The most recent measurements on the Taiwan-sourced Glasgow-style cantilver (see CRYO:1213) are encouraging, but the best Q measurement at low temperature is still a couple orders of magnitude worse than what is theoretically achievable, and about one order of magnitude worse than our conservative clamp loss estimates. Also, I've done some measurements on other modes (that have different expected clamp loss contributions due to the relative strain energy ratios) to try and sort out what is going on, with little success. Finally, some modes---including the 2nd bending mode at ~650 Hz---exhibited very low Q for no known reason.
One thing I thought about is that, since the Taiwan cantilever did not fit in the groove that was built into the block for the Glasgow-style cantilevers and therefore is just sandwiched between the two large pieces making up the clamp (see CRYO:1211), the clamp is likely pushing down at somewhat of an angle, which could lead to all sorts of non-idealities. Since the other Si samples we have lying around are roughly the size of the clamping region of this cantilever (~300-500 um), I opened up the cryostat today and reclamped the cantilever using a spare broken-off 300-um-thick cantilever piece as a spacer on the other side:

Pumping it all back down, I immediately measured Qs a bit higher than what we saw last time around at room temperature. The last measurement I made before leaving was tau ~ 135 s ==> Q ~ 46000, though it had been increasing up to that point, likely from the residual pressure, which was at ~10-3 Torr when I left. Compare this with the Q of ~14000 from the last time around, though admittedly I did not record the pressure at which this was measured.

|