  318   Thu Feb 14 17:21:53 2008 Max JonesUpdateComputersNoise budget code changes
In cvs/cds/caltech/NB/matlab/utilities/LSCmodel.m at line 146
I have hardwired in changes to struct lsc. Please see code.
  317   Thu Feb 14 15:05:18 2008 robDAQDMFseisBLRMS 1.1
> 
> Also, for reasons of aesthetics, I changed the data delay from 6 minutes to 5 minutes.  We'll see if that's enough.

5 minutes didn't work.  
  316   Thu Feb 14 15:04:53 2008 robDAQDMFDMF delay

Sometime ago I edited seisBLRMS to keep of track of how long it was taking to write RMS data (that is, the delay between the accelerometer data and the write of the EPICS rms data). Here's a plot of that info, showing how the delay increases over time. I think this indicates a logical flaw in the timing of the seisBLRMS program, which sort of relies on everything running well consistently; this should not be difficult to fix. I'll maybe try increasing the delay to ~10 minutes, and making it relatively inflexible.
  315   Wed Feb 13 20:37:11 2008 JohnUpdateLSCFibre locking - Fiber
Sam and I observed fringes in the light reflected from the Y arm. These fringes are due to the sidebands and not the carrier. To improve matters we plan to reduce the RF AM and increase our modulation index.
  314   Wed Feb 13 11:41:00 2008 AlbertoUpdateElectronicsSome characterization of the RF Monitor Box (StocMon)
I'm attaching a table with some measurements and the power spectrum from the pd to help evaluate the numbers.

The box output ranges from 0.5V to 2.1V. The coefficient between power and voltage is negative so higher voltage means lower power.

The red numbers are the outputs from each channel at their resonant frequencies. As one can see these are not very well centered on the dynamic range of the power detectors.

The cross coupling seems to be not a problem.

Even if the 166 filter, which handles the smallest of the frequencies and is also the most lossy (for construction reason), mounts a preamplifier, the output is still rather small. this explain also the high bias due to the noise amplification at the maximum power (13dB). A better insertion loss either remaking the filter or re-tuning that one would simplify many problems, i.e. there is not much room in the metal pomona box to fit the amplifier. I might want to consider, after everything else is ready and if I have time before leaving next week, to work on a new 166 filter.
  313   Tue Feb 12 16:39:52 2008 robUpdateLockingreport

Did some locking work on DRFPMI on sunday and (with John) on monday nights. So far progress has not been terribly encouraging.

Problems include the DD_handoffs not working and the CARM->MCL handoff not working so well. To get around the DD signals trouble, I decided for now to just ignore 67% of the DD signals. We should be able to run with PRC & MICH on single demod signals, and SRC on a DD signal. This seems to work well in a DRMI state, and it also works well in a DRMI+2ARMs state.

The CARM->MCL handoff actually works, but it doesn't take kindly to the AO path and it doesn't work very stably. I guess this was always the most fragile part of the whole locking procedure, and it's fragility is really coming to light now. Investigation continues.
  312   Tue Feb 12 16:34:07 2008 robDAQDMFseisBLRMS 1.1
The compiled version of seisBLRMS had been running ~2 weeks without crashing as of last night, when I killed it
so it wouldn't interfere with alignment scripts.  I added an EPICS channel C1:DMF-ENABLE, and updated the DMF
executables to check this channel while running.  So far it seems to work.  When you're running alignment acripts,
simply click the DISABLE button on the C1DMF_MASTER.adl screen, and then re-ENABLE when the scripts finish. 
It's not clear why this is necessary though.  Theories include the constant disk access is keeping the
framebuilder busy, reducing its ability to deal with ezcademod commands and DMF programs just flooding the
network with so much traffic that ezcademod-related packets run late and get ignored.

Also, for reasons of aesthetics, I changed the data delay from 6 minutes to 5 minutes.  We'll see if that's enough.
  311   Tue Feb 12 16:18:29 2008 robConfigurationComputer Scripts / Programsautoburt cron moved to op340m

I moved the autoburt cron job from op440m to op340m. For some reason, burtrb requires gcc to run (I gather it uses the C-preprocessor to parse input files), so I had to install that on op340m to get it to work properly.

There are no more cron jobs running on op440m now.
  310   Tue Feb 12 13:53:27 2008 robOmnistructureVACReturn of the RGA

The new RGA head was installed a few days ago. I just ran the RGAlogger script to see if it works, which it does. I also edited the crontab file on op340m to run the RGAlogger script every night at 1:25 AM. It should run tonight.
  309   Mon Feb 11 22:44:29 2008 robConfigurationDAQChange in channel trending on C1SUS2

I removed the MC2 optical lever related channels from trends, and the SRM POUT and YOUT (as these are redundant if we're also trending PERROR and YERROR). I did this because the c1susvme2 processor was having bursts of un-syncy lateness every ~15 seconds or so, and I suspected this might interfere with locking activities. This behaviour appears to have been happening for a month or so, and has been getting steadily worse. Rebooting did not fix the issue, but it appears that removing some trends has actually helped. Attached is a 50 day trend of the c1susvme2 sync-fe monitor.
  308   Mon Feb 11 14:24:19 2008 steveUpdatePEMmore earthquakes
ITMX and ITMY sus damping restored after Baja earthquake 5.1 mag at 10:29 this morning.

The ground preparation for The ITS building is almost finished.
Activity is winding down, however the Baja California_ Mexico earthquake zone
"Guadala Victoria" started acting up on Friday.

http://earthquake.usgs.gov/eqcenter/recenteqsus/Maps/special/California_Nevada_eqs.php
  307   Sun Feb 10 21:43:16 2008 robConfigurationIOOMC alignment tweaked

I adjusted the alignment of the free hanging mode cleaner to best transmit the PSL beam.
  306   Sun Feb 10 20:47:01 2008 AlanSummarySUSAll watchdogs tripped
A moderate earthquake occurred at 11:12:06 PM (PST) on Friday, February 8, 2008.
The magnitude 5.1 event occurred 21 km (13 miles) NW of Guadalupe Victoria, Baja California, Mexico.
http://quake.wr.usgs.gov/recenteqs/Quakes/ci14346868.html
  305   Sat Feb 9 13:32:07 2008 JohnSummarySUSAll watchdogs tripped
When I arrived this afternoon the watchdogs had tripped on all optics. I reset them and enabled the coil currents.

I had to adjust the alignment of the mode cleaner to get it to lock.
  304   Sat Feb 9 13:05:48 2008 JohnSummaryIOOPMC camera/ HEPA
I replaced the Gig-E camera on the PMC trans beam. The PZT was close to railing and I wanted to adjust it. I just did a quick job, there is a little scattered light on the image. If Joe is finished with the Gig-E I'll take another look at it.

The HEPA in the PSL table was turned off for some reason. I turned it back on.
  303   Fri Feb 8 17:55:53 2008 joshConfigurationPEMPEM-AS_MIC now in PSL enclosure
I have moved the microphone to the PSL enclosure, hanging near the south (Y) side from a support rod for the overhanging storage area so that it's reasonably close to the PMC. I've fastened it in many places using cable ties to make sure that it won't fall.

Alberto helped me solder together a female BNC-female 3.5 mm stereo adapter so that I can use the DAQ to output through BNC to PC speakers. My plan is do sweep sine output through PC speakers to find the transfer function of sound from outside the enclosure to inside the enclosure and by moving the microphone more centrally over the PSL table, check if there are any strong resonances. Hopefully I can use this technique at other places around the interferometer or measure the effect of installing acoustic foam.
  302   Fri Feb 8 17:09:52 2008 Max JonesUpdateComputersChanges to NoiseBudget
Today I altered the following files

caltech/NB/matlab/utilities/get_dtt_dataset
In DARM_CTRL case I changed the second channel name to DARM_ERR. Messy but it may be effective.

caltech/NB/matlab/noise/getUGF.m
I commented out lines of code specifically pertaining to non-existent DARM_DAQ channel. Marked in
code around approximately line 60.

Please address all comments or concerns to me (williamj(at symbl)caltech.edu). Thank you.
  301   Wed Feb 6 19:39:11 2008 ranaConfigurationCamerasRegions of Interest and max frame rate
We really need to look into making the 40m CDS network have an all GigE backbone so that we can have cooler cameras as well as collect multiple datastreams...
  300   Wed Feb 6 16:50:47 2008 josephbConfigurationCamerasRegions of Interest and max frame rate
The Snap code has once again been modified such that setting the -l option to 0 will take images as fast as possible. Also, the -H and -W options set the height and width, while in principle the -Y and -X options set the position in pixels of the top edge and left edge of the image. It also seems possible to set these values such that the saved image wraps around. I'll be adding some command checking so that the user can't do this in the near future.

Doing some timed runs, using a -H 350 and -W 350 (as opposed to the full 752x480), 100 images can be saved in roughly 8 seconds, and 1000 images took about 73 seconds. This corresponds to a frame rate of about 12-13 frames per second (or a 12-13 Hz display). The size of this area was sufficient to cover the current PMC transmission beam.

The command line I used was

time ./Snap -l 0 -m 1000 -f 'test' -W 350 -H 350 -Y 50 -X 350 -E 2000

Interestingly enough, there would be bursts of failed frame saves if I executed commands in another terminal (such as using ls on the directory where the files were being stored).

As always, this code is available in /cvs/cds/caltech/target/Prosilica/40mCode/.
  299   Wed Feb 6 09:17:31 2008 steveUpdatePEMIST building construction continoues
The bulldosers at work
  298   Tue Feb 5 17:39:05 2008 jweinerConfigurationPEMPEM-AS_MIC taken down from AS table, will put in PSL enclosure soon
I took down the microphone that Andrey hung above the AS table his first week in lab. I want to hang the microphone above the PMC to check the effect of acoustic noise on the PMC. The cables were a little more tangled than I thought so I've only taken the microphone down and haven't hung it back up yet, but on Thursday I'll have enough time to carefully put it up inside the PSL and see what I can find out about acoustic noise inside the PSL. I think the microphone should be sensitive enough for the frequencies we're interested in, and I'll hopefully find out if it's sufficient once I put it up in the PSL. The microphone cable and microphone are on top of the PSL for now.
  297   Tue Feb 5 15:32:29 2008 josephbConfigurationCamerasPMC and the GigE Camera
The PMC transmission video camera has been removed and replaced with the GigE GC750 camera for the moment.

A ND4.0 filter has been added in the path to that camera to reduce saturation for the moment.

The old camera has been placed on the elevated section inside the enclosure, and the cable for it is still on the table proper.

The Gige camera is currently running the Snap code on Linux3 with the following command line:

./Snap -E 2000 -l 60 -m 1440 -f './pmc_trans/pmc_trans'

So its going to be taking tiff images every minute for the next 24 hours into the cvs/cds/caltech/target/Prosilica/40mCode/pmc_trans/ directory.

Attached is an example image with exposure set to 2000, loaded into matlab and plotted with the surf command. 2500 microseconds looked like it was still saturating, but this seems to be a good level (with a max of 58560 out of 65535).
  296   Mon Feb 4 22:01:57 2008 JohnSummaryLSCFibres auxiliary locking - Fibers
I managed to couple ~75% of the light transmitted from the y arm, through the fibre, back to the SP table. I hoped that this would be a good way to match the beam from the fibre into the arm. Still no flashes. It looks like the cameras just aren't sensitive enough.
  295   Sun Feb 3 05:02:41 2008 ranaUpdatePEMSeism 4 day
  294   Sat Feb 2 14:11:27 2008 JohnSummaryComputersOPTLEVmaster screen
I changed the layout of the optlev master screen. The old version is /cvs/cds/caltech/medm/old/C1ASC_OPTLEVmaster080202.adl
  293   Fri Feb 1 17:17:05 2008 JohnSummaryPSLMach-Zender tweaking
I helped Rob adjust the alignment of the Mach-Zender to try and reduce any AM on the laser light. Our goal was to reduce the large offsets in the DD signals.

We reduced the MZ refl from 0.54 to 0.39. We were able to re-lock the mode cleaner without problems. We then centred the WFS heads.

No change in the DD signal offsets.
  292   Fri Feb 1 15:04:54 2008 josephbConfigurationCamerasSnap with looping functionality available
New GiGE camera code is available in /cvs/cds/caltech/target/Prosilica/40mCode/. Currently only runs on Mafalda.

Snap has expanded functionality to continuously loop infinitely or for a maximum number of images set by the user. File names generated with the loop option have the current Unix time and .tiff appended to them. So -f './test' will produce tiff files with format "test1234567.tiff". The -l option sets the number of seconds between images.

"./Snap -l 5 -i -f './test' " will cause the program to infinitely loop, saving images every 5 seconds. Using "-m 10" instead of "-i" will take a series of 10 images every 5 seconds (so taking a total of 50 seconds to run).

It also now defaults to 16-bit (in reality only 10 bit) output instead of 8 bit output. You can select between the two with -F 'Mono8' or -F 'Mono16'.

Use --help for a full list of options.

Note that if you ctrl-c out of the loop, you may need to run ./ResetCamera 131.215.113.104 (or whatever the IP is - use ./ListCameras to determine IP if necessary) in order to reset the camera because it doesn't close out elegantly at the moment.
  291   Fri Feb 1 12:37:39 2008 robUpdateDMFseisBLRMS trends

Here are DV trends of the output of seisBLRMS over the last ~36 hours (which is how long it's been running), and another of the last 2 hours (which show the construction crew taking what appears to be a lunch break).
  290   Fri Feb 1 10:43:05 2008 JohnUpdateEnvironmentConstruction work
The boys next door have some bigger noisier toys.
  289   Thu Jan 31 16:53:41 2008 josephbConfigurationCamerasImproving camera user interface
There's a new and improved version of Snap program at the moment people are free to play with.

Located in /cvs/cds/caltech/target/Prosilica/40mCode/

The program Snap now has a -h or --help option which describes some basic command line arguments. The height (in pixels), width (in pixels), exposure time (in micro seconds), file name to be saved to (in .tiff format), and packet size can all be set. The format type (i.e. pixel format such as Mono8 or Mono16) doesn't work at the moment.

At the moment, it only runs on mafalda.

Currently in the process of adding a loop option which will take images every X seconds, saving them to a given file name and then appending the time of capture to the file name.

After that need to add the ability to identify and choose the camera you want (as opposed to the first one it finds).

Lastly, I've been finding on occassion that the frame fails to save. However if you try again a few seconds later with the exact same parameters, it generally does save the second time. Not sure whats causing this, whether on the camera or network side of things.

I've attached two images, the first at default exposure time (15,000 microseconds) and the second at 1/5th that time (3,000 microseconds).
The command line used was "./Snap -E 3000 -F 'Camera_exp_3000.tiff' "
  288   Thu Jan 31 12:39:14 2008 JohnConfigurationGeneralY arm test mass cameras
I've adjusted the test mass cameras on the y arm to make the beam injected through ETMY more visible.
  287   Wed Jan 30 20:39:31 2008 robUpdateDMFseisBLRMS


In order to reduce the probability of seisBLRMS crashing due to unavailability of data, I edited seisBLRMS.m so that it displays data from 6 minutes in the past, rather than 3. After compiling, this version ran for ~8 hours without crashing. I've killed the process now because it seems to interfere with alignment scripts that use ezcademod, causing "DATA RECEIVING ERROR 4608" messages. These don't cause ezcademod to crash, but so many of them are spit out that the scripts don't work very well. I guess running DMF constantly is just making the framebuilder work too hard with disk access. In the near term, we can maybe work around this by having DMF programs check the AutoDithering bit of the IFO state vector, and just not try to get data when we're running these sorts of scripts.
  286   Wed Jan 30 13:09:55 2008 AndreyUpdateSUSNew results for XARM (pdf)

See attachments: pdf-presentation with plots in "true" axes Q_ETMX and Q_ITMX, and seismic backgound measurement.

Results that were shown a week ago turned out to be not sad at all!
  285   Wed Jan 30 11:49:30 2008 AlbertoSummaryElectronicsRF monitor's filters final schematics and transfer functions
These are the final schematics for the 6th order Chebyshev filters of the RF monitor board. I'm also attaching the TF as I measured. The tuning is probably not optimal, less insertion noise could be achieved.
  284   Tue Jan 29 14:56:39 2008 robUpdateDMFseisBLRMS 1.0

The seisBLRMS 1.0 program crashed at ~7:20 pm last night, so we didn't get data from overnight. It crashed when framecaching failed. I added
a try-catch-end statement around the call to dttfft2 to let the program survive this, then compiled it and started it on mafalda. After ~45 minutes, the compiled version encountered the same error, and while it didn't crash per se, after 20 minutes it still wasn't able to read data. We may have to dig deeper into the guts of mDV to make this stuff run more robustly.
  283   Mon Jan 28 19:35:55 2008 ranaSummaryPEMAccelerometer and Seismometer Coherences
The attached PDF shows that there is some strange behavior at low frequencies.

From the plot it looks like to me that the Wilcoxon accelerometers (which are supposed to have good response down to 0.05 Hz) are not displaying real seismic motion below 0.3 Hz. Because the coherence length for seismic waves at those frequencies should be 100's of meters we should expect that the accelerometers would have good coherence (>0.8) down there. Instead, my guess is that its all air currents, temperature, or electronics noise. These sensors are not reliable indicators for the microseism.

The Ranger seismometer, however, seems to work fine down to just below the microseism. The Ranger is mounted down around the X end and pointing in the z-direction. The coherence I plotted between it and EX_Z is larger than any other acc/seis pair (as expected).

JM and I discussed what could be done; if we get a SURF student who's into building stuff we can ask them to make a styrofoam hut for the Wilcoxons to see if that helps anything. JM also asked what the point of all this is.

IF we want to do good Adaptive Noise subtraction then we need sensors which can sense the motion which disturbs the mirrors and they need to sense it with a good SNR to get a good subtraction ratio. If the styrofoam thing doesn't work, we should probably look into getting a Guralp 3-axis seismometer for the corner area and just move the accelerometers down to the ends. The sites have Guralp CMG-40T units (~ 8k$). I think we should check out the CMG-3T or the CMG-3ESP.

Does anyone know someone in the Geo depts that we can borrow one from?
  282   Mon Jan 28 18:56:47 2008 ranaUpdateDMFseisBLRMS 1.0
I made all of the updates I aludded to before:

  • Expanded the dmf.db file on c1aux to include all accelerometers and the seismometer.
  • Added the channels to the C0EDCU.ini file and restarted the framebuilder daqd.
  • Modified seisBLRMS.m to use .conf files for the channels. The .conf files are now residing in the compiled matlab directory that Rob made.

I still have yet to compile and test the new version. Its running on linux3 right now, but feel free to kill it and compile it to run on Mafalda.

It should be making trends overnight and so we can finally see what the undergrads are really up to.
  281   Mon Jan 28 17:16:54 2008 AndreyConfigurationComputersMatlab libraries DO NOT WORK properly sometimes

Working in Matlab, I encountered at two different times today the license distribution problem:

??? License checkout failed.
License Manager Error -4
Maximum number of users for Curve_Fitting_Toolbox reached.
Try again later.
To see a list of current users use the lmstat utility or contact your License Administrator.

Troubleshoot this issue by visiting:
http://www.mathworks.com/support/lme4a
  280   Mon Jan 28 15:11:38 2008 robHowToDMFrunning compiled matlab DMF tools

I compiled Rana's seisBLRMS monitor, and it's now running on mafalda. To start your own DMF tools, here is a procedure:

1) build your tool in mDV, get it working the way you'd like.
2) Make a new directory /cvs/cds/caltech/apps/DMF/compiled_matlab/{your_new_directory} and copy the *.m file there.
3) Make the *.m in your new directory into a function with no args (just add a function line at the top)
4) compile it (from within a fully mDV-configured matlab) with mcc -m -R -nojvm {yourfile}.m at the matlab command line.
5) add a line corresponding to your new tool to the script /cvs/cds/caltech/apps/DMF/scripts/start_all
6) Run the start_all script referenced in part (5).

NB: Steps (4) and (6) must be carried out on mafalda.
  279   Mon Jan 28 12:42:48 2008 DmassBureaucracyTMICoffee
There is tea in the coffee carafe @ the 40m. It is sitting as though it were fresh coffee. There is also nothing on the post it.
  278   Sun Jan 27 21:44:48 2008 ranaUpdateCDSSeismic BLRMS on Matlab
I wrote a matlab script to produce band limited RMS trends from our accelerometers. It mimics the code written
by Ed Daw which makes the seismic FOMs at the sites.

Here's how it works:
  • Use mDV to get data by reading directly from frames.
  • Use the Matlab pwelch function to produce a power spectrum of the channels.
  • Use the Matlab find function and rms.m to get the RMS in user-defined frequency bands.
  • Makes a tdswrite command string which writes all the values to EPICS channels.
  • The EPICS channels are just a list of simple names in a database file.
  • The channel names are (will be) added to the C0EDCU.ini file so that its all trended.

The code is in the mDV/extra/C1 directory; its ~20 lines of code (excluding comments and spaces).

Next up is to add more DMF trend channels to the database and upgrade the code to use a .conf file
instead of hardcoded channel names. We should also evaluate if the bands I used are appropriate for the 40m;
I just used Ed's choices (0.1-0.3, 0.3-1, 1-3, 3-10, and 10-30 Hz).

In the medium term, we should make this compiled (like what RW did with the linetracker), and explore if we
want it to write values faster than 1/minute.
  277   Sun Jan 27 13:13:21 2008 tobinMetaphysicsGeneraldeparture
It's been grand. Thanks for having me!

GWAVES IN '08!

Sugar napoleons may be forwarded to T. F., c/o LLO, P.O. Box 940, Livingston, LA 70754-0940.
  276   Sat Jan 26 22:00:03 2008 JohnUpdateGeneralLSC-TRY_OUT and ETMY-QPD
In the path from the ETM to the trans PD and QPD at the Y end I have replaced a BS1-1064-10-2037-45P with a polariser. The power falling on these diodes has been reduced. When the arm is locked in its nominal state the transmitted power is now less than 1.

This polariser should serve as an injection point for the auxiliary arm locking. I am attempting to use crossed polarisations to separate this loop from the main arm light.
  275   Sat Jan 26 18:48:43 2008 JohnSummaryComputersRestart iscepics
iscepics died this afternoon. We restarted it and restored settings from yesterday. I've written up instructions in the wiki.
  274   Sat Jan 26 18:12:45 2008 ranaHowToComputersextra processes and crashes
There were a load of extra processes on op440m; they were all related to rogue DV processes (frameMemRead, framer4, xmgrace, etc.)

You can check this by running top. IF there's lots of them you can kill them by using 'pkill'.

I'm attaching the screen cap of a 'top' session. You can also see that the cron of updatedb is running and taking upp all the CCPU. Thee result is that the delay puts misspellings and doouble characters intoo my elog entry. Clearly wwe'll have to fix the cron setup.
  273   Sat Jan 26 02:34:34 2008 AndreyUpdateComputer Scripts / ProgramsOvernight Measuremts in XARM

I am running the program for measuring RMS of peaks in XARM tonight. I just started it, and it will run for about 9 hours until noon on Saturday. Please do not disturb the interferometer. Now the XARM is locked, it should stay locked over the night.

Andrey.
  272   Sat Jan 26 02:08:53 2008 JohnOmnistructureLSCFibres
There is now a fibre running from the SP table to the ISCT at the Y-end. In the coming days I will try to mode match the beam from this fibre into the arm through ETMY. To achieve this I will be altering the optical layout of this table.
  271   Sat Jan 26 02:02:43 2008 JohnSummaryGeneralNew Channels
I added the following channels.

# C1ASC_QPDs
[C1:SUS-ETMY_QPDSUM_MON]
[C1:SUS-ETMY_QPDYAW_MON]
[C1:SUS-ETMY_QPDPIT_MON]

[C1:SUS-ETMX_QPDSUM_MON]
[C1:SUS-ETMX_QPDYAW_MON]
[C1:SUS-ETMX_QPDPIT_MON]

The old .ini file is /cvs/cds/caltech/chans/daq/C0EDCU_26_1_2008.ini
  270   Fri Jan 25 21:36:40 2008 ranaUpdateCDSmDV / channel issues
Fri Jan 25 21:30:00 2008

As it turns out, the residual problem with the mDV stuff was not to do with our button pushing episode but instead fallout from the 'turning off of the computers' during the water leak caused by the rain and construction.

The /frames partition from fb0 (the FrameBuilder) is not mounted to the control machines via vfstab; it does not remount on bootup. I originally did this because Ben Johnson and Dave Barker had warned me that during a power outage, fb0 may not come up right away. This could make the control room machines hang up for awhile. I elected to have the mount be by hand.

So the thing to do is to put the mount command into the cold start procedures (Andrey). Its in an old elog entry of mine from Feb '07.
  269   Fri Jan 25 17:11:07 2008 Max , AndreyConfigurationGeneralNEW_FETCH_SHOUROV and GET_DATA do not work

The problem which started yesterday after Andrey's framebuilder restart still persists.

It is still impossible to read data in the past from the channels using "get_data" which in turn uses "new_fetch_shourov".

Max was trying to read data from the channel
"C1:LSC-DARM_CTRL",

and he got the same error messages as Andrey.

Andrey tried earlier today to read data from "C1:SUS-ITMS_SUS" or "C1:SUS-ETMX_SUS" with the error meassge
Error in ==> new_fetch_shourov at 22
at (start_time+duration) > stops(end)

So, it seems that Robert Ward fixed just one problem out of two problems.

Robert revived the realtime signals in Dataviewer,
but did not revive the memory of channels for new_fetch_shourov.

To be more precise, channels have memory (it is possible to see the "Playback" curves in Dataviewer"),
but "get_data" and "new_fetch_shourov" do not see the data from those channels. The problem appeared immediately after Andrey's clicking on blue buttons to restart the framebuilder.

Andrey again apologizes.
ELOG V3.1.3-