ID Date Author Type Categoryup Subject Text Attachments
  11292   Fri May 15 16:18:28 2015 SteveUpdateVACVac Operation GuideVacuum
Operation Guide is up loaded into the
40m-wiki. This is an old master copy. Not
  
  11352   Wed Jun 10 15:54:14 2015 SteveUpdateVACVacuum comp. rebootedKoji and Steve succeded rebooting C1vac1,
C1vac2 and pressure reading is working now

More tomorrow .........
 afterReb061015.png 
  11353   Thu Jun 11 19:40:59 2015 KojiUpdateVACVacuum comp. rebootedThe serial connections to the vacuum gauges
were recovered by rebooting c1vac1 and c1vac2.

Steve claimed that the vacuum screen
  
  11354   Fri Jun 12 08:40:17 2015 SteveUpdateVACVacuum comp. rebootedKoji and Steve,

One computer expert and one vacuum
expert required.
 vacMonNoGauges.pngprepReboot.png 
  11387   Wed Jul 1 10:01:25 2015 SteveUpdateVACRGA scan pd78 day 275 

 
 rga275d.png 
  11483   Thu Aug 6 09:16:57 2015 SteveUpdateVACChiara gets new Ethernet card
Yesterday we did put the vacuum
system into
 ChiaraNetCard.png23.png 
  11593   Mon Sep 14 10:41:03 2015 SteveUpdateVACTP2 dry pump replacedTP2 dry fore pump sn PLE10082 was replaced
at pressure 717 mTorr,  TP2 50K rpm
0.33A @ 112,677 hrs
 TP2_dry_pump_replaced.png 
  11625   Mon Sep 21 11:12:14 2015 SteveUpdateVACcold cathode is flakyCC4 cold cathode gauge jump triggered interlock
to close VM1 valve to protect the RGA.

The IFO pressure is 1e-5 Torr
 cc4jumps.png 
  11633   Tue Sep 22 08:58:38 2015 SteveUpdateVACcold cathode is flakyThe cold cathode gauge is back to normal.
cc4 is the last gauge is "functioning"

MKS is not responding. The spare
 4and80days.png 
  11644   Fri Sep 25 17:00:38 2015 SteveUpdateVACcold cathode is flakyThe IFO pressure is estimated ~1E-6 Torr
with modified Vac normal valve configuration.

CC4 is in a jumping mode between
 flakyCC4.png 
  11666   Mon Oct 5 10:11:35 2015 SteveUpdateVACRGA scan pd78 day 386RGA background scan.

The IFO is closed off from RGA
with VM1
 RGA_background.png 
  11678   Fri Oct 9 11:41:09 2015 ericqUpdateVACN2 Pressure fellAt 10:02AM, the N2 Pressure fell below
60 PSI. The watch script saw this happen,
but I did not recieve the email it is supposed
  
  11695   Fri Oct 16 18:27:52 2015 SteveUpdateVAC vacuum normal conditionVacuum normal configuration with VM1-closed,
VM2-open valve positions. Power load normal
24V 0.2A
 oct16Fpm2015.png 
  11697   Mon Oct 19 11:20:34 2015 gautamUpdateVACRGA scan resetSteve pointed out that in the aftermath
of the Nitrogen running out a couple of times
last week, the RGA had shut itself off thinking
 RGAscan_151019.png 
  11703   Tue Oct 20 16:27:04 2015 ericqUpdateVACvacuum VME machines rebooted[ericq, Gautam, Steve]

Following roughly the same procedure
as ELOG
  
  11705   Wed Oct 21 10:03:15 2015 SteveUpdateVACafter running out of N2 




 1d_after_reboot.png 
  11706   Wed Oct 21 15:42:56 2015 SteveUpdateVACvacuum gauge checkConvectron gauge check:

Brand new 10 years old convectron
gauge at atm was swapped into the place of
 Gauges.png 
  11707   Thu Oct 22 08:52:04 2015 SteveUpdateVACclean RGA scan after sweaty MaglevClean comparable scan at vacuum normal.
There was no backstreaming.
 clean_scan.png 
  11719   Tue Oct 27 10:08:12 2015 SteveUpdateVAC CC vacuum gaugesCC1 and CC2 are working again. Why did
they start working  again ?

 
 gauges_5y_vs_8d.png13.9yCC1&4.png 
  11832   Tue Dec 1 16:55:04 2015 SteveUpdateVACDataviewer fixedQ adjusted the Dataviewer so it is not
chopping of data any more. Thanks.

 
 10yrsCC1&4M.png 
  11858   Mon Dec 7 11:17:51 2015 SteveUpdateVACnoisy RGA scan at day 433 




 433dRGAscan.png 
  11863   Tue Dec 8 15:40:48 2015 SteveUpdateVACglitchy RGA scan at day 434The noise floor of the Rga scan is glitching
less today

 
 lessGlichingToday.png 
  11866   Wed Dec 9 11:25:55 2015 SteveUpdateVACnormal RGA scan at day 435-436Glitches are gone. Rga scan is good

 
 glichesGone.pngd436.png 
  11960   Fri Jan 29 09:59:23 2016 SteveUpdateVACvacuum control upgradeRana stated yesterday that there will be
a vacuum control update in the close future.
Witnesses : Rich, Chris and Dave
 1100days.png 
  11996   Wed Feb 17 09:05:07 2016 SteveUpdateVACforepump replacedTP3 drypump replaced after 10,344 hrs at
750 mTorr foreline pressure.

The foreline pressure is 13 mTorr
 jetStoreCrashedAgain.png 
  12014   Tue Mar 1 09:37:15 2016 steveUpdateVACvac rack UPS batteries replacedAmstron batteries replaced after 11 months
with SP-12-5.5HR,  2 years warranty
from replaceUPSbattery.com
 vacRackBat.jpg 
  12086   Thu Apr 21 15:12:38 2016 SteveUpdateVACRGA is not working 




  
  12115   Wed May 11 16:39:01 2016 ericqUpdateVACc0rga alive, output wonky


  
  12116   Thu May 12 14:29:58 2016 gautamUpdateVACRGA back up and runningIt looks like the hardware reset did the
trick. Previously, I had just tried ssh-ing
into c0rga and rebooting it. At the time,
 RGAscan_12May2016.png 
  12117   Sun May 15 19:48:08 2016 SteveUpdateVACrun out of N23-4 hrs ago we run out of nitrogen. We
are back to Vacuum Normal

 
 noN2.png 
  12119   Tue May 17 14:46:51 2016 SteveUpdateVACRGA scan at day 595We have good RGA scan now. There was no
scan for 3 months.
 RGAscan595d.pngpd78-560Hz-d600.png 
  12174   Tue Jun 14 10:13:34 2016 SteveUpdateVACtemporary N2 supply lineThe drill room floor will be retiled Thursday,
June 16. Temporary nitrogen line set up will
allow emptying the hole area.
 afterN2work.png 
  12195   Fri Jun 17 15:22:31 2016 gautamUpdateVACN2 supply line restored after retiling


 c0vac_06172016.png 
  12237   Fri Jul 1 09:20:44 2016 SteveUpdateVACvent 78  VENT
OBJECTIVES:

1, Fix ETMX sus "jump issue"
 vent78.pngvent78at250T.png 
  12238   Fri Jul 1 15:48:10 2016 SteveUpdateVACvent 78 at 710 TorrI just disconnected the 6th instrument
grade air cylinder from the vacuum envelope
at 720 Torr. Now it will reach equilibrium
 pd78at710T.png 
  12248   Wed Jul 6 08:34:00 2016 SteveUpdateVACRGA scan at day 639The last RGA scan of this pumpdown 78

Pressure plot of 640 days long
pd 78
 RGAscan639d.pngpd78d640.png 
  12250   Wed Jul 6 10:58:07 2016 SteveUpdateVACvent 78 valve configurationVacuum Status: Chamber Open

All chamber annuloses are vented. 
Vac Monitor screen is not communicating with
 CamberOpen.png 
  12293   Tue Jul 12 18:13:19 2016 KojiUpdateVACN2 bottle replacedGautam, Koji

We replaced the right N2 bottle
as it was empty.
  
  12314   Wed Jul 20 11:39:07 2016 SteveUpdateVACvent 78 day 20RGA background scan




 vent78-d20.png 
  12337   Tue Jul 26 14:24:38 2016 KojiSummaryVACPurge compressed air system at LHOI've visited the purge clean
air system at LHO Yarm mid-station with John
Worden.
 12x 
  12338   Tue Jul 26 16:01:32 2016 SteveSummaryVACPurge compressed air systemThanks for checking this out Koji

The builder in 1996 was Process
System International, Inc ( Westborough,
 cleanAirSupl.pdfmobileHEPA.jpg 
  12339   Tue Jul 26 17:41:59 2016 KojiSummaryVACPurge compressed air systemWe have no number for the CFM without calculation.
We can't assume a random number like
10-15
  
  12483   Mon Sep 12 17:54:24 2016 SteveUpdateVACpumpdown 79 We are pumping down. The annuloses
are below 10 mTorr

The vacuum gauges are not communicating
  
  12487   Tue Sep 13 16:16:28 2016 ChrisOmnistructureVACvacuum interlock bypassed(Steve, Chris)

The pumpdown had stalled because
of some ancient vacuum interlock code that
  
  12488   Tue Sep 13 17:47:04 2016 SteveUpdateVACpumpdown 79 completedP1 IFO pressure is 1 mTorr, valve configuration: vacuum normal,
annulosses are pumped, RGA is off, not pumped.
  
  12492   Wed Sep 14 09:44:09 2016 SteveUpdateVACpumpdown 79 completedThe pumpdown started at
9-12-2016

The IFO is at 5e-5 Torr vacuum
 pd79.pngpd79d2.jpgCCcg.jpg 
  12496   Thu Sep 15 10:36:32 2016 SteveUpdateVACpumpdown 79 day 3IFO pressure 3.7e-5 Torr at new cold cathode
InstruTech - Hornet

New
 vent78for73d.png 
  12525   Fri Sep 30 10:37:57 2016 SteveUpdateVACvacuum VME machines rebooted[ Gautam and Steve ]

c1vac1 and c1vac2 were rebooted
and the gauges are communicating now. V1,
 rebooted.png 
  12530   Tue Oct 4 11:02:31 2016 SteveUpdateVACRGA is out of orderThe last good rga scan at vent 78 
day 38

 lastBGscan.png 
  12537   Fri Oct 7 10:29:57 2016 SteveUpdateVACRGA removedThe RGA is removed for repaire. It's
volume at atmophere and sealed.. P4 reading
of 38 Torr is not correct.
 RGAremoved.png 