Akhil | Summary | Electronics | Timing Issues of Mini Circuits UFC-6000: Solved |
Main Problem:
The frequency counter (FC) takes in an analog RF input(signal) and outputs the frequency of the signal(Ranging from 1 MHz- 6000 MHz) in the digital domain (into a processor). The FC samples the data with a given sample rate( user defined) which ranges from 0.1 s to 1 s(faced problems in fixing this initially). For data acquisition, we have been using a Raspberry Pi(as a processor) which is connected to the martian network and can communicate with the computers inside the 40m. The ultimate challenge which I faced( and been knocking my head off from past two-three weeks) is the synchronization of clocks between the Raspberry Pi and the FC i.e the clock which the FC uses to sample and dump data( every 'x' s) and the clock inside the raspberry pi( used in the loop to wait for a particular amount of time the frequency counter takes to dump successive data).
Steps Taken:
- To address this problem, first I added an external clock circuit which monitors the Raspberry Pi and the FC to dump and read data at a particular rate(which is equal to the sampling rate of the FC)In detail: http://nodus.ligo.caltech.edu:8080/40m/10129.
- While doing so, at first the level trigger algorithm was used which means that the external clock frequency was half as that of the reciprocal of the sampling rate and a trigger was seen every time the level shifts from +DC to -DC(of the external square wave).
- But this did not completely mitigate the issues and there were still few issues on how quickly the ADC reads the signal and R Pi processes it.
- To minimize these issues completely, an edge trigger algorithm which detects a pos edge(rising) of the clock was used. The clock frequency is now equal to the reciprocal of the sampling rate. This algorithm showed better results and greatly minimized the drift of the sampling time.
Psuedo Code(code attached):
open device : FC via USB-HID;
open device : ADC via I2C;
always(for t= recording time):
read data from ADC(external clock);
if pos edge detected:
read data from FC and store it in a register;
else read data from ADC;
end
write data stored in the register to a file( can be an Epics channel or a text file);
Results:
The attached are the plots showing the time between samples for a large number of samples taken for different sampling times of the FC. The percentage error is the percentage of standard error in the timing between two samples for the data for the entire measurement. It can be inferred that this error has been cut down to the order of ms.
To do next:
- I have started taking phase measurements( analysis and plots will follow this elog) and also the PSD plots with the improved timing characteristics.
|
Attachment 1: 0.2timinganalysis.png
|
|
Attachment 2: 0.3timinganalysis.png
|
|
Attachment 3: 0.5timinganalysis.png
|
|
Attachment 4: 1stiminganalysis.png
|
|
Attachment 5: pdf.zip
|
10194
|
Mon Jul 14 14:28:27 2014 |
Koji | Summary | Electronics | Timing Issues of Mini Circuits UFC-6000: Solved |
Looks good. Now you have the internal timer to verify the external clock.
If you can realize the constant rate sampling without employing the external clock, that's quite handy. |
10199
|
Tue Jul 15 01:31:13 2014 |
Akhil | Summary | Electronics | Timing Issues of Mini Circuits UFC-6000: Solved |
The attached are the PSD plots with improved FC timing(with the same code as in http://nodus.ligo.caltech.edu:8080/40m/10151). More plots(Phase and PSD) to follow.
|
Attachment 1: qnoise.png
|
|
Attachment 2: qnoise.pdf
|
|
10209
|
Wed Jul 16 01:18:02 2014 |
rana | Summary | LSC | Python Wavelet peak finding for dramatic ALS - Red Resonance finding speedup |
New script called scripts/ALS/findRedResonance.py finds the IR resonance in less than 20 seconds.
- Do a ~2 FSR scan of the arm over 10 seconds using the Phase Tracker servo offset ramp. (dt = 10 s)
- Load the frame data for the TR_OUT and the Phase Tracker phase. (dt = 2 s)
- Use Python (modern) SciPy to find all peaks with moderate SNR (dt = 3 s)
- Take the top 3 peaks (all presumably TEM00) and choose the one closest to zero and go there.
The above is the gist of what goes on, but there were several issues to overcome to get the code to run.
- None of our workstations have a modern enough Python distro to run either the ipython notebooks or the new SciPy with wavelets, so I installed Anaconda Python in the controls home directory.
- In order to get the peak finder function to work well I gave it a range of peak widths to look for. If we change the speed of the ALS scan by more than 3x, we probably have to change the width array in the function.
- I've used the find_peaks_cwt function in SciPy. This uses a Ricker ('Mexican Hat') wavelet to find peaks by doing multi-res transforms and looking for ridges in the wavelet space (I think)
- To make the code run fast, I downsample from 16 kHz to 1 kHz right at the top. Would be better if we had downsampling in v2 of the getData function.
|
Attachment 1: findRedResonance.pdf
|
|
10211
|
Wed Jul 16 01:35:16 2014 |
Koji | Summary | LSC | Python Wavelet peak finding for dramatic ALS - Red Resonance finding speedup |
From the last plot:
- Subtracting the offset of 0.0095, the modulation depth were estimated to be 0.20 for 11MHz, 0.25 for 55MHz
- Carrier TEM00 1.0, 1st order 0.01, 2nd order 0.05, 3rd order 0.002, 4th order 0.004
==> mode matching ~93%, dominat higher order is the 2nd order (5%).
Eric: now we have the number for the mode matching. How much did the cavity round-trip loss be using this number? |
10219
|
Wed Jul 16 19:38:37 2014 |
manasa | Summary | PSL | AOM alignment issues and removed from beam path |
AOM removed from the beampath and PMC relocked.
AOM alignment:
1. Measured the initial power after PMC as 1.30W and reduced it down to 130mW.
2. Checked the power in the AOM zero order transmission before touching it. For 0-1V modulation input, the power dropped from 125uW to 98.3uW.
3. Steered the mirror right before the AOM to increase AOM zero order transmission and then carefully moved the AOM around to obtain maximum power attenuation. I repeated this a few times and the maximum attenuation that I could obtain was 125uW to 89.2uW (~30% attenuation).
Although this is not the right way to align the AOM, we do not have much options with the current setup as there is not enough separation between the zero order and first order beams and the AOM is on a fixed rigid mount.
4. I tried to dump the first order beam from the AOM and it wasn't satisfactory as well. There is barely any separation between the zero order and first order beams.
PMC relocking:
1. SInce the alignment to the PMC was disturbed by moving the AOM and the steering mirror in front of it, the PMC alignment was lost.
2. I could not relock the PMC at low power or high power. Rana had to come to rescue and fixed the alignment so that we could see flashes of PMC on the trans camera (This was done by aligning refl beam to the PMC REFL PD while giving a triangular ramp to the PMC PZT voltage).
Also I should not have tried to lock the PMC at high power as I could have been steering the beam at high power to the edges of the PMC mirrors that way and burning stuff easily.
3. Before fine tuning the alignment, I decided to remove the AOM from the beam path as there needs some work done on it to make it useful.
4. I removed the AOM from the beam path and relocked the PMC.
5. PMC is relocked with 0.79 counts in TRANS and I measured the power after PMC 1.30W
Attachment: picture showing AOM removed from the beampath. |
Attachment 1: AOMremoved.jpg
|
|
10222
|
Wed Jul 16 22:17:40 2014 |
Akhil | Summary | Electronics | Bode Plots and complete Characterization of Frequency Counter |
Goal:
To estimate the transfer function and the noise in the FC that is a part of the FOL-PID loop.
Measurements Taken:
The setup used for the measurements is described in my previous elogs.
The input modulation signal and the FC output were recorded simultaneously for a certain period of time and the phase and gain are estimated from the data.
Analysis(Data and code attached):
The recordings must contain equal number of data points(around 6000 data points in my measurements) for analysis.
The steps I followed to generate these plots are:
- Took the FFT of both FC out data(from FC) and Modulation input(from SRS via ADC).
- Estimated the phase angles at the particular modulation frequencies from the FFT data(in Matlab using angle(x) for phase at the frequency f(x);x: is the frequency bin)
- Then for the phase of the system at a particular modulation frequency,
Phase(system) =Phase(FC Signal) - Phase(Input Signal)
- Plotted the acquired phase vs the modulation frequency on a Semi-log graph.
Results:
From the plots its can be inferred that :the delay of the FC is almost 0 until the modulation of 0.1 Hz. Then there are phase shifts of +/- 180 degrees showing that the system has multiple poles and zeroes(will be estimated after I have phase plots at few more carrier frequencies).
To Do Next:
Phase plots for varying carrier frequencies and different sampling times.
Installation of FC inside the 40m. |
Attachment 1: Phase_Data.zip
|
Attachment 2: Bode100MHz.png
|
|
10223
|
Wed Jul 16 23:02:16 2014 |
Koji | Summary | Electronics | Bode Plots and complete Characterization of Frequency Counter |
If I assume 1sample delay for 0.1s sampling rate, the delay is Exp[-I 2 pi f T], where T is the sampling period.
This means that you expect only 36 deg phase delay at 1Hz. In reality, it's 90deg. Huge!
Also there are suspicious zeros at ~1.6Hz and ~3Hz. This may suggest that the freq counter is doing some
internal averaging like a moving average.
It would be interesting to apply a theoretical curve on the plot. It's an intellectual puzzle. |
10238
|
Fri Jul 18 17:10:57 2014 |
Nichin | Summary | Electronics | Characterization of demodulator boards. |
Rack 1Y2, I took transfer function measurements for each of the following demodulator boards: REFL11, REFL33, REFL55, REFL165, AS55, POP22, POX11 and POY11.
What I did:
1) Removed the wire at PD Input to demodulator board.
2) Put the MOD output from network analyzer into PD input of board.
3) Ran a sweep from 100kHz to 100MHz.
4) Measured the transfer function between PD RF MON and PD Input. (The PD RF MON signal came out of the RF multiplexer, so the mux is included as well )
5) Put the original wire back at PD Input.
Results:
The plots clearly show an attenuation of 20dB (factor of 10) for all the demodulator boards. This explains why my transimpedance measurements are off by 10 times.
Note: for REFL 165, there was an extra 100MHz high pass filter installed at PD Input. I did not remove this and made my measurements along with this.
To Do:
a) Modify the PDFR system to calibrate out this attenuation.
b) Measure the transfer function between the input and output of RF mux, so that we can have just the transfer function between PD input an PD RF MON (for documentation's sake)
|
Attachment 1: Demodulators_TF.pdf
|
|
10239
|
Fri Jul 18 19:32:50 2014 |
Akhil | Summary | Electronics | Filters used inside the Frequency Counter |
Thanks Koji , for your hint for the brain teasing puzzle. I was looking into Filters that are usually used in devices like counters, DSO and other scopes. I found that , to improve the quality of the measurement one of the best approach is averaging. I looked deeper into averaging and found out this:
There are two general use-cases for averaging . The first, successive sample averaging, takes a single acquisition and averages between its samples. The second, successive capture averaging, combines the corresponding samples of multiple captures to create a single capture. Successive sample averaging is also called boxcar filtering or moving average filtering. In an implementation of this type of averaging each output sample represents the average value of M consecutive input samples. This type of averaging removes noise (one of the reasons the noise level was not bad: http://nodus.ligo.caltech.edu:8080/40m/10151) by decreasing the device's bandwidth(could be one of the reasons why the FC operates in 4 different frequency ranges). It applies an LPF function with a 3dB point approximated by 0.433 * s / M, where M is the number of samples to be averaged, and s is the sample rate in samples per second.
Now I tried verifying the 3 dB points in the gain plots I generated :
For 1 s Sampling time : the 3 dB point for such a Boxcar filter should be at 0.433* 1/M. If we assume that it averages for 2 samples, M=2 which gives the 3dB point at 0.288 Hz but occurs somewhere between 0.3 and 0.4 Hz. (http://nodus.ligo.caltech.edu:8080/40m/140619_120548/GainVsFreq.png)
For 0.1s Sampling time: the 3dB point should be at 2.17 Hz and in reality is 2.5 Hz(http://nodus.ligo.caltech.edu:8080/40m/140701_211904/gain.png).
Also, This type of filter will have very sharp nulls at frequencies corresponding to signals whose periods are integer sub-multiples of M/s. As seen my previous plots (http://nodus.ligo.caltech.edu:8080/40m/10118 , http://nodus.ligo.caltech.edu:8080/40m/10070) there are sharp nulls at frequencies
0.4 Hz for 1S sampling time and
at 1.5 Hz,3 Hz for 0.1 S sampling time as correctly predicted.
The moving average filter is L-sample moving average FIR, with the frequency response as: H(ω) = (1/L) (1 − e− jω L)/(1 − e− jω)..
There is an overall delay of (M - 1)/2 samples from such a length-M causal FIR filter.
The expected bode plots for such a filter with L= 5 is attached(attachment 2). |
Attachment 1: TheoreticalGainPlot.png
|
|
Attachment 2: TFexpected.png
|
|
10246
|
Mon Jul 21 12:16:27 2014 |
Akhil | Summary | Electronics | Filters used inside the Frequency Counter |
The expected bode plots for such a filter with L= 4 is attached and compared with the measured.
RXA: When comparing two things, please put them onto the same plot so that they can be compared. |
Attachment 1: FC_TF_Characterization.png
|
|
10248
|
Mon Jul 21 17:32:43 2014 |
ericq | Summary | LSC | Arm losses |
Quote: |
From the last plot:
- Subtracting the offset of 0.0095, the modulation depth were estimated to be 0.20 for 11MHz, 0.25 for 55MHz
- Carrier TEM00 1.0, 1st order 0.01, 2nd order 0.05, 3rd order 0.002, 4th order 0.004
==> mode matching ~93%, dominat higher order is the 2nd order (5%).
Eric: now we have the number for the mode matching. How much did the cavity round-trip loss be using this number?
|
Using these numbers for both arms (Modulation takes away .2*.25 = 5% power, mode matching takes away 7% after that), I get the following from my data from March:
Xarm loss is 561.19 +/- 14.57 ppm
Yarm loss is 130.67 +/- 18.97 ppm
Obviously, the Xarm number looks very fishy, but its behavior was qualitatively very different when I took the data. ASDC would change from ~0.298 to ~0.306 when the Yarm was locked vs. misaligned, whereas the xarm numbers were .240 to .275.
In any case, I'll do the measurement again tomorrow, being careful with offsets and alignment; it won't take too long. |
10252
|
Tue Jul 22 15:50:35 2014 |
Nichin | Summary | Electronics | Characterization of demodulator boards. |
Quote: |
Rack 1Y2, I took transfer function measurements for each of the following demodulator boards: REFL11, REFL33, REFL55, REFL165, AS55, POP22, POX11 and POY11.
What I did:
1) Removed the wire at PD Input to demodulator board.
2) Put the MOD output from network analyzer into PD input of board.
3) Ran a sweep from 100kHz to 100MHz.
4) Measured the transfer function between PD RF MON and PD Input. (The PD RF MON signal came out of the RF multiplexer, so the mux is included as well )
5) Put the original wire back at PD Input.
Results:
The plots clearly show an attenuation of 20dB (factor of 10) for all the demodulator boards. This explains why my transimpedance measurements are off by 10 times.
Note: for REFL 165, there was an extra 100MHz high pass filter installed at PD Input. I did not remove this and made my measurements along with this.
To Do:
a) Modify the PDFR system to calibrate out this attenuation.
b) Measure the transfer function between the input and output of RF mux, so that we can have just the transfer function between PD input an PD RF MON (for documentation's sake)
|
I repeated the exact steps above and made sure everything was back where it should be after I was done.
Reason I had to retake the measurements:
My script for acquiring data from the AG4395A network analyzer was such that it first acquired the magnitude data from channel 1 and then recorded phase data from channel 2 without holding its trace. Hence the phase and magnitude data were not exactly in sync with each other. So, when I tried to fit the data to a model using vector fitting, I ended up with very bad results.
I have now changed every single script relating to the network analyzer to just get the real and imaginary data in one go and then calculate the phase using this data.
The fitting process is now in progress and results will be up shortly. |
Attachment 1: Demodulators_TF.pdf
|
|
10298
|
Wed Jul 30 15:33:48 2014 |
Akhil | Summary | General | Calibrated Thermal Actuator TFs |
The goal of the measurements we made ( my previous 3 elogs) was to characterize the laser frequency thermal actuator that is a part of the FOL- PID loop.
For this we made indirect TF measurements for the thermal actuator by looking at the PZT response by 1)arm cavity( ETM ,ITM) displacement and 2) temperature offset excitation. The goal was to do something like getting G1=TF3/TF1 and G2=TF3/TF2 and ultimately dividing G2/G1 to get TF2/TF1 with correct calibration. The final TFs obtained are the X and Y arm TFs for Laser frequency response vs temperature offset in(HZ/count). The calculations in detail are:
Obtained G1 = PZT response/ Temperature Offset (count/count): (in detail here )
Obtained G2 = PZT response/ X and Y arm displacement( count/ count) : (in detail here)
Calibrated G2 to count/m ( in detail here)
Divided G2/G1 to get X and Y arm displacement/ Temperature Offset( m/ count) to get G3
Did these calculations:
dL/ L = dF /F
F = c/lambda ;Lambda = 532 nm ; L =
X arm length = 37.79 +/- 0.05 m
Y arm length = 37.81 +/- 0.01 m
TF: Laser Freq/ Temperature Offset = G3 *F/L (HZ/Count)
The calibration coefficients for the ends are :
X End: [23.04 +/- 0.23 ]* 10^3 (HZ/Count)
Y End: [18.71 +/- 0.2 ]* 10^3 (HZ/Count)
For the TFs of the temperature actuator on laser frequency I used ITMs for both the arms. The bode plots for the calibrated( HZ/Temp Count) are attached.
For the X-Arm Thermal Actuator, I calculated the TFs at two different frequency ranges and combined the results where the coherence is high(>0.7). At 1 Hz the coherence was not as good as the other frequencies(due to the suspension resonance at 0.977 Hz).
The poles and zeroes are estimated after fitting this data using Matlab vectfit tool.The graphs showing fit and measured values are attached.
Y arm Thermal Actuator:
5th order TF fitted:
Gain: 9000
Zeroes:
z1 = -0.9799;
z2 = 2.1655;
z3 = -2.9746- i * 3.7697
z4 = -2.9746+ i * 3.7697
z5 = 95.7703 + 0.0000i
Poles:
p1 = -0.0985- i* -0.0845
p2 = -0.0985+ i* -0.0845
p3 = -0.6673- i* -0.7084
p4 = -0.6673+ i* -0.7084
p5 = -8.7979.
X-arm Thermal Actuator:
5th order TF fitted:
Gain = 20
Zeroes:
z1= -305.7766
z2 = -18.2774
z3 = -16.6167
z4 = -1.2486
z5 = 28.1080
Poles:
p1 = -0.1311 - 0.1287i
p2 = -0.1311 + 0.1287i
p3 = -8.3797 + 0.0000i
p4 = -4.0588 - 7.5613i
p5 = -4.0588 + 7.5613i
I will use get the poles and zeroes from these fitted bode plots and use it to build the PID loop.
|
Attachment 1: Y_Arm_TA_TF.pdf
|
|
Attachment 2: X_Arm_TA_TF.pdf
|
|
Attachment 3: Y_Arm_TA_with_fit.pdf
|
|
Attachment 4: X_Arm_TA_with_fit.pdf
|
|
10304
|
Thu Jul 31 11:54:54 2014 |
Akhil | Summary | Electronics | PZT Calibration |
Koji asked me to get the calibration of the PZT counts to Volts for the the X and Y ends. Yesterday, I went inside the lab and took some measurements from the digital readout of the PZT by giving in a DC offset(-5 to +5 volts) to PZT_Out and read out from these channels:
For X-end: C1:ALS-X-SLOW_SERVO1_IN1
For Y-end: C1:ALS-Y-SLOW_SERVO1_IN1
Since a 20dB attenuator was placed in the path of X-arm readout while taking the Transfer functions(Detail), I did the calibration measurements without removing it from the path. However, for the Y arm there was no attenuator in the readout path.
The obtained calibration values are :
X- arm PZT : [146.3 +/- 2.37 ] counts/Volt
Y- arm PZT : [ 755.1 +/- 3.6] counts/Volt
The attached are the fit and data plots for the above calibration. |
Attachment 1: PZT_Y_Calibration.pdf
|
|
Attachment 2: PZT_X_Calibration.pdf
|
|
10306
|
Thu Jul 31 12:23:38 2014 |
Koji | Summary | Electronics | PZT Calibration |
1) Don't be brainless. Redo the fitting of the Y arm. Obviously the fit is not good.
2) How can you explain the value from the ADC bit and range?
e.g. +/-10V range 16bit ADC => 2^16/20 = 3276.8 count/V |
10307
|
Thu Jul 31 14:23:28 2014 |
Akhil | Summary | Electronics | PZT Calibration |
The PZT seems to saturate at around +/- 3500 counts. So for the Y arm, I excluded the saturated points and fitted the data points again.
As for the calibration number, we expect the 3276.8 count/V for +/- 10 V range of a 16 bit ADC but the number is ~800 count/V. I couldn't figure out a reason why the number is so different.
The new calibration values are :
X- arm PZT : [146.3 +/- 2.37 ] counts/Volt (with a 20 dB attenuator included in the path)
Y- arm PZT : [ 797 +/- 3.6] counts/Volt
I will get the calibration in MHz/V of PZT actuation and check whether these numbers make any sense. |
Attachment 1: PZT_Y_Calibration.pdf
|
|
10317
|
Fri Aug 1 01:57:24 2014 |
Koji | Summary | IOO | MC auto locker |
To make MC auto locker running correctly, mcdown and mcup were revised
I tried it by unlocking MC several times. It seems OK. Let's see how it works.
Nominal gains for locking (to be taken care by mcdown)
C1:IOO-MC_REFL_GAIN
was 16 and is 19 now.
C1:IOO-MC_VCO_GAIN
was 9 and is 9 now too.
C1:PSL-FSS_MGAIN
was missing and now +13
C1:PSL-FSS_FASTGAIN
was +23.5 and is now +20.0
Nominal gains for operation ( to be taken care by mcup.
C1:IOO-MC_REFL_GAIN
was 19 and is 19 now too.
C1:IOO-MC_VCO_GAIN
was 25 and now uses ezcastep (ezcastep C1:IOO-MC_VCO_GAIN=9 +1,16 -s 0.1)
C1:PSL-FSS_MGAIN
C1:PSL-FSS_FASTGAIN
ezcawrite C1:PSL-FSS_MGAIN `ezcaread -n C1:PSL-STAT_FSS_NOM_C_GAIN`
ezcawrite C1:PSL-FSS_FASTGAIN `ezcaread -n C1:PSL-STAT_FSS_NOM_F_GAIN`
C1:PSL-STAT_FSS_NOM_C_GAIN` is +18
C1:PSL-STAT_FSS_NOM_F_GAIN` is +20 |
10318
|
Fri Aug 1 03:49:26 2014 |
Koji | Summary | General | Koji - to do |
- Put the circuit diagram of the sum amp on/in the circuit enclosure and associate it with an elog [done].
- Update the circuit diagram of the pomona box [done]
ALL DONE
|
10319
|
Fri Aug 1 08:55:34 2014 |
Koji | Summary | IOO | MC auto locker |
It seems that the MC auto locker and the FSSSlow PID servo survived a night.
PC Drive is still angry occasionally. We want to know what this is. |
Attachment 1: MC.png
|
|
10320
|
Fri Aug 1 10:40:48 2014 |
Koji | Summary | IOO | MC servo summing amp |
The summing amp is prepared to allow up to use bipolar full range of the FSS box output
This means that the FSS fast PZT output is now nominally 0V and can range +/-10V.
- FSS Box has the output range of +/-10V
- Thorlabs HV amp MDT694 accepts 0V ~ +10V
- This circuit add an offset of +5V while the main signal is attenuated by a factor of 2. The offset voltage is produced from the voltage reference IC AD586.
- In addition, a summing node and voltage monitors before and after the summing node are provided. They are useful to test the crossover frequency of the fast/PC loops.
- The output noise level at 10kHz was ~60 nV/rtHz. The transfer function of the circuit was measured and flat up to 100kHz. The phase delay is negligible at 10kHz and less than 3deg at 100kHz
- Although the schematic was drawn in Altium, the board is a universal 1U eurocard and all wires were hand soldered. |
Attachment 1: Fast_PZT_IF.PDF
|
|
10322
|
Fri Aug 1 12:49:06 2014 |
Koji | Summary | IOO | MC servo analysis |
Reasoning to choose the current parameters:
FSS Common: 18dB
FSS Fast: 20dB
Attachment 1:
Openloop transfer function of the IMC loop with the nominal gain setting. The UGF is 176kHz and the phase margin is 48 deg.
This is about 3 time more bandwidth than the previous setting. (Good)
It is visible that the TF has sharp roll off around 1MHz. I wonder if this comes from the demodboard LPF and/or the PMC cav pole.
In fact, according to Manasa, the PMC has the ringdown of 164.6ns which corresponds to the cavity pole of 967kHz. So this must
be there in the OLTF.
From the plot, the order of the low pass is about 5. Subtracting the slope by the cavity pole, the order is four. If I look at the TF of the minicircuits
LPFs (this entry), the phase delay of the filter at 1/10 of the cut off freq is ~30deg. And the order of the filters are maybe 6th elliptic?
So it's not yet clear if the LPF is causing a significant phase delay at 180kHz.
More significantly, the gain margin at ~1MHz is way too small. This is causing a big servo bump at that frequency as seen in Attachment 2.
In total, my recommendation is to move the LPF freq up by x2 or x3, and give a mild LPF above 500kHz.
This requires some modeling as well as try and error.
Attachment 2:
This figure is to explain how the common FSS gain was set. By increasing the gain, the UGF is increased and we can enjoy more supression (from red to purple).
The more gain, however, the more servo bump we observe above the UGF. The gain was chosen so that the total PC feedback does not exceed 3V.
Attachment 3/4:
This figure explains how the fast FSS gain (namely crossover frequency between fast and PC) was set. When the fast is low (red) the phase margin between two loops
are plenty and therefore the openloop TF is smooth. But the PC's frequency domain is large and has to work more (in rms). As the fast gain is increased, the actuation
by the PC is offloaded to the fast PZT (that's good). But eventually the phase margin is not enough and the dip start to show up (purple). This dip cause worse closed loop TF,
as seen in Attachment 4, or even an instability of the loop eventually. So the fast gain was set somewhere in between (green). |
Attachment 1: MC_OLTF.pdf
|
|
Attachment 2: MC_Error_Common.pdf
|
|
Attachment 3: MC_Crossover.pdf
|
|
Attachment 4: MC_CLTF_Fast.pdf
|
|
10324
|
Fri Aug 1 18:48:46 2014 |
Akhil | Summary | Electronics | PZT Calibration |
The PZT actuation on the laser frequency in MHz/V ( assuming the previous calibration here of the PZT count/V) is :
X- arm: 33.7 MHz/V
Y- arm: 14.59 MHz/V
This number seems to be wrong by a factor of 10.
So we[I and EricQ] decided to trace the cables that run into the ADC from the PZT Out. We found a black LEMO box in the path to ADC,which is an anti-aliasing filter for each input channel. However,in theory the response of this filter should be flat up until a few kHz i.e. for the DC gain it should be 1. But we will manually test it and look at the DC gain of the LEMO box.
|
10343
|
Thu Aug 7 11:57:59 2014 |
Koji | Summary | IOO | MC servo analysis |
LISO Fit for the IMC open loop TF. The data and liso source for the fitting were attached in the ZIP file.
I noticed now that the open loop TF I measured has too less phase delay.
I used the closed loop TF to estimate the openloop TF.
Looking at this comparison, I'm afraid that the superboost was not on during the measurement.
I need a new measurement to design MC loop modification to give the AO path for broader bandwidth. |
Attachment 1: MC_OLTF_Fit.pdf
|
|
Attachment 2: IMC_OLTF.zip
|
Attachment 3: MC_OLTF_estimated.pdf
|
|
10350
|
Fri Aug 8 11:22:35 2014 |
steve | Summary | SUS | oplev laser summary |
2005 ALL oplev servos use Coherent DIODE LASERS # 31-0425-000, 670 nm, 1 mW
Sep. 28, 2006 optical lever noise budget with DC readout in 40m, LIGO- T060234-00-R, Reinecke & Rana
May 22, 2007 BS, SRM & PRM He Ne 1103P takes over from diode
May 29, 2007 low RIN He Ne JDSU 1103P selected, 5 purchased sn: T8078254, T8078256, T8078257, T8078258 & T8077178 in Sep. 2007
Nov 30, 2007 Uniphase 1103P divergence measured
Nov. 30, 2007 ETMX old Uniphase 1103P from 2002 dies: .............., running time not known......~3-5 years?
May 19, 2008 ETMY old Uniphase 1103P from 1999 dies;.....................running time not known.....~ ?
Oct. 2, 2008 ITMX & ITMY are still diodes, meaning others are converted to 1103P earlier
JDSU 1103P were replaced as follows:
May 11, 2011 ETMX replaced, life time 1,258 days or 3.4 years
May 13, 2014 ETMX , LT 1,098 days or 3 y
May 22, 2012 ETMY, LT 1,464 days or 4 y
Oct. 5, 2011 BS & PRM, LT 4 years, laser in place at 1,037 days or 2.8 y
Sep. 13, 2011 ITMY old 1103P & SRM diode laser replaced by 1125P ..........old He life time is not known, 1125P in place 1,059 days or 2.9 y
June 26, 2013 ITMX 622 days or 1.7 y note: we changed because of beam quality.........................laser in place 420 days or 1.2 y
Sep. 27, 2013 purchased 3 JDSU 1103P lasers, sn: P893516, P893518, P893519 ......2 spares ( also 2 spares of 1125P of 5 mW & larger body )
|
10351
|
Fri Aug 8 12:39:19 2014 |
ericq | Summary | IOO | MC servo analysis |
I have measured the current boosted MC CLG below 100kHz with an SR785. Swept sine only could get me down to 10kHz, but I was able to get down to 5kHz with a noise-injection measurement.

I am attaching the SR785 outputs, which are in dB and Degrees. Additionally I pruned the areas of bad coherence out of these, and merged them to provide data files for the CLG and OLG in Real,Imaginary format. |
Attachment 1: mcLoopAug8.zip
|
10354
|
Fri Aug 8 15:57:29 2014 |
ericq | Summary | IOO | MC servo analysis |
I did some further measurements, to try and see what corresponds to what. In the end I performed four measurements:
- Closed loop gain measurement on SR785: Source to MC exc, T'd to channel one. Test 2 to channel two.
- Open loop gain measurement on SR785: Source to MCexc, Test 2 to channel one, Test 1 to channel two.
- Closed loop gain measurement on AG4395: RF Source to MC exc, T'd to R input. Test 2 to A input.
- Open loop gain measurement on AG4395: RF Source to MC exc, Test 2 to R input. Test 1 to A input.
I then converted OLGs to CLG and vice-versa with CLG = 1/(1-OLG)
Here are two plots showing the measured and inferred loop TFs for both closed and open.
 
The best agreement seems to be between the directly measured OLGs. Maybe I did something weird with the CLG measurements, or input impedances are distorting things ...
All data is attached, along with code used to generate the plots. |
Attachment 3: mcLoopAug8.zip
|
10356
|
Fri Aug 8 18:08:12 2014 |
Koji | Summary | IOO | MC servo analysis |
The closed gain I meant is the AO path: Use IN2 to excite the MC loop and measure IN1 using MON2(?).
In order to obtain the open loop gain from this meausrement, the gain mismatching needs to be compensated, though.
This measurement is to correctly predict the AO path response from the open loop transfer function.
Anyway, the openloop gain seems nicely measured. I'll try to predict AO path response from this. |
10359
|
Sat Aug 9 14:35:28 2014 |
Koji | Summary | IOO | MC servo analysis |
Eric's OLTF turned out consistent with the AO path TF that has been measured by me on Jul 31 (entry 10322).
Attachment 1:
Updated empirical fit of the open loop TF by LISO.
In this fit, I gave some of the poles/zeros associated with the boost manually set so that I can use them for the servo design.
LISO itself can make better fitting if all of the variables are moved.
Atatchment 2:
The OLTF data and LISO source for the fitting.
Attachment 3:
Comparison of the AO path TFs. The red one was measured directly on Jul 31. The TF is normalized at the low frequency.
The blue was estimated from the OLTF model given above. They are well consistent now.
Attachment 4:
Now some servo design was tried. In the new design (blue), zeros of the super boost frequency was moved from 20kHz to 30kHz
with the hope of having flatter AO response. The improvement is very little while costing costing above 100kHz. Note that the vertical
axis is intentionally in a linear scale. In fact, the AO response is much improved compared to the one before the MC UGF was increased
(shown in magenta). We have a flatter response both in magnitude and phase.
Therefore I think there is no need to tweak the boost frequency for the AO path.
I'd rather recommend to inspect the high frequency LPFs to earn more gain margin at 1MHz as explained in entry 10322.
Attachment 5:
This figure shows the comparison of the TFs for the current and new design trial, just in case someone is interested in to see.
|
Attachment 1: MC_OLTF_Fit.pdf
|
|
Attachment 2: liso.zip
|
Attachment 3: MC_CLTF_Fit.pdf
|
|
Attachment 4: MC_CLTF_new.pdf
|
|
Attachment 5: MC_OLTF_new.pdf
|
|
10362
|
Mon Aug 11 10:23:39 2014 |
steve | Summary | SUS | oplev laser summary updated |
Quote: |
2005 ALL oplev servos use Coherent DIODE LASERS # 31-0425-000, 670 nm, 1 mW
Sep. 28, 2006 optical lever noise budget with DC readout in 40m, LIGO- T060234-00-R, Reinecke & Rana
May 22, 2007 BS, SRM & PRM He Ne 1103P takes over from diode
May 29, 2007 low RIN He Ne JDSU 1103P selected, 5 purchased sn: T8078254, T8078256, T8078257, T8078258 & T8077178 in Sep. 2007
Nov 30, 2007 Uniphase 1103P divergence measured
Nov. 30, 2007 ETMX old Uniphase 1103P from 2002 dies: .............., running time not known......~3-5 years?
May 19, 2008 ETMY old Uniphase 1103P from 1999 dies;.....................running time not known.....~ ?
Oct. 2, 2008 ITMX & ITMY are still diodes, meaning others are converted to 1103P earlier
JDSU 1103P were replaced as follows:
May 11, 2011 ETMX replaced, life time 1,258 days or 3.4 years
May 13, 2014 ETMX , LT 1,098 days or 3 y
May 22, 2012 ETMY, LT 1,464 days or 4 y
Oct. 5, 2011 BS & PRM, LT 4 years, laser in place at 1,037 days or 2.8 y
Sep. 13, 2011 ITMY old 1103P & SRM diode laser replaced by 1125P ..........old He life time is not known, 1125P in place 1,059 days or 2.9 y
June 26, 2013 ITMX 622 days or 1.7 y note: we changed because of beam quality.........................laser in place 420 days or 1.2 y
Sep. 27, 2013 purchased 3 JDSU 1103P lasers, sn: P893516, P893518, P893519 ......2 spares ( also 2 spares of 1125P of 5 mW & larger body )
|
May 13, 2014 ETMX, .............laser in place 90 d
May 22, 2012 ETMY,
Oct. 7, 2013 ETMY, LT 503 d or 1.4 y............bad beam quality ?
Aug. 8, 2014 ETMY, .............laser in place 425 days or 1.2 y
|
10363
|
Mon Aug 11 21:03:48 2014 |
ericq, rana | Summary | IOO | MC demod measurement |
We measured the TF of the MC Demod board today.
We set the Marconi to +3dBm and drove the PD IN port of the demod board, starting at 29.5 MHz. Then we looked at the beat signal amplitude in the output of the demod board. So this is a transfer function but with mag only. Plots from Q below.
Rana took the demod board out and took pictures of it. Inside, the post mixer low pass is a SCLF-5 from mini-circuits. This has a lot of cutoff down low. Since the purpose of this filter is only to cutoff the 2f-1f and the 3f-2f products, we need to have a lot of attenuation at 29.5 MHz. One day, we may want to re-instate that notch for the (3*f1- f_MC) beat frequency, but for now we want stability.
So, I recommend that we (Steve) get 3 each of the SCLF-10 and SCLF-10.7 from Mini-Circuits Tuesday morning. Maybe we can put them into a spare board?
Also, we should probably remove the 140kHz:70kHz lead filter which is in the MC servo board. Its out of date. I think it would be fine for us to get a 7-15 kHz UGF for the CM servo and the MC can basically do that already. Mainly we want to fix the high frequency shape to get more stability.
After the measurements and photos, we had to reset the MCWFS offsets to get the WFS to not break the lock. Seems very sensitive to offsets. Hopefully Andres will give us a new Gouy phase telescope. |
10364
|
Mon Aug 11 22:07:31 2014 |
Koji | Summary | IOO | MC demod measurement |
SCLF-5!? It's surprising as the cut off of the OLTF is just above 1Hz. cf this entry
This means that not the demod board but MC or FSS boards seem to have large attenuation above 1MHz.
In this situation, does SCLF-10/10.7 really help us? |
10365
|
Mon Aug 11 23:32:54 2014 |
ericq | Summary | IOO | MC demod measurement |
Here's the magnitude plot of the board TF. As mentioned above, this was done with Marconi+Scope, so we were not able to get the phase of this transfer function.

Oddly enough, the bump that I saw is not included in Minicircuit's data on the SCLF-5. |
Attachment 2: demodLP.txt
|
# F(Hz) RMS(mV)
1035 38.6
2031 38.47
4031 38.47
8032 38.38
16030 38.10
32030 38.10
64030 38.16
128000 38.10
256000 38.22
... 12 more lines ...
|
10396
|
Thu Aug 14 22:58:59 2014 |
rana, jenne | Summary | Green Locking | ALS DIFF tuning |
We've been having trouble tuning the ALS DIFF matrix. Trying to see if the MC2 EXC can be cancelled in ALS DARM by adjusting the relative gains in ALSX and ALSY Phase Tracker outputs.
There's a bunch of intermittent behavior. Between different ALS locks, we get more or less cancellation. We were checking this by driving MC2 at ~100-400 Hz and checking the ALS response (with the ALS loops closed). We noticed that the X and Y readbacks were different by ~5-10 degrees and that we could not cancel this MC2 signal in DARM by more than a factor of 4-5 or so. In the middle of this, we had one lock loss and it came back up with 100x cancellation?
Attached is a PDF showing a swept sine measurement of the ALSX, ALSY, and DARM signals. You can see that there is some phase shift between the two repsonses leading to imperfect cancellation. Any ideas? Whitening filters? HOM resonance? Alignment? |
Attachment 1: sweep.pdf
|
|
10397
|
Thu Aug 14 23:19:49 2014 |
rana | Summary | LSC | ETM Violin fundamental filters moved to LSC |
We used to do violin mode and test mass body mode notches in the SUS-LSC filter modules. Now we want them balanced in the LSC and triggered by the LSC, so they're in the filter modules which go from the the LSC output matrix to the SUS.

Today, we were getting ETM violin mode ringups while doing ALS hunt and so we moved the bandstops into the LSC. I also changed the bandstop from a wide one which missed the ETMX mode to a double bandstop which gets both the ETMX and the ETMY mode. See attached image of the Bode mag.

|
10406
|
Mon Aug 18 09:42:50 2014 |
Koji | Summary | IOO | MCREFL PD charcterization |
Riju did the measurement of the MCREFL PD.
I found data files in her directory on the control machine.
I was not sure how much was the transimpedance of the DC out.
I assumed the default number from the circuit diagram which was 66.7Ohm.
This may cause the error in absolute caribration of the transimpedance but the shape does not change.
The RF preamp is gain-peeking at 250MHz.
Here is further characterization of the PD response.
As you can see in the second attachment, the 3dB cut off of the resonance is about 2.3MHz.
The game plan file in dropbox was also modified. |
Attachment 1: MCREFLPD_transimpedance.pdf
|
|
Attachment 2: MCREFLPD_transimpedance_zoom.pdf
|
|
10421
|
Thu Aug 21 22:10:52 2014 |
rana | Summary | Computer Scripts / Programs | network movements |
To help development of the data visualization project, we've assigned the .101 and .102 IP to DataVis. This is being used by the iMac in the control room via port 8 of the CDS switch near the Blue Plataeu Tournant.
We tried using one of the free ports, but Jamie realized that we had to use one of the already assigned ones due to some 'Smart' switch management software. So for the moment, please leave the iMac alone so that Bill can use it. |
10424
|
Fri Aug 22 15:11:55 2014 |
andres, nicolas | Summary | IOO | MC WFS activity |
1. Before doing anything, we centered the IOO QPDs.
2. With the WFS enabled, we offloaded the control signals onto the bias sliders. Then we saved the slider values. The MC LSC diode had a DC value of ~0.5
3. Turned down power with half wave plate before PMC. Power injected to vacuum ~ 100mW.
4. We did a beam scan of MC REFL, it looks smaller than what Andres predicted based on the MC eigenmode by 10-20%.
5. We made many changes on the table, pictures to be added by Andres.
6. We didn't have the 80% reflector we wanted to increase the WFS power, so it's still a 98%.
6. Beams were aligned on MC REFL PL, camera, beam dumps, WFSs.
7. Clean up
8. PSL power increased to 1.2W, MC locked right away.
9 We didn't change the IOO WFS output matrix, but we changed some signs and gains to make everything stable. MC autolocker brings it back from cold just fine.
10. All time bombs that we've left will be E.Q.'s to clean up. Sorry.\
11. Yay |
10425
|
Fri Aug 22 15:58:02 2014 |
Steve | Summary | IOO | MC WFS activity |
|
Attachment 1: GoyphaseSet.png
|
|
10442
|
Tue Sep 2 22:54:27 2014 |
Koji | Summary | LSC | phase tracker UGF |
FYI and FMI
Phase tracker UGF is Q_AMP * G * 2 PI / 360 where Q_AMP is the amplitude of the Q_ERR output and G is the gain of the phase tracker.
For example: Q_AMP = 270, G = 4000\ => UGF = 1.9kHz |
10454
|
Thu Sep 4 18:30:13 2014 |
Gabriele | Summary | ASC | Optimal Gouy phase for POP QPD |
Jenne asked me to simulate the signals on POP QPD when moving different mirrors, as a function of the Gouy phase where the QPD is placed.
I used the opportunity to create a MIST simulation file of the entire 40m interferometer, essentially based on my aLIGO configuration file. I used the recycling cavity lengths obtained from our survey, and other parameters from the wiki page. The configuration file is attached (fortymeters.mist).
Coming back to the main simulation, here is the result, both for the "regular" POP QPD and for a 22MHz demodulated one. The Gouy phase is measured starting from PR2. Cavity mirrors are easily decoupled from PRM in the "regular" QPD. As already demonstrated in a previous simulation, ETMs signals are very small in the 22 MHz QPD. Moreover, it is possible to zero the contribution from ITMs by choosing the right Gouy phase, at the price of a reduction of the PRM signal by a factor of 3-4. Simulation files are attached.

|
Attachment 2: fortymeters.mist
|
###########################################################################
# Configuration file for full dual recycled 40m interferometer
classname FortyMeters
################################################################ Parameters
# General parameters
const Pin 1 # input power
# Mirror parameters
const T_ITM 0.01384 # ITM transmission [from https://wiki-40m.ligo.caltech.edu/Core_Optics]
... 143 more lines ...
|
Attachment 3: fortymeters_pop_qpd.mist
|
###########################################################################
# Configuration file for full dual recycled 40m interferometer
classname FortyMetersPOP_QPD
################################################################ Parameters
# General parameters
const Pin 1 # input power
# Mirror parameters
const T_ITM 0.01384 # ITM transmission [from https://wiki-40m.ligo.caltech.edu/Core_Optics]
... 148 more lines ...
|
Attachment 4: pop_qpd.m
|
% compile and create simulation class
clear classes
MIST('fortymeters_pop_qpd.mist');
s = FortyMetersPOP_QPD(4);
% set angular motion of ITMs, ETMs and PRM
s.ETMX.setMotionShape('pitch');
s.ETMY.setMotionShape('pitch');
s.ITMX.setMotionShape('pitch');
s.ITMY.setMotionShape('pitch');
... 47 more lines ...
|
10455
|
Fri Sep 5 00:56:00 2014 |
rana | Summary | Optical Levers | ITM OLs recentered: violations found |
I re-centered the ITMX & ITMY Optical lever beams today since they were off. First I aligned the beam into the vacuum so that it went through the center of the on table optics and then tweaked the receiver optics alignment.
There are several bad practices on these which probably makes them drift:
- plastic bases on some lens mounts
- some lens mounts are fastened with a single dog instead of two
- there is no need to use dogs on mounts that have screw holes. Just put the mount so that 2 screws with washers can be used. The placement for these is not so critical.
- Use less steering mirrors! The ITMY OL path has 5 optics the beam enters the vacuum!!!
According to the datasheets, the laser has a beam diameter of 0.6 mm and a divergence angle of 1.3/2 mrad. So we can just calculate the right lens positions next time and not have to experiment with the whole visible laser lens kit.
For next Wednesday's cleanup, someone should volunteer to make the mounts more stable for the ITMs. |
10492
|
Wed Sep 10 22:17:29 2014 |
Koji | Summary | LSC | X/Y green beat mode overlap measurement |
[Koji Manasa]
We made quantitative inspection of the X/Y green beat setup on the PSL table.
DC output of the BBPD for each arm was measured by blockiing the beams at either or both side of the recombination BS.
The power over lap for the X arm beat note setup was 7.8% and is now 53%.
There is 3dB of headroom for the improvement of the mode overlap.
The power over lap for the Y arm beat note setup was 1.2% and is now 35%.
There is 4dB of headroom for the improvement of the mode overlap.
The RF analyzer monitor for the beat power is about 10dB lower than expected. Can we explain this only by the cable loss?
If not it there something causing the big attenuation?
XARM YARM
o BBPD DC output (mV)
V_DARK: - 3.3 + 1.9
V_PSL: + 4.3 +22.5
V_ARM: +187.0 + 8.4
o BBPD DC photocurrent (uA)
I_DC = V_DC / R_DC ... R_DC: DC transimpedance (2kOhm)
I_PSL: 3.8 10.3
I_ARM: 95.0 3.3
o Expected beat note amplitude
I_beat_full = I1 + I2 + 2 sqrt(e I1 I2) cos(w t) ... e: mode overwrap (in power)
I_beat_RF = 2 sqrt(e I1 I2)
V_RF = 2 R sqrt(e I1 I2) ... R: RF transimpedance (2kOhm)
P_RF = V_RF^2/2/50 [Watt]
= 10 log10(V_RF^2/2/50*1000) [dBm]
= 10 log10(e I1 I2) + 82.0412 [dBm]
= 10 log10(e) +10 log10(I1 I2) + 82.0412 [dBm]
for e=1, the expected RF power at the PDs [dBm]
P_RF: -12.4 -22.6
o Measured beat note power (before the alignment)
P_RF: -23.5 -41.7 [dBm] (38.3MHz and 34.4MHz)
e: 7.8 1.2 [%]
o Measured beat note power (after the alignment)
P_RF: -15.2 -27.1 [dBm] (26.6MHz and 26.8MHz)
e: 53 35 [%]
Measured beat note power at the RF analyzer in the control room
P_CR: -25 -20 [dBm]
Expected -17 - 9 [dBm]
Expected Power:
Pin + External Amp Gain (0dB for X, 20dB for Y)
- Isolation trans (1dB)
+ GAV81 amp (10dB)
- Coupler (10.5dB)
|
10518
|
Thu Sep 18 10:08:07 2014 |
steve | Summary | SUS | oplev laser summary updated |
Quote: |
Quote: |
2005 ALL oplev servos use Coherent DIODE LASERS # 31-0425-000, 670 nm, 1 mW
Sep. 28, 2006 optical lever noise budget with DC readout in 40m, LIGO- T060234-00-R, Reinecke & Rana
May 22, 2007 BS, SRM & PRM He Ne 1103P takes over from diode
May 29, 2007 low RIN He Ne JDSU 1103P selected, 5 purchased sn: T8078254, T8078256, T8078257, T8078258 & T8077178 in Sep. 2007
Nov 30, 2007 Uniphase 1103P divergence measured
Nov. 30, 2007 ETMX old Uniphase 1103P from 2002 dies: .............., running time not known......~3-5 years?
May 19, 2008 ETMY old Uniphase 1103P from 1999 dies;.....................running time not known.....~ ?
Oct. 2, 2008 ITMX & ITMY are still diodes, meaning others are converted to 1103P earlier
JDSU 1103P were replaced as follows:
May 11, 2011 ETMX replaced, life time 1,258 days or 3.4 years
May 13, 2014 ETMX , LT 1,098 days or 3 y
May 22, 2012 ETMY, LT 1,464 days or 4 y
Oct. 5, 2011 BS & PRM, LT 4 years, laser in place at 1,037 days or 2.8 y
Sep. 13, 2011 ITMY old 1103P & SRM diode laser replaced by 1125P ..........old He life time is not known, 1125P in place 1,059 days or 2.9 y
June 26, 2013 ITMX 622 days or 1.7 y note: we changed because of beam quality.........................laser in place 420 days or 1.2 y
Sep. 27, 2013 purchased 3 JDSU 1103P lasers, sn: P893516, P893518, P893519 ......2 spares ( also 2 spares of 1125P of 5 mW & larger body )
|
May 13, 2014 ETMX, .............laser in place 90 d
May 22, 2012 ETMY,
Oct. 7, 2013 ETMY, LT 503 d or 1.4 y............bad beam quality ?
Aug. 8, 2014 ETMY, .............laser in place 425 days or 1.2 y
|
Sept. 5, 2014 new 1103P, sn P893516 installed at SP table for aLIGO oplev use qualification
|
10694
|
Mon Nov 10 23:14:20 2014 |
rana | Summary | SUS | SRM: damping gains & Optical Lever servo Tune-up |
- tweaked gains for POS, PIT, YAW, SIDE by ~10-20% to get nice ringdowns with Q~5.
- Measured bounce rool freqs = 16.43 / 23.99 Hz. Updated the Mech Res Wiki page. Tightened up the bandstops to get back a few deg of phase. Propagated these new bandstops into the SRM SUS damping filter banks.
- Made and turned on a LP filters for the loops.
- Added a ~0.3 Hz gain boost / bubble.
- Set UGF to be ~2.5x below where it rings up. Estimate this to be ~3.5 Hz.
- First PDF shows bounce/roll peaks in OSEMs. Notice how f_roll is > sqrt(2)*f_bounce. ??
- Second PDF shows the OL spectra with the loops on/off. Previously there was no Boost and no LP (!!) turned on.
- 3rd PDF shows the modeled Bode plot of the OLG. yellow/blue is boost off/on
|
Attachment 1: SRM-BR.pdf
|
|
Attachment 2: SRM_err_141110.pdf
|
|
Attachment 3: SRM-OLG.pdf
|
|
10706
|
Wed Nov 12 22:22:11 2014 |
Koji | Summary | IOO | Estimation of the angular jitter imposed by the TTs |
[Koji, Rana, Jenne]
One coil of the TT produce 36nrad/rtHz at DC.
- C1:IOO-TT2_UL_EXC was excited with 5 count_pk at 0.04Hz.
- LSC_TRY exhibited the symmetric reduction of the transmission from 0.95 to 0.90.
1 - (theta/theta0)^2 /2 = 0.90 / 0.95
=> theta / theta0 = 0.32
- 40m beam waist radius is 3.1mm. This means the divergence angle is 1.1e-4 rad.
=> 1.1e-4*0.32 = 3.6e-5 rad
=> 3.6e-5/5 = 7.2 urad/count (per coil)
- DAC noise 1/sqrt(12 fs), where fs is the sampling rate (fs = 16384)
=> 0.002 cnt/rtHz
- One coil causes 7.2u*0.002 = 14 nrad/rtHz (at DC)
- One suspension cause 29 nrad/rtHz (at DC) |
Attachment 1: 03.png
|
|
10722
|
Mon Nov 17 20:28:17 2014 |
rana | Summary | IOO | MC servo summing amp |
I modified the /cvs/cds/caltech/target/c1psl/psl.db file to adjust the records for the FSS-FAST signal (to make it go yellow / red at the correct voltages). This was needed to match 5V offset which Koji added to the output of the FSS board back in August.
I also manually adjusted the alarm levels with caput so that we don't have to reboot c1psl. Beware of potential tiimebomb / boot issues if I made a typo! psl.db update in the SVN (also, there were ~12 uncomitted changes in that directory....please be responsible and commit ALL changes you make in the scripts directory, even if its just a little thing and you are too cool for SVN) |
10760
|
Sun Dec 7 13:11:57 2014 |
manasa | Summary | General | Frequency Offset Locking - To Do List |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours. |
Attachment 1: FOLtodolist.pdf
|
|
10765
|
Mon Dec 8 15:54:39 2014 |
manasa | Summary | General | Dec 8 - Check Frequency Counter module |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
[Diego, Manasa]
We looked into the configuration and settings that the frequency counters (FC) and Domenica (the R pi to which the FCs talk to) were left at . After poking around for a few hours, we were able to readout the FC output and see it on StripTool as well.
We have made a list of modifications that should be done on Domenica and to the readout scripts to make the FC module automated and user-friendly.
I will prepare a user manual that will go on the wiki once these changes are made.
|
10766
|
Mon Dec 8 20:53:51 2014 |
diego | Summary | General | Dec 8 - Check Frequency Counter module |
Quote: |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
[Diego, Manasa]
We looked into the configuration and settings that the frequency counters (FC) and Domenica (the R pi to which the FCs talk to) were left at . After poking around for a few hours, we were able to readout the FC output and see it on StripTool as well.
We have made a list of modifications that should be done on Domenica and to the readout scripts to make the FC module automated and user-friendly.
I will prepare a user manual that will go on the wiki once these changes are made.
|
I started working on the scripts/FOL directory (I did a backup before tampering around!):
- I still need to make some serious polishing in the folder, and into the Raspberry Pi itself, in order to have a clean and understandable environment;
- as of now, I created an single armFC.c program, which takes as arguments the device (/dev/hidraw0 for the X arm, and /dev/hidraw1 for the Y arm) and the value to write into the frequency counter (0x3 for initialization and 0x2 for actual use); hence, no more need for recompilation!
- I improved the codetorun.py script (and gave the fellow a proper name, epics_channels.py) which handles the initialization AND the availability of the channels;
- On the Raspberry Pi, I created two init scripts, /etc/init.d/epics_server.sh and /etc/init.d/epics_channels.sh, which start at the end of the boot process with default runlevels; the former starts the softIOc process (epics itself), while the latter executes the constantly running epics_channels.py script; as they are services, they can be started/stopped with the usual sudo /etc/init.d/NAME start|stop|restart
As a result, as soon as the Raspberry Pi completes its boot process, the two beatnote channels are immediately available.
|