ID Date Author Type Category Subject Text Attachments
  13672   Thu Mar 8 18:15:42 2018 gautamUpdateGeneralCDS recovery after work at LSC rackI was forced into a simultaneous power-cycle
rebooting of the three vertex FEs just now.
I took the opportunity to completely disconnect
  
  13671   Thu Mar 8 15:23:16 2018 gautamUpdateElectronicsNew DC power ports at c1lsc[Koji,
Gautam]
  
  13670   Thu Mar 8 14:41:25 2018 gautamUpdateGeneralCDS recovery after work at LSC rackAs I had found before,
restarting the c1oaf
 CDS-recovery.pngc1sus_timing.pngc1sus_crashed.png 
  13669   Thu Mar 8 01:10:22 2018 gautamUpdateGeneralCDS recovery after work at LSC rackThis required multiple hard reboots, but
seems like all the RT models are back for
now. The only indicator I can't explain
 CDS-recovery.png 
  13668   Thu Mar 8 00:40:25 2018 gautamUpdateALSnew look ALS electronics - characterizationI am almost ready for a digital test of
the new ALS electronics. Today, Koji and
I spent some time tapping new +/-24VDC DIN
 BeatMouthX_20180305_diffOut.pdf 
  13667   Wed Mar 7 12:04:14 2018 gautamUpdateElectronicsThree opamps walked onto an AA boardHere are the plots. Comments:


Measurement
and model agree quite well 
 AA_TF.pdfAA_noise.pdfD070081_LTspiceFiles.zip 
  13666   Mon Mar 5 17:27:34 2018 gautamUpdateALSnew look ALS electronics - characterizationI did a quick test of the noise of the
new ALS electronics with the X arm ALS. Attachment
#1 shows the results - but something looks
 BeatMouthX_20180305_diffOut.pdf 
  13665   Mon Mar 5 11:58:24 2018 gautamUpdateElectronicsThree opamps walked onto an AA boardFor testing the new IR ALS noise, we had
decided that we would like to use the differential
output of the demodulated ALS beat signal,
 AA_TF.pdfAA_noise.pdf 
  13664   Mon Mar 5 10:13:21 2018 gautamUpdateVACvacuum healthIn Steve's absence, I've tried
to keep an eye on the health of the vacuum
system. From Attachment #1, the pressure
 VacuumStatus.png 
  13663   Fri Mar 2 01:45:06 2018 gautamUpdateALSnew look ALS electronicsI spent today making another daughter board
(so that we can use the new scheme for I
and Q for one arm), testing it (i.e. measuring
 IMG_6916.JPG 
  13662   Wed Feb 28 21:14:34 2018 gautamSummaryPEMChannel adminSince we decided to use the Acromag for
readback of the temperature sensor for Kira's
seismometer temperature control, I enabled
 tempSensData.png 
  13661   Wed Feb 28 19:13:25 2018 gautamUpdateALSADC test for differential receiving in c1lsc[koji,
gautam]

we did a bunch of tests to figure
 c1lsc_ADCtest.png 
  13660   Wed Feb 28 12:31:28 2018 KiraUpdatePEMtemp sensor inputI switched out the DIN fuses for the long
cables and it fixed the issue of them
not showing any votage on the other end.
  
  13658   Tue Feb 27 21:10:45 2018 gautamUpdateALSDaughter board testingI thought a little bit about the next steps
in testing the daughter board. The idea is
to install this into the existing 1U chassis
 IMG_6912.JPGIMG_6913.JPGIMG_6914.JPG 
  13657   Mon Feb 26 20:55:56 2018 ranaUpdateALSDaughter board prototypingLooks good.

* for bypass type applications,
you don't have to use Wima caps (which
  
  13656   Mon Feb 26 16:22:10 2018 KiraUpdatePEMtemp sensor input[Kira, Gautam]

We began the setup for the lab
temperature sensor today. First, we needed
 1.jpgIMG_20180226_154649.jpgIMG_20180226_154720.jpg 
  13655   Sun Feb 25 00:03:12 2018 gautamUpdateALSDaughter board prototypingUsing one of the prototype  schematic.pdfdaughterBoard_TF.pdfdaughterBoard_noise.pdfTEK00000.PNGdaughterBoard_noise.pdf 
  13654   Fri Feb 23 20:46:04 2018 Udit KhandelwalSummaryGeneralCAD Summary 2018/02/23I have more or less finished cadding the
test mass chamber by referring to the drawings
Steve gave me. Finer details like lugs and
 tmc_20180223.JPG 
  13653   Fri Feb 23 07:47:54 2018 SteveUpdateVACCC1 HornetWe have the IFO pressure logged again!
Thanks Johannes and Gautam

This
 InstruTech_Hornet_CC1.pngCC1_readback_updated.png 
  13652   Thu Feb 22 17:19:47 2018 KojiUpdateGeneralModulation depth measurement for an aLIGO EOMaLIGO
EOM test: Setup


The modulation signal was
 IMG_3436.JPGmodulation_depth.pdfmodulation_linearity.pdfmodulation_24MHz.pdf 
  13651   Thu Feb 22 16:16:43 2018 KiraUpdatePEMtemp sensor inputRewired the temperature sensor inputs to
Molex connectors so that we can now attach
them to the +/- 15V Sorensens for input instead
 IMG_20180222_160602.jpg 
  13650   Thu Feb 22 16:11:14 2018 KojiUpdateGeneralaLIGO EOM crystal replacementaLIGO
EOM crystal replacement


The entire operation has
 IMG_3421.JPGIMG_3426.JPGIMG_3427.JPG 
  13649   Thu Feb 22 10:49:11 2018 SteveUpdateElectronicsrack power supplies checkedAll rack power supplies labeled if their
load changed.

 
 8x 
  13648   Thu Feb 22 00:09:11 2018 gautamUpdateALSD0902745 in-situ testingI thought a little bit about the design
of the preamp we want for the demodulated
ALS signals today. The requirements are:
 daughterBoard_TF.pdfdaughterBoard_noise.pdfIMG_6906.JPG 
  13647   Wed Feb 21 17:20:32 2018 johannesUpdateVACHornet gauge connected to DAQ.I wired the six available BNC connectors
on the front panel of the new XEND slow DAQ
to physical Acromag channels. There were
 VacLog.png 
  13646   Wed Feb 21 12:17:04 2018 gautamUpdateCDSLO Power mon channels added to c1lscTo make this setup more permanent, I modified
the c1lsc
  
  13645   Wed Feb 21 00:04:27 2018 gautamUpdateElectronicsTemporary RF power monitor setupI made a voltage divider using a 20.47kohm
and 1.07kohm (both values measured with a
DMM). The whole thing is packaged inside
 RFbeatAmp.pngBeatMouthX_RFAM_20180221.pdf 
  13644   Tue Feb 20 23:08:27 2018 gautamUpdateALSD0902745 in-situ testingAttachment #1 shows
the ALS noise measurement today. Main differences
from the spectrum posted last week is that
 BeatMouthX_20180220_diffOut.pdf 
  13643   Tue Feb 20 21:14:59 2018 gautamUpdateCDSRFM network errorsI wanted to lock the single arm POX/POY
config to do some tests on the BeatMouth.
But I was unable to.
  
  13642   Tue Feb 20 13:59:30 2018 KojiUpdateGeneralModulation depth measurement for an aLIGO EOMLast night I worked at the PSL table for
the modulation depth measurement for an aLIGO
EOM. Let me know if the IFO behavior is unusual.
  
  13641   Mon Feb 19 14:27:25 2018 gautamUpdateGeneralFibel ALS input polarization tuningSummary:

Current configuration of PSL free-space
to fiber coupling is:
 IMG_6900.JPG 
  13640   Fri Feb 16 22:19:07 2018 gautamUpdateGeneralFibel ALS input polarization tuningAfter discussing with Koji, I decided to
try and align the input beam polarization
at the PSL fiber coupler to one of the special
 PER_setup.JPG 
  13639   Fri Feb 16 22:15:30 2018 gautamUpdateGeneralc1mcs model restartedc1mcs
had died for some reason. Looking at dmesg,
  
  13638   Fri Feb 16 21:03:17 2018 Udit KhandelwalSummaryGeneralSummary 2018/02/1640m Lab Cad:
Updated the dimensions of and fleshed
out the chambers in greater detail, by referring
  
  13637   Fri Feb 16 15:57:58 2018 SteveUpdateVACTP3 drypump replacedThe forline pressure of TP3 was 399 mTorr

It was replaced this morning at 
TP3 controller 134,638hrs with the "failed
  
  13636   Fri Feb 16 01:34:40 2018 gautamUpdateALSD0902745 in-situ testingHaving implemented the changes to the audio
amplifier stage, I re-installed this unit
at the LSC rack, and did some testing. 
 BeatMouthX_20180216.pdfIMG_5134.JPG 
  13635   Fri Feb 16 01:09:55 2018 gautamUpdateALSEX green locking duty cycleI have been puzzled as to why the duty
cycle of the EX
green locks are much
 GreenLock8hrs.png 
  13634   Thu Feb 15 16:03:57 2018 KiraUpdatePEMPID test planI checked channels 6 and 7 on the ADC and
they have long wires leading to BNC ends
and are currently not being used, so we could
  
  13633   Wed Feb 14 17:49:22 2018 gautamUpdateElectronicsPSL table power supply cleanup[steve,
gautam]
 IMG_5135.JPGSorensens_1X1_before.JPGSorensens_1X1_after.JPG 
  13632   Tue Feb 13 22:35:21 2018 gautamUpdateElectronicsPSL table power supply cleanupThe main motivation
for this work is that I want +15VDC power
  
  13631   Tue Feb 13 21:22:44 2018 SteveUpdateSEIone load cells testedGautam and Steve,

The "called 225 lbs"
steel crane load measured right on 102 kg
 as_measured_102kg.jpgsensor.jpg1500lbs_load__cell.jpg 
  13630   Mon Feb 12 14:56:00 2018 SteveUpdatesafety crane inspection 2018Our 3 cranes passed  professional 
inspection. Fred Goodbar of Konacrane with
450 lbs load at full extension.
 ETMY.jpgVetex.jpgETMX.jpg 
  13629   Fri Feb 9 15:29:32 2018 KiraUpdatePEMPID test plan[Kira, Steve]

We installed and labeled the
Sorensens today.
 IMG_20180209_152158.jpg 
  13628   Fri Feb 9 13:37:44 2018 gautamUpdateALSTHD measurement trialI quickly put together some code that calculates
the THD from CDS data and generates a plot
(see e.g. Attachment #1).
 THD_trial.pdf 
  13627   Thu Feb 8 18:10:36 2018 gautamUpdateALSD990694 pulled outThis is proving much more challenging than
I thought - while Cut #1 was easy to identify
and execute, my initial plan for Cut #2 seems
  
  13626   Thu Feb 8 17:32:44 2018 KiraUpdatePEMPID test plan[Kira, Steve]

We set up a new rail for the Sorensens
(attachment 1) and placed one of them down
 IMG_20180208_171423.jpgIMG_20180208_172107.jpgIMG_20180208_171853.jpgIMG_20180208_171932.jpg 
  13625   Thu Feb 8 13:13:14 2018 gautamUpdateALSD990694 pulled outAfter labeling all cables, I pulled out
one of the D990694s in the LSC rack (the
one used for the ALS X signals, it is Rev-B1,
 IMG_5131.JPGIMG_5133.JPGD990694-B_cuttingPlan.pdf 
  13624   Thu Feb 8 12:24:37 2018 KiraUpdatePEMPID test planSome points before we can set up the can:


Cable length and type

For
the DAC, we can use the LEMO outputs and
  
  13623   Thu Feb 8 12:00:09 2018 gautamUpdateALSD990694 is NOT differential receivingCorrecting a mistake   
  13622   Thu Feb 8 01:27:16 2018 KojiUpdateALSD990694 characterization / THD measurement plan>
So my question is - should we just cut the
  