40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 9 of 346  Not logged in ELOG logo
ID Date Author Type Category Subject
  17030   Mon Jul 25 09:05:50 2022 PacoSummaryGeneralTesting 950nm laser found in trash pile

[Paco, Yehonathan]

==== Late elog from Friday ====

Koji provided us with a QFLD-950-3S (QPHOTONICS) salvaged from Aidan's junk pile (LD is alive according to him). We tested the Jenne laser setup with this just to decide if we should order another one, and it worked.

The laser driver anode and cathode pins (8/9, 4/5 respectively) on the rear DB9 port from the  ILX Lightwave LDX-3412 driver were connected to the corresponding anode and cathode pins in the laser package (5, and 9; note the numbers are reversed between driver and laser). Then, interlock pins 1 and 2 in the driver were shorted to enable operation. This is all illustrated in Attachments #1-2.

After setting a limit of 27.6 mA current in the driver, we slowly increased the actual current to ~ 19 mA until we could see light on a beam card. We can go ahead and get a 1060 nm replacement.

Attachment 1: PXL_20220722_234600124.jpg
PXL_20220722_234600124.jpg
Attachment 2: PXL_20220722_234551918.jpg
PXL_20220722_234551918.jpg
  17029   Sun Jul 24 08:56:01 2022 HangUpdateCalibrationError propagation to astrophysical parameters from detector calibration uncertainty

Sorry I forgot to put tc & phic in the example. 

I modified astroFisherLib.py to include these parameters. Please note that their meaning is that we don't know when the signal happens and at which phase it merges.

It does not mean the time & phase from a reference frequency to the merger. This part is not free to vary because it is fixed by the intrinsic parameters.  

It might be good to have a quick scan through the Cutler & Flanagan 94 paper to better understand their physical meanings.

 

  17028   Fri Jul 22 17:46:10 2022 yutaConfigurationBHDc1sus2 watchdog update and DCPD ERR channels

[Tega, Yuta]

We have added C1:HPC-DCPD_A_ERR and C1:HPC-DCPD_B_ERR testpoints, which can be used as A+B, A-B etc.
Restarting c1hpc crashed c1sus2, and also made c1lsc/ioo/sus models red.
We run /opt/rtcds/caltech/c1/Git/40m/scripts/cds/restartAllModels.sh to restart all the machines. It worked perfectly without manually pressing power buttons! Wow!heart

We have also edited /opt/rtcds/caltech/c1/medm/c1su2/C1SU2_WATCHDOGS.adl so that it will use new /opt/rtcds/caltech/c1/Git/40m/scripts/SUS/medm/resetFromWatchdogTrip.sh instead of old /opt/rtcds/caltech/c1/scripts/SUS/damprestore.py.

Attachment 1: Screenshot_2022-07-22_17-48-25.png
Screenshot_2022-07-22_17-48-25.png
  17027   Fri Jul 22 17:43:19 2022 KojiUpdateGeneralObtained a functional CRT

[Koji Paco]

Koji went to Downs and found a CRT labeled "for 40m Rana?". So I decided to salvage it to the 40m after getting approval from Rich/Todd.

Paco and I tried this unit with the control room CCD signal and it just worked fine. So we can use this as a spare for any purpose in the lab.

Attachment 1: PXL_20220723_003631871.jpg
PXL_20220723_003631871.jpg
  17026   Fri Jul 22 15:05:26 2022 TegaConfigurationBHDc1sus2 shared memory and ADC fix

[Tega, Yuta]

We were able to fix the shared memory issue by updating the receiver model name from ''SUS' to 'SU2' and the ADC zero issue by including both ADC0 and ADC1 in the c1hpc and c1bac models as well as removing the grounding of the unused ADC channels (including chn#16 and chn#17 which are actually used in c1hpc) in c1su2. We also used shared memory to move the DCPD_A/B error signals (after signal conditioning and mixing A/B; now named A_ERR and B_ERR) from c1hpc to c1bac.
C1:HPC-DCPD_A_IN1 and C1:HPC-DCPD_B_IN1 are now availableangel (they are essentially the same as C1:LSC-DCPD_A_IN1 and C1:LSC-DCPD_B_IN1, except for they are ADC-ed with different ADC; see elog 40m/16954 and Attachment #1).
Dolphin IPC error in seding signal from c1hpc to c1lsc still remains.crying

Attachment 1: Screenshot_2022-07-22_15-04-33_DCPD.png
Screenshot_2022-07-22_15-04-33_DCPD.png
Attachment 2: Screenshot_2022-07-22_15-12-19_models.png
Screenshot_2022-07-22_15-12-19_models.png
Attachment 3: Screenshot_2022-07-22_15-15-11_ERR.png
Screenshot_2022-07-22_15-15-11_ERR.png
Attachment 4: Screenshot_2022-07-22_15-32-19_GDS.png
Screenshot_2022-07-22_15-32-19_GDS.png
  17025   Thu Jul 21 21:50:47 2022 TegaConfigurationBHDc1sus2 IPC update

IPC issue still unresolved.

Updated shared memory tag so that 'SUS' -> 'SU2' in c1hpc, c1bac and c1su2. Removed obsolete 'HPC/BAC-SUS' references from IPC file, C1.ipc. Restarted the FE models but the c1sus2 machine froze, so I did a manual reboot. This brought down the vertex machines---which I restarted using /opt/rtcds/caltech/c1/scripts/cds/rebootC1LSC.sh---and the end machines which I restarted manually. Everything but the BHD optics now have their previous values. So need to burtrestore these.
 

# IPC file:
/opt/rtcds/caltech/c1/chans/ipc/C1.ipc

# Model file locations:
/opt/rtcds/userapps/release/isc/c1/models/isc/c1hpc.mdl
/opt/rtcds/userapps/release/sus/c1/models/c1su2.mdl
/opt/rtcds/userapps/release/isc/c1/models/isc/c1bac.mdl

# Log files:
/cvs/cds/rtcds/caltech/c1/rtbuild/3.4/c1hpc.log
/cvs/cds/rtcds/caltech/c1/rtbuild/3.4/c1su2.log
/cvs/cds/rtcds/caltech/c1/rtbuild/3.4/c1bac.log


SUS overview medm screen :

  • Reduced the entire screen width
  • Revert to old screen style watchdog layout
  17024   Wed Jul 20 18:07:52 2022 PacoUpdateBHDBHD MICH test

[Paco, Yuta, JC]

We did some easy tests on the BHD readout in preparation for BHD MICH. With the arm cavities and LO beam misaligned, but the MICH aligned, we measured the transfer function from C1:LSC-DCPD_A_OUT to C1:LSC-DCPD_B_OUT to get a rough estimate of the gain balance: 1.8 * DCPD_A = DCPD_B. We then locked MICH using REFL55_Q and looked at

  • A=C1:LSC-DCPD_A_OUT
  • B=C1:LSC-DCPD_B_OUT
  • 1.8 * A - B (which we encoded using C1:LSC-PRCL_A_IN1)
  • 1.8 * A + B (which we encoded using C1:LSC-PRCL_B_IN1)

namely the DCPD BHD signals. After turning the MICH_OSC on (2000 gain @ 311.1 Hz), we took some power spectra under the following three configurations:

  1. LO misaligned, no MICH offset.
  2. LO overlap, no MICH offset.
  3. LO overlap and MICH offset.

For 1. the expectation was that since LO is misaligned and the AS port is dark, we would get no signal. In 2., however both A and B would might see some incoherent signal, but still no MICH. Finally in 3. all signals should be able to see MICH, including A-B. Attachment #1 shows the measurements 1, 2, and 3 (offset = -5.0). Then, with increasing offset values, the BHD MICH signals increased as well; discussion to follow.

Attachment 1: BHD_MICH_OSC.pdf
BHD_MICH_OSC.pdf
  17023   Wed Jul 20 15:58:52 2022 KojiSummaryGeneralJenne laser kaput!

For troubleshooting, the proper laser driver (found beneath the AG network analyzer) was connected.
The current ~1mA was provided and the driver detected the "open circuit", which means the laser diode was busted.

https://dcc.ligo.org/LIGO-T060240

The laser diode in the parts list is: "GTRAN GaAs Strained QW Laser Diode, Part # LD-1060".

  17022   Wed Jul 20 14:12:07 2022 PacoSummaryGeneralJenne laser kaput!

[Koji, Yehonathan, Paco]

Koji pointed out that this laser was always driven with a current driver (which was not nearby), and after finding it on one of the rolling carts, we hooked up the system but found that the laser driver displayed open circuit near the usual 20mA operating point. We therefore have to conclude that this laser is no more. We will look for a reasonable replacement.

Quote:

[Paco, Yehonathan, JC]

We were trying to setup the Jenne laser to characterize the response of three 1811s that Yehonathan is using for his WOPA experiment (in QIL). We hooked up a ~ 5 VDC power supply to the bias tee and looked to see if there was any DC response in the REF PD. We used a DB9 breakout board and a DB9 cable, and saw some current being drawn. The DC current was a bit too high (500 mA), so we turned the DC voltage off, and realized the VDC power was reversed, probably along the DB9 cable which we didn't check before. As we flipped the power supply leads and turned power back on, we could no longer see any current even though the voltage was now right (or was it???). We would like to debug this laser, and continue using it if it still works (!), but there is negligible documentation either here or in the wiki, so if there are any known places to look at it would be helpful to know them.

 

  17021   Wed Jul 20 11:58:45 2022 PacoSummaryGeneralJenne laser kaput?

[Paco, Yehonathan, JC]

We were trying to setup the Jenne laser to characterize the response of three 1811s that Yehonathan is using for his WOPA experiment (in QIL). We hooked up a ~ 5 VDC power supply to the bias tee and looked to see if there was any DC response in the REF PD. We used a DB9 breakout board and a DB9 cable, and saw some current being drawn. The DC current was a bit too high (500 mA), so we turned the DC voltage off, and realized the VDC power was reversed, probably along the DB9 cable which we didn't check before. As we flipped the power supply leads and turned power back on, we could no longer see any current even though the voltage was now right (or was it???). We would like to debug this laser, and continue using it if it still works (!), but there is negligible documentation either here or in the wiki, so if there are any known places to look at it would be helpful to know them.

  17020   Tue Jul 19 18:41:42 2022 yutaUpdateBHDContrast measurements for Michelson and ITM-LO

[Paco, Yuta]

We measured contrast of Michelson fringe in both arms locked and mis-aligned. It was around 90%.
We also measured the contrast of ITM single bounce vs LO beam using BHD DC PDs. It was around 43%.
The measurement depends on the alignment and how to measure the maximum and minimum of the fringe. ITM-LO fringe was also not stable because motions of AS/LO mirrors are large. More tuning necessary.

Background
 - As measured in elog 40m/17012, we see a lot of CARM in AS, which indicates large contrast defect.
 - We want to check mode-matching of LO beam to AS beam.

BHD DC PD conditioning
 - We added DCPD_A and DCPD_B to /opt/rtcds/caltech/c1/scripts/LSC/LSCoffsets3 script, which zeros the offsets when shutters are closed.
 - We also set C1:LSC-DCPD_(A|B)_GAIN = -1 since they are inverted.

Contrast measurement
 - Contrast was measured using channels ['C1:LSC-ASDC_OUT','C1:LSC-POPDC_OUT','C1:LSC-REFLDC_OUT','C1:LSC-DCPD_A_OUT','C1:LSC-DCPD_B_OUT']. For LO, only DCPD_(A|B) are used.
 - We took 15%-percentile (40% for ITM-LO fringe) from the maximum and minimum of the data, and took the median to estimate the maximum value and the minimum value (see Attachment).
 - Contrast = (Imax - Imin) / (Imax + Imin)
 - We measured three times in each configuration to estimate the standard error.
 - Jupyter notebook: https://git.ligo.org/40m/scripts/-/blob/main/CAL/BHD/measureContrast.ipynb

Results
Both arms locked, MICH fringe (15% percentile)
Contrast measured by C1:LSC-ASDC_OUT is 89.75 +/- 0.17 %
Contrast measured by C1:LSC-POPDC_OUT is 79.41 +/- 0.86 %
Contrast measured by C1:LSC-REFLDC_OUT is 97.34 +/- 0.34 %
Contrast measured by C1:LSC-DCPD_A_OUT is 95.41 +/- 1.55 %
Contrast measured by C1:LSC-DCPD_B_OUT is 89.76 +/- 1.49 %
Contrast measured by all is 90.34 +/- 1.68 %


Both arms mis-aligned, MICH fringe (15% percentile)
Contrast measured by C1:LSC-ASDC_OUT is 89.32 +/- 0.57 %
Contrast measured by C1:LSC-POPDC_OUT is 94.55 +/- 0.62 %
Contrast measured by C1:LSC-REFLDC_OUT is 97.95 +/- 1.37 %
Contrast measured by C1:LSC-DCPD_A_OUT is 96.40 +/- 1.04 %
Contrast measured by C1:LSC-DCPD_B_OUT is 90.98 +/- 1.07 %
Contrast measured by all is 93.84 +/- 0.94 %


ITMY-LO fringe (40% percentile)
Contrast measured by C1:LSC-DCPD_A_OUT is 45.51 +/- 0.45 %
Contrast measured by C1:LSC-DCPD_B_OUT is 38.69 +/- 0.43 %
Contrast measured by all is 42.10 +/- 1.03 %


ITMX-LO fringe (40% percentile)
Contrast measured by C1:LSC-DCPD_A_OUT is 46.65 +/- 0.65 %
Contrast measured by C1:LSC-DCPD_B_OUT is 39.82 +/- 0.51 %
Contrast measured by all is 43.24 +/- 1.45 %


Discussion
 - As you can see from the attachment, REFLDC is noisy and over estimating the contrast. ASDC is reliable. We need to tune the threshold to measure the maximum value and minimum value. We should also use the mode instead of median.
 - Contrast depends very much on the alignment. We didn't tweak too much today.
 - ITM-LO fringe was not stable, probably due to too much motion in AS1, AS4, LO1, LO2. Their damping needs to be re-tuned.

Next:
 - Model FPMI sensing matrix with measured contrast defect
 - Estimate AS-LO mode-mismatch using the measured contrast
 - Lock ITM-LO fringe using DCPD_(A|B) as error signal, and ITM or LO1/2 as actuator
 - Lock MICH with DCPD_(A|B), and with LO beam
 - Get better contrast data with better alignment and better AS1, AS4, LO1, LO2 damping

Attachment 1: ContrastMeasurements.pdf
ContrastMeasurements.pdf ContrastMeasurements.pdf ContrastMeasurements.pdf ContrastMeasurements.pdf ContrastMeasurements.pdf ContrastMeasurements.pdf ContrastMeasurements.pdf ContrastMeasurements.pdf
  17019   Tue Jul 19 17:18:34 2022 JCUpdateElectronicsNew Coil Driver on Rack 1X3

[Yehonathan, JC]

Yehonathan and I began to put the electronics on Rack 1X3. To do this, we had to move the monitor over the the PD testing table. Before mounting the Coil Drivers, we added numbers to the spaces to follow the rack plan Koji has provided. The drivers which have been mounted are PRM (Slots 10,11), BS (Slots 15, 16), ITMX (Slots 26, 27), and ITMY (34, 35).

Attachment 1: 22DC1767-6073-4D82-BEED-915318B57C03.jpeg
22DC1767-6073-4D82-BEED-915318B57C03.jpeg
  17018   Tue Jul 19 16:00:34 2022 yutaConfigurationBHDFast channels for BHD DCPDs now available in c1lsc but not in c1hpc

[Paco, Anchal-remote-support, Yuta]

We added fast channels to BHD DC PDs.
C1:LSC-DCPD_(A|B)_IN1 are now available, but C1:HPC-DCPD_(A|B)_IN1 still gives us zero.

c1hpc situation -> not good
 - We can see the slow signal at C1:X07-MADC1_EPICS_CH16 (DC PD A) and CH17 (DC PD B)
 - C1:HPC-DCPD_(A|B)_IN1 is there, but zero.
 - We have modified c1hpc model to add DCPD_(A|B) filters in front of the input matrix (see Attachment #1).
 - After modifying the model, we run
ssh c1sus2
rtcds make c1hpc
rtcds install c1hpc
ssh fb1
sudo systemctl restart daqd_*

 - After this, we got 0x2000 error. So, we ran the following. This removed 0x2000 error, but DCPD signals are still zero. They are also not available in C1HPC-MONITOR_ADC1.adl screen (see Attachment #3).
ssh c1sus2
rtcds restart c1hpc


c1lsc situation -> good
 - We could see the slow signal at C1:X04-MADC1_EPICS_CH4 (DC PD A) and CH5 (DC PD B), and also C1:LSC-DCPD_(A|B)_NORM after making C1:LSC-DCPD_(A|B)_POW_NORM=1. The ADC channel and DCPD channel are exactly the same.
 - After confirming the above, we modified the c1lsc model to add DCPD_(A|B) filters in front of the input matrix (see Attachment #2).
 - After modifying the model, we run
ssh c1lsc
rtcds make c1lsc
rtcds install c1lsc
ssh fb1
sudo systemctl restart daqd_*

 - After this, we also got 0x2000 error. We also noticed that, for example, C1:X04-MADC0_EPICS_CH31 and C1:LSC-ASDC_INMON are different, which used to be the same (ASDC_INMON was largely attenuated).
 - In the end, we run the following to remove 0x2000 error, but it crashed c1lsc, as well as c1sus, c1ioo.
ssh c1lsc
rtcds restart c1lsc

 - So, we did rebootC1LSC.sh. This made c1lsc, c1ioo and c1sus as green as before, except for RFM issue in TRX/TRY, like we saw in June. We followed the steps in 40m/16887 to hard reboot c1iscex/c1iscey and ran rebootC1LSC.sh again. This made C1CDS_FE_STATUS.adl screen as green as before (see Attachment #3).

 - Fast channels C1:LSC-DCPD_(A|B)_IN1 are now available. They are also available in C1LSC-MONITOR_ADC1.adl screen (see Attachment #3).

Attachment 1: Screenshot_2022-07-19_14-26-39_c1hpc.png
Screenshot_2022-07-19_14-26-39_c1hpc.png
Attachment 2: Screenshot_2022-07-19_14-24-49_c1lsc.png
Screenshot_2022-07-19_14-24-49_c1lsc.png
Attachment 3: Screenshot_2022-07-19_15-51-25_GreenGreen.png
Screenshot_2022-07-19_15-51-25_GreenGreen.png
  17017   Tue Jul 19 07:34:46 2022 AnchalUpdateCalibrationError propagation to astrophysical parameters from detector calibration uncertainty

Addressing the comments as numbered:

  1. Yeah, that's correct, that equation normally \Delta \Theta = -\mathbf{H}^{-1} \mathbf{M} \Delta \Lambda but it is different if I define \Gamma bit differently that I did in the code, correct my definition of \Gamma to :
    \Gamma_{ij} = \mu_i \mu_j \left( \frac{\partial g}{\partial \mu_i} | \frac{\partial g}{\partial \mu_j} \right )
    then the relation between fractional errors of detector parameter and astrophysical parameters is:
    \frac{\Delta \Theta}{\Theta} = - \mathbf{H}^{-1} \mathbf{M} \frac{\Delta \Lambda}{\Lambda}
    I prefer this as the relation between fractional errors is a dimensionless way to see it.
  2. Thanks for pointing this out. I didn't see these parameters used anywhere in the examples (in fact there is no t_c in documentation even though it works). Using these did not affect the shape of error propagation slope function vs frequency but reduced the slope for chirped Mass M_c by a couple of order of magnitudes.
    1. I used the get_t_merger(f_gw, M1, M2) function from Hang's work to calculate t_c by assuming f_{gw} must be the lowest frequency that comes within the detection band during inspiral. This function is:
      t_c = \frac{5}{256 \pi^{8/3}} \left(\frac{c^3}{G M_c}\right)^{5/3} f_{gw}^{-8/3}
      For my calculations, I've taken f_{gw} as 20 Hz.
    2. I used the get_f_gw_2(f_gw_1, M1, M2, t) function from Hang's work to calculate the evolution of the frequency of the IMR defined as:
      f_{gw}(t) = \left( f_{gw0}^{-8/3} - \frac{768}{15} \pi^{8/3} \left(\frac{G M_c}{c^3}\right)^{5/3} t \right)^{-3/8}
      where f_{gw0} is the frequency at t=0. I integrated this frequency evolution for t_c time to get the coalescence phase phi_c as:
      \phi_c = \int^{t_c}_0 2 \pi f_{gw}(t) dt
  3. In Fig 1, which representation makes more sense, loglog of linear axis plot? Regarding the affect of uncertainties on Tidal amplitude below 500 Hz, I agree that I was also expecting more contribution from higher frequencies. I did find one bug in my code that I corrected but it did not affect this point. Maybe the SNR of chosen BNS parameters (which is ~28) is too low for tidal information to come reliably anyways and the curve is just an inverse of the strain noise PSD, that is all the information is dumped below statistical noise. Maybe someone else can also take a look at get_fisher2() function that I wrote to do this calculation.
  4. Now, I have made BBH parameters such that the spin of the two black holes would be assumed the same along z. You were right, the gamma matrix was degenerate before. To your second point, I think the curve also shows that above ~200 Hz, there is not much contribution to the uncertainty of any parameter, and it rolls-off very steeply. I've reduced the yspan of the plot to see the details of the curve in the relevant region.
Quote:

1. In the error propogation equation, it should be \Delta \Theta = -H^{-1} M \Delta \Lambda, instead of the fractional error. 

2. For the astro parameters, in general you would need t_c for the time of coalescence and \phi_c for the phase. See, e.g., https://ui.adsabs.harvard.edu/abs/1994PhRvD..49.2658C/abstract.

3. Fig. 1 looks very nice to me, yet I don't understand Fig. 3... Why would phase or amplitude uncertainties at 30 Hz affect the tidal deformability? The tide should be visible only > 500 Hz.

4. For BBH, we don't measure individual spin well but only their mass-weighted sum, \chi_eff = (m_1*a_1 + m_2*a_2)/(m_1 + m_2). If you treat S1z and S2z as free parameters, your matrix is likely degenerate. Might want to double-check. Also, for a BBH, you don't need to extend the signal much higher than \omega ~ 0.4/M_tot ~ 10^4 Hz * (Ms/M_tot). So if the total mass is ~ 100 Ms, then the highest frequency should be ~ 100 Hz. Above this number there is no signal.

 

Attachment 1: BNSparamsErrorwrtfdError.pdf
BNSparamsErrorwrtfdError.pdf BNSparamsErrorwrtfdError.pdf
Attachment 2: BBHparamsErrorwrtfdError.pdf
BBHparamsErrorwrtfdError.pdf BBHparamsErrorwrtfdError.pdf
Attachment 3: BNSparamsEPSwrtCalError.pdf
BNSparamsEPSwrtCalError.pdf
Attachment 4: BBHparamsEPSwrtCalError.pdf
BBHparamsEPSwrtCalError.pdf
  17016   Mon Jul 18 21:41:42 2022 AnchalSummaryLSCFPMI locking procedure using REFL55 and AS55

Now that you have found a working configuration, I suggest we update CARM and DARM filter banks so that they are used in locking those degrees of freedom instead of repurposing XARM/YARM banks. It would be bit easier to understand and leaves room for future changes for one configuration while keeping single arm lock configurations untouched.

  17015   Mon Jul 18 18:33:38 2022 KojiUpdateBHDadd Laser RIN to MICH budget

You should measure the coupling by noise injection. Noise budgeting does not need any modeling:

1) Measure the power spectrum density of the target signal (i.e. DARM) and the source noise (i.e. RIN this case)

2) Calibrate both using a calibration peak to convert 1) into the physical units (m/rtHz, 1/rtHz, etc)

3) Measure the transfer function from source to target using the noise injection. (i.e. RIN injection this case and look at the injection to RIN and injection to DARM)

4) Measure open-loop transfer functions if necessary. (i.e. DARM control open-loop transfer function to convert the error signal into the free running noise level)

Primarily, these are measured noise levels and noise couplings there is no room to involve a model there.
Once the noise budget was done, you can compare it with the model and say "the coupling is big/small/comparable".
 

Also, why don't you use C1:MC_TRANS_SUMFILT_IN1_DQ instead? Your _OUT signal seems affected by the bunch of comb notch filters to artificially remove the 60Hz harmonics. It's not a fair RIN measurement.

  17014   Mon Jul 18 17:07:12 2022 yutaUpdateLSCx4.12 added to ETMX coil outputs to balance with ETMY

To balance the actuation on ETMX and ETMY, x4.12 was aded to C1:SUS-ETMX_(UL|UR|LR|LL|SD)COIL FM1. OSEM damping filter gains, oplev loop gains, and alignment offsets were divided by this factor.
C1:LSC-ETMX_GAIN is now 1.

To do:
 - Balance ETM and ITM. It should make ASS more sensible.
 - Re-commission Xarm ASS and Yarm ASS.

  17013   Mon Jul 18 16:49:57 2022 YehonathanUpdateBHDadd Laser RIN to MICH budget

I measured the RIN by taking the spectrum of C1:MC_TRANS_SUMFILT_OUT and dividing it by the mean count on that channel (~13800 cts). Attachment 1 shows the result.

I updated the MICH AS55 noise budget but got a very low contribution (gold trace in attachment 2).

It seems too low I think. What could've gone wrong? Finesse calculates that the transfer function from laser amplitude modulation to AS55 is ~ 1.5e-9 at DC. If I turn off HOMs I get 1e-11 at DC, so this coupling is a result of some contrast defect. Should I include some RMS imbalances in the optics to account for this? Should I include it as a second-order effect due to MICH RMS deviation from zero crossing?

Quote:

the main laser noise coupling for a Michelson is because of the RIN, not the frequency noise. You can measure the RIN, in MC trans or at the AS port by getting a single bounce beam from a single ITM.

 

Attachment 1: Laser_RIN.pdf
Laser_RIN.pdf
Attachment 2: MICH_AS55_Noise_Budget.pdf
MICH_AS55_Noise_Budget.pdf
  17012   Mon Jul 18 16:39:07 2022 PacoSummaryLSCFPMI locking procedure using REFL55 and AS55

[Yuta, Paco]

In summary, we locked FPMI using REFL55_I, REFL55_Q, and AS55_Q. The key to success was to mix POX11_I and POY11_I in the right way to emulate CARM/DARM, and to find out the correct demodulation phase for AS55.


Procedure

  1. Close PSL shutter and zero offsets in AS55, REFL55, POX11, POY11, and ASDC
    • For ASDC run python3 resetOffsets.py -c C1:LSC-ASDC_IN1, otherwise use the zer offsets on I and Q inputs from the RFPD medm screen.
  2. Lock XARM/YARM using POX/POY to tune demodulation phase.
    • Today, the demode phase in POX11 changed to 104.801, and POY11 to -11.256 deg.
  3. XARM and YARM are used in the following configuration
    • INMAT
      • 0.5 * POX11_I - 0.5 * POY --> XARM
      • 0.5 * POX + 0.5*POY --> YARM
      • REFL55_Q --> MICH (** this should be turned on after POX11/POY11)
    • LSC Filter gains
      • XARM = 0.012
      • YARM = 0.012
      • MICH = +40 (note the sign flip from last time)
    • OUTMAT
      • XARM --> 0.5 * ETMX - 0.5 * ETMY
      • YARM --> MC2
      • MICH --> BS
    • UGFs (sanity check)
      • XARM (DARM) ~ 100 Hz
      • YARM (CARM) ~ 200 Hz
      • MICH (MICH) ~ 40 Hz
  4. Run MICHOpticalGainCalibration.ipynb to see if ASDC vs REFL55_Q looks nice (ellipse in the XY plot), and find any residual offset in REFL55_Q.
    • If the plot doesn't look nice in this regard, the IFO needs to be aligned.
  5. Sensing matrix for CARM/DARM and MICH.
    • With the DARM, CARM and MICH lines on, verify the demod error signals look ok both in mag and phase.
    • For example, we found that CARM error signals were correctly represented by either 0.5 * POX11_I + 0.5 * POY11_I or 0.5 * REFL55_I.
    • Similarly, we found that DARM error signal was correctly represented by either 0.5 * POX11_I - 0.5 * POY11_I or 2.5 * AS55_Q.
    • To find this, we minimized CARM content in AS55_Q, as well as CARM content in REFL55_Q.
  6. We acquired the lock by re-configuring the error point as below:
    • INMAT
      • 0.5*REFL55_I --> YARM (CARM)
      • 2.5 * AS55_Q --> XARM (DARM)
    • During the hand-off trials, we repeatedly ran the sensing matrix and UGF measurements while stopping at various intermediate mixed error points to check how the error signal calibrations changed if at all.
      • Attachment #1 shows the DARM OLTF using POX/POY (blue), only with CARM handoff (green), and after DARM handoff (red)
      • Attachment #2 shows the CARM OLTF using POX/POY (blue), only with CARM handoff (green), and after DARM handoff (red)
      • Attachment #3 shows the MICH OLTF using POX/POY (blue), only with CARM handoff (green), and after DARM handoff (red)
    • The sensing matrix after handoff is below:
Sensing Matrix with the following demodulation phases
{'AS55': 192.8, 'REFL55': 95.63177865911078, 'POX11': 104.80089727128349, 'POY11': -11.256509422276006}
Sensors          	           DARM     	           CARM     	            MICH     	
C1:LSC-AS55_I_ERR_DQ	5.09e-02 (89.6761 deg)	2.03e-01 (-114.513 deg)	1.28e-04 (-28.9254 deg)	
C1:LSC-AS55_Q_ERR_DQ	4.78e-02 (88.7876 deg)	3.61e-03 (-68.7198 deg)	8.34e-05 (-39.193 deg)	
C1:LSC-REFL55_I_ERR_DQ	5.18e-02 (-92.2555 deg)	1.20e+00 (65.2507 deg)	1.15e-04 (-102.027 deg)	
C1:LSC-REFL55_Q_ERR_DQ	1.81e-04 (59.0854 deg)	1.09e-02 (-114.716 deg)	1.77e-05 (-23.6485 deg)	
C1:LSC-POX11_I_ERR_DQ	8.51e-02 (91.2844 deg)	4.77e-01 (67.1709 deg)	7.97e-05 (-72.5252 deg)	
C1:LSC-POX11_Q_ERR_DQ	2.63e-04 (114.584 deg)	1.32e-03 (-113.505 deg)	2.10e-06 (118.146 deg)	
C1:LSC-POY11_I_ERR_DQ	1.58e-01 (-88.9295 deg)	6.16e-01 (67.6098 deg)	8.71e-05 (172.73 deg)	
C1:LSC-POY11_Q_ERR_DQ	2.89e-04 (-89.1114 deg)	1.09e-03 (70.2784 deg)	3.77e-07 (110.206 deg)	

Lock gpstimes:

  1. [1342220242, 1342220260]
  2. [1342220420, 1342220890]
  3. [1342221426, 1342221574]
  4. [1342222753, 1342223230]

Sensitivity estimate (NANB)

Using diaggui, we look at the AS55_Q error point and the DARM control point (C1:LSC-XARM_OUT). We roughly calibrate the error point using the sensing matrix element and actuation gain at the DARM oscillator freq 4.78e-2 / (10.91e-9 / 307.880^2). The control point is calibrated with a 0.95 Hz SUS pole. Attachment #4 shows the sensitivity estimate.

Attachment 1: DARM_07_18_2022_FMPI.pdf
DARM_07_18_2022_FMPI.pdf DARM_07_18_2022_FMPI.pdf DARM_07_18_2022_FMPI.pdf
Attachment 2: CARM_07_18_2022_FPMI.pdf
CARM_07_18_2022_FPMI.pdf CARM_07_18_2022_FPMI.pdf CARM_07_18_2022_FPMI.pdf
Attachment 3: MICH_07_18_2022_FPMI.pdf
MICH_07_18_2022_FPMI.pdf MICH_07_18_2022_FPMI.pdf MICH_07_18_2022_FPMI.pdf
Attachment 4: fpmi_darm_nb_2022_07.pdf
fpmi_darm_nb_2022_07.pdf
  17011   Mon Jul 18 15:17:51 2022 HangUpdateCalibrationError propagation to astrophysical parameters from detector calibration uncertainty

1. In the error propogation equation, it should be \Delta \Theta = -H^{-1} M \Delta \Lambda, instead of the fractional error. 

2. For the astro parameters, in general you would need t_c for the time of coalescence and \phi_c for the phase. See, e.g., https://ui.adsabs.harvard.edu/abs/1994PhRvD..49.2658C/abstract.

3. Fig. 1 looks very nice to me, yet I don't understand Fig. 3... Why would phase or amplitude uncertainties at 30 Hz affect the tidal deformability? The tide should be visible only > 500 Hz. 

4. For BBH, we don't measure individual spin well but only their mass-weighted sum, \chi_eff = (m_1*a_1 + m_2*a_2)/(m_1 + m_2). If you treat S1z and S2z as free parameters, your matrix is likely degenerate. Might want to double-check. Also, for a BBH, you don't need to extend the signal much higher than \omega ~ 0.4/M_tot ~ 10^4 Hz * (Ms/M_tot). So if the total mass is ~ 100 Ms, then the highest frequency should be ~ 100 Hz. Above this number there is no signal. 

 

  17010   Mon Jul 18 04:42:54 2022 AnchalUpdateCalibrationError propagation to astrophysical parameters from detector calibration uncertainty

We can calculate how much detector calibration uncertainty affects the estimation of astrophysical parameters using the following method:

Let \overrightarrow{\Theta} be set of astrophysical parameters (like component masses, distance etc), \overrightarrow{\Lambda}be set of detector parameters (like detector pole, gain or simply transfer function vaue for each frequency bin). If true GW waveform is given by h(f; \overrightarrow{\Theta}), and the detector transfer function is given by \mathcal{R}(f; \overrightarrow{\Lambda}), then the detected gravitational waveform becomes:
g(f; \Theta, \Lambda) = \frac{\mathcal{R}(f; \overrightarrow{\Lambda_t})}{\mathcal{R}(f; \overrightarrow{\Lambda})} h(f; \overrightarrow{\Theta})

One can calculate a derivative of waveform with respect to the different parameters and calculate Fisher matrix as (see correction in 40m/17017):

\Gamma_{ij} = \left( \frac{\partial g}{\partial \mu_i} | \frac{\partial g}{\partial \mu_j}\right )

where the bracket denotes iner product defined as:

\left( k_1 | k_2 \right) = 4 Re \left( \int df \frac{k_1(f)^* k_2(f))}{S_{det}(f)}\right)

where S_{det}(f) is strain noise PSD of the detector.

With the gamma matrix in hand, the error propagation from detector parameter fractional errors \frac{\Delta \Lambda_j}{\Lambda_j}to astrophysical paramter fractional errors \frac{\Delta \Theta_i}{\Theta_i}is given by (eq 26 in Evan et al 2019 Class. Quantum Grav. 36 205006):

\frac{\Delta \Theta_j}{\Theta_j} = - \mathbf{H}^{-1} \mathbf{M} \frac{\Delta \Lambda_j}{\Lambda_j}

where \mathbf{H}_{ij} = \left( \frac{\partial g}{\partial \Theta_i} | \frac{\partial g}{\partial \Theta_j}\right ) and \mathbf{M}_{ij} = \left( \frac{\partial g}{\partial \Lambda_i} | \frac{\partial g}{\partial \Theta_j}\right ).


Using the above mentioned formalism, I looked into two ways of calculating error propagation from detector calibration error to astrophysical paramter estimations:

Using detector response function model:

If we assume detector response function as a simple DC gain (4.2 W/nm) and one pole (500 Hz) transfer function, we can plot conversion of pole frequency error into astrophysical parameter errors. I took two cases:

  • Binary Neutron Star merger with star masses of 1.3 and 1.35 solar masses at 100 Mpc distance with a \tilde{\Lambda} of 500. (Attachment 1)
  • Binary black hole merger with black masses of 35 and 30 at 400 MPc distance with spin along z direction of 0.5 and 0.8. (I do not fully understand the meaning of these spin components but a pycbc waveform generation model still lets me calculate the effect of detector errors) (Attachment 2)

The plots are plotted in both loglog and linear plots to show the order of magnitude effect and how the error propsagation slope is different for different parameters. 'm still not sure which way is the best to convey the information. The way to read this plot is for a given error say 4% in pole frequency determination, what is the expected error in component masses, merger distance etc. I

Note that the overall gain of detector response is not sensitive to astrophysical error estimation.

Using detector transfer function as frequency bin wise multi-parameter function

Alternatively, we can choose to not fit any model to the detector transfer function and simply use the errors in magnitude and phase at each frequency point as an independent parameter in the above formalism. This then lets us see what is the error propagation slope for each frequency point. The hope is to identify which parts of the calibration function are more important to calibrate with low uncertainty to have the least effect on astrophysical parameter estimation. Attachment 3 and 4 show these plots for BNS and BBH cases mentioned above. The top panel is the error propagation slope at each frequency due to error in magnitude of the detector transfer function at that frequency and the bottom panel is the error propagation slope at each frequency due to error in phase of the detector transfer function.

The calibration error in magnitude and phase as a function of frequency would be multiplied by the curves and summed together, to get total uncertainty in each parameter estimation.


This is my first attempt at this problem, so I expect to have made some mistakes. Please let me know if you can point out any. Like, do the order of magnitude and shape of error propagation makes sense? Also, comments/suggestions on the inference of these plots would be helpful.

Finally, I haven't yet tried seeing how these curves change for different true values of the merger event parameters. I'm not yet sure what is the best way to extract some general information for a variety of merger parameters.

Future goals are to utilize this information in informing system identification method i.e. multicolor calibration scheme parameters like calibration line frequencies and strength.

Code location

Attachment 1: BNSparamsErrorwrtfdError-merged.pdf
BNSparamsErrorwrtfdError-merged.pdf BNSparamsErrorwrtfdError-merged.pdf
Attachment 2: BBHparamsErrorwrtfdError-merged.pdf
BBHparamsErrorwrtfdError-merged.pdf BBHparamsErrorwrtfdError-merged.pdf
Attachment 3: BNSparamsEPSwrtCalError.pdf
BNSparamsEPSwrtCalError.pdf
Attachment 4: BBHparamsEPSwrtCalError.pdf
BBHparamsEPSwrtCalError.pdf
  17009   Sat Jul 16 02:44:10 2022 KojiUpdateIOOIMC servo tuning

I wasn't sure how the IMC servo was optimized recently. We used to have the FSS over all gain (C1:PSL-FSS_MGAIN) of +6dB a few years back. It is not 0dB. So I decided to do a couple of measurements.

1) Default setting:

C1:IOO-MC_REFL_GAIN +4
C1:IOO-MC_BOOST2 +3
C1:IOO-MC_VCO_GAIN +13
C1:PSL-FSS_MGAIN +0
C1:PSL-FSS_FASTGAIN +19

2) Looked at the power spectrum at TEST1A output (error signal)
TEST1A is the signal right after the input gain stage (C1:IOO-MC_REFL_GAIN). Prior to the measurement, I've confirmed that the UGF is ~100Hz even at +0dB (see next section). It was not too bad even with the current default. Just wanted to check if we can increase the gain a bit more.
The input gain was fixed at +4dB and the FSS overall gain C1:PSL-FSS_MGAIN was swept from +0 to +6.
At +5dB and +6dB, the servo bump was very much visible (Attachment 1).
I decided to set the default to be +4dB (Attachment 3).

3) Took OLTF at 0dB and 4dB for the FSS overall gain.

Now the comparison of the opel loop transfer functions (OLTF) for C1:PSL-FSS_MGAIN at 0dB and 4dB. The OLTF were taken by injectiong the network analyzer signal into EXCA and measure the ratio between TEST1A and TEST1B (A/B).

C1:PSL-FSS_MGAIN +0 -> UGF 100kHz / Phase Margin ~50deg
C1:PSL-FSS_MGAIN +4 -> UGF 200kHz / Phase Margin 25~30deg

The phase margin was a bit less but it was acceptable.

4) IMC FSR

Took the opportunity to check the FSR of the IMC. Connected a cable to the RF MON of the IMC REFL demod board. Looked at the peak at 40.56MHz (29.5MHz + 11.066MHz). The peak was not so clear at 11.066195MHz (see 40m ELOG 15845). The peak was anyway minimized and the new modulation frequency was set to be 11.066081MHz (new FSR). The change is 10ppm level and it is within the range of the temp drift.

Attachment 1: ErrorPSD.pdf
ErrorPSD.pdf
Attachment 2: OLTF.pdf
OLTF.pdf
Attachment 3: Screen_Shot_2022-07-16_at_03.59.05.png
Screen_Shot_2022-07-16_at_03.59.05.png
  17008   Fri Jul 15 22:36:04 2022 ranaSummaryLSCFPMI with REFL/AS55 demod phase adjust

Very nice!

DARM feedback should go to ETMY - ETMX, not just a single mirror: Differential ARM.

For it to work with 1 mirror the UGF of the CARM loop must be much larger than DARM UGF. But in our case, both have a UGF of ~150 Hz.

In principle, you could run the CARM loop with higher gain by using the CM servo board, but maybe that can wait until the X,Y -> CARM, DARM handoff.

 

  17007   Fri Jul 15 19:13:22 2022 PacoSummaryLSCFPMI with REFL/AS55 demod phase adjust

[Yuta, Paco]

  • We first zero the offsets in ASDC, AS55, REFL55, POX11, and POY11 when PSL shutter is closed.
    • After this, we checked the offsets with only ITMX aligned. Some of RFPDs had ~2 counts of offsets, which indicate some RFAM of sidebands, but we decided not to tune Marconi frequencies since the offsets were small enough.
  • We went over the demod phases for AS55, REFL55, POX11, and POY11.
    • For POX11/POY11 first we just minimized the Q in each locked XARM/YARM individually. The newfound values were
      • C1:LSC-POX11_PHASE_R = 106.991
      • C1:LSC-POY11_PHASE_R = -12.820
    • Then we misaligned the XARM by getting rid of the MICH fringe in the ASDC port with ITMX yaw offset, and locked YARM using AS55_Q and REFL55_I and found the demod phase that minimized the AS55_I and REFL55_Q. The newfound values were
      • C1:LSC-AS55_PHASE_R = -65.9586
      • C1:LSC-REFL55_PHASE_R = -78.6254
    • Repeating the above, but now misaligning YARM with ITMY yaw offset, locking XARM with AS55_Q and REFL55_I, we found the demod phases that minimized AS55_1 and REFL55_Q. The newfound values were
      • C1:LSC-AS55_PHASE_R = -61.4361
      • C1:LSC-REFL55_PHASE_R = -71.0434
  • The above demod phases difference, Schnupp asymmetry between X and Y were measured. We repeated the measurement three times to derive the error.
    • Optimal demod phase difference between X arm and Y arm for both AS55 and REFL55 were measured to be -4.5 +/- 0.1 deg, which means that lx-ly = 3.39 +/- 0.05 cm (Marconi frequency: 11.066195 MHz).
  • We measured the gain difference between AS55_Q and POX11/POY11 = -0.5
  • We measured the gain difference between REFL55_I and POX11/POY11 = -2.5

After this, we locked DARM, CARM and MICH using POX11_I, POY11_I and AS55 error signals respectively, and actuating on ETMX, MC2, and BS with NO TRIGGERS (but FM triggers were on for boosts as usual). Under this condition, FM5 is used for lock acquisition, and FM1, FM2, FM3, FM6 are turned on with FM triggers. No FM4 was on. We also noticed:

  • CARM FM6 "BounceRoll" is slightly different than "YARM" FM6 "Bounce". The absent roll resonant gain actually makes it easier to control the CARM, we just had to use YARM filter for locking it.
  • When CARM is controlled, we often just kick the ETMX to bring it near resonance, since the frequency noise drops and we otherwise have to wait long.
  17006   Fri Jul 15 16:20:16 2022 Cici HannaUpdateGeneralFinding UGF

I have temporarily abandoned vectfit and aaa since I've been pretty unsuccessful with them and I don't need poles/zeroes to find the unity gain frequency. Instead I'm just fitting the transfer function linearly (on a log-log scale). I've found the UGF at about 5.5 kHz right now, using old data - next step is to get the Red Pitaya working so I can take data with that. Also need to move this code from matlab to python. Uncertainty's propagated using the 95% confidence bounds given by the fit, using curvefit - so just from the standard error, and all points are weighted equally. Ideally would like to propagate uncertainty accounting for the coherence data too, but haven't figured out how to do that correctly yet.

 

[UPDATE 7/22/2022: added raw data files]

Attachment 1: UGF_4042.png
UGF_4042.png
Attachment 2: UGF_5650.png
UGF_5650.png
Attachment 3: TFSR785_29-06-2022_114042.txt
# SR785 Measurement - Timestamp: Jun 29 2022 - 11:40:42
# Parameter File: TFSR785template.yml
#---------- Measurement Setup ------------
# Start frequency (Hz) = 100000.000000
# Stop frequency (Hz) = 100.000000
# Number of frequency points = 30
# Excitation amplitude (mV) = 10.000000
# Settling cycles = 5
# Integration cycles = 100
#---------- Measurement Parameters ----------
... 52 more lines ...
Attachment 4: TFSR785_29-06-2022_115650.txt
# SR785 Measurement - Timestamp: Jun 29 2022 - 11:56:50
# Parameter File: TFSR785template.yml
#---------- Measurement Setup ------------
# Start frequency (Hz) = 100000.000000
# Stop frequency (Hz) = 2000.000000
# Number of frequency points = 300
# Excitation amplitude (mV) = 5.000000
# Settling cycles = 5
# Integration cycles = 200
#---------- Measurement Parameters ----------
... 322 more lines ...
  17005   Fri Jul 15 12:21:58 2022 JCUpdateElectronicsChecking Sorensen Power Supplies

Of the 7 Sorenson Power Supplies I tested, 5 are working fine, 1 cannot output voltage more than 20 Volts before shorting, and other does not output current. Six Sorensons are behind the X-Arm.

Quote:

[JC]

I went around 40m picking up any Sorensens that were laying around to test if they worked, or in need of repair. I gathered up a total of 7 Sorensens and each one with a Voltmeter. I made sure the voltage would rise on the Sorenson as well as the voltmeter, maxing out at ~33.4 Volts. For the current, the voltmeter can only rise to 10 Amps before it is fused. Many of the Sorensons that I found did not have their own wall connection, so I had to use the same one for multiple.

From these 7, I have found 5 that are well. One Sorenson I have tested has a output shortage above 20V and the other has yet to be tested.

 

Attachment 1: 50DF21D7-D61A-4674-B0DA-463378B00ADB.jpeg
50DF21D7-D61A-4674-B0DA-463378B00ADB.jpeg
Attachment 2: FA4CF579-6C1E-48D5-B152-74F35B4EE90B.jpeg
FA4CF579-6C1E-48D5-B152-74F35B4EE90B.jpeg
  17004   Thu Jul 14 19:56:15 2022 ranaUpdateIOOmc wfs demod

It looks like Tomislav's measurements of the WFS demod board noise were actually of the cable that goes from the whitening to the ADC. So the huge low frequency excess that he saw is not due to wind, but just the inverse whitening of the digital system?

In any case, today, I looked at the connections from the Whitening to the ADC. It goes through an interface chassis to go from ribbon to SCSI. The D-Sub connectors there have the common problem in many of the LIGO D-sub connectors: namely that the strain relief nuts are too tall and so the D connector doesn't seat firmly - its always about to fall out. JC, can you please take a look at this and order a set of low profile nuts so that we can rework this chassis? Its the one between the WFS whitening and the SCSI cables which go to the ADCs.

After pushing them in, I confirmed that the WFS are working, by moving all 6 DoF of the MC mirrors via bias slider, and looking at the step responses (attached). As you can see, all sensors see all mirrors, even if they are noisy.

Next up: get a breakout for the demod output connector and measure the noise there.

For today, I aligned the IMC by hand, then centred the WFS beams by unlocking the IMC and aligning the bright beam. I noticed that the WFS1 beam was being dumped randomly, so I angled the WFS1 by ~3 deg and dumped the specular reflection on a razor blade dump. To handle the sign change in the MC1 actuation (?), I changed the sign in the MC1 ASC filter banks. MCWFS loops still nto closing, but they respond to mirror alignment.

Attachment 1: mcwfs-steps.pdf
mcwfs-steps.pdf
  17003   Thu Jul 14 19:09:51 2022 ranaUpdateGeneralEQ recovery

There was a EQ in Ridgecrest (approximately 200 km north of Caltech). It was around 6:20 PM local time.

All the suspensions tripped. I have recovered them (after some struggle with the weird profusion of multiple conflicting scripts/ directories that have appeared in the recent past...)

ETMY is still giving me some trouble. Maybe because of the HUGE bias on that within the fast CDS system, it had some trouble damping. Also the 'reenable watchdog' script in one of the many scripts directories seems to do a bad job. It re-enables optics, btu doesn't make sure that the beams are on the optical lever QPD, and so the OL servo can smash the optic around. This is not good.

Also what's up with the bashrc.d/ in some workstations and not others? Was there something wrong with the .bashrc files we had for the past 15 years? I will revert them unless someone puts in an elog with some justification for this "upgrade".

This new SUS screen is coming along well, but some of the fields are white. Are they omitted or is there something non-functional in the CDS? Also, the PD variances should not be in the line between the servo outputs and the coil. It may mislead people into thinking that the variances are of the coils. Instead, they should be placed elsewhere as we had it in the old screens.

Attachment 1: ETMY-screen.png
ETMY-screen.png
  17002   Thu Jul 14 00:10:08 2022 yutaSummaryLSCFPMI with REFL/AS55 trial continued

[Paco, Koji, Yuta]

We managed to lock MICH using REFL55_Q by setting the demodulation phases and offsets right.
The following is the current FPMI locking configuration we achieved so far.

DARM: POX11_I / gain 0.007 / 0.5*ETMX-0.5*ETMY (or 1*ETMX) / UGF of ~100 Hz
CARM: POY11_I / gain 0.018 / 1*MC2 / UGF of ~200 Hz
MICH: REFL55_Q / gain -10 / 0.5*BS / UGF of ~30 Hz

Transitioning DARM error signal from POX11_I to 0.5*POX11_I+0.5*POY11_I was possible with FM4 filter off in DARM filter bank, but not to AS55_Q yet.

REFL55 and AS55 demodulation phase tuning:
 - We found that both AS55 and REFL55 are contaminated by large non-MICH signal, by making a ASDC vs RF plot (see 40m/16929).
 - After both arms are locked with POX and POY, MICH was locked with AS55_Q. ASDC was minimized by putting an offset to MICH filter.
 - With this, REFL55 offsets were zeroed and demodulation phase was tuned to minimize REFL55_Q.
 - Locked MICH with REFL55_Q, and did the same thing for AS55_Q.
 - Resulting ASDC vs RF plots were attached. REFL55_Q now looks great, but REFL55_I and AS55 are noisy (due to signals from the arms?).

Jupyter notebook: https://git.ligo.org/40m/scripts/-/blob/main/CAL/MICH/MICHOpticalGainCalibration.ipynb

Sensing matrix:
 - With FPMI locked using POX/POY, DARM and CARM lines were injected at around 300 Hz to measure the sensing gains. For line injection, C1:CAL-SENSMAT was used, but for the demodulation we used a script. The following is the result.

 Sensors              DARM (ETMX)         CARM (MC2)        
C1:LSC-AS55_I_ERR    3.10e+00 (-34.1143 deg)    1.09e+01 (-14.907 deg)    
C1:LSC-AS55_Q_ERR    9.96e-01 (-33.9848 deg)    3.30e+00 (-27.9468 deg)    
C1:LSC-REFL55_I_ERR    6.75e+00 (-33.7723 deg)    2.92e+01 (-34.0958 deg)    
C1:LSC-REFL55_Q_ERR    7.07e-01 (-33.4296 deg)    3.08e+00 (-33.4437 deg)    
C1:LSC-POX11_I_ERR    3.97e+00 (-33.9164 deg)    1.51e+01 (-30.7586 deg)    
C1:LSC-POY11_I_ERR    6.25e-02 (-20.3946 deg)    3.59e+00 (38.4207 deg)

Jupyter notebook: https://git.ligo.org/40m/scripts/-/blob/main/CAL/SensingMatrix/MeasureSensMat.ipynb

 - By taking the ratios of POX11_I and AS55_Q for DARM, POY11_I and REFL55_I for CARM, we tried to find the correct gains for REFL55 and AS55 for DARM and CARM. x3.96 more gain for AS55_Q than POX11_I and x0.123 less gain for REFL55_I than POY11_I.

Next:
 - Try locking the arms with no triggering, and then try locking FPMI with REFL/AS without triggering. No FM4 for this, since FM4 kills gain margin.
 - Lock single arm with AS55_Q and make a noise budget. Make sure to misalign ITMX(Y) completely when locking Y(X)arm.
 - Lock single arm with REFL55_I and make a noise budget.
 - Repeat Xarm noise budget with Yarm locked with POY11_I and MC2 (40m/16975).
 - Check IMC to reduce frequency noise (40m/17001)

Attachment 1: AS55_I.png
AS55_I.png
Attachment 2: AS55_Q.png
AS55_Q.png
Attachment 3: REFL55_I.png
REFL55_I.png
Attachment 4: REFL55_Q.png
REFL55_Q.png
  17001   Wed Jul 13 18:58:17 2022 KojiUpdateIOOIMC suspecion

This is just my intuition but the IMC servo seems not so optimized. I can increase the servo gain by 6~10dB easily. And I couldn't see that the PC drive went mad (red) as I increase the gain (=UGF).
The IMC needs careful OLTF measurements as well as the high freq spectrum observation.

It seems that I have worked on the IMC servo tuning in 2014 July/Aug. Checking these elogs would be helpful.

  17000   Wed Jul 13 17:30:19 2022 KojiUpdateCDSToo huge script_archive

I wanted to check the script archive to see some old settings. I found that the script archive inflated to huge volume (~1TB).
The size of the common NFS volume (/cvs/cds) is 3TB. So it is really significant.

- The scripts living in /opt/rtcds/caltech/c1/scripts are archived daily in /cvs/cds/caltech/scripts_archive as bz2 files. This is done by crontab of megatron (see https://wiki-40m.ligo.caltech.edu/Computers_and_Scripts/CRON)

- In fact, the script folder (say old script folder) /opt/rtcds/caltech/c1/scripts has the size of 10GB. And we have the compressed copy of thi s everyday.

- This large script folder is due to a couple of huge files/folders listed below

  • (scripts)/MEDMtab is 5.3GB / This is probably related to the web MEDM view (on nodus) but I don't think the web page is not updated. (i.e. the images are unused)
  • (scripts)/MC/logs/AutoLocker.log 2.9GB / This is just the accumulated MC autolocker log.
  • (scripts)/GigE 780M / This does not look like scripts but source and object files
  • (scripts)/Admin/n2Check.log 224M / This is important but increases every minute.
  • (scripts)/ZI 316MB / Zurich Instrument installation. This should not be here.

Here I propose some changes.
For the script archive

  • We can remove most of the scripts for the past (say ~2019). We leave an archive file per month.
  • For the scripts in 2020, we leave a weekly archive.
  • For 2021 and 2022, we leave all the archive files.

For the existing large files/folders

  • MEDMtab: the stored files are redundant with the burt snapshots. Remove the image files. Also, we want to move the image-saving location.
  • Autolocker.log: simply zap it
  • n2Check.log: we should move the saving location
  • GigE /ZI: they need a new home where the daily copy is not taken.
  16999   Wed Jul 13 13:30:48 2022 YehonathanUpdateBHDadd Laser RIN to MICH budget

the main laser noise coupling for a Michelson is because of the RIN, not the frequency noise. You can measure the RIN, in MC trans or at the AS port by getting a single bounce beam from a single ITM.

  16998   Wed Jul 13 13:26:44 2022 ranaSummaryElectronicsElectronics noise measurements

as I said to you yesterday, I don't think image 2a shows the output of the demod board. The output of the demod board is actually the output connector ON the demod board. What you are showing in 2a, is the signal that goes from the whitening board to the ADC I believe. I may be msitaken, so please check with Tega for the signal chain.

  16997   Wed Jul 13 12:49:25 2022 PacoSummarySUSSUS frozen

[Paco, JC, Yuta]

This morning, while investigating the source of a burning smell, we turned off the c1SUS 1X4 power strip powering the sorensens. After this, we noticed the MC1 refl was not on the camera, and in general other vertex SUS were misaligned even though JC had aligned the IFO in the morning to almost optimum arm cavity flashing. After a c1susaux modbusIOC service restart and burt restore, the problem persisted.

We started to debug the sus rack chain for PRM since the oplev beam was still near its alignment so we could use it as a sensor. The first weird thing we noticed was that no matter how much we "kicked" PRM, we wouldn't see any motion on the oplev. We repeatedly kicked UL coil and looked at the coil driver inputs and outputs, and also verified the eurocard had DC power on which it did. Somehow disconnecting the acromag inputs didn't affect the medm screen values, so that made us suspicious that something was weird with these ADCs.


Because all the slow channels were in a frozen state, we tried restarting c1susaux and the acromag chassis and this fixed the issue.

  16996   Wed Jul 13 10:54:39 2022 YehonathanUpdateBHDMICH AS55 noise budget

I fixed some mistakes in the budget:

1. The BS pendulum resonance was corrected from 0.8Hz to 1Hz

2. Added missing X3 filter in the coil filters

3. Optical gain is now computed from MICH to AS55 instead of BS to AS55 and is calculated to be: 9.95e8 cts/m.

4. Coil driver gain is still unmeasured but it is found to be 1.333 to make the actuation calibration from BS to MICH match the measurement (see attachment 1).

Attachment 2 shows the resulting MICH OLTF.

Laser noise was added to the budget in a slightly ad-hoc fashion (will fix later): Yuta and I measured MC_F and computed MC_F*(Schnupp asymmetry)/(Laser frequency). Attachment 3 shows the updated noise budget.

Attachment 1: BS_MICH_ACtuation_Calibration.pdf
BS_MICH_ACtuation_Calibration.pdf
Attachment 2: MICH_AS55_Model_Measurement_Comparison.pdf
MICH_AS55_Model_Measurement_Comparison.pdf
Attachment 3: MICH_AS55_Noise_Budget.pdf
MICH_AS55_Noise_Budget.pdf
  16995   Wed Jul 13 07:16:48 2022 JCUpdateElectronicsChecking Sorensen Power Supplies

[JC]

I went around 40m picking up any Sorensens that were laying around to test if they worked, or in need of repair. I gathered up a total of 7 Sorensens and each one with a Voltmeter. I made sure the voltage would rise on the Sorenson as well as the voltmeter, maxing out at ~33.4 Volts. For the current, the voltmeter can only rise to 10 Amps before it is fused. Many of the Sorensons that I found did not have their own wall connection, so I had to use the same one for multiple.

From these 7, I have found 5 that are well. One Sorenson I have tested has a output shortage above 20V and the other has yet to be tested.

Attachment 1: 658C5D39-11BD-4EE3-90E2-34CBBC1DBD3C.jpeg
658C5D39-11BD-4EE3-90E2-34CBBC1DBD3C.jpeg
Attachment 2: 5328312A-7918-44CC-82B7-54B57840A336.jpeg
5328312A-7918-44CC-82B7-54B57840A336.jpeg
  16994   Tue Jul 12 19:46:54 2022 PacoSummaryALSHow (not) to take NPRO PZT transfer function

[Paco, Deeksha, rana]

Quick elog for this evening:

  • Rana disabled MC servo .
  • Slow loop also got disengaged.
  • AUX PSL beatnote is best taken with *free running lasers* since their relative frequency fluctuations are lowest than when locked to cavities.
  • DFD may be better to get PZT transfer funcs, or get higher bandwidth phase meter.
  • Multi instrument to be done with updated moku
  • Deeksha will take care of updated moku
  16993   Tue Jul 12 18:35:31 2022 Cici HannaSummaryGeneralFinding Zeros/Poles With Vectfit

Am still working on using vectfit to find my zeros/poles of a transfer function - now have a more specific project in mind, which is to have a Red Pitaya use the zero/pole data of the transfer function to find the UGF, so we can check what the UGF is at any given time and plot it as a function of time to see if it drifts (hopefully it doesn't). Wrestled with vectfit more on matlab, found out I was converting from dB's incorrectly (should be 10^(dB/20)....) Intend to read a bit of a book by Bendat and Piersol to learn a bit more about how I should be weighting my vectfit. May also check out an algorithm called AAA for fitting instead.

  16992   Tue Jul 12 14:56:17 2022 TomislavSummaryElectronicsElectronics noise measurements

[Paco, Tomislav]

We measured the electronics noise of the demodulation board, whitening board, and ADC for WFSs, and OPLEV board and ADC for DC QPD in MC2 transmission. We were using SR785.

Regarding the demodulation board, we did 2 series of measurements. For the first series of measurements, we were blocking WFS (attachment 1) and measuring noise at the output of the demod board (attachment 2a). This measurement includes dark noise of the WFS, electronics noise of demod board, and phase noise from LO. For the second series of the measurements, we were unplugging input to the demod board (attachment 2b & 2c is how they looked like before unplugging) (the mistake we made here is not putting 50-ohm terminator) and again measuring at the output of the demod board. This measurement doesn't include the dark noise of the WFS. We were measuring it for all 8 segments (I1, I2, I3, I4, Q1, Q2, Q3, Q4). The dark noise contribution is negligible with respect to demod board noise. In attachments 3 & 4 please find plots that include detection and demodulation contributions for both WFSs.

For whitening board electronics noise measurement, we were terminating the inputs (attachment 5) and measuring the outputs (attachment 6). Electronics noise of the whitening board is in the attachments 7 & 8.

For ADC electronics noise we terminated ADC input and measured noise using diaggui (attachments 9 & 10). Please find these spectra for WFS1, WFS2, and MC TRANS in attachments 11, 12 & 13.

For MC2 TRANS we measured OPLEV board noise. We did two sets of measurements, as for demod board of WFSs (with and without QPD dark noise) (attachments 14, 15 & 16). In the case of OPLEV board noise without dark noise, we were terminating the OPLEV input. Please find the electronics noise of OPLEV's segment 1 (including dark noise which is again much smaller with respect to the OPLEV's electronics noise) in attachment 17.

For the transfer functions, demod board has flat tf, whitening board tf please find in attachment 18, ADC tf is flat and it is (2**16 - 1)/20 [cts/V], and dewhitening tf please find in attachment 19. Also please find the ASD of the spectral analyzer noise (attachment_20).

Measurements for WFS1 demod and whitening were done on 5th of July between 15h and 18h local time. Measurements for WFS2 demod and whitening were done on 6th of July between 15h and 17h local time. All the rest were done on July 7th between 14h and 19h. In attachment 21 also find the comparison between electronics noise for WFSs and cds error signal (taken on the 28th of June between 17h and 18h). Sorry for bad quality of some pictures.

Attachment 1: attachment_1.jpg
attachment_1.jpg
Attachment 2: attachment_2a.jpg
attachment_2a.jpg
Attachment 3: attachment_2b.jpg
attachment_2b.jpg
Attachment 4: attachment_2c.jpg
attachment_2c.jpg
Attachment 5: attachment_3.png
attachment_3.png
Attachment 6: attachment_4.png
attachment_4.png
Attachment 7: attachment_5.jpg
attachment_5.jpg
Attachment 8: attachment_6.jpg
attachment_6.jpg
Attachment 9: attachment_7.png
attachment_7.png
Attachment 10: attachment_8.png
attachment_8.png
Attachment 11: attachment_9.jpg
attachment_9.jpg
Attachment 12: attachment_10.jpg
attachment_10.jpg
Attachment 13: attachment_11.png
attachment_11.png
Attachment 14: attachment_12.png
attachment_12.png
Attachment 15: attachment_13.png
attachment_13.png
Attachment 16: attachment_14.jpg
attachment_14.jpg
Attachment 17: attachment_15.jpg
attachment_15.jpg
Attachment 18: attachment_16.jpg
attachment_16.jpg
Attachment 19: attachment_17.png
attachment_17.png
Attachment 20: attachment_18.png
attachment_18.png
Attachment 21: attachment_19.png
attachment_19.png
Attachment 22: attachment_20.png
attachment_20.png
Attachment 23: attachment_21.png
attachment_21.png
  16991   Tue Jul 12 13:59:12 2022 ranaSummaryComputersprocess monitoring: Monit

I've installed Monit on megatron and nodus just now, and will set it up to monitor some of our common processes. I'm hoping that it can give us a nice web view of what's running where in the Martian network.

  16990   Tue Jul 12 09:25:09 2022 ranaUpdateIOOIMC WFS

MC WFS Demod board needs some attention.

Tomislav has been measuring a very high noise level in the MC WFS demod output (which he promised to elog today!). I thought this was a bogus measurement, but when he, and Paco and I tried to measure the MC WFS sensing matrix, we noticed that there is no response in any WFS, although there are beams on the WFS heads. There is a large response in MC2 TRANS QPD, so we know that there is real motion.

I suspect that the demod board needs to be reset somehow. Maybe the PLL is unlocked or some cable is wonky. Hopefully not both demod boards are fried.

Please leave the WFS loops off until demod board has been assessed.

  16989   Tue Jul 12 09:14:50 2022 ranaUpdateBHDMICH AS55 noise budget

Looking good:

  • I think the notches you see in he measured noise are a clue as to the excess noise source. You can try turning some notches on/off.
  • Laser noise does matter a bit more subtley: the low freq noise couples to AS55 through the RMS deviation of the MICH loop from the zero crossing, and the noise of the 55 MHz modulation.
  • Jitter in the IMC couples to MICH through the misalignment of the Michelson.
  • As you rightly note, the optical lever feedback on the ITMs and BS also make length noise through the suspension actuator imbalance and the spot mis-centering.
  16988   Mon Jul 11 19:29:23 2022 PacoSummaryGeneralFinalizing recovery -- timing issues, cds, MC1

[Yuta, Koji, Paco]

Restarting CDS

We were having some trouble restarting all the models on the FEs. The error was the famous 0x4000 DC error, which has to do with time de-synchronization between fb1 and a given FE. We tried a combination of things haphazardly, such as reloading the gpstime process using

controls@fb1:~ 0$ sudo systemctl stop daqd_*
controls@fb1
:~ 0$ sudo modprobe -r gpstime
controls@fb1:~ 0$ sudo modprobe gpstime
controls@fb1:~ 0$ sudo systemctl start daqd_*
controls@fb1:~ 0$ sudo systemctl restart open-mx.service

without much success, even when doing this again after hard rebooting FE + IO chassis combinations around the lab. Koji prompted us to check the local times as reported by the gpstime module, and comparing it to network reported times we saw the expected offset of ~ 3.5 s. On a given FE ("c1***") and fb1 separately, we ran:

controls@c1***:~ 0$ timedatectl
  Local time: Mon 2022-07-11 16:22:39 PDT
  Universal time: Tue 2022-07-11 23:22:39 UTC
       Time zone: America/Los_Angeles (PDT, -0700)
       NTP enabled: yes
       NTP synchronized: no
 RTC in local TZ: no
       DST active: yes
 Last DST change: DST began at
                  Sun 2022-03-13 01:59:59 PST
                  Sun 2022-03-13 03:00:00 PDT
 Next DST change: DST ends (the clock jumps one hour backwards) at
                  Sun 2022-11-06 01:59:59 PDT
                  Sun 2022-11-06 01:00:00 PST
controls@fb1:~ 0$ ntpq -p
     remote           refid      st t when poll reach   delay   offset  jitter
==============================================================================
 192.168.123.255 .BCST.          16 u    -   64    0    0.000    0.000   0.000

which meant a couple of things:

  1. fb1 was serving its time (broadcast to local (martian) network)
  2. fb1 was not getting its time from the internet
  3. c1*** was not synchronized even though fb1 was serving the time

By looking at previous elogs with similar issues, we tried two things;

  1. First, from the FEs, run sudo systemctl restart systemd-timesyncd to get the FE in sync; this didn't immediately solve anything.
  2. Then, from fb1, we tried pinging google.com and failed! The fb1 was not connected to the internet!!!

We tried rebooting fb1 to see if it connected, but eventually what solved this was restarting the bind9 service on chiara! Now we could ping google, and saw this output

controls@fb1:~ 0$ ntpq -p
     remote           refid      st t when poll reach   delay   offset  jitter
==============================================================================
+tor.viarouge.ne 85.199.214.102   2 u  244 1024  377  144.478    0.761   0.566
*ntp.exact-time. .GPS.            1 u   93 1024  377  174.450   -1.741   0.613
 time.nullrouten .STEP.          16 u    - 1024    0    0.000    0.000   0.000
+ntp.as43588.net 129.6.15.28      2 u  39m 1024  314  189.152    4.244   0.733
 192.168.123.255 .BCST.          16 u    -   64    0    0.000    0.000   0.000 

meaning fb1 was getting its time served. Going back to the FEs, we still couldn't see the ntp synchronized flag up, but it just took time after a few minutes we saw the FEs in sync! This also meant that we could finally restart all FE models, which we successfully did following the script described in the wiki. Then we had to reload the modbusIOC service in all the slow machines (sometimes this required us to call sudo systemctl daemon-reload) and performed burt restore to a last Friday's snap file collection.


IMC realign and MC1 glitch?

With Koji's help PMC locked, and then Yuta and Paco manually increased the input power to the IFO by rotating the waveplate picomotor to 37.0 deg. After this, we noticed that the MC REFL spot was not hitting the camera, so maybe MC1 was misaligned. Paco checked the AP table and saw the spot horizontally misaligned on the camera, which gave us the initial YAW correction on MC1. After some IMC recovery, we saw only MC1 got spontaneously kicked along both PIT and YAW, making our alignment futile. Though not hard to recover, we wondered why this happened.

We went into the 1X4 rack and pushed MC1 suspension cables in to disregard loose connections, but as we came back into the control room we again saw it being kicked randomly! We even turned damping off for a little while and this random kicking didn't stop. There was no significant seismic motion at the time so it is still unclear of what is happening.

  16987   Mon Jul 11 17:41:52 2022 KojiHowToVACStartup after Power Outage

- Once the FRG gauge readings are back (see next elog by Tega), I could open V1 to pump down the main vacuum manifold.
- TP2/TP3 were brought back to stand-by mode (slower spinning)
- V7 was closed to separate the annuli side and TP1

During the vacuum recovery, I saw TPs were automatically turned on as soon as the backing pumps were engaged. I could not figure out what caused this automation.

Also, I saw some gate valve states changed while I was not touching them. e.g. V7 was close / VM3 was open / etc
I really had no idea what/who was handling these.

As of ~18:00 local, the main volume pressure is ~2e-5 torr and ready to open the PSL shutter.

Attachment 1: Screen_Shot_2022-07-11_at_18.13.00.png
Screen_Shot_2022-07-11_at_18.13.00.png
  16986   Mon Jul 11 17:25:43 2022 TegaUpdateVACfixed obsolete reference bug in serial_XGS600 service

Koji noticed that the FRG sensors were not updating due to reference to an obsolete modbusIOC_XGS service, which was used temporarily to test the operation of the serial XGS sensor readout to EPICS. The information in this service was later moved into modbusIOC.service but the dependence on the modbusIOC_XGS.service was not removed from the serial_XGS600.service. This did not present any issue before the shutdown, probably bcos the obsolete service was already loaded but after the restart of c1vac, the obsolete service file modbusIOC_XGS.service was no longer available. This resulted in  serial_XGS600.service throwing a failure to load error for the missing obsolete modbusIOC_XGS service. The fix involved replacing two references to 'modbusIOC_XGS' with 'modbusIOC' in  /opt/target/services/serial_XGS600.service.

I also noticed that the date logged in the commit message was Oct 2010 and that I could not do a push from c1vac due to an error in resolving git.ligo.org. I was able to push the commit from my laptop git repo but was unable to do a pull on c1vac to keep it synced with the remote repo.

  16985   Mon Jul 11 15:26:12 2022 JCHowToVACStartup after Power Outage

[Koji, Jc]

Koji and I began starting that Vacuum system up.

  1. Reverse order step 2 of shutting down electronics. Anthing after, turn on manually.
  2. If C1vac does not come back, then restart by holding the reset button.
  3. Open VA6
  4. Open VASE, VASV,VABSSCI, VABS, VABSSCO, VAEV, and VAEE
  5. Open V7
  6. Check P3 and P2, if they are at high pressure, approx. 1 Torr range, then you must use the roughing pumps.
  7. Connect Rotary pump tube. (Manually)
  8. Turn on AUX Pump
  9. Manually open TP2 and TP3 valves.
  10. Turn on TP2 and TP3, when the pumps finish startup, turn off Standby to bring to nominal speed.
  11. Turn on RP1 and RP3
  12. Open V6
  13. Once P3 reaches <<1 Torr, close V6 to isolate the Roughing pumps.
  14. When TP2 and TP3 are at nominal speed, open V5 and V4.
  15. Now TP1 is well backed, turn on TP1.
  16. When TP1 is at nominal speed, Open V1.
  16984   Mon Jul 11 11:56:40 2022 he YehonathanUpdateBHDMICH AS55 noise budget

I calculated a noise budget for the MICH using AS55 as a sensor. The calculation includes closed-loop TF calculations.

The notebook and associated files can be found on https://git.ligo.org/40m/bhd/-/blob/master/controls/compute_MICH_noisebudget.ipynb.

Attachment 1 shows the loop diagram I was using. The equation describing the steady-state of the loop is

\left[\mathbb{I}-G \right]\begin{pmatrix} \gamma \\ \delta \\ \Delta\end{pmatrix} = \begin{pmatrix} \alpha \\ \beta \\ \epsilon\end{pmatrix}

, where G is the adjacency matrix given by

G=\begin{pmatrix} 0 & 0 & AE_2\\ 0 & 0 & BE_2 \\ E_1C & E_1D & 0 \end{pmatrix}

First, the adjacency matrix G is constructed by stitching the small ABCDE matrices together. Once the inverse of (I-G) is calculated we can simply propagate any noise source to \delta and then calculate \left[\mathbb{I}-E(CA+DB)\right]B^{-1}\delta to estimate the displacement of the optics. 

Attachment 2 shows the calculated noise budget together with Yuta's measurement.

All the input and output electronics are clumped together for now. Laser noise is irrelevant as this is a heterodyne measurement at 55MHz.

It seems like there is some mismatch in the calibration of the optical gain between the measurement and model. The missing noise at 3-30Hz could be due to angle-to-length coupling which I haven't included in the model.

Attachment 1: Control_Diagram.pdf
Control_Diagram.pdf
Attachment 2: MICH_AS55_Noise_Budget.pdf
MICH_AS55_Noise_Budget.pdf
  16983   Mon Jul 11 11:16:45 2022 JCSummaryElectronicsStartup after Shutdown

[Paco, Yehonathan, JC]

We began starting up all the electronics this morning beginning in the Y-end. After following the steps on the Complete_Power_Shutdown_Procedures on the 40m wiki, we only came across 2 issues.

  1. The Green beam at the Y-End : Turn on the controller and the indicator light began flashing. After waiting until the blinking light becomes constant, turn on the beam. 
  2. C1lsc "could not find operating system"-unable to SSH from Rossa : We found an Elog of how to restart Chiara and this worked. We proceeded by adding this to the procedures of startup.
  16982   Fri Jul 8 23:10:04 2022 KojiSummaryGeneralJuly 9th, 2022 Power Outage Prep

The 40m team worked on the power outage preparation. The detailed is summarized on this wiki page. We will still be able to access the wiki page during the power outage as it is hosted some where in Downs.

https://wiki-40m.ligo.caltech.edu/Complete_power_shutdown_2022_07

  16981   Fri Jul 8 16:18:35 2022 ranaUpdateLSCActuator calibration of MC2 using Yarm

although I know that Yuta knows this, I will just put this here to be clear: the NNN/f^2 calibration is only accurate abouve the pendulum POS eiegenfrequency, so when we estimate the DC part (in diaggui, for example), we have to assume that we have a pendulum with f = 1 Hz and Q ~5, to get the value of DC gain to put into the diaggui Gain field in the calibration tab.

ELOG V3.1.3-