40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 335 of 349  Not logged in ELOG logo
ID Date Author Type Categoryup Subject Text Attachments
  12337   Tue Jul 26 14:24:38 2016 KojiSummaryVACPurge compressed air system at LHOI've visited the purge clean
air system at LHO Yarm mid-station with John
Worden.
 12x 
  12338   Tue Jul 26 16:01:32 2016 SteveSummaryVACPurge compressed air systemThanks for checking this out Koji

The builder in 1996 was Process
System International, Inc ( Westborough,
 cleanAirSupl.pdfmobileHEPA.jpg 
  12339   Tue Jul 26 17:41:59 2016 KojiSummaryVACPurge compressed air systemWe have no number for the CFM without calculation.
We can't assume a random number like
10-15
  
  12483   Mon Sep 12 17:54:24 2016 SteveUpdateVACpumpdown 79 We are pumping down. The annuloses
are below 10 mTorr

The vacuum gauges are not communicating
  
  12487   Tue Sep 13 16:16:28 2016 ChrisOmnistructureVACvacuum interlock bypassed(Steve, Chris)

The pumpdown had stalled because
of some ancient vacuum interlock code that
  
  12488   Tue Sep 13 17:47:04 2016 SteveUpdateVACpumpdown 79 completedP1 IFO pressure is 1 mTorr, valve configuration: vacuum normal,
annulosses are pumped, RGA is off, not pumped.
  
  12492   Wed Sep 14 09:44:09 2016 SteveUpdateVACpumpdown 79 completedThe pumpdown started at
9-12-2016

The IFO is at 5e-5 Torr vacuum
 pd79.pngpd79d2.jpgCCcg.jpg 
  12496   Thu Sep 15 10:36:32 2016 SteveUpdateVACpumpdown 79 day 3IFO pressure 3.7e-5 Torr at new cold cathode
InstruTech - Hornet

New
 vent78for73d.png 
  12525   Fri Sep 30 10:37:57 2016 SteveUpdateVACvacuum VME machines rebooted[ Gautam and Steve ]

c1vac1 and c1vac2 were rebooted
and the gauges are communicating now. V1,
 rebooted.png 
  12530   Tue Oct 4 11:02:31 2016 SteveUpdateVACRGA is out of orderThe last good rga scan at vent 78 
day 38

 lastBGscan.png 
  12537   Fri Oct 7 10:29:57 2016 SteveUpdateVACRGA removedThe RGA is removed for repaire. It's
volume at atmophere and sealed.. P4 reading
of 38 Torr is not correct.
 RGAremoved.png 
  12556   Thu Oct 13 00:23:54 2016 ericqUpdateVACPre-vent prepI have completed the following non-Steve
portions of the pre-vent checklist [wiki-40m.ligo.caltech.edu]


Center all oplevs, transom
  
  12559   Thu Oct 13 16:48:34 2016 SteveUpdateVACvent 79 is completedIFO is at atmosphere. The MC can be locked
in air now.

The doors will be coming off tomorrow
 vent_79.png 
  12574   Thu Oct 20 09:17:08 2016 SteveUpdateVACPumpdown 80 has startedPumping again after 7 days at atmosphere.

BS, ITMY and OMC chambers were
open only.
 readyPd80.pngbeforePd80.pngpd80at1h.png7d_atm.png 
  12575   Thu Oct 20 15:59:52 2016 SteveUpdateVACPumpdown 80 has completedIFO pressure 1.7E-4 Torr on new not logged
cold cathode gauge. P1 <7E-4 Torr

 7hPd80.png8hPd80.png9hPd80.png 
  12577   Fri Oct 21 09:28:21 2016 SteveUpdateVACVac Normal reachedDry pump of TP3 replaced after 9.5 months
of operation.[ 45 mTorr d3 ]

The annulosses are pumped.
 VacNormal.png 
  12589   Mon Oct 31 16:20:50 2016 SteveUpdateVACRGA is reinstalledQuad rods and  ionizer kit: consisting
of repeller cage, anode grid, focus plate
and  filament were replaced....... under
  
  12633   Tue Nov 22 11:31:52 2016 SteveUpdateVACRGA is running againThe Rga was turned on yesterday.




 RGAisBack33d.pngpd80-560Hz-d33.pngafterRepair1d.png 
  12636   Wed Nov 23 11:32:13 2016 SteveUpdateVACTP3 drypump replaced againTP3 drypump replaced at 655 mTorr, no load,
tp3 0.3A 

This seal lasted only for 33 days
  
  12642   Mon Nov 28 09:45:44 2016 SteveUpdateVACRGA scan at pumpdown day 40The vacuum envelope ~1 C warmer today than
7 days ago. Bad peaks are coming down as
normal.
 RGAscan_d40.pngd2_&_d7scan.pngRGAd2.png 
  12681   Thu Dec 22 09:37:20 2016 SteveUpdateVACRGA scan at day 63Valve configuration: vacuum normal

RGA head temp: 43.5 C

Vac envelope temp: 23 C
 pd80-d63.pngpd80-580Hz-d63.png 
  12698   Tue Jan 10 14:24:09 2017 SteveUpdateVACRGA scan at day 82Valve configuration: vacuum normal

Vacuum envelope: 23C

Rga head: 44C
 pd80VNd82.png 
  12774   Tue Jan 31 14:14:29 2017 ranaUpdateVACPressure watch scriptI think this cron job is running on NODUS
(our gateway) instead of our scripts machine:


p.p1 {margin:
  
  12853   Mon Feb 27 15:33:10 2017 SteveUpdateVACRGA scan at day 130Valve configuration: vacuum normal

Vacuum envelope: 23.5 C

RGA head: 46.6 C
 rgascan@130d.png 
  13047   Wed Jun 7 11:32:56 2017 SteveUpdateVACsmooth vac rebootGautam and Steve,

 

The medm monitor & vac control
 vac_reboot.png 
  13075   Tue Jun 20 16:28:23 2017 SteveUpdateVACRGA scan   RGAscan243d.pngRGAscan.png 
  13132   Sun Jul 23 15:00:28 2017 JamieOmnistructureVACstrange sound around X arm vacuum pumpsWhile walking down to the X end to reset
c1iscex I heard what I would call a "rythmic
squnching" sound coming from under the
  
  13134   Mon Jul 24 09:55:29 2017 SteveOmnistructureVACcontroller failerIfo pressure was 5.5 mTorr this Monday
morning. The PSL shutter was still open.
TP2 controller failed. Interlock closed V1,
 TP2controllerFails.png 
  13142   Tue Jul 25 08:48:57 2017 SteveUpdateVACRGA scan at d278The RGA did not shut down at the turbo
pump controller failing.

 RGA278d.png 
  13143   Tue Jul 25 14:04:06 2017 SteveUpdateVACturbo controller installed and we are running at vac normalGautam and Steve,

Spare Varian turbo-V 70 controller,
Model 969-9505, sn 21612 was swapped in.
 TP2@75Krpm.png 
  13171   Tue Aug 8 17:04:26 2017 SteveUpdateVACunintended pump downIFO pressure 2 Torr,    PSL 
shutter closed.  I'm pumping down
with 2 roughing pumps with ion pump gate
 pumping_down.png 
  13172   Tue Aug 8 17:44:11 2017 SteveUpdateVACunintended pump downPumpdown stopped for over night  at
~  1 Torr

The roughing line disconnected. 
 stopped_pumping.png 
  13175   Wed Aug 9 11:59:51 2017 SteveUpdateVACunintended pump down at vacuum normalpd80b has reached Vac Normal. IFO pressure
0.5 mTorr

We need our vauum channels back
 pdc.pngpd80b@vacnormal.png 
  13179   Wed Aug 9 16:34:46 2017 ranaUpdateVACVacuum Document recoveredSteve and I found the previous draft of
the 40m Vacuum Document. Someone in 2015
had browsed into the Docs history and then
  
  13186   Thu Aug 10 15:45:34 2017 SteveUpdateVACaccidental vent to 17 TorrFinally we got the cold cathode gauge working.
IFO pressure 7e-6 Torr-it, vacuum normal
valve configuration with all 4 ion pump gate
 Mag_UPS.jpgUPSbatteryPack.jpgclosing_V1.jpg 
  13233   Mon Aug 21 14:53:32 2017 gautamUpdateVACRGA reset[gautam, steve]

In the aftermath of the accidental
vent, it looks like the RGA was shutdown.
  
  13234   Mon Aug 21 16:35:48 2017 gautamUpdateVACUPS checkup[steve, gautam]

At Rolf/Rich Abbott's request,
we performed a check of the UPS today.
  
  13245   Wed Aug 23 10:11:46 2017 SteveUpdateVACvacuum valve specificationsThe V1
gate valve specs installed at 40m wiki
page. VAT model number 10846-UE44-0007 Our
  
  13260   Mon Aug 28 10:28:07 2017 SteveUpdateVACRGA scan 20 day after 17 TorrThe RGA was turned on 7 days ago. 
It's 46 C now.   The X-arm room
tem ~20 C
 20d_after_17_torr.png 
  13268   Tue Aug 29 15:35:19 2017 SteveUpdateVACvacuum pump specifications & manualsRP1 and RP3 roughing pump manual of
Leybold D30A oily rotory pump

Fore pump of
 1759489_995814a9-55c1-4fa1-b333-f446d272c1cc.jpg 
  13321   Tue Sep 19 11:05:26 2017 SteveUpdateVACRGA scan at day 334  

 
 RGAscan334d.png 
  13403   Fri Oct 27 10:14:11 2017 SteveUpdateVACRGA scan at day 372Valve configuration: Vacuum normal

Note: Tp2 running at 75Krpm 0.25A
26C has a  load high pitch sound today.
 RGA_scan_d372.png 
  13407   Mon Oct 30 10:09:41 2017 SteveUpdateVACTP2 failed IFO pressure 1.2e-5 Torr at 9:30am




 tp2failed.pngifo_1.0E-5_Torrit.pngtp2failed2dago.png4days.png 
  13409   Mon Nov 6 09:09:43 2017 SteveUpdateVACsetting up new TP2 turboOur new Agilent Technology TwisTorr
84FS AG rack controller ( English Manual
pages 195-297 )  RS232/485, product
  
  13423   Fri Nov 10 08:52:21 2017 SteveUpdateVACTP3 drypump replacedPSL shutter closed at 6e-6 Torr-it    

   The foreline pressure
of the drypump is 850 mTorr at 8,446 hrs
  
  13454   Sun Nov 26 19:38:40 2017 SteveUpdateVACTP3 drypump replaced againThe TP3 foreline pressure was 4.8 Torr,
50K rpm 0.54A and 31C........Maglev rotation
normal 560 Hz.......    IFO pressure
 after_replacement.png 
  13459   Thu Nov 30 10:38:39 2017 SteveUpdateVACannuloses are not pumpedAnnuloses are not pumped for 30 days, since
TP2 failed.  IFO pressure 7e-6 Torr
it,  Rga 2.6e-6 Torr
 AnnulosesNotPumped.png 
  13495   Tue Jan 2 15:43:35 2018 SteveUpdateVACpumpdown after power outage 




 pumpdown_from_44_mTorr.png 
  13588   Tue Jan 30 10:22:20 2018 SteveUpdateVACRGA scan at day 175pd80b rga scan at 175 day.  IFO pressure
7.3e-6 Torr-it

Condition: vacuum normal, annuloses
 RGApd80-b-468-175d.pngRgaOn20hrs.png 
  13615   Wed Feb 7 15:50:42 2018 SteveUpdateVACIFO pressure monitoringHornet cold cathode gauge analoge output 
[ DSub9 pin 3 and 7 ] are wired to go ETMX
Acromag. It was reading 4.9V at 7.8e-6 Torr
 CC_analog.png 
ELOG V3.1.3-