    Reply  Fri Mar 28 16:26:20 2014, steve, Update, VAC, RGA scan after power failure afterPowerFailure.png


    
        
            Quote:
        
        
            

            Out gassing plus leak rate   0.15  mTorr / hour
    Reply  Wed Apr 2 16:34:15 2014, Steve, Update, VAC, Maglev controller needs service 


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Thu Apr 3 17:05:52 2014, Steve, Update, VAC, Maglev controller swapped loanerControllerIn.png


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Thu Apr 10 16:09:29 2014, Steve, Update, VAC, RGA scan at 75% pumping speed 520HzpumpingSpeed.png


    
        
            Quote:
        
        
            

            
    Reply  Mon Apr 21 10:14:00 2014, Steve, Update, VAC, Maglev controller serviced ourControllerBack560Hz.png


    
        
            Quote:
        
        
            

            
    Reply  Tue Apr 29 16:42:29 2014, Steve, Update, VAC, TP2 drypump replaced dryforepumpreplaced.png


    
        
            Quote:
        
        
            

            
    Reply  Tue May 27 11:00:43 2014, Steve, Update, VAC, RGA scan at day 111 pd77d111RGA.png
 
 Rga scan at pump down 77, vacuum normal valve configuration, maglev rotation 560 Hz and day 111
Entry  Mon Jun 30 10:06:39 2014, Steve, Update, VAC, low on pneumatic pressure of vacuum valves runoutofN2.pnglowN2.pngN2d100.png
This morning valve condition: V1, VM1, V4 and V5 valves were closed. IFO pressure rose to 1.3 mTorr
It was caused by low N2 pressure.  Our vacuum valves are moved-controlled by 60-70 PSI of nitrogen.
When this supply drops below 50-60 PSI the interlock closes V1 valve. This is the minimum pressure required to move the large valves.
    Reply  Wed Jul 2 11:47:26 2014, Steve, Update, VAC, broken changeover regulator of N2 supply 


    
        
            Quote:
        
        
            

           This morning valve condition: V1, VM1, V4 and V5 valves were closed. IFO pressure rose to 1.3 mTorr
Entry  Mon Jul 21 10:51:06 2014, Steve, Update, VAC, N2 supply run out outofN2.png
Interlock closed valve V1, V4, V5 and VM1 when the nitrogen supply run out. The IFO pressure rose to P1 1 mTorr
In order to recover Vacuum Normal valve configuration I did the following:
Replaced both nitrogen cylinders. Confirmed pneumatic nitrogen pressure 70 PSI.   Opened valves V4 and V5
    Reply  Thu Aug 21 15:07:48 2014, Steve, Update, VAC, RGA scan at day 197 RGAscan@d197.png
 
 
    Reply  Mon Sep 22 15:20:32 2014, Steve, Update, VAC, RGA scan at day 229 RGA@229day.pngRGAscan229d.png
 
 
    Reply  Wed Sep 24 08:40:29 2014, Steve, Update, VAC, Vent has started vent77.pngbvent77.png
 
 Jam nuts checked. Oplev servos turned off. Particle count checked. Vertex crane functionality checked.
Entry  Wed Sep 24 13:28:48 2014, Steve, Update, VAC, vent is completed ventiscompleted.pngatAtm.png
 The vent is completed. ITMX was kicked up accidentally. Valve configuration: chamber open, RGA is pumped through VM2  Maglev
    Reply  Wed Sep 24 16:02:58 2014, Jenne, Update, VAC, vent is completed 
[Steve, EricQ, Jenne]
ITMY and BS heavy doors are off, light doors are on.  Q is aligning the IFO.
    Reply  Thu Sep 25 08:21:15 2014, Steve, Update, VAC, vent day 1 ventDay1.png


    
        
            Quote:
        
        
            

           [Steve, EricQ, Jenne]
    Reply  Mon Sep 29 09:11:25 2014, Steve, Update, VAC, vent day 5 RGA scan v77_bg-mag-d5.png
 
 
    Reply  Mon Sep 29 10:29:25 2014, Steve, Update, VAC, RGA is not running rgascanTimeStamp.png
 
 The RGA time stamp was correct last at 20140527
 
    Reply  Tue Sep 30 16:18:57 2014, Steve, Update, VAC, pump down #78 has started pd78hastarted.png
 
Q checked the earth quake stops of SRM and we put the ITMY & BS doors on. 
    Reply  Tue Sep 30 18:02:53 2014, Steve, Update, VAC, pump down #78 stops at 320 Torr 3h.png
 
We  stopped pumping just short of 3 hours at 320 Torr.  Pumping speed was 2.7 Torr / min with partially closed RV1 and butterfly valve/
RP1&3 roughing pump hose is disconnected. Butterfly valve removed. The vac envelope is closed. 
    Reply  Wed Oct 1 10:21:19 2014, Steve, Update, VAC, pump down #78 stops at 3.5 Torr pd78secondStop.png


    
        
            Quote:
        
        
            

            
    Reply  Wed Oct 1 16:26:53 2014, Steve, Update, VAC, pump down #78 completed pd78completedin25hrs.png
 
 Pump down reached "vacuum normal" state. IFO _P1 pressure 1e-4 torr in 8 hrs actual pumping time
 PSL shutter is opened.
    Reply  Thu Oct 2 09:23:23 2014, Steve, Update, VAC, vent 77 vent77SRMmoved.png


    
        
            Quote:
        
        
            

            
    Reply  Fri Oct 3 10:10:37 2014, Steve, Update, VAC, cold cathode gauge reading switched CC1switched.png
 
We have two cold cathode gauges at the pump spool and one  signal cable to controller. CC1  in horizontal position and CC1 in vertical
position.  
    Reply  Mon Oct 6 10:23:43 2014, Steve, Update, VAC, Unexpected power shutdown 


    
        
            Quote:
        
        
            

           We had an unexpected power shutdown for 5 sec at ~ 9:15 AM.
    Reply  Tue Oct 7 16:55:16 2014, Steve, Update, VAC, Unexpected sweaty valves sweatyV5.jpgsweatyV5f.jpg
 Pump  spool valves V5, V4, V3 sweating a lot. VM3 and VC2 not so much. 

They are VAT valves F28-62887-03, 11, 14 and so on ~15-16 years old.

 I'm speculating that some plastic is aging-braking down
at the     atmospheric-pneumatic side of valves.
    Reply  Wed Oct 8 08:46:57 2014, Steve, Update, VAC, VAT valves actuator lubricant 


    
        
            Quote:
        
        
            

            Pump  spool valves V5, V4, V3 sweating a lot. VM3 and VC2 not so much.
    Reply  Thu Oct 9 11:56:35 2014, Steve, Update, VAC, Power outage II & recovery 


    
        
            Quote:
        
        
            

           Post 30-40min unexpected power outage this morning, Steve checked the status of the vacuum and I powered up Chiara.
    Reply  Thu Oct 9 17:33:28 2014, Steve, Update, VAC, Power outage II & recovery pumpingAgain.png


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Fri Oct 10 14:41:04 2014, Steve, Update, VAC, Power outage II & recovery recovery_poweroutage.png


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Mon Oct 13 14:44:52 2014, Steve, Update, VAC, RGA scan pd78 -day 13 RGAscan13Oct2014.png
 Our first RGA scan since May 27, 2014 elog10585
 The Rga is still warming up. It was turned on 3 days ago as we recovered from the second power outage.
 
    Reply  Mon Dec 1 17:13:57 2014, Steve, Update, VAC, RGA scan pd78 -day 63 RGAscanD63.png


    
        
            Quote:
        
        
            

            Our first RGA scan since May 27, 2014 elog10585
    Reply  Tue Dec 23 14:33:24 2014, Steve, Update, VAC, Chiara gets UPS prepUP.png


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Tue Dec 23 15:37:32 2014, Steve, Update, VAC, TP3 drypump replaced 


    
        
            Quote:
        
        
            

           
            
                
           
Entry  Fri Dec 26 17:45:21 2014, Steve, Update, VAC, vac pressure rose to 1.3 mTorr noN2.pngbacktoVacNormal.png
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things.
    Reply  Sun Dec 28 07:29:16 2014, Steve, Update, VAC, vacuum is normal 3d.png


    
        
            Quote:
        
        
            

           We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor
    Reply  Tue Feb 3 15:20:13 2015, Steve, Koji, Update, VAC, Pneumatic pressure is being read again N2pressureReadingIsBack.png
[Koji, Steve]

Summary

The N2 pressure reading (C1:VAC-N2PRES) is now up-to-date after rebooting c1vac1.
    Reply  Sun Mar 22 18:42:40 2015, Steve, Update, VAC, vac pressure rose to 1.2mTorr NoN2.pngPressureReadingWorks.png
 




Quote:



We run out of N2 for
    Reply  Mon Mar 23 10:36:55 2015, ericq, Update, VAC, Pressure watch script 
Based on Jenne's chiara disk usage monitoring script, I made a script that checks the N2 pressure, which will send an email to myself, Jenne, Rana,
Koji, and Steve, should the pressure fall below 60psi. I also updated the chiara disk checking script to work on the new Nodus setup. I tested the
two, only emailing myself, and they appear to work as expected. 
    Reply  Fri Mar 27 10:38:40 2015, steve, Update, VAC, vac rack UPS battery replaced UPSvac.jpg
Batteries replaced after 3.5 years with Amstron AP-1250F2,  8x 12V 6Ah




Quote:



APC Smart -UPS 2200  
    Reply  Tue Mar 31 09:03:32 2015, Steve, Update, VAC, RGA scan pd78 day 182 RGApd78d182.png
 

 
    Reply  Fri Apr 24 17:30:32 2015, Jenne, Update, VAC, Pressure watch script broken 



Quote:



I made a script that
checks the N2 pressure, which will send an email to myself, Jenne, Rana, Koji, and Steve, should the pressure fall below 60psi.
    Reply  Sat Apr 25 18:50:47 2015, ericq, Update, VAC, Pressure watch script broken 
Ugh, this turns out to be because cron doesn't source the controls bashrc that defines where to find caget and all that jazz that many commands depend
on. This is probably also why the AutoMX cron job isn't working either. 

Also, cron automatically emails everything from stderr to the email address that is configured for the user, which is why the n2 script blew
    Reply  Mon Apr 27 14:37:55 2015, Steve, Update, VAC, N2 pneumatic pressure watch  
 




Quote:



Based on Jenne's
    Reply  Mon Apr 27 21:42:07 2015, rana, Update, VAC, Summary pages 
We want to have a VAC page in the summaries, so Steve
- please put a list of important channel names for the vacuum system into the elog so that we can start monitoring for trouble.

Also, anyone that has any ideas can feel free to just add a comment to the summary pages DisQus comment section with the 40m shared account or
    Reply  Tue Apr 28 09:49:26 2015, Steve, Update, VAC, Vac Summery Channels 
 

 





      
    Reply  Thu Apr 30 16:30:25 2015, Steve, Update, VAC, N2 pneumatic pressure watch set up 
We have 2 transduser PX303-3KG5V   http://www.omega.com/pressure/pdf/PX303.pdf They 
will be installed on the out put of the N2 cylinders to read the supply pressure.

I will order one DC power supply  http://www.omega.com/pptst/PSU93_FPW15.html     
    Reply  Tue May 5 16:02:57 2015, Steve, Update, VAC, Vac Summery Channels with discription 
As it was requested by the Bos.

It would be nice to read from the epic screen C1:Vac-state_mon.......Current State: Vacuum Normal, valve configuration




    Reply  Fri May 8 14:30:09 2015, Steve, Update, VAC, CC1 cold cathode gauges are baked now cc1bake95C.jpg
CC1s  are not reading any longer. It is an attempt to clean them over the weekend at 85C

These brand new gauges "10421002" sn 11823-vertical, sn 11837 horizontal replaced 11 years old 421 http://nsei.missouri.edu/manuals/hps-mks/421%20Cold%20Cathode%20Ionization%20Guage.pdf 
on 09-06-2012 http://nodus.ligo.caltech.edu:8080/40m/7441
    Reply  Tue May 12 14:57:52 2015, Steve, Update, VAC, CC1 cold cathode gauges are baked now 
Baking both CC1 at 85 C for 60 hrs did not help.

The temperature is increased to 125 C and it is being repeated.