges are covered with an outer layer of foil, followed by cardboard, followed by an inner layer of foil. This is the inner layer of foil
  3. prepping the blank and gasket before installation. I wiped the knife edges and gasket with a lint free cloth
  4. After installing the blank. Bolts are finger tight
  5. Cryo cantilevers optics "before"
Attachment 1: E04D8FF9-A1E2-4950-8441-6EB4826ECC9B.jpeg
E04D8FF9-A1E2-4950-8441-6EB4826ECC9B.jpeg
Attachment 2: C4804B3C-C72B-4D79-84DD-2F363EEBD154.jpeg
C4804B3C-C72B-4D79-84DD-2F363EEBD154.jpeg
Attachment 3: 96621ED3-587E-4BFC-B051-0F1EF60592CC.jpeg
96621ED3-587E-4BFC-B051-0F1EF60592CC.jpeg
Attachment 4: ED779D82-85F1-4D77-8487-EEE14651F0CC.jpeg
ED779D82-85F1-4D77-8487-EEE14651F0CC.jpeg
Attachment 5: 3B6E57F5-DCFA-4B7F-9080-3A81A4919C56.jpeg
3B6E57F5-DCFA-4B7F-9080-3A81A4919C56.jpeg
  3054   Wed Dec 21 17:26:24 2022 aaronDailyProgressPSOMAhow much AM do we expect?

I'm trying to understand the changes in DC power on BEAT MON and TRANS MON channels when the resonant probe field is tuned to constructively or destructively interfere with the pump. In particular, the change in DC power are larger than I would expect based on the sideband amplitude in the beat note spectrum.

Here are some observations. In all cases, I assume that minimizing power on BEAT mon (maximizing power on TRANS mon) implies the pump and probe are in-phase.

Channel State Value
Beat mon (DC) resonant probe in-phase with pump, PDH servo off, -15 dBm RF @ 150 MHz on probe EOM -3.08 V
Beat mon (DC) resonant probe 90 deg detuned from pump, PDH servo off, -15 dBm RF @ 150 MHz on probe EOM -3.16 V
Beat mon (DC) resonant probe 180 deg detuned from pump, PDH servo off, -15 dBm RF @ 150 MHz on probe EOM -3.25 V
Beat @ 150 MHz PLL locked at 150 MHz, PDH servo off, -15 dBm RF @ 100 MHz on probe EOM -27 dBm
Beat @ 50 MHz PLL locked at 150 MHz, PDH servo off, -15 dBm RF @ 100 MHz on probe EOM -54 dBm
Beat mon (DC) Pump laser on, probe laser off -1.88 V
Beat mon (DC) Probe laser on, Pump laser off -1.22 V
Beat mon (DC) Both lasers off (dark current) 45 mV

 

The beat note at the PLL LO frequency is proportional to \sqrt{P_\mathrm{pump}P_\mathrm{probe, 0}}, while the probe's sideband is proportional to \sqrt{P_\mathrm{pump}P_\mathrm{probe, 1}}. Therefore, the ratio of field amplitudes for the probe's carrier and first sideband is

\frac{\sqrt{P_\mathrm{pump}P_\mathrm{probe,0}}}{\sqrt{P_\mathrm{pump}P_\mathrm{probe,1}}}=\frac{E_\mathrm{probe,0}}{E_\mathrm{probe,1}}=27 \mathrm{dB}\approx 500.

When one of the sidebands is made homodyne with the pump by driving the probe EOM at the PLL LO frequency, the probe's sideband field interferes with the pump with some detuning set by the relative phase (\phi) of the PLL LO and EOM drive tone. Beat mon voltage is proportional to the incident power at DC, which is

P_\mathrm{beat}(0\mathrm{Hz})\approx (E_\mathrm{pump}+E_\mathrm{probe,1}\cos(\phi))^2+P_\mathrm{probe,0}+P_\mathrm{probe,1}

Since the pump and probe lasers have approximately the same power, I'd expect the sideband to cause up to a ~1/250 fractional change in power at the pump carrier frequency. Even doing the calculation more carefully finds that the ratio of Beat mon levels when the pump and probe are constructively / destructively interfering should be

\frac{P_\mathrm{beat}^\mathrm{\phi=0}(0 \mathrm{Hz})}{P_\mathrm{beat}^\mathrm{\phi=\pi}(0 \mathrm{Hz})}\approx 1.0039

However, I instead measured

\frac{P_\mathrm{beat}^\mathrm{\phi=0}(0 \mathrm{Hz})}{P_\mathrm{beat}^\mathrm{\phi=\pi}(0 \mathrm{Hz})}\approx1.054

I still can't explain this discrepancy. When I made an analogous measurement with the cavity locked and comparing TRANS mon levels instead of Beat mon, I saw a similar discrepancy. I was driving the probe EOM with -10 dBm, and I saw ~10% difference between \phi=0 and \phi=\pi. The measurement above confirmed for me that the scaling is approximately linear and not a result of unexpected coupling between the PLL and PDH loops. It's great that I can see a lot of interference between my pump and probe, but disconcerting that I can't explain the effect. Have I missed something in this calculation or measurement?

idea for the measurement...

While making this measurement I was able to drive the probe EOM with up to 3 dBm RF power and maintain PLL lock and even (noisy) PDH lock. I also drove the probe EOM at the PLL LO frequency detuned by <kHz and saw low frequency beating on Beat mon. If the PDH can remain locked with a frequency-detuned probe, this might let us make a lock-in measurement of the gain rather than controlling the pump-probe phase.

 

Update

Chris found my error -- electronic voltage is proportional to optical power, so beat note power on the spectrum analyzer is proportional to P_1P_2 not \sqrt{P_1P_2}. Everything above is consistent with this modification.

  3053   Tue Dec 20 17:48:48 2022 aaronDailyProgressPSOMAsaw some signal in amplifier output, but system not fully understood

I saw a signature of amplification, but I can't explain everything I observed so don't want to attribute it definitively to ponderomotive effects.

Attachments 1 and 2 show the PDH control signal in the top display (measured at pump's HF mon) and the MON of the beat note PD (amplitude fluctuations into the cavity) on the bottom display. In attachment 1, I've tuned the pump-probe phase to minimize the signal in the amplitude quadrature entering the cavity. Attachment 2 adds 90 degrees to the pump-probe phase, so the signal should be maximally in the amplitude quadrature (up to drifts during the measurent time).

When the signal is in the phase quadrature, it couples directly to the "output" of the amplifier since the PDH servo compensates phase fluctuations on the pump laser. When the signal is in the amplitude quadrature, it should only couple to the "output" of the amplifier if the cavity reflection transfer function rotates amplitude into phase (due to ponderomotive effects, or possibly due to the cavity sitting off-resonance).

The attachments show that the signal peak at the amplifier output is 2 dB higher when injected in the amplitude quadrature compared to the phase quadrature.

 

It's a bit premature to claim this is a ponderomotive effect because there are a number of unexplained features...

  • I still can't explain the changes I see in HF mon. When the signal is close to fully in phase, the TRANS MON signal is either maximized or minimized, and the case with signal in amplitude quadrature has TRANS MON halfway between these extrema. However the span is too large (roughly 10% change between max and min).
    • Here's a possible explanation: I was using the upper sideband at ~300 MHz to estimate the size of the lower sideband at DC. However, my EOM only has a 150 MHz bandwidth! So the 300 MHz peak could experience >>3 dB rolloff. Will check again with a lower PLL frequency.
  • I think the signal strengths at amplifier input are identical whether rotated into phase or amplitude, since the power in the sidebands are identical. However, I need to think through at least once whether this works out in dimensionless units.
  • Cavity detuning could cause amplitude-to-phase coupling
  • The pump-probe phase drifts. I observed that if I integrated longer than ~5 min, the peak height started to drift with each new average.
  • What is the extra feature in the PDH control spectrum in attachment 2? If it's residual phase noise from the PLL, why doesn't it appear in either input spectrum?
Attachment 1: phase_quadrature.jpg
phase_quadrature.jpg
Attachment 2: amp_quadrature.jpg
amp_quadrature.jpg
  3052   Mon Dec 19 11:03:32 2022 aaronDailyProgressLasersignal injection, loop tuning

I was remiss in my eloging, but here's the update from last week...

tldr: I saw my AM signal before the amplifier above the noise floor, but wasn't able to see the optomechanically-induced PM signal after the amplifier. Then, I messed around with the PLL loop shape longer than I should.

Setup

I removed the amplitude modulator from the pump path to get us back to ~1 W circulating pump power. I also changed the PLL LO frequency (offset between the pump and signal lasers) to 150 MHz, the bandwidth of our fiber EOM. The loop seemed to work fine at 300 MHz offset, but if the EOM response is falling off there could be some unnecessary phase delay, and 150 MHz seems far enough from the cavity resonance and PDH sidebands.

I tried two signal injection schemes. In both cases, the Moku and Marconi are locked to the same 10 MHz timing signal.

Audio signal phase modulates the probe after the PLL

  1. Phase lock the pump and probe lasers 150 MHz apart. The LO for the PLL is generated internally by the Moku, and has a user-adjustable DC phase.
  2. Use the Marconi and a fiber EOM to phase modulate the probe at the same frequency as the PLL's LO, generating a sideband at the pump frequency and phase-locked to the pump. The relative phase of the Marconi and Moku LO sets the phase of the homodyne probe sideband.
  3. Phase modulate the Marconi LO to generate audio sidebands on the homodyne probe. The audio sidebands are AM when the probe is out of phase with the pump.

Audio signal phase modulates the probe in the PLL

  1. Phase lock the pump and probe lasers 150 MHz apart, with the LO for the PLL provided by the Marconi.
  2. Phase modulate the Marconi LO to generate audio sidebands around the probe.
  3. Use the Moku to generate a pure tone at the PLL LO's carrier frequency, and modulate the probe laser at the EOM with this tone. For example, if the PLL LO is 150 MHz with some audio sidebands, the Moku drives the EOM with a pure 150 MHz tone.

In both cases, the signal can appear as either PM or AM on the pump according to the relative phase of the LOs used for the PLL and for driving the EOM.

I didn't actually notice that one of these methods generated more AM around the pump, but I didn't make a careful enough comparison. I expected that modulating the probe after the PLL would lead to the PLL suppressing the signal. Will have to re-measure and re-think this.

Is our signal injection working at all?

To find out, I'm looking at REFL and BEAT MONs on the SR785 spectrum analyzer. I was able to produce AM sidebands above the noise floor between 40 Hz and 10 kHz, and could change the height of these AM sidebands by roughly tuning the relative phase of pump and homodyne probe.

With both the PLL and PDH servos locked, I can change the transmitted light level by changing the relative phase of the pump and homodyne probe. This makes sense: when the pump is in-phase with the pump, it just looks like more carrier power to the cavity. However, the magnitude of the changes in transmission power is larger than I expect (O(10%) changes in transmitted power, but I think the homodyne probe field amplitude is only 1% that of the pump).

I couldn't find our signal after the amplifier

Assuming the signal is AM, it would only appear in the PDH loop after being rotated to PM by pushing on the cavity mirror. Below the UGF of the PDH servo, I would expect to see a signal peak in the PDH control spectrum as the loop compensates. I couldn't find this peak in the pump laser's HF mon when the cavity was locked. I tried integrating the spectrum for O(5-10 minutes), as well as using a lock-in amplifier with similar integration times.

Thursday

Messed around with the PLL, and tried to understand the open questions.

Friday

I tuned up the PLL servo for several hours by measuring the PLL open loop transfer function then adjusting a two-stage PID filter (with up to 4 integrators and 2 derivatives total) a) increase gain below the UGF in the region with significant residual error, b) increase the UGF frequency, c) increase the phase margin at the UGF, and d) increase the gain margin where the phase margin vanishes. I made and unmade many marginal improvements, and found a few heuristics about our PLL.

We can push the UGF above 150 kHz and maintain lock, but the intrinsic phase delays in the loop limit the stability of this loop. For example, I could get a marginally stable loop with UGF above 200 kHz by adding a second derivative, but since this reduces the gain rolloff above the UGF I still get gain peaking wherever the phase margin vanishes.

There is some

  3051   Tue Dec 13 10:58:12 2022 aaronDailyProgressOpticstest of intensity modulator S/N 6800-03

I checked the general function of the intensity modulator. The internal PD port responds to input laser power (reads ~10 mV with S laser off, ~865 mV with S laser on). I'm using the beat note PD to monitor transmitted laser power, and checked that the changing the DC bias level with a function generator can change the DC output of the beat note PD from 0 V to ~325 mV.

I created a couple of new cds filters to monitor and control the intensity modulator

  • X1:OMA-PUMP_AM_MON, whose input monitors the internal PD of the MXAN
  • X1:OMA-PUMP_AM_BIAS, whose output goes to the DC bias port of the MXAN

I make'd and installed the new x1oma model, and ran rtreset to restart the front ends. Rather than wait for Chris' cronjob to run this week, I ran "./mdl2adl.sh" from /opt/rtcds/tst/x1/scripts/mdl2adl directory on cominaux to generate updated medm screens. Note that running the script only on the x1oma model (as in "./mdl2adl.sh x1oma") threw an error.

I used a function generator to sweep the DC bias level with a +- 10 V triangle wave, and monitored the DC power level on the beat note PD and internal PD of the modulator (ndscope traces in attachment 1, all y-axis units in counts). The internal PD is supposed to monitor a 90:10 or 99:1 tap fiber on the transmission side, so I'm surprised to see a nonlinear output.

Attachment 1: Screenshot_from_2022-12-13_15-20-24.png
Screenshot_from_2022-12-13_15-20-24.png
  3050   Tue Dec 13 10:06:19 2022 aaronDailyProgressVacuumvacuum testing

I've left the system under vacuum for about a week. On early Sunday morning ~4am (nobody was in the lab), the pressure apparently dropped to 1e-7 torr for several minutes then increased to ~600 utorr. Each step change was in about a minute. The pump is still operating as normal, and I wasn't able to find a leak by spritzing methanol around the flanges. The vacuum pressure over the last 7 days is attached.

Attachment 1: Screenshot_from_2022-12-13_10-05-46.png
Screenshot_from_2022-12-13_10-05-46.png
  3049   Fri Dec 9 17:24:48 2022 aaronDailyProgressOpticsamplitude modulator in pump path

I added the MXAN-LN-10 amplitude modulator to the pump (S laser) fiber path, just after its EOM. I do see the expected ~3dB of insertion loss due to the AOM (though I should measure this more carefully). I was able to lock the cavity with no trouble with the new modulator in the path (no modulation applied). I'll measure the modulator's transfer functions and Mach-Zehnder DC tuning next time.

I connected the RF ports of the modulator to the front panel with LM-200 SMA cables. While I had the box open, I swapped out some lossier SMA cables for LM-200 cables: between the beat note PD RF port and the front panel, and between the S EOM RF port and the front panel. The cables were not of the same length, so I expect the PDH phase will have changed.

Due to space constraints, I mounted the amplitude modulator on top of the S EOM, separated by the breakout breadboard and a heat sink as shown in attached photos. I used rubber to soften the clamp, and placed another small heat sink between the rubber and the amplitude modulator. I didn't use any conductive interface between the heat sinks and the modulators.

 

Attachments

1. Fiber setup before modification

2. fiber setup after modification

3. Clamp

4. heat sink 

5. breakout PCB is supported not dangling 

 

Attachment 1: CA948CB1-07B5-4141-AD69-BC94908B29C2.jpeg
CA948CB1-07B5-4141-AD69-BC94908B29C2.jpeg
Attachment 2: B0307390-69BB-4442-868E-41BFEAFA2B90.jpeg
B0307390-69BB-4442-868E-41BFEAFA2B90.jpeg
Attachment 3: B87F0703-5F13-404A-BCCD-E0E62CFAC14B.jpeg
B87F0703-5F13-404A-BCCD-E0E62CFAC14B.jpeg
Attachment 4: C2146338-F2E0-40EA-BBDF-F1F3C0B19866.jpeg
C2146338-F2E0-40EA-BBDF-F1F3C0B19866.jpeg
Attachment 5: 44B069F8-E331-4A5D-B27F-260CC0C51096.jpeg
44B069F8-E331-4A5D-B27F-260CC0C51096.jpeg
  3048   Thu Dec 8 17:26:44 2022 aaronDailyProgressElectronicssma breakout for fiber amplitude modulator

I'd like to add the MXAN-LN-10 fiber-based amplitude modulator to our pump path to allow us to amplitude modulate our pump independent of our PLL-and-probe-laser scheme for signal injection. Since we aren't yet using a Mach-Zehnder to reject pump RIN, we won't really be taking a hit from the perspective of SNR, and it would be nice to test the optomechanical response of the cavity independent of the PLL and transfer functions involving the probe laser.

Zach had been using wire clips to tune the DC electrodes of our MXAN. Since this is a semi-permanent installation, I made a PCB board to connect the electrode pins to SMA connectors. The result is attached.

I'm not sure whether our model actually has an internal photodiode, but I included a connector for pins 3 and 4 anyway. I checked that the electrical connections work and there isn't unexpected shorting. I also found some heat sinks that I can place under the MXAN, since the body of the modulator won't press directly on the optical breadboard.

Attachment 1: AD667D97-9CAF-4368-921F-2EE0C90E287B.jpeg
AD667D97-9CAF-4368-921F-2EE0C90E287B.jpeg
Attachment 2: 5FF5A44D-7A26-4E1C-9A4D-4F5C10529196.jpeg
5FF5A44D-7A26-4E1C-9A4D-4F5C10529196.jpeg
  3047   Thu Dec 8 09:45:56 2022 aaronDailyProgressVacuumvacuum testing

minor observation about the vacuum pressure

For about 9 hours after turning off the heater and removing insulating foil, the vacuum pressure was below 1e-7 torr. Then, the pressure increased to several 1e-7 torr, where it has remained. Not sure why this order of magnitude increase in vacuum pressure.

The vacuum pressure minute trend for the last day is in attachment 1, and the raw data for the hour or so where the pressure was rising is in attachment 2.

possible explanation?

This vacuum handbook from Agilent that I return to often notes that at high vacuum, most of the remaining gas in the system is adsorbed on the walls rather than residing in the volume. Slide 7 notes that at 1e-9 torr, there are 500,000 molecules on the surface for every one in the volume, and it takes 2200 seconds for a monolayer to form (with both of these numbers inversely proportional to vacuum pressure in the molecular flow regime). Maybe for the first ~9 hours after the heater was turned off, the vacuum pressure was determined by (pumping speed + adsorption to the walls) - (desorption from the walls + permeation through KF flanges + backflow + outgassing + etc), with the adsorption rate limited by the surface area of the vacuum system. If there's some relation between number of monolayers and adsorption/desorption rate, at some point the number of monolayers could limit the net adsorption rate and lead to a higher equilibrium pressure.

At a pressure of ~4e-8 torr, the walls would form 1 monolayer of gas in about a minute. So in 9 hours, ~500 monolayers would form. I'd sort of expect the adsorption/desorption rate to equilibrate with just 1 monolayer of gas, so maybe I need another explanation.

Attachment 1: Screenshot_from_2022-12-08_10-16-47.png
Screenshot_from_2022-12-08_10-16-47.png
Attachment 2: Screenshot_from_2022-12-08_10-17-25.png
Screenshot_from_2022-12-08_10-17-25.png
  3046   Wed Dec 7 18:25:05 2022 aaronDailyProgressNoise Budgetupdated noise budget

I tuned up the PDH servo by widening the region with positive phase above 100 kHz and setting the PI corner near the cavity pole. I measured the plant transfer function from 10 kHz to 100 kHz and found that the phase is -45 degrees relative to ~DC near 37 kHz; the recent PDH sweep implied cavity pole ~39 kHz, so this is reasonable.

I measured the transfer functions and noise spectra (at error and control points) for the S laser locked to the (higher finesse) cavity. I kept the N laser off during these measurements, so it's the same 1-loop system we worked with in September. The cavity pole is now lower, ~37 kHz, and the UGF is ~115 kHz.

I plotted the results, but wasn't completely satisfied that I caught all factors of few in the units from the HP spectrum analyzer, so I'll update with attachments when I do so.

 

Update: The results are attached.

  1. Transfer function measurements. Below 100 kHz taken with SR785, above 100 kHz taken with HP spectrum analyzer. In the future maybe I'll divide out the 'actuator' transfer function (HF mon to laser frequency) and instead plot just the plant (cavity transfer function). THe UGF is 116 kHz, and phase margin is 62 degrees.
  2. Noise spectra. These are consistent with the UGF ~116 kHz. There is a 14 dB discrepency between the spectra measured at error mon and HF mon, probably due to my accidentally taking data with the Moku +14 dB output gain stage on for some of the measurements... but I couldn't figure out where the mistake was made, so I left it uncorrected. I suspect the control mon is the correctly calibrated curve.
  3. Stage 1 control filter
  4. stage 2 control filter 

 

I pushed this updated noise spectrum to the gitlab, and debugged the pipeline so the results are again available on this webpage.

Attachment 1: LoopTFs.pdf
LoopTFs.pdf
Attachment 2: AllSpectra_HzrtHz_20221207.pdf
AllSpectra_HzrtHz_20221207.pdf
Attachment 3: 0339DB09-10BA-4506-9368-1B36D5715105.png
0339DB09-10BA-4506-9368-1B36D5715105.png
Attachment 4: 07D4744E-1903-4B22-883E-B5B082014652.png
07D4744E-1903-4B22-883E-B5B082014652.png
  3045   Wed Dec 7 11:59:54 2022 aaronDailyProgressVacuumvacuum testing

I turned off the heater and let the system return to room temperature. The pressure settled between 2-6e-8 torr (attachment 1  shows pressure decreasing for a couple hours after turning off the heater). This far exceeds our requirement for this vacuum system.

Attachment 1: Screenshot_from_2022-12-07_12-02-46.png
Screenshot_from_2022-12-07_12-02-46.png
  3044   Tue Dec 6 11:56:36 2022 aaronDailyProgressVacuumvacuum testing

The pressure did reach 6e-6 torr, but then jumped to ~mtorr for after less than an hour.

I could not identify leaks by spritzing methanol around all flanges, but after going up to atmosphere noticed one KF flange was somewhat loose.

I noticed that the gauge reports >1000 torr at atmospheric pressure. After spinning down the turbo and venting to atmosphere, I calibrated the atmosphere reading by shorting pin 7 of the gauge to its negative supply. Even after the atmosphere calibration, the gauge read >900 torr, so I introduced an offset voltage on my calc channel to send the gauge's pressure reading at atmosphere to 10 V (the nominal gauge output at atmosphere). The -0.082 V 'ad hoc' offset decreases the absolute pressure reading by 25% at any given pressure.There is a Vacuum gauge manual available online

I then pumped down to several utorr and again shorted pin 7 to the negative supply to provide a 'vacuum' reading for the pirani. I only somewhat trust the absolute pressure readings (maybe to 25%), but presumably the pumpdown curves can now be consistently compared going forward.

After tightening the KF flanges and calibrating the gauge, I pumped down again and maintained at < 1 utorr for over 1 hour. I tried tightening the CF flange, but in doing so caused a leak that limited vacuum pressure to ~600 utorr (one of the KF flanges was torqued while tightening the CF bolts, and I'm not sure whether the leak is at the CF or one of the KF flanges since methanol reveals no leaks at any flange). I vented back up to atmosphere, cleaned the effected KF flanges, and replaced the copper gasket and resealed the CF flange. Then, I pumped down again.

The latest pumpdown curve is in attachment 1. I wrapped the cross in heat strap and foil, and when the pressure reached ~utorr turned the heater on to equilibrate at 85 C.

Attachment 1: Screenshot_from_2022-12-06_17-35-24.png
Screenshot_from_2022-12-06_17-35-24.png
  3043   Mon Dec 5 18:18:22 2022 aaronDailyProgressVacuumvacuum testing

With the Edwards WRG-S gauge borrowed from QIL, I'm continuing with planned tests of the PSOMA chamber.

I don't have a controller for this gauge, so I cut off an RJ45 cable and added BNC connectors to the appropriate wires for +18V power supply and pressure signal output.

I'm testing the gauge by pumping on a minimal system: right angle valve -> KF40 and KF16 cross -> [KF16 blank, up to air valve, KF40-to-DN40CF adapter] -> WRG-S with DN40CF connection. Pumpdown curve is attached, I'll leave it pumping overnight to see how low the pressure gets.

Attachment 1: Screenshot_from_2022-12-05_18-18-01.png
Screenshot_from_2022-12-05_18-18-01.png
  3042   Mon Dec 5 18:00:13 2022 aaronLab InfrastructureElectronicsoverdrawn Sorensen on +18V supply

On Friday, I noticed that the PSOMA S laser diode had suddenly reduced power (we've been seeing 1.7 mW at the beam launch, and suddenly I was getting only 0.5 mW). I was close to wrapping up, and was able to recover full power by hopping to a different hysterisis curve with TEC set to ~10 kOhm. Lee also pointed out that he wasn't seeing as much laser power as expected out of the cryo cavs diodes.

However, today I'm seeing odd behavior in the Acromag channels (such at X1:AUX-ACROXT_AI_1, which is the analog input channel for PSOMA vacuum pressure). The channels alternate between reading 0 and reading the correct value, for 10s of seconds at a time in either state.

On investigating the acromag chassis' power supply, I found that the +18 V rail is only supplying ~11 V. This could explain the sudden loss in laser power -- the current driver is probably supplying less current than it would with full voltage supplied. Probably turning on the cryo cavs lasers (plus whatever else we've plugged in since they were last used) drew too much current.

I increased the current limit on that Sorensen, and the DC current increased from 3.5 mA to 3.8 mA. The acromag-related channels are now fully functioning.

  3041   Thu Dec 1 11:07:34 2022 aaronDailyProgressPSOMAcleaning cavity optics

I'm measuring the beam properties at a few more locations

Distance from front of fiber launch x-axis beam diameter (13.5% clip) y-axis beam diameter (13.5% clip) ellipticity
9 mm 856.7 um 883.4 um 1.03
188 mm 1078.1 um 1083.9 um 1.01
232 mm 1142.9 um 1145.8 um 1.00
252 mm 1183.5 um 1184.3 um 1.00
268 mm 1218.2 um 1216.5 um 1.00
296 mm 1267.7 um 1259.9 um 0.99
313 mm 1296.3 um 1286.8 um 0.99

These measurements and the alamode fit for the beam profile are in attachment 1. The x data best fit waist radius is 405.6 um at -103 mm from the front plane of the fiber launch. The y data best fit are 421.3 um at -103 mm.

Attachment 2 shows the >0.99% mode matching solution least sensitive to errors in modematching lens position. It uses a 500 mm FL lens at 0.3023 m from the launch plane, and a 850 mm FL lens 0.4608 m from the launch plane. The target beam waist is larger than yesterday's plot because I looked up the quote for our coastline concave lens, and found it has a 1m ROC (not a 1m FL).

After these changes, I am seeing a bit more transmitted power. Power measurements suggest 72% mode matching and 450 ppm loss. I added a second stage ID filter to give some extra gain between 10 Hz and 100 Hz to suppress the cantilever motion. The PDH servo filters are in attachment 3 and 4. I measured the open loop transfer function to determine that the UGF is between 100 kHz and 125 kHz, which helped in choosing the derivative corner and saturation. When adjusting mode matching lenses, I increased the servo gain by 6dB and tuned the output HWP to adjust the amount of light sent to REFL PD; this allowed me to lock on the 00 mode most of the time just by roughly placing the lenses such that the beam spot maintains the same position on REFL camera.

I'm not convinced my mode matching solution is any better than the previous one, or perhaps the lenses are misplaced. The reflected beam still looks like a 1st order Laguerre-Gauss mode. I'm still placing the first mode matching lens at a 30-45 degree angle to correct astygmatism. THe reflected beam is the top monitor in attachment 5. Since there seems to be >1W of circulating power, I'll move on to signal injection tomorrow, but it would be good to know how to diagnose and correct this mode mismatch.

Attachment 1: 221201_alaPSOMA_fit.pdf
221201_alaPSOMA_fit.pdf
Attachment 2: 221201_alaPSOMA_MM.pdf
221201_alaPSOMA_MM.pdf
Attachment 3: B3046E97-3811-419D-9886-24EAFF8FE138.png
B3046E97-3811-419D-9886-24EAFF8FE138.png
Attachment 4: F223FF47-81F4-4D3E-8E7F-4AC41941BBC5.png
F223FF47-81F4-4D3E-8E7F-4AC41941BBC5.png
Attachment 5: B7BE18D5-4A78-4B77-9D99-312547CD02E7.jpeg
B7BE18D5-4A78-4B77-9D99-312547CD02E7.jpeg
  3040   Wed Nov 30 12:50:26 2022 aaronDailyProgressPSOMAcleaning cavity optics

Thanks Koji! Shruti also suggested that WinCamD might be Si-based... I was looking at their TEL Phosphor response curve! I'm using the InGaAs camera on the Beam'R2-DD now.

I recorded the following beam parameters, with no lenses in the beam path. I left one steering mirror and the input HWP in the beam path. I need to check out the manual to understand some of these measurements... I think it's something like the camera scans a slit along two orthogonal directions (parameterized by 'theta'), and uses this to measure the beam power along the slices defined by the slit.

parameter Value 8 mm from front surface of fiber launch Value 308 mm from front surface of fiber launch
Ellipticity 1.03 1.00
Beam width at 13.5% clip, x-axis 854.4 um 1284.0 um
Beam width at 50% clip, x-axis 459.3 um 800.7 um
Beam width at 13.5% clip, y-axis 883.7 um 1281.8 um
Beam width at 50% clip, y-axis 501.8 um 771.3 um

I measured the distance from the front surface of the beam launch to the HR surface of MC1 to be 523 mm. MC1 to MC2 is 230 mm. MC2 to MC3 is 230 mm. And MC3 to MC1 is 53 mm.

I found some mode matching solutions with a la mode (eg attached), and will implement one tomorrow.

Attachment 1: alaPSOMA.pdf
alaPSOMA.pdf
  3039   Wed Nov 30 11:39:24 2022 KojiDailyProgressPSOMAWincamD/Beam'R2 (Re: cleaning cavity optics)

Regarding WincamD:
WincamD is a silicon-based CCD camera (I believe). And may not have sufficient sensitivity at 1550nm.

Instead, use Beam'R2, which you probably brought back together with the WincamD combo. It has Silicon PD and exInGaAs PD and the exInGaAs can measure the beam profile even at the wavelength of 2um. So, your success is promised. You just need to make sure to use the exInGaAs side.

ELOG V3.1.3-