ID |
Date |
Author |
Type |
Category |
Subject |
12749
|
Tue Jan 24 07:36:56 2017 |
Max Isi | Update | Summary Pages | Cluster maintenance |
System-wide CIT LDAS cluster maintenance may cause disruptions to summary pages today. |
12748
|
Tue Jan 24 01:04:16 2017 |
gautam | Summary | IOO | IMC WFS RF power levels |
Summary:
I got around to doing this measurement today, using a minicircuits bi-directional coupler (ZFBDC20-61-HP-S+), along with some SMA-LEMO cables.
- With the IMC "well aligned" (MC transmission maximized, WFS control signals ~0), the RF power per quadrant into the Demod board is of the order of tens of pW up to a 100pW.
- With MC1 misaligned such that the MC transmission dropped by ~10%, the power per quadrant into the demod board is of the order of hundreds of pW.
- In both cases, the peak at 29.5MHz was well above the analyzer noise floor (>20dB for the smaller RF signals), which was all that was visible in the 1MHz span centered around 29.5 MHz (except for the side-lobes described later).
- There is anomalously large reflection from Quadrant 2 input to the Demod board for both WFS
- The LO levels are ~-12dBm, ~2dBm lower than the 10dBm that I gather is the recommended level from the AD831 datasheet
Quote: |
We should insert a bi-directional coupler (if we can find some LEMO to SMA converters) and find out how much actual RF is getting into the demod board.
|
Details:
I first aligned the mode cleaner, and offloaded the DC offsets from the WFS servos.
The bi-directional coupler has 4 ports: Input, Output, Coupled forward RF and Coupled Reverse RF. I connected the LEMO going to the input of the Demod board to the Input, and connected the output of the coupler to the Demod board (via some SMA-LEMO adaptor cables). The two (20dB) coupled ports were connected to the Agilent spectrum analyzer, which have input impedance 50ohms and hence should be impedance matched to the coupled outputs. I set the analyzer to span 1MHz (29-30MHz), IF BW 30Hz, 0dB input attenuation. It was not necessary to turn on averaging to resolve the peaks at ~29.5MHz since the IF bandwidth was fine enough.
I took two sets of measurements, one with the IMC well aligned (I maximized the MC Trans as best as I could to ~15,000 cts), and one with a macroscopic misalignment to MC1 such that the MC Trans fell to 90% of its usual value (~13,500 cts). The peak function on the analyzer was used to read off the peak height in dBm. I then converted this to RF power, which is summarized in the table below. I did not account for the main line loss of the coupler, but according to the datasheet, the maximum value is 0.25dB so there numbers should be accurate to ~10% (so I'm really quoting more S.Fs than I should be).
WFS |
Quadrant |
Pin (pW) |
Preflected(pW) |
Pin-demod board (pW) |
IMC well aligned
1 |
1 |
50.1 |
12.6 |
37.5 |
2 |
20.0 |
199.5 |
-179.6 |
3 |
28.2 |
10.0 |
18.2 |
4 |
70.8 |
5.0 |
65.8
|
2 |
5 |
100 |
19.6 |
80.0 |
6 |
56.2 |
158.5 |
-102.3 |
7 |
125.9 |
6.3 |
11.5 |
8 |
17.8 |
6.3 |
119.6
|
WFS |
Quadrant |
Pin (pW) |
Preflected(pW) |
Pin-demod board (pW) |
MC1 Misaligned
1 |
1 |
501.2 |
5.0 |
496.2 |
2 |
630.6 |
208.9 |
422 |
3 |
871.0 |
5.0 |
866 |
4 |
407.4 |
16.6 |
190.8
|
2 |
5 |
407.4 |
28.2 |
379.2 |
6 |
316.2 |
141.3 |
175.0 |
7 |
199.5 |
15.8 |
183.7 |
8 |
446.7 |
10.0 |
436.7 |
For the well aligned measurement, there was ~0.4mW incident on WFS1, and ~0.3mW incident on WFS2 (measured with Ophir power meter, filter out).
I am not sure how to interpret the numbers for quadrants #2 and #6 in the first table, where the reverse coupled RF power was greater than the forward coupled RF power. But this measurement was repeatable, and even in the second table, the reverse coupled power from these quadrants are more than 10x the other quadrants. The peaks were also well above (>10dBm) the analyzer noise floor
I haven't gone through the full misalginment -> Power coupled to TEM10 mode algebra to see if these numbers make sense, but assuming a photodetector responsivity of 0.8A/W, the product (P1P2) of the powers of the beating modes works out to ~tens of pW (for the IMC well aligned case), which seems reasonable as something like P1~10uW, P2 ~ 5uW would lead to P1P2~50pW. This discussion was based on me wrongly looking at numbers for the aLIGO WFS heads, and Koji pointed out that we have a much older generation here. I will try and find numbers for the version we have and update this discussion.
Misc:
- For the sake of completeness, the LO levels are ~ -12.1dBm for both WFS demod boards (reflected coupling was negligible)
- In the input signal coupled spectrum, there were side lobes (about 10dB lower than the central peak) at 29.44875 MHz and 29.52125 MHz (central peak at 29.485MHz) for all of the quadrants. These were not seen for the LO spectra.
- Attached is a plot of the OSEM sensor signals during the time I misaligned MC1 (in both pitch and yaw approximately by equal amounts). Assuming 2V/mm for the OSEM calibration, the approximate misalignment was by ~10urad in each direction.
- No IMC suspension glitching the whole time I was working today

|
Attachment 1: MC1_misalignment.png
|
|
12747
|
Mon Jan 23 17:24:26 2017 |
Steve | Update | Optical Levers | ETMY Oplev HeNe is running hot |
ETMY He/Ne 1103P body temp is ~45 C The laser was seated loosely in the V-mount with black rubber padding.
The enclosure has a stinky plastic smell from this black plastic. This laser was installed on Oct 5, 2016 See 1 year plot.
Oplev servo turned off. Thermocouple attached to the He/Ne
It will be replaced tomorrow morning.
Quote: |
On the control room monitors, I noticed that the IR TEM00 spot was moving around rather a lot in the Y arm. The last time this happened had something to do with the ETMY Oplev, so I took a look at the 30 day trend of the QPD sum, and saw that it was decaying steeply (Steve will update with a long term trend plot shortly). I noticed the RIN also seemed rather high, judging by how much the EPICS channel reading for the QPD sum was jumping around. Attached are the RIN spectra, taken with the OL spot well centered on the QPD and the arms locked to IR. Steve will swap the laser out if it is indeed the cluprit.
|
|
Attachment 1: ETMY_oplev__lasers.png
|
|
12746
|
Mon Jan 23 15:16:52 2017 |
gautam | Update | Optical Levers | ETMY Oplev HeNe needs to be replaced |
On the control room monitors, I noticed that the IR TEM00 spot was moving around rather a lot in the Y arm. The last time this happened had something to do with the ETMY Oplev, so I took a look at the 30 day trend of the QPD sum, and saw that it was decaying steeply (Steve will update with a long term trend plot shortly). I noticed the RIN also seemed rather high, judging by how much the EPICS channel reading for the QPD sum was jumping around. Attached are the RIN spectra, taken with the OL spot well centered on the QPD and the arms locked to IR. Steve will swap the laser out if it is indeed the cluprit. |
Attachment 1: ETMY_Oplev.pdf
|
|
12745
|
Mon Jan 23 10:24:01 2017 |
Steve | Update | SUS | MC1 SUS electronics investigation |
Two day plot of glitching suspentions: MC3, ITMY and ETMX |
Attachment 1: 3glitchingSUS.png
|
|
12744
|
Fri Jan 20 17:12:37 2017 |
Steve | Update | PEM | EM172 mic is hooked up in the PSL |
Gautam and Steve,
It is hanging in the midle of the PSL enclosure. Wired to 1X1 to get plus and minus 15V through fuse. It's output is connected to FB c17 input.
GV: C17 corresponds to "MIC 1" in the PEM model. So the output is saved as "C1:PEM-MIC_1_OUT_DQ"
Quote: |
I added an EM172 to my soldered circuit and it seems to be working so far. I have taken a spectra using the EM172 in ambient noise in the control room as well as in white noise from Audacity. My computer's speakers are not very good so the white noise results aren't great but this was mainly to confirm that the microphone is actually working.

white_v_ambient.pdf
|
|
Attachment 1: EM172c.jpg
|
|
12743
|
Fri Jan 20 14:42:12 2017 |
Steve | Update | PEM | Caltech weather station |
We should be able to connect to this station |
12742
|
Fri Jan 20 11:16:30 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
Both suspensions have been relatively well behaved for the best part of the last two days, since I effected the Satellite Box swap. Today morning, I set about re-enabling the damping and locking the MC. Judging by the wall StripTool, it stayed locked for about 30 mins or so, after which the glitching returned.
Attached is a screenshot of the sensor signals from MC1 and MC3 (second trend), and also the highest band (>30Hz) BLRMS output for the same 10 channels (full data sampled at 16Hz). Note that MC1 and MC3 satellite boxes remain swapped. So the glitches now have migrated to the MC3 channels.

I need to think about whether this is just coincidence, or if me re-enabling the damping has something to do with the re-occurrence of the glitching...
Addendum 4.30pm: I've also re-aligned the Y arm. Its alignment has been stable over the last few hours, despite several mode cleaner lock losses in between, it recovers good IR transmission. The X arm has been re-aligned to green, but I can't get it locked to the IR - everytime I turn the LSC to ETMX on, there seems to be some large misalignment applied to it. c1iscaux was dead, I restarted it by keying the crate. I haven't had time to investigate the X arm locking in detail, I will continue to debug this. |
12741
|
Thu Jan 19 19:56:09 2017 |
rana | Update | SUS | MC1 SUS electronics investigation |
Might be. Or it might be in the satellite box cabling. Hard to tell without a tester. I recommend you squish the cables on there and lock the MC and get back to the usual business. I'll check on sat. box with Ben.
Quote: |
Is this sufficient evidence to conclude that the Satellite boxes are to blame? It's hard to explain why the glitches come and go in this fashion, and also the apparent difference in the length of time for which the glitches persist. Here, in almost 24 hours, there is one incidence of glitching, but in yesterday's trend plot, the glitching remains present over several hours... The amplitude of the glitches, and their coincidence in all 5 channels, seems consistent with what we have been seeing though...
|
|
12740
|
Thu Jan 19 16:36:35 2017 |
ericq | Update | Computer Scripts / Programs | nodus web apache simlinks too soft |
This was done by adding "Options +Indexes" to /etc/apache/sites-available/nodus
I've added a little more info about the apache configuration on the wiki: ApacheOnNodus |
12739
|
Thu Jan 19 12:00:10 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
Going through the last ~20 hours of data, the MC1 sensor channels look glitch free the entire period. However, there is a ~10min period around 1PM UTC today when there were a couple of glitches ~80 counts in size in all the MC3 sensor channels. The attached shows the full 2k data from all 10 channels (MC1 and MC3 sensors) around this time.

Is this sufficient evidence to conclude that the Satellite boxes are to blame? It's hard to explain why the glitches come and go in this fashion, and also the apparent difference in the length of time for which the glitches persist. Here, in almost 24 hours, there is one incidence of glitching, but in yesterday's trend plot, the glitching remains present over several hours... The amplitude of the glitches, and their coincidence in all 5 channels, seems consistent with what we have been seeing though...
|
12738
|
Thu Jan 19 10:21:54 2017 |
Ashley | Update | General | Preliminary Microphone Data |
Brief Summary: I am currently looking at the acoustic noise around both arms to see if there are any frequencies from machinery around the lab that stand out and to see what we can remove/change. I am using a Bluebird microphone suspended with surgical tubing from the cable trays to isolate it from vibrations. I am also using a preamp and the SR875 spectrum analyzer taking 6 sets of data every 1.5 meters (0 to 200Hz, 200Hz to 400Hz, 400z to 800Hz, 800Hz to 3200Hz, 3.2kHz to 12kHz, 12kHz to 100kHz).
· Attachment 1 is a PSD of the first 3 measurements (from 0 to 12kHz) that I took every 1.5 meters along the x arm with the preamp and spectrum analyzer
· Attachment 2 is a blrms color map of the first 6 sets of data I took (from 2.4m to 9.9m)
· Attachmetn 3 is a picture of the microphone set up with the surgical tubing
Problems that occurred: settings on the preamp made the first set of data I took significantly smaller than the data I took with the 0dB button off and the last problem I had was the spectrum analyzer reading only from -50 to -50 dBVpk
|
Attachment 1: xend_psd.png
|
|
Attachment 2: xblrms.png
|
|
Attachment 3: IMG_3734.JPG
|
|
12737
|
Thu Jan 19 08:25:12 2017 |
Steve | Update | SUS | MC1 SUS electronics investigation |
Quote: |
Quote: |
As part of the ongoing debugging, I've switched the MC1 and MC3 satellite boxes. Both MC1 and MC3 have their watchdogs shudown for the moment.
|
In the last 3.5 hours, there has been nothing conclusive - no evidence of any glitching in either MC1 or MC3 sensor channels. I am going to hold off on doing the LEMO cable swap test for a few more hours, to see if we can rule out the satellite box.

|
No change. |
Attachment 1: MC1_MC3_ITMY_ETMX_sensors.png
|
|
Attachment 2: sensors_UL.png
|
|
12736
|
Wed Jan 18 18:44:53 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
Quote: |
As part of the ongoing debugging, I've switched the MC1 and MC3 satellite boxes. Both MC1 and MC3 have their watchdogs shudown for the moment.
|
In the last 3.5 hours, there has been nothing conclusive - no evidence of any glitching in either MC1 or MC3 sensor channels. I am going to hold off on doing the LEMO cable swap test for a few more hours, to see if we can rule out the satellite box.

|
12735
|
Wed Jan 18 15:17:38 2017 |
rana | Update | Computer Scripts / Programs | nodus web apache simlinks too soft |
I suppose before directory listings were turned off we should have fixed the script to make an index.html, but that's how it goes with "up"-grades. How about re-allow directory listing so that our old links for webview work again?
EQ: https://nodus.ligo.caltech.edu:30889/FE is live |
12734
|
Wed Jan 18 14:23:47 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
As part of the ongoing debugging, I've switched the MC1 and MC3 satellite boxes. Both MC1 and MC3 have their watchdogs shudown for the moment. |
12733
|
Wed Jan 18 12:46:47 2017 |
ericq | Update | Computer Scripts / Programs | nodus web apache simlinks too soft |
Quote: |
I tried to follow these instructions today to make the Simulink Webview accessible:
controls@nodus|public_html > ln -sfn /users/public_html/FE /export/home/
But...I got a "403 Forbidden" message. What is the secret handshake to get this to work? And why have we added this extra step of security?
|
This link works for me: https://nodus.ligo.caltech.edu:30889/FE/c1als_slwebview.html. The problem with just /FE/ is that there is no index.html, and we have turned off automatic directory listings.
IIRC, this arrangement was due to the fact that authentication of some of the folders (maybe the wikis) was broken during the nodus upgrade, so there was sensitive information being publicly displayed. This setup gives us discretion over what gets exposed. |
12732
|
Wed Jan 18 12:34:21 2017 |
ericq | Summary | IOO | MCL / MCF / Calibration |
Quote: |
In the filter banks there were various version of a 'dewhite' filter. They were all approximately z=150, p=15, g =1 @ DC, but with ~1% differences. I don't trust their provenance and so I've enforced symmetry and fixed their names to reflect what they are (150:15).
|
The filters were made in response to a measurement of the pentek whitening boards in 2015 (ELOG 11550), but this level of accuracy probably isn't important. |
12731
|
Wed Jan 18 11:40:54 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
After the repair of the faulty LEMO cable, I left MC1 with it's watchdog off overnight. Unfortunately, it looks like the problem still persists. The first attachment shows a second trend plot for the past 15 hours. Towards the left end of the plot, you can see where I re-connected the LEMO cable for the LR/UR channels.

A couple of months ago, I added a BLRMS block for the IMC optics that calculates BLRMS for the shadow sensor output as well as the coil output. Looking at this trend overnight, I noticed that the glitches appear in the coil outputs as well, as shown in the plot below, which is for a 1 hour stretch last night (I used the full data from a 16Hz coil output channel and not the BLRMS, I am not sure if there is a DQ'ed version of the coil outputs?).

Zooming in further to one of these glitches, we can see that the glitches in the coil and shadow sensor signals are in fact coincident.

But given that the watchdog was turned off all this time, the only voltage going to the coils should be the DC bias voltages. So does this not support the hypothesis that the problem lies in the part of the signal chain that supplies the bias voltage to the coils?
Never mind, the "coil output" channel isn't a true readback of the voltage to the coil, but is the calculated damping output (which is not sent to the coils when the watchdog is shutdown... |
12730
|
Wed Jan 18 10:41:14 2017 |
gautam | Update | General | ETMX suspension electronics problems? |
Summary pages show no kicking in the ETMX watchdogs from midnight to 6 AM (0800 - 1400 UTC):
https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20170118/sus/watchdogs/ |
12729
|
Tue Jan 17 21:31:57 2017 |
gautam | Update | General | ETMX suspension electronics problems? |
Last night, I plugged the ETMX suspension coils back into the satellite box. Tonight, we turned on the damping loops for ETMX. Rana centered the Oplev so we can use that as an additional diagnostic to see if the optic gets kicked around overnight. We will re-assess the situation tomorrow.
Sometime earlier today, Lydia noticed that the +/- 5V Sorensens at the X end were not displaying their nominal voltage/current values (as per the stickers on them). She corrected this. |
12728
|
Tue Jan 17 21:29:52 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
Quote: |
After some investigation, Rana found that on the AA filter board end, one of the 4pin LEMOs from the whitening board had one of its wires come unstuck from where it was soldered (this presumably happened while we were squishing cables tonight, as the LR channel was fine before that). Also, there was no heat shrink used on any of the solder joints.
|
The faulty cable has been re-soldered (with heat shrink) and replaced. All 5 sensor signals appear normal on dataviewer now. I am leaving things in this state for the night, let us see if the glitches return overnight.
PSL shutter remains closed |
12727
|
Tue Jan 17 20:47:23 2017 |
rana | Update | CDS | Simulink Webview updated |
Seems like this stops working every ~2 years. Its been busted since early 2016 according to cron, so I fixed up the paths and restored some missing files and committed things to the SVN (with comments!) and now its working and grabbing the Web viewable versions of the front end models. Just need to restore its viewability and then the world can watch our models any time.
Quote: |
Back in 2011, JoeB wrote some entries on how to automatically update the Simulink webview stuff.
Somehow, the cron broke down over the years. I reran the matlab file by hand today and it worked fine, so now you can see the up to date models using the internet.
https://nodus.ligo.caltech.edu:30889/FE/
|
|
12726
|
Tue Jan 17 20:39:30 2017 |
rana | Update | Computer Scripts / Programs | nodus web apache simlinks too soft |
I tried to follow these instructions today to make the Simulink Webview accessible:
controls@nodus|public_html > ln -sfn /users/public_html/FE /export/home/
Quote: |
The story is: we currently don't expose the whole /users/public_html folder. Instead, we are symlinking the folders from public_html to /export/home/ on nodus, which is where apache looks for things
So, I fixed the links on the Core Optics page by running:
controls@nodus|~ > ln -sfn /users/public_html/40m_phasemap /export/home/
|
But...I got a "403 Forbidden" message. What is the secret handshake to get this to work? And why have we added this extra step of security? |
12725
|
Mon Jan 16 23:25:07 2017 |
gautam | Update | SUS | MC1 SUS electronics investigation |
[rana,gautam]
Summary:
- MC1 glitchy behaviour is back
- Found a broken LEMO cable, left unplugged for the night -> to be repaired tomorrow
- Further diagnosis to follow
During the course of Rana's inspection of the general state of the IFO, he commented that there seemed to be several seismic-related IMC lock losses in the time that he had been observing it. This issue looked suspiciously like the the MC1 glitches I had noticed sometime late last year, especially since each time the IMC would unlock, we could see significant amounts of motion on MC REFL. To diagnose, we did the following:
- Closed PSL shutter
- Ramped down the gains of the MC1 damping loops by factor of 1000 in ~4 secs using z step
- Shut down the watchdog for MC1
- Observed dataviewer traces for glitches
Sure enough, there were several glitches that occurred in all 5 sensor channels. These glitches varied in size from a few counts (the smaller ones) to 60-70counts for the bigger ones. In the past, squishing the LEMO connector on the front of the PD whitening board (D000210) had apparently made the glitching go away. So tonight, for starters, we squished everything else - Sat. Box connectors, the breakout board from Sat. Box to whitening board on the back of 1X6, and the DB connector on the front of the whitening board. This had no effect - the glitching remained consistent.
Next, Rana pulled out two of the three 4pin LEMOs, and left only those coresponding to UL/LL plugged in - but the glitching persisted in these two channels. We then pulled out the board. It was installed in 1998, but has a sticker on it that says "fixed in 2003". Not sure what the fix was. Visual inspection of the circuit didn't show anything obviously faulty, but it did look like the two MAX333A quad switches (these control whether the whitening is bypassed or not) had been replaced at some point. There are other undesirable features, such as the use of thick film resistors, but nothing that would explain the glitchy behaviour.
Next, we re-inserted the whitening board back into its original slot in the Eurocrate, but switched the cables (both D sub and LEMO, but only on the whitening board end) between the boards for MC1 and MC3 (i.e. MC1 cables were routed through the whitening board that was originally used for MC3, and vice-versa). But the glitches remained consistent on the MC1 channels. So it looks like the board is not a likely culprit.
Finally, we went in and squished all the cables from the PD whitening board to the ADC (via an AA filter board). For some of the LEMO cables from the whitening board, the LEMO backshells were not properly tightened. Rana fixed these before putting them back in. Some of the connectors were also not pushed in tightly enough, Rana heard the click when he pushed them in. The cables from the adaptor board to the ADC itself looked fine, it was screwed on at both ends, and all these connections looked snug enough. In the interest of completeness, Rana also pushed in the backplane connectors on the Eurocrate (these supply the signals from the BIO cards to switch the whitening ON/OFF). The one corresponding to MC1 was indeed a little loose.
Coming back to the control room, we saw that the MC1 LR sensor was dead. After some investigation, Rana found that on the AA filter board end, one of the 4pin LEMOs from the whitening board had one of its wires come unstuck from where it was soldered (this presumably happened while we were squishing cables tonight, as the LR channel was fine before that). Also, there was no heat shrink used on any of the solder joints. Could this explain the glitchy behaviour? Perhaps, but the glitches remained in the 3 channels that were connected. Anyways, I will repair this cable tomorrow, and we can see if this has fixed the problem or not..
Some misc points:
- Regarding the adaptor boards that take the PD signals from the satellite box and route it to the whitening board, there are some clamps that hold the IDE connectors in place for MC1, MC2 and MC3 boards, but not for the others (see attached picture). Steve, can we install clamps for all of the boards? [taken care of, see here]
- The whitening boards are not screwed in place into the Eurocrate. This should be rectified.
PSL shutter is closed, MC1 watchdog is shutdown for the night. |
Attachment 1: 20170116_231625.png
|
|
Attachment 2: IMG_7175.JPG
|
|
Attachment 3: IMG_7174.JPG
|
|
12724
|
Mon Jan 16 22:03:30 2017 |
jamie | Configuration | Computers | Megatron update |
Quote: |
We should consider upgrading a few of our workstations to Ubuntu 14 LTS to see how painful it is to run our scripts and DTT and DV. Better to upgrade a bit before we are forced to by circumstance.
|
I would recommend upgrading the workstations to one of the reference operating systems, either SL7 or Debian squeeze, since that's what the sites are moving towards. If you do that you can just install all the control room software from the supported repos, and not worry about having to compile things from source anymore. |
12723
|
Mon Jan 16 21:03:47 2017 |
rana | Summary | IOO | MCL / MCF / Calibration |
Oot on the streets and in the chat rooms, people often ask, "What is up with the MC_F calibration?".
Not being sure of the wiring in the c1ioo model, I have formed this screencap of today's model and put it here. The MC_LENGTH and MC_FREQ are the filter banks which would calibrate these channels. In the filter banks there were various version of a 'dewhite' filter. They were all approximately z=150, p=15, g =1 @ DC, but with ~1% differences. I don't trust their provenance and so I've enforced symmetry and fixed their names to reflect what they are (150:15). I have also turned on one filter in MC_FREQ so that now the whitening of the Pentek Interface board is compensated.
Why is this TF 1/f? It should be -20 dB/decade if MC_F is in units of Hz* and MCL is a pendulum response. Perhaps its because the combination of the Koji summing box, the Thorlabs HV driver, and the Pomona box forms an additional 1/f ? IF so, this would explain the TF we see. Once we get confirmation from Koji, we can load the TF into the MC_FREQ filter bank and then MC_F will be in units of Hz (as will the summary pages).
(along the way I've also turned off the craaaazzzy servo input enable tickling that gets put in the MC AutoLocker every April Fool's leap year - resist the temptation)
Since we have a frequency counter system here and some oscillators, I wonder if we can just calibrate the MC_L and MC_F directly using a mixer lashed up to one of the counters. If so, and we can get the stabilized laser frequency noise down below 10 mHz/rHz, maybe this is a viable alternative method to the photon calibrators. Counting zero crossings is more honest than counting photons. |
Attachment 1: c1ioo_zoom_MCLF.png
|
|
Attachment 2: MCL.pdf
|
|
12722
|
Mon Jan 16 18:54:01 2017 |
rana | Update | SUS | BS: whitening re-engaged |
Found that the BS whitening was off. Gautam says that "it has always been that way" and "there's nothing in the elog about this" and "I have no special relationship with Putin".
I looked at DV and DTT while turning the OSEM whitening back on. As expected, the sensor noise improved by 10x above 10 Hz. The time series shows no problems - its just less fuzzy now.
All OSEM spectra after the switch show on upper panel of plot. Lower panel shows comparison of BS UL before/after. To rotate the DTT PDF landscape output I typed this:
pdftk BS-white.pdf cat 1N output BSwhite.pdf
"if you see something, do something" |
Attachment 1: BSwhite.pdf
|
|
12721
|
Mon Jan 16 12:49:06 2017 |
rana | Configuration | Computers | Megatron update |
The "apt-get update" was failing on some machines because it couldn't find the 'Debian squeeze' repos, so I made some changes so that Megatron could be upgraded.
I think Jamie set this up for us a long time ago, but now the LSC has stopped supporting these versions of the software. We're running Ubuntu12 and 'squeeze' is meant to support Ubuntu10. Ubuntu12 (which is what LLO is running) corresponds to 'Debian-wheezy' and Ubuntu14 to 'Debian-Jessie' and Ubuntu16 to 'debian-stretch'.
We should consider upgrading a few of our workstations to Ubuntu 14 LTS to see how painful it is to run our scripts and DTT and DV. Better to upgrade a bit before we are forced to by circumstance.
I followed the instructions from software.ligo.org (https://wiki.ligo.org/DASWG/DebianWheezy) and put the recommended lines into the /etc/apt/sources.list.d/lsc-debian.list file.
but I still got 1 error (previously there were ~7 errors):
W: Failed to fetch http://software.ligo.org/lscsoft/debian/dists/wheezy/Release Unable to find expected entry 'contrib/binary-i386/Packages' in Release file (Wrong sources.list entry or malformed file)
Restarting now to see if things work. If its OK, we ought to change our squeeze lines into wheezy for all workstations so that our LSC software can be upgraded. |
12720
|
Sat Jan 14 22:39:30 2017 |
rana | Summary | SUS | ITMY is drifting ? |
https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20170114/sus/susdrift/
ITMY is not like the others. Real or just OSEM madness? |
12719
|
Sat Jan 14 12:36:57 2017 |
ericq | Update | DAQ | minute trends missing |
Yes, writing minute trends causes hourly FB crashes in the current state of things. The "raw" minute trending is turned on, but I think that these are unknown to nds. |
12718
|
Sat Jan 14 12:12:03 2017 |
rana | Update | DAQ | minute trends missing |
Did we turn off minute trend writing in one of the recent FrameBuilder debug sessions? Seems we only have second trends in 2016. Maybe this explains why its so slow to get minute trends? Dataviewer has to rebuild it from second trend.
controls@nodus|frames > l
total 64
drwx------ 2 root root 16384 Jun 8 2009 lost+found/
drwxr-xr-x 2 controls controls 4096 Jul 14 2015 tmp/
-rw-r--r-- 1 controls controls 0 Jul 14 2015 test-file
drwxr-xr-x 5 controls controls 4096 Apr 7 2016 trend/
drwxr-xr-x 4 root root 4096 Apr 11 2016 archive/
drwxr-xr-x 789 controls controls 36864 Jan 13 19:34 full/
controls@nodus|frames > cd trend
controls@nodus|trend > l
total 3340
drwxr-xr-x 258 controls controls 3342336 Jul 6 2015 minute_raw/
drwxr-xr-x 387 controls controls 36864 Nov 5 2015 minute/
drwxr-xr-x 969 controls controls 36864 Jan 13 19:49 second/
|
12717
|
Sat Jan 14 00:53:05 2017 |
rana | HowTo | DAQ | Get 40m data using NDS2 and Python |
Minute trend data seems not available using the NDS2 server. Its super slow using dataviewer from the control room.
Did some digging into the NDS2 config on megatron. It hasn't been updated in 2 years.
All of the stuff is run by the user 'nds2mgr'. The CronTab for this user was running all the channel name updates and server restarts at 3 AM each day; I've moved it to 5:05 AM. I don't know the password for this user, so I just did 'sudo su nds2mgr' to become him.
On megatron, in /home/nds2mgr/nds2-megatron/ there is a list of channels and configs. The file for the minute trend (C-M-ChanList.txt), hasn't been updated since Nov-2015. ??? |
12716
|
Fri Jan 13 23:39:46 2017 |
gautam | Update | General | ETMX suspension electronics problems? |
[Koji,gautam]
After Koji's leap second fix, we were playing around with the X arm locking. In particular, we were playing around with the limit value on the X arm LSC filter bank - the nominal value is 4000, we wanted to see if we could increase this without kicking the optic while acquiring arm lock. We initially increased it to 8000, and then turned it off altogether. Then we rapidly turned the output of the servo ON/OFF, and looked at the arm transmission to see if it came back to the level before unlocking, as an indication of whether the optic was kicked.
These trials suggested a value of 8000 for the limiter was OK, so we left the LSC mode on with the limiter set to 8000. But just as we were about to leave for the night, I noticed on the wall Striptool that the X arm was unlocked. Investigating, we found that the green wasn't even locking to a HOM. Further investigation of the Oplev spot showed that ETMX had received a large kick (both pitch and law errors were ~200urad). ITMX was unaffected.
We initially tried lowering the LSC limit value back to 4000, then used first the Oplev spot and then the green to align the arm. But turning on LSC misaligned the arm after acquiring lock. So we decided to leave LSC off, thinking that the notorious ETMX suspension problems have resurfaced. As a diagnostic, we figured we'd leave the watchdog tripped, and use the Oplev to see if the optic was getting kicked. But the act of turning the watchdog off kicked the optic again (WHY?!).
Looking at the ETMX sus screen, turning off all the damping and LSC (but watchdog on) still leaves a non-zero offset in the "Vmon" field, between 0.02-0.05V depending on the coil. Turning the watchdog OFF takes all these to 0.009V, although I can see the LR value fluctuating between 0.004V and 0.009V. I went to the Xend and squished all the cables on the Sat. Box, but the problem persisted.
At this time, I can't think of any explanation, so I am giving up for the night. To avoid unnecessarily kicking the optic, I am going to unplug the suspension from the Sat. Box and leave one of our tester boxes plugged in, lets see if that sheds any light on the situation...
Notes:
- The +/-20V sorensens at this end were "tripped" for a few days after the power glitch until they were reset and turned back on yesterday. But this should not affect Vmon, as these Sorensens only supply the DC voltage for the coil bias, which is a slow machine channel?
- The X arm was staying locked and well aligned for hours on end earlier this afternon - in fact it was locked for about 2 hours 6-8 hours ago, I can still see the trace on the wall StripTool....
|
12715
|
Fri Jan 13 21:41:23 2017 |
Koji | Update | CDS | DC errors |
I think I fixed the DC error issue
1. I added the leap second (leapsecond ?) entry for 2016/12/31, 23:60:00 UTC to daqdrc
[OLD]
set gps_leaps = 820108813 914803214 1119744016;
[NEW]
set gps_leaps = 820108813 914803214 1119744016 1167264018;
2. Restarted FB and all realtime models
Now I don't see any RED light. |
12714
|
Fri Jan 13 21:32:49 2017 |
rana | HowTo | DAQ | Get 40m data using NDS2 and Python |
The attached file is a python notebook that you can use to get data. Minimal syntax. |
Attachment 1: get40mData.ipynb
|
{
"cells": [
{
"cell_type": "markdown",
"metadata": {},
"source": [
"## Get some 40m data using NDS"
]
},
{
... 137 more lines ...
|
12713
|
Fri Jan 13 14:33:00 2017 |
MAX (not Rana) | Update | Summary Pages | December outage |
PEM config file was also lacking a section named "summary", which is necessary for all parent tabs; this has now been solved. I have deactivated the MEDM pages because Praful's screencap script seemed to be broken (I should have logged this, I apologize).
Quote: |
Pages still not working: PEM and MEDM blank.
- Committed existing MEDM grabbing scripts to SVN. Ran the cron job on megatron by hand. It grabs PNG files, but somehow its not getting into the summary pages.
- Changed the MEDM grabbing scripts to use '/usr/bin/env'.
- GW summary log files were numbering in the many thousands, so I moved everything over 320 days old into the OLD/ sub-directory using 'find . -type f -mtime +320 -exec mv {} OLD/ \;' (the semi-colon is needed)
- Did apt-get upgrade on Megatron.
- pinged Max
- Stared at GWsumm docs to see if there's a clue about what (if anything) is wrong with the .ini file.
|
|
12712
|
Fri Jan 13 14:18:28 2017 |
Steve | Update | PEM | air condition fixed |
The old control room AC has been stick in heating mode for about 2 months. It's thermostate and fan belt was finally replaced. It was calibrated and set to 71 F ( just behind 1X6 on west wall ) around 1pm.
Out belt; sad inside
at 4 pm Rana cried
It must be too tight. |
Attachment 1: PEM_120d.png
|
|
12711
|
Fri Jan 13 10:53:03 2017 |
Steve | Update | PEM | doors are fixed |
Control room to outside door was realigned.
It is self closing now.
Control room to IFO door lock optimized to soft closing.
All other doors lubricated by Alex of the key shop. |
12710
|
Fri Jan 13 08:54:32 2017 |
Johannes | Update | General | DC PD installed |
I installed a DC PD (Thorlabs PDA 520) in the beam path to AS55. I placed a 2" 90/10 BS on a flip mount that picks of enough light for the PD to spit out ~8V when the port is bright. Single arm continuous signal will be ~2V. While most of the light still continues towards AS55, the displacement from the BS moves the beam off AS55, so I used the flip mount in case anyone needs to use AS55. The current configuration is UP.
When we're done with loss investigations the flip mount should be removed from the bench.
 
I hooked the PD up to an ethernet-enabled scope and started scripting the loss map measurement (scope can receive commands via http so we can automate the data acquisition). The scope that was present at the bench and had been used for the MC ringdown measurements had a 'scrambled' screen that I couldn't fix so I had to retrieve another scope ("scope1"). I'll try to find out what's wrong with it but we may have to send it in for repair.

|
12709
|
Thu Jan 12 23:22:34 2017 |
rana | Update | Summary Pages | December outage |
Pages still not working: PEM and MEDM blank.
- Committed existing MEDM grabbing scripts to SVN. Ran the cron job on megatron by hand. It grabs PNG files, but somehow its not getting into the summary pages.
- Changed the MEDM grabbing scripts to use '/usr/bin/env'.
- GW summary log files were numbering in the many thousands, so I moved everything over 320 days old into the OLD/ sub-directory using 'find . -type f -mtime +320 -exec mv {} OLD/ \;' (the semi-colon is needed)
- Did apt-get upgrade on Megatron.
- pinged Max
- Stared at GWsumm docs to see if there's a clue about what (if anything) is wrong with the .ini file.
|
12708
|
Thu Jan 12 17:31:51 2017 |
gautam | Update | CDS | DC errors |
The IFO is more or less back to an operational state. Some details:
- The IMC mirror excess motion alluded to in the previous elog was due to some timing issues on c1sus. The "DAC" and "DK" blocks in the c1x02 diag word were red instead of green. Restarting all the models on c1sus fixed the problem
- When c1ioo was restarted, all of Koji's changes (digital) to the MC WFS servo where lost as they were not committed to the SDF. Eric suggested that I could just restore them from burt snapshots, which is what I did. I used the c1iooepics.snap file from 12:19PM PST on 26 December 2016, which was a time when the WFS servo was working well as per this elog by Koji. I have also committed all the changes to the SDF. IMC alignment has been stable for the last 4 hours.
- Johannes aligned and locked the arms today. There was a large DC offset on POX11, which was zeroed out by closing the PSL shutter and running LSC offsets. Both arms lock and stay aligned now.
- The doubling oven controller at the Y end was switched off. Johannes turned it on.
- Eric and I started a data consistency check on the RAID array yesterday, it has completed today and indicated no issues
- NDS2 is now running again on megatron so channel access from outside should(???) be possible again.
One error persists - the "DC" indicator (data concentrator?) on the CDS medm screen for the various models spontaneously go red and return to green often. Is this a known issue with an easy fix?

|
12707
|
Thu Jan 12 13:45:58 2017 |
Steve | HowTo | safety | closing a door |
It was requested this morning.
Quote: |
This is one of those unsolved door lock acquisition problems. Its been happening for years.
Please ask facilities to increase the strength of the door tensioner so that it closes with more force.
|
|
12706
|
Thu Jan 12 13:43:02 2017 |
rana | HowTo | safety | closing a door |
This is one of those unsolved door lock acquisition problems. Its been happening for years.
Please ask facilities to increase the strength of the door tensioner so that it closes with more force. |
12705
|
Thu Jan 12 10:14:49 2017 |
Steve | HowTo | safety | closing a door |
The door was not locked this morning.
Please do not use this door if you can not close it!
Last person leaving the lab should check that the latch is cut by the strike plate. |
Attachment 1: odc.jpg
|
|
Attachment 2: howtocloseadoor.jpg
|
|
12704
|
Thu Jan 12 02:45:53 2017 |
Johannes | Update | General | Next armloss steps |
As stated in elog 12618, using an oscilloscope to average the reflected powers and thus circumventing all filtering yielded much better results than before:
XARM: 21 +/- 35 ppm
YARM: 69 +/- 45 ppm
We can probably decrease the measurement uncertainty further by using a larger photodiode that is more suited for DC measurements. It will be placed in the AS pathtemporarily. If we get below 10 ppm systematic errors will begin to matter. To get those under control I will have to re-determine the visibility in the arm cavities and the modulation indices. The numbers to match from an estimate via the power recycing gain are <= 50 ppm arm average from elog 12586. Once the measurement scheme is up and running, we can proceed to generate ETM lossmaps. ITM will still be tricky but let's see what we can do.
Following Yutaro's approach, we can move the beams on the optcs in a deterministic way by several mm on the ETMs. Moving the beam is achieved by introducing offsets into the ASS auto alignment. As an example, the Yaw dither for ETMY is shown:

Each of the 8 test mass rotational degrees of freedom is driven by a particular frequency, and 2 signals are digitally demodulated in the real-time system: The arm transmission ("T") and the LSC arm length feedback signal to the ETM (L). The T signal feeds back to the input pointing, aka Tip Tilts and BS. This maximizes the transmission for a given test mass orientation. The L feedback controls the beam position on the mirrors in the arms. It minimizes the coupling of the dither to the length feedback, which is achieved when the beam goes through the axis of the rotational motion. This is where we introduce the offset:
The signal C1:ASS-YARM_ETM_YAW_L_DEMOD_I_OFFSET (for this example) moves the locking point of the dither-to-length coupling and thus moves the beam around on the ETM. This is true for the PIT and YAW of all test masses except ITMX. In the current configuration the TTs optimize the alignment into the YARM, and for the X we only have the BS, which is why the beam spot on ITMX cannot be independently controlled as-is. We could, however, for the sake of this measurement, temporarily temporarily give TT authority to the XARM feedback to control the ITMX beam position. I imagine something like dither-aligning with ASS the normal way, and then run a customized script in which the XARM is treated as the YARM, feecback to the BS is cut, and the YAW signals are inverted due to the reflection on BS.
Knowing the angle of the offset gives us a way to calculate the beam spot displacement with the cavity geometry. For best results I want to make sure our OpLev calibration is still good (laser power decay, although last time this was done was only about a year ago), which would be analogous to elog 11831.
As for ITM beam position, this scheme only works partially, because it would require the beam to steer further off its axis than in the ETM case. This is problematic because of the spacing between tip tilts and ITMs. I summarize:
- Place larger DCPD in AS path
- Confirm mode-matching and mod-indices
- Assess loss in center with zero offsets
- Uncertainty low enough? If not get better.
- Calibrate OpLevs
- Introduce calibrated offsets in dither alignment
- Wander beam on test masses, recording arm losses
- ???
- Profit
|
Attachment 1: ass_illustration.pdf
|
|
12703
|
Wed Jan 11 19:20:23 2017 |
Max Isi | Update | Summary Pages | December outage |
The summary pages were not successfully generated for a long period of time at the end of 2016 due to syntax errors in the PEM and Weather configuration files.
These errors caused the INI parser to crash and brought down the whole gwsumm system. It seems that changes in the configuration of the Condor daemon at the CIT clusters may have made our infrastructure less robust against these kinds of problems (which would explain why there wasn't a better error message/alert), but this requires further investigation.
In any case, the solution was as simple as correcting the typos in the config side (on the nodus side) and restarting the cron jobs (on the cluster side, by doing `condor_rm 40m && condor_submit DetectorChar/condor/gw_daily_summary.sub`). Producing pages for the missing days will take some time (how to do so for a particular day is explained in the wiki https://wiki-40m.ligo.caltech.edu/DailySummaryHelp).
RXA: later, Max sent us this secret note:
However, I realize it might not be clear from the page which are the key steps. These are just running:
1) ./DetectorChar/bin/gw_daily_summary --day YYYYMMDD --file-tag some_custom_tag To create pages for day YYYYMMDD (the file-tag option is not strictly necessary but will prevent conflict with other instances of the code running simultaneously).
2) sync those days back to nodus by doing, eg: ./DetectorChar/bin/pushnodus 20160701 20160702
This must all be done from the cluster using the 40m shared account. |
12702
|
Wed Jan 11 16:35:03 2017 |
gautam | Update | CDS | power glitch - recovery progress |
[lydia, ericq, gautam]
We set about following the instructions linked in the previous elog. A few notes/remarks:
- It is important to run the ntpdate commands before restarting the models. Sometimes, multiple restarts of the models were required to turn all the indicator blocks on the MEDM screen green.
- There was also an issue of multiple ntpd processes running on the same machine, which obviously caused all sorts of timing havoc. EricQ helped us diagnose and fix these. At the moment, all the lights are green on the CDS status MEDM screen

- On the hardware side, apart from the usual suspects of frontends/megatron/optimus/fb needing to be rebooted, I noticed that the ETMX OSEM lights were off on the control room monitors. Investigation pointed to the 2 20V sorensens at the X end outputting 0V, 0A after the power glitch. We turned down both dials, and then gradually ramped them up again. Both Sorensens now read +/-20V, 0.3A, which is in agreement with the label stuck onto them.
- Restarted MC autolocker and FSS Slow scripts on megatron. I have not yet looked at the status of the nds2 server on megatron.
- 11 MHz Marconi has yet to be restarted - but I am unable to get even the IMC locked at the moment. For some reason, the RMS of the MC1 and MC3 coils are way higher than what I am used to seeing (~5mV rms as compared to the <1mV rms I am used to seeing for a damped optic). I will investigate further. Leaving MC autolocker disabled for now.
|
12701
|
Tue Jan 10 22:55:43 2017 |
gautam | Update | CDS | power glitch - recovery steps |
Here is a link to an elog with the steps I had to follow the last time there was a similar power glitch.
The RAID array restart was also done not too long ago, we should also do a data consistency check as detailed here, if not already..
If someone hasn't found the time to do this, I can take care of it tomorrow afternoon after I am back.
Quote: |
Does "done" mean they are OK or they are somehow damaged? Do you mean the workstations or the front end machines?
The computers are all done.
|
megatron and optimus are not responding to ping commands or ssh -- please power them up if they are off; we need them to get data remotely
|
|
12700
|
Tue Jan 10 21:47:00 2017 |
rana | Update | CDS | power glitch |
Does "done" mean they are OK or they are somehow damaged? Do you mean the workstations or the front end machines?
The computers are all done.
|
megatron and optimus are not responding to ping commands or ssh -- please power them up if they are off; we need them to get data remotely |