  12530   Tue Oct 4 11:02:31 2016 SteveUpdateVACRGA is out of orderThe last good rga scan at vent 78 
day 38

 lastBGscan.png 
  12537   Fri Oct 7 10:29:57 2016 SteveUpdateVACRGA removedThe RGA is removed for repaire. It's
volume at atmophere and sealed.. P4 reading
of 38 Torr is not correct.
 RGAremoved.png 
  12556   Thu Oct 13 00:23:54 2016 ericqUpdateVACPre-vent prepI have completed the following non-Steve
portions of the pre-vent checklist [wiki-40m.ligo.caltech.edu]


Center all oplevs, transom
  
  12559   Thu Oct 13 16:48:34 2016 SteveUpdateVACvent 79 is completedIFO is at atmosphere. The MC can be locked
in air now.

The doors will be coming off tomorrow
 vent_79.png 
  12574   Thu Oct 20 09:17:08 2016 SteveUpdateVACPumpdown 80 has startedPumping again after 7 days at atmosphere.

BS, ITMY and OMC chambers were
open only.
 readyPd80.pngbeforePd80.pngpd80at1h.png7d_atm.png 
  12575   Thu Oct 20 15:59:52 2016 SteveUpdateVACPumpdown 80 has completedIFO pressure 1.7E-4 Torr on new not logged
cold cathode gauge. P1 <7E-4 Torr

 7hPd80.png8hPd80.png9hPd80.png 
  12577   Fri Oct 21 09:28:21 2016 SteveUpdateVACVac Normal reachedDry pump of TP3 replaced after 9.5 months
of operation.[ 45 mTorr d3 ]

The annulosses are pumped.
 VacNormal.png 
  12589   Mon Oct 31 16:20:50 2016 SteveUpdateVACRGA is reinstalledQuad rods and  ionizer kit: consisting
of repeller cage, anode grid, focus plate
and  filament were replaced....... under
  
  12633   Tue Nov 22 11:31:52 2016 SteveUpdateVACRGA is running againThe Rga was turned on yesterday.




 RGAisBack33d.pngpd80-560Hz-d33.pngafterRepair1d.png 
  12636   Wed Nov 23 11:32:13 2016 SteveUpdateVACTP3 drypump replaced againTP3 drypump replaced at 655 mTorr, no load,
tp3 0.3A 

This seal lasted only for 33 days
  
  12642   Mon Nov 28 09:45:44 2016 SteveUpdateVACRGA scan at pumpdown day 40The vacuum envelope ~1 C warmer today than
7 days ago. Bad peaks are coming down as
normal.
 RGAscan_d40.pngd2_&_d7scan.pngRGAd2.png 
  12681   Thu Dec 22 09:37:20 2016 SteveUpdateVACRGA scan at day 63Valve configuration: vacuum normal

RGA head temp: 43.5 C

Vac envelope temp: 23 C
 pd80-d63.pngpd80-580Hz-d63.png 
  12698   Tue Jan 10 14:24:09 2017 SteveUpdateVACRGA scan at day 82Valve configuration: vacuum normal

Vacuum envelope: 23C

Rga head: 44C
 pd80VNd82.png 
  12774   Tue Jan 31 14:14:29 2017 ranaUpdateVACPressure watch scriptI think this cron job is running on NODUS
(our gateway) instead of our scripts machine:


p.p1 {margin:
  
  12853   Mon Feb 27 15:33:10 2017 SteveUpdateVACRGA scan at day 130Valve configuration: vacuum normal

Vacuum envelope: 23.5 C

RGA head: 46.6 C
 rgascan@130d.png 
  13047   Wed Jun 7 11:32:56 2017 SteveUpdateVACsmooth vac rebootGautam and Steve,

 

The medm monitor & vac control
 vac_reboot.png 
  13075   Tue Jun 20 16:28:23 2017 SteveUpdateVACRGA scan   RGAscan243d.pngRGAscan.png 
  13132   Sun Jul 23 15:00:28 2017 JamieOmnistructureVACstrange sound around X arm vacuum pumpsWhile walking down to the X end to reset
c1iscex I heard what I would call a "rythmic
squnching" sound coming from under the
  
  13134   Mon Jul 24 09:55:29 2017 SteveOmnistructureVACcontroller failerIfo pressure was 5.5 mTorr this Monday
morning. The PSL shutter was still open.
TP2 controller failed. Interlock closed V1,
 TP2controllerFails.png 
  13142   Tue Jul 25 08:48:57 2017 SteveUpdateVACRGA scan at d278The RGA did not shut down at the turbo
pump controller failing.

 RGA278d.png 
  13143   Tue Jul 25 14:04:06 2017 SteveUpdateVACturbo controller installed and we are running at vac normalGautam and Steve,

Spare Varian turbo-V 70 controller,
Model 969-9505, sn 21612 was swapped in.
 TP2@75Krpm.png 
  13171   Tue Aug 8 17:04:26 2017 SteveUpdateVACunintended pump downIFO pressure 2 Torr,    PSL 
shutter closed.  I'm pumping down
with 2 roughing pumps with ion pump gate
 pumping_down.png 
  13172   Tue Aug 8 17:44:11 2017 SteveUpdateVACunintended pump downPumpdown stopped for over night  at
~  1 Torr

The roughing line disconnected. 
 stopped_pumping.png 
  13175   Wed Aug 9 11:59:51 2017 SteveUpdateVACunintended pump down at vacuum normalpd80b has reached Vac Normal. IFO pressure
0.5 mTorr

We need our vauum channels back
 pdc.pngpd80b@vacnormal.png 
  13179   Wed Aug 9 16:34:46 2017 ranaUpdateVACVacuum Document recoveredSteve and I found the previous draft of
the 40m Vacuum Document. Someone in 2015
had browsed into the Docs history and then
  
  13186   Thu Aug 10 15:45:34 2017 SteveUpdateVACaccidental vent to 17 TorrFinally we got the cold cathode gauge working.
IFO pressure 7e-6 Torr-it, vacuum normal
valve configuration with all 4 ion pump gate
 Mag_UPS.jpgUPSbatteryPack.jpgclosing_V1.jpg 
  13233   Mon Aug 21 14:53:32 2017 gautamUpdateVACRGA reset[gautam, steve]

In the aftermath of the accidental
vent, it looks like the RGA was shutdown.
  
  13234   Mon Aug 21 16:35:48 2017 gautamUpdateVACUPS checkup[steve, gautam]

At Rolf/Rich Abbott's request,
we performed a check of the UPS today.
  
  13245   Wed Aug 23 10:11:46 2017 SteveUpdateVACvacuum valve specificationsThe V1
gate valve specs installed at 40m wiki
page. VAT model number 10846-UE44-0007 Our
  
  13260   Mon Aug 28 10:28:07 2017 SteveUpdateVACRGA scan 20 day after 17 TorrThe RGA was turned on 7 days ago. 
It's 46 C now.   The X-arm room
tem ~20 C
 20d_after_17_torr.png 
  13268   Tue Aug 29 15:35:19 2017 SteveUpdateVACvacuum pump specifications & manualsRP1 and RP3 roughing pump manual of
Leybold D30A oily rotory pump

Fore pump of
 1759489_995814a9-55c1-4fa1-b333-f446d272c1cc.jpg 
  13321   Tue Sep 19 11:05:26 2017 SteveUpdateVACRGA scan at day 334  

 
 RGAscan334d.png 
  13403   Fri Oct 27 10:14:11 2017 SteveUpdateVACRGA scan at day 372Valve configuration: Vacuum normal

Note: Tp2 running at 75Krpm 0.25A
26C has a  load high pitch sound today.
 RGA_scan_d372.png 
  13407   Mon Oct 30 10:09:41 2017 SteveUpdateVACTP2 failed IFO pressure 1.2e-5 Torr at 9:30am




 tp2failed.pngifo_1.0E-5_Torrit.pngtp2failed2dago.png4days.png 
  13409   Mon Nov 6 09:09:43 2017 SteveUpdateVACsetting up new TP2 turboOur new Agilent Technology TwisTorr
84FS AG rack controller ( English Manual
pages 195-297 )  RS232/485, product
  
  13423   Fri Nov 10 08:52:21 2017 SteveUpdateVACTP3 drypump replacedPSL shutter closed at 6e-6 Torr-it    

   The foreline pressure
of the drypump is 850 mTorr at 8,446 hrs
  
  13454   Sun Nov 26 19:38:40 2017 SteveUpdateVACTP3 drypump replaced againThe TP3 foreline pressure was 4.8 Torr,
50K rpm 0.54A and 31C........Maglev rotation
normal 560 Hz.......    IFO pressure
 after_replacement.png 
  13459   Thu Nov 30 10:38:39 2017 SteveUpdateVACannuloses are not pumpedAnnuloses are not pumped for 30 days, since
TP2 failed.  IFO pressure 7e-6 Torr
it,  Rga 2.6e-6 Torr
 AnnulosesNotPumped.png 
  13495   Tue Jan 2 15:43:35 2018 SteveUpdateVACpumpdown after power outage 




 pumpdown_from_44_mTorr.png 
  13588   Tue Jan 30 10:22:20 2018 SteveUpdateVACRGA scan at day 175pd80b rga scan at 175 day.  IFO pressure
7.3e-6 Torr-it

Condition: vacuum normal, annuloses
 RGApd80-b-468-175d.pngRgaOn20hrs.png 
  13615   Wed Feb 7 15:50:42 2018 SteveUpdateVACIFO pressure monitoringHornet cold cathode gauge analoge output 
[ DSub9 pin 3 and 7 ] are wired to go ETMX
Acromag. It was reading 4.9V at 7.8e-6 Torr
 CC_analog.png 
  13637   Fri Feb 16 15:57:58 2018 SteveUpdateVACTP3 drypump replacedThe forline pressure of TP3 was 399 mTorr

It was replaced this morning at 
TP3 controller 134,638hrs with the "failed
  
  13647   Wed Feb 21 17:20:32 2018 johannesUpdateVACHornet gauge connected to DAQ.I wired the six available BNC connectors
on the front panel of the new XEND slow DAQ
to physical Acromag channels. There were
 VacLog.png 
  13653   Fri Feb 23 07:47:54 2018 SteveUpdateVACCC1 HornetWe have the IFO pressure logged again!
Thanks Johannes and Gautam

This
 InstruTech_Hornet_CC1.pngCC1_readback_updated.png 
  13664   Mon Mar 5 10:13:21 2018 gautamUpdateVACvacuum healthIn Steve's absence, I've tried
to keep an eye on the health of the vacuum
system. From Attachment #1, the pressure
 VacuumStatus.png 
  13685   Fri Mar 16 09:36:56 2018 SteveUpdateVACRGA scan at day 511,218dPumpdown 80 at 511 days and pd80b at 218
days

Valve configuration:  special
 RgaScan511d.png08.png 
  13702   Mon Mar 26 09:25:18 2018 SteveUpdateVACVM1 openedCC1 old MKS cold cathode gauge randomly
turns on- off. This makes software interlock
close VM1 to protect RGA  So the closed
 CC4VM1.png 
  13709   Tue Mar 27 08:58:21 2018 SteveUpdateVACVM1 opened.......scan fine 




 rga2018march27.png 
  13918   Wed Jun 6 10:02:52 2018 SteveUpdateVACRGA scan   RGA302d.pngannuloses_NOT_pumped.pngtemp_vac.png 
  13919   Wed Jun 6 10:44:52 2018 gautamUpdateVACAnnulus pressure channels added to frames[steve,
gautam]

We added the following channels
 AnsPressureLogged.png 