40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 81 of 335  Not logged in ELOG logo
ID Date Author Typeup Category Subject
  864   Wed Aug 20 18:09:48 2008 YoichiUpdateIOOMC still unlocks
Being suspicious of FSS PC path as the culprit of the MC unlocks, I opened the FSS box and connected a probe to the TP7,
which is a test point in the PC path (before high voltage amplifier).
The signal is routed to an unused fast DAQ channel in the IOO rack. It is named C1:IOO-MC_TMP1 and recorded by the frame
builder. You can use this channel as a generic test DAQ channel later.

By looking at the attachment, the PC path (C1:IOO-MC_TMP1) goes crazy at the same time as other channels. So probably
it is not the trigger for the MC unlock.

Then I noticed the WFS signals drift away just before the unlock as shown in the attached plot. So now the WFS is the
main suspect.
Rob tweaked MC1 pitch to center the WFS QPDs while the MC is not locked. It improved the shape of the MC reflection.
However, the sudden MC unlock still happens. We then lowered the WFS gain from 0.5 to 0.3. Did not change the situation.
It looks like the MC length loop starts oscillating after the WFS signals drift away.
We will measure the WFS and MC OPLTF to see the stability of the loops tomorrow.

Attachment 1: MC-unlock.png
MC-unlock.png
  868   Thu Aug 21 18:13:24 2008 ranaUpdateIOOMC WFS Control signals not responsible for lock losses
This is a 4 hour, second-trend of the MC WFS error and control signals.

There is no sign that the MC loses lock because of feedback signal saturations.
Attachment 1: Untitled.png
Untitled.png
  869   Fri Aug 22 10:39:41 2008 JenneUpdateSUSTaking Free Swinging spectra of PRM, SRM, ITMX, BS
I'm taking free swinging spectra of PRM, SRM, ITMX and BS, so I've turned off their watchdogs for now. I should be done around 11:15am, so I'll turn them back on then.
  871   Fri Aug 22 16:06:29 2008 steveUpdatePEMparticle counter replaced, flowbenches & HEPAs checked
MetOne #2 counter was swapped in (on the top of IOC, facing SW direction, at ~75 deg upwards)
with channel one size 0.3 micron and channel two size 1.0 micron.
Sampling time was reduced from 60s to 6 sec at 0.1 cf/min at 25 min rate.

This means that displayed number needs to be multiplied by x100 to get particles/cf/min


HEPA filters and flow benches were checked:

PSL enclosure closed, HEPA speed at 60% 0-0 particles on optical table NW corner

AP covered optical table 1,000 particles of 0.3micron and 10 of 1.0 micron at NE corner

Flow bench at SE 0-0 particle (p)

on the top of SP cover at SE corner 60,000 p of 0.3 micron and 530 p of 1.0 micron

Mobile HEPAs 10cm from output screen in the center 800 p of 0.3 micron and 0 p of 1 micron
These filters will be replaced.

Clean assembly room:
both flow benches 0-0 p for 0.3-1.0 micron
east side bench 520 p of 0.3 micron and 210 p of 1.0 micron

Large hood in baking room with fan on 1.7 million p of 0.3 micron
and 16,000 p of 1.0 micron

Pasadena air just outside of main entrance:
3 million p of 0.3 micron and 30,000 p of 1.0 micron

My desk 743,000 p of 0.3micron, 63,000 p of 0.5 micron and 5,500 p of 1.0 micron cf/min

NOTE: existing COCHECKLIST.adl PEM displays needs to be corrected so it shows the 10 fold increase
and change particle size on this screen to 0.3 micron
  872   Fri Aug 22 17:03:41 2008 YoichiUpdateIOOMC open loop TF
I measured the open loop TF of the overall MC loop using the sum-amp A of the MC board.
I used the Agilent 4395A network analyzer and saved the data into a floppy disk. However, the data was corrupted when
I read it with my computer. I had the same problem before. The floppy is not reliable. Anyway, I have to re-measure the TF.
From what I remember, the UGF was around 25kHz and the phase margin was less than 15deg.
Above this frequency, the open loop gain was almost flat and had a small bump around 100kHz.
This bump has a gain margin of less than 4dB (the phase is more than 180deg delayed here).
So the MC is marginally stable and either decreasing or increasing the gain will make it unstable easily.
Probably, the broken FSS is responsible for this. We have to fix it.

During the measurement, I also found that the input connectors (IN1 and IN2) of the MC board are freaky.
These are TNC connectors directory mounted on the board. Gently touching the cables hooked up to those connectors
caused a large offset change in the output.
When Rana pulled the board out and pushed it in firmly, the strange behavior went away. Probably, the board was
not correctly inserted into the backplane.
This could have been the reason for the MC unlocks.
  873   Sat Aug 23 09:39:51 2008 rana, jenneUpdatePSLPMC Survey
Jenne, Rana

We scoped out the PMC situation yesterday.

Summary: Not broke. UGF ~ 500 Hz. Needs some electronics work (notches, boosts, LPFs)

Ever since we swapped out the PMC because of the broken PZT of the previous one, the UGF has been
limited to a low value. This is because the notches no longer match the mechanical resonant
frequencies of the body. The old one had a resonance at 31.3 kHz which we were notching using
the LC notch on the board as well as a dangling Pomona box in the HV line to the PZT. The one
has a resonance at ~14.5 kHz which we don't yet have a notch for. Jenne has all the real numbers and
will update this entry with them.

Todo:
  • Implement the 4th order Grote low pass after the mixer.
  • Replace the AD797 with an OP27.
  • Change servo filter to have a boost (need DC gain)
  • Make a 14.5 kHz notch for the bode mode.
  • Put a 20 lb. gold-foil wrapped lead brick on the PMC.

Here's the link about the modified PMC board which we installed at LHO:
LHO PMC elog 2006
  874   Mon Aug 25 10:07:35 2008 JenneUpdatePSLNumbers for the PMC servo board (Re: entry # 873)
Jenne, Rana

These are the numbers that go along with Rana's entry #873:

The existing notch in the PMC servo is at 31.41kHz.

The power spectrum of the PMC has a peak at 14.683kHz when it is just sitting on the PSL table (no extra mass). When we put a pile of steel and aluminum (~20lbs) on top of the PMC, the body resonance moves to 14.622kHz, but is decreased by about 40 dB!

Rana has ordered a lead brick + foil that should arrive sometime this week. To complete the mechanical part of this installation, we need to extend the earthquake mounts around the PMC so that the lead brick can't fall off of the PMC onto the rest of the table.
  876   Mon Aug 25 10:51:06 2008 steveUpdatePSLpsl headtemp is coming down
The laser water cooler was overflowing this morning.
I removed 500 cc water from the chiller.

The 4 days plot shows clearly:
that the capacity of the chiller is depending on the water level.
Overflowing water is a heat load for the chiller, so laser head temp goes up.
Attachment 1: ht4dw.jpg
ht4dw.jpg
  878   Mon Aug 25 12:13:49 2008 JenneUpdatePSLBroken PMC Servo Board
I broke the PMC servo board (on accident).

I was trying to measure the resistance of the extra resistor that someone put between the board and the HV OUT connector, since this is part of an RC filter (where C is the capacitance of the PZT on the PMC) that I need to know the values of as part of my mission to make a 14.6kHz notch for the PMC body mode. The resistance is 63.6k. I had to pull the board to get in to measure this resistance.

This resistor between the board and the center pin of the panel-mount HV OUT connector made a rigid connection between the board and the panel. When I was putting the board back in, I must have strained this connection enough that it broke. We don't have any of the same kind of resistor here at the 40m, so I'm waiting until after lunch to go to Wilson house and see if they've got any. The IFO is down until I get this sorted out.
  879   Mon Aug 25 14:18:36 2008 JenneUpdatePSLPMC servo board is fixed
The PMC servo board is back in place, all fixed up with a shiny new resistor. The PMC locks, and the MC locks (I'm not saying anything either way about how long the MC will stay locked, but it is locked for now). The resistor is connected to the connector using a short piece of wire, so this problem won't happen again, at least with this connector on this board.
  892   Wed Aug 27 13:55:43 2008 rana,jenneUpdatePSLPMC Servo Board
Board is back in. PMC is locked.

Nominal gain is now 15 dB with brick. We need to do more studies:

  • Find out why there is still 35 MHz signal at the error point. Order some low pass filters to cut off above 35 MHz.
  • Explore brick + no-brick loop shapes and error spectra.
  • Measure and set the OLG.

We've left the copper-wrapped lead brick installed to let it slowly conform to the glass better.
  895   Fri Aug 29 02:40:43 2008 rana,jenneUpdatePSLPMC Servo Board

Quote:
Board is back in. PMC is locked.


This entry has details about the low pass filter after the PMC mixer. This filter has a few purposes:

1] Remove the beat signal (at 2*f_mod) between the PD RF signal at f_mod and the LO signal at f_mod.
2] Remove the beat signal (at f_mod) between the PD RF signal at 2*f_mod (which comes from the
beating of the upper and lower RF sidebands) and the LO signal at f_mod.
3] Remove other RF signals from non-ideal behavior of the LO drive signal and distortion in the RF PD pre-amp.


So its important to have a very good rejection at 35 MHz and higher. I used the Hartmut LC network design which is
installed on H1, H2, & L1. Since there is a high gain in the audio amps right after the mixer we have to get rid of
the RF or else we'll get slew rate limited or otherwise rectified downconversion of the RF signal into our audio band.

Of course, what everyone immediately realizes from the above 3 points, is that this filter can't protect the PMC
noise performance from homodyne mixing (e.g. 2*f_mod in the LO and 2*f_mod in the RF PD). To get around that, we're
ordering some filters from Mini-Circuits to remove the 2f from those signals by ~30 dB. As long as we install
the same filters on the RF and LO legs, there should be no significant phase shift in the demodulated signal.

The attached 2 page PDF shows the calculated before and after TFs of this filter. The 2 attached .m files
calculate the TF's and have ascii art which shows how the filter works.

Here's a comparison of the attenuation (in dB) of 2 candidate Mini-circuits filters:

f(MHz)SLP-30SLP-50
31 0.5 0.4
35 1.3 0.4
38 6.1 0.4
40 10.8 0.42
61 46.3 14.8
71 60 29
91 76.9 48
10780 60

We don't have tabulated data at the same frequencies for both filters so I just made up some of the points by eye-balling the
plots from the catalog - but you get the idea: we can get away with using the SLP-30 at 35 MHz since it only attenuates the
signals by ~1.5 dB. So if someone can find 4 of these then Steve doesn't have to order any from Mini-Circuits.
Attachment 1: pmclp-07.pdf
pmclp-07.pdf pmclp-07.pdf
Attachment 2: pmclp_40m_080824.m
% PMCLP is a TF of the IF filter after the PMC mixer
%       

% Mixer_Voltage -- Rs -- L1 --- L2 ---------Vout
%                            |      |   |
%                           C1     C2   Rl                   
%                            |      |   |
%                           GND    GND  GND
%

... 58 more lines ...
Attachment 3: pmclp.m
% PMCLP is a TF of the IF filter after the PMC mixer
%       

% Mixer_Voltage -- Rs -- L1 --- L2 ---------Vout
%                            |      |   |
%                           C1     C2   Rl                   
%                            |      |   |
%                           GND    GND  GND
%

... 57 more lines ...
  901   Fri Aug 29 15:01:45 2008 steveUpdatePSLMOPA_HTEMP in increasing
The laser chiller temp is 21.9C ( it should be 20.0C )
Control room temp 73F ok, no obvious block

Ops, there is a piece of paper blocking the intake of the chiller

This is a four day plot. The paper was blocking the air flow all day.
Attachment 1: htcl.jpg
htcl.jpg
  905   Fri Aug 29 22:57:48 2008 YoichiUpdatePSLFSS loop transfer functions
I've been measuring a bunch of transfer functions of the FSS related stuffs.
There are a lot to be analyzed yet, but here I put one mystery I'm having now.
Maybe I'm missing something stupid, so your suggestions are welcome.

Here is a conceptual diagram of the FSS control board

                                                          TP3             TP4
                                                           ^               ^
                                                           |               |
RF PD -->--[Mixer]-----[Sum Amp]------>--[Common Gain]--->----[Fast Gain]----[Filter]--> NPRO PZT
              ^     |      ^        |                  |     
              |     V      |        V                  |
LO ---->-------    TP1     IN      TP2                 -->---[Filter]--[High Volt. Amp.] --> Phase Corrector

What I did was first to measure a "normal" openloop transfer function of the FSS servo.
The FSS was operated in the normal gain settings, and a signal was injected from "IN" port.
The open loop gain was measured by TP1/TP2.
Now, I disconnected the BNC cable going to the phase corrector to disable the PC path and locked the ref. cav. 
only using the PZT. This was done by reducing the "Common Gain" and "Fast Gain" by some 80dB.
Then I measured the open loop gain of this configuration. The UGF in this case was about 10kHz.
I also measured the gain difference between the "normal" and "PZT only" configurations by injecting 
a signal from "IN" and measuring TP3/TP2 and TP4/TP3 with both configurations (The signal from the Mixer was
disconnected in this measurement). 

The first attachment shows the normal open loop gain (purple) and the PZT only open loop gain scaled by the 
gain difference (about 80dB). The scaled PZT open loop gain should represent the open loop gain of the PZT
path in the normal configuration. So I expected that, at low frequencies, the scaled PZT loop TF overlaps the normal
open loop TF.
However, it is actually much larger than the normal open loop gain.
When I scale the PZT only TF by -30dB, it looks like the attachment #2.
The PZT loop gain and the total open loop gain match nicely between 20kHz and 70kHz.
Closer look will show you that small structures (e.g. around 30kHz and 200kHz) of the two
TFs also overlap very well. I repeated measurements many times and those small structures are always there (the phase is
also consistently the same). So these are not random noise.

I don't know where this 30dB discrepancy comes from. Is it the PC path eating the PZT gain ?

I have measured many other TFs. I'm analyzing these.
Here is the TO DO list:

* Cavity response plot from AOM excitation measurements.
* Cavity optical gain plot.
* Reconstruct the open loop gain from the electric gain measurements and the optical gain above.
* Using a mixer and SR560(s), make a separate feedback circuit for the PZT lock. Then use the PC path
  to measure the PC path response.
* See the response of the FSS board to large impulse/step inputs to find the cause of the PC path craziness.
etc ...
Attachment 1: OPLTFs.pdf
OPLTFs.pdf
Attachment 2: OPLTFsScaled.pdf
OPLTFsScaled.pdf
  907   Mon Sep 1 04:34:00 2008 ranaUpdatePSLFSS loop transfer functions
I started from 6th item in Yoichi's todo list.

1) Increased the setpoint of the thermostat next to the framebuilder from 73F to 79F. Its freezing over there
in the room with the drill press. Steve's illegal mercury thermometer is reading 19 C.


2) Looked the RFPD's output spectrum using the 20 dB coupled output from the coupler that's in-line.
The first attached PDF file (n.pdf) has several plots:
page 1: 0-500 MHz anomolous peaks at 138 & 181 MHz but nothing too crazy
page 2: 0-100 MHz 80 MHz peak is RF pickup from the VCO Driver - not on the light
page 3: 10-30 MHz totally nuts
page 4: 18-25 MHz that's just wrong

The RF spectrum should only have some action around 21.5 MHz and a little peak at 2x 21.5 MHz. All that extra
junk means that something is broken!


3) To see if I could rid of any of the 80 MHz signal or any of that other trash from 18-25 MHz, I wound the RF cable
around a large toroidal ferrite core. This should have given us many uH of inductance for any signals common to
both the center and shield of the cable with no effect on the differential RF signals. There was no effect.


4) Next went to look at the 21.5 MHz Crystal Oscillator Reference card (D980353...I bet you can't figure out how
this one works). These have the Mini-Circuits SMA 30 MHz low pass (SLP-30) filters on both the LO and EOM outputs.

FSSLO.PNG shows the waveform after 20 dB attenuation going into a scope terminated with 50 Ohms.
FSSLO-Spec.png shows the spectrum of this signal - its pretty distorted. Here's the levels
   f (MHz) |  before filter (dBm) | after filter (dBm)
   ---------------------------------------------------
     21.5  |       -12.8               -13.1       
     43            -24                 -46
     64.5          -50               < -80
     86            -64               < -80

This would be OK after the filter, but the level is very low. Only 7dBm (accounting for my 20 dB att) !!
The FSS uses a JMS-1H mixer which needs, as everyone knows, a +17 dBm LO signal. Que lastima.

There seems to be something wrong already, but wait...


5) PC25.PNG shows the output signal going to the EOM from 0 - 25 MHz. The step that's visible there at
around 10 MHz is just something inherent to the analyzer (??). But see all that crap there down below
5 MHz ? That is NOT supposed to be there.

pc.pdf shows on the first page the comparison in EOM drive with 2 different slider values on the
RF AM adjust screen for the FSS. But page 2 is the punchline of this long entry: There is a bunch of
excess junk on the drive signal going to the FSS's phase modulator.
The FSS is then trying to handle
this extra frequency noise and getting into trouble.

We have to fix this board. I have also ordered a few SBP-21.4 from mini-circuits (SMA bandpass around 21.4 MHz)
just in case. Another option is to just replace this thing with a Marconi and an RF amp.






Attachment 1: n.pdf
n.pdf n.pdf n.pdf n.pdf
Attachment 2: FSSLO.PNG
FSSLO.PNG
Attachment 3: FSSLO-Spec.png
FSSLO-Spec.png
Attachment 4: PC25.png
PC25.png
Attachment 5: pc.pdf
pc.pdf pc.pdf
  911   Tue Sep 2 10:09:03 2008 steveUpdatePSLhead temp is cooling down
The chiller was over flowing this morning.
800 cc of water was removed.
PSL-126MOPA_HTEMP peaked at 20.7 C (normal is 18.7 C)
  912   Tue Sep 2 14:28:41 2008 YoichiUpdatePSLFSS EOM driving signal spectra
Rich advised me to change the +10V input of the FSS crystal frequency reference board from whatever voltage supply we use now to a nice one.
This voltage is directory connected to the signal lines of both LO and RF output amps. Therefore, fluctuations in the voltage directly appear
in the outputs, though DC components are cut off by the AC coupling capacitors.

I changed the source of this voltage from the existing Sorensen one to a power supply sitting next to the rack.
The attached plots shows the difference of the RF output spectra between the two 10V sources.
The low frequency crap is almost gone in the new 10V spectrum.

I tried to increase the FSS gain with the new 10V, but still it goes crazy. I suspect it is because the LO power is too low.
Attachment 1: RFDrive1.png
RFDrive1.png
Attachment 2: RFDrive2.png
RFDrive2.png
  918   Thu Sep 4 00:38:14 2008 ranaUpdatePSLc1iovme power cycled
Entry 663 has a plot of this using the PSL/FSS/SLOWscan script. It shows that the SB's were ~8x smaller than the carrier.
P_carrier   J_0(Gamma)^2 
--------- = ------------
P_SB        J_1(Gamma)^2

Which I guess we have to solve numerically for large Gamma?
  919   Thu Sep 4 07:29:52 2008 YoichiUpdatePSLc1iovme power cycled

Quote:
Entry 663 has a plot of this using the PSL/FSS/SLOWscan script. It shows that the SB's were ~8x smaller than the carrier.
P_carrier   J_0(Gamma)^2 
--------- = ------------
P_SB        J_1(Gamma)^2

Which I guess we have to solve numerically for large Gamma?


P_carrier/P_SB = 8 yields gamma=0.67.
  920   Thu Sep 4 07:46:10 2008 YoichiUpdateIOOMC is now happy
The MC has been locked for more than 12 hours continuously now !
Changes I made yesterday were:
(1) Removed the 20dB attenuator before the EOM.
(2) Reduced the Fast Gain from 21dB to 16dB, which made the PC to be a little bit more loaded (~0.6Vrms).

As Rana pointed out in the meeting, setting the Fast Gain a bit lower may have put the FSS in a stabler state.
Attachment 1: MC-lock.png
MC-lock.png
  921   Thu Sep 4 10:13:48 2008 JenneUpdateIOOWe unlocked the MC temporarily
[Joe, Eric, Jenne]

While trying to diagnose some DAQ/PD problems (look for Joe and Eric's entry later), we unlocked the PMC, which caused (of course) the MC to unlock. So if you're looking back in the data, the unlock at ~10:08am is caused by us, not whatever problems may have been going on with the FSS. It is now locked again, and looking good.
  923   Thu Sep 4 13:48:50 2008 YoichiUpdatePSLFSS modulation depth
I scanned the reference cavity with the NPRO temperature (see the attached plot).
The power ratio between the carrier and the sideband resonances is about 26.8.
It corresponds to gamma=0.38.
The RF power fed into the EOM is now 14.75dBm (i.e. 1.7V amplitude). The NewFocus catalog says 0.1-0.3rad/V. So
gamma=0.38 is a reasonable number.




Attachment 1: RCScan.png
RCScan.png
  924   Thu Sep 4 14:43:58 2008 JenneUpdatePSLPMC Open Loop Gain
I have measured the PMC's open loop gain. UGF is 629.7Hz, with a phase margin of 53 degrees.

I injected into FP2 on the front panel, and measured MixOut/Source from 100Hz to 100kHz using the SR785. I did this both when the loop was open, and when the loop was closed (open the loop by enabling FP1, which breaks the loop).

We have 2 transfer functions involved: The actual open loop gain of the PMC servo loop (G1), and the gain between FP2 and the MixerOut monitor point (G2). This gives us:

TF(closed loop) = G2*(1+G1)
TF(broken loop) = G2

G1 = TF(closed)/TF(broken) - 1

This G1 is the final open loop gain, and it is plotted below.
Attachment 1: OpenLoopTF04Sept2008.png
OpenLoopTF04Sept2008.png
  926   Thu Sep 4 17:03:25 2008 YoichiUpdatePSLRF oscillator noise comparison
I measured current spectra of the RF signal going to the FSS EOM.
The attachment compares the spectra between a Stanford signal generator and a Marconi.
I borrowed the Marconi from the abs. length measurement experiment temporarily.
The measurement was done using the signal going to the EOM. That means the spectra include
noise contributions from the RF amp., splitter and cables.

21.5MHz peak was not included because that would overload the ADC and I would have to use a large attenuation.
This means the measurement would be totally limited by ADC noise everywhere except for 21.5MHz.

I noticed that with the Marconi, the FSS is a little bit happier, i.e. the PC path is less loaded
(0.9Vrms with Stanford vs. 0.7Vrms with Marconi). But the difference is small.
Probably the contribution from the 77kHz harmonics in the laser light is more significant (see entry #929).
Also the peaks in the Stanford spectrum are not harmonics of 77kHz, which we see in the FSS error signal.

I returned the Marconi after the measurement to let Alberto work on the abs. length measurement.
Attachment 1: RFSpectra.png
RFSpectra.png
  927   Thu Sep 4 17:12:57 2008 YoichiUpdatePSLFSS open loop TF
I changed the gain settings of the FSS servo.
Now the Common Gain is 5dB (the last night it was 2dB) and the Fast Gain is 12dB (formerly 16dB).
I measured the open loop TF with this setting (the attachment).
I also plotted the OPLTF when CG=2dB, FG=20.5dB. With this setting, the MC looses lock every 30min.

You can see that the OPLTF is smoother with FG=12dB.
When the FG is high, you can see some structure around 250kHz. This structure is reproducible.
This may be some interruption from the fast path to the PC path through a spurious coupling.
Attachment 1: FSS-OPLTFs.png
FSS-OPLTFs.png
  928   Thu Sep 4 17:17:03 2008 YoichiUpdateIOOMC open loop TF
I measured open loop transfer functions of the MC servo.
The UGF was about 30kHz. Since there was some gain margin at higher frequencies, I increased
the input gain of the MC servo board from 19dB to 22dB. Now the UGF is 40kHz and we have more
phase margin (~30deg).
Attachment 1: MC-OPLTF.png
MC-OPLTF.png
  929   Thu Sep 4 17:44:27 2008 YoichiUpdatePSLFSS error signal spectrum
Attached is a spectrum of the FSS error signal.
There are a lot of sharp peaks above 100kHz (the UGF of the servo is about 200kHz).
These are mostly harmonics of 77kHz. They are the current suspects of the FSS slew rate saturation.
I remember when I blocked the light to the PD, these peak went away. So these noises must be
in the light. But I checked it a few weeks ago. So I will re-check it later.

One possible source of the lines is a DC-DC converter in the NPRO near the crystal.
We will try to move the converter outside of the box.
Attachment 1: FSS-Error-Spe.png
FSS-Error-Spe.png
  931   Fri Sep 5 08:34:03 2008 steveUpdatePSLMZ locked
The MC is happy.
The MZ can be locked if you move the slider by hand.
Attachment 1: mzhv.jpg
mzhv.jpg
  933   Fri Sep 5 10:36:34 2008 steveUpdatePEMthermostate setting changed
Some one changed the thermostat (old control room ) setting behind 1Y6 from 73 to 79F
It should be in the elog.
The temp changed from freezing 20 to sunny 25 C
  934   Fri Sep 5 15:09:50 2008 ranaUpdatePEMthermostate setting changed

Quote:
Some one changed the thermostat (old control room ) setting behind 1Y6 from 73 to 79F
It should be in the elog.

In fact, it is. I demand satisfaction for the injury to my elogging reputation!
  935   Mon Sep 8 10:57:49 2008 steveUpdateIOOthe psl and mc are back to normal
The alarm handler is silent this morning.
This is almost unbelievably pleasant after two mount of harassment.
The MC did not lose lock for three days.

Atm1: the new fss layout
Atm2: PMC with lead brick
Atm3: 3 days plot
Attachment 1: fss.png
fss.png
Attachment 2: pmcbrick.png
pmcbrick.png
Attachment 3: brick.jpg
brick.jpg
  936   Mon Sep 8 13:47:35 2008 steveUpdatePEMthermostate setting changed

Quote:

Quote:
Some one changed the thermostat (old control room ) setting behind 1Y6 from 73 to 79F
It should be in the elog.

In fact, it is. I demand satisfaction for the injury to my elogging reputation!




Thermostate setting was changed from 79F to 77F behind 1Y6
  938   Wed Sep 10 08:57:03 2008 steveUpdateGeneraletmy illuminator turned off
The ETMY illuminator was left on yesterday.
I just turned it off.
  940   Wed Sep 10 19:53:53 2008 AlbertoUpdateGeneralabs length experiment
Update of the last days work on the experiment to measure the absolute length of the cavities.

I'm trying to repeat the same measurement that Koji did on the Y arm, before switching to the X arm.

I switched to the PHD universal box for the PLL control between the main laser and the secondary laser. I found a good gain value for the servo and now I can set the frequency of the beat to any value as long as I do it slowly turning the LO frequency from the knob on the Marconi.

I laid down a 50m BNC cable from the Y end to near the BS chamber, where all the abs length equipment is. I matched the two laser beams changing the alignment of the injection steering mirror at the the dark port on the AP table. I then locked the Y arm cavity. When I first tried to do that, the locking script didn't work because the beam was off of the 'sweet spot' where Rob had set it on Monday. It turned out that aborting the script during one of its previous run, had changed the alignment of the PZT steering mirrors. So with Rob I brought them back near the positions as in the snapshot and then saved a new one with the latest values.

Eventually I could set the beat frequency to the FSR of the arm cavity and saw it in transmission at the ETMY.

Now I'm working on the LabView interface for the GPIB data acquisition board.
  943   Thu Sep 11 23:28:35 2008 albertoUpdateGeneralabs cavity length experiment
The MC lost lock for some reason not related to either the FSS or the PMC I'm done with my measurement for tonight. I've shut the NPRO beam before leaving.
  944   Fri Sep 12 11:09:20 2008 AlbertoUpdateGeneralabs cavity length experiment
I'm leaving the lab for a couple of hours. I shut the NPRO. The interferometer is available to anyone.
  945   Sat Sep 13 23:13:01 2008 AlbertoUpdateGeneralabs cavity length experiment
The Y arm was locked all time today but, suddenly, this afternoon it lost lock and since then I've been unable to restore it. I tried unsuccessfully the Restore and the Align scripts several times, although the position of PZT steering mirrors were good (as in the snapshot). I tried things like unlocking/locking the MC, the FSS reference cavity, the PMC but it didn't work. Then I decided to switch to the X arm. Locking it was a piece of cake compare to Y. I'm going to start measuring the FSR of the X arm.
  946   Sun Sep 14 18:30:32 2008 AlbertoUpdateGeneralABSL: measured X arm
Today I measured the X arm FSR.
Hi moved the fast PD (Thor Labs PDA255) from the Y end table to the X end table. I had to use a beam splitter to pick out the transmitted beam from the cavity beam and send it to the PD. I did not want to interpose the BS before the TRANS X PD, so I had to move the ETMXT camera to an other place in the table to gain some room. Now the beam that used to go directly to the camera is 50% split and goes also to the PD. I had to put a lens to focus the beam on the PD. The transmitted beam is currently not aligned to the ETMXT camera, I need to fix the alignment of the BS before.
I'm now doing a rough scan of a frequency range 5 times as large as the FSR. I'll post the results soon.
  947   Sun Sep 14 19:29:07 2008 AlbertoUpdateGeneralABSL: measured X arm

Quote:
Today I measured the X arm FSR.
Hi moved the fast PD (Thor Labs PDA255) from the Y end table to the X end table. I had to use a beam splitter to pick out the transmitted beam from the cavity beam and send it to the PD. I did not want to interpose the BS before the TRANS X PD, so I had to move the ETMXT camera to an other place in the table to gain some room. Now the beam that used to go directly to the camera is 50% split and goes also to the PD. I had to put a lens to focus the beam on the PD. The transmitted beam is currently not aligned to the ETMXT camera, I need to fix the alignment of the BS before.
I'm now doing a rough scan of a frequency range 5 times as large as the FSR. I'll post the results soon.


I'm leaving a long measurement running. I should be back later on. If I won't, whoever wanted to use the interferometer has just to shut the NPRO laser in the AP table.
  953   Wed Sep 17 12:58:12 2008 robUpdateLockingbad

Locking was pretty unsuccessful last night. All the subparts were locked (ARMs, PRM, DRM) and
aligned, but no DRMI+2ARMs locks. The alignment may have drifted significantly by the time I
got around to working the full shebang, however.

We should get back into the habit of clicking the
yellow "Restore last auto-alignment" button when we finish using the interferometer.
  956   Wed Sep 17 13:58:36 2008 AlbertoUpdateGeneralABSL: results from the X arm
Today I repeated the measurement of the FSR on the X arm cavity. The noise in the transmitted power that made the measures fluctuate was much reduced after last night Rob worked on the interferometer. The X arm cavity length is now: (38.4580+/-0.0003)m. I'm attaching a summary of the data I've taken.

I'm now preparing the setup to measure the transverse mode spacing.
Attachment 1: Sept17_XarmFSRmeasurement_report.ps
Sept17_XarmFSRmeasurement_report.ps
  958   Wed Sep 17 17:31:24 2008 YoichiUpdatePSLFSS calibration
I calibrated the reference cavity error signal with the following procedure.

(1) I disconnected the PC path BNC cable and locked the RC only using the PZT. To do so, I had to insert a 20dB attenuator
in the RF signal path going to the EOM to reduce the gain of the loop sufficiently.
The normal RF level going to the EOM is 17dBm. With the attenuator it is of course -3dBm.

(2) Using the SR785, I injected signal into the Test-IN2 (a sum-amp after the mixer) of the FSS box and measured the TF from the Ramp-IN to the IN1.
When the Ramp-In switch is off, the Ramp-IN port can be used as a test point connected to the PZT drive signal path just before the output.
There is a RC low-pass filter after the Ramp-IN. IN1 is the direct output from the mixer (before the sum-amp).
The attm1 is the measured transfer function along with the fitting by a first order LPF.
From this measurement, the DC transfer function from the applied voltage on the PZT to the error signal is determined to be 163.6 (V/V).
Since the RF level is lowered by 20dB, the cavity gain in the normal operation mode is 10 times larger (assuming that the modulation depth is
linearly proportional to the applied voltage to the EOM).

(3) According to elog:791, the conversion factor from the voltage on the PZT to the frequency change of the NPRO is 11.172MHz/V. Combining this with the
number obtained above, we get 6.83kHz/V as the calibration factor for converting the error signal (mixer output) to the frequency at DC.
Using 38kHz cavity pole frequency, the calibration factor is plotted as a function of frequency in the attm2.

(4) I took a spectrum of the error signal of the FSS and calibrated it with the obtained calibration factor. See attm3.
The spectrum was measured by SR785. I will measure wide band spectra with an RF analyzer later.

TO DO:
1: Use the actual modulation depth difference to extrapolate the calibration factor obtained by with a low RF signal for the EOM.
The cavity sweep was already done.

2: I assumed 38kHz cavity pole. I will measure the actual cavity pole frequency by cavity ringdown.

3: Measure out-of-the-loop spectrum of the frequency noise using PMC and MC.
Attachment 1: PZTresp.png
PZTresp.png
Attachment 2: Calibration.png
Calibration.png
Attachment 3: FreqNoiseSpectrum.png
FreqNoiseSpectrum.png
  960   Wed Sep 17 19:13:47 2008 AlbertoUpdateGeneralABSL: status
I installed the setup for measuring TEM01/10 on the X end table.
I'm leaving. I shut the laser, flipped down the flipper mirror, disconnected the pzt drive signal from the laser.
For Jenne. The power cable for the Guralps' board is now connected to the PDH box on my instruments cart but you can take it.
  962   Thu Sep 18 09:30:12 2008 steveUpdateGenerallow noise metal film resistors are in
Low noise metal film resistor and capacitor kits from www.garrettelec.com are in.

manufacturer: Dale, 289 values, 25ea, surface mount,1206, 0.1% from 100 to 100K, 1/8 or 1/4W
additional values below 100 ohm and above 100K were purchased from Mouser with the same Dale specification

Ceramic capacitor kit from AVX
67 values, 25ea, surface mount, 1206 from 1.0 pF up

atm2: our new storage cabinet pick and put together by Jenni
Attachment 1: mfr&cap.png
mfr&cap.png
Attachment 2: storcab.png
storcab.png
  963   Thu Sep 18 12:16:01 2008 YoichiUpdateComputersEPICS BACK

Quote:

Somehow the EPICS system got hosed tonight. We're pretty much dead in the water till we can get it sorted.


The problem was caused by the installation of a DNS server into linux1 by Joe.
Joe removed /etc/hosts file after running the DNS server (bind). This somehow prevented proper boot of
frontend computers.
Joe and I confirmed that putting back /etc/hosts file resolved the problem.
Right now, the DNS server is also running on linux1.

We are not sure why /etc/hosts file is still necessary. My guess is that the NFS server somehow reads /etc/hosts
when he decides which computer to allow mounting. We will check this later.

Anyway, now the computers are mostly running fine. The X-arm locks.
The Y-arm doesn't, because one of the digital filters for the Y-arm lock fails to be loaded to the frontend.
I'm working on it now.
  964   Thu Sep 18 13:05:05 2008 YoichiUpdateComputersEPICS BACK

Quote:

The Y-arm doesn't, because one of the digital filters for the Y-arm lock fails to be loaded to the frontend.
I'm working on it now.


Rob told me that the filter "3^2:20^2" is switched on/off dynamically by the front end code for the LSC.
Therefore, the failure to manually load it was not actually a problem.
The Y-arm did not lock just because the alignment was bad.
Now the Y-arm alignment is ok and the arm locks.
  967   Thu Sep 18 23:31:26 2008 ranaUpdatePSLISS: Saturating too often at nominal gain
The ISS has been saturating whenever the MC relocks and puts the gain up to +8dB. I have
lowered the gain to +1 dB for now to stop this, but we need to revisit the ISS loop and
performance. Stefan can fix it up for us as penance when he returns from the hedonism of Amsterdam.
Attachment 1: FIRE_BLOWER.jpg
FIRE_BLOWER.jpg
  968   Fri Sep 19 00:06:54 2008 ranaUpdateIOOMC_F: Too much frequency noise around 100 Hz
WE noticed this excess again in MC_F. We tried recentering the WFS, but no effect.

Also no effect from changing the FSS gain, PMC gain, or ISS gain.

Actually, there IS a change when changing the PMC gain -- the ISS can be made to saturate
by lowering the PMC gain by 10 dB. Jenne and I need to finish off the PMC loop.

10 kHz UGF or bust!
Attachment 1: mcf.png
mcf.png
  969   Fri Sep 19 00:18:14 2008 ranaUpdateComputerssvn is old
linux2:mDV>ssh nodus
Password:
Last login: Fri Sep 19 00:11:44 2008 from gwave-69.ligo.c
Sun Microsystems Inc.   SunOS 5.9       Generic May 2002
nodus:~>c
nodus:caltech>cd apps/
nodus:apps>cd mDV
nodus:mDV>svn update
svn: This client is too old to work with working copy '.'; please get a newer Subversion client
nodus:mDV>whoami
controls
nodus:mDV>uname -a
SunOS nodus 5.9 Generic_118558-39 sun4u sparc SUNW,A70 Solaris
nodus:mDV>pwd
/cvs/cds/caltech/apps/mDV
nodus:mDV>
Frown
  971   Fri Sep 19 08:09:55 2008 steveUpdatePSLpsl HEPAs turned on
I have just turned on the PSL HEPA filters at 60% operational speed.
ELOG V3.1.3-