ID |
Date |
Author |
Type |
Category |
Subject |
3458
|
Mon Aug 23 19:55:31 2010 |
kiwamu | Update | CDS | some progress |
{Joe, Kiwamu}
Today we did the following works in order to get ready for the new CDS test.
- solved the DAC issue.
- checked all the channel assignments of the ADC and the DAC.
- preparation for modification of the AA filter chassis.
- checked DAC cable length.
- connected the power cables of the BO boards to Sorensens.
Although we performed those works, we still couldn't do the actual damping tests.
To do the damping tests, we have to modify the AA chassis to let the SCSIs go in it. Now Joe and Steve are working for this issue.
Also we found that we should make three more 37pin Dsub - 40pin IDC cables.
But this is not a critical issue because the cables and the connectors are already in our hands. So we can make them any time later.
(Notes)
(DAC issue)
Now all the DAC channels are working correctly.
There had been a combination of some issues.
When I posted the elog entry last time, the DAC was not working at all (see here).
But in the last week Joe found that the IO process didn't correctly run. He modified the IOP file named 'c1x02.mdl' and ran it after compiling and installing it.
This made the situation better because we then were able to see the most of the signals coming out from the DACs.
However we never saw any signals associated with SIDE_COILs.
We checked the DAC cards, their slots and their timing signals. But they all were fine.
At that time we were bit confused and spent a couple of days because the DAC signals appeared at a different slot some time after we rebooted the computer. Actually this issue still remains unsolved...
Finally we found that SIDE_COILs had an input matrix which didn't show up in the medm screen.
We put 1 in the matrix and we successfully got the signal coming out from the DAC.
(channel assignments)
We checked all the channel assignments of the DACs and the ADCs.
All the channels are assigned correctly (i.e. pin number and channel name).
(AA chassis)
We have been planning to put the SCSI cables into the AA chassis to get the ADC signals.
As Joe said in the past entry (see here) , we need a modification on the AA chassis to let the SCSIs go in it.
Joe and Steve will put an extension component so that we can make the chassis wider and eventually SCSI can go in.
(DAC cable length)
In the default plan we are going to reuse some DAC cables which are connected to the existing systems.
To reuse them we had to make sure that the length of those cables are long enough for the new CDS.
After stopping the watchdogs, we disconnected those DAC cables and confirmed they were long enough.
Now those cables are connected to the original place where they have been before.
The same test will be performed for the binary outputs.
(power cables to Sorensens)
Since the binary output boards need +/- 15V power, we hooked up the power cables to Sorensens sitting on the new 1X5 rack.
After cabling them, we turned on the power and successfully saw the green LEDs shining on the back panel of the boards.
|
8196
|
Thu Feb 28 02:43:49 2013 |
yuta | Update | LSC | some qualitative evidence of PRMI sideband lock |
[Manasa, Yuta]
Since we have setup POP22 PD now(elog #8192), we could confirm that sideband power builds up when PRMI is sideband locked.
Plot:
Here's some plot of PRC intra-cavity powers and MICH,PRCL error signals. As you can see from POP22, we locked at the peak of 11MHz sideband. There was oscillation at ~500 Hz, but we couldn't optimize the gain yet.

Movie:
Here's 30 sec movie of AS, POP, REFL when acquiring (and losing) PRMI sideband lock. It was pretty hard to take a movie because it locks pretty seldom (~1 lock / 10 min).
Locking details:
For MICH lock, we used ITMs instead of BS for reducing coupling between PRCL.
Also, AS55 phase rotation angle was coarsely optimized by minimizing MICH signal in I.
For PRCL lock, we used REFL55_I_ERR instead of REFL33_I_ERR. It had better PDH signal and we coarsely optimized phase rotation angle by minimizing PRCL PDH signal in Q.
== PRMI sideband ==
MICH: AS55_Q_ERR, AS55_PHASE_R = -12 deg, MICH_GAIN = -0.1, feedback to ITMX(-1),ITMY(+1)
PRCL: REFL55_I_ERR, REFL55_PHASE_R = 70 deg, PRCL_GAIN = -15, feedback to PRM
We set POP22_PHASE_R = -170 deg by minimizing Q.
Issues:
- We tried to use REFL55_Q_ERR to lock MICH, but couldn't. It looks like REFL error signals are bad.
- We tried to use POP22_I_ERR to trigger PRCL lock, but it didn't work. |
14907
|
Thu Sep 26 17:56:28 2019 |
Koji | Update | CDS | some rebooting |
Yesterday (Sep 25) evening: I had to reboot c1psl, c1iool0, and c1aux to recover nominal IMC locking
Today megatron had no response and I had to reboot it with the reset button. MCautolocker and FSSSlow were recovered and the IMC is locking as usual. |
5702
|
Wed Oct 19 16:53:38 2011 |
kiwamu | Update | CDS | some screens need labels |

Some of the sub-suspension screens need labels to describe what those row and column are. |
11294
|
Sat May 16 21:05:24 2015 |
rana | Update | General | some status |
1) Checked the N2 pressures: the unregulated cylinder pressures are both around 1500 PSI. How long until they get to 1000?
2) The IMC has been flaky for a day or so; don't know why. I moved the gains in the autolocker so now the input gain slider to the MC board is 10 dB higher and the output slider is 10 dB lower. This is updated in the mcdown and mcup scripts and both committed to SVN. The trend shows that the MC was wandering away after ~15 minutes of lock, so I suspected the WFS offsets. I ran the offsets script (after flipping the z servo signs and adding 'C1:' prefix). So far powers are good and stable.
3) pianosa was unresponsive and I couldn't ssh to it. I powered it off and then it came back.
4) Noticed that DAQD is restarting once per hour on the hour. Why?
5) Many (but not all) EPICS readbacks are whiting out every several minutes. I remote booted c1susaux since it was one of the victims, but it didn't change any behavior.
6) The ETMX and ITMX have very different bounce mode response: should add to our Vent Todo List. Double checked that the bounce/roll bandstop is on and at the right frequency for the bounce mode. Increased the stopband from 40 to 50 dB to see if that helps.
7) op340 is still running ! The only reason to keep it alive is its crontab:
op340m:SUS>crontab -l
07 * * * * /opt/rtcds/caltech/c1/burt/autoburt/burt.cron >> /opt/rtcds/caltech/c1/burt/burtcron.log
#46 * * * * /opt/rtcds/caltech/c1/scripts/general/scripto_cron /opt/rtcds/caltech/c1/scripts/PSL/FSS/FSSSlowServo > /cvs/cds/caltech/logs/scripts/FSSslow.cronlog 2>&1
#14,44 * * * * /cvs/cds/caltech/conlog/bin/check_conlogger_and_restart_if_dead
15,45 * * * * /opt/rtcds/caltech/c1/scripts/SUS/rampdown.pl > /dev/null 2>&1
#10 * * * * /opt/rtcds/caltech/c1/scripts/general/scripto_cron /opt/rtcds/caltech/c1/scripts/MC/autolockMCmain40m >/cvs/cds/caltech/logs/scripts/mclock.cronlog 2>&1
#27 * * * * /opt/rtcds/caltech/c1/scripts/general/scripto_cron /opt/rtcds/caltech/c1/scripts/PSL/FSS/RCthermalPID.pl >/cvs/cds/caltech/logs/scripts/RCthermalPID.cronlog 2>&1
00 0 * * * /var/scripts/ntp.sh > /dev/null 2>&1
#00 4 * * * /opt/rtcds/caltech/c1/scripts/RGA/RGAlogger.cron >> /cvs/cds/caltech/users/rward/RGA/RGAcron.out 2>&1
#00 6 * * * /cvs/cds/scripts/backupScripts.pl
00 7 * * * /opt/rtcds/caltech/c1/scripts/AutoUpdate/update_conlog.cron
00 8 * * * /opt/rtcds/caltech/c1/scripts/crontab/backupCrontab
added a new script (scripts/SUS/rampdown.py) which decrements every 30 minutes if needed. Added this to the megatron crontab and commented out the op340m crontab line. IF this works for awhile we can retire our last Solaris machine.
8) To see if we could get rid of the wandering PCDRIVE noise, I looked into the NPRO temperatures: was - T_crystal = 30.89 C, T_diode1 = 21 C, T_diode2 = 22 C. I moved up the crystal temp to 33.0 C, to see if it could make the noise more stable. Then I used the trimpots on the front of the controller to maximimize the laseroutput at these temperatures; it was basically maximized already. Lets see if there's any qualitative difference after a week. I'm attaching the pinout for the DSUB25 diagnostics connector on the back of the box. Aidan is going to help us record this stuff with AcroMag tech so that we can see if there's any correlation with PCDRIVE. The shifts in FSS_SLOW coincident with PCDRIVE noise corresponds to ~100 MHz, so it seems like it could be NPRO related.
|
11297
|
Mon May 18 09:50:00 2015 |
ericq | Update | General | some status |
Quote: |
Added this to the megatron crontab and commented out the op340m crontab line. IF this works for awhile we can retire our last Solaris machine.
|
For some reason, my email address is the one that megatron complains to when cron commands fail; since 11:15PM last night, I've been getting emails that the rampdown.py line is failing, with the super-helpful message: expr: syntax error |
11298
|
Mon May 18 11:59:07 2015 |
rana | Update | General | some status |
Yes - my rampdown.py script correctly ramps down the watchdog thresholds. This replaces the old rampdown.pl Perl script that Rob and Dave Barker wrote.
Unfortunately, cron doesn't correctly inherit the bashrc environment variables so its having trouble running.
On a positive note, I've resurrected the MEDM Screenshot taking cron job, so now this webpage is alive (mostly) and you can check screens from remote:
https://nodus.ligo.caltech.edu:30889/medm/screenshot.html |
11301
|
Mon May 18 16:28:18 2015 |
ericq | Update | General | some status |
Quote: |
4) Noticed that DAQD is restarting once per hour on the hour. Why?
|
It looks like daqd isn't being restarted, but in fact crashing every hour.
Going into the logs in target/fb/logs/old, it looks like at 10 seconds past the hour, every hour, daqd starts spitting out:
[Mon May 18 12:00:10 2015] main profiler warning: 1 empty blocks in the buffer
[Mon May 18 12:00:11 2015] main profiler warning: 0 empty blocks in the buffer
[Mon May 18 12:00:12 2015] main profiler warning: 0 empty blocks in the buffer
[Mon May 18 12:00:13 2015] main profiler warning: 0 empty blocks in the buffer
...
***CRASH***
An ELOG search on this kind of phrase will get you a lot of talk about FB transfer problems.
I noticed the framebuilder had 100% usage on its internal, non-RAID, non /frames/, HDD, which hosts the root filesystem (OS files, home directory, diskless boot files, etc), largely due to a ~110GB directory of frames from our first RF lock that had been copied over to the home directory. The HDD only has 135GB capacity. I thought that maybe this was somehow a bottleneck for files moving around, but after deleting the huge directory, daqd still died at 4PM.
The offsite LDAS rsync happens at ten minutes past the hour, so is unlikely to be the culprit. I don't have any other clues at this point. |
11303
|
Mon May 18 17:42:14 2015 |
rana, ericQ | Update | General | some status |
Today at 5 PM we replaced the east N2 cylinder. The east pressure was 500 and the west cylinder pressure was 1000. Since Steve's elogs say that the consumption can be as high as 800 per day we wanted to be safe.
- We closed the black valve before the regulator and closed the valve on the cylinder.
- We unscrewed the brass fill line to the cylinder.
- We unchained the cylinder and put in the dolly (and attached the chains on there).
- We rolled in a fresh cylinder from outside using the red dolly (it should have chains).
- We put it in place, hooked up the chains, and screwed on the brass nozzle with the large adjustable wrench (need to put a non-adjustable here).
- Opened up the cylinder valve.
- Opened up the black valve.
- New east pressure reading is 2500 PSI. Regulated N2 pressure is 68 PSI.
Quote: |
1) Checked the N2 pressures: the unregulated cylinder pressures are both around 1500 PSI. How long until they get to 1000?
|
|
11305
|
Mon May 18 18:03:12 2015 |
rana | Update | General | some status |
The c1cal model was maxing out its CPU meter so I logged onto c1lsc and did 'rtcds c1cal stop'. Let's see if this changes any of our FB / DAQD problems. |
11306
|
Tue May 19 00:19:23 2015 |
rana | Update | General | some status |
There's a few hours so far after today's c1cal shut off that the summary page shows no dropouts. I'm not yet sure that this is related, but it seems like a clue. |
11315
|
Tue May 19 18:55:12 2015 |
rana, ericQ | Update | General | some status |
After one day the pressures are east/west = 2200/450 PSI
Quote: |
- New east pressure reading is 2500 PSI. Regulated N2 pressure is 68 PSI.
|
|
11317
|
Wed May 20 03:08:27 2015 |
rana | Update | General | some status |
I think that the real clue was that the dropouts are in some channels and not in others:
https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20150519/psl/
As it turns out, the channel with no dropouts is the RAW PSL RMTEMP channel. All the others are the minute trends. So something is up with the trend making or the trend reading in the cluster.
Quote: |
There's a few hours so far after today's c1cal shut off that the summary page shows no dropouts. I'm not yet sure that this is related, but it seems like a clue.
|
|
11318
|
Wed May 20 11:41:59 2015 |
ericq | Update | General | some status |
West cyclinder is empty, east is at 2000 psi; regulated N2 pressure is 64psi. I'll replace the west one after the meeting. |
1222
|
Mon Jan 12 10:57:38 2009 |
rob | Update | General | some stuff |
The AS beam was not hitting the AS166 diode, so I aligned the last little steering mirror and adjusted the phase for MICH locking.
I turned on the HV supplies for the OMC.
Then I realigned the beam onto the AS166 diode, since the steering mirrors came on when I turned on the HV supplies.
It took awhile to find the alignment of the beam into the OMC. Once that was done, the output beam alignment was set, so I aligned onto the AS166 diode a third time.
The bottom two Sorensens in the OMC voltage supply don't look right. They have stickers that say +-24V, but each is sitting at 17.5V and showing no current draw. What's going on here? |
15237
|
Mon Mar 2 16:14:47 2020 |
gautam | Update | CDS | some target directory cleanup |
$TARGET_DIR = /cvs/cds/caltech/target
- $TARGET_DIR/c1psl and $TARGET_DIR/c1iool0 moved to $TARGET_DIR/preAcromag_oldVME/
- $TARGET_DIR/c1psl1 moved to $TARGET_DIR/c1psl
- $TARGET_DIR/c1psl/*.service and $TARGET_DIR/C1_PSL.cmd modified - i executed :%s/c1psl1/c1psl/g in vim.
- $TARGET_DIR/preAcromag_oldVME/c1psl/autoBurt.req and $TARGET_DIR/preAcromag_oldVME/c1iool0/autoBurt.req catenated into $TARGET_DIR/c1psl/autoBurt.req. The first snapshot at 16:19 has been verified.
It remains to (Jon is taking care of these)
add a line to modbusIOC.service on the new c1psl machine that restores the latest burt snapshot on startup (this necessitated installation of a debian jessie libXp6 package on our debian buster machine because our shared EPICS is soooooooooooooo oooooooold)
change the hostname from c1psl1 to c1psl
update martian.hosts
|
4458
|
Tue Mar 29 22:29:16 2011 |
kiwamu | Update | General | some tasks tomorrow |
* Temporary strain relief for the heliax cables on 1X2 (Steve)
* RF diagrams and check lists (Suresh)
=> In the lunch meeting we will discuss the details about what we will do for the RF installation.
* Electronics design and plan for Green locking (Aidan / Kiwamu)
=> In the lunch meeting we will discuss the details.
* LSC model (Koji)
* Video cable session (team)
* LPF for the laser temperature control (Larisa) |
11071
|
Wed Feb 25 23:48:57 2015 |
rana | Update | LSC | some thoughts |
- Comparing just the 2 cases with locking on 33, it seems that the 55 MHz gain has changed by 14 dB instead of the 10 dB that we expected. Is it that we need to measure the modulation drive change more carefully, or just that the PRMI was aligned differently?
- The 165 signal changed by a factor of 60 (35 dB) which is more consistent with a ~12 dB change in Gamma2, so not so far off.
- The fact that the whole sensing matrix increases in amplitude between 33 and 55 lock makes me think that either the alignment was very bad for the 33 lock, or that the 33 signals have a significant offset; if that's the case, then we should do as LLO and set the digital offsets in the 33 signals by locking first on 55.
- How does the REFL33 demod phase change by 70 deg?
|
5595
|
Sun Oct 2 02:33:32 2011 |
kiwamu | Update | LSC | something funny with AS55 |
Just a quick report.
The AS55 signal contains more noise than the REFL signals.
Why ? Is this the reason of the instability in PRMI ? 

I locked the Power-Precycled ITMY with REFL33.
As shown in the plot above, I compared the in-loop signal (REFL33) and out-of-loop signals (REFL11 and AS55).
All the signals are calibrated into the displacements of the PR-ITMY cavity by injecting a calibration peak at 283 Hz through the actuator of PRM.
AS55 (blue curve) showed a structure around 3 Hz and higher flat noise below 1 Hz.
Quote from #5582 |
I am suspecting that something funny (or stupid) is going on with the MICH control rather than the PRCL control.
|
|
1170
|
Wed Dec 3 12:49:11 2008 |
jenne | Update | Computers | something sketchy with NDS ... or something |
Never mind...I had forgotten that you have to run mdv_config every time you open matlab, not just every time you boot a computer.
I am not able to get channels using get_data from the mDV toolbox on Allegra, Megatron or Rosalba.
The error I get while running the "hello_world" test program is:
hello_world
setting up configuration...
added paths for nds
added paths for qscan
couldn't add path for matapps_SDE
couldn't add path for matapps_path
couldn't add path for framecache
couldn't add path for ligotools_matlab
added paths for home_pwd
fetching channels for C...
Warning: get_channel_list() failed.
??? Error using ==> NDS_GetChannels
Failed to get channel list.
Error in ==> fetch_nds at 47
eval(['CONFIG.chl.' server ' = NDS_GetChannels(ab);']);
Error in ==> get_data at 100
out = fetch_nds(channels,dtype,start_time,duration);
Error in ==> hello_world at 6
aa = get_data('C1:LSC-DARM_ERR', 'raw', gps('now - 1 hour'), 32); |
4898
|
Tue Jun 28 14:21:41 2011 |
kiwamu | Update | IOO | something wrong ? : Power incident on REFL11 and REFL55 |
The measured change in the REFL DC power with and without PRM aligned seems unacceptably small. Something wrong ?
The difference in the power with and without PRM aligned should be more than a factor of 300.
[difference in power] = [single bounce from PRM] / [two times of transmission through PRM ]
= (1-T) / T^2 ~ 310,
where T is the transmissivity of PRM and T = 5.5% is assumed in the calculation.
Also the reflectivity of MICH is assumed to be 1 for simplicity.
Quote from #4894 |
We now have (with the PRM misaligned):
REFL11: Power incident = 7.60 mW ; DC out = 0.330 V => efficiency = 0.87 A/W
REFL55: Power incident = 23 mW ; DC out = 0.850 V => efficiency = 0.74 A/W
and with the PRM aligned::
REFL11: DC out = 0.35 V => 8 mW is incident
REFL55: DC out = 0.975 V => 26 mW is incident
|
|
4954
|
Thu Jul 7 14:27:16 2011 |
Suresh | Update | IOO | something wrong ? : Power incident on REFL11 and REFL55 |
Just tying up a loose end. The next day Kiwamu and I checked to see what the trouble was. We concluded that the PRM had not moved during my measurement though I had 'Misaligned' it from the medm screen. So all the power levels measured here were with the PRM aligned. The power level change was subsequently measured and e-logged
Quote: |
The measured change in the REFL DC power with and without PRM aligned seems unacceptably small. Something wrong ?
The difference in the power with and without PRM aligned should be more than a factor of 300.
[difference in power] = [single bounce from PRM] / [two times of transmission through PRM ]
= (1-T) / T^2 ~ 310,
where T is the transmissivity of PRM and T = 5.5% is assumed in the calculation.
Also the reflectivity of MICH is assumed to be 1 for simplicity.
Quote from #4894 |
We now have (with the PRM misaligned):
REFL11: Power incident = 7.60 mW ; DC out = 0.330 V => efficiency = 0.87 A/W
REFL55: Power incident = 23 mW ; DC out = 0.850 V => efficiency = 0.74 A/W
and with the PRM aligned::
REFL11: DC out = 0.35 V => 8 mW is incident
REFL55: DC out = 0.975 V => 26 mW is incident
|
|
|
5025
|
Mon Jul 25 00:35:44 2011 |
rana | Update | SUS | something wrong with ETMY LR sensor |

Looks like either the LR OSEM is totally mis adjusted in its holder or the whitening eletronics are broken.
Also looks like the ETMY is just not damped at 1 Hz? How can this be?
I look at the SUS_SUMMARY screen which apparently only Steve and I look at:

Looks like the suspensions have factor of 10-100 different gains. Why?
** The ETMY just doesn't behave correctly when I bias it. Both pitch and yaw seem to make it do yaw. I leave this for Jamie to debug in the morning.
*** Also, the BIAS buttons are still broken - the HOPR/LOPR limits ought to be 5000 and the default slider increment be 100. Also the YAW button readback doesn't correctly show the state of the BIAS.
**** And.....we have also lost the DAQ channels that used to be associated with the _IN1 of the SUSPOS/PIT/YAW filter modules. Please put them back; our templates don't work without them. |
4269
|
Thu Feb 10 11:16:31 2011 |
steve | Update | PEM | south arm AC turned on |
The air condition was off for the south arm. I turned it on. |
3513
|
Thu Sep 2 14:11:17 2010 |
steve | Summary | PEM | south end crane balancing is completed |
The crane I -beam now leveled at all degrees of rotation. The lower hinge was moved southward about 1/4 of an inch. Performance was tested at 2000 lbs
Atm1, work in progress
Atm2, load test at 1 Ton
Atm3, service report |
12187
|
Thu Jun 16 11:10:17 2016 |
Steve | Update | SUS | spare ETMX optics preparation to be hanged |
D Location
Number on
Drawing
|
Component
Name
|
Baked Clean
Pieces Needed
|
Pieces
In Stock
(on tower )
|
Notes |
|
|
|
|
|
31,20,19,13 |
viton tips |
|
not cut yet |
baked material in stock |
30 |
6-32x0.75" stops |
4 |
(4) |
|
29 |
steel music wire 0.017" |
not baked |
on roll |
needs good wipes |
28 |
1/4 "washer |
4 |
(4) |
|
27 |
lock washer |
4 |
50 |
install |
26 |
Ag plated 1/4-20x1.25 |
4 |
(4) |
|
25 |
Ag plated1/4-25x.75 & not plated |
20 |
(20) |
|
24 |
SS 4-40x.5 |
2 |
(2) |
|
23 |
SS 4-40x.38 |
4 |
(4) |
|
22 |
spring plunger |
4+2 |
(4+2) |
|
|
|
|
|
|
|
|
|
|
|
18 |
magnets, 1.9mm od, length 3.2 mm |
5 |
~30 not coated, rusty |
buy Ni coated ones for future use from www.electroenergy.com
|
17 |
guide rod, 0.635 mm od, 3.3 mm |
3 |
6 |
Al |
16 |
wire standoff, 1 mm od, 4.8 mm |
2 |
2 |
Al and ruby (ruby groove not centered) |
15 |
short OSEMs |
5 |
6 |
|
14 |
spare ETMX in 40m wiki |
1 |
1 |
confirmed in cabinet |
|
|
|
|
|
12 |
dumbbell standoff |
5 |
6 |
|
11 |
Al stiffening plate |
1 |
(1) |
|
10 |
wire clamp B in sus block |
2 |
(2) |
|
9 |
wire clamp A |
1 |
(1) |
|
8-7 |
lower clamp for lifting optics |
2 |
(2) |
|
6 |
upper clamp to hold down optic |
1 |
(1) |
|
5-4 |
left-right side of tower |
1 ea |
(1ea) |
|
3 |
tower base |
1 |
(1) |
|
2 |
sus block |
1 |
(1) |
|
1 |
lower and upper OSEM holders |
1ea |
(1ea) |
|
|
|
|
|
|
48 |
sandind fixture for magnet&dumbbell |
|
|
|
45 |
magnet-dumbbell assemblly fixture |
1 |
1 |
|
43 |
guide-wirestandoff gluing fixture |
1 |
1 |
Ok for larger RUBY,
unit is not in perfect condtition but usefull
|
|
|
|
|
|
|
First contact |
|
3-15-2013 purchase |
Pick up FC from Gary with purchasing date 7-7-2015
or later
|
|
FCPEEK peeler ring disk for TM cleaning |
10 front,10 back side |
have sheets only |
32&19mm ID punches ordered |
|
GordonBrush custom for LIGO optical cleaning |
~5 |
1 (3/8wide nylonSS) |
more from Calum available |
|
EP30-2 epoxy |
have |
have |
expiration date 9-24-2016 |
NOT finished, last edited 7-7 |
13158
|
Wed Aug 2 09:40:55 2017 |
Steve | Update | Electronics | spare ILIGO electronics |
Spare ILIGO electronics temporarly stored in the east arm. We need cabinet space. |
1387
|
Wed Mar 11 16:41:22 2009 |
steve | Update | MOPA | spare NPRO power |
The spare M126N-1064-700, sn 5519 of Dec 2006 rebuilt NPRO's power output
measured 750mW at DC2.06A with Ohpir meter.
Alberto's controller unit 125/126-OPN-PS, sn516m was disconnected from lenght measurment NPRO on the AP table.
5519 NPRO was clamp to the optical table without heatsink and it was on for 15 minutes. |
12088
|
Mon Apr 25 11:07:06 2016 |
Steve | Update | SUS | spare SOS tower |
Earth quake stops need viton tips.
Wirestandoffs are still aluminum. |
12092
|
Wed Apr 27 09:45:56 2016 |
rana | Update | SUS | spare SOS tower |
Bah, we need ruby slippers for all future suspensions. Prism with curved backside and smooth grooves.
No aluminum, no cry.
Quote: |
Earth quake stops need viton tips.
Wirestandoffs are still aluminum.
|
|
7071
|
Wed Aug 1 16:38:46 2012 |
steve | Update | SUS | spare SRMU 03 optic moved |
SOS optics prepared to be hanged are moved from the South Flow Bench to S15 Clean Cabinet.
SRMU 03 (1-25-2010) specification summery E080460-05-D, older vintage SRM 01 and PRM 02 (need more specification)
There was one NOT MARKED SOS with two broken magnets on its face. This is labeled ???
This was done to prepare clean space for TIP-TILT drive- test set up.
The existing cable from 1X5 can reach only to the south end: from whitening filter to satelite amp. This will be good for future test of suspensions.
We need to make new cable from 1 X 1 to the south end = 40m long |
7072
|
Wed Aug 1 17:04:22 2012 |
Jenne | Update | SUS | spare SRMU 03 optic moved |
Quote: |
There was one NOT MARKED SOS with two broken magnets on its face. This is labeled ???
|
While I'm not sure what specific optic this is, I think it's an older optic. (a) All of the new optics we got from Ramin were enscribed with their #. (b) This optic appears to have a short arrow scribe line (about the length of the guiderod), and then no scribe line (that I could see through the glass dish) on the other side. The new optics all have a long arrow scribe line, ~1/2 the full width of the optic, and have clear scribe lines on the opposite side. |
7755
|
Tue Nov 27 17:45:42 2012 |
Steve | Update | General | spare optics of AP table moved to cabinet |
Spare optics from the AP table were moved to glass cabinet in the east arm. I'm not sure this is the right place. We'll see what everybody thinks.
There were two UNMARKED optics. Shame on you! No pencil marks on the optics either. These optics were shipped to the FBI for finger tip analysis.
|
632
|
Thu Jul 3 16:18:51 2008 |
rob | Summary | Locking | specgrams |
I used ligoDV to make some spectrograms of DARM_ERR (1), QPDX (2), and QPDY (3). These show the massive instability from 30-40Hz growing in the XARM in the last two minutes of a reasonably high power lock (arm powers up to 30). It's strange that it only shows up in one arm.
CARM is on PO-DC, for both the MCL and the AO path.
DARM is on AS166Q.
|
7517
|
Wed Oct 10 08:36:47 2012 |
Steve | Update | COC | special mirror mounts holder |
Quote: |
After looking at the in-vacuum layout I think we should make two changes during the next vent:
1) Reduce the number of mirrors between the FI and its camera. We install a large silvered mirror in the vacuum flange which holds the Faraday cam (in the inside of the viewport). That points directly at the input to the Faraday. We get to remove all of the steering mirror junk on the IO stack.
2) Take the Faraday output (IFO REFL) out onto the little table holding the BS and PRM Oplevs. We then relocate all 4 of the REFL RFPDs as well as the REFL OSA and the REFL camera onto this table. This will reduce the path length from the FI REFL port to the diodes and reduce the beam clutter on the AS table.
|
1) Mirror mount holder for "large silvered mirror" inside of the 8" OD tube vacuum envelope. |
1684
|
Thu Jun 18 23:08:46 2009 |
rob | Update | Locking | spectrum |
Here's a noise spectrum of the RSE interferometer, in anti-spring mode, with RF readout. I'd say the calibration is "loose."
I used the Buonanno & Chen modification of the KLMTV IFO transfer functions to model the DARM opto-mechanical response. I just guessed at the quadrature, and normalized the optical gain at the frequency of the calibration line used (927Hz, not visible on the plot). |
6183
|
Tue Jan 10 00:09:33 2012 |
kiwamu | Update | LSC | spike hunting in REFL33 |
[John / Kiwamu]
We tried to figure out what is causing spikes in the REFL33 signal, which is used to lock PRCL.
No useful information was obtained tonight and it is still under investigation.
(Background)
One thing preventing us from doing smooth measurements of the noise budget and the sensing matrix is some sharp spikes in the LSC error signals.
For example when we lock PRMI with REFL33 and AS55 fedback to PRCL and MICH respectively, both the REFL33 and AS55 signals show some spikes in time series.
Those spikes then bring the noise spectra higher than how they should be.
So for the reason, taking the noise budget doesn't give us much information about the interferometer rather than there are spikes.
Also the sensing matrix measurement has been suffered from those spikes, which excite the impulse responses of the low pass filters in the LOCKIN detection systems a lot.
(What we did)
We looked into the actual analog signals to see if there are indeed spikes or not before they are acquired to the ADCs.
But we didn't find any corresponding spikes in the signals that are after the mixers.
It maybe because the signals we looked into didn't have high enough SNR because they were coming out from the monitor lemo outputs on the demod boards.
Then we thought the spikes are from the whitening circuits, due to some kind of saturation.
We decreased the gain of the whitening filters by a factor of 10, but it didn't help and the spikes were still there. |
6327
|
Mon Feb 27 19:04:13 2012 |
jamie | Update | CDS | spontaneous timing glitch in c1lsc IO chassis? |
For some reason there appears to have been a spontaneous timing glitch in the c1lsc IO chassis that caused all models running on c1lsc to loose timing sync with the framebuilder. All the models were reporting "0x4000" ("Timing mismatch between DAQ and FE application") in the DAQ status indicator. Looking in the front end logs and dmesg on the c1lsc front end machine I could see no obvious indication why this would have happened. The timing seemed to be hooked up fine, and the indicator lights on the various timing cards were nominal.
I restarted all the models on c1lsc, including and most importantly the c1x04 IOP, and things came back fine. Below is the restart procedure I used. Note I killed all the control models first, since the IOP can't be restarted if they're still running. I then restarted the IOP, followed by all the other control models.
controls@c1lsc ~ 0$ for m in lsc ass oaf; do /opt/rtcds/caltech/c1/scripts/killc1${m}; done
controls@c1lsc ~ 0$ /opt/rtcds/caltech/c1/scripts/startc1x04
c1x04epics C1 IOC Server started
* Stopping IOP awgtpman ... [ ok ]
controls@c1lsc ~ 0$ for m in lsc ass oaf; do /opt/rtcds/caltech/c1/scripts/startc1${m}; done
c1lscepics: no process found
ERROR: Module c1lscfe does not exist in /proc/modules
c1lscepics C1 IOC Server started
* WARNING: awgtpman_c1lsc has not yet been started.
c1assepics: no process found
ERROR: Module c1assfe does not exist in /proc/modules
c1assepics C1 IOC Server started
* WARNING: awgtpman_c1ass has not yet been started.
c1oafepics: no process found
ERROR: Module c1oaffe does not exist in /proc/modules
c1oafepics C1 IOC Server started
* WARNING: awgtpman_c1oaf has not yet been started.
controls@c1lsc ~ 0$
|
5371
|
Fri Sep 9 16:16:59 2011 |
kiwamu | Update | IOO | spot positions on MC mirrors (in-vac) |
The spot positions on the MC mirrors were measured in the vacuum condition.
The obtained spot positions are quite bad and roughly at 2-3 mm level. We have to realign the beam axis and the MC mirrors.
|
Feb 26 2011 |
May 08 2011 |
Aug 2 2011 |
Aug 10 2011 (in air) |
Aug 14 2011 (in air) |
Aug 23 2011 (in air)
|
[NEW] Sep 9 2011 |
MC1 pit [mm] |
1.6 |
1.9 |
1.93 |
-0.858 |
-0.2 |
0.1354 |
-1.55 |
MC2 pit [mm] |
6.4 |
9.0 |
9.03 |
-0.844 |
-0.8 |
-0.2522 |
-2.28 |
MC3 pit [mm] |
1.4 |
2.0 |
2.01 |
-1.03 |
-0.1 |
-0.1383 |
-1.8 |
MC1 yaw [mm] |
-1.5 |
-1.7 |
-1.72 |
-0.847 |
-1.1 |
-1.0893 |
-0.02 |
MC2 yaw [mm] |
1.0 |
0.2 |
0.178 |
0.582 |
0.6 |
0.7122 |
3.52 |
MC3 yaw [mm] |
-1.3 |
-1.9 |
-1.87 |
-1.06 |
-1.1 |
-1.5587 |
-2.145 |
|
5373
|
Fri Sep 9 20:52:59 2011 |
kiwamu | Update | IOO | spot positions on MC mirrors (in-vac) |
The spot positions on the MC mirrors were readjusted.
All the amount of the off-center became smaller than 2 mm, which meet requirements of the beam clearance on the Faraday.
|
Feb 26 2011 |
May 08 2011 |
Aug 2 2011 |
Aug 10 2011 (in air) |
Aug 14 2011 (in air) |
Aug 23 2011 (in air)
|
Sep 9 2011 |
[NEW] Sep 9 2011 |
MC1 pit [mm] |
1.6 |
1.9 |
1.93 |
-0.858 |
-0.2 |
0.1354 |
-1.55 |
0.22 |
MC2 pit [mm] |
6.4 |
9.0 |
9.03 |
-0.844 |
-0.8 |
-0.2522 |
-2.28 |
-0.34 |
MC3 pit [mm] |
1.4 |
2.0 |
2.01 |
-1.03 |
-0.1 |
-0.1383 |
-1.8 |
-0.21 |
MC1 yaw [mm] |
-1.5 |
-1.7 |
-1.72 |
-0.847 |
-1.1 |
-1.0893 |
-0.02 |
-1.15 |
MC2 yaw [mm] |
1.0 |
0.2 |
0.178 |
0.582 |
0.6 |
0.7122 |
3.52 |
0.07 |
MC3 yaw [mm] |
-1.3 |
-1.9 |
-1.87 |
-1.06 |
-1.1 |
-1.5587 |
-2.145 |
-1.07 |
In order to improve the MC1-YAW and MC3-YAW spot positions, the angle of the incident beam has to be shifted by approximately 1/100 rad.
However it turned out to be very difficult to introduce such amount of angle only with the steering mirrors on the PSL table since we have to keep the same translation while changing the angle.
Quote from #5371 |
The obtained spot positions are quite bad and roughly at 2-3 mm level. We have to realign the beam axis and the MC mirrors.
|
|
529
|
Wed Jun 11 11:45:25 2008 |
steve | Update | PSL | ss trap works |
The trap works well at 3 W level. No back reflected beam coming out of the trap on low power
sensing card level. The back scattering was not measured. The trap is insensitive to small pointing variations.
The SS surface did not show any visible degradation after 16 hrs of 3w exposure at elliptical beam size 4x8 mm
It is ready to be placed into the 35 W beam. |
1372
|
Mon Mar 9 10:59:05 2009 |
Alan | Omnistructure | Computers | ssh agent on fb40m restarted for backup |
After the boot-fest, the nightly backup to Powell-Booth failed, and an automatic email got sent to me. I restarted the ssh agent, following the instructions in /cvs/cds/caltech/scripts/backup/000README.txt . |
2081
|
Mon Oct 12 17:14:39 2009 |
rob | Update | Locking | stability |
Last night, 2+ hour lock, probably broken by me driving too hard (DARM_EXC). |
11708
|
Fri Oct 23 09:55:50 2015 |
Steve | Update | LSC | stable days |
|
8212
|
Sat Mar 2 05:53:15 2013 |
yuta | Update | LSC | stable lock of PRMI |
I tuned alignment, gains and filters to achieve stable lock of PRMI.
It now locks quite stably with UGF of ~100 Hz. Measured power recycling gain at maximum is ~ 25.
Locking details:
== PRMI carrier ==
MICH: AS55_Q_ERR, AS55_PHASE_R = -12 deg, MICH_GAIN = -1, feedback to ITMX(-1),ITMY(+1)
PRCL: REFL55_I_ERR, REFL55_PHASE_R = 70 deg, PRCL_GAIN = 0.3, feedback to PRM
MICH servo is always on. PRCL loop turns on by trigger using POP DC. Boost filters and resonant gains turn on by triggers using POP DC.
Gain normalization was not used.
Openloop transferfunctions:
MICH: UGF ~90 Hz, phase margin ~40 deg
PRCL: UGF ~100 Hz, phase margin ~50 deg (cf. Fitted gain was same as half-PRC: elog #8053)

Power recycling gain:
POP DC when unlocked is 6, when locked is 2200-2500, and when dark is 0. So, power recycling gain is ~ 22 to 25. Value without any loss in PRMI is 45 (elog #6947). Alignment was pretty critical to achieve this recycling gain and stable lock.
There was oscillation at 630 Hz when locked, which is similar to the one we saw in POX11 (elog #8203).
Youtube:
AS(top left), POP(top right), REFL(bottom left), and ETMYT(bottom right). ETMY was mis-aligned, but you can see the beam at ETMYT after PRMI was carrier locked.
MICH/PRCL coupling:
I measured "sensing matrix" of PRMI by tickling PRM/ITMs/BS at 8.5 Hz and measuring 8.5 Hz peak height of AS55_Q, REFL55_I spectra during PRMI lock (attached is an example measurement of PRM). Below table is the result. AS55_Q has ~5% of sensitivity to PRCL compared with MICH. Also, BS introduces REFL55_I signal considerably. And also, there seems to be an imbalance in actuation efficiency between ITMX and ITMY.
actuation AS55_Q_ERR REFL55_I_ERR
ITMX +11.4 +0.80
ITMY +33.0 +1.06
BS +50.8 +1.90
PRM - 0.7 +1.05
AS clipping:
AS was clipped inside the vaccuum the other day(elog #8198). So, I tried to avoid AS clipping by aligning BS this morning. But it turned out that avoiding AS clipping by BS makes ITMX beam centering worse. Maybe we should center the beam on Yarm first next week.
Next:
- calculate expected PRMI recycling gain with loss, PR2 filpped
- beam centering on the arms
- IPANG, IPPOS, Y green, X green
- PRMI g-factor measurement |
8213
|
Sat Mar 2 14:52:02 2013 |
rana | Update | LSC | stable lock of PRMI |
Whereas the sensing matrix coefficients for ITMX/ITMY in REFL_I indicate an actuation imbalance, the disparity in the ITMX/ITMY to AS_Q elements does not. However, they do indicate why there is a PRM to AS_Q coupling at all.
I would recommend setting up the triggering so that the REFL & AS whitening is turned on AFTER lock acquisition and using a frequency of ~100-300 Hz for the sensing matrix measurement to fix these issues. |
5882
|
Sat Nov 12 02:46:13 2011 |
Den | Update | Adaptive Filtering | stacks and ground |
We measured the coherence between the seismometer near the MC2 stack and accelerometers on the vacuum tank where MC2 is. Because accelerometers produce small signals at low frequencies, which are comparable with adc noise, we amplified the accelerometer signal by a factor of 20. We could not do it more because though adc has 40 V range, the black box that follows the channel sockets can transmit only 2.5 V max amplitude signal. Probably, this was done because old adc accepted 2 V max amplitude.


We were able to found some coherence at 0.1-1 Hz though the accelerometer signal is rather noisy. So to consider stack as a noise source is still possible. This measurement should be better done with two seismometers, one on the floor, the other on the stack. From the figure we can also see that tilt affects the x and y seismometer signals from 0.1 Hz. Green line (z-component) is much lower that red and blue lines (x and y). Tilt affects on horisontal axes of the seismometer much more than on vertical.
What we also think about is that at low frequencies mirrors start to move approximately the same and seismometers can help us to figure out small reletive displacement of the mirrors which form the MC length. We can estimate the critical frequency by presenting the ground motion as interference of surface waves with different velocities and amplitudes. For only 1 wave we have for the relation of MC length to the seismometer read out ~sin (2*pi*f/v*L). f - frequency of the wave, v -speed, L - length between the mirrors. We can see that below 1 Hz we have ~sin (f/2). At this point seismometer signal could lose coherence with MC length signal. We could try to subtract seismometer signals from corresponding axes, but gur1 and sts1 has different calibrations. Moreover, the noise floor of the seismometers might not allow us to measure the differential signal. We'll try to simulate this scenario and find seismometer calibration or measure it. We are basicly interested only in the ratio of calibraion fucntions of 2 seismometers. |
3704
|
Wed Oct 13 09:35:41 2010 |
rana | Update | elog | start script edited |
The existing elog restart script was running the kill process in the background using the '&' symbol before starting new elog process. This is a BAD idea since there's no way to make sure that the background process has actually worked before the new one tries to start.
That's why you sometimes had to run the script twice. I've removed all of the background "cleverness" so now it will take ~2s more for the script to run - however, it now actually works. We may also upgrade from v2.7.5 to 2.8 today. |
4311
|
Thu Feb 17 11:20:04 2011 |
josephb | Update | CDS | start scripts no longer need sudo |
I've modified the rc.local file to run the IOC codes as controls, which means they no longer write root permission log files on startup.
The awgtpman, which was the other permission issue with the start scripts, is started by a run script now. This new version seems to be content to keep the permissions of the current log file, which is set to controls.
This should prevent the issue of sudo wiping your path environment variable for just that command. (Try "sudo which burtwb" versus "which burtwb" for example). This apparently a security feature of sudo.
If you should happen to use sudo to run a start script, the easiest solution to fix the permissions is just got to the target directory (type "target") and run "sudo chown controls:controls -R *" on one of the workstations (the front ends don't handle the groups properly at the moment).
This should allow the scripts to properly use burtrb and burtwb to write and backup burt files. |
4935
|
Sun Jul 3 21:18:06 2011 |
rana | Update | Computer Scripts / Programs | statScreen scripts dead since Feb 4 / now revived |
This CSHRC mangling on Feb 4 did more than re-arrange FB binaries.
It broke the path to MEDM for the 32-bit machines in the lab (e.g. mafalda) and stopped the MEDM snapshots from being posted onto our MEDM Status Web Page.
This is because, in addition to the paths mentioned in the above elog, the paths to the EPICS directories were also commented out. I've re-inserted them into our
.cshrc file in the 32-bit section; the statScreen CRON that Yoichi set up is now back in business.
* for some reason, the 'cronjob.sh' script is wiping out its own log file. It would be great if someone who understands stderr output re-direction can fix it so that the log-file from each run is retained until the next time cron runs. |