Entry  Wed Apr 3 01:32:04 2013, Jenne, Update, Computers, updated EPICS database (channels selected for saving) 
I modified /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini to include the C1:LSC-DegreeOfFreedom_TRIG_MON channels.  These are the same
channel that cause the LSC screen trigger indicators to light up. 
I vaguely followed Koji's directions in elog 5991, although I didn't add new grecords,
    Reply  Wed Apr 3 14:45:34 2013, Jamie, Update, Computers, updated EPICS database (channels selected for saving) 


    
        
            Quote:
        
        
            

           I modified /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini to include the C1:LSC-DegreeOfFreedom_TRIG_MON channels.  These are
Entry  Tue May 12 17:51:25 2020, rana, Summary, Computer Scripts / Programs, updated LESS syntax highlight on nodus 
apt install source-highlight

then modified bashrc to point to /usr/share instead of /usr/bin
Entry  Wed Mar 20 14:44:47 2013, Jamie, Update, Optics, updated calculations of PRC/SRC g-factors and ARM mode matching 
Below are new alamode calculations of the PRC and SRC g-factors and arm mode matchings.  These include fixes to the ABCD matrices for the flipped
folding mirrors that properly (hopefully) take into account the focusing effect of passing through the optic substrates.
I've used nominal curvatures of -600 m for the G&H PR2/SR2 optics, and -700 m for the Laseroptik PR3/SR3 dichroics.
    Reply  Mon Mar 25 16:51:59 2013, Jenne, Update, Optics, updated calculations of PRC/SRC g-factors and ARM mode matching 
I think we like the idea of flipping SR2 better than SR3 for the same ghost beam reasons as PR2 is better than PR3.  There isn't very much free
space in the BS chamber, and if we flip PR/SR3, we have to deal with green ghosts as well as IR.
The flipping SR2 case seems okay to me - flipping either one of the SR folding mirrors gives us a slightly better g-factor than the design with
    Reply  Thu Jun 6 16:33:19 2013, steve, Update, safety, updated laser inventory ELSOconnected.jpg
 Updated laser inventory and operator list. They are posted in the 40m wiki and entry doors of the 40m IFO room.
Let me know if this list needs correction.
Lightwave  M126N-1064-700 NPRO sn 337 in the PSL enclosure got connected to local emergency shut off switch. This is a LIGO operational safety
    Reply  Tue Jul 31 18:59:59 2012, Jamie, Update, CDS, updated medm paths for MISC (IFO,CDS,VIDEO), IFO alignment scripts updated accordingly 
In an attempt to clean up the medm situation, I did a bunch of further rearrangement and cleanup.  Instead of having c1ifo, I moved a bunch of stuff
 into a MISC directory, including all of the CDS, IFO, and VIDEO  screens:
controls@rossa:~ 0$ ls -1 /opt/rtcds/caltech/c1/medm/MISC | grep -v _BAK
    Reply  Fri Mar 1 18:23:28 2013, Jamie, Update, Computer Scripts / Programs, updated version of "getdata"  
I updated the getdata script so that it can now handle downloading long stretches of data.

  /opt/rtcds/caltech/c1/scripts/general/getdata

It now writes the data to disk incrementally while it's downloading from the server,
so it doesn't fill up memory.
Entry  Wed Sep 24 19:13:06 2008, rana, Configuration, Computer Scripts / Programs, updatedb & locate: megatron & rosalba 
I ran [I]updatedb[/I] as root today on megatron and rosalba just before the meeting. 
It finished at ~14:10 on both machines so that's ~20 minutes total.
    Reply  Wed Aug 10 00:21:10 2011, jamie, Update, CDS, updates to peakFit scripts 
I updated the peakFit routines to make them a bit more user friendly:

    modified so that any subset of optics can be processed at a time, instead of just all
    broke out tweakable fit parameters
into a separate parameters.m file
    added a README that describes use

These changes were committed to the 40m svn.
Entry  Mon Jun 3 23:18:59 2019, Milind, Update, Computer Scripts / Programs, updating bashrc 
I was working with the git repo in the SnapPy_pypylon folder (/cvs/cds/rtcds/caltech/c1/scripts/GigE/SnapPy_pypylon) and needed to create a branch. To
avoid any confusion, I modified the PS1 variable and that alone in the bashrc file to reflect the git branch so that the prompt now displays the git branch
if you enter a repository. This is just an update.
    Reply  Thu Jun 16 16:11:11 2016, jamie, Update, CDS, upgrade aborted for now 
After poking at the new configuration more, it also started to show instability.  I couldn't figure out how to make test points or excitations
available in this configuration, and adding in the full set of test point channels, and trying to do simple things like plotting channels with dtt, the
frame writer (fw) would fall over, apparetnly unable to keep up with the broadcast from the dc.
Entry  Wed Dec 15 13:57:59 2021, Paco, Update, Computers, upgraded ubuntu on zita 
[Paco]

Upgraded zita's ubuntu and restarted the striptool script.
    Reply  Tue Sep 29 18:03:11 2015, rana, Update, LSC, use LISO Screen_Shot_2015-09-29_at_5.55.34_PM.png
Use LISO - see what it tells you. I would think that you should make a differential RC filter to get the right behavior. (e.g. 1K on each leg and 1 uF
between them)

Each leg of the diff input of the board has a 4k input impedance.
    Reply  Thu Jul 19 15:55:21 2012, rana, Update, Computers, used 'aptitude' to install software (TexLive) on rosalba 
I used APTITUDE to install texlive-full on rosalba so that I could work
on a paper. pdflatex was not found on pianosa, rosalba, megatron, etc. I used this command:

sudo aptitude install texlive-full

While this was installing, I read a bunch of forums which claim that its better to bypass the apt-get
    Reply  Wed Sep 30 10:00:02 2015, ericq, Update, LSC, used LISO 
LISO confirms that I did my algebra right in picking the component values, and shows no extra zeros. 

I also took some TFs with the SR785 and confirmed that both CM board inputs behave the same, and that including the LPF on the input gives
the expected 1/f shape at the slow and fast outputs.
    Reply  Fri Mar 5 11:21:18 2010, steve, Update, SUS, used OSEMs are magnetic 


    
        
            Quote:
        
        
            
            Jenne and Koji
    Reply  Fri Mar 5 18:04:56 2010, rana, Update, SUS, used OSEMs are magnetic 
The OSEM LEDs and PDs from Honeywell have always had some ferromagnetic material in them. These are the same OSEMs we had since 2000.
You must be thinking of the really old 20th century plastic OSEMs.
Entry  Sat Jun 16 01:03:21 2012, yuta, Update, Green Locking, used longer delay line for mode scan YarmScan20120615.png
[Mengyao, Yuta]

Last night, I used 1.5 m delay line COARSE and got 5FSR mode scan. The range 5FSR was limited by the range of SR560.
Entry  Sun Aug 18 20:07:41 2013, rana, Update, Computer Scripts / Programs, userapps SVN up 
JoeB and JamieR are working somewhat coherently on a set of python libraries to fulfill all of our command line CDS wants. This is being done mostly
to satisfy The Guardian and the SkunkTools project.
I did an 'svn up' in /opt/rtcds/userapps (it
    Reply  Wed May 20 16:03:52 2009, steve, Omnistructure, Environment, using OSEMs to look at seismic activity eqOSEM5m4.jpg
Rana suggested using OSEM sensing voltages as guide lines to look seismic activity.
As you see todays drilling and tumping activity was nothing compared to the EQ of mag 5 and 4
Optic level servos are turned back on.
    Reply  Thu Oct 25 09:32:17 2012, Steve, Omnistructure, IOO, using access connector acclosing.jpg


    
        
            Quote:
        
        
            

           
            
                
           
Entry  Tue Dec 18 16:50:35 2007, tobin, Update, SAFETY, uvex laser safety glasses defective bad-glasses.jpgbad-glasses-zoom.jpg
A few days ago we noticed what appeared to be a blotched, speckled fracturing of the coating of the "UVEX" laser safety glasses.  These are the glasses
with "transparent" (reflective to 1064nm) lenses and white frames that we keep in a box on top of a filing cabinet in the control room.  Today Steve measured
the transmission of these glasses and found [b]80% transmission of 1064nm[/b] in several cases.
Entry  Mon Oct 29 09:46:50 2007, steve, Routine, VAC, vac & pem trend pd64md125.jpg
Pumpdown 64 pumped by maglev for 125 days
pd64-m-d125
Entry  Sun Dec 21 13:00:06 2008, steve, Update, all down cond., vac and laser back on 
There was a power outage sometimes early Saturday.

All things are down.
Entry  Thu Apr 17 15:00:59 2008, steve, Update, VAC, vac controller computer to be replaced 
Now that Rosalba is up and running someone should step forward to volunteer for this job.

Old desk top at the Vacuum Rack should be replaced by a functional less old computer so
Entry  Thu Sep 8 13:20:07 2011, steve, Update, PEM, vac envelope at atm for 36 days vent70.jpg
We tried not to open chambers above 10,000 particles of 0.5 micron cf/min
New items going in:                       2 rasor beam traps, 5 badly oxidized
old silver plated  setscrew with spring loaded tips......to be replaced in the future, viton tips for eq screws....some are lose, gold plated allen
Entry  Wed Feb 11 11:39:17 2009, steve, Configuration, General, vac envelope ground 
The vacuum envelope was grounded from the IOO chamber to master ground north (under electric breaker panel L)
with 4GA  25ft red battery
cable.
 
    Reply  Tue Mar 23 09:37:36 2010, steve, Update, IOO, vac envelope has to be sealed as antproof for overnight pc3.JPGpc4.JPG


    
        
            Quote:
        
        
            
            This is the first touch to the MC mirrors after the earthquake on 16th.
    Reply  Tue Mar 23 09:55:20 2010, Koji, Update, IOO, vac envelope has to be sealed as antproof for overnight 
Roger.


    
        
            Quote:
        
        
            
    Reply  Wed Mar 24 02:06:24 2010, Koji, Update, IOO, vac envelope has to be sealed as antproof for overnight 
Matt and Koji:
We closed the light doors of the chambers.


    
        
            
Entry  Thu Apr 9 11:27:19 2009, steve, Update, VAC, vac gauge reading problem vacgflsec.jpgvacgflmin.jpg
Cold cathode gauge CC4  is reading normal.
CC1 is glitching, it is probably dirty.
CC2 is fluctuating too much and it is cutting out for 6-7 minutes. It must be insulated by deposits and there is no emission current.
Entry  Fri Jun 20 14:42:57 2014, steve, Update, SUS, vac illuminators turned off 
 
    Reply  Wed May 6 19:04:11 2009, rana, Summary, VAC, vac images vacrack.pdf
Since there's no documentation on this besides Steve's paper notebooks...

and BTW, since when did the elog start giving us PNG previews of PDFs?
Entry  Thu Jan 17 17:04:49 2013, Steve, Update, VAC, vac monitor screen changed screenmod.png
The COVAC_MONITOR.adl was changed. Ion pump labeled as disconnected means: ion pump controller power turned off and ion pump gate valve control connection
disconnected.
RP2 roughing pump labeled disconnected means:  hardware  disconnected.
    Reply  Sun Mar 22 18:42:40 2015, Steve, Update, VAC, vac pressure rose to 1.2mTorr NoN2.pngPressureReadingWorks.png
 




Quote:



We run out of N2 for
Entry  Fri Dec 26 17:45:21 2014, Steve, Update, VAC, vac pressure rose to 1.3 mTorr noN2.pngbacktoVacNormal.png
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things.
    Reply  Tue Mar 1 09:37:15 2016, steve, Update, VAC, vac rack UPS batteries replaced vacRackBat.jpg
Amstron batteries replaced after 11 months with SP-12-5.5HR,  2 years warranty from replaceUPSbattery.com




Quote:



Batteries replaced after
Entry  Fri Sep 23 16:25:01 2011, steve, Update, VAC, vac rack UPS battery replaced P1080252.JPGP1080254.JPG
APC Smart -UPS 2200   model: SUA2200RM2U   batteries were replaced by compatible RBC43, 8x  12V5A
    Reply  Fri Mar 27 10:38:40 2015, steve, Update, VAC, vac rack UPS battery replaced UPSvac.jpg
Batteries replaced after 3.5 years with Amstron AP-1250F2,  8x 12V 6Ah




Quote:



APC Smart -UPS 2200  
Entry  Thu Apr 23 16:38:33 2009, steve, Update, VAC, vac valve relay box is shorting vacrel.png
Ben and I found this vacuum valve relay box intermittently shorthing yesterday.
It effects V4, V5, VA6 and VM1........   Please do not touch this box under the beam pipe next to the vac rack!
The function of this box to send 120VAC to the vacuum valve to move.
    Reply  Mon Oct 15 13:07:51 2012, jamie, Update, VAC, vacuum VME crate broken, replaced, minor vacuum mayhem ensues 
Steve and I managed to access the fuse in the vacuum VME crate, but replacing it did not bring it back up.  We decided to replace the entire crate.
We manually checked that the most important valves, VC1, VM1 and V1, were all closed.  We disconnected their power so that they would automatically
close, and we wouldn't have to worry about them accidentally opening when we rebooted the system.
    Reply  Tue Oct 20 16:27:04 2015, ericq, Update, VAC, vacuum VME machines rebooted 
[ericq, Gautam, Steve]

Following roughly the same procedure as ELOG 11354, c1vac1 and c1vac2 were rebooted.
The symptoms were identical to the situation in that ELOG; c1vac1 could be pinged and telneted to, but c1vac2 was totally unresponsive. 
    Reply  Fri Sep 30 10:37:57 2016, Steve, Update, VAC, vacuum VME machines rebooted rebooted.png
[ Gautam and Steve ]

c1vac1 and c1vac2 were rebooted and the gauges are communicating now. V1, VA6, V5 and V4 were closed and disconnected to avoid unexpected valve
switching. All went smoothly.
    Reply  Mon Mar 1 11:21:50 2010, steve, Configuration, VAC, vacuum contoller is back to normal 


    
        
            Quote:
        
        
            
            There is a planned power outage tomorrow, Saturday from 7am till midnight.
    Reply  Thu Jan 23 08:04:04 2014, Steve, Update, VAC, vacuum control screen is blank pd76m170d2dRgaOn.png
 
There is BLANK VacControl_BAK.adl 
screen only. 
    Reply  Fri Jan 29 09:59:23 2016, Steve, Update, VAC, vacuum control upgrade 1100days.png
Rana stated yesterday that there will be a vacuum control update in the close future. Witnesses : Rich, Chris and Dave

Can you give me this in writing?

 
    Reply  Wed Oct 21 15:42:56 2015, Steve, Update, VAC, vacuum gauge check Gauges.png
Convectron gauge check:

Brand new 10 years old convectron gauge at atm was swapped into the place of existing gauges to see if they read close to 760 Torrs

They did reasonable well at the low end. I tried to imitate calibration and bring down the high end with little success.
    Reply  Mon Mar 5 10:13:21 2018, gautam, Update, VAC, vacuum health VacuumStatus.png
In Steve's absence, I've tried to keep an eye on the health of the vacuum system. From Attachment #1, the pressure of the main volume seems stable,
no red flags there. I also don't here any anomalously loud sounds near the vacuum pumps. I've changed the N2 cylinders that keep V1 open twice,
on Wednesday and Sunday of last week. So
Entry  Tue Sep 13 16:16:28 2016, Chris, Omnistructure, VAC, vacuum interlock bypassed 
(Steve, Chris)

The pumpdown had stalled because of some ancient vacuum interlock code that prevented opening the valve V1 between the turbo pump and the main
volume.