  10924   Tue Jan 20 19:32:06 2015 JenneUpdateSUSSUS Drift restore scriptsI made little scripts to go with the sus
driftmon buttons, that will servo the alignment
sliders until the susyaw and suspit values
  
  10925   Tue Jan 20 20:03:17 2015 JenneUpdateLSCALS lock not staying?The Xarm ALS has been a little funky today. 

First, the green and the arm-axis
would not stay co-aligned.  I'm
  
  10927   Wed Jan 21 15:27:47 2015 JenneUpdateLSCFixed LSC model bugThis problem with the CARM loop last night
was the fault of a bug that I had put into
the LSC model last week.  When I gave
  
  10929   Thu Jan 22 03:21:24 2015 JenneUpdateLSCLocks with large MICH offsets[Jenne, Diego, EricQ]

Tonight we worked on the acquisition
sequence (including re-re-re-commissioning
 MICHoffset65percent.pnglockloss_21Jan2015_MICHhittingRails.png 
  10933   Fri Jan 23 02:11:40 2015 JenneUpdateLSCCARM filters modified slightly[Jenne, Diego]

One of tonight's goals was
to tweak the CARM filters, so that we could
 plot.pngCARM_22Jan2015.pdf 
  10937   Fri Jan 23 18:08:10 2015 JenneUpdateCDSEPICS freezesSo, I neglected to elog this yesterday,
but yesterday we had one of those EPICS freezes
that only affects slow channels that come
  
  10938   Fri Jan 23 19:38:02 2015 JenneUpdateLSCcarm_cm_up supports both signs of CARM offsetA small change, but now the carm_up script
supports both sides of the CARM offset. 
After the arms are locked with ALS it asks
  
  10939   Mon Jan 26 13:07:28 2015 JenneUpdatePEMSeismometers back in nominal placesI have just put the seismometers back into
their nominal positions, on the concreted
slabs.  The T-240 is in the vertex,
 IMG_1800.JPGIMG_1805.JPG 
  10941   Mon Jan 26 21:10:04 2015 JenneUpdateModern ControlWaking up the OAFI had a look at the OAF model today. 

Somehow, the screens that we had
weren't matching up with the model. 
  
  10942   Tue Jan 27 04:11:21 2015 JenneUpdateLSCSmall tweaks to the locking[Jenne, Diego, EricQ]

We did a series of small things
that may have helped with the locking, although
 CARM_27Jan2015_JCD.pdfDARM_27Jan2015.pdf 
  10943   Tue Jan 27 14:00:05 2015 JenneUpdateModern ControlPRMI lockedPRMI is locked on sideband, starting ~4
minutes ago, to collect ASC/seismic data
for feedforward.
  
  10945   Tue Jan 27 17:58:21 2015 JenneConfigurationTreasureWelcome, Donatella!Welcome to your new abode, Donatella!  IMG_1806.JPG 
  10947   Wed Jan 28 03:01:24 2015 JenneUpdateLSCTransitioned DARM to AS55Q[Jenne, Diego]

Tonight
we were able to transition DARM from DC transmission
  
  10959   Fri Jan 30 02:57:03 2015 JenneUpdateModern ControlFirst try with PRCL ASC Wiener filtering[Aside - How do you rotate plots in the
new elog?  It's showing them correctly
in the attachments list below the entry,
 9x 
  10962   Sat Jan 31 01:34:12 2015 JenneUpdateLSCNot able to engage AO pathNothing earth-shattering today.

A few things of note:


I checked the coherence
 PRCL_CARM_Coupling_30Jan2015.pngPRCL_CARM_Coupling_lockloss_30Jan2015.png 
  10963   Mon Feb 2 12:24:27 2015 JenneUpdateASCPOP QPD centeredAfter aligning the PRC, I centered the
POP QPD.
  
  10967   Tue Feb 3 04:07:49 2015 JenneUpdateModern ControlMeasured PRM -> POP QPD TFsToday (after centering the POP QPD), I
measured the PRM to POP QPD transfer functions. 
I am suspicious that this was part of my
 PIT_measured_TF.pngYAW_measured_TF.png 
  10976   Wed Feb 4 19:21:45 2015 JenneUpdatePEMSeismometer Vertex is coveredI opened up the spaghetti pot over the
vertex seismometer, and taped the cable to
the slab.  The way the cable is coiled,
  
  10987   Sat Feb 7 21:30:45 2015 JenneUpdateLSCPRC alignedI'm leaving the PRC aligned and locked. 
Feel free to unlock it, or do whatever with
the IFO.
  
  10992   Tue Feb 10 02:40:54 2015 JenneUpdateLSCSome locking thoughts on PRMI[EricQ, Jenne]

We wanted to make the PRMI lock
more stable tonight, which would hopefully
  
  10993   Tue Feb 10 02:55:29 2015 JenneConfigurationModern ControlQuacky filtersI discovered that somehow my Wiener filters
that show up in Foton are not the same as
what I have in Matlab-land. 
 Mismatch_wienerQuackFoton_xaxis_9Feb2015.pngMismatch_wienerQuackFoton_yaxis_9Feb2015.pngMismatch_wienerQuackFoton_zaxis_9Feb2015.png 
  10999   Wed Feb 11 02:42:05 2015 JenneUpdateLSCPRC error signal RF spectraSince we're having trouble keeping
the PRC locked as we reduce the CARM offset,
and we saw that the POP22 power is significantly
 REFL33_25.pdfREFL33_15.pdf 
  11001   Wed Feb 11 04:08:53 2015 JenneUpdateLSCNew Locking Paradigm?[Rana, Jenne]

While meditating over what to do
about the fact that we can't seem to
  
  11007   Wed Feb 11 22:13:44 2015 JenneUpdateLSCNew Locking Paradigm - LSC model changesIn order to try out the new locking scheme
tonight, I have modified the LSC model. 
Screens have not yet been made.
  
  11011   Thu Feb 12 11:14:29 2015 JenneUpdateLSCNew Locking Paradigm - Loop-gebraI have calculated the response of this
new 2.5 loop system.

The first attachment is my block
 ALS_REFL_comboLockingCartoon_11Feb2015.PDFALS_REFL_comboLocking_11Feb2015.zip 
  11016   Thu Feb 12 19:18:49 2015 JenneUpdateLSCNew Locking Paradigm - Loop-gebraEDIT, JCD, 17Feb2015:  Updated loop
diagram and calculation: http://131.215.115.52:8080/40m/11043


Okay, Koji and I talked
  
  11017   Thu Feb 12 22:28:16 2015 JenneUpdateLSCNew Locking Paradigm - LSC model changes, screens modifiedI have modified the LSC trigger matrix
screen, as well as the LSC overview screen,
to reflect the modifications to the model
  
  11021   Fri Feb 13 03:44:56 2015 JenneUpdateLSCHeld using ALS for a while at "0" CARM offset with PRMI[Jenne, Rana]

We wanted to jump right in and
see if we were ready to try the new "ALS
 QPDsaturation_12Feb2015.pdfALSfool_12Feb2015.pdf 
  11024   Fri Feb 13 17:07:51 2015 JenneUpdateElectronicsSecond QPD Whitening Switch enabledI first updated the DCC branches for the
Xend and Yend to reflect the as-built situation
from December 2014, and then I updated the
  
  11030   Sat Feb 14 20:20:24 2015 JenneUpdateLSCALS fool cartoonThe ALS fool scheme is now diagrammed up
in OmniGraffle, including its new official
icon.  The mathematica notebook has
 ALSfool_LoopDiagram.pngALSfool_LoopDiagram.graffle.zip 
  11037   Mon Feb 16 02:49:57 2015 JenneUpdateLSCALS fool measured decoupling TFI have measured very, very carefully the
transfer function from pushing on
MC2 to the Yarm ALS beatnote. 
 ALSfool_measuredActuatorTF_YarmOnly_15Feb2015.pngYARM_actTF_compareActuators.pdf 
  11039   Mon Feb 16 15:08:26 2015 JenneUpdateLSCALS fool measured decoupling TFDang it, yes. You're right.  I
should have caught that. 

Since Dcpl and Srefl are both zero
  
  11042   Tue Feb 17 04:04:32 2015 JenneUpdateLSCALS fool mathI re-did the Mathematica notebook according
to the most current diagram (note to daytime
self: attach .nb file!!!), and found that
 OpenLoopGainComparison_16Feb2015.pngCancellationTFs_DecouplingIsArefl.png 
  11043   Tue Feb 17 16:36:08 2015 JenneUpdateLSCALS fool mathHere is an updated cartoon, with the ALS
sensor explicitly shown as the beatbox times
the closed loop response of the phase tracker
 ALSfool_LoopDiagram_17Feb2015.pngALS_REFL_comboLocking_16Feb2015.nb.zip 
  11050   Thu Feb 19 04:16:45 2015 JenneUpdateLSCALS fool attempt with PRFPMI[Jenne, EricQ]

We tried several times tonight
to engage the Fool path with the PRFPMI. 
 BEATYcrazy_duringCARMdarmLock_18Feb2015.png19Feb.png 
  11060   Tue Feb 24 01:57:01 2015 JenneUpdateLSCModelled effect of relative modulation phaseI have used Optickle to model the effect
of changing the phase between the 11MHz and
55MHz EOMs.  Also, I have looked at
 ChangeInAngle_PRCLvsMICH.pngREFL33_MICH_SensMat_vsModulationOrder_vsModulationPhase.pngREFL33_SensMat_vsModulationPhase.pngREFL33_PRCL_nominalVsLoweredGamma.pngREFL33_MICH_nominalVsLoweredGamma.png 
  11062   Tue Feb 24 23:39:16 2015 JenneUpdateLSCDelay line re-installed againAbout 5-10 minutes ago I just put in the
modulation cancellation setup according to
the recipe in http://nodus.ligo.caltech.edu:8080/40m/11032
  
  11064   Wed Feb 25 04:59:47 2015 JenneUpdateLSCDelay line re-installed, measurements round 2[Jenne, EricQ, Rana]

We spent this evening measuring
and thinking about our 3f signals, and the
 SensMat_PRMI_24Feb2015.pdf 
  11069   Wed Feb 25 14:51:13 2015 JenneUpdateLSCDelay line un-installed againAnd now I've removed the delay line,
and am in the process of reverting the demod
phases, etc.
  
  11070   Wed Feb 25 20:00:39 2015 JenneUpdateLSCDelay line un-installed again - sensing matrix comparisonsI have measured the sensing matrix for
the PRMI at the REFL photodiodes for both
the nominal configuration and the 33MHz cancellation
  
  11080   Thu Feb 26 20:02:46 2015 JenneUpdateLSCre: some thoughtsI have clarified my elog from last night
to indicate that the sensing matrix in the
"33MHz cancellation" configuration
  
  11082   Fri Feb 27 03:48:29 2015 JenneUpdateLSCPRMI locked during CARM scanOne of the things that I looked at tonight
was whether or not I could hold the PRMI
on REFL165 at CARM offset of 0, and it turns
 REFL33Lock.pngREFL165Lock.png 
  11086   Fri Feb 27 21:16:10 2015 JenneUpdateLSCPRMI locked during CARM scanHere are a bunch of PDFs of time series
from last night's CARM sweeps. 
The y-axes are all calibrated (except for
 8x 
  11088   Mon Mar 2 18:34:12 2015 JenneUpdateSUSETMX moving during single arm lockThis has been edited several times over
the last several hours, as I try to change
different parameters, to see if they affect
 ETMX_moving.pngETMXmoving_ALSlock.pngStartsMoving_stopsItself_ranASS.png80mHz.pngIndependFromFSStemp.png 
  11089   Tue Mar 3 02:43:06 2015 JenneOmnistructurePEMNo heat??It's super cold in the control room
and EE bench area tonight.  I'm
wondering if, similar to what happened on
  
  11090   Tue Mar 3 04:41:45 2015 JenneUpdateLSCTransitioned DARM to AS55Q, some other workBetter elog tomorrow - notes for now:

REFL165 for PRMI has been "a
champ" (quote from Q).  We're
  
  11091   Tue Mar 3 04:43:12 2015 JenneOmnistructurePEMOttavia is dead? After some searching, including help from
4 security guys (I think they don't have
a lot to do at 4:30am :), we found that 
  
  11095   Tue Mar 3 19:19:54 2015 JenneUpdateLSCTransitioned DARM to AS55Q, some other work[Jenne, EricQ]

A slightly more coherent elog for
last night's work.
 ALSfool_PRFPMI_2Mar2015.pdfALSfool_kindaEngaged_2March2015_noFrames.png 
  11097   Wed Mar 4 03:42:14 2015 JenneUpdateLSCTried a few CARM / ALS fool things, no successMuch of tonight was spent fighting with
ETMX.  This time, ASC was definitely
off, there was nothing coming out of the
 PRFPMI_DARMonAS55Q_3Mar2015.pdf 
  11104   Thu Mar 5 20:44:30 2015 JenneUpdateSUSdamprestore script updatedI just realized that the "damprestore"
script that can be called from the watchdog
screen did not have the new oplev names. 
  