17296
|
Mon Nov 21 18:43:46 2022 |
yuta | Update | BHD | c1hpc and c1lsc modified to send BHD_DIFF and BHD_SUM | [Anchal, Yuta]
To send BHD signals from c1hpc after |
|
17295
|
Mon Nov 21 18:33:05 2022 |
Anchal | Update | SUS | MC2 OSEMs calibrated using MC_F | Repeated this for MC2 using the error measurement
technique mentioned in 40m/17286
using this notebook. |
|
17294
|
Mon Nov 21 17:44:00 2022 |
yuta | Summary | BHD | MICH BHD displacement sensitivity with AS55_Q and BHD_DIFF | [Paco, Yuta]
MICH displacement sensitivity was
compared under AS55_Q locking and BHD_DIFF |
 |
17293
|
Mon Nov 21 15:16:12 2022 |
Tega | Update | CDS | workstation upgrade | We are currently working on donatella workstation
upgrade, which we are calling donatellaclone
for now. After installing Debian 11, we install |
|
17292
|
Mon Nov 21 14:55:58 2022 |
JC | Configuration | General | LED Instead of Incandescent Lights | Electrical came by today to see the lights.
The issue may be the switches, but they will
come by tomorrow to solve the issue. A couple |
|
17291
|
Mon Nov 21 11:52:50 2022 |
Radhika | Summary | Calibration | Single arm cal with 5 lines | [Paco, Radhika]
We set out to realign the YARM
AUX laser input into the arm cavity. |
|
17290
|
Mon Nov 21 09:13:31 2022 |
JC | Update | PSL | PMC input beam aligned again, IMC | [JC, Paco]
I attempted to align PMC Beam from
a transmission of 0.72. I failed to do so |
|
17289
|
Sun Nov 20 13:42:21 2022 |
Anchal | Update | SUS | IMC test | I repeated this test for the following
configuration:
PSL shutter closed at good |
|
17288
|
Fri Nov 18 23:21:54 2022 |
rana | Update | ASC | IMC WFS ongoing | On Wednesday, I did some rework of the
MC WFS gains. I think it should still work
as before as long as the overall input gain |
|
17287
|
Fri Nov 18 22:46:02 2022 |
yuta | Summary | BHD | MICH optical gain measurements with different LO phases, with signs | MICH
optical gain with a sign was measured with
different LO phases over ~180 degrees, with |
 |
17286
|
Fri Nov 18 17:00:15 2022 |
Anchal | Update | SUS | MC1 and MC3 OSEMs calibrated using MC_F | After the MC1 osem dewhitening was fixed,
I did the calibration of MC1 OSEM signals
using MC_F using this notebook. |
|
17285
|
Fri Nov 18 16:58:39 2022 |
yuta | Summary | BHD | Actuator calibrations for MICH BHD | As there is some confusion in actuator
calibration, we have done the measurement
again from scratch. |
7x |
17284
|
Fri Nov 18 13:05:00 2022 |
yuta | Summary | BHD | Gains adjusted for bandstop filters for BHD optics | [Paco, Yuta]
We realized that bandstop filters |
|
17283
|
Fri Nov 18 09:00:27 2022 |
JC | Update | VAC | Pressure Gauge Information | I bought the spare Full-Range Pirani Gauge
a while ago and realized that I never logged
this. The Pirani Gauges we are using is described |
|
17282
|
Thu Nov 17 20:02:10 2022 |
yuta | Summary | BHD | MICH optical gain measurements with different LO phases | MICH
optical gain was measured with different
LO phases over ~90 degrees. |
 |
17281
|
Thu Nov 17 16:48:07 2022 |
Anchal | Frogs | Computer Scripts / Programs | FSS SLOW servo running Now | I've moved the FSS Slow PID script
running to megatron through systemd daemons.
The script is working as expected right now. |
|
17280
|
Thu Nov 17 15:53:47 2022 |
JC | Configuration | Cameras | ITMX Camera -- attempt at fix | The issue was the power supply.
|
|
17279
|
Thu Nov 17 14:12:58 2022 |
yuta | Summary | BHD | Optical gain calibrations for BHD MICH with lower UGF | [Paco, Yuta]
We found that MICH UGF was unexpectedly |
 |
17278
|
Thu Nov 17 12:24:48 2022 |
Paco | Configuration | Cameras | ITMX Camera -- attempt at fix | I found that an old BNC cable for ITMXF
video existed so I first tried swapping both
ends of the cable, one on the ITMX viewport |
|
17277
|
Thu Nov 17 11:24:39 2022 |
JC | HowTo | LSC | Locking MICH | [Yuta, JC]
Here is the Yuta's Alignment
Scheme from elog |
|
17276
|
Thu Nov 17 09:06:35 2022 |
JC | Summary | General | Light Replace | Electrical Shop came in today to replace
the lightbulbs around 40m. I supervised the
entire visit, inside and outside the lab |
|
17275
|
Thu Nov 17 07:39:01 2022 |
JC | Configuration | Cameras | ITMX Camera | Coming in this morning, I found ITMX Camera
malfunctioning. |
|
17274
|
Wed Nov 16 18:41:17 2022 |
yuta | Summary | BHD | Optical gain calibrations for BHD MICH | Optical gains of AS55 and BH55 are calibrated
for BHD MICH.
LO-ITM single bounce: |
  |
17273
|
Wed Nov 16 15:09:08 2022 |
yuta | Update | BHD | BHD fringe contrast measured with unwhitening filters | BHD fringe visibility was measured again
with unwhitening filters on on BHDC_A and
B, which removed signal leakage to zero ( |
|
17272
|
Wed Nov 16 12:53:36 2022 |
rana | Update | ASC | IMC WFS ongoing | In the middle of aportioning gains and
signs in the IMC WFS screen, so beware. More
updates soon. |
|
17271
|
Wed Nov 16 11:56:21 2022 |
Radhika | Update | PSL | PMC input beam aligned again, IMC | [Yuta, JC, Radhika]
PMC input beam was aligned again,
bringing transmission from 0.70 to ~0.75. |
|
17270
|
Tue Nov 15 19:00:56 2022 |
yuta | Summary | BHD | MICH locked with balanced homodyne readout at some LO phase | [Paco, Yuta]
MICH
was locked with BHD DCPD A-B signal with |
  |
17269
|
Tue Nov 15 17:58:00 2022 |
Paco | Configuration | Cameras | POP camera realignment after IFO alignment | [Paco, Yuta]
I swapped the 1 inch BS and lenses
along the POP beam to clear the apertures |
  |
17268
|
Tue Nov 15 17:08:59 2022 |
Paco | Update | BHD | Request for estimates | [Yehonathan, Yuta, Paco]
We would like to estimate:
LO phase sensitivty (for |
|
17267
|
Tue Nov 15 16:34:55 2022 |
alex | Update | General | Debian with Sensoray | I have confirmed the ability to install
the sensoray drivers on Debian 11 in a virtual
machine environment. I will |
|
17266
|
Tue Nov 15 11:04:01 2022 |
Paco | Summary | Calibration | Single arm cal with 5 lines | YALS Hardware inspection
The
ALS Cal for ITMY actuation was off by ~ 1000,
so I decided I don't trust / understand |
|
17265
|
Mon Nov 14 17:45:02 2022 |
yuta | Update | BHD | BHD DC PD unwhitening and removing cables to c1lsc | [Paco, Yuta]
We removed splitter to route BHD DC |
|
17264
|
Mon Nov 14 14:52:56 2022 |
Paco | Configuration | SUS | BHD SUS Coil output balance | [JC, Paco]
We installed a steering mirror
intersecting the BHD beam path and put the |
    |
17263
|
Sat Nov 12 21:59:24 2022 |
Anchal | Frogs | Computer Scripts / Programs | FSS SLOW servo not running | I stopped the Docker PID script and started
the old python script on megatron. Instructions
on how to do this are here. |
|
17262
|
Fri Nov 11 20:59:13 2022 |
Chris | Frogs | Computer Scripts / Programs | FSS SLOW servo not running | The problem with trends was due to the
epics data collection process (standalone_edc)
that runs on c1sus. When all the FEs were |
|
17261
|
Fri Nov 11 20:01:56 2022 |
rana | Frogs | Computer Scripts / Programs | FSS SLOW servo not running | I was trying to debug why the NPRO PZT
is all over the place, and it turns out that
the new FSS SLOW script is not actually running. |
|
17260
|
Fri Nov 11 19:47:04 2022 |
rana | Update | SUS | MC SUS were overdamped: damping gains all reduced | After clicking on the extra MC1 unwhitening
filter, I was going to retune the damping
gains to give a Q ~ 5 for the suspension. |
|
17259
|
Fri Nov 11 19:20:23 2022 |
rana | Update | SUS | MC1 OSEMs output is weird | I
turned on the extra un-whitening filter (not |
|
17258
|
Fri Nov 11 19:11:50 2022 |
JC | Update | General | WFS Whitening and Demod Boards | WFS Whitening and Demod boards were scavenged.
' |
 |
17257
|
Fri Nov 11 14:15:45 2022 |
Paco | Summary | Calibration | Single arm cal with 5 lines | I turned all the LSC oscillators on and
used the digital demod for BEATYF (fine y
als beat) to grab the data. For this I added |
|
17256
|
Fri Nov 11 11:29:11 2022 |
Anchal | Update | SUS | MC1 OSEMs output is weird | Late elog; original time Thursday, Nov
10 16:00 2022
MC1 is using a new satellite |
|
17255
|
Thu Nov 10 20:46:32 2022 |
rana | Update | ASC | IMC WFS servo diagnosis | To check out the bandwidths and
cross-coupling in the WFS loops, I made a
script (attached) to step the offsets around, |
 |
17254
|
Thu Nov 10 18:09:28 2022 |
rana | Update | SUS | MC WFS settings all weird | Today I found the MC WFS settings all weird.
Not only was the input slider set to zero,
but the filter bank outputs were off, and |
|
17253
|
Thu Nov 10 17:40:31 2022 |
Paco | Summary | Calibration | Calibration Plan | Plan to calibrate single arm actuation
strength
Lock single arm cavity |
 |
17252
|
Wed Nov 9 20:29:20 2022 |
Anchal | Update | SUS | MC2 and MC3 set to undergo free swing test tonight | I have set a free swing test for MC2 and
MC3 to trigger at 1 am tonight. The test
should last for about 4.5 hrs upto 5:30 am. |
|
17251
|
Wed Nov 9 20:01:38 2022 |
Anchal | Update | SUS | MC1 OSEMs output is weird | I took a coil to OSEM transfer function
for MC1 osems (LL, UR) today and again
the slope of the transfer function was -1.4 |
  |
17250
|
Wed Nov 9 14:19:18 2022 |
Tega | Update | CDS | fb1 OS migration | [Tega, Chris]
We migrated fb1 OS from the
teststand fb1 drive to the internal 2TB RAID |
|
17249
|
Wed Nov 9 11:07:16 2022 |
Anchal | Update | SUS | IMC test | Following configurations were kept today
morning:
|
|
17248
|
Wed Nov 9 07:49:59 2022 |
rana | Update | SUS | MC2 OSEMs calibrated using MC_F | It would be good if you can post the uncertainties
with these measurements (and for generally
any measurements). This is so that we can |
|
17247
|
Tue Nov 8 21:39:12 2022 |
alex | Summary | General | Sensoray & SDI Video Encoder selection | I have been looking at various replacements
for the sensoray, and have found that the
majority of new usb video encoders don't |
|