  40m Log, Page 48 of 335  Not logged in ELOG logo
ID Date Author Type Category Subject Text Attachments
  14489   Wed Mar 20 20:07:22 2019 JonUpdateVACDoing vac controls workWork is completed and the vac system is
back in its nominal state.

  
  14488   Wed Mar 20 19:26:25 2019 JonUpdateVACProtection against AC power lossToday I implemented protection of the vac
system against extended power losses. Previously,
the vac controls system (both old and new)
 IMG_3146.jpg 
  14487   Wed Mar 20 12:31:30 2019 JonUpdateVACDoing vac controls workI'm rebooting the IOLAN server to load
new serial ports. The interlocks might trip
when the pressure gauge readbacks cut out.
  
  14486   Mon Mar 18 20:22:28 2019 gautamUpdateALSALS stability testI'm running a test to see how stable
the EX green lock is. For this purpose, I've
left the slow temperature tuning servo on
  
  14485   Mon Mar 18 18:10:14 2019 KojiSummaryGeneralTask items and priority[Gautam/Chub/Koji] ~ Mini discussion

Maintenance / Upgrade Items

(Priority high to low)
  
  14484   Mon Mar 18 17:06:12 2019 gautamUpdateOptical LeversITMY HeNe replacedOplev HeNe was replaced this afternoon.
We did some HeNe shuffling:


A new HeNe was being used
 OLRIN.pdfOL_PIT.pdfOL_YAW.pdf 
  14483   Mon Mar 18 12:27:42 2019 gautamUpdateGeneralIFO status
c1iscaux2
VME crate is damaged -
 IMG_7343.JPGITMXOL.png 
  14482   Sun Mar 17 21:06:17 2019 AnjaliUpdateALSAmplifier characterisationThe goal was to characterise the new amplifier
(AP1053). For a practice, I did the characterisation
of the old amplifier.This test is similar
 Gain_measurement.pdfAmplifier_gain.pdfnoise_measurement.pdfnoise_characterisation.pdf 
  14481   Sun Mar 17 13:35:39 2019 AnjaliUpdateALSPower splitter characterizationWe characterized the power splitter ( Minicircuit-
ZAPD-2-252-S+). The schematic of the measurement
setup is shown in attachment #1. The network/spectrum/impedance
 Measurement_setup.pdfTotal_loss.pdfAmplitude_unbalance.pdfPhase_unbalance.pdf 
  14480   Sun Mar 17 00:42:20 2019 gautamUpdateALSNF1611 cannot be shot-noise limited?Summary:

Per the manual (pg12)
of the NF 1611 photodiode, the "Input
 PDnoise.pdf 
  14479   Thu Mar 14 23:26:47 2019 AnjaliUpdateALSALS delay line electronicsAttachment #1 shows the schematic of the
test setup. Signal generator (Marconi) was
used to supply the RF input. We observed
 test_setup_ALS_delay_line_electronics.pdf 
  14478   Wed Mar 13 01:27:30 2019 gautamUpdateALSALS delay line electronicsThis test was done, and I determine the
frequency discriminant to be 
 DFDcal.pdfDFDnoise.pdf 
  14477   Tue Mar 12 22:51:25 2019 gautamUpdateALSALS delay line electronicsThis
Hanford alog may be of relevance as we
are using the aLIGO AA chassis for the IR
  
  14476   Fri Mar 8 08:40:26 2019 AnjaliConfiguration Frequency stabilization of 1 micron sourceThe schematic of the homodyne configuration
is shown below.

Following are the list of components
 Homodyne_setup.png 
  14475   Thu Mar 7 01:06:38 2019 gautamUpdateALSALS delay line electronicsSummary:

The restoration of the delay-line
electronics is complete. The chassis has
  
  14474   Tue Mar 5 15:56:27 2019 gautamSummaryTip-TIltDiscussion points about TT re-designChub, Koji and I have been talking about
Udit's re-design. Here are a few points
that were raised. Chub/Koji can add to/correct
  
  14473   Sun Mar 3 14:16:31 2019 gautamUpdateIOOMegatron hard-rebootedIMC was not locked for the past several
hours. Turned out MC autolocker was stuck,
and I could not ssh into megatron because
  
  14472   Sat Mar 2 14:19:35 2019 gautamUpdateCDSFSS Slow servo gains not burt-edPSL NPRO PZT voltage showed large low frequency
(hour timescale) excursions on the control
room StripTool trace, leading me to suspect
  
  14471   Wed Feb 27 21:34:21 2019 gautamUpdateGeneralSuspension diagnosisIn my effort to understand what's going
on with the suspensions, I've kicked
all the suspensions and shutdown the watchdogs
  
  14470   Mon Feb 25 20:20:07 2019 KojiUpdateSUSDIN 41612 (96pin) shrouds installed to vertex SUS coil driversThe forthcoming Acromag c1susaux is
supposed to use the backplane connectors
of the sus euro card modules.
 P_20190222_175058.jpg 
  14469   Fri Feb 22 12:19:46 2019 gautamUpdateIOOTT coil driver VmonTo debug the issue of the suspected drifting
TTs further, I temporarily hijacked CH0-CH8
of ADC1 in the c1lsc expansion chassis, and
  
  14468   Wed Feb 20 23:55:51 2019 gautamUpdateALSALS delay line electronicsSummary:

Last year, I worked
on the ALS delay line electronics, thinking
 delayLineDemod.pdf.zip 
  14467   Wed Feb 20 18:26:05 2019 gautamUpdateIOOIPPOS recommissionedI've suspected that the TTs are drifting
significantly over the course of the last
couple of days, because despite repeated
 IMG_7330.JPGScreenshot_from_2019-02-20_19-43-27.png 
  14466   Tue Feb 19 22:52:17 2019 gautamUpdateASSY arm clipping doubtfulIn an earlier
elog, I had claimed that the suspected
clipping of the cavity axis in the Y arm
 overheadComparison.pdfclipping.pdfrearComparison.pdf 
  14465   Tue Feb 19 19:03:18 2019 ranaUpdateComputersMartian router -> WPA2I have swapped our martian router's
WiFi security over to WPA2 (AES) from the
previous, less-secure, system. Creds are
  
  14464   Mon Feb 18 19:16:55 2019 ranaSummaryComputersnew laptop setup: ASIAThe old IBM laptop (Asia) has died from
a fan error after 7 years. WE have a new
Lenovo 330 IdeaPad to replace it:
  
  14463   Sun Feb 17 17:35:04 2019 gautamSummaryLoss MeasurementInferred X arm loss Summary:

To complete the story before moving
on to ALS, I decided to measure the X arm
 DQcheck_XARM.pdfconsolidated.pdfposterior_modelParams_XARM.pdfposterior_Loss_XARM.pdf 
  14462   Fri Feb 15 21:15:42 2019 gautamUpdateVACdd backup of c1vac made
Connected one of the solid-state
drives to c1vac. It was /dev/sdb.
Formatted
the drive using sudo mkfs -t
  
  14461   Fri Feb 15 20:07:02 2019 JonUpdateVACUpdated vacuum punch listWhile working on the vac controls
today, I also took care of some of the remaining
to-do items. Below is a summary of what was done,
  
  14460   Fri Feb 15 19:50:09 2019 ranaUpdateVACVac system is back upThe acromags are on the UPS. I suspect
the transient came in on one of the signal
lines. Chub tells me he unplugged one of
  
  14459   Fri Feb 15 18:42:57 2019 gautamUpdateLSCTRY 60 Hz solved A more permanent fix than a crocodile clip was
implemented. Should probably look to do this
for the X end unit as well.
 IMG_7323.JPG 
  14458   Fri Feb 15 18:41:18 2019 ranaUpdateVACVac system is back upIf the acromags lock up whenever there
is an electrical spike, shouldn't we
have them on UPS to smooth out these ripples?
  
  14457   Fri Feb 15 15:22:08 2019 gautamUpdateCDSc1rfm errors persistI restarted c1scy, 
 Screenshot_from_2019-02-15_15-21-47.png 
  14456   Fri Feb 15 11:58:45 2019 JonUpdateVACVac system is back upThe problem encountered with the vac controls
was indeed resolved via the recommendation
I posted yesterday. The Acromags had gone
  
  14455   Thu Feb 14 23:14:12 2019 gautamUpdateCDSc1rfm errorsThe pressure is still 2e-4 torr according
to CC1 so I thought I'd give ASS debugging
a go tonight. But the arm transmission signal
  
  14454   Thu Feb 14 21:29:24 2019 gautamSummaryLoss MeasurementInferred Y arm loss Summary:

From the measurements I have, the
Y arm loss is estimated to be 58
 modelPerturb.pdfposterior_modelParams.pdfposterior_Loss.pdf 
  14453   Thu Feb 14 18:16:24 2019 JonUpdateVACVacromag failureI sent Gautam instructions to first try
stopping the modbus service, power
cycling the Acromag chassis, then
  
  14452   Thu Feb 14 15:37:35 2019 gautamUpdateVACVacromag failure[chub, gautam]

Sumary:

One of the XT1111 units (XT1111a)
 Screenshot_from_2019-02-14_15-40-36.pngIMG_7320.JPGScreenshot_from_2019-02-14_20-43-15.png 
  14451   Wed Feb 13 02:28:58 2019 gautamSummaryLoss MeasurementY arm lossAttachment #1 shows estimated systematic
uncertainty contributions due to 


ITM transmission by +/-
 systUnc.pdf 
  14450   Tue Feb 12 22:59:17 2019 gautamSummaryLoss MeasurementY arm lossSummary:

There are still several data quality
issues that can be improved. I think there
 segmented.pdfconsolidated.pdf 
  14449   Tue Feb 12 18:00:32 2019 gautamSummaryLoss MeasurementLoss measurement setupAnother
arm loss measurement started at 6pm.
  
  14448   Mon Feb 11 19:53:59 2019 gautamSummaryLoss MeasurementLoss measurement setupTo measure the Y-arm loss, I decided to
start with the classic reflectivity method.
To prepare for this measurement, I did the
 DQcheck.pdfinitialData.pdf 
  14447   Mon Feb 11 16:38:34 2019 gautamUpdateLSCETMY OL calibration updatedSince we changed the HeNe, I updated the
calibration factors, and accepted the changes
in the SDF.
 OL_calib_ETMY_PERROR.pdfOL_calib_ETMY_YERROR.pdf 
  14446   Mon Feb 11 15:41:49 2019 gautamUpdateLSCTRY 60 Hz solved - but clipping persistsRich came by the 40m to photocopy some
pages from Hobbs, and saw me working on the
60 Hz hunting. As I suspected, the problem
  
  14445   Fri Feb 8 20:48:52 2019 gautamUpdateLSCIFO recoverySeveral housekeeping tasks were carried
out today in preparation for the Y-arm loss
measurement.
 IMG_7307.JPG 
  14444   Fri Feb 8 20:35:57 2019 gautamSummaryTip-TIltCoating spec[Attachment
#1]: Computed spectral
 specRefl.pdfAoIscan.pdf 
  14443   Fri Feb 8 02:00:34 2019 gautamUpdateSUSITMY has tendency of getting stuckAs it turns out, now ITMY has a tendency
to get stuck. I found it MUCH
more difficult to release the optic using
 Screenshot_from_2019-02-08_02-04-22.png 
  14442   Fri Feb 8 00:20:56 2019 gautamSummaryTip-TIltFive FiveNine Optics Optics deliveredThey have been stored on the 3rd shelf
from top in the clean optics cabinet at the
south end. EX
  
  14441   Thu Feb 7 19:34:18 2019 gautamUpdateSUSETMY suspension oddnessI did some tests of the electronics chain
today.


Drove a sine-wave using
 ETMY_sensors.pdfpostDiag.pdf 
  14440   Thu Feb 7 19:28:46 2019 gautamUpdateVACIFO recovery[rana, gautam]

The full 1 W is again being sent
into the IMC. We have left the PBS+HWP combo
  