ID |
Date |
Author |
Type |
Category |
Subject |
Text |
 |
16652
|
Wed Feb 9 11:56:24 2022 |
Anchal | Update | General | Bringing back CDS | [Anchal, Paco]
Bringing back CDS took a lot of
work yesterday. I'm gonna try to summarize |
|
16653
|
Wed Feb 9 13:55:05 2022 |
Koji | Update | General | Bringing back CDS | Great recovery work and cleaning of the
rebooting process.
I'm just curious: Did you observe |
|
16655
|
Wed Feb 9 16:43:35 2022 |
Paco | Update | General | Scheduled power outage recovery - Locking mode cleaner(s) | [Paco, Anchal]
We went in and measured
the power after the power splitting HWP at |
|
16657
|
Thu Feb 10 15:41:00 2022 |
Anchal | Update | General | Scheduled power outage recovery - Locking mode cleaner(s) | I found out that the ESP300 service needs
to be run in root mode for it to be able
to connect to the USB port of HWP motor controller. |
|
16658
|
Thu Feb 10 17:57:48 2022 |
Anchal | Update | General | Scheduled power outage recovery - Locking mode cleaner(s) | Something is wrong with the Video MUX.
The system did not turn back on with full
functionality. Even though we see the screens |
|
16659
|
Thu Feb 10 19:03:23 2022 |
Koji | Update | General | Scheduled power outage recovery - Locking mode cleaner(s) | I came back to the 40m and started the
investigation.
If I ping 192.168.113.92, it responds. |
|
16660
|
Thu Feb 10 19:46:37 2022 |
Koji | Update | General | Scheduled power outage recovery - Locking mode cleaner(s) | ==
Assign new IP address to c1sus2 ==
cf: [40m |
|
16661
|
Thu Feb 10 21:10:43 2022 |
Koji | Update | General | Video Mux setting reset | Now the video matrix is responding correctly
and the web interface shows up. (Attachment
1) |
 |
16663
|
Thu Feb 10 21:51:02 2022 |
Koji | Update | CDS | [Solved] Huge random numbers flowing into ETMX/ETMY ASC PIT/YAW | Huge random numbers are flowing into ETMX/ETMY
ASC PIT/YAW. Because of this, I could not
damp the ETMX/ETMY suspension at the beginning |
  |
16664
|
Fri Feb 11 10:56:38 2022 |
Anchal | Update | CDS | [Solved] Huge random numbers flowing into ETMX/ETMY ASC PIT/YAW | Yeah, this is a known issue actually. We
go to ASC screen and manually swich off all
the outputs after every reboot. We haven't |
|
16665
|
Fri Feb 11 11:17:00 2022 |
Anchal | Update | General | Scheduled power outage recovery | I found that two computers are not powering
up in the control room, Ottavia and Allegra.
Allegra was important for us as it had the |
|
16666
|
Fri Feb 11 12:22:19 2022 |
rana | Update | CDS | [Solved] Huge random numbers flowing into ETMX/ETMY ASC PIT/YAW | you
can hand edit the autoBurt file which the
FE uses to set the values after boot up. |
|
16667
|
Fri Feb 11 16:09:11 2022 |
Anchal | Update | General | Scheduled power outage recovery - Input power increased | We increased the input power to IMC by
replacing the 98% transmission BS by a 10%
transmission BS on the detection table (reverse |
|
16668
|
Fri Feb 11 17:07:19 2022 |
Anchal | Update | CDS | [Solved] Huge random numbers flowing into ETMX/ETMY ASC PIT/YAW | The autoBurt file for FE already has the
C1:ASC-ETMX_PIT_SW2 (and other channels for
ETMY, ITMX, ITMY, BS and for YAW) present, |
|
16669
|
Mon Feb 14 18:31:50 2022 |
Paco | Update | General | Scheduled power outage recovery - IMC recovery progress | [Paco, Anchal, Tega]
We have been realigning the IMC
as of last Friday (02/11). Today we made |
|
16671
|
Mon Feb 14 21:03:25 2022 |
Koji | Update | General | Scheduled power outage recovery | I opened the boxes. Allegra has obvious
vent of at least 4 caps. And the power supply
did not respond even a paper clip test was |
|
16672
|
Tue Feb 15 19:32:50 2022 |
Koji | Update | General | Scheduled power outage recovery - IMC recovery progress | Reduced the IMC power to 100mW
Setup: The power
meter was placed right before the final aperture |
    |
16673
|
Tue Feb 15 19:40:02 2022 |
Koji | Update | General | IMC locking | IMC
is locking now. There was
nothing wrong: just a careful alignment + |
|
16674
|
Wed Feb 16 15:19:41 2022 |
Anchal | Update | General | Reconfigured MC reflection path for low power | I reconfigured the MC reflection path for
low power. This meant the following changes:
Replaced the 10% reflection |
|
16675
|
Tue Feb 22 18:47:51 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement | [Ian, Koji]
In preparation for the replacement
of the suspension electronics that control |
 |
16676
|
Wed Feb 23 15:08:57 2022 |
Anchal | Update | General | Removed extra beamsplitter in MC WFS path | As discussed in the meeting, I removed
the extra beam splitter that dumps most of
the beam going towards WFS photodiodes. |
|
16677
|
Thu Feb 24 14:32:57 2022 |
Anchal | Update | General | MC RFPD DCMON channel got stuck to 0 | I found a peculiar issue today. The C1:IOO-MC_RFPD_DCMON
remains constantly 0. I wonder if the RFPF
output is being read properly. I opened the |
|
16678
|
Thu Feb 24 18:05:58 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 | {Yehonathan, Anchal, Paco}
Yesterday, Anchal and Paco removed
AS1 from the vacuum chamber and moved it |
 |
16679
|
Thu Feb 24 19:26:32 2022 |
Anchal | Update | General | IMC Locking | I think I have aligned the cavity, including
MC1 such that we are seeing flashing of fundamental
mode and significant transmission sum value |
|
16680
|
Fri Feb 25 14:00:08 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement | [Koji, Ian]
We looked at a few power supplies
and we found one that was marked "CHECK |
 |
16681
|
Fri Feb 25 14:48:53 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement | I moved the network-enabled power strip
from above the power supplies on rack 1y4
to below. Nothing was powered through the |
 |
16682
|
Sat Feb 26 01:01:40 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | I will make a detailed elog later today
giving a detailed outline of the connection from
the Agilent gauge controller to the vacuum |
|
16683
|
Sat Feb 26 15:45:14 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | I have attached a flow diagram of my understanding
of how the gauges are connected to the network.
Earlier today, I connected the |
|
16684
|
Sat Feb 26 23:48:14 2022 |
Koji | Update | SUS | ETMY SUS Electronics Replacement | [Ian, Koji] - Activity
on 25th (Fri)
We continued working on |
|
16685
|
Sun Feb 27 00:37:00 2022 |
Koji | Update | General | IMC Locking Recovery | Summary:
- IMC was locked.
- Some alignment change in the output |
    |
16686
|
Sun Feb 27 01:12:46 2022 |
Koji | Update | General | IMC manual alignment procedure | We expect that the MC sus are susceptible
to the temperature change and the alignment
drifts away with time. |
|
16687
|
Mon Feb 28 15:51:07 2022 |
Ian MacMillan | Update | SUS | ETMY 1Y4 Electronics Replacement | [Paco, Ian]
paco helped me wire the ETMY 1Y4
rack. We wired the following (copied from |
   |
16688
|
Mon Feb 28 19:15:10 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | I decided to create an independent service
for the XGS data readout so we can get this
to work first before trying to integrate |
|
16689
|
Tue Mar 1 16:01:14 2022 |
Paco | Update | Electronics | RFSoC 2x2 board -- setup for remote work & BALUN saga | [Tommy, Paco]
Since last week I've worked
with tommy on getting the RFSoC 2x2 board |
 |
16690
|
Tue Mar 1 19:26:24 2022 |
Koji | Update | SUS | ETMY SUS Electronics Replacement | The replacement key switches and Ne Indicators
came in. They were replaced and work fine
now. |
 |
16691
|
Tue Mar 1 20:38:49 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | During my investigation, I inadvertently
overwrote the serial port configuration for
the connected devices. So I am now working |
   |
16692
|
Wed Mar 2 11:50:39 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | Here is the IOLAN SDS TCP socket setting
and the USBserial setting for comparison.
I have also included the python |
    |
16693
|
Wed Mar 2 12:40:08 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | Connector Test:
A quick test to rule out any issue
with the Ethernet to Serial adapter was done |
6x |
16694
|
Wed Mar 2 14:02:43 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 | Yesterday, I rebuilt the OpLev setup in
the cleanroom in order to suspend AS1. It
took me a while to find all the necessary |
|
16695
|
Thu Mar 3 04:11:36 2022 |
Koji | Update | SUS | ETMY 1Y4 Electronics Replacement | For the Y-end electronics replacement,
we want to remove unused power supplies.
In fact, we already removed the +/-5V supplies |
  |
16696
|
Thu Mar 3 04:24:23 2022 |
Koji | Update | SUS | ETMY 1Y4 Electronics Replacement | The DC power strip at Y-end was connected
to the bottom two Sorensen power supplies.
They are configured to provide +/-18V. |
 |
16698
|
Thu Mar 3 17:09:46 2022 |
Paco | Update | BHD | Re-susspension of AS1 | [Anchal, Paco]
Wire clamp spare was installed,
furthermore AS1 was reinstalled on adapter, |
|
16699
|
Thu Mar 3 17:21:11 2022 |
Ian MacMillan | Update | SUS | ETMY 1Y4 Electronics Replacement | [Koji, Ian]
1) We attached the 30 coil driving
cables to the vacuum feed through to the |
|
16701
|
Fri Mar 4 18:12:44 2022 |
Koji | Update | VAC | RGA pumping down | 1. Jordan reported that the newly installed
Pirani gauge for P2 shows 850Torr while PTP2
show 680 Torr. Because of this, the vacuum |
|
16703
|
Sat Mar 5 02:03:46 2022 |
Koji | Update | SUS | ETMY 1Y4 Electronics Replacement | Oplev saga
Summary
- The new coil driver had not enough |
6x |
16704
|
Sun Mar 6 18:14:45 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | Following repeated failure to establish
communication between c1vac and the XGS600
controller via Perle IOLAN serial device |
|
16705
|
Mon Mar 7 10:06:32 2022 |
Yehonathan | Update | IOO | IMC unlocked again, completely misaligned | Came this morning and saw that the IMC
is unlocked.
Went into MC Lock screen and see |
|
16706
|
Mon Mar 7 13:53:40 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | So it appears that my deduction from
the pictures of needing a cable swap was
correct, however, it turns out that the installed |
|
16707
|
Mon Mar 7 14:52:34 2022 |
Koji | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels | Great trouble shoot!
> I guess, the only remaining
issue now is the incorrect atmospheric |
|
16708
|
Mon Mar 7 14:55:33 2022 |
Koji | Update | IOO | IMC unlocked again, completely misaligned | Hmm, the bias values were reset at 2022-03-03-20:01UTC
which is 2022-03-03-12:01 PST with no apparent