ID |
Date |
Author |
Type |
Category |
Subject |
10524
|
Mon Sep 22 15:20:32 2014 |
Steve | Update | VAC | RGA scan at day 229 |
|
10530
|
Wed Sep 24 08:40:29 2014 |
Steve | Update | VAC | Vent has started |
Jam nuts checked. Oplev servos turned off. Particle count checked. Vertex crane functionality checked. |
10532
|
Wed Sep 24 13:28:48 2014 |
Steve | Update | VAC | vent is completed |
The vent is completed. ITMX was kicked up accidentally. Valve configuration: chamber open, RGA is pumped through VM2 Maglev |
10533
|
Wed Sep 24 16:02:58 2014 |
Jenne | Update | VAC | vent is completed |
[Steve, EricQ, Jenne]
ITMY and BS heavy doors are off, light doors are on. Q is aligning the IFO. |
10536
|
Thu Sep 25 08:21:15 2014 |
Steve | Update | VAC | vent day 1 |
Quote: |
[Steve, EricQ, Jenne]
ITMY and BS heavy doors are off, light doors are on. Q is aligning the IFO.
|
|
10547
|
Mon Sep 29 09:11:25 2014 |
Steve | Update | VAC | vent day 5 RGA scan |
|
10548
|
Mon Sep 29 10:29:25 2014 |
Steve | Update | VAC | RGA is not running |
The RGA time stamp was correct last at 20140527
Rga stopped scanning at 20140530 |
10553
|
Tue Sep 30 16:18:57 2014 |
Steve | Update | VAC | pump down #78 has started |
Q checked the earth quake stops of SRM and we put the ITMY & BS doors on. |
10555
|
Tue Sep 30 18:02:53 2014 |
Steve | Update | VAC | pump down #78 stops at 320 Torr |
We stopped pumping just short of 3 hours at 320 Torr. Pumping speed was 2.7 Torr / min with partially closed RV1 and butterfly valve/
RP1&3 roughing pump hose is disconnected. Butterfly valve removed. The vac envelope is closed. |
10556
|
Wed Oct 1 10:21:19 2014 |
Steve | Update | VAC | pump down #78 stops at 3.5 Torr |
Quote: |
We stopped pumping just short of 3 hours at 320 Torr. Pumping speed was 2.7 Torr / min with partially closed RV1 and butterfly valve/
RP1&3 roughing pump hose is disconnected. Butterfly valve removed. The vac envelope is closed.
|
This is our second stop. I will be back this afternoon. IFO P1 3.5 Torr
|
10557
|
Wed Oct 1 16:26:53 2014 |
Steve | Update | VAC | pump down #78 completed |
Pump down reached "vacuum normal" state. IFO _P1 pressure 1e-4 torr in 8 hrs actual pumping time
PSL shutter is opened. |
10559
|
Thu Oct 2 09:23:23 2014 |
Steve | Update | VAC | vent 77 |
Quote: |
Pump down reached "vacuum normal" state. IFO _P1 pressure 1e-4 torr
PSL shutter is opened.
|
IFO_P1 pressure 1.6e-5 torr after 6 days at atm
PS: PSL sliding door 11 was left open overnight. The PSL particle count will reach room counts in 20 seconds at low speed of HEPA |
10563
|
Fri Oct 3 10:10:37 2014 |
Steve | Update | VAC | cold cathode gauge reading switched |
We have two cold cathode gauges at the pump spool and one signal cable to controller. CC1 in horizontal position and CC1 in vertical position.
CC1 h started not reading so I moved cable over to CC1 v |
10568
|
Mon Oct 6 10:23:43 2014 |
Steve | Update | VAC | Unexpected power shutdown |
Quote: |
We had an unexpected power shutdown for 5 sec at ~ 9:15 AM.
Chiara had to be powered up and am in the process of getting everything else back up again.
Steve checked the vacuum and everything looks fine with the vacuum system.
|
PSL Innolight laser and the 3 units of IFO air conditions turned on.
The vacuum system reaction to losing power: V1 closed and Maglev shut down. Maglev is running on 220VAC so it is not connected to VAC-UPS. V1 interlock was triggered by Maglev "failure" message.
Maglev was reset and started. After Chiara was turned on manually I could bring up the vac control screen through Nodus and opened V1
"Vacuum Normal" valve configuration was recovered instantly.
Chiara needs UPS
It is arriving Thursday |
10579
|
Tue Oct 7 16:55:16 2014 |
Steve | Update | VAC | Unexpected sweaty valves |
Pump spool valves V5, V4, V3 sweating a lot. VM3 and VC2 not so much.
They are VAT valves F28-62887-03, 11, 14 and so on ~15-16 years old.
I'm speculating that some plastic is aging-braking down at the atmospheric-pneumatic side of valves.
The vacuum side is not effected, according to vacuum pressure readings.
May be some condensation from the small turbos? No
I'm looking for an identical valve to examine, but I can not find one.
We are using industrial grade 99.96% Nitrogen to actuate these valves.
Valves are not effected are dry: VA6, V6, V7 and all annuloses.
|
10584
|
Wed Oct 8 08:46:57 2014 |
Steve | Update | VAC | VAT valves actuator lubricant |
Quote: |
Pump spool valves V5, V4, V3 sweating a lot. VM3 and VC2 not so much.
They are VAT valves F28-62887-03, 11, 14 and so on ~15-16 years old.
I'm speculating that some plastic is aging-braking down at the atmospheric-pneumatic side of valves.
The vacuum side is not effected, according to vacuum pressure readings.
May be some condensation from the small turbos? No
I'm looking for an identical valve to examine, but I can not find one.
We are using industrial grade 99.96% Nitrogen to actuate these valves.
Valves are not effected are dry: VA6, V6, V7 and all annuloses.
|
VAT's answer:
Yes, our engineers are aware of this issue. They say:
The pneumatic actuator needs lubricant as the O-ring (Viton) slides in the cylinder. Without grease the O-ring would be abraded and leaking after only a relatively few cycles. The lubricant used in our pneumatic actuators is an emulsion of oil and Teflon flakes. Vibration, many cycles and sometimes high temperature lead to the separation of the oil and Teflon. That is apparently the issue you are seeing.
VAT is and has been testing and qualifying new lubricants, and this is one of the factors we are always looking to improve. The formula we used 15 years ago in these valves seems to have performed reasonable well. Our formula today should perform even better.
We realize this explanation does not help you with these existing valves, but 15 years of service is not too bad is it?
Steve -NOTE:bonnet seal is metal so there is no way this oil can get into our vacuum ( only if the bellow leaks ) |
10587
|
Thu Oct 9 11:56:35 2014 |
Steve | Update | VAC | Power outage II & recovery |
Quote: |
Post 30-40min unexpected power outage this morning, Steve checked the status of the vacuum and I powered up Chiara.
I brought back the FE machines and keyed all the crates to bring back the slow machines but for the vac computers.
c1vac1 is not responding as of now. All other computers have come back and are alive.
|
IFO vacuum, air condition and PMC HV are still down. PSL out put beam is blocked on the table. |
10590
|
Thu Oct 9 17:33:28 2014 |
Steve | Update | VAC | Power outage II & recovery |
Quote: |
Quote: |
Post 30-40min unexpected power outage this morning, Steve checked the status of the vacuum and I powered up Chiara.
I brought back the FE machines and keyed all the crates to bring back the slow machines but for the vac computers.
c1vac1 is not responding as of now. All other computers have come back and are alive.
|
IFO vacuum, air condition and PMC HV are still down. PSL out put beam is blocked on the table.
|
We are pumping again. This is a temporary configuration. The annuloses are at atmosphere. The reset reboot of c1Vac1 and 2 opened everything except the valves that were disconnected.
TP2 lost it's vent solenoid power supply and dry pump during the power outage.
They were replaced but the new small turbo controller is not set up as the old TP2 was so it does not allow V4 to open.
Tomorrow I will swap back the old controller, pump down the annuloses and close off the ion pumps.
I removed the beam block from the PSL table and opened the shutter. CC4 has the real pressure 2e-5 Torr
CC1 is not real. |
10597
|
Fri Oct 10 14:41:04 2014 |
Steve | Update | VAC | Power outage II & recovery |
Quote: |
Quote: |
Quote: |
Post 30-40min unexpected power outage this morning, Steve checked the status of the vacuum and I powered up Chiara.
I brought back the FE machines and keyed all the crates to bring back the slow machines but for the vac computers.
c1vac1 is not responding as of now. All other computers have come back and are alive.
|
IFO vacuum, air condition and PMC HV are still down. PSL out put beam is blocked on the table.
|
We are pumping again. This is a temporary configuration. The annuloses are at atmosphere. The reset reboot of c1Vac1 and 2 opened everything except the valves that were disconnected.
TP2 lost it's vent solenoid power supply and dry pump during the power outage.
They were replaced but the new small turbo controller is not set up as the old TP2 was so it does not allow V4 to open.
Tomorrow I will swap back the old controller, pump down the annuloses and close off the ion pumps.
I removed the beam block from the PSL table and opened the shutter. CC4 has the real pressure 2e-5 Torr
CC1 is not real.
|
Tp2 is controlled by old controller. Annuloses pumped down. Valve configuration: "vacuum normal "
Ion pumps closed at <1e-4 mT |
10599
|
Mon Oct 13 14:44:52 2014 |
Steve | Update | VAC | RGA scan pd78 -day 13 |
Our first RGA scan since May 27, 2014 elog10585
The Rga is still warming up. It was turned on 3 days ago as we recovered from the second power outage.
|
10741
|
Mon Dec 1 17:13:57 2014 |
Steve | Update | VAC | RGA scan pd78 -day 63 |
Quote: |
Our first RGA scan since May 27, 2014 elog10585
The Rga is still warming up. It was turned on 3 days ago as we recovered from the second power outage.
|
|
10837
|
Tue Dec 23 14:33:24 2014 |
Steve | Update | VAC | Chiara gets UPS |
Quote: |
Quote: |
We had an unexpected power shutdown for 5 sec at ~ 9:15 AM.
Chiara had to be powered up and am in the process of getting everything else back up again.
Steve checked the vacuum and everything looks fine with the vacuum system.
|
PSL Innolight laser and the 3 units of IFO air conditions turned on.
The vacuum system reaction to losing power: V1 closed and Maglev shut down. Maglev is running on 220VAC so it is not connected to VAC-UPS. V1 interlock was triggered by Maglev "failure" message.
Maglev was reset and started. After Chiara was turned on manually I could bring up the vac control screen through Nodus and opened V1
"Vacuum Normal" valve configuration was recovered instantly.
Chiara needs UPS
It is arriving Thursday
|
EricQ and Steve,
Steve preset the vacuum for safe-reboot mode with C1vac1 and C1vac2 running normal: closed valves as shown, stopped Maglev & disconnected valves V1 plus valves with moving labels.
(The position indicator of the valves changes to " moving " when its cable disconnected )
Eric shut down Chiara, installed APC's UPS Pro 1000 and restarted it.
All went well. Nothing unexpected happened. So we can conclude that the vacuum system with running C1vac1 and C1vac2 is not effected by Chiara's losing AC power. |
10838
|
Tue Dec 23 15:37:32 2014 |
Steve | Update | VAC | TP3 drypump replaced |
Quote: |
Quote: |
TP2's fore line - dry pump replaced at performance level 600 mTorr after 10,377 hrs of continuous operation.
Where are the foreline pressure gauges? These values are not on the vac.medm screen.
The new tip seal dry pump lowered the small turbo foreline pressure 10x
TP2fl after 2 day of pumping 65mTorr
|
TP2 dry pump replaced at fore pump pressure 1 Torr, TP2 50K_rpm 0.34A
Top seal life 6,362 hrs
New seal performance at 1 hr 36 mTorr,
Maglev at 560 Hz, cc1 6e-6 Torr
|
TP3 dry pump replaced at 540 mT as TP3 50K_rpm 0.3A with annulos load. It's top seal life time was 11,252 hrs
|
10843
|
Fri Dec 26 17:45:21 2014 |
Steve | Update | VAC | vac pressure rose to 1.3 mTorr |
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things. |
10845
|
Sun Dec 28 07:29:16 2014 |
Steve | Update | VAC | vacuum is normal |
Quote: |
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things.
|
ITMX damping restored. |
10968
|
Tue Feb 3 15:20:13 2015 |
Steve, Koji | Update | VAC | Pneumatic pressure is being read again |
[Koji, Steve]
Summary
The N2 pressure reading (C1:VAC-N2PRES) is now up-to-date after rebooting c1vac1.
The vaccum system is "Vacuum normal". We now have a space pressure transducer.
Introduction
Our vacuum valves are manipulated with 60~75 PSI of nitrogen. All the valves are configured to be closed in the case of low N2 supply pressure.
In order to avoid this safety shutdown accidentally triggered, we have two N2 cylinders to sustain the vacuum valves. When one cylinder goes to low
the mechanical valve switches over to the other cylinder.
We have the monitor channel for this (combined) cylinder pressure. One shoulbe be able to see periodical pressure variation when the auto cylinder
switch is operating. However, the nirogen pressure reading got stuck at 66 PSI on Dec.16, 2014 (See attached 60-day plot of N2 supply pressure).
What we did
This morning we tracked down the cause of the trouble. We first closed the valves on EPICS and started to vary the N2 pressure.
Our first guess was the pressure transducer (Omega #236PC100GW) that was already 15 yrs old. We even has a sensor spare for replacement.
But it turned out that the direct voltage reading (to be 1mV/PSI) is changing correctly. The second guess was Omega Controller-Monitor
#DPiS32-C24 that is reading the voltage from the tranceducer. The display on this small black unit was changing corresponding to the
pressure change.
So our thought was
1) RS232C of the monitor unit is not working correctly
or
2) c1vac1 is not communicating with the monitor unit.
We wondered what could cause c1vac1 not communicating with the monitor unit, but we were afraid that some function got stuck
during either the nodus upgrade or chiara rebooting (or something else). So we decided to reboot c1vac1
In order to avoid any glitch in the main vacuum pressure, Steve disconnected some of the controller connectors for the closed valves.
We did this treatment before and it was successful.
Then c1vac1 was rebooted just by telnet and type reboot in the terminal.
Once the target is back in action, we noticed that the monitor value started to move.
Steve reverted the cables to the valves and operated the valves to recover "Vacuum Normal" state. Everything is now nicely settled. |
11156
|
Sun Mar 22 18:42:40 2015 |
Steve | Update | VAC | vac pressure rose to 1.2mTorr |
Quote: |
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things.
|
We run out of N2 for the vacuum system 6 hrs ago. The pressure rose to 1.2 mTorr with V1 closed. The interlock worked! See Nirogen presure reading fixed at http://nodus.ligo.caltech.edu:8080/40m/10968
The vacuum interlock: Nitrogen pressure transducer is reading the pneumatic pressure continously at the pump spool and c1vac1 processing it. When it drops below 60 PSI it closes V1 gate valve and V4 & V5. Gate valve V1 needs minimum 60 PSI to close. It is critical that V1 is closed before you run out of Nitrogen so the IFO pressure is contained.
IFO vacuum is back to Vac Normal. The MC is locked.
cc4 = 2E-6 Torr with VM1 open.
Daily N2 consumption measured to be 530PSI as 3 days on 3-27-2015 but note: it does vary !
I have seen it as high as 900 psi The long term average ~750 psi |
11159
|
Mon Mar 23 10:36:55 2015 |
ericq | Update | VAC | Pressure watch script |
Based on Jenne's chiara disk usage monitoring script, I made a script that checks the N2 pressure, which will send an email to myself, Jenne, Rana, Koji, and Steve, should the pressure fall below 60psi. I also updated the chiara disk checking script to work on the new Nodus setup. I tested the two, only emailing myself, and they appear to work as expected.
The scripts are committed to the svn. Nodus' crontab now includes these two scripts, as well as the crontab backup script. (It occurs to me that the crontab backup script could be a little smarter, only backing it up if a change is made, but the archive is only a few MB, so it's probably not so important...) |
11178
|
Fri Mar 27 10:38:40 2015 |
steve | Update | VAC | vac rack UPS battery replaced |
Batteries replaced after 3.5 years with Amstron AP-1250F2, 8x 12V 6Ah
Quote: |
APC Smart -UPS 2200 model: SUA2200RM2U batteries were replaced by compatible RBC43, 8x 12V5A
|
|
11184
|
Tue Mar 31 09:03:32 2015 |
Steve | Update | VAC | RGA scan pd78 day 182 |
|
11243
|
Fri Apr 24 17:30:32 2015 |
Jenne | Update | VAC | Pressure watch script broken |
Quote: |
I made a script that checks the N2 pressure, which will send an email to myself, Jenne, Rana, Koji, and Steve, should the pressure fall below 60psi.
|
The script checking the N2 pressure is not working. I signed into the foteee account to look at some of the picasa photos, and there are thousands of emails (one every 10 minutes for the past month!) with error messages. Q, can you please make it stop (having errors)?
The error looks like it's mad about a "caget" command. I don't have time to investigate further though. |
11249
|
Sat Apr 25 18:50:47 2015 |
ericq | Update | VAC | Pressure watch script broken |
Ugh, this turns out to be because cron doesn't source the controls bashrc that defines where to find caget and all that jazz that many commands depend on. This is probably also why the AutoMX cron job isn't working either.
Also, cron automatically emails everything from stderr to the email address that is configured for the user, which is why the n2 script blew up the foteee account and why the AutoMX script was blowing up my email yesterday. This can be avoided by doing something like this in the crontab:
0 8 * * * /bin/somecommand >> somefile.log 2>&1
(The >> part means that the standard output is appended to some log file, while the 2>&1 means send the standard error stream to the same place as stdout)
I made this change for the n2 script, so the foteee email account should be safe from this script. I haven't figured out the right way to set up cron to have all the right $PATH and other environment stuff, such as epics may need, so the script is still not working. |
11260
|
Mon Apr 27 14:37:55 2015 |
Steve | Update | VAC | N2 pneumatic pressure watch |
Quote: |
Based on Jenne's chiara disk usage monitoring script, I made a script that checks the N2 pressure, which will send an email to myself, Jenne, Rana, Koji, and Steve, should the pressure fall below 60psi. I also updated the chiara disk checking script to work on the new Nodus setup. I tested the two, only emailing myself, and they appear to work as expected.
The scripts are committed to the svn. Nodus' crontab now includes these two scripts, as well as the crontab backup script. (It occurs to me that the crontab backup script could be a little smarter, only backing it up if a change is made, but the archive is only a few MB, so it's probably not so important...)
|
This watch script gives little time to replace N2 cylinder. When the regulated supply drops below 60 psi the cylinder pressure is 60 psi too.
It is more of a statement that V1 is closed and act accordingly. It's only practical if you are in the lab.
Rana pointed it out correctly that we need this message 24 hrs before it happens. This requires monitoring of total supplies , not the regulated one.
So we need pressure transducers on each nitrogen cylinder, before the regulator. The sum of the two N2 cylinder when they are full 4000 - 4500 psi
The first email should be send out at 1000 psi as sum of the two cylinders. This means that you have 1 day to replace nitrogen cylinder.
Most of the time the daily consumption is 750 +-50 psi
However sometimes this variation goes up ~750 +-150 psi |
11261
|
Mon Apr 27 21:42:07 2015 |
rana | Update | VAC | Summary pages |
We want to have a VAC page in the summaries, so Steve - please put a list of important channel names for the vacuum system into the elog so that we can start monitoring for trouble.
Also, anyone that has any ideas can feel free to just add a comment to the summary pages DisQus comment section with the 40m shared account or make your own account. |
11262
|
Tue Apr 28 09:49:26 2015 |
Steve | Update | VAC | Vac Summery Channels |
Channel
|
Function |
|
Interlock action |
|
|
|
|
C1:Vac-P1_pressure
|
IFO vac envelope pressure |
|
at 3 mT close V1 and PSL shutter |
C1:Vac-P2_pressure |
Maglev foreline pressure |
|
at 6 Torr close V1 |
C1:Vac-P3_pressure |
annuloses |
|
|
C1:Vac-CC1_pressure |
IFO pressure |
|
at 1e-5 Torr close VM1 |
C1:Vac-CC4_pressure |
RGA pressure |
|
|
C1:Vac-N2pres |
valve's drive pneumatic 60-80PSI |
|
at 55 PSI close V1, at 45 PSI close all |
It does not exist yet |
2 N2 cylinder sum pressure |
|
|
11264
|
Thu Apr 30 16:30:25 2015 |
Steve | Update | VAC | N2 pneumatic pressure watch set up |
We have 2 transduser PX303-3KG5V http://www.omega.com/pressure/pdf/PX303.pdf They will be installed on the out put of the N2 cylinders to read the supply pressure.
I will order one DC power supply http://www.omega.com/pptst/PSU93_FPW15.html PSU-93
One full cylinder pressure is ~ 2400 PSI max so two of them will give us ~9Vdc
The email reminder should be send at 1000 PSI = 1.8 V
|
11274
|
Tue May 5 16:02:57 2015 |
Steve | Update | VAC | Vac Summery Channels with discription |
As it was requested by the Bos.
It would be nice to read from the epic screen C1:Vac-state_mon.......Current State: Vacuum Normal, valve configuration
Quote: |
Channel
|
Function |
Description |
Interlock |
|
|
|
|
C1:Vac-P1_pressure
|
Main volume of 40m interferro meter |
P=Pirani gauge, Pressure range: ATM-760 to 1e-4 Torr |
at 3 mT close V1 and PSL shutter |
C1:Vac-P2_pressure |
Maglev foreline pressure |
Maglev is the main pump of our vacuum system below 500 mTorr
It's long term pressure has to be <500 mTorr
|
at 6 Torr close V1 |
C1:Vac-P3_pressure |
annuloses |
Each chamber has it's own annulos. These small volumes are indipendent from main volume. Their pressure ranges are <5 mTorr at vac. normal valve configuration.
|
|
C1:Vac-CC1_pressure |
IFO main volume |
CC1=cold cathode gauge (low emmision), Pressure range: 1e-4 to 1e-10 Torr,
In vac- normal configuration CC1= 2e-6 Torr
|
at 1e-5 Torr close VM1 |
C1:Vac-CC4_pressure |
RGA pressure |
In vac-normal configuration CC1=CC4 |
|
C1:Vac-N2pres |
valve's drive pneumatic |
The N2 supply is regulated to 60-80 PSI out put at the auto cylinder changer.
|
at 55 PSI close V1, at 45 PSI close all |
It does not exist yet |
2 N2 cylinder sum pressure |
Each cylinder pressure will be measured before the regulator and summed for warning message to be send
at 1000 PSI
|
|
|
11276
|
Fri May 8 14:30:09 2015 |
Steve | Update | VAC | CC1 cold cathode gauges are baked now |
CC1s are not reading any longer. It is an attempt to clean them over the weekend at 85C
These brand new gauges "10421002" sn 11823-vertical, sn 11837 horizontal replaced 11 years old 421 http://nsei.missouri.edu/manuals/hps-mks/421%20Cold%20Cathode%20Ionization%20Guage.pdf on 09-06-2012 http://nodus.ligo.caltech.edu:8080/40m/7441
Quote: |
We have two cold cathode gauges at the pump spool and one signal cable to controller. CC1 in horizontal position and CC1 in vertical position.
CC1 h started not reading so I moved cable over to CC1 v
|
|
11287
|
Tue May 12 14:57:52 2015 |
Steve | Update | VAC | CC1 cold cathode gauges are baked now |
Baking both CC1 at 85 C for 60 hrs did not help.
The temperature is increased to 125 C and it is being repeated.
|
11292
|
Fri May 15 16:18:28 2015 |
Steve | Update | VAC | Vac Operation Guide |
Vacuum Operation Guide is up loaded into the 40m-wiki. This is an old master copy. Not exact in terms of real action, but it is still a good guide of logic.
Rana has promissed to watch the N2 supply and change cylinder when it is empty. I will be Hanford next week. |
11352
|
Wed Jun 10 15:54:14 2015 |
Steve | Update | VAC | Vacuum comp. rebooted |
Koji and Steve succeded rebooting C1vac1, C1vac2 and pressure reading is working now
More tomorrow .........
|
11353
|
Thu Jun 11 19:40:59 2015 |
Koji | Update | VAC | Vacuum comp. rebooted |
The serial connections to the vacuum gauges were recovered by rebooting c1vac1 and c1vac2.
Steve claimed that the vacuum screen had showed "NO COMM" at the vacuum pressure values.
The epics connection to c1vac was fine. We could logged in to c1vac1 with telnet too although c1vac2 had no response.
After some inspection, we decided to reboot the slow machines. Steve manually XXXed YYY valves (to be described)
to prepare for any possible unwanted switching. Initially Koji thought only c1vac2 can be rebooted. But it was wrong.
If the reset button is pushed, all of the modules on the same crate is reset. So everything was reset. After ~3min we still
don't have the connection to c1vac1 restored. We decided to another reboot. This time I pushed c1vac1 reset button.
After waiting about two minutes, the ADCs started to show green lights and the switch box started scanning.
We recovered the telnet connection to c1vac1 and epics functions. c1vac2 is still note responding to telnet, and
the values associated with c1vac2 are still blank.
Steve restored the valves and everything was back to normal. |
11354
|
Fri Jun 12 08:40:17 2015 |
Steve | Update | VAC | Vacuum comp. rebooted |
Koji and Steve,
One computer expert and one vacuum expert required.
Quote: |
The serial connections to the vacuum gauges were recovered by rebooting c1vac1 and c1vac2.
Steve claimed that the vacuum screen had showed "NO COMM" at the vacuum pressure values.
The epics connection to c1vac was fine. We could logged in to c1vac1 with telnet too although c1vac2 had no response.
After some inspection, we decided to reboot the slow machines. Steve manually XXXed YYY valves (to be described)
to prepare for any possible unwanted switching. Initially Koji thought only c1vac2 can be rebooted. But it was wrong.
If the reset button is pushed, all of the modules on the same crate is reset. So everything was reset. After ~3min we still
don't have the connection to c1vac1 restored. We decided to another reboot. This time I pushed c1vac1 reset button.
After waiting about two minutes, the ADCs started to show green lights and the switch box started scanning.
We recovered the telnet connection to c1vac1 and epics functions. c1vac2 is still note responding to telnet, and
the values associated with c1vac2 are still blank.
Steve restored the valves and everything was back to normal.
|
Atm 1, problem condition: gauges are not reading for a week, error message "NO COMM" and all computer LEDs are green
Atm 2, prepare to safe reboot:
a, close V1, disconnect it's power cable and turn off Maglev, wait till rotation stops
b, close PSL shutter ( take adrenaline if needed )
c, close V4, V5, VA6 valves and disconnect their cables. "Moving" error message indicating this condition.
V1 is not showing "Moving" because its power cable disconnected only! It will show it if its position indicator cable is disconnected too. There is no need for that.
These valves closed and disabled will not allow accidental venting of main volume.
d, push reset, reseting c1vac2 will reset c1vac1 also, wait ~ 6 minutes
"Vacuum Normal" valve configuration was restored after succesful reboot as follows:
a, reconnect cable and open V4 and V5 at P2 & P3 <1e-1 Torr
b, observe that P2 < 1e-3 Torr and retsart Maglev
c, wait till Maglev reaches full speed of 560 Hz and reconnect-open V1
d, reconnect-open VA6 at P3 <1e-3 Torr
NOTE: VM1 valve was locked in open position and it was not responding before and after reboot
Error message on Atm2 is indicating this locked condition: "opening VM1 will vent IFO"
This is a fauls message. The valve is frozen in open position. We need a softwear expert help.
|
11387
|
Wed Jul 1 10:01:25 2015 |
Steve | Update | VAC | RGA scan pd78 day 275 |
|
11483
|
Thu Aug 6 09:16:57 2015 |
Steve | Update | VAC | Chiara gets new Ethernet card |
- Yesterday we did put the vacuum system into safe to reboot mode for the ethernet card swap.
- Atm1, IFO pressure
- Atm2, as prepared valve configuration where " moving " means closed and disconnected
|
11593
|
Mon Sep 14 10:41:03 2015 |
Steve | Update | VAC | TP2 dry pump replaced |
TP2 dry fore pump sn PLE10082 was replaced at pressure 717 mTorr, TP2 50K rpm 0.33A @ 112,677 hrs
Top seal life was 8,160 hrs
Model SH110, Sn LP1007L556 was installed. It's fore line pressure after 30 minutes of running 38 mTorr, TP2 turbo at 50K rpm 0.18A
|
11625
|
Mon Sep 21 11:12:14 2015 |
Steve | Update | VAC | cold cathode is flaky |
CC4 cold cathode gauge jump triggered interlock to close VM1 valve to protect the RGA.
The IFO pressure is 1e-5 Torr
Vac normal was recovered by opening VM1
|
11633
|
Tue Sep 22 08:58:38 2015 |
Steve | Update | VAC | cold cathode is flaky |
The cold cathode gauge is back to normal. cc4 is the last gauge is "functioning"
MKS is not responding. The spare controller and gauges are back for repair.
|
11644
|
Fri Sep 25 17:00:38 2015 |
Steve | Update | VAC | cold cathode is flaky |
The IFO pressure is estimated ~1E-6 Torr with modified Vac normal valve configuration.
CC4 is in a jumping mode between 2e-5 and 1e-6 Torr
Pressure based interlock kicks in to close VM1 at 2e-5 Torr to protect the RGA.
I did open VM1 repeatedly in the last few mornings but as cc4 jumps VM1 closes.
As VM1 closed to RGA scans are not seeing the IFO. I will look at some scans on Monday.
Mean while I opened VM2 to lower the pressure for the RGA. This change will be read by "Current Status: Undefined State"
So do not panic, the IFO pressure is normal.
I need someone's help to raise the interlock threshold to 5e-5 Torr
I'm buying a new cold cathode gauge on Monday.
note: cc1 is out of order!
just read P1 the pressure is < 7e-4 Torr This gauge is very reliable and it is at the low end of it's range.
|
11666
|
Mon Oct 5 10:11:35 2015 |
Steve | Update | VAC | RGA scan pd78 day 386 |
RGA background scan.
The IFO is closed off from RGA with VM1
CC4 ( and CC1) is still flaky and it's interlock closes VM1
|