  14845   Tue Aug 13 14:36:17 2019 gautamUpdateCDSP1--->P2As it turns out, only one extra shroud
needed to be installed - I did this and migrated
the cables for the 4 whitening boards from
  
  14844   Tue Aug 13 08:07:09 2019 gautamUpdateCDSP1--->P2This morning, I wanted to move the existing
cables going to the P1 connectors of the
iLIGO whitening boards to the P2 connector,
  
  14843   Mon Aug 12 21:25:19 2019 KojiUpdateCDSMore bench test of c1iscaux1.

> Looking through the manual,
I found a recommendation (pg10) that the
  
  14842   Mon Aug 12 19:58:23 2019 gautamUpdateIOOMC1 suspension oddnessRepair plan:


Get "spare" satellite
box working --- Chub

According
  
  14841   Mon Aug 12 17:36:04 2019 gautamUpdateCDSMore bench test of c1iscaux[chub, gautam]

With Chub's help, most of the
problems have been resolved. Summary: I
 iscauxCheclist.pdf 
  14840   Sun Aug 11 11:47:42 2019 gautamUpdateCDSBench test of c1iscauxI bench tested the functionality of all
the c1iscaux Acromag crate channels. Summary:
we are not ready for a Monday install, much
  
  14839   Fri Aug 9 20:58:33 2019 JonUpdateElectronicsBorrowed Variac transformerI borrowed an old-looking Variac variable
transformer from the power supplies cabinet
along the y-arm. It is currently in
  
  14838   Fri Aug 9 16:37:39 2019 gautamUpdateALSMore EY table workSummary:


220 uW / 600 uW (~36 %
mode-matching) of IR light coupled into fiber
 IMG_7791.JPGfiberCleaning.pdf 
  14837   Fri Aug 9 08:59:04 2019 gautamUpdateCDSPrep for install of c1iscaux[chub, gautam]

We scoped out the 1Y3 rack this
morning to figure out what needs to be done
  
  14836   Thu Aug 8 12:01:12 2019 gautamUpdateIOOMC1 suspension oddnessAt ~1am PDT today, all the MC1 shadow sensor
readbacks (fast CDS channels and Slow Acromag
channels, latter not shown here) went to
 MC1_suspension.pngMC1_suspension.pdf 
  14835   Tue Aug 6 23:09:20 2019 gautamUpdateALSEY table work
Removed power monitoring PD (It
was off anyways)
Installed Steering
mirror and collimator in K6XS mount (fast
  
  14834   Tue Aug 6 16:44:50 2019 yehonathanUpdateLoss Measurement I grab 2 hours of the PD measurements using
dlData_simple.ipynb in the misaligned state.

I get pretty much a normally distributed
 figure_1.pngfigure_1-1.png 
  14833   Tue Aug 6 15:52:06 2019 gautamUpdateBHDPreliminary BHD calculationsSummary:

The requirement on the phase noise
on the direct backscatter
 OMCbackscatter.pdf 
  14832   Tue Aug 6 14:55:23 2019 gautamUpdateCDSMaking Matlab R2015b the defaultML2013 is unable to open Simulink on any
of the workstations. We decided to make the
default version of Matlab R2015b (the default
 ML2015b.png 
  14831   Tue Aug 6 14:12:02 2019 yehonathanUpdateComputersmaking rossa great againcdsutils is not working on rossa.

Import cdsutils produces this error:

In [2]: import cdsutils
  
  14830   Mon Aug 5 17:36:04 2019 yehonathanUpdateLoss Measurement We check for unexpected drifts in the PD
reading (clipping and such). We put a pickoff
mirror where the PD used to be and place
 20190805_171511.jpg 
  14829   Mon Aug 5 17:23:26 2019 gautamSummaryComputersWiFi Settings on asiaThe VEA laptop asia was configured to be
able to connect to too many WiFi networks
- it was getting conflicted in its default
  
  Draft   Mon Aug 5 16:28:41 2019 yehonathanUpdateLoss Measurementwhat is going on with the loss measurements ?We hypothesize that the systematic error
in the loss measurement can come from the
fact that the requirement on the alignment
  
  14827   Mon Aug 5 14:47:36 2019 yehonathanUpdateLoss Measurement Summary:

I analyze the 100 reps loss measurement
of the Y arm using the AnalyzeLossData.ipynb
  
  14826   Sun Aug 4 14:39:41 2019 gautamUpdateGeneralsome lab activity
Unresponsive c1psl, c1iool0, c1auxey
and c1iscaux VME crates were keyed.
c1psl
channels were burt-restored, did a burtrestore,
 ETMX_OL.pngSeismic_TRXTRYandMC_4Aug2019.pdf 
  14825   Fri Aug 2 17:07:33 2019 yehonathan, gautamUpdateLoss Measurement We run a loss measurement on the Y arm
with 50 repetitions.
  
  14824   Fri Aug 2 16:46:09 2019 KruthiUpdateCamerasClean upI've put the analog camera back and
disconnected the 151 unit GigE. But I ran
out of time and wasn't able to replace
  
  14823   Fri Aug 2 11:37:38 2019 gautamUpdateALSEY IR ALS AssaySummary:

I'd like to confirm that the
IR ALS scheme will work for locking. The
 IMG_7780.JPGEy_MM_20190802.pdf 
  14822   Thu Aug 1 13:55:34 2019 DuoBureaucracyEquipment loanGpib module taken to QIL labvanna --> QIL.

gautam
20190804: The GPIB module + power supply
  
  14821   Wed Jul 31 17:57:35 2019 KojiUpdateBHDOMC cavity geometry4 deg is not an optimized number optimized
for criteria, but to keep the cavity short
width to 0.1m. But the justification of 4deg
  
  14820   Wed Jul 31 14:44:11 2019 gautamUpdateComputersSupermicro inventoryChub brought the replacement Supermicro
we ordered to the 40m today. I stored
it at the SW entrance to the VEA,
  
  14819   Wed Jul 31 09:41:12 2019 gautamUpdateBHDOMC cavity geometrySummary:

We need to determine the geometry
(= round-trip length and RoC of curved mirrors) of
 paramSpaceHeatMap.pdf 
  14818   Tue Jul 30 20:11:12 2019 ranaSummaryIMCIMC ASC: thoughts and hopesOne of the
biggest challenges in LIGO is reducing the
alignment control noise.
  
  14817   Tue Jul 30 09:13:31 2019 gautamUpdatePSLc1psl keyed, Agilent setup cleared
IMC would not lock. c1psl EPICS
channels were unresponsive. I keyed the crate
and went through the usual burtrestore/PMC-relocking
  
  14816   Mon Jul 29 19:08:55 2019 yehonathanUpdateLoss MeasurementReviving loss measurement by reflection1. X arm is totally misaligned in order
to measure the Y arm loss using the reflection
method. Each measurement round consists of
  
  14815   Mon Jul 29 13:32:56 2019 gautamUpdateLoss MeasurementLoss measurement PD installed in AS path[yehonathan, gautam]


we placed a PDA520 photodiode in
the AS beampath, so AS110 and AS55 no longer
  
  14814   Fri Jul 26 19:53:53 2019 JonOmnistructureCamerasGigE Camera ServerI've started setting up the last
new rackmount SuperMicro as a dedicated server
for the GigE cameras. The new machine is
  
  14813   Thu Jul 25 20:08:36 2019 gautamUpdateComputersSolidworks machineI brought one CPU (Dell T3500) and one
28" monitor from Mike Pedraza's
office in Downs to the 40m. It is on Steve's
  
  14812   Thu Jul 25 14:28:03 2019 gautamConfigurationComputersfirewalld disabled for EPICS CAI think rana did some more changes to this
workstation to make it useful for commissioning
activities - but the MEDM screens were still
 donatellaCommissioning.png 
  14811   Thu Jul 25 12:25:56 2019 gautamUpdateALSIR ALSX noiseSummary:


There are some broad peaks
in the ALS out-of-loop noise, centered at
 ALS_broadPeaks.pdf 
  14810   Thu Jul 25 09:19:32 2019 JonUpdateCamerasUpgraded Pylon from 5.0.12 to 5.2.0I'll keep developing the camera server
on a parallel track using the "new_..."
directory naming convention. One thing I
  
  14809   Thu Jul 25 00:26:47 2019 MilindUpdateCamerasConvolutional neural networks for beam trackingSomehow I never got around to doing the
pixel sum thing for the new real data from
the GigE. Since I have to do it for the presentation,
 centroid.pdfsubplot_yaw_test.pdf 
  14808   Wed Jul 24 20:23:52 2019 gautamUpdateCamerasUpgraded Pylon from 5.0.12 to 5.2.0Since there are multiple SURF projects
that rely on the cameras:


I moved the new installs
  
  14807   Wed Jul 24 20:05:47 2019 MilindUpdateCamerasCNNs for beam tracking || Tales of desperationAt the lab meeting today, Rana suggested
that I use the Pylon app to collect more
data if that's what I need. Following
 saturation_percentage.pdf 
  14806   Wed Jul 24 16:45:32 2019 JonUpdateCamerasUpgraded Pylon from 5.0.12 to 5.2.0I upgraded Pylon, the C/C++ API for the
GigE cameras, to the latest release, 5.2.0.
It is installed in the same location as before,
 IMG_3525.jpg 
  14805   Wed Jul 24 12:24:43 2019 MilindUpdateIOOunstick.py and ifotestMoved the unstick.py code to the ifotest
repository here.
It now handles signals like those generated
  
  14804   Wed Jul 24 04:20:35 2019 KruthiUpdateCalibration-RepairMC2 pitch and yaw calibrationSummary:  I calibrated MC2
pitch and yaw offsets to spot position in
mm. Here's what I did:
 Pitchyaw_calibration.png 
  14803   Wed Jul 24 02:06:05 2019 KruthiUpdateCamerasHDR imagesI have been trying a couple of HDR algorithms,
all of them seem to give very different results.
I don't know how suitable these algorithms
 6x 
  14802   Wed Jul 24 00:22:24 2019 gautamUpdateALSPSL frequency locked to XARM length using ALSSummary:

I succeeded in locking the PSL
frequency to the XARM cavity length, with
 ALS_X_noise_POX.pdfBR_comparison.pdfALS_CARM_OLG.pdf 
  14801   Tue Jul 23 21:59:08 2019 JonUpdateCamerasPlan for GigE camerasThis afternoon Gautam and I assessed what
to do about restoring the GigE camera software.
Here's what I propose:
  
  14800   Mon Jul 22 23:53:16 2019 gautamUpdateALSIR ALS locking attemptSummary:

My goal tonight was to lock the
PSL frequency to be resonant in the
 ALS_X_outOfLoopnoise.pdfALS_X_inLoopnoise.pdfCARM_loopShape.pdf 
  14799   Mon Jul 22 21:04:40 2019 ranaUpdateComputersmaking rossa great again
copied over /etc/fstab lines from
pianosa sothat the NFS mounts work correctly
added
symlinks so that the NFS dirs mount in the
 seis.pdf 
  14798   Mon Jul 22 13:32:55 2019 KruthiUpdateSUSTest mass pitch adjustment test[Kruthi, Milind]

On Friday, Milind and I performed
the pitch adjustment test Rana had asked
 6x 
  14797   Mon Jul 22 13:26:41 2019 KruthiUpdateIOOMC lockedThe MC2 has 2
GigE cameras right now. I'll
put back the analog asap.
  
  14796   Mon Jul 22 12:57:35 2019 KruthiUpdateLoss MeasurementMC2 loss mapIn my script I have used "CAM-ETMX_SNAP"
only; while entering it in the elog
I made a mistake, my bad!
  