40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 41 of 344  Not logged in ELOG logo
ID Date Authorup Type Category Subject
  5196   Thu Aug 11 16:15:59 2011 Ishwita, ManuelUpdatePEMCalibration of Guralp and STS2

 

Finally, we have found the correct calibration of Guralp and STS2 seismometers.

ADC: 216counts = 20V Hence, calibration of ADC is 3.2768e+03 counts/V.


 

GURALP

Sensitivity of seismometer = 800 V/ms-1

Gain of the guralp breakout box (reference elog entry) = 20

Calibration = 3.2768e+03 counts/V x 800  V/ms-1 x 20 = 52428800 counts/ms-1 -----> 1.9073e-08 ms-1/count


STS

Sensitivity = 1500 V/ms-1

Gain of the STS electronic breakout box = 10

Calibration = 3.2768e+03 counts/V x 1500  V/ms-1 x 10 = 49152000 counts/ms-1 -----> 2.0345e-08 ms-1/count

 

  5201   Fri Aug 12 00:18:30 2011 Ishwita, ManuelUpdatePEMCoherence of Guralp1 and STS2(Bacardi, Serial NR 100151)

 

 We moved the seismometer STS2(Bacardi, Serial NR 100151) as we told in this Elog Entry, so the distance between Guralp1 and STS2 is 31.1m. Following is the coherence plot for this case:

coher_gur1_sts1_31m.png

then we also moved the Guralp1 under the BS and plugged it with the Guralp2 cable (at 7:35pm PDT), so now the distance between the two seismometers is 38.5m. Following is the coherence plot for this case:

coher_gur1_sts1_38.png

  14049   Tue Jul 10 16:59:12 2018 Izabella PastranaHowToComputer Scripts / ProgramsTaking Remote TF Measurements with the Agilent 4395A

I copied the netgpibdata folder onto rossa (under the directory ~/Agilent/), which contains all the necessary scripts and templates you'll need to remotely set up, run, and download the results of measurements taken on the AG4395A network analyzer. The computer will communicate with the network analyzer through the GPIB device (plugged into the back of the Agilent, and whose communication protocol is found in the AG4395A.py file in the directory ~/Agilent/netgpibdata/).

The parameter template file you'll be concerned with is TFAG4395Atemplate.yml (again, under ~/Agilent/netgpibdata/), which you can edit to fit your measurement needs. (The parameters you can change are all helpfully commented, so it's pretty straightforward to use! Note: this template file should remain in the same directory as AGmeasure, which is the executable python script you'll be using). Then, to actually set up, run, and download your measurement, you'll want to navigate to the ~/Agilent/netgpibdata/ directory, where you can run on the command line the following: python AGmeasure TFAG4395Atemplate.yml

The above command will run the measurement defined in your template file and then save a .txt file of your measured data points to the directory specified in your parameters. If you set up the template file such that the data is also plotted and saved after the measurement, a .pdf of the plot will be saved along with your .txt file.

Now if you want to just download the data currently on the instrument display, you can run: python AGmeasure -i 192.168.113.105 -a 10 --getdata

Those are the big points, but you can also run python AGmeasure --help to learn about all the other functions of AGmeasure (alternatively, you can read through the actual python script).

Happy remote measuring! :)

 

 

 

 

  16794   Thu Apr 21 11:31:35 2022 JCUpdateVACGauges P3/P4

[Jordan, JC]

It was brought to attention during yesterday's meeting that the pressures in the vacuum system were not equivalent althought the valve were open. So this morning, Jordan and I reviewed the pressure gauges P3 and P4. We attempted to recalibrate, but the gauges were unresponsive. Following this, we proceeded to connect new gauges on the outside to test for a calibration. The two gauges successfully calibrated at atmosperic pressure. We then proceeded to remove the old gauges and install the new ones. 

  16808   Mon Apr 25 14:19:51 2022 JCUpdateGeneralNitrogen Tank

Coming in this morning, I checked on the Nitrogen tanks to check the level. One of the tanks were empty, so I went ahead and swapped it out. One tank is at 946 PSI, the other is at 2573 PSI. I checked for leaks and found none.

  16814   Wed Apr 27 10:05:55 2022 JCUpdateCoil DriversCoil Drivers Update

18 (9 pairs) Coil Drivers have been modified. Namely ETMX/ITMX/ITMY/BS/PRM/SRM/MC1/MC2/MC3.

 

ETMX Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3        S2100624                     ETMX Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100631

ITMX Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100620                      IMTX Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100633

ITMY Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100623                   ITMY Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100632

BS Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3            S2100625                     BS Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100649

PRM Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100627                      PRM Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100650

SRM Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100626                        SRM Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100648

MC1 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100628                        MC1 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100651

MC2 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3       S2100629                        MC2 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3         S2100652

MC3 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3        S2100630                        MC3 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3         S2100653

 

 

Will be updating this linking each coil driver to the DCC

  16820   Fri Apr 29 08:34:40 2022 JCUpdateVACRGA Pump Down

Jordan and I, in order to start pumpig down the RGA Volume, we began by opening V7 and VM. Afterwards, we started RP1 and RP3. After this, the pressure in the line between RP1, RP3, and V6 dropped to 3.4 mTorr. Next, we tried to open V6, although an error message popped up. We haven't been able to erase it since. But we were able to turn on TP2 with V4 closed. The pressure in that line is reporting 1.4 mTorr.

 

PRP on the sitemap is giving off an incorrect pressure for the line between RP1, RP3, and V6. This is verified by the pressure by the control screen and the physical controller as well. 

  16823   Mon May 2 13:30:52 2022 JCUpdateCoil DriversCoil Drivers Update

The DCC has been updated, along with the modified schematic. Links have been attached.

Quote:

18 (9 pairs) Coil Drivers have been modified. Namely ETMX/ITMX/ITMY/BS/PRM/SRM/MC1/MC2/MC3.

 

ETMX Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3        S2100624                     ETMX Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100631

ITMX Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100620                      IMTX Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100633

ITMY Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100623                   ITMY Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100632

BS Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3            S2100625                     BS Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100649

PRM Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100627                      PRM Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100650

SRM Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100626                        SRM Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100648

MC1 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3         S2100628                        MC1 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3      S2100651

MC2 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3       S2100629                        MC2 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3         S2100652

MC3 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3        S2100630                        MC3 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3         S2100653

 

 

Will be updating this linking each coil driver to the DCC

 

  16825   Tue May 3 13:18:47 2022 JCUpdateVACRGA Pump Down

Jordan, Tega, JC

Issue has been resolved. Breaker on RP1 was tripped so the RP1 button was reporting ON, but was not actually on which continuously tripped the V6 interlock. Breaker was reset, RP1 and RP3 turned on. The V6 was opened to rough out the RGA volume. Once, pressure was at ~100mtorr, V4 was opened to pump the RGA with TP2. V6 was closed and RP1/3 were turned off.

RGA is pumping down and will take scans next week to determine if a bakeout is needed

Quote:

Jordan and I, in order to start pumpig down the RGA Volume, we began by opening V7 and VM. Afterwards, we started RP1 and RP3. After this, the pressure in the line between RP1, RP3, and V6 dropped to 3.4 mTorr. Next, we tried to open V6, although an error message popped up. We haven't been able to erase it since. But we were able to turn on TP2 with V4 closed. The pressure in that line is reporting 1.4 mTorr.

 

PRP on the sitemap is giving off an incorrect pressure for the line between RP1, RP3, and V6. This is verified by the pressure by the control screen and the physical controller as well. 

 

  16842   Tue May 10 15:46:38 2022 JCUpdateBHDRelocate green TRX and TRY components from PSL table to BS table

[JC, Tega]

Tega and I cleaned up the BS OPLEV Table and took out a couple of mirrors and an extra PD. The PD which was removed is "IP-POS - X/Y  Reversed". In addition to this, the cable is zip-tied to the others located on the outside of the table in case this is required later on. 

Next, we placed the cameras and mirrors for the green beam into their postions. A beam splitter and 4 mirrors were relocated from PSL table and placed onto the BS Oplev table to complete this. I will upload the picture of the newly updated photo with arrows of the beam routes.

  16845   Wed May 11 15:49:42 2022 JCUpdateOPLEV TablesGreen Beam OPLEV Alignment

[Paco, JC]

Paco and I began aligning the Green Beam in the BS Oplev Table. while aligning the GRN-TRX, the initial beam was entering the table a bit low. To fix this, Paco went into the chamber and correcting the pitch with the steering mirror. The GRN-TRX is now set up, both the PD and Camera. Paco is continuing to work on the GRN-TRY and will update later on today. 

In the morning, I will update this post with photos of the new arrangement of the BS OPLEV Table.


Update Wed May 11 16:54:49 2022

[Paco]

GRY is now better mode matched to the YARM and is on the edge of locking, but it more work is needed to improve the alignment. The key difference this time with respect to previous attempts was to scan the two lenses on translation stages along the green injection path. This improved the GTRY level by a factor of 2.5, and I know it can be further improved. Anyways, the locked HOMs are nicely centered on the GTRY PD, so we are likely done with the in-vac GTRY GTRX alignment.


Update Wed May 12 10:59:22 2022

[JC]

The GTRX PD is now set up and connected. The camera have been set to an angle because the cable to connect it is too thick for the camera to maintain its original position along the side. 

 

  16851   Fri May 13 14:26:00 2022 JCUpdateAlignmentLO2 Beam

[Yehonathan, JC]

Yehonathan and I attempted to align the LO2 beam today through the BS chamber and ITMX Chamber. We found the LO2 beam was blocked by the POKM1 Mirror. During this attempt, I tapped TT2 with the Laser Card. This caused the mirror to shake and dampen into a new postion. Afterwards, when putting the door back on ITMX, one of the older cables were pulled and the insulation was torn. This caused some major issues and we have been able to regain either of the arms to their original standings.

  16871   Tue May 24 11:04:53 2022 JCUpdateVACBeginning Pumpdown

[JC, Jordan, Paco, Chub]

We began with the pumpdown this morning. We started with the annulus volume and proceeded by using the following:

1. Isolate the RGA Volume by closing of valves VM3 and V7.

2. Opened valves VASE, VASV, VABSSCT, VABS, VABSSCO, VAEV, and VAEE, in that order.

3. Open VA6 to allow P3, FRG3, and PAN to equalize.

4. Turn on RP1 and RP3, rough out annulus volume, once <1 torr turn on TP3. Close V6. Open V5 to pump the annulus volume with TP3.

5. Re route pumping from RP1 and RP3 to the main volume by opening V3 and slowly opening RV1.

6. After ~3.5 hours the pressure in the arms was <500mtorr on both FRG1 and P1a. Turn on TP1 and wait to reach full speed 560 Hz

7. Open V1 with RV2 barely open. The pressure diff between P1a and P2/FRG2 needs to be below 1 torr. This took a couple attempts with the manual valve in different positions. The interlocks were tripped for this reason. Repeat step 7 until the manual gate valve was in a position that throttled pumping enough to maintain the <1 torr differential.

8. Slowly open the manual gate valve over the course of ~ 1 hour. Once the manual gate valve fully opened, pressure in the arms was <1mtorr.

9. V7 was closed, leaving only TP2 to back TP1, while TP3 was used to continue pumping the annuli. Left in that configuration overnight (see attached)

 

We did have to replace gauge PAN becuase it was reading a signal error. In addition, we found the cable is a bit sketchy and has a sharp bend. The signal comes in and out when the cable is fiddled with.

  16878   Fri May 27 12:15:30 2022 JCUpdateElectronicsCRT TV / Monitor 6

[Yehonathan, Paco, Yuta, JC]

As we were cleaning up this morning, we heard a high pitch sound that turned into a buzz. After searching for where the sound came from, we noticed the CRT TV went out. We swapped this out with a moniter and used a BNC to VGA adapter to display the cameras.

  16882   Tue May 31 14:44:02 2022 JCUpdateElectronicsCRT TV / Monitor 6

[Paco, JC]

Paco and I fixed the ethernet cable which was hanging. We stopped models c1x07 and c1su2, realigned the cable to follow the shelf from top, and returned to turn on the computers.

 

Note: There was not a long enough ethernet cable, so we used a female to female adapter and attached 2 ethernet cables.

Quote:

[Yehonathan, Paco, Yuta, JC]

As we were cleaning up this morning, we heard a high pitch sound that turned into a buzz. After searching for where the sound came from, we noticed the CRT TV went out. We swapped this out with a moniter and used a BNC to VGA adapter to display the cameras.

 

  16891   Mon Jun 6 09:37:16 2022 JCUpdateSUSInMatCalc

[Paco, JC]

Paco and I attempted to calculate the input matrices from May 24, 2022, but the OSEM data was all saturated and not very useful. Therefore, we decided to manually investigate the appropriate coil offsets for all BHD SUS. Before, the default offset kick was 30000 counts, but we found that LO1, AS1, AS4, and PR2 cannot take more than 5000 counts. As for LO2, SR2, and PR3 cannot take more than 2000 counts before saturating. Note that all these kick test were taken by kicking OSEM UL on all BHD Optics.

 

We started the freeSwing.py script on tmux freeSwing session for tomorrow at 1:00 am for only the 5000 count offset SUS.

 

  16906   Fri Jun 10 13:52:22 2022 JCUpdateOPLEV TablesITMX, ITMY, and Vertex Table Beam Paths

I have at taken photos and added arrows which signify the beam paths for ITMX, ITMY, and Vertex Oplev tables.

  16912   Tue Jun 14 08:41:36 2022 JCUpdateOPLEV TablesBS Oplev Table Sketch

[JC]

Lately, I have been working on a 3d sketch of the BS OPLEV Table on SolidWorks. This is my progress so far, a few of the components I will have to sketch myself, such as the HeNe laser and photodiodes. This will just be a general layout of the HeNe laser, optics, and photodiodes.

  16916   Wed Jun 15 07:26:35 2022 JCUpdateVACBeginning Pumpdown

[Jordan, JC]

Jordan and I went in to retore the Vacuum System back to it's original state before the power loss on June 8, 2022. The process went smoothly as we first closed V7 and opened VM3 (in that order).

The RP1/3 line did not have the KF blank installed. That was added and the RP flex line was capped off.

Quote:

[JC, Jordan, Paco, Chub]

We began with the pumpdown this morning. We started with the annulus volume and proceeded by using the following:

1. Isolate the RGA Volume by closing of valves VM3 and V7.

2. Opened valves VASE, VASV, VABSSCT, VABS, VABSSCO, VAEV, and VAEE, in that order.

3. Open VA6 to allow P3, FRG3, and PAN to equalize.

4. Turn on RP1 and RP3, rough out annulus volume, once <1 torr turn on TP3. Close V6. Open V5 to pump the annulus volume with TP3.

5. Re route pumping from RP1 and RP3 to the main volume by opening V3 and slowly opening RV1.

6. After ~3.5 hours the pressure in the arms was <500mtorr on both FRG1 and P1a. Turn on TP1 and wait to reach full speed 560 Hz

7. Open V1 with RV2 barely open. The pressure diff between P1a and P2/FRG2 needs to be below 1 torr. This took a couple attempts with the manual valve in different positions. The interlocks were tripped for this reason. Repeat step 7 until the manual gate valve was in a position that throttled pumping enough to maintain the <1 torr differential.

8. Slowly open the manual gate valve over the course of ~ 1 hour. Once the manual gate valve fully opened, pressure in the arms was <1mtorr.

9. V7 was closed, leaving only TP2 to back TP1, while TP3 was used to continue pumping the annuli. Left in that configuration overnight (see attached)

 

We did have to replace gauge PAN becuase it was reading a signal error. In addition, we found the cable is a bit sketchy and has a sharp bend. The signal comes in and out when the cable is fiddled with.

 

  16943   Fri Jun 24 12:13:16 2022 JCUpdateIOOWFS issues

[Yuta, JC]

It seems that early this morning MC got very misaligned. Yuta was able to align the Mode Cleaner again by individually adjusting the MC1 MC2, and MC3. Once transmission reach ~12000, we went ahead and turned on WFS. Oddly enough, the transmission began plummeting and MC fell out of lock. After this, Yuta reset the WFS offsets and realigned the WFS QPDs. We then locked MC and turned on WFS once again, but the same issue happened. After fiddeling around with this, we found the if we set C1:IOO-MC2_TRANS_PIT_OUTPUT and C1:IOO-WFS1_YAW_OUTPUT equal to 0, WFS does not cause this issue. Is there a proper to reset WFS, aside from only zeroing the offsets?

  16953   Tue Jun 28 09:03:58 2022 JCUpdateGeneralOrganizing and Cleaning

The plan for the tools in 40m

As of right now, there are 4 tool boxes. X-end, Y-end, Vertex, and the main tool box along the X-arm. The plan is the give each toolbox a set of their own tools. The tools of X-end, Y-end, and Vertex toolboxes will be very similar containing the basic tools such as pliers, screwdrivers, allen ball drivers. Along with this, each tool box will have a tape measure, caliper, level, and other measuring tools we find convinient. 

As for the new toolbox, I have done research and found a few good selections. The only problem I have ran into with this is the width of the tool box corresponding with the prices. The tool cabinet we have now is 41" wide. The issue I have is not in finding another toolbox of the same width, but for a similar price we can find a 54" wide tool cabinet. Would anyone be objected to making a bit more space for this?

How the tools will stay organized.

I the original idea I had was to use a specified color of electrical tape for each tool box. Then to wrap the corresponding tools tools with the same color tape. But it was brought to my attention that the electrical tape would become sticky over time. So, I think the using the label maker would be the best idea. with the labels being 'X' for X-end, 'Y' for Y-end, 'V' for vertex, and 'M' for main toolboxes.

An idea for the optical tables:

Anchal brought it up to me that it is a hassle to go back and forth searching for the correct sizes of Hex Keys and Allen Wrenches. The idea of a pouch on the outside of each optical table was mentioned so I brought this up to Paco. Paco also gave me the idea of a 3D printed stand we could make for allen ball drives. Does anyone have a preference or an idea of what would be the best choice and why? 


A few sidenotes: 

Anchal mentioned to me a while back that there are many cables that are laying on the racks that are not being used. Is there a way we could identify which ones are being used? 

I noticed that when we were vented that a few of the chamber doors were leaning up against the wall and not on a wooden stand like others. Although, the seats for the chamber doors are pretty spacious and do not give us much clearance. For the future ones, could we make something more sleek and put the wider seats at the end chambers?

The cabinets along the Y-Arm are labelled, but do not correspond with all the materials inside or are too full to take in more items. Could I organize these? 
 

  16980   Fri Jul 8 14:03:33 2022 JCHowToVACVacuum Preparation for Power Shutdown

[Koji, JC]

Koji and I have prepared the vacuum system for the power outage on Saturday.

  1. Closed V1 to isolate the main volume.
  2. Closed of VASE, VASV, VABSSCI,VABS, VABSSCO, VAEV, and VAEE.
  3. Closed V6, then close VM3 to isolate RGA
  4. Turn off TP1 (You must check the RPMs on the TP1 Turbo Controller Module)
  5. Close V5
  6. Turn off TP3 (There is no way to check the RPMs, so be patient)
  7. Close V4 (System State changes to 'All pneumatic valves are closed)
  8. Turn off TP2 (There is no way to check the RPMs, so be patient)
  9. Close Vacuum Valves (on TP2 and TP3) which connect to the AUX Pump.
  10. Turn of AUX Pump with the breaker switch wall plug.

From here, we shutdown electronics.

  1. Run /sbin/shutdown -h now on c1vac to shut the host down.
  2. Manually turn off power to electronic modules on the rack.
    • GP316a
    • GP316b
    • Vacuum Acromags
    • PTP3
    • PTP2
    • TP1
    • TP2 (Unplugged)
    • TP3 (Unplugged)

 

  16983   Mon Jul 11 11:16:45 2022 JCSummaryElectronicsStartup after Shutdown

[Paco, Yehonathan, JC]

We began starting up all the electronics this morning beginning in the Y-end. After following the steps on the Complete_Power_Shutdown_Procedures on the 40m wiki, we only came across 2 issues.

  1. The Green beam at the Y-End : Turn on the controller and the indicator light began flashing. After waiting until the blinking light becomes constant, turn on the beam. 
  2. C1lsc "could not find operating system"-unable to SSH from Rossa : We found an Elog of how to restart Chiara and this worked. We proceeded by adding this to the procedures of startup.
  16985   Mon Jul 11 15:26:12 2022 JCHowToVACStartup after Power Outage

[Koji, Jc]

Koji and I began starting that Vacuum system up.

  1. Reverse order step 2 of shutting down electronics. Anthing after, turn on manually.
  2. If C1vac does not come back, then restart by holding the reset button.
  3. Open VA6
  4. Open VASE, VASV,VABSSCI, VABS, VABSSCO, VAEV, and VAEE
  5. Open V7
  6. Check P3 and P2, if they are at high pressure, approx. 1 Torr range, then you must use the roughing pumps.
  7. Connect Rotary pump tube. (Manually)
  8. Turn on AUX Pump
  9. Manually open TP2 and TP3 valves.
  10. Turn on TP2 and TP3, when the pumps finish startup, turn off Standby to bring to nominal speed.
  11. Turn on RP1 and RP3
  12. Open V6
  13. Once P3 reaches <<1 Torr, close V6 to isolate the Roughing pumps.
  14. When TP2 and TP3 are at nominal speed, open V5 and V4.
  15. Now TP1 is well backed, turn on TP1.
  16. When TP1 is at nominal speed, Open V1.
  16995   Wed Jul 13 07:16:48 2022 JCUpdateElectronicsChecking Sorensen Power Supplies

[JC]

I went around 40m picking up any Sorensens that were laying around to test if they worked, or in need of repair. I gathered up a total of 7 Sorensens and each one with a Voltmeter. I made sure the voltage would rise on the Sorenson as well as the voltmeter, maxing out at ~33.4 Volts. For the current, the voltmeter can only rise to 10 Amps before it is fused. Many of the Sorensons that I found did not have their own wall connection, so I had to use the same one for multiple.

From these 7, I have found 5 that are well. One Sorenson I have tested has a output shortage above 20V and the other has yet to be tested.

  17005   Fri Jul 15 12:21:58 2022 JCUpdateElectronicsChecking Sorensen Power Supplies

Of the 7 Sorenson Power Supplies I tested, 5 are working fine, 1 cannot output voltage more than 20 Volts before shorting, and other does not output current. Six Sorensons are behind the X-Arm.

Quote:

[JC]

I went around 40m picking up any Sorensens that were laying around to test if they worked, or in need of repair. I gathered up a total of 7 Sorensens and each one with a Voltmeter. I made sure the voltage would rise on the Sorenson as well as the voltmeter, maxing out at ~33.4 Volts. For the current, the voltmeter can only rise to 10 Amps before it is fused. Many of the Sorensons that I found did not have their own wall connection, so I had to use the same one for multiple.

From these 7, I have found 5 that are well. One Sorenson I have tested has a output shortage above 20V and the other has yet to be tested.

 

  17019   Tue Jul 19 17:18:34 2022 JCUpdateElectronicsNew Coil Driver on Rack 1X3

[Yehonathan, JC]

Yehonathan and I began to put the electronics on Rack 1X3. To do this, we had to move the monitor over the the PD testing table. Before mounting the Coil Drivers, we added numbers to the spaces to follow the rack plan Koji has provided. The drivers which have been mounted are PRM (Slots 10,11), BS (Slots 15, 16), ITMX (Slots 26, 27), and ITMY (34, 35).

  17078   Fri Aug 12 13:40:36 2022 JCUpdateGeneralPreparing for Shutdown on Saturday, Aug 13

[Yehonathan, JC]

Our first step in preparing for the Shutdown was to center all the OpLevs. Next is to prepare the Vacuum System for the shutdown.

 

  17099   Tue Aug 23 14:59:15 2022 JCUpdateToolsNew Toolbox at Y-End

A new tool box has been placed at the Y-end! Each drawer has its label so PLEASE put the tools back in their correct location. In addition to this, Each tool has its assigned tool box, so PLEASE RETURN all tools to their designated tool box. The tools can be distinguished by a writing or heat shrink which corresponds to the color of the tool chest or location. Photo #2 is an example of how the tools have been marked.

Each toolbox from now on will contain a drawer for the folllowing: Measurements, Allen Keys, Pliers and Cutters, Screwdrivers, Zipties and Tapes, Allen Ball Drivers, Crescent Wrenches, Clamps, and Torque Wrenches/ Ratchets.

  17110   Mon Aug 29 13:33:09 2022 JCUpdateGeneralLab Cleanup

The machine shop looked a mess this morning, so I cleaned it up. All power tools are now placed in the drawers in the machine shop. Let me know if there are any questions of where anything here is placed. 

  17126   Thu Sep 1 09:00:02 2022 JCConfigurationDaily ProgressLocked both arms and aligned Op Levs

Each morning now, I am going to try to align both arms and lock. Along with that, sometime at towards the end of each week, we should align the OpLevs. This is a good habit that should be practiced more often, not only by me. As for the Y Arm, Yehonathan and I had to adjust the gain to 0.15 in order to stabilize the lock.

  17132   Tue Sep 6 09:57:26 2022 JCSummaryGeneralLab cleaning

DB9 Cables have been assorted and placed behind the Y-Arm. Long BNC Cables and Ethernet Cables have been stored under the Y-Arm. 

Quote:

We held the lab cleaning for the first time since the campus reopening (Attachment 1).
Now we can use some of the desks for the people to live! Thanks for the cooperation.

We relocated a lot of items into the lab.

  • The entrance area was cleaned up. We believe that there is no 40m lab stuff left.
    • BHD BS optics was moved to the south optics cabinet. (Attachment 2)
    • DSUB feedthrough flanges were moved to the vacuum area (Attachment 3)
  • Some instruments were moved into the lab.
    • The Zurich instrument box
    • KEPCO HV supplies
    • Matsusada HV supplies
  • We moved the large pile of SUPERMICROs in the lab. They are around MC2 while the PPE boxes there were moved behind the tube around MC2 area. (Attachment 4)
  • We have moved PPE boxes behind the beam tube on XARM behind the SUPERMICRO computer boxes. (Attachment 7)
  • ISC/WFS left over components were moved to the pile of the BHD electronics.
    • Front panels (Attachment 5)
    • Components in the boxes (Attachment 6)

We still want to make some more cleaning:

- Electronics workbenches
- Stray setup (cart/wagon in the lab)
- Some leftover on the desks
- Instruments scattered all over the lab
- Ewaste removal

 

  17135   Thu Sep 8 11:54:37 2022 JCConfigurationLab OrganizationLab Organization

The arms in the 40m laboratory have now been sectioned off. Each arm has been divided up into 15 sections. Along the Y arm, the section are labelled "Section Y1 - Section Y15". For the X arm, they are labelled "Section X1- Section X15". Anything changed or moved will now be updated into the elog with their appropriate section.

 

Below is an example of Section X6.

  17136   Thu Sep 8 12:01:02 2022 JCConfigurationLab OrganizationLab Organization

The floor cable cover has been changed out for a new one. This is in Section X11.

  17139   Wed Sep 14 15:40:51 2022 JCUpdateTreasurePlastic Containers

There are brand new empty plastic containers located inside the shed that is outside in the cage. These can be used for organizing new equipment for CDS, or cleanup after Wednesday meetings

  17154   Fri Sep 23 10:05:38 2022 JCUpdateVACN2 Interlocks triggered

[Chub, Anchal, Tega, JC]

After replacing an empty tank this morning, I heard a hissing sound coming from the nozzle area. It turns out that this was from the copper tubing. The tubing we slightly broken and this was comfirmed with soapy water bubbles. This caused the N2 pressure to drop and the Vac interlocks to be triggered. Chub and I went ahead and replaced the fitting in connected this back to normal. Anchal and Tega have used the Vacuum StartUp procedures to restore the vacuum to normal operation.

Adding screenshot as the pressure is decreasing now.

  17164   Thu Sep 29 15:12:02 2022 JCUpdateComputersSetup the 6 new front-ends to boot off the FB1 clone

[Jamie, Christopher, JC]

This morning we decided to label the the fiber optic cables. While doing this, we noticed that the ends had different label, 'Host' and 'Target'. Come to find out, the fiber optic cables are directional. Four out of Six of the cables were reversed. Luckily, 1 cable for the 1Y3 IO Chassis has a spare already laid (The cable we are currently using).  Chris, Jamie, and I have begun reversing these cable to there correct position.

Quote:

[Tega, JC]

We laid 4 out of 6 fiber cables today. The remaining 2 cables are for the I/O chassis on the vertex so we would test the cables the lay it tomorrow. We were also able to identify the problems with the 2 supposedly faulty cable, which are not faulty. One of them had a small bend in the connector that I was able to straighten out with a small plier and the other was a loose connection in the switch part. So there was no faulty cable, which is great! Chris wrote a matlab script that does the migration of all the model files. I am going through them, i.e. looking at the CDS parameter block to check that all is well. Next task is to build and install the updated models. Also need to update the '/opt/rtcds' and '/opt/rtapps' directory to the latest in the 40m chiara system.

 

 

  17187   Thu Oct 13 14:46:34 2022 JCUpdateLab OrganizationLab Cleanup 10/12/2022

During Wednesday’s lab clean up, we made a ton of progress in organization. Our main focus was to tackle CDS debris from the ongoing upgrade. We proceeded with the following tasks.

  • Loop the OneStop Cables and mark ‘good’ or ‘bad’.
  • Clean materials from the PD testing table.
  • Removed the SuperMicro boxes from the lab
  • Vacuum area organization.
  • Remove old plastic containers from the laboratory.
  • Relocate Koji’s electronics underneath Y-Arm
  • Arrange cabling to the TestStand and create clearance to the Machine Shop/Laboratory exit.

Attachment #2 shows that all the CDS equipment has been relocated behind Section X3 of the X-Arm.

  17194   Mon Oct 17 17:42:35 2022 JCHowToOPLEV TablesIMC Reflected beam sketch

I sketched up a quick drawing with estimated length for the IMC reflected beam. This includes the distances and focal length. Distances are from optic to optic. 

  17208   Tue Oct 25 08:25:00 2022 JCUpdateBHDBHD fringe aligned with reduced LO and AS beam clipping

I aligned today using this scheme. I couldn't seem to get C1:IOO-MC_TRANS_SUM above 13400 by using WFS or manually aligning. The original state before was the following:
                  Pitch       Yaw     
C1:SUS-MC1: 
    -0.4672     -0.7714
C1:SUS-MC2:      4.0446     -1.3558
C1:SUS-MC3:     -2.0006      1.6001

 

  17229   Fri Nov 4 15:43:05 2022 JCUpdateLab OrganizationNew Toolbox Organization

The new tool box has came in! I have spent serious time organizing these tools and making it look pretty, so please take care of it! A few things I would like to note.

  • The tool box has a power strip, so this is now where we will be charging the power tools. There is also a selected cabinet for the fully charged batteries, bits, and the power tools themselves. 
  • The cart next to the tool box had to turned 90 degrees in order to fit the tool box in. While moving this, I noticed this cart is piled with random stuff, let’s tackle this on Wednesdays clean up.
  • The tools from this tool box are specifically assigned to the VERTEX area (and the X-End until we get another toolbox.) Please return all the tools in their right tool box and cabinet after usage. 
  • The heavier duty equipment such as pipe wrenches, hammers, and power tools will remained stored here in the bottom cabinets. 

Hope you all like it!

  17230   Mon Nov 7 08:36:10 2022 JCUpdateSUSIMC out of alignment -- f2A failure

[Paco, JC]

We came in this morning and noted the IMC was grossly misaligned, with MC3 still damped but with >= 100 rms motion in all coil monitors no (a lot but not enough to trip the WD)... Turning off the WFS didn't do much so it was obviously an issue with the recent f2A output filters, so we turned all off (though only MC3 had this excess motion). After this we aligned IMC, engaged the lock and turned WFS back on.

There was no elog about f2A beyond this test scheduled to run Friday, I guess the filters were meant to stay on long term?

  17258   Fri Nov 11 19:11:50 2022 JCUpdateGeneralWFS Whitening and Demod Boards

WFS Whitening and Demod boards were scavenged. ' iLIGO WFS cards are in big plastic boxes placed on the north wall around Section Y5 or Y6 (not under the arm). The WFS head PCBs, empty WFS housing, WFS components, I/Q demod components are at SectionY10 under the arm tube. ' - Koji . I have taken pictures of the boards and will upload them to the DCC once I find a way to add a Serial Number . (I will upload this to the eLog as a HowTo) The next step is to search for at least 2 extender boards to troubleshoot these board and find if there are any issues. We may have replace some components and retune the boards. Attachment #1 is an example of the WFS Whitening Filter and Attachment #2 is and example of the the WFS Demod Board.

If you use the Camera DO NOT remove the strap. I have also purchased lens caps for the camera, so after usage PUT THE LENS CAP BACK ON. 

  17275   Thu Nov 17 07:39:01 2022 JCConfigurationCamerasITMX Camera

Coming in this morning, I found ITMX Camera malfunctioning.

  17276   Thu Nov 17 09:06:35 2022 JCSummaryGeneralLight Replace

Electrical Shop came in today to replace the lightbulbs around 40m. I supervised the entire visit, inside and outside the lab area. I was sure to use the ladders which were already inside the lab area. Although, while I was crawling under the MC beam tube, I came up too quickly and bumped it. This may have caused MC to become misaligned. Anyways, Yehonathan and I are realigning now.

 

  17277   Thu Nov 17 11:24:39 2022 JCHowToLSCLocking MICH

[Yuta, JC]

Here is the Yuta's Alignment Scheme from elog 17056  with these slight adjustments:

Current alignment scheme:
Current alignment scheme I figured out is the following.
 - Check Y green. If it is transmitted at good spot on GTRY camera, Yarm is OK. If not, tweak ITMY/ETMY. alignment.
 - Mis-align AS4, align TT1, TT2, LO1 to have BHDC_A_OUT of ~115 and BHDC_B_OUT of ~95.
 - Align PR3, PR2 to maximize TRY_OUT to ~1.0
 - Tweak ITMY/ETMY if the beam spot on them are not good.
 - Align BS, ITMX to have good MICH fringe and TRX_OUT to ~1.0.
 - Tweak ITMX/ETMX if the beam spot on them are not good.
 - Misalign ETMY, ETMX, ITMY to have LO-ITMX fringe in BHD DCPDs. From Sitemap --> BHD --> Homodyne Phase Ctrl --> LO_PHASE --> Turn ON LO Phase Servo (This will lock LO and ITMX fringe) --> align AS beam with SR2 and AS4 differentially, with ratio of AS4/SR2=3.6 for YAW, with ratio of AS4/SR2= 4.5 for PITCH to have BHDC_A_OUT ~ 50.
 - Restore ITMY alignment, and lock MICH.

  17280   Thu Nov 17 15:53:47 2022 JCConfigurationCamerasITMX Camera -- attempt at fix

The issue was the power supply.

Quote:

I found that an old BNC cable for ITMXF video existed so I first tried swapping both ends of the cable, one on the ITMX viewport and the other one in the video MUX input in the rear. This didn't fix the issue.

I searched around in the CCD cabinet by XARM and found an identical analog camera so I swapped it and got the same image ...

I then searched for a AC/DC supply cable, but couldn't find one.

 

  17283   Fri Nov 18 09:00:27 2022 JCUpdateVACPressure Gauge Information

I bought the spare Full-Range Pirani Gauge a while ago and realized that I never logged this. The Pirani Gauges we are using is described below.

QTY Product Description Serial No.
1 FRG702CF35 -702 FULL RANGE PIRANI/IMG GA.,2.75CF

LI2218F003

I purchased this gauge from Agilent through TechMart. The spare is located inside the Vac Equipment cabinet (The only brown cabinet.) along the X-Arm.

  17290   Mon Nov 21 09:13:31 2022 JCUpdatePSLPMC input beam aligned again, IMC

[JC, Paco]

I attempted to align PMC Beam from a transmission of 0.72. I failed to do so on my own, but Paco arrived and helped me out. Transmission has gone from from 0.72 to ~0.73.

  17292   Mon Nov 21 14:55:58 2022 JCConfigurationGeneralLED Instead of Incandescent Lights

Electrical came by today to see the lights. The issue may be the switches, but they will come by tomorrow to solve the issue. A couple light bulbs were noticed to be out, but they no longer have incandescent lights. . . only LED. I figured this would be preffered because of the reduction on noise. I would prefer to go ahead and ask to change all the incandescent bulbs to LED bulbs. Are there any objections to this?

ELOG V3.1.3-