    Reply  Tue Sep 13 15:00:25 2011, Keiko, Update, LSC, 3f demodulation board check WF33.pdfWF165.pdf
Whitening filters for the REFL33 & 165 demodulated channels were measured and confirmed that they are working. They can be turned on and off by un-white
filter switches on the MEDM screen because they are properly linked. The measured filter responses are showen below. (Sorry, apparentyl the thumbnails
are not shown here. Please click the attachments.) 
    Reply  Tue Sep 13 23:08:51 2011, Keiko, Update, LSC, 3f demodulation board check REFL33-modified.pngREFL165-modified.png
Keiko, Jamie , Kiwamu
The I and Q orthogonalities of REFL33 and 165 demodulation board were measured by "orthogonality.py"
 Python package scipy were addied on Pianosa to run this code. Please note that
    Reply  Thu Sep 15 01:06:20 2011, Keiko, Update, LSC, 3f demodulation board check 
In addition to REFL 33 ans 165, I checked the orthogonality for the other existing three channels.
 
AS11
    Reply  Fri Sep 16 21:26:12 2011, Keiko, Update, LSC, 3f demodulation board check 
The demodulation phases and gains for the all existing channels, AS11, REFL11,REFL55, REFL165, and REFL33, were
adjusted by the command "ezcawrite" commands. 
Entry  Tue May 20 16:48:52 2014, ericq, Update, LSC, 3f Stability 3fs.pdfCARMsweepPrcFields.pdf
So, I really should have done this as soon as Manasa measured the arm lengths... I've
updated my MIST model with the real arm lengths, but still am using assumed identical losses of 75ppm on each mirror. (I've tried measuring the arm losses
for real, but got numbers in the hundreds of ppms, so I need to reexamine things...) 
    Reply  Tue May 20 22:42:28 2014, ericq, Update, LSC, 3f Stability PRCangles.pdf
 Here's the angles of MICH and PRCL from the my earlier plot by themselves; this shows that the individual demod angles in REFL165 aren't changing
much either. 
Entry  Wed Sep 14 01:19:20 2011, Anamaria, Configuration, LSC, 3f PD Install in Progress 
I have reconfigured the refl beam path on the AP table to include REFL33 and REFL165. Would be done if we hadn't prepared P BSs instead of S, which required
some serious digging to find two others. And if someone hadn't stolen our two 3m SMA cables that Keiko and I made on our previous visit and I had
left with the 3f PDs. I don't expect them to reappear but if they do, it would be grand.
    Reply  Tue Nov 11 21:41:09 2014, Koji, Update, LSC, 3f DRMI sensing mat DRMI_radar.pdf
Sensing matrix calculation using DTT + Matlab
Note: If the signal phase is, for example,  '47 deg', the phase rotation angle is -47deg in order to bring this signal to 'I' phase.
Note2: As I didn't have the DQ channels for the actuation, only the relative signs between the PDs are used to produce the radar chart.
    Reply  Wed Nov 12 03:22:23 2014, Jenne, Update, LSC, 3f DRMI sensing mat 8x
Koji pointed out something to me that I think he had told me ages ago, and Rana alluded to last night:  Since I'm not tuning my "demod phase"
for the sensing matrix lockins, unless I happened to get very lucky, I was throwing away most of the signal.  Lame.  
So, now the magnitude is sqrt(real^2 + imag^2), where real and imag here are the I and Q outputs of the lockin demodulator, after the 0.1Hz lowpass. 
Entry  Tue Nov 11 03:48:46 2014, Jenne, Update, LSC, 3f DRMI SensMatMeas_10Nov2014_DRMI.pngSensMatMeas_10Nov2014_PRMI.png
I was able to lock the DRMI on 3f signals this evening, although the loops are not stable, and I can hear oscillations
in the speakers.  I think the big key to making this work was the placement of the SHP-150 high pass filter at the REFL165 PD, and also
the installation of Koji's 110 MHz notch filter just before the amplifier, which is before the demod board for REFL165.  These were done to prevent
Entry  Wed Nov 5 01:58:54 2014, ericq, Update, LSC, 3F RFPD RF spectra Nov52014_3fPD_DRMIspectra.zipREFL165_DRMIspectrum.pngREFL33_DRMIspectrum.png
Given the checkout of the aLIGO BBPDs happening (aLOG link), wherein the
PDs were acting funny, and Koji has made some measurements determining that intermodulation/nonlinearity of circuitry can corrupt 3F signals, I've made
a similar measurement of the RF spectra of REFL165 when we're locked on DRMI using 1F signals. Maybe this could give us insight to our bad luck using REFL165...
    Reply  Wed Nov 5 11:09:44 2014, Koji, Update, LSC, 3F RFPD RF spectra 
If you look at the intermodulation at 14 (4+10) and 16 (6+10), 15 (5+10) would make any problem, thanks to the notch at 1f and 5f.
BUT, this absolute level of 165MHz is too tiny for the demodulator. From the level of the demodulated signal, I can say REFL165 has
too little SNR. We want to amplify it before the demodulator.
    Reply  Wed Nov 5 22:25:42 2014, ericq, Update, LSC, 3F RFPD RF spectra REFL165_demod_DRMIspectrum.png
 
Now that I have followed the chain, the PD signal is indeed being amplified at the LSC rack. It goes into a ZFL-1000LN+ amplifier (~23dB gain at
165MHz and 15V supply), followed by a SHP-100 high pass filter, and then enters the RF IN of the demod board. 
    Reply  Thu Nov 6 01:58:55 2014, Koji, Update, LSC, 3F RFPD RF spectra 
Where is the PD out spectrum measured with the coupler??? 
    Reply  Thu Nov 6 11:49:58 2014, ericq, Update, LSC, 3F RFPD RF spectra 


    
        
            Quote:
        
        
            

           Where is the PD out spectrum measured with the coupler???
    Reply  Thu Nov 6 14:41:49 2014, Koij, Update, LSC, 3F RFPD RF spectra 
That's not what I'm asking.
Also additional cables are left connected to the signal path. I removed it.
    Reply  Fri Nov 7 02:21:12 2014, ericq, Update, LSC, 3F RFPD RF spectra 
 
After some enlightening conversation with Koji, we figured that the RF amplifier in the REFL165 chain is probably being saturated (the amp's 1dB
compression is at +3dBm, has 23dB gain, and there are multiple lines above -20dBm coming out of the PD). I took a few more spectrum measurements to quantify
    Reply  Fri Nov 7 14:41:18 2014, ericq, Update, LSC, 3F RFPD RF spectra 165_ampSaturation.png165_HPatPD.png


    
        
            Quote:
        
        
            

            After some enlightening conversation with Koji, we figured that the RF amplifier in the REFL165 chain is probably being saturated.
    Reply  Fri Nov 7 16:15:53 2014, Jenne, Update, LSC, 3F RFPD RF spectra 
I have found an SHP-150, but no SHP-175's (also, several 200's, and a couple of 500's).
Why do you say the SHP-150 isn't enough?  The blue peak at 10*fmod in your plot looks like it's at -12 dBm.  -13 dB on top of that will
leave that peak at -25 dBm.  That should be enough to keep us from saturation, right?  It's not a lot of headroom, but we can give it a twirl
    Reply  Sat Nov 8 11:35:05 2014, rana, Update, LSC, 3F RFPD RF spectra 
 
 I think 'saturation' here is a misleading term to think about. In the RF amplifiers, there is always saturation. What we're
trying to minimize is the amount of distorted waveforms appearing at 3f and 15f from the other large peaks. Usually for saturation we are
    Reply  Mon Nov 10 18:11:57 2014, ericq, Update, LSC, 3F RFPD RF spectra 165signals.pdf
 Jenne and I measured the situation using a SHP-150 directly attached to the REFL165 RF output, and at first glance, the magnitude of the 165MHz
signal seems to not be distorted by the amplifier. 
 
    Reply  Thu Feb 12 03:43:54 2015, ericq, Update, LSC, 3F PRMI at zero ALS CARM Feb12_negativeTR.png
I have been able to recover the ability to sit at zero CARM offset while the PRMI is locked on RELF33 and CARM/DARM are on ALS, effectively indefinitely.
However, I feel like the transmon QPDs are not behaving ideally, because the reported arm powers freqently go negative as the interferometer is "buzzing"
through resonance, so I'm not sure how useful they'll be as normalizing signals for REFL11. I tried tweaking the DARM offset to help the buildup,
Entry  Tue Oct 20 16:35:24 2009, rob, Configuration, LSC, 33MHz Mod depth 
The 33MHz mod depth is now controlled by the OMC (C1:OMC-SPARE_DAC_CH_15).  The setting to give us the same modulation depth as before is 14000
(in the offset field).
Entry  Thu Jan 13 16:34:02 2011, josephb, Update, Computers, 32 bit matlab updated 
There was a problem with running the webview report generator in matlab on Mafalada.  It complained of not having
a spare report generator license to use, even though the report generator was working before and after on other machines such as Rosalba.  So I moved
the old 32 bit matlab directory from /cvs/cds/caltech/apps/Linux/matlab to /cvs/cds/caltech/apps/Linux/matlab_old.  I installed the latest R2010b
    Reply  Fri Sep 6 09:27:10 2013, Steve, Update, VAC, 31 days after pumpdown day31vac.png


    
        
            Quote:
        
        
            

            Valve configuration: Vacuum Normal
Entry  Wed Apr 1 15:47:48 2009, Yoichi, Update, Locking, 3.8kHz peak looks like a real optical response of the interferometer CARM-ETM-EXC.pngAOpath-TFs.png
Yoichi, Peter

To see where the 3.8kHz peak comes from, we locked the interferometer with the CARM fed back only to ETM and increased the arm power to 4.
Entry  Wed Apr 1 16:14:36 2009, Yoichi, Update, Locking, 3.8kHz peak does not change with SRC offset 
Yoichi, Peter

We suspected that maybe the 3.8kHz peak is the DARM RSE somehow coupled to the CARM.
Entry  Thu Mar 26 04:27:26 2009, Yoichi, Update, Locking, 3.8kHz peak as a function of the arm power CARM-PODC.pdf
During the power ramp-up, I actuated CARM using ETMs and measured the transfer functions to the PO_DC at several arm powers.
The peak grows rapidly with the power. It also seems like the frequency shifts slightly as the power goes up, but not much.
Entry  Thu Jun 26 22:08:33 2008, Jenne, Update, Electronics, 3.7MHz bandstop filter in MC Servo  MCLoopGainBoth.pngTP1ASpectrumBoth.pngQmonWithFilt.pngMCOut2WithFilt.png
The 3.7MHz elliptical bandstop filter that I made during my SURF summer is now installed in the MC servo loop to reduce the noise at 3.7MHz. 

I have taken transfer functions with and without the filter between TP1A and TP2A, with EXCA at -20dBm, using the HP4195A Network Analyzer.  I have also
Entry  Fri May 3 08:55:55 2013, Steve, Update, PEM, 3.2 M earthquake 3.2eqChannelIland.png3.2eq.png
 
Entry  Wed Mar 17 10:52:48 2021, Jordan, Update, SUS, 3" Ring Adpater for SOS  Side_View.pngFront_View.pngRing_with_Modifed_Bracket.pngBack_of_ring.pngFront_of_Ring.png
I have added a .1", 45deg chamfer to the bottom of the adapter ring. This was added for a new placement of the eq stops, since the barrel screws
are hard to access/adjust.

This also required a modification to the eq stop bracket, D960008-v2, with 1/4-20 screws angled at 45 deg to line up with the chamfer.
    Reply  Mon Jan 13 11:08:48 2014, Steve, Update, PSL, 3 good days of IOO pointing 3gdPSLpointing.png
 Three good days of IOO pointing: Friday, Sat and Sun    What was changed?  May be the clamping on Friday?
IOO vertical changes recovering as tempeture. IP is clipping at plastic enclosure of ETMY
 
    Reply  Mon Aug 29 15:15:46 2022, Tega, Update, Computers, 3 FEs from LLO got delivered today IMG_20220829_145533452.jpgIMG_20220829_144801365.jpg
[JC, Tega]

We got the 3 front-ends from LLO today. The contents of each box are:


FE machine
OSS adapter card for connecting to I/O chassis
PCI riser cards (x2)
Timing Card and cable
Power
    Reply  Tue Aug 30 15:21:27 2022, Tega, Update, Computers, 3 FEs from LHO got delivered today IMG_20220830_144925325.jpgIMG_20220830_142307495.jpgIMG_20220830_143059443.jpg
[Tega, JC]

We received the remaining 3 front-ends from LHO today. They each have a timing card and an OSS host adapter card installed. We also receive 3
dolphin DX cards. As with the previous packages from LLO, each box contains a rack mounting kit for the supermicro machine.
    Reply  Mon Nov 21 16:08:04 2011, kiwamu, Update, Green Locking, 2nd trial of Y arm ALS noise budget : broad band noise gone Yarm_ALS_2011Nov19.png


    
        
            Quote from #5930
        
    
   
            
            Right now the fluctuation
Entry  Mon Aug 10 14:58:01 2009, Jenne, Update, PEM, 2nd set of Guralp channels plugged into ADCU 
The second set of Guralp channels is now plugged into the PEM ADCU, into channels which are confirmed to be working.  (Method: 1Vpp sine wave into
channel, check with DataViewer).
 
    Reply  Mon Aug 10 18:12:25 2009, Jenne, Update, PEM, 2nd set of Guralp channels plugged into ADCU 


    
        
            Quote:
        
        
            
            The second set of Guralp channels is now plugged into the PEM ADCU, into channels which are confirmed to be working. 
Entry  Fri Oct 8 07:41:42 2010, steve, Configuration, SAFETY, 2W laser shutter is closed 
The 2 W Innolight shutter is closed and enclosure door removed. Beam path blocked. Do not change this condition.
    Reply  Fri Oct 8 10:38:03 2010, steve, Configuration, SAFETY, 2W laser shutter is closed P1060905.JPGP1060910.JPG


    
        
            Quote:
        
        
            
            The 2 W Innolight shutter is closed and enclosure door removed. Beam path blocked. Do not change this condition.
Entry  Wed May 5 16:58:21 2010, Koji, Update, PSL, 2W hooked up to the interlock service 
Ben, Steve, and Koji
Ben came to the 40m and hooked up a cable to the main interlock service.
We have tested the interlock and confirmed it's working.
Entry  Sat Apr 24 15:05:41 2010, Kevin, Update, PSL, 2W Vertical Beam Profile vbp.jpg
The vertical beam profile of the Innolight 2W laser was measured at eight points along the axis of the laser.
These measurements were made with the laser crystal temperature at 25.04°C and the injection current at 2.091A. z is the distance from the
razor blade to the flat black face of the front of the laser.
Entry  Wed Jun 2 22:58:04 2010, Kevin, Update, PSL, 2W Second Reflected Beam Profile profile_2nd.png
[Koji, Kevin]
The profile of the Innolight 2W was previously measured by measuring the reflected beam from the front surface of a W2 window (see
entry). To investigate thermal effects, Rana suggested also measuring the profile of the beam reflected from the back surface of the W2.
Entry  Thu Oct 14 23:33:45 2010, Suresh, Update, IOO, 2W NPRO laser output power versus temperature tempscan.pngScan2010.zip
Steve measured an apparent power drop in the 2W NPRO output from 2.1W to 1.6W(elog entry no 3698)
at 2.1A of diode current in the laser (elog entry:  2822).  It was later noticed that the laser temperature was set to about 45 degC while
    Reply  Fri Oct 15 00:15:52 2010, Koji, Update, IOO, 2W NPRO laser output power versus temperature 
From the plot, you observed the reduction of the output power only by 1% between 25deg to 45deg.
This does not agree with the reduction from 2.1W to 1.6W.
Is there any cause of this discrepancy?
Entry  Fri Oct 15 21:25:23 2010, Suresh, Update, IOO, 2W NPRO laser output power drop question 
  The power meter used in the measurements of elog entries 2822, 3698 and 3709 was the Ophir PD300-3W.  This power head has several damaged
patches  and a slight movement of the laser spot changes the reading considerably.  To verify I checked the power out with another power meter
(the Vector S310) and found that there is no significant variation of the power output with the temperature of the laser.  And the power at 2.1A of
Entry  Mon Aug 30 23:21:36 2010, rana, HowTo, PSL, 2W NPRO Mount designed with emachineshop.com software LaserMount.png
To test out this website - emachineshop.com, Jenne and I are designing some of the mounts for the new beam height.
    Reply  Wed Sep 1 08:36:59 2010, Koji, HowTo, PSL, 2W NPRO Mount designed with emachineshop.com software 
1. I can not see whether the attaching surface is flat or not.
It should have ~1mm step to avoid "the legs" of the laser at the four corners.
Otherwise we will have ~0.5mm space between the block and the laser 
Entry  Thu Jun 3 00:47:17 2010, Kevin, Update, PSL, 2W Beam Profile of Second Reflected Beam profile_2nd.png
[Koji, Kevin]

The profile of the Innolight 2W was previously measured by measuring the reflected beam from the front surface of a W2 window (see
Entry  Wed Jun 2 03:24:22 2010, Kevin, Update, PSL, 2W Beam Profile profile.pngerrors.pngLayout.jpg
[Rana, Kiwamu, Kevin]
The Innolight 2W beam profile was measured with the beam scan. A W2-IF-1025-C-1064-45P window was used to reflect a small amount of the main beam.
A 5101 VIS mirror was used to direct just the beam reflected from the front surface of the W2 down the table (the beam reflected from the back surface
    Reply  Wed Jun 2 04:27:02 2010, Koji, Update, PSL, 2W Beam Profile 
This is what I already told to Kevin and Rana:
A direct output beam is one of the most difficult measurements for the mode profiling.
I worried about the thermal lensing.