40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 81 of 335  Not logged in ELOG logo
ID Date Author Typeup Category Subject Text Attachments
  864   Wed Aug 20 18:09:48 2008 YoichiUpdateIOOMC still unlocksBeing suspicious of FSS PC path as the culprit
of the MC unlocks, I opened the FSS box and
connected a probe to the TP7, 
 MC-unlock.png 
  868   Thu Aug 21 18:13:24 2008 ranaUpdateIOOMC WFS Control signals not responsible for lock lossesThis is a 4 hour, second-trend of the MC WFS
error and control signals.
 Untitled.png 
  869   Fri Aug 22 10:39:41 2008 JenneUpdateSUSTaking Free Swinging spectra of PRM, SRM, ITMX, BSI'm taking free swinging spectra of PRM, SRM,
ITMX and BS, so I've turned off their watchdogs
for now. I should be done around 11:15am,
  
  871   Fri Aug 22 16:06:29 2008 steveUpdatePEMparticle counter replaced, flowbenches & HEPAs checkedMetOne #2 counter was swapped in (on the top
of IOC, facing SW direction, at ~75 deg upwards)
with channel one size 0.3 micron and channel
  
  872   Fri Aug 22 17:03:41 2008 YoichiUpdateIOOMC open loop TFI measured the open loop TF of the overall
MC loop using the sum-amp A of the MC board.
I used the Agilent 4395A network analyzer
  
  873   Sat Aug 23 09:39:51 2008 rana, jenneUpdatePSLPMC SurveyJenne, Rana

We scoped out the PMC situation yesterday.
  
  874   Mon Aug 25 10:07:35 2008 JenneUpdatePSLNumbers for the PMC servo board (Re: entry # 873)Jenne, Rana

These are the numbers that go along with
  
  876   Mon Aug 25 10:51:06 2008 steveUpdatePSLpsl headtemp is coming downThe laser water cooler was overflowing this
morning.
I removed 500 cc water from the chiller.
 ht4dw.jpg 
  878   Mon Aug 25 12:13:49 2008 JenneUpdatePSLBroken PMC Servo Board[SIZE=3]I broke the PMC servo board[/SIZE]
(on accident).
  
  879   Mon Aug 25 14:18:36 2008 JenneUpdatePSLPMC servo board is fixedThe PMC servo board is back in place, all
fixed up with a shiny new resistor.  The
PMC locks, and the MC locks (I'm not saying
  
  892   Wed Aug 27 13:55:43 2008 rana,jenneUpdatePSLPMC Servo BoardBoard is back in. PMC is locked.

Nominal gain is now 15 dB with brick. We
  
  895   Fri Aug 29 02:40:43 2008 rana,jenneUpdatePSLPMC Servo Board[quote]Board is back in. PMC is locked.
[/quote]
 pmclp-07.pdfpmclp_40m_080824.mpmclp.m 
  901   Fri Aug 29 15:01:45 2008 steveUpdatePSLMOPA_HTEMP in increasingThe laser chiller temp is 21.9C ( it should
be 20.0C )
Control room temp 73F ok, no obvious block
 htcl.jpg 
  905   Fri Aug 29 22:57:48 2008 YoichiUpdatePSLFSS loop transfer functionsI've been measuring a bunch of transfer functions
of the FSS related stuffs.
There are a lot to be analyzed yet, but here
 OPLTFs.pdfOPLTFsScaled.pdf 
  907   Mon Sep 1 04:34:00 2008 ranaUpdatePSLFSS loop transfer functionsI started from 6th item in Yoichi's todo list.

1) Increased the setpoint of the thermostat
 n.pdfFSSLO.PNGFSSLO-Spec.pngPC25.pngpc.pdf 
  911   Tue Sep 2 10:09:03 2008 steveUpdatePSLhead temp is cooling downThe chiller was over flowing this morning.
800 cc of water was removed.
PSL-126MOPA_HTEMP peaked at 20.7 C (normal
  
  912   Tue Sep 2 14:28:41 2008 YoichiUpdatePSLFSS EOM driving signal spectraRich advised me to change the +10V input of
the FSS crystal frequency reference board
from whatever voltage supply we use now to
 RFDrive1.pngRFDrive2.png 
  918   Thu Sep 4 00:38:14 2008 ranaUpdatePSLc1iovme power cycledEntry 663 has a plot of this using the PSL/FSS/SLOWscan
script. It shows that the SB's were ~8x smaller
than the carrier.
  
  919   Thu Sep 4 07:29:52 2008 YoichiUpdatePSLc1iovme power cycled[quote]Entry 663 has a plot of this using
the PSL/FSS/SLOWscan script. It shows that
the SB's were ~8x smaller than the carrier.
  
  920   Thu Sep 4 07:46:10 2008 YoichiUpdateIOOMC is now happyThe MC has been locked for more than 12 hours
continuously now !
Changes I made yesterday were:
 MC-lock.png 
  921   Thu Sep 4 10:13:48 2008 JenneUpdateIOOWe unlocked the MC temporarily[Joe, Eric, Jenne]

While trying to diagnose some DAQ/PD problems
  
  923   Thu Sep 4 13:48:50 2008 YoichiUpdatePSLFSS modulation depthI scanned the reference cavity with the NPRO
temperature (see the attached plot).
The power ratio between the carrier and the
 RCScan.png 
  924   Thu Sep 4 14:43:58 2008 JenneUpdatePSLPMC Open Loop GainI have measured the PMC's open loop gain.
 [SIZE=3][COLOR=green]UGF is 629.7Hz, with
a phase margin of 53 degrees.
 OpenLoopTF04Sept2008.png 
  926   Thu Sep 4 17:03:25 2008 YoichiUpdatePSLRF oscillator noise comparisonI measured current spectra of the RF signal
going to the FSS EOM.
The attachment compares the spectra between
 RFSpectra.png 
  927   Thu Sep 4 17:12:57 2008 YoichiUpdatePSLFSS open loop TFI changed the gain settings of the FSS servo.
Now the Common Gain is 5dB (the last night
it was 2dB) and the Fast Gain is 12dB (formerly
 FSS-OPLTFs.png 
  928   Thu Sep 4 17:17:03 2008 YoichiUpdateIOOMC open loop TFI measured open loop transfer functions of
the MC servo.
The UGF was about 30kHz. Since there was
 MC-OPLTF.png 
  929   Thu Sep 4 17:44:27 2008 YoichiUpdatePSLFSS error signal spectrumAttached is a spectrum of the FSS error signal.
There are a lot of sharp peaks above 100kHz
(the UGF of the servo is about 200kHz).
 FSS-Error-Spe.png 
  931   Fri Sep 5 08:34:03 2008 steveUpdatePSLMZ lockedThe MC is happy.
The MZ can be locked if you move the slider
by hand.
 mzhv.jpg 
  933   Fri Sep 5 10:36:34 2008 steveUpdatePEMthermostate setting changedSome one changed the thermostat (old control
room ) setting behind 1Y6 from 73 to 79F
It should be in the elog.
  
  934   Fri Sep 5 15:09:50 2008 ranaUpdatePEMthermostate setting changed[quote]Some one changed the thermostat (old
control room ) setting behind 1Y6 from 73
to 79F
  
  935   Mon Sep 8 10:57:49 2008 steveUpdateIOOthe psl and mc are back to normalThe alarm handler is silent this morning.
This is almost unbelievably pleasant after
two mount of harassment.
 fss.pngpmcbrick.pngbrick.jpg 
  936   Mon Sep 8 13:47:35 2008 steveUpdatePEMthermostate setting changed[quote][quote]Some one changed the thermostat
(old control room ) setting behind 1Y6 from
73 to 79F
  
  938   Wed Sep 10 08:57:03 2008 steveUpdateGeneraletmy illuminator turned offThe ETMY illuminator was left on yesterday.
I just turned it off.
  
  940   Wed Sep 10 19:53:53 2008 AlbertoUpdateGeneralabs length experimentUpdate of the last days work on the experiment
to measure the absolute length of the cavities.
  
  943   Thu Sep 11 23:28:35 2008 albertoUpdateGeneralabs cavity length experimentThe MC lost lock for some reason not related
to either the FSS or the PMC I'm done with
my measurement for tonight. I've shut the
  
  944   Fri Sep 12 11:09:20 2008 AlbertoUpdateGeneralabs cavity length experimentI'm leaving the lab for a couple of hours.
I shut the NPRO. The interferometer is available
to anyone.
  
  945   Sat Sep 13 23:13:01 2008 AlbertoUpdateGeneralabs cavity length experimentThe Y arm was locked all time today but, suddenly,
this afternoon it lost lock and since then
I've been unable to restore it. I tried unsuccessfully
  
  946   Sun Sep 14 18:30:32 2008 AlbertoUpdateGeneralABSL: measured X armToday I measured the X arm FSR.
Hi moved the fast PD (Thor Labs PDA255) from
the Y end table to the X end table. I had
  
  947   Sun Sep 14 19:29:07 2008 AlbertoUpdateGeneralABSL: measured X arm[quote]Today I measured the X arm FSR.
Hi moved the fast PD (Thor Labs PDA255) from
the Y end table to the X end table. I had
  
  953   Wed Sep 17 12:58:12 2008 robUpdateLockingbad
Locking was pretty unsuccessful last night.
 All the subparts were locked (ARMs, PRM,
  
  956   Wed Sep 17 13:58:36 2008 AlbertoUpdateGeneralABSL: results from the X armToday I repeated the measurement of the FSR
on the X arm cavity. The noise in the transmitted
power that made the measures fluctuate was
 Sept17_XarmFSRmeasurement_report.ps 
  958   Wed Sep 17 17:31:24 2008 YoichiUpdatePSLFSS calibrationI calibrated the reference cavity error signal
with the following procedure.
 PZTresp.pngCalibration.pngFreqNoiseSpectrum.png 
  960   Wed Sep 17 19:13:47 2008 AlbertoUpdateGeneralABSL: statusI installed the setup for measuring TEM01/10
on the X end table. 
I'm leaving. I shut the laser, flipped down
  
  962   Thu Sep 18 09:30:12 2008 steveUpdateGenerallow noise metal film resistors are inLow noise metal film resistor and capacitor
kits from www.garrettelec.com are in.
 mfr&cap.pngstorcab.png 
  963   Thu Sep 18 12:16:01 2008 YoichiUpdateComputersEPICS BACK[quote]
Somehow the EPICS system got hosed tonight.
 We're pretty much dead in the water till
  
  964   Thu Sep 18 13:05:05 2008 YoichiUpdateComputersEPICS BACK[quote]
The Y-arm doesn't, because one of the digital
filters for the Y-arm lock fails to be loaded
  
  967   Thu Sep 18 23:31:26 2008 ranaUpdatePSLISS: Saturating too often at nominal gainThe ISS has been saturating whenever the MC
relocks and puts the gain up to +8dB. I have
lowered the gain to +1 dB for now to stop
 FIRE_BLOWER.jpg 
  968   Fri Sep 19 00:06:54 2008 ranaUpdateIOOMC_F: Too much frequency noise around 100 HzWE noticed this excess again in MC_F. We tried
recentering the WFS, but no effect.
 mcf.png 
  969   Fri Sep 19 00:18:14 2008 ranaUpdateComputerssvn is old[CODE]linux2:mDV>ssh nodus
Password:
Last login: Fri Sep 19 00:11:44 2008 from
  
  971   Fri Sep 19 08:09:55 2008 steveUpdatePSLpsl HEPAs turned onI have just turned on the PSL HEPA filters
at 60% operational speed.
  