  814   Fri Aug 8 11:04:34 2008 SharonUpdate MCL Wiener filter
I took some old data from Rana and converted the units of the Weiner filter to m/m so to make the effect of the seismometer and accelerometers more obvious.

The data is in counts, and so to convert to m this is what I did:

%%% MC_L calibration
v_per_counts = 5/32768;
v_per_v = 3;
amp_per_N = 1/0.064;

%%% Accelerometers calibration
v_per_counts_acc = 61.045e-6;
g_per_v = 9.8/100;

%%% Seismometer calibration
v_per_counts_seis = 61.045e-6;
m_per_s_per_s_per_volt = 9.8/100;
m_per_v_per_s = 1/345;



for jj=1:6
hfmatm(:,jj)=hfmat(:,jj).*(v_per_counts.*v_per_v.*amp_per_N.*f)./(v_per_counts_acc*g_per_v); %%% accelerometers' data
end
hfmatm(:,7)=hfmat(:,7).*(v_per_counts.*v_per_v.*amp_per_N)./(v_per_counts_seis*m_per_v_per_s); %%% Seismometer data
  813   Fri Aug 8 10:58:05 2008 josephbConfigurationCamerasCameras and gstreamer
In regards to camera failure:
1) I forgot to reconnect that particular camera to the network (my fault) so thats why it was failing.

2) Even with the correct camera connected, I've realized at full frame rate, op440m is going to get a few frames and then fail, as I don't think it has a fast enough ethernet card. It will work on Rosalba, and will also work ssh-ing from Rosalba because it is using a new ethernet card. It also works on my laptop, which is where I originally tested the command. One way to get around this is to increase the time between pictures, by changing -l 0 to -l 1 (or higher), where the number after the "ell" is the number of seconds to wait between frame captures.

3) What I should do is figure out the UDP transmission plugins for gstreamers and compress first (using the theoraenc since it gets compression ratios of better than 100:1) and transmit that over the network.

I have since reconnected the camera, so it should work on Rosalba and any sufficiently well connected computer. For other machines like linux2 or op440, try the following line:

Running the following command on Mafalda (via ssh -X mafalda) or Rosalba while in /cvs/cds/caltech/target/Prosilica/40mCode/SnapCode/

CamSnap -F 'Mono8' -c 44058 -E 10000 -X 0 -Y 0 -H 480 -W 752 -l 1 -m 100 | gst-launch-0.10 fdsrc fd=0 blocksize=360960 ! video/x-raw-gray, height=480, width=752, bpp=8,depth=8,framerate=1/1 ! ffmpegcolorspace ! ximagesink

This will be at a much slower frame rate (1 per second) but should work on any of the machines. (Tested on linux2).
  812   Fri Aug 8 09:54:10 2008 ranaUpdateCamerasNew code + gstreamer allows for easy saving and compression of images

Quote:
Modified the CamSnap code to output the image data stream to standard out. This can then be piped into a gstreamer plugin and then be used


Didn't work; Prosilica has only 1 "l". Even so, sshing from op440m to mafalda, I got this:
mafalda:SnapCode>CamSnap -F 'Mono8' -c 44058 -E 5000 -X 0 -Y 0 -H 480 -W 752 -l 0 -m 300 | gst-launch-0.10 fdsrc fd=0 blocksize=360960 ! video/x-raw
-gray, height=480, width=752, bpp=8,depth=8,framerate=30/1 ! ffmpegcolorspace ! theoraenc ! oggmux ! filesink location="./testVideo.ogm"
Setting pipeline to PAUSED ...
Pipeline is PREROLLING ...

** (gst-launch-0.10:27121): WARNING **: Size 60 is not a multiple of unit size 360960
Caught SIGSEGV accessing address 0x487c
ERROR: from element /pipeline0/ffmpegcsp0: subclass did not specify output size
Additional debug info:
gstbasetransform.c(1495): gst_base_transform_handle_buffer (): /pipeline0/ffmpegcsp0:
subclass did not specify output size
ERROR: pipeline doesn't want to preroll.
Setting pipeline to NULL ...
#0  0xffffe410 in __kernel_vsyscall ()
#1  0xb7deddae in __lll_mutex_lock_wait ()
#2  0xb7de9aac in _L_mutex_lock_51 () from /lib/tls/i686/cmov/libpthread.so.0
#3  0xb7de949d in pthread_mutex_lock ()
#4  0xb7e452e0 in g_static_rec_mutex_lock () from /usr/lib/libglib-2.0.so.0
#5  0xb7f1fa08 in ?? () from /usr/lib/libgstreamer-0.10.so.0
#6  0x080c1220 in ?? ()
#7  0x00000001 in ?? ()
#8  0x0809586c in ?? ()
#9  0x00000001 in ?? ()
#10 0x08095868 in ?? ()
#11 0xb7f7a2a8 in ?? () from /usr/lib/libgstreamer-0.10.so.0
#12 0xb7e8da80 in ?? () from /usr/lib/libglib-2.0.so.0
#13 0xb7f7a2a8 in ?? () from /usr/lib/libgstreamer-0.10.so.0
#14 0xb7f7a2a8 in ?? () from /usr/lib/libgstreamer-0.10.so.0
#15 0x00000000 in ?? ()
Spinning.  Please run 'gdb gst-launch 27121' to continue debugging, Ctrl-C to quit, or Ctrl-\ to dump core.
Caught interrupt -- 
  811   Thu Aug 7 17:32:23 2008 JenneUpdateSUSAfternoon PRM activities
Rana, Jenne, Yoichi, Dmass

After Yoichi confirmed this morning that the wire was in both grooves, Rana attempted to lift the PRM a tiny bit, and twist it around (very gently of course) to see if we could make the wire slip back to its nominal position underneath the optic. On the first attempt, the wire ended up slipping the wrong way, causing slightly more tilt. On another attempt, the wire came out of the groove nearest the chamber door by about 0.5mm. We got the wire back in the groove by slightly lifting the optic, and pushing the wire back in. Then, on further attempts at making the wire slip back to its nominal position, the wire came out of the groove farthest from the chamber door. It is very difficult to get at that side of the PRM, because the table is crowded, and it is on the far side of the optical table from the chamber door. We decided to pull the PRM out of the chamber. Rana clamped the mirror into its cage using the earthquake stops and removed the OSEMS, and then we pulled the mirror out. We put it on a cart that was covered with foil and had a little foil house for the optic cage. We rolled the mirror+cage over to the flow bench at the end of the y-arm.

We saw that the wire is no longer even on the standoff (~3mm away from the groove) on the side that was farthest from the chamber door.

Since we have not confirmed that we have spare wire and spare magnets (and due to the time of day), we have decided to cover the cage with some foil, while it is sitting on the flow bench, and we'll fix the wire in the morning.
  810   Thu Aug 7 12:20:52 2008 YoichiUpdateSUSPRM stand-offs and wire
We removed the side OSEM of the PRM so that we can see the stand-off on the farther side.

Attachment 1: Farther side stand-off from an angle before removing the OSEM
Attachment 2: Farther side stand-off through the empty OSEM hole.
Attachment 3: Near side stand-off

The wire is definitely in the near side stand-off groove.
Probably the wire is in the groove also on the farther side.
  809   Thu Aug 7 11:54:26 2008 josephbConfigurationCamerasNew code + gstreamer allows for easy saving and compression of images
Modified the CamSnap code to output the image data stream to standard out. This can then be piped into a gstreamer plugin and then be used to save, encode, transmit, receive, slice, dice and or mangle video (or virtually any type of data stream).

The gstreamer webpage can be found at: http://www.gstreamer.net/

Under documentation you can find a list off all available plug-ins. Some good, some bad, some ugly.

Running the following command on Mafalda (via ssh -X mafalda) or Rosalba while in /cvs/cds/caltech/target/Prosilica/40mCode/SnapCode/

CamSnap -F 'Mono8' -c 44058 -E 15000 -X 0 -Y 0 -H 480 -W 752 -l 0 -m 1000 | gst-launch-0.10 fdsrc fd=0 blocksize=360960 ! video/x-raw-gray, height=480, width=752, bpp=8,depth=8,framerate=1/1 ! ffmpegcolorspace ! ximagesink

This command will create a window which displays what the camera with UID 44058 is looking at. It will display 1000 images, then quit. (You can switch the -m 100 to -i to just have it continue until the process is stopped).

You can also encode the data into compressed format and save it in a media file. The following command line will encode the images into an ogg media file (.ogm), which can be played with the totem viewer (available on Rosalba or almost any machine running Ubuntu or Centos) or any other viewer capable of handling ogm files. By switching the plugins you can generate other formats as well.

The compression is good, putting 300 images normally about 500K individually uncompressed to about 580K as a single file.

The following command line was used to generate the attached video file:

CamSnap -F 'Mono8' -c 44058 -E 5000 -X 0 -Y 0 -H 480 -W 752 -l 0 -m 300 | gst-launch-0.10 fdsrc fd=0 blocksize=360960 ! video/x-raw-gray, height=480, width=752, bpp=8,depth=8,framerate=30/1 ! ffmpegcolorspace ! theoraenc ! oggmux ! filesink location="./testVideo.ogm"

Currently looking into plugins which allow you to pull individual frames out of a video file and display or save them in a variety of formats. This would allow us to save long term images in compressed video format, and then pull out individual frames as needed.

Also need to look into how to "T" the streams, so one can be displaying while another encodes and saves.
  808   Thu Aug 7 10:27:59 2008 ranaUpdateSUSFree swinging OSEM spectra
Sometimes we see extra peaks in the OSEM spectra coming from a beat between the regular eigenmodes.
This probably comes from the OSEM shadow sensor not being entirely linear - the nonlinearity is
greatly increased if the magnet is not perfectly centered in the LED beam. So the beats are
probably there at some level in all of them; usually below the noise.
  807   Thu Aug 7 10:07:13 2008 YoichiUpdateSUSFree swinging OSEM spectra
Looks like there are more extra peaks in the SRM than other optics.
Maybe because it is closer to the door ?
  806   Wed Aug 6 22:19:07 2008 YoichiUpdateSUSBS alignment
Koji, Yoichi

We realized that we did not pay attention to the BS alignment while working on the alignment of the ITMX today. Because we were injecting the ALM laser (absolute length measurement laser) from the AS port, the ITMX alignment depends on the BS alignment.
The BS optical lever was not centered and the sum was about 2000cnt, which is low compared, for example, to the SRM oplev.
So we were not sure if the BS was in a good alignment or not.
So we decided to move the BS to center the QPD.
In doing so, we also moved the ITMX so that we do not lose the ALM laser beam coming back to the AS port.
When the BS oplev was centered, the sum of the QPD was still about 2000. So it was not far off centered.
After the tweaking, we were able to see some interference between the light reflected by the ITMY and ITMX at the AS port (actually this is the bright port for the ALM laser). By tweaking the ITMY, we were able to see Michelson fringes at the AS port.
If we believe the ALM laser alignment is still good after the vent, the ITMX, ITMY, BS and SRM should be now in a good alignment condition.
The OSEM values for the ITMX, BS, SRM seem to be ok (0.9+/-0.2). The ITMY LL is a bit low (~ 0.45).
  805   Wed Aug 6 19:01:15 2008 AlbertoUpdateGeneralITMX and SRM OSEM post-earthquake diagnostic
Koji, Yoichi, Alberto

Today we reset the OSEMs on ITMX and SRM in order to be centered when the mirrors are aligned to the IFO beam. Since the PRM is still out of order, we used the beam from NPRO laser of the absolute length measurement experiment as it is injected through the AS port.
Thatís how we did it:

1) We aligned the SRM so that the reflected beam from the NPRO was at the camera after at the AS port.

2) We traded off the alignment of SRM in order for the reflected beam at the camera to have a nice shape, avoiding any clipping from the optics, and for the optical lever to be not too far from zero. The final alignment for SRM, as read on the sliders on the MDM screen, is: Pitch=1.1650, Yaw=1.4674.

3) We aligned ITMX checking out by an IR card that the incoming and the reflected main beam in between ITMX and the BS matched. The alignment of the two beams was improved checking the matching after the SRM. The final alignment for ITMX, as read on the sliders on the MDM screen, is: Pitch=-1.2937, Yaw=-0.9890.

4) After the alignment of SRM and ITMX these were the voltages at the OSEMs:

SRM
UL=0.957
UR=1.254
LR=0.768
LL=0.620
Side=0.958

ITMX
UL=1.144
UR=1.360
LR=0.591
LL=0.325
Side=-----

5) Finally we centered the OSEMs on both mirrors and we read these voltages:

SRM
UL=0.939
UR=0.994
LR=0.782
LL=0.938
Side=0.953

ITMX
UL=0.918
UR=0.891
LR=0.887
LL=0.875
Side=0.883
  804   Wed Aug 6 13:57:44 2008 MashaSummaryAuxiliary lockingweekly summary
Finished second progress report.

Working on improving the sensitivity of the Mach Zehnder to more accurately measure the fiber noise. Making more stable mounts that have fewer degrees of freedom/springs and are more solid should get rid of their vibrational modes and help with the noise. I found a good mount for the Faraday Isolator, and John and Aidan helped me to make the solid aluminum blocks to mount the fiber couplers. I'll also replace the laser feet with a similar solid mount. I will also get a plastic box to block out acoustic noise/air currents/etc. Am starting to couple into the fiber again; now I am using polarization maintaining fiber.

I am starting to plan the fiber noise cancellation setup, and thinking about the noise sources and their effects.
  803   Wed Aug 6 13:15:57 2008 YoichiUpdateSUSSRM ETMX freeswing spectra
After yesterday's work on the SRM, I took free swinging spectra of SRM.
The eigen modes look ok. But there are many other peaks which were not present in vacuum.
Some of those peaks may be resonances of the air inside the chambers and the pipes.
However, the peaks around 0.2Hz are too low frequency for those air compression modes.
I took the ETMX spectra at roughly the same time. I chose ETMX because we have not touched it after the vent.
ETMX also shows some extra peaks but the frequencies are different.
  802   Wed Aug 6 11:43:52 2008 KojiUpdateGeneralAbs. Len. Meas. ~ analysis of the TEM01 scan
Analysis of the data on August 3th ~ Part 2

o I already have reported that the resonant freq of TEM10 and TEM01 split.

o Again, note that TEM10/01 were arranged almost in the horizontal/vertical by the observation of the video.

o The peaks of TEM10 and TEM01 were fitted with the same method as of TEM00.

o The peak freqs were:
f_TEM10: 5087040 Hz +/- 20 Hz
f_TEM01: 5068322 Hz +/- 15 Hz
The split is 18.7kHz.

o The additional parameter from the previous entry:
f_TEM00: 3879252 Hz +/- 9 Hz
L_yarm: 38.6462 m +/- 0.0003 m

o Radius of curvature
Rx = L /(1-Cos^2(Pi (f_TEM10 - f_TEM00) / (c/L/2) ))
Ry = L /(1-Cos^2(Pi (f_TEM01 - f_TEM00) / (c/L/2) ))

from these formula we get the value
Rx = 56.1620 +/- 0.0013 [m]
Ry = 57.3395 +/- 0.0011 [m]
  801   Wed Aug 6 11:10:34 2008 KojiUpdateGeneralAbs. Len. Meas. ~ analysis of the TEM00 scan
Analysis of the data on August 3th ~ Part 1

From the measurement of the 5 FSRs, the FSR frequency for the Yarm cavity was estimated as
f_FSR = 3878678 Hz +/- 30 Hz
and the Yarm length is
L_yarm = 38.6462 m +/- 0.0003 m
This is the precision of 8ppm. In my opinion, this is a satisfactory result for our purpose.
Y-arm length
e-log    length [m]
-----------------------------
 556(2008-Jun-24)  38.70    +/- 0.08    Cavity swinging measurement
 556(2008-Jun-24)  38.67    +/- 0.03    Tape & photo
 776(2008-Jul-31)  38.640   +/- 0.007   Beam injection, poor PLL, Transmitted DC
 782(2008-Aug-02)  38.6455  +/- 0.0012  Beam injection, independent PLL, Transmitted DC
 787(2008-Aug-04)  38.64575 +/- 0.00037 Beam injection, independent PLL, Transmitted RF
this(2008-Aug-04)  38.6462  +/- 0.0003  Beam injection, independent PLL, Transmitted RF, five FSRs, freq calibrated
-----------------------------
----------------
o According to the entry 795, all of the scan frequency was calibrated.
o The five peaks of the scanned data for TEM00 were fitted. Each peak was fitted by the following formula:

V(f) = A / Sqrt(1 - ((f-f0)/fc)^2)

Variable
f: scan frequency

Parameters
A: peak amplitude
f0: center frequency
fc: half bandwidth of the peak for -3dB

o The results are shown in the attached figure 1. They look very similar each other but they are different plot! The fittings were extremely good. The center frequencies estimated were as follows:
FSR1:  3879251.9 Hz +/-  8.8 Hz
FSR2:  7757968.1 Hz +/- 10.8 Hz
FSR3: 11636612.9 Hz +/- 10.2 Hz
FSR4: 15515308.1 Hz +/-  8.7 Hz
FSR5: 19393968.7 Hz +/-  8.4 Hz
o The FSR frequencies were fitted by a line. The fitting and the residuals are shown in the attached figure 2.
The fitting results were

f_FSR(n) = 586.4 + 3878678 * n

This means that:
o FSR frequency was 3878678 [Hz].
o The lock of the carrier had detuning of 586 [Hz].

The detuning of the carrier from the resonance can be explained by the alignment drift. In deed, at the end of the measurement, decrease of the transmitted power by -15% was found. Then, the frequency of the 1st FSR was measured before and after the alignment adjustment. This changed the frequency of the FSR1 by 350Hz. This change could not be explained by the cavity length change as this is too big (~3.5mm).

Actually, the spacing of the cavity length is more stable. The residual is rather scattered with in 20-30Hz. So, I took the error of 30Hz as the whole precision of the frequency measurement that includes the fluctuation of the alignment, the cavity length itself, and so on. This yields the FSR and the cavity length of
f_FSR = 3878678 Hz +/- 30 Hz
L_yarm = 38.6462 m +/- 0.0003 m .
  800   Tue Aug 5 17:56:23 2008 AlbertoConfigurationGeneralSRM and PRM inspection
Yoichi, Koji, Rana, Steve, Alberto

Today we opened the BSC to inspect the optics, and in particular the SRM and PRM.
We found that one of the side magnets of the SRM was broken and a piece of it fell and got stuck to the LR magnet.
We removed the LR OSEM and took off the broken part with tweezers. Since we couldnít replace the magnet on the side,
we decided to just switch the OSEM to the other side were a second magnet was available. Then we centered the OSEMs.
Using the optical levers we aligned both the ITMX and the SRM so that now we have to center again the OSEMs on both.

The PRM was visibly tilted and it was out of the range of the OSEMs. To try to fix the tilt we lift it up a little
with the screws on the bottom and pushed it with the third screw on top. That had the effect of making the mirror
tilt to the opposite direction. We looked at the wires (see attached picture) and it seemed centered on the side
of the mirror.

Tomorrow we are going to reset the OSEMs on ITMX and SRM and then weíre going to try to fix the tilt on PRM.
  799   Tue Aug 5 12:52:28 2008 YoichiUpdateSUSITMX, SRM OSEM spectra
Free swinging spectra of ITMX and SRM.
ITMX seems to be ok after yesterday's work, though the OSEM DC values are still a bit off from the normal value of 0.9.
(ITMX OSEM values: UL=1.12, UR=1.38, LR=0.66, LL=0.41, SIDE=0.66)
SRM is still clearly wrong.
  798   Tue Aug 5 10:56:05 2008 AlbertoConfigurationGeneralITMX chamber opened and mirror released
D-Mass, Steve, Rana, Koji, Yoichi, Alberto,
We opened the ITMX chamber to check the optics after last week earthquake. In particular, from the spectra, ITMX seemed to be stuck and had to be released again. When we inspected the mirror, we found that it wasnít necessary to touch it. It had become free again during the vent thanks to the change of conductivity in the air inside during the vent.
We checked the magnets and they seemed to be fine.
A couple of stop screws had lost the rubber on their tips, although we donít know if that was due to the earthquake.
We also took advantage of the opening to center the LR and the left OSEMs in the mirror to their zero.
Inspecting the table we found a couple of things not totally clear on the configuration of the optics in the table. In particular we found a beam dump located too close to the ifo beam. Eventually we found out that the dump was meant to block a ghost beam coming from the ITM. A better location should probably be figured out for that. We also found that the POXM1 mirror designed to have the maximum reflectivity for the P polarization of the beam at 45 degrees is mounted so that the incident beam is at 22 degrees. This cause the beam to be 90% transmitted and only 10 percent reflected to POX. The transmitted beam appears at ther BSC chamber.

The ifo beam passes so close to the POXM1 mirror so that it can be clipped by its large metal frame ring. The beam at that point is about 6mm large and the ring is about 1cm thick so that we could gain some distance with a different mount.
  797   Tue Aug 5 10:23:00 2008 steveUpdateSUSearthquake and venting effects
atm 1, EQ
atm 2, vent 7 days later: venting kicks optic into place to be free,
PRM: LR magnet gets pushed in and it is stocked, side in free
  796   Tue Aug 5 02:39:55 2008 KojiConfigurationGeneralAbs. Len. Meas. ~ Optical Layout on the AP / PSL table 2008-Aug-05
Here are the PDF and the PNG of the AP and PSL table layouts.
After this photo, the squeezing setup at the AP table was removed.
  795   Tue Aug 5 00:05:57 2008 KojiUpdateGeneralAbs. Len. Meas. ~ IFR2023A calibration
Work log on August 4th

o IFR2023A (Marconi) was calibrated by the SR620 frequency counter which is locked to the GPS signal.
o The frequency of the IFR2023A was scanned from 1MHz to 20MHz with 1MHz interval. The readout of the frequency counter was recorded.
o The linear fit was taken.
f_freq_count = K0 + K1 * f_IFR [Hz]
K0 = 0.00        +/- 0.02
K1 = 0.999999470 +/- 0.000000001

o So, the IFR seems to have -0.5ppm systematic error.
  794   Mon Aug 4 22:31:10 2008 KojiUpdateGeneralAbs. Len. Meas. ~ Simple Test for TEM01/10 split
Work log on August 3rd - Part3

Question:
o The TEM01 and TEM10 of the Yarm were found to split with 19kHz separation. Is this true?
o In which direction the eigenmodes are?

Thought:
o The separation of 19kHz is a kind of too big because the cavity bandwidth is several kHz.
o This means that "TEM01 and TEM10 can not resonate at the same time (by the PSL beam)".

Test:
o Imagine we are just using the PSL beam and playing with an arm cavity.
o Tilt the end mirror in pitch. Resonate the TEM01 mode (8-shaped).
o Then tilt the end mirror in yaw.
o a) If the resonances are degenerated within the bandwidth of the laser, it rotates freely.
o b) If the resonances splits, the tilt in yaw does not change the shape. Then suddenly jumps to TEM10 (by an accident).

Result:
o The shape does not change. Just jumps to the other mode. (The case above b.)
o The eigenmode looked like quite horizontal and vertical.

Conclusion: the mode really splits.
  793   Mon Aug 4 21:48:24 2008 KojiUpdateGeneralAbs. Len. Meas. ~ Scan for TEM00/01/10
Work log on August 3rd - Part2

o I tried to measure the frequency of the FSRs using TEM00 resonances. Also search of TEM01/TEM10 resonances were tried.

-----------
Measurement for TEM00

o The frequency of the injection beam was scanned from 2MHz to 20MHz using the LabVIEW panel with GPIB. The 1st figure attached below is the result of the scan. Equispaced peaks were found as expected. The interval of the peaks are about 3.89MHz. Each peaks were measured with freq intervals of down to 50Hz. I will analyze the center frequency of the peaks precisely later in order to have a final result.

Measurements for TEM01/TEM10

o The beam injection technique is thought to be useful for measureing the frequency of the higher-order resonances. In order to measure the higher-order resonances the modifications of the experimental setup were applied as below.

1) For TEM10 (the beam like "OO" shape), a razor blade which blocked the horizontal half of the transmitted beam was placed. We needed to disturb half of the beam because the beat between the PSL TEM00 and the injection TEM01 cancels if the PD receives all of the light.

2) The injection beam is slightly misaligned in the horizontal direction in order to enhance the coupling of the injection beam to the cavity TEM01 mode.

3) For TEM01 (the beam like "8" shape), a razor blade cutting the vertical half and the misalignment of the inj beam in the vertical direction are applied.

o The frequency of the injection beam was scaned from 1st FSR of TEM00 in the upward direction. The alignment of the arm cavity was left untouched during the measurement. As shown in the 2nd figure attached below, the resonances were found about 1.19MHz away from the TEM00, but they are separated by about 19kHz(!). This could be split of the degenerated modes which corresponds to the difference of the mirror curvature in two directions! This difference is something like 56 m and 57 m. Can you believe this?

(To be continued to the next entry)
  792   Mon Aug 4 16:20:20 2008 DmassConfigurationPhotosITMX magnet position relative to OSEMS
We have vented, and taken the following pics of the magnets to document their position before we ruin everything.
  791   Mon Aug 4 13:43:02 2008 YoichiSummaryPSLFSS loop calibration
As a part of the effort to repair the FSS loop bandwidth, I tried to calibrate the FSS loop.

First, I scanned the MOPA frequency by injecting a triangular wave into the ramp-in of the FSS box, which goes to the PZT of the NPRO.
The first attachment shows the transmitted light curve (pink one) along with the PDH signal (light blue).
The sweep was very slow (0.1Hz for 2Vp-p). From this measurement, the FWHM was 6.8e-3V. Then fpol = FWHM/2=3.4e-3V, where fpol is the cavity pole frequency.
So the PZT's DC response is 294*fpol/V. If we use the canonical fpol=38kHz, it is 11.172MHz/V.

Then I tried to measure the cavity pole. First I tried the cavity ring down measurement, by blocking the beam abruptly. Unfortunately, my hand was not fast enough.
The ring down shape was not an exponential decay.
I then locked the reference cavity only using the PZT with very narrow bandwidth (UGF=2kHz). I injected signal into the external modulation input of the 80MHz VCO
for the AOM. The second attachment shows the transfer function from this input to the IN2 (mixer output monitor port) of the FSS servo box.
To plot this, I corrected the measurement for the open loop TF (i.e. multiplying the measured TF with (1+G)), and other filters in the path (8MHz LPF after the ext. mod.
input of the 80MHz VCO, and an RCL network after the mixter). The gain looks like a cavity pole, but the phase decreases very rapidly.
If you look at the third attachment showing a wider band transfer function, there are notches at 1.8MHz and above. I couldn't find this kind of filter in the schematic.
Maybe this is the RFPD's bandpass filter. I will check this later. From these plots, it is difficult to tell the cavity pole frequency. From the -3dB point, fpol is around 83kHz,
but from the phase=-45deg point, fpol is around 40kHz.

Finally, I calibrated the cavity's optical gain by locking the Ref. Cavity with only PZT, and injecting a signal into the loop.
The signal was injected from Test-In2 of the FSS servo box and the transfer function from the PZT output signal (TP10) to IN1 (mixer output) was measured.
The transfer function was corrected for a 10Hz LPF after TP10.
The attachment4 shows a nice flat response up to 30kHz. Above 30kHz, the measurement is too noisy. The optical gain at DC is about 22dB from the PZT drive to the error signal (IN1).
Using fpol=38kHz, it means 887kHz/V calibration factor for the signal at IN1. There is a mixer output monitor DAQ channel in the FSS but it seems to be not working at the
moment. I will look into this later. There is a gain of 10dB between IN1 and the mixer monitor channel.
By looking at the phase response of the attachement4, there is a cavity pole like behavior around 30kHz. If we assume the PZT response is flat up to this frequency, it is
roughly consistent with fpol=38kHz.

I was not able to take a sensible spectrum of IN1 using the network analyzer. When the FSS servo was engaged, the signal was too small.
I will try to use an AF spectrum analyzer later to get a calibrated spectrum.
  790   Mon Aug 4 12:12:20 2008 steveConfigurationVACthe ifo is at atm
The 40m vac envelope was vented this morning.
P1 is at 760 Torr
  789   Mon Aug 4 05:23:57 2008 KojiUpdateGeneralAbs. Len. Meas. ~ Measurement for Y-arm completed
Finally, I have completed the abs length and g-factor measurements for Y-arm.
>>>GO FOR THE VENT<<<

I will report the results later.

Some notes on the status:
o Y-arm was aligned at the end of the experiment by the script. The values were saved.

o At the AP table, the injection beam and the flipper were left aligned so that the inj. beam can be used as a reference of the SRM and the ITMs. But the shutter of the NPRO was closed.

o The experiment setup was mostly left at the side of the AP table. I tried not to disturb the walk as much as possible.

o The long cable from the Y-end was wound and placed at the Y-end. The knife-edge was left on the Y-end bench. It is not disturbing any beam.
  788   Mon Aug 4 00:56:07 2008 KojiHowToGeneralAbs. Len. Meas. ~ Auto freq scanner with GPIB
Work log on August 3rd - Part1

o Yesterday I was too much tired of changing the RF frequency, reading peaks on the RF spectrum, and writing the values. Rana saw me and thought I was such poor that he gave me an USB-GPIB adapter.

o I dig into the internet for the manuals of the adapter, IFR2023A(Marconi), and HP8591E(RF spectrum analyzer) in order to learn how to use them.

o I had LabVIEW installed on my laptop. Finally I understand how to use that adapter (by Agilent) with LabVIEW. I made a small program to scan the frequency of IFR2023A, and read the peak values from HP8591E. It is unfortunate that there is no LabVIEW in the 40m lab. I think I can make an independent executable which does not need the LabVIEW itself. Give me some time to understand how to do it.
  787   Mon Aug 4 00:37:58 2008 KojiUpdateGeneralAbs. Len. Meas. ~ RF PD at the Y end / Manual frequency scan
Work log on August 2nd

o Just remind you:
The idea of the absolute length measurement was to detect an RF beat between the injection beam and the PSL beam by resonating both of the beams to the cavity at the same time, but on different londitudinal modes. From the frequency separation between the two beams, we get the FSR of the cavity. In order to have an injection beam with stable frequency separation, a heterodne interferometer was built at the PSL table, and the PLL servo is used to control and stabilize the frequency of the inj. beam.

----------

o An RF PD (Tholab PDA255) and a steering mirror were placed at the Yarm END. Fortunately, I found that an unused BS was already in the optical path. There was a beam block which dump the reflection of the BS and some stray lights of the OPLEV. I moved the beam block to make the BS reflection available, as well as to block the OPLEV stray light still (Photo1). In order to have the RF signal from the PD, a long BNC cable was laid along the Yarm. I did't know any better idea than this. Don't blame me.

o To have an intuitive interpretation of the beat frequency, the injection beam was set to be at higher frequency than the PSL beam. How did I confirm this? When the crystal temp (LT) of the NPRO was tuned to be higher, the beat frequency got lower.

o Frequency of the PLL was manually swept at around 15.51MHz where the 4th FSR was expected to be found. I could see strong RF peak at that frequency! When I tuned the PLL frequency, the peak height changes dramatically! Too cool!

o The amplitude of the RF peak was measured by an RF spectrum analyzer. I did all of this scan by my hands and eyes. The center frequency of the 4th FSR was 15.5149MHz. From the eye I would say the error is +/-150Hz. It is OK so far although I am not sure statistically this is correct or not. This corresponds to the length of 38.64575 +/- 0.00037 [m].

o All of the past measurements are fairly consistent.
Y-arm length
e-log              length [m]           Measurement Conditions
----------------------------------------------------------------------------------------
 556(2008-Jun-24)  38.67    +/- 0.03    Cavity swinging measurement
 776(2008-Jul-31)  38.640   +/- 0.007   Beam injection, poor PLL, Transmitted DC
 782(2008-Aug-02)  38.6455  +/- 0.0012  Beam injection, independent PLL, Transmitted DC
this(2008-Aug-04)  38.64575 +/- 0.00037 Beam injection, independent PLL, Transmitted RF
----------------------------------------------------------------------------------------
  786   Sun Aug 3 20:53:54 2008 ranaConfigurationPEMGuralp
We got our repaired Guralp back in the mail from England (s/n T4157). I plugged it in
to Ben's 3-Guralp breakout box (http://www.ligo.caltech.edu/docs/D/D060506-00.pdf) and
verified that it is not oscillating (like it was before) and that it responds to us
jumping around.

The breakout box has way too much gain, however. The ADC wants +/-2 V and the box puts out
~5 Vpk in the night time.

Looking at the schematic, it has a DC gain of 200 and a double whitener (50,50:10,10) so that
there's a gain of 5000 from 50-2000 Hz. The Guralp has a transduction gain of 800 V/(m/s) and
so we can just calculate what the frequency dependent noise figure of the box has to be. I've
pulled it out, put it on the bench, and started reworking it. I am looking for a soldering/
testing volunteer.

The other kink in the problem is that since we want to use this for the adaptive noise cancellation,
we have to make the noise floor of the readout better than the ambient noise by the same factor
with which we want to cancel the noise.
  785   Sat Aug 2 18:37:41 2008 ranaUpdateSUSOSEM Spectra

The attached PDF file is from the .xml files that I found from 7/30. Looks like someone
took some free swinging data and even made nice plots but didn't elog it. Raspberry for you.
The data files are saved in Templates/FreeSwinging/{ETMX,ETMY,etc.}/2008_07_30.xml

The top left plot on the multi-page file all have the same scale so you can see what's happened.
The peaks should all be as measured by Busby in Sep '06
but instead they are as you see here.
  784   Sat Aug 2 16:05:38 2008 ranaConfigurationComputer Scripts / ProgramsmDV update
I did an svn update on our mDV directory. Justin has improved it so that the NDS client binaries
are included for solaris, mac, linux32, and linux64. Now you can just use this version without
having to worry about any path definitions.
  783   Sat Aug 2 13:07:23 2008 KojiConfigurationGeneralThe AP table cleaned
During the construction of the independent PLL I cleaned up some of the unused optics from the AP table. Essentially this should be harmless as they had already been isolated from any beam. They were related to Go's squeezing project and Osamu's MC Transmitted beam measurement.

Nevertherless, if you find any problem on the signals at the AP table (when the ifo returns), I am the person to be blamed.

I am going to update the table layout later next week.
  782   Sat Aug 2 12:53:43 2008 KojiUpdateGeneralAbs. Len. Meas. ~ New PLL at the PSL table
Report of the work last night:
The new heterodyne interferometer on the PSL table was built.
The length of the Yarm cavity was measured with better precision.

-------------
Yarm is locked. The injection beam was aligned. The beat was there at around LT=48.9 [C_deg] of the NPRO.

The new PLL setup on the PSL table has been built. The two beams from the MC incident beam and the injection beam are
mode-matched with lenses. I measured the Rayleigh ranges of the beams by a sensor card and my eyes, and then placed
appropriate lenses so that they can have 5~6 [m] Rayleigh range. They looks a bit too thick but just ok for an inch
optics. The new PLL setup shows ~70% intensity modulation which is enormous. The servo is still SR560-based so far.

Now the PLL has no singular frequency within its range. I could sweep the 4th FSR of the cavity with 500Hz interval. I
was still observing at the transmitted DC.

At each freqency from 15.51MHz to 15.52MHz, a timeseries data of the Yarm transmitted was recorded at sampling of 32Hz for 10
seconds. The figure shows the averaged values of the transmitted DC with errors. An increase of the transmitted power by
3-4% was found. If we consider the resonance is at f_PLL = 15.515 +/- 0.0005 [MHz], this indicates the
arm cavity length of 38.6455 +/- 0.0012 [m].
Y-arm length
e-log    length [m]
-----------------------------
556      38.70   +/- 0.08   Cavity swinging measurement
556      38.67   +/- 0.03   tape & photo
776      38.640  +/- 0.007  Beam injection, poor PLL, Transmitted DC
this     38.6455 +/- 0.0012 Beam injection, independent PLL, Transmitted DC
-----------------------------

NEXT STEPS:
o RF detection at the transmitted
o Better PLL: PLL stability (in-loop / out-of-loop)
o Measurement for the 1st~3rd FSR
o Reproducibility of the measurement
o Higher order mode search
o Check the acuuracy and presicion of the Marconi
  781   Fri Aug 1 16:33:52 2008 ranaConfigurationPSLPSL Quad change and new script
Here's the sensor ringdown trend from the kick.
  780   Fri Aug 1 11:51:15 2008 justingOmnistructureComputersadded /cvs/cds/site directory
I added a /cvs/cds/site directory. This is the same as is dicsussed here. Right now it just has the text file 'cit' in it, but eventually the other scripts should be added. I'll probably use it in the next version of mDV.
  779   Fri Aug 1 10:45:46 2008 josephbConfigurationComputersMegatron now running tcsh
At Rana's request, I've remotely switched Megatron over to using tcsh. I had to ssh -X in order ot use the "/sbin/system-config-users" program which is a graphical UI for modifying users. I had to go to preferences and uncheck hide system users, which then allowed me to see the controls user (at the bottom of the list), and edit it.

I also created a .tcshrc file in the /home/controls directory and copied the information from the .bashrc file, and also moved the matlab path definition into the PATH environment variable.

Does anyone know if sourcing /cvs/cds/caltech/cshrc.40m would be usable on a 64 bit machine, or does a new one need to be made for Megatron and/or Rosalba?
  778   Fri Aug 1 01:13:32 2008 ranaConfigurationPSLPSL Quad change and new script
Koji and I changed a few optics so that now ~60% of the beam that went to the PSL POS QPD
now goes to the west side of the table for the aux. laser locking PLL. The beam is sort of
on the QPD again but needs a centering.

After this work I wrote a script SUS/freeswing-all.csh which puts a 30000 count offset into
the UL coil of each suspension and then disables it. This is just good enough to kick it up
so that the eigenfrequency can be measured. I ran it and it worked -- it finished running at

Fri Aug 1 00:44:30 PDT 2008
  777   Thu Jul 31 16:11:22 2008 josephbConfigurationComputersMatlab on Megatron
Matlab now works on megatron.

I did a few things:

1) Added to the PATH environment variable. Did this in .bash_profile in the /home/controls directory by adding the line

PATH=$PATH:/cvs/cds/caltech/apps/linux64/matlab/bin/
export PATH

This probably should be somewhere else up further up the line, but I was too lazy to figure it out.

2)Fixed a gateway mistake I had added earlier so the megatron could use the NAT router and see the outside world so yum worked.

3) Removed the i386 based libXp and openmotif packages.

4) Installed the x86_64 based libXp and openmotif packages.

Edit: Forgot that I also added the following line to the /etc/fstab file in order to mount the shared code. This was stolen directly from Rosalba's /etc/fstab file. This was so that it could see the matlab code.
linux1:/home/cds/ /cvs/cds nfs rw,bg,soft 0 0
  776   Thu Jul 31 11:19:30 2008 KojiUpdateGeneralAbs. Len. Meas. ~ Resonance search trial
Last night, I tried to find the resonance of Yarm by sweeping the frequency of the injection beam.

A strong beat was present at LT_NPRO=48.7856[C_deg], the power coupling of the injection beam was estimated to be 35%. 
(Vmax_beat = 1.060[V], Vmin_beat = 0.460[V], Vno_inject = 0.664[V])

The Yarm was locked and the alignment script was executed. The PLL between the PSL beam and the injection beam was 
somehow locked.

I tried to scan the freq offset (f_PLL) at around 3.88MHz first, then at around 15.52MHz. They are supporsed to be the 
first and fourth FSR of the Yarm cavity. The Yarm transmitted power (DC) was observed to find the resonance of the 
injection beam. It would have been better to use the RF power, but so far I didnot have the RF PD prepared at the end 
transmission. I just used the DC power.

I think I saw the increase of the transmitted power by 10%, at f_PLL = 15.517 +/- 0.003 [MHz]. This corresponds to the 
arm cavity length of 38.640 +/- 0.007 [m]. The previous measurement was not so bad!

Y-arm length
e-log             length [m]
-----------------------
556(2008-Jun-24)  38.70    +/- 0.08    Cavity swinging measurement
556(2008-Jun-24)  38.67    +/- 0.03    tape & photo
This              38.640   +/- 0.007

However, I had difficulties to have more precise measurement mainly because of two reasons:
o The PLL servo is too naive, and the freqency stability of the inj beam is not enough.
  The injected beam should have the linewidth (=freq stability) narrower than the cavity linewidth.

o The PLL servo may experience change of the transfer function at around the resonance. The PLL works the other 
frequencies. However, close to the resonance, it starts to be unstable.

So the next stuffs we should do is 
o Build the PLL just using the incident beams to the ifo, not by the reflected beams.
o Build sophisticated servo to have better frequency stability.
o RF PD at the transmission.

Left the lab with Yarm locked, flipper down, shutter for the NPRO closed.
  775   Thu Jul 31 10:27:17 2008 ranaUpdatePSLPMC Scan Graphs

Quote:
Graphs of the PMC scan data that I got earlier today.

On the UNIX computers, one can use 'convert' to change these to PNG. A DC offset should be added to the transmitted
light so that the scan can be plotted with a log y-scale. And, of course, Acrobat can be used to make it into a
single PDF file.

The PMC scan always has this distortion and so the input power has to be decreased to a few mW to reduce the
thermal expansion effect; the expansion coefficient for SiO2 is ~5 x 10^-7 / K and we're worried about nm level
expansions.
  774   Thu Jul 31 10:24:32 2008 KojiUpdateGeneralIFO status
Last night I used the Y-arm for the abs length measurement. The Yarm was aligned by the script.
I left the ifo with the Yarm locked as it is the only meaningful configuration so far.
  773   Wed Jul 30 18:45:01 2008 ranaConfigurationSUSNew SUS Drift Technology
I updated the SUS DRIFT screen again, this time with a new feature.

I used Rolf's idea for the AdvLIGO status system and just made the nominals be an
unused sub-field (.SVAL) of the main INMON record. Then I wrote a .csh script to
use tdsavg to average the current INMON vals and insert that as the .SVAL. The next
script should read the .SVAL and set the HIHI and LOLO based on this.

Of course, the values I have just entered are no good because our suspensions are in
a bad state but we can run this script (SUS/setDriftNoms) next time things are good.

And...even better would be if someone were to do the same but used mDV to grab the
minute trend in the past good times instead of the tdsavg (which can't go in the past).
  772   Wed Jul 30 16:35:56 2008 EricUpdatePSLPMC Scan Graphs
Graphs of the PMC scan data that I got earlier today.

PMCLongScanWide.tiff shows the transmission intensity and PZT voltage plotted against time for a longer scan of the PMC (~120 seconds for one sweep).

PMCLongScanPeak.tiff is the same scan zoomed in on the primary peak. This scan was done with the laser power at around 1/3 its original value. However, scans done at around 1/6 the original value have peaks that are just as messy.

PMCShortScanWide.tiff shows the intensity and voltage for a more rapid scan (~30 second for one sweep). The black lines show how the peak positions are at very different PZT voltages (a difference of ~10 volts in both cases).

PMCShortScanPeak.tiff is zoomed in on the primary peak. The peak is much cleaner than for the long scan (less time for the laser's heat to expand the mirror?), though it is likely still too messy to reliably fit to a lorentzian.
  771   Wed Jul 30 15:28:08 2008 robUpdateLSCY arm locked

By using a combination of the SUS-DRIFT mon screen and the optical levers (which turned out pretty well) I steered the BS, ITMY, and ETMY back to their previous positions, and was able to lock the Y arm. The "Restore Y Arm" script on the IFO_CONFIGURE screen works. I couldn't test the alignment script, as a dump truck/construction vehicle showed up and started unlocking the MC.
  770   Wed Jul 30 15:12:08 2008 ranaSummaryIOOHistory of the MC abs length
> I was notified by Rob and Rana that there were many measurements of the MC abs length (i.e. modulation
> frequencies for the IFO.) between 2002 and now.

I will just add that I think that the Marconi/IFR has always been off by ~150-200 Hz
in that the frequency measured by the GPS locked frequency counter is different from
what's reported by the Marconi's front panel. We should, in the future, clearly indicate
which display is being used.
  769   Wed Jul 30 13:52:41 2008 EricSummaryCamerasWeekly Summary
I tracked the tendency for ezcaPut to fail and sometimes seg-fault in the camera code to a conflict between the camera API and ezca, either on the 
network level or the thread level.  Since neither are sophisticated enough to provide controls over how they handle these two things, I instead 
separated the call to ezcaPut out into a small, separate script (a stripped down ezcawrite), which the camera code calls at the system level.  This is a 
bit hacky of a solution, but its the only thing that seems to work.

I've developed a transformation based on Euler angles that should be able to take the 4 OSEMs in a picture of the end mirror and use their relative 
positions to determine the angle of the camera to the optic.  This would allow the position data determined by the fitting software to be converted 
from pixels to meaningful lengths, and should aid any servo-ing done on the beams position.  I've yet to actually test if the equations work, though.

The servo code needs to have slew rate limiters and maximums/minimums to protect the mirrors written in to it before it can be tested again, but I 
have no idea what reasonable values for these limits are.

Joe and I recently scanned the PMC by driving C1:PSL-PMC_RAMP with the trianglewave script over a range of -3.5 to -1.25 (around 50 to 150 volts 
to the PZT) and read out C1:PSL-ISS_INMONPD to measure the transmission intensity.  This included slightly under 2 FSRs.  For slow scans (covering 
the range in 150 to 300 s), the peaks were very messy (even with the laser power at 1/6 its normal value), and it was difficult to place where the 
actual peak center occurred.  For faster sans (covering the range in 30 seconds or so), the peaks were very clean and nearly symmetric, but were 
not placed logically (the same peak showed up at two very different values for the PZT voltage in two separate runs).  I don't have time to put 
together graphs of the scans at the moment; I'll have that up sometime this afternoon.
  768   Wed Jul 30 13:14:03 2008 KojiSummaryIOOHistory of the MC abs length
I was notified by Rob and Rana that there were many measurements of the MC abs length (i.e. modulation 
frequencies for the IFO.) between 2002 and now.

So, I dig the new and old e-logs and collected the measured values of the MC length, as shown below. 

I checked the presence of the vent for two big steps in the MC length. Each actually has a vent. 
The elog said that the tilt of the table was changed at the OMC installation in 2006 Oct.
It is told that the MC mirrors were moved a lot during the vent in 2007 Nov.

Note:
o The current modulation freq setting is the highest ever.
o Rob commented that the Marconi may drift in a long time.
o Apparently we need another measurement as we had the big earthquake.

My curiosity is now satified so far.

Local Time	3xFSR[MHz]	5xFSR[MHz]	MC round trip[m]	Measured by
----------------------------------------------------------------------------
2002/09/12	33.195400 	165.977000 	27.09343		Osamu
2002/10/16	33.194871 	165.974355 	27.09387		Osamu
2003/10/10	33.194929 	165.974645 	27.09382		Osamu
2004/12/14	33.194609 	165.973045 	27.09408		Osamu
2005/02/11	33.195123 	165.975615 	27.09366		Osamu
2005/02/14	33.195152 	165.975760 	27.09364		Osamu
2006/08/08	33.194700 	165.973500 	27.09401		Sam
2006/09/07	33.194490 	165.972450 	27.09418		Sam/Rana
2006/09/08	33.194550 	165.972750 	27.09413		Sam/Rana
----2006/10 VENT OMC installation	
2006/10/26	33.192985 	165.964925 	27.09541		Kirk/Sam
2006/10/27	33.192955 	165.964775 	27.09543		Kirk/Sam
2007/01/17	33.192833 	165.964165 	27.09553		Tobin/Kirk
2007/08/29	33.192120 	165.960600 	27.09611		Keita/Andrey/Rana
----2007/11 VENT Cleaning of the MC mirrors
2007/11/06	33.195439 	165.977195 	27.09340		Rob/Tobin
2008/07/29	33.196629 	165.983145 	27.09243		Rob/Yoichi
  767   Wed Jul 30 13:09:40 2008 josephb, EricConfigurationPSLPMC scan experiment
We turned the PSL power down by a factor of 4, blocked one half of the Mach Zehnder and scanned the PMC by applying a ramp signal to PMC PZT. Eric will adding plots later today of those results.

We returned the power to close to original level and removed the block on the Mach Zehnder, and then relocked the PMC.
  766   Wed Jul 30 13:08:44 2008 Max JonesUpdateComputer Scripts / ProgramsWeekly Summary
This week I've been working on the noise budget script. The goal is to add Siesmic, Darm, Mich, Prc and magnetometer noise. I believe I've added Seismic noise in a reasonable and 40m specific manner (please see the attached graph). The seismic noise in the noise budget at 100 Hz was 10 times higher than that predicted by Rana in elog #718. This could be due to the fact that graph is taken from data today when the device is unlocked and construction workers are busy next door. I am currently trying to fix the getDarm.m file to add the DARM source to the noise budget. I have run into several problems, the most pressing of which was that the C1:LSC-DARM_ERR channel is zero except with the interferometer is being locked. According to Rob, we only save data for approximately a day (we save trends for much longer but this is insufficient for the noise budget script) and sometimes we are not locked the night before. Rob showed me how I may introduce an artificial noise in the DARM_ERR signal but I'm having trouble making the script output a graphic. I'm still unsure how to make the getDarm function 40m specific.

Today I will start working on my second progress report and abstract.
  765   Wed Jul 30 12:36:19 2008 SharonUpdate Weekly update
This week included many computer's issues. I tested Alex's new C code (the one that saves the FIR coefficients and restores them when you start running the code again). Seems there is an improvement in the adaptation time, but not a significant one (more details on the coming report). I had to recompile simulink and the FB whenever I wanted to find a solution for taking the record of those coefficients. This is so I could simulate the adaptive filter with a regular IIR filter and compare the two.
After Rob tried to help and it seems to be an impossible to a huge hassle mission, we thought of a different method to do this. I re-compiled the old simulink file and restored the .ini file and all should be back in place. Instead of finding the FIR coefficients, I am going to use one noise source in the adaptive filter, stop the adaptation (by setting mu and tau to 0), and put excitation instead of the noise signal. The transfer function I will get between the exc. and MC1_IN1 is the filter I am looking for.

Also seems that whenever I get the MC unstable, and the adaptive code stops itself, it doesn't come back. Setting the reset flag to a different number (anything other than 0) and pressing the reset button will get it working again, but the CPU will always flip and the ASS computer needed a restart. Still haven't found a problem in the C code, but that's the plan. Moreover, I want to change Alex's code, so that instead of starting from zero like in Matt's code, or starting from the old coefficients like in Alex's, it is going to calculate a Wiener Filter as the first set of coefficients. This will hopefully reduce the adaptation time.

I have also been working on my progress report, and stood in line for the MC... Still standing...
ELOG V3.1.3-