Fri Mar 5 11:21:18 2010 |
steve | Update | SUS | used OSEMs are magnetic |
Quote: |
Jenne and Koji
We successfully hung ITMX on the SOS. Side magnet is ~2mm off from the center of the OSEM. ITMX aligned using the QPD. The OSEMs changes the alignment. It looks that something magnetic is inside the OSEM PD or LED.
Reguled ITMY side magnet.
Cleaned up the lab for the safety inspection.
|
The brand new OSEM LED and PD can be picked up with a weak magnet. These ferrous metals of LEDs and PDs will be magnetized by sitting in the sus next to the
magnets for years. I hanged optics with new OSEMs and never saw this effect before.
We have to demagnetize them. |
2659
|
Fri Mar 5 18:04:56 2010 |
rana | Update | SUS | used OSEMs are magnetic |
The OSEM LEDs and PDs from Honeywell have always had some ferromagnetic material in them. These are the same OSEMs we had since 2000.
You must be thinking of the really old 20th century plastic OSEMs. |
6822
|
Sat Jun 16 01:03:21 2012 |
yuta | Update | Green Locking | used longer delay line for mode scan |
[Mengyao, Yuta]
Last night, I used 1.5 m delay line COARSE and got 5FSR mode scan. The range 5FSR was limited by the range of SR560.
So, this time, we used 6.4 m(21 feet) cable as a delay line for FINE servo. This should increase the sensitivity by factor of 4. But the range will be 4 tmes smaller, ~ 1.3FSR.
Below is the plot of the mode scan.
You can see the peak height difference between TEM00s, but it's just from the resolution of pixels.
You still can see noisiness goes up when blue plot goes down. But this time, 2000 stands for 27 MHz and -2000 stands for 15 MHz in the beat frequency because we flipped the filter gain this time.
Last night, the top of the triangle was about 40 MHz and bottom was about 60 MHz.

We are going to derive mode-matching and some cavity parameters using this plot. |
9023
|
Sun Aug 18 20:07:41 2013 |
rana | Update | Computer Scripts / Programs | userapps SVN up |
JoeB and JamieR are working somewhat coherently on a set of python libraries to fulfill all of our command line CDS wants. This is being done mostly to satisfy The Guardian and the SkunkTools project.
I did an 'svn up' in /opt/rtcds/userapps (it might finish in ~1000 years) to get the things that they have so far (in particular, Joe's 'pyavg'). There's going to be some issues since the pylib stuff written by Yuta/Kiwamu has never been integrated with anything and is imported as 'epics' in many python scripts. As we move over to the new stuff there will be a lot of broken script functions since the new libraries are also used in that way. |
1614
|
Wed May 20 16:03:52 2009 |
steve | Omnistructure | Environment | using OSEMs to look at seismic activity |
Rana suggested using OSEM sensing voltages as guide lines to look seismic activity.
As you see todays drilling and tumping activity was nothing compared to the EQ of mag 5 and 4
Optic level servos are turned back on.
What Steve means is that there is some drilling going on in the CES shop to accomodate the new water flume group. We want to
make sure that the mirrors don't move enough to break the magnets. On the dataviewer we should look to make sure that the
sensor channels stay between 0-2 V. -Rana
|
Attachment 1: eqOSEM5m4.jpg
|
|
7620
|
Thu Oct 25 09:32:17 2012 |
Steve | Omnistructure | IOO | using access connector |
Quote: |
Quote: |
Quote: |
Quote: |
We really need something better to replace the access connector when we're at air. This tin foil tunnel crap is dumb. We can't do any locking in the evening after we've put on the light doors. We need something that we can put in place of the access connector that allows us access to the OMC and IOO tables, while still allowing IMC locking, and can be left in place at night.
|
It is in the shop. It will be ready for the next vent. Koji's dream comes through.
|
24" diameter clear acetate access connector is in place. The 0.01" thick plastic is wrapped around twice to insure air and bug tight barrier for the MC to lock overnight. The acetate transmission for 1064 nm is 90 % This was measured at 150 mW 2.5 mm beam size.
|
Aluminum sheet as shown will replace the acetate. Side entries for your arms and "window" on the top will be covered with acetate using double- sided removable-no residue tape 3M 9425
|
The second loop of the bungee cord should be on the top of the acrylic and still on the supporting aluminum tube as shown. |
Attachment 1: acclosing.jpg
|
|
196
|
Tue Dec 18 16:50:35 2007 |
tobin | Update | SAFETY | uvex laser safety glasses defective |
A few days ago we noticed what appeared to be a blotched, speckled fracturing of the coating of the "UVEX" laser safety glasses. These are the glasses with "transparent" (reflective to 1064nm) lenses and white frames that we keep in a box on top of a filing cabinet in the control room. Today Steve measured the transmission of these glasses and found 80% transmission of 1064nm in several cases.
Do not use the white, transparent "uvex" laser safety glasses until further notice. Steve has hidden them away so that you won't be tempted.
Below is attached a photo of a bad lens. |
Attachment 1: bad-glasses.jpg
|
|
Attachment 2: bad-glasses-zoom.jpg
|
|
24
|
Mon Oct 29 09:46:50 2007 |
steve | Routine | VAC | vac & pem trend |
Pumpdown 64 pumped by maglev for 125 days
pd64-m-d125
Rob, can you tell me, when did the fire start on this plot? |
Attachment 1: pd64md125.jpg
|
|
1199
|
Sun Dec 21 13:00:06 2008 |
steve | Update | all down cond. | vac and laser back on |
There was a power outage sometimes early Saturday.
All things are down.
The Maglev was started and reached normal operating condition.
V1 was opened at P1 3.8 mTorr and cc1 is back to 3e-6 Torr now
The MOPA was turned on.
Ion pump HV on to FSS cavity. |
423
|
Thu Apr 17 15:00:59 2008 |
steve | Update | VAC | vac controller computer to be replaced |
Now that Rosalba is up and running someone should step forward to volunteer for this job.
Old desk top at the Vacuum Rack should be replaced by a functional less old computer so
the vacuum controller can count on an emergency situation. It's only has to be able to run medm
The existing one is dying. |
5366
|
Thu Sep 8 13:20:07 2011 |
steve | Update | PEM | vac envelope at atm for 36 days |
We tried not to open chambers above 10,000 particles of 0.5 micron cf/min
New items going in: 2 rasor beam traps, 5 badly oxidized old silver plated setscrew with spring loaded tips......to be replaced in the future, viton tips for eq screws....some are lose, gold plated allen wrench installed at ITMX bottom, reglued magnet on ITMX
Bad hardware things found: nylon ball "locking elements" on OSEM locking set screws with screwdriver slot, lose 1064 nm filter on OSEM pd |
Attachment 1: vent70.jpg
|
|
1293
|
Wed Feb 11 11:39:17 2009 |
steve | Configuration | General | vac envelope ground |
The vacuum envelope was grounded from the IOO chamber to master ground north (under electric breaker panel L)
with 4GA 25ft red battery cable.
|
2699
|
Tue Mar 23 09:37:36 2010 |
steve | Update | IOO | vac envelope has to be sealed as antproof for overnight |
Quote: |
This is the first touch to the MC mirrors after the earthquake on 16th.
- I made an aluminum access connector so that we can work on the MC even the door is open. We still can be able to open the aluminum tube. The photos are attached. Steve, could you please look it at a glance whether the seal is enough or not.
- MC resonances were flashing. Align MC2 and MC3 so that we have many TEM00s.
- Found c1vmesus2 gone mad. Restarted remotely according to the wiki entry.
- Reset the MC coil output matrix to 1. (Previously, balance was adjusted so that A2L was minimized.)
- Excite MC2 Pitch/Yaw at 8 and 9 Hz, looking at the peaks in the MC-MCL output. Move MC2 Pitch/Yaw so that the peak
is reduced. (*)
- MC1/MC3 were aligned so that we get the maximum transmission (or minimum reflection). (**)
- Repeat (*) and (**)
So far, I have aligned in Yaw such that the yaw peak is minimized.
|
This seal is good for daily use- operation only. The IFO has to be sealed with light metal doors every night so ants and other bugs can not find their way in.
Our janitor Kevin is mopping the hole IFO room floor area with 5% ant killing solution in water in order to discourage bugs getting close to our openings of the vented chamber.
You may be sensitive to this chemical too. Do not open chamber till after lunch. |
Attachment 1: pc3.JPG
|
|
Attachment 2: pc4.JPG
|
|
2700
|
Tue Mar 23 09:55:20 2010 |
Koji | Update | IOO | vac envelope has to be sealed as antproof for overnight |
Roger.
Quote: |
This seal is good for daily use- operation only. The IFO has to be sealed with light metal doors every night so ants and other bugs can not find their way in.
|
|
2705
|
Wed Mar 24 02:06:24 2010 |
Koji | Update | IOO | vac envelope has to be sealed as antproof for overnight |
Matt and Koji:
We closed the light doors of the chambers.
Quote: |
Roger.
Quote: |
This seal is good for daily use- operation only. The IFO has to be sealed with light metal doors every night so ants and other bugs can not find their way in.
|
|
|
1462
|
Thu Apr 9 11:27:19 2009 |
steve | Update | VAC | vac gauge reading problem |
Cold cathode gauge CC4 is reading normal.
CC1 is glitching, it is probably dirty.
CC2 is fluctuating too much and it is cutting out for 6-7 minutes. It must be insulated by deposits and there is no emission current.
I think the same goes for P1
They will have to be replaced at the next vent |
Attachment 1: vacgflsec.jpg
|
|
Attachment 2: vacgflmin.jpg
|
|
10081
|
Fri Jun 20 14:42:57 2014 |
steve | Update | SUS | vac illuminators turned off |
|
1552
|
Wed May 6 19:04:11 2009 |
rana | Summary | VAC | vac images |
Since there's no documentation on this besides Steve's paper notebooks...
and BTW, since when did the elog start giving us PNG previews of PDFs? |
Attachment 1: vacrack.pdf
|
|
7914
|
Thu Jan 17 17:04:49 2013 |
Steve | Update | VAC | vac monitor screen changed |
The COVAC_MONITOR.adl was changed. Ion pump labeled as disconnected means: ion pump controller power turned off and ion pump gate valve control connection disconnected.
RP2 roughing pump labeled disconnected means: hardware disconnected.
The actual operational, valve control screen has not been changed yet. |
Attachment 1: screenmod.png
|
|
11156
|
Sun Mar 22 18:42:40 2015 |
Steve | Update | VAC | vac pressure rose to 1.2mTorr |
Quote: |
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things.
|
We run out of N2 for the vacuum system 6 hrs ago. The pressure rose to 1.2 mTorr with V1 closed. The interlock worked! See Nirogen presure reading fixed at http://nodus.ligo.caltech.edu:8080/40m/10968
The vacuum interlock: Nitrogen pressure transducer is reading the pneumatic pressure continously at the pump spool and c1vac1 processing it. When it drops below 60 PSI it closes V1 gate valve and V4 & V5. Gate valve V1 needs minimum 60 PSI to close. It is critical that V1 is closed before you run out of Nitrogen so the IFO pressure is contained.
IFO vacuum is back to Vac Normal. The MC is locked.
cc4 = 2E-6 Torr with VM1 open.
Daily N2 consumption measured to be 530PSI as 3 days on 3-27-2015 but note: it does vary !
I have seen it as high as 900 psi The long term average ~750 psi |
Attachment 1: NoN2.png
|
|
Attachment 2: PressureReadingWorks.png
|
|
10843
|
Fri Dec 26 17:45:21 2014 |
Steve | Update | VAC | vac pressure rose to 1.3 mTorr |
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things. |
Attachment 1: noN2.png
|
|
Attachment 2: backtoVacNormal.png
|
|
12014
|
Tue Mar 1 09:37:15 2016 |
steve | Update | VAC | vac rack UPS batteries replaced |
Amstron batteries replaced after 11 months with SP-12-5.5HR, 2 years warranty from replaceUPSbattery.com
Quote: |
Batteries replaced after 3.5 years with Amstron AP-1250F2, 8x 12V 6Ah
Quote: |
APC Smart -UPS 2200 model: SUA2200RM2U batteries were replaced by compatible RBC43, 8x 12V5A
|
|
Note: the replace battery LED did not go out ( well pasted 24 hrs ) till the self test bottom was hold down for 2-3 sec |
Attachment 1: vacRackBat.jpg
|
|
5529
|
Fri Sep 23 16:25:01 2011 |
steve | Update | VAC | vac rack UPS battery replaced |
APC Smart -UPS 2200 model: SUA2200RM2U batteries were replaced by compatible RBC43, 8x 12V5A |
Attachment 1: P1080252.JPG
|
|
Attachment 2: P1080254.JPG
|
|
11178
|
Fri Mar 27 10:38:40 2015 |
steve | Update | VAC | vac rack UPS battery replaced |
Batteries replaced after 3.5 years with Amstron AP-1250F2, 8x 12V 6Ah
Quote: |
APC Smart -UPS 2200 model: SUA2200RM2U batteries were replaced by compatible RBC43, 8x 12V5A
|
|
Attachment 1: UPSvac.jpg
|
|
1511
|
Thu Apr 23 16:38:33 2009 |
steve | Update | VAC | vac valve relay box is shorting |
Ben and I found this vacuum valve relay box intermittently shorthing yesterday.
It effects V4, V5, VA6 and VM1........ Please do not touch this box under the beam pipe next to the vac rack!
The function of this box to send 120VAC to the vacuum valve to move. |
Attachment 1: vacrel.png
|
|
7547
|
Mon Oct 15 13:07:51 2012 |
jamie | Update | VAC | vacuum VME crate broken, replaced, minor vacuum mayhem ensues |
Steve and I managed to access the fuse in the vacuum VME crate, but replacing it did not bring it back up. We decided to replace the entire crate.
We manually checked that the most important valves, VC1, VM1 and V1, were all closed. We disconnected their power so that they would automatically close, and we wouldn't have to worry about them accidentally opening when we rebooted the system.
We noted where all the cables were, disconnected everything, and removed the crate. We noted that one of the values switched when we disconnected one of the IPC cables from a VME card. We'll note which one it was in a followup post. We thought that was a little strange, since the VME crate was completely unpowered.
Anyway, we removed the crate, swapped in a spare, replaced all the cards and connections, double checked everything, then powered up the crate. That's when minor chaos ensued.
When the system came back online after about 20 seconds, we heard a whole bunch of valves switching. Luckily we were able to get the medm screens back up so that we could see what was going on.
Apparently all of the ION pump valves (VIPEE, VIPEV, VIPSV, VIPSE) opened, which vented the main volume up to 62 mTorr. All of the annulus valves (VAVSE, VAVSV, VAVBS, VAVEV, VAVEE) also appeared to be open. One of the roughing pumps was also turned on. Other stuff we didn't notice? Bad.
We ran around and manually unplugged all of the ION pump valves, since I couldn't immediately pull up the vacuum control screen. Once that was done and we could see that the main volume was closed off we went back to figure out what was going on.
We got the medm vacuum control screen back (/cvs/cds/caltech/medm/c0/ve/VacControl_BAK.adl. really??) There was a lot of inconsistency between the readback states of the valves and the switch settings. Toggling the switches seemed to bring things back in line. At this point it seemed that we had control of the system again. The epics readings were consistent with what we were seeing in the vacuum rack.
We went through and closed everything that should have been closed. The line pressure between the big turbo pump TP1 and the rest of the pumps was up at atmosphere, 700 Torr. We connected the roughing pumps and pumped down the lines so that we could turn the turbos back on. Once TP2 and TP3 were up to speed, we turned on TP1 and opened V1 to start pumping the main volume back town. The main volume is at 7e-4 Torr right now.
So there are a couple of problems with the vacuum system.
- Why the hell did valves open when we rebooted the VME crate? That's very bad. That should never happen. If the system is set to come up to an unsafe state that needs to be fixed ASAP. The ION pump valves should never have opened. Nor the annulus valves.
- Why were the switches and the readbacks showing different states?
- Apparently there is no control of the turbo pumps through MEDM. This should be fixed.
I connect belledona, the laptop at the vacuum station to the wired network, so that it's connection would be less flaky.
|
11703
|
Tue Oct 20 16:27:04 2015 |
ericq | Update | VAC | vacuum VME machines rebooted |
[ericq, Gautam, Steve]
Following roughly the same procedure as ELOG 11354, c1vac1 and c1vac2 were rebooted. The symptoms were identical to the situation in that ELOG; c1vac1 could be pinged and telneted to, but c1vac2 was totally unresponsive.
The only change in the linked procedure was that we did not shut down the maglev. Since I unwittingly had it running for days without V4 open while Steve was away, we now know that it can handle shorter periods of time than that...
Upon reboot, many channels were readable again, unfortunately the channels for TP2 and TP3 are still blank. We were able to return to "Vacuum normal state," but because of unknowned communication problems with VM1's interlock, we can't open VM1 for the RGA. Instead we opened VM2 to expose the RGA to the main IFO volumn, but this isn't part of the "Normal" state definite, so things currently read "Undefined state". |
12525
|
Fri Sep 30 10:37:57 2016 |
Steve | Update | VAC | vacuum VME machines rebooted |
[ Gautam and Steve ]
c1vac1 and c1vac2 were rebooted and the gauges are communicating now. V1, VA6, V5 and V4 were closed and disconnected to avoid unexpected valve switching. All went smoothly.
The new ITcc gauge is at 1e-5 Torr as CC1 This is the gauge that should be logged in slow channel.
TP2 fore line dry pump was replaced this morning after 382 day of operation.
TP3 dry pump is very noisy, but it's pressure still 47 mTorr
Quote: |
[ericq, Gautam, Steve]
Following roughly the same procedure as ELOG 11354, c1vac1 and c1vac2 were rebooted. The symptoms were identical to the situation in that ELOG; c1vac1 could be pinged and telneted to, but c1vac2 was totally unresponsive.
The only change in the linked procedure was that we did not shut down the maglev. Since I unwittingly had it running for days without V4 open while Steve was away, we now know that it can handle shorter periods of time than that...
Upon reboot, many channels were readable again, unfortunately the channels for TP2 and TP3 are still blank. We were able to return to "Vacuum normal state," but because of unknowned communication problems with VM1's interlock, we can't open VM1 for the RGA. Instead we opened VM2 to expose the RGA to the main IFO volumn, but this isn't part of the "Normal" state definite, so things currently read "Undefined state".
|
|
Attachment 1: rebooted.png
|
|
2648
|
Mon Mar 1 11:21:50 2010 |
steve | Configuration | VAC | vacuum contoller is back to normal |
Quote: |
There is a planned power outage tomorrow, Saturday from 7am till midnight.
I vented all annulies and switched to ALL OFF configuration. The small region of the RGA is still under vacuum.
The vac-rack: gauges, c1vac1 and UPS turned off.
|
Vac- rack is powered back up. UPS first, than all other power switches from top to bottom of the rack, except Maglev
Manually started one by one TP2 and TP3 to accelerate to 50 KRPM
Brought up vac.control screen on lap-top at /cvs/cds/caltech/medm/c0/ve/VacControl_BAK.adl
V5 and VM3 were opened so TP3 can pump on the RGA
V4 was opened so TP2 can pump on the Maglev-TP1. The Maglev power was turned on and started acceleration.
The vac control screen positions indicators were checked for true position and annulies vent valves were opened.
RGA manual on/off switch was turned at the top of the RGA-head. Ubuntu copmuter was started at cc4 1.1e-6 Torr
The RGA communication was started with: ssh c0rga from control room
The rga-script was started ./RGAset.py This script turns on the filament, rs-232 and scan parameters etc
Vac -configuration: IFO-P1 at atm, RGA is pumped and running in background mode, all annulos at atm
|
9570
|
Thu Jan 23 08:04:04 2014 |
Steve | Update | VAC | vacuum control screen is blank |
There is BLANK VacControl_BAK.adl screen only.
I can move a valve by disconnecting it's solenoid power if it's position is normally open.
I will close V1 and check computer cable connections and move on with manual - hand disconnect ea valve to be moved into the right position for vent. Valve positions will be confirmed by looking manual indicators on valves. |
Attachment 1: pd76m170d2dRgaOn.png
|
|
11960
|
Fri Jan 29 09:59:23 2016 |
Steve | Update | VAC | vacuum control upgrade |
Rana stated yesterday that there will be a vacuum control update in the close future. Witnesses : Rich, Chris and Dave
Can you give me this in writing?
|
Attachment 1: 1100days.png
|
|
11706
|
Wed Oct 21 15:42:56 2015 |
Steve | Update | VAC | vacuum gauge check |
Convectron gauge check:
Brand new 10 years old convectron gauge at atm was swapped into the place of existing gauges to see if they read close to 760 Torrs
They did reasonable well at the low end. I tried to imitate calibration and bring down the high end with little success.
( P1 and P2 were reading 7e-4 to ~660 Torr The correction of the upper end pushed up the lower end too. I will correct this later
P3 and P4 high ends are way off )
The point is that they work. Convectron gauges will be replaced and calibrated at the next vent.
Interlocks were not triggered during this test. I was expected to close the PSL shutter when P1 was reading 760 Torr
This hide some problem or not understanding.
It was good to see CC1 and CC4 working at the moment
Somehow I succeded opening VM1 and closed VM2 = vacuum normal
I just hope it stays open overnight to get comparable RGA scan.
|
Attachment 1: Gauges.png
|
|
13664
|
Mon Mar 5 10:13:21 2018 |
gautam | Update | VAC | vacuum health |
In Steve's absence, I've tried to keep an eye on the health of the vacuum system. From Attachment #1, the pressure of the main volume seems stable, no red flags there. I also don't here any anomalously loud sounds near the vacuum pumps. I've changed the N2 cylinders that keep V1 open twice, on Wednesday and Sunday of last week. So in summary, the vacuum system looks fine based on all the metrics I know of. |
Attachment 1: VacuumStatus.png
|
|
12487
|
Tue Sep 13 16:16:28 2016 |
Chris | Omnistructure | VAC | vacuum interlock bypassed |
(Steve, Chris)
The pumpdown had stalled because of some ancient vacuum interlock code that prevented opening the valve V1 between the turbo pump and the main volume.
This interlock [0] compares the channels C1:Vac-P1_pressure and C1:Vac-PTP1_pressure, neither of which is functioning at the moment. The P1 channel apparently stopped reading sometime during the vent, and contained a value of ~700 torr, while the PTP1 channel contained 0. So the interlock code saw this huge apparent pressure difference and refused to move the valve.
To bypass this check, we used caput to enter a pressure of 0 for P1.
[0] /cvs/cds/caltech/state/from_luna/VacInterlock.st |
2624
|
Mon Feb 22 11:38:05 2010 |
joe, jenne and steve | Configuration | VAC | vacuum is back to normal |
Morning condition: vacuum rack power is still off, no MEDM screen reading.....meaning unknown vacuum pressure.We closed PSL shutter immediately.
Joe restored c1iscepis and Jenne powered up the vac-rack UPS. Now the rest of the vac-rack power were restored from starting at the top to bottom.
P1 was reading 15 mTorr. We restarted pumps and set vacuum valve positions. V1 opening required Rob's recipe of elog # 1863 to defeat interlock that
has a non communicating gauge: PTP1
CC1 pressure just reached 1e-6 Torr at VAC NORMAL configuration. |
5962
|
Sun Nov 20 13:51:54 2011 |
steve | Update | VAC | vacuum is back up after ac interruption |
State condition: Vac Normal, CC1 ~8e-6 Torr
The IFO pressure peaked at 8.3 mTorr after 2days of not pumping. |
Attachment 1: notpumping.png
|
|
7549
|
Mon Oct 15 17:08:01 2012 |
Steve | Update | VAC | vacuum is back with some disabled valves |
Quote: |
Quote: |
Apparently all of the ION pump valves (VIPEE, VIPEV, VIPSV, VIPSE) opened, which vented the main volume up to 62 mTorr. All of the annulus valves (VAVSE, VAVSV, VAVBS, VAVEV, VAVEE) also appeared to be open. One of the roughing pumps was also turned on. Other stuff we didn't notice? Bad.
|
Several of the suspensions were kicked pretty hard (600+ mV on some sensors) as a result of this quick vent wind. All of the suspensions are damped now, so it doesn't look like we suffered any damage to suspensions.
|
CLOSE CALL on the vacuum system:
Jamie and I disabled V1, VM2 and VM3 gate valves by disconnecting their 120V solenoid actuator before the swap of the VME crate.
The vacuum controller unexpectedly lost control over the swap as Jamie described it. We were lucky not to do any damage! The ion pumps were cold and clean. We have not used them for years so their outgassing possibly accumulated to reach ~10-50 Torr
I disconnected_ immobilized and labelled the following 6 valves: the 4 large ion pump gate valves and VC1, VC2 of the cryo pump. Note: the valves on the cryo pump stayed closed. It is crucial that a warm cry pump is kept closed!
This will not allow the same thing to happen again and protect the IFO from warm cryo contamination.
The down side of this that the computer can not identify vacuum states any longer.
This vacuum system badly needs an upgrade. I will make a list. |
Attachment 1: VMEcrateout.png
|
|
Attachment 2: IMG_1725.JPG
|
|
Attachment 3: warning.png
|
|
Attachment 4: shortcomings.jpg
|
|
10845
|
Sun Dec 28 07:29:16 2014 |
Steve | Update | VAC | vacuum is normal |
Quote: |
We run out of N2 for the vacuum system. The pressure peaked at 1.3 mTorr with MC locked. V1 did not closed because the N2 pressure sensor failed.
We are back to vac normal. I will be here tomorrow to check on things.
|
ITMX damping restored. |
Attachment 1: 3d.png
|
|
9749
|
Tue Mar 25 14:52:57 2014 |
steve | Update | VAC | vacuum is recovered |
Out gassing plus leak rate 0.15 mTorr / hour
The pressure rose to 2.5 mTorr in 17 hours
V1 was opened at 1:56pm
VM2 opened at 2:10 so the RGA region is back to 1e-5 torr
|
Attachment 1: 17hrsNoPumping.png
|
|
9615
|
Mon Feb 10 09:45:53 2014 |
Steve | Update | VAC | vacuum normal is reached |
Quote: |
Pumpdown completed. IR shutter opened at P1 1 mTorr The block is still in the beam path.
Remember to protect MCR pd before crack up the PSL power.
The ion pump gate valves were just closed by cc1 triggered interlock
The cry pump was "regenerated" during the vent and it's outgassing rate minimized.
CC3 cold cathode gauge was replaced.
Valve configuration for week end:
1, VA6 disconnected to avoid accidental venting the IFO through the annulos
2, VC2 disconnected to insure that the cryo stays closed
3, RGA is not running, It's pressure limit 1e-5 Torr
|
Valve configuration: Vacuum Normal is reached in really 4 days if we do not count overnight rest of roughing.
VA6 and VC2 are reconnected. I'm turning on the RGA next
All 4 ion pumps were vented with air and pumped down to ~ 1e-4 Torr
Ion pumps gate valve control cables are connected and their pumps are disconnected.
The black relay box was tested repeatedly and it stopped misbehaiving.
We were at atmosphere for 13 days. Chamber BS, ITMX, ITMY and ETMY were opened.
Al foil "cups" were placed on the back side OSEMs of PRM. |
Attachment 1: pd77m5d.png
|
|
Attachment 2: 13daysAtm.png
|
|
13268
|
Tue Aug 29 15:35:19 2017 |
Steve | Update | VAC | vacuum pump specifications & manuals |
RP1 and RP3 roughing pump manual of Leybold D30A oily rotory pump
Fore pump of TP2 & TP3 Varian SH-100 Dry Scroll
TP2 and TP3 small turbo drag pump Varian 969-9361
TP2 and TP3 turbo controller Varian 969-9505
TP1 magnetically suspended turbo pump Osaka TG390MCAB, sn360 and controller TC010M and note : this pump running on 208VAC single phaseIt is not on the UPS !
Osaka Maglev Manual and Osaka Controller Communication Wiring
VC1 cryo pump CTI-Cryogenics Cryo Torr 8 sn 8g23925 SAFETY note: compressor single phase 208VAC and the head driver 3 phase 208VAC Compressor and driver have each separate power cord!
Installed at 40m wiki also

Quote: |
The V1 gate valve specs installed at 40m wiki page. VAT model number 10846-UE44-0007 Our main volume pumping goes through this 8" id gate valve V1 to Maglev turbo or Cryo pump to VC1
The ion pumps have 6" id gate valves:VAT 10844-UE44-AAY1, Pneumatic actuator with position indicator and double acting solenoid valve 115V 60Hz Purchased 1999 Dec 22
UHV gate valves 2.5" id. VAT 10836-UE44 Pneumatic actuator with position indicator and double acting solenoid valve 115V 60 hz, IFO to RGA VM1 & RGA to Maglev VM2
mini UHV gate valve 1.5" id. VAT 01032-UE01 2016 cataloge page 14, manual - no position indicator, VM4 next to manual adjustable fine leak valve to RGA
UHV angle valve 1.5" id, model VAT 28432-GE41, Viton plate seal, pneumatic actuator with position indicator & solenoid valve 115V & single acting closing spring MEDM screen: VM3,VC2, V3,V4,V5,V6,VA6,V7 & annuloses Each chamber annulos has 2 valves.
UHV angle valve 1.5" id, model VAT 57132-GE05 go page 208, Metal tip seal, manual actuating only with position indicator, MEDM screen: roughing RV1 and venting VV1 hand wheel needed to close to torque spec
UHV angle valve 1.5" id. model VAT 28432-GE01 Viton plate seal, manual operation only at IT gauges Hornet & Super Bee and ion pumps roughing ports. These are not labeled.
The Cryo pump interlock wiring was added too
Note: all moving valve plate seals are single.
|
|
7605
|
Wed Oct 24 09:15:12 2012 |
Steve | Update | VAC | vacuum rack 24V ps was current limited |
Quote: |
Quote: |
Quote: |
Apparently all of the ION pump valves (VIPEE, VIPEV, VIPSV, VIPSE) opened, which vented the main volume up to 62 mTorr. All of the annulus valves (VAVSE, VAVSV, VAVBS, VAVEV, VAVEE) also appeared to be open. One of the roughing pumps was also turned on. Other stuff we didn't notice? Bad.
|
Several of the suspensions were kicked pretty hard (600+ mV on some sensors) as a result of this quick vent wind. All of the suspensions are damped now, so it doesn't look like we suffered any damage to suspensions.
|
CLOSE CALL on the vacuum system:
Jamie and I disabled V1, VM2 and VM3 gate valves by disconnecting their 120V solenoid actuator before the swap of the VME crate.
The vacuum controller unexpectedly lost control over the swap as Jamie described it. We were lucky not to do any damage! The ion pumps were cold and clean. We have not used them for years so their outgassing possibly accumulated to reach ~10-50 Torr
I disconnected_ immobilized and labelled the following 6 valves: the 4 large ion pump gate valves and VC1, VC2 of the cryo pump. Note: the valves on the cryo pump stayed closed. It is crucial that a warm cry pump is kept closed!
This will not allow the same thing to happen again and protect the IFO from warm cryo contamination.
The down side of this that the computer can not identify vacuum states any longer.
This vacuum system badly needs an upgrade. I will make a list.
|
While I was doing the oil change of the roughing pumps I accidentally touched the 24 V adjustment knob on the power supply.
All valve closed to default condition. I realized that the current indicator was red at 0.2A and the voltage fluctuated from 3-13V
Increased current limiter to 0.4A and set voltage to 24V I think this was the reason for the caos of valve switching during the VME swap.
|
Attachment 1: currentlimited24V.jpg
|
|
5590
|
Fri Sep 30 18:35:42 2011 |
steve, kiwamu | Update | VAC | vacuum set for poweroutage |
We did the following:
1, closed V1, VM1, annuloses: VASE, VASV, VABS, VAEV, VAEE and VA6
2, stop rotation of Maglev-TP1, waited to decellerate and turned off power to it
3, closed V4, stoped rotation of TP2, waited to decellerate and turned power off
4, opened VM3 to RGA that is still running
I will come in tomorrow 9-10am to restart pumping. |
Attachment 1: prepforpoweroutage.png
|
|
13245
|
Wed Aug 23 10:11:46 2017 |
Steve | Update | VAC | vacuum valve specifications |
The V1 gate valve specs installed at 40m wiki page. VAT model number 10846-UE44-0007 Our main volume pumping goes through this 8" id gate valve V1 to Maglev turbo or Cryo pump to VC1
The ion pumps have 6" id gate valves:VAT 10844-UE44-AAY1, Pneumatic actuator with position indicator and double acting solenoid valve 115V 60Hz Purchased 1999 Dec 22
UHV gate valves 2.5" id. VAT 10836-UE44 Pneumatic actuator with position indicator and double acting solenoid valve 115V 60 hz, IFO to RGA VM1 & RGA to Maglev VM2
mini UHV gate valve 1.5" id. VAT 01032-UE01 2016 cataloge page 14, manual - no position indicator, VM4 next to manual adjustable fine leak valve to RGA
UHV angle valve 1.5" id, model VAT 28432-GE41, Viton plate seal, pneumatic actuator with position indicator & solenoid valve 115V & single acting closing spring MEDM screen: VM3,VC2, V3,V4,V5,V6,VA6,V7 & annuloses Each chamber annulos has 2 valves.
UHV angle valve 1.5" id, model VAT 57132-GE05 go page 208, Metal tip seal, manual actuating only with position indicator, MEDM screen: roughing RV1 and venting VV1 hand wheel needed to close to torque spec
UHV angle valve 1.5" id. model VAT 28432-GE01 Viton plate seal, manual operation only at IT gauges Hornet & Super Bee and ion pumps roughing ports. These are not labeled.
The Cryo pump interlock wiring was added too
Note: all moving valve plate seals are single. |
4059
|
Wed Dec 15 16:04:49 2010 |
steve | Configuration | VAC | vacuum valves found in default condition |
This condition is created when we run out of nitrogen that holds the valves in set positions. The Maglev and other turbos are still running in this configuration so one has to pay attention that the foreline valves are open.
Reset error message on vacuum screen and opened V4 & V5 at 3pm today. |
3009
|
Fri May 28 13:32:01 2010 |
alberto, kiwamu | Update | VAC | vacuum work |
We started a vacuum work in this morning. And still it's going on.
Although the last night the green team replaced a steering mirror by an 80% reflector on the PLS table, the beam axis to the MC looks fine.
The MC refl beam successfully goes into the MCrefl PD, and we can see the MC flashing as usual.
We started measuring the distance of the optics inside the vacuum chamber, found the distance from MC3 to MMT1(curved mirror) is ~13cm shorter than the design.
We moved the positions of the flat mirror after the Faraday and the MMT1, but could not track the beam very well because we did not completely lock the MC.
Now we are trying to get the lock of the MC by steering the MC mirrors.
P.S.
Kevin suceeded in locking it !! |
3010
|
Fri May 28 13:51:51 2010 |
Jenne | Update | VAC | vacuum work |
Quote: |
We started a vacuum work in this morning. And still it's going on.
Although the last night the green team replaced a steering mirror by an 80% reflector on the PLS table, the beam axis to the MC looks fine.
The MC refl beam successfully goes into the MCrefl PD, and we can see the MC flashing as usual.
We started measuring the distance of the optics inside the vacuum chamber, found the distance from MC3 to MMT1(curved mirror) is ~13cm shorter than the design.
We moved the positions of the flat mirror after the Faraday and the MMT1, but could not track the beam very well because we did not completely lock the MC.
Now we are trying to get the lock of the MC by steering the MC mirrors.
|
Hey guys,
I just finished redoing the calc based on the measurements that happened last week. Using the average of the Vert and Horz measurements in Kevin's elog 2986, I find that we need to make the MMT telescope ~8cm longer. So, can you please place the flat mirror after the Faraday in the same place as the drawing, but move the MMT1 79mm farther away from that flat mirror? Looking at the table layouts that Koji has on the wiki, this should still (barely) fit.
New distances:
d2a = 884.0mm (no change) ------ MC3 to Flat after Faraday
d2b = 1123.2mm (move MMT1 farther toward center of BS table) -------- Flat after Faraday (SM1) to MMT1
d3 = 1955.0mm (result of moving MMT1) --------- MMT1 to MMT2
d4a = 1007.9mm (no chnage) ----------- MMT2 to SM2
d4b = 495.6mm (no change) ------------ SM2 to PRM |
3011
|
Fri May 28 14:24:30 2010 |
Jenne | Update | VAC | vacuum work |
I just got off the phone with Alberto and Kiwamu, and I'm going to try to recalculate things based on their measurements of the distances between MC3 and SM1. It sounds like the CAD drawings we have aren't totally correct. I know that when we opened doors just before Christmas we measured the distances between the BS table and the ITM tables, but I don't think we measured the distance between the IOO table and the BS table. Hopefully we can fit everything in our chambers..... |
2613
|
Thu Feb 18 15:39:16 2010 |
Koji and Steve | Configuration | VAC | valve condition: ALL OFF |
As preparation for the upcoming planned power outage we turned turbos, RGA off and closed valves.
IFO chamber is not pumped now. Small leaks and out gassing will push the pressure up slowly. At 3 mTorr of P1 the PSL output shutter
will be closed by the interlock.
It is OK to use light in the IFO up to this point. |
5104
|
Wed Aug 3 09:40:13 2011 |
steve | Update | VAC | vent |
The vacuum system is coming up to atm. The vent was started with slow N2 flow. The PSL output is blocked 2 places on the table. |
Attachment 1: pzt2011Aug3.png
|
|