  5415   Thu Sep 15 07:28:08 2011 steveUpdateSUSPRM damping restored


   
        
            
  
  5434   Fri Sep 16 16:07:28 2011 steveUpdateSAFETYvisitors safety trainingPaul, Mirko and Katrin visiting grad students
received the 40m basic safety training.
 P1080241.JPG 
  5497   Wed Sep 21 11:35:07 2011 steveUpdateVACRGA scanRGA scan with maglev pumping speed at day
14 of the pump down.
The larger inserted box contains
 rgascanpd71m14d.jpg 
  5529   Fri Sep 23 16:25:01 2011 steveUpdateVACvac rack UPS battery replacedAPC Smart -UPS 2200   model: SUA2200RM2U
  batteries were replaced by compatible
RBC43, 8x  12V5A
 P1080252.JPGP1080254.JPG 
  5542   Mon Sep 26 11:35:44 2011 steveUpdateCamerasarms cameras upgradedThe arm's CCD cameras were reset as picture
shows last week.
The height adjustment only works
 P1080251.JPG 
  5562   Wed Sep 28 07:36:41 2011 steveUpdateSUSITMY damping restoredITMY suspention damping restored   
  5563   Wed Sep 28 07:46:42 2011 steveUpdateSUSDsub connections at the back of 1X5 are fixed 
 I'm turning  the sus damping 
off for Dsub connection fix at  the
  
  5587   Fri Sep 30 17:01:29 2011 steveUpdateSUSETMX oplev intensity effect on noiseKiwamu and Steve,
This is one of the better oplev set
up we have. Single bounce from TM, no mirror
 ETMXoplintens.jpgP1080267.JPG 
  5599   Mon Oct 3 08:38:21 2011 steveSummaryVACRecovery from the power shutdown is completed 
I failed to start Rana's favorit
anciant desktop computer at the vac rack.
 poweroutage.jpg 
  5610   Tue Oct 4 07:51:36 2011 steveUpdateCDSc1lsc and c1sus are still down 
 c1lsc and c1sus are still down.
Only ETMX and ETMY are damped
  
  5620   Wed Oct 5 11:33:25 2011 steveUpdateSUSPRM and BS oplev laser replaced 
JDSU 1103P died after 4 years of
service. It was replaced with new identical
  
  5622   Wed Oct 5 17:08:49 2011 steveUpdateSUSBS oplev spectraKiwamu and Steve,
The He/Ne oplev shows no coherece
so relative intensity noise is not limiting
 BSoplservON2.png 
  5630   Fri Oct 7 14:04:48 2011 steveUpdateSUSHe/Ne intensity noise effect on oplevsThe SRM bounce peak 18.33 Hz. Suresh helped
me to retune filter through Foton, but we
failed to remove it.
 7x 
  5645   Mon Oct 10 16:32:18 2011 steveUpdateSUSUL sensor of ETMY is recovered I lost UL osem voltage this morning
when I was checking the actual connection
at rack ETMY
 ETMY_UL.pngETMY_OSEM_UL.pngnoisyETMY_UL.png 
  5655   Wed Oct 12 08:43:30 2011 steveUpdateSUSITMX oplev improved a bit Atm2 is before optical path adjustment.
The idea was to remove possible clipping
in vacuum.
 ITMXoplev.pngITMXoplevservo_ON.png 
  5660   Thu Oct 13 14:23:09 2011 steveUpdateSUSITMX oplev with 3 mm beam on qpd I replaced the JDSU-Uniphase 1125P
by 1103P He/Ne laser. This new laser had
2.8 mW output yesterday. It degraded to 0.5
 ITMXoplev1103p#2.pngITMXoplev.png 
  5675   Mon Oct 17 07:57:24 2011 steveUpdateSUSETMX sus damping restored    
  5680   Mon Oct 17 17:07:30 2011 steveUpdateSUSETMX oplev returning beam od 3 mm ETMX oplev had 6 mm diameter beam
on the qpd.  I relayed the beam path
with 2 lenses  to get back  3 mm
 ETMX-oplev.jpg 
  5692   Wed Oct 19 08:34:16 2011 steveUpdateSUSSRM sus damping restoredSorry Kiwamu, I realized too late that
you were freeswigging. Hopefully 4 hrs was
enough.
 SRM.png 
  5693   Wed Oct 19 09:44:10 2011 steveUpdateSUSETMX oplev power spectrum


   
        
            
 ETMXoplev2lens.png 
  5701   Wed Oct 19 16:28:35 2011 steveUpdateGreen LockingETMY mechanical shutter is working


   
        
            
  
  5712   Thu Oct 20 12:43:19 2011 steveUpdateSAFETYrefilled first aid kits & their locationsFirst aid kits are located close vicinity
of entry doors and under circuit breaker
panels.
 8x 
  5719   Fri Oct 21 09:08:33 2011 steveUpdateSUSMC2 Misalignment Thinks to do before the NEXT realignment:
B,  tie 4 ancher bolts on table
legs to the floor
 mc2.jpg 
  5720   Fri Oct 21 10:32:14 2011 steveUpdatePSLPSL laser turned offIn order to move the emergency shut off
switch in room 103 I had to turn off the
2 W Innolight laser. This job will take an
  
  5721   Fri Oct 21 11:02:47 2011 steveUpdatePSLPSL laser turned ON


   
        
            
 laserofon.png 
  5728   Mon Oct 24 14:26:01 2011 steveUpdateGeneralno electricity on Nov. 19
    
       

            
  
  5741   Wed Oct 26 10:07:16 2011 steveUpdateGeneralno electricity ....... will be recheduled

    

       
            
  
  5759   Fri Oct 28 18:33:59 2011 steveUpdateVACleaking nitrogen line fixedI was lucky to notice that the nitrogen
supply line to the vacuum valves was leaking.
Closed ALL valves. Open supply line to atm.
 allvalvesclosed.png 
  5767   Mon Oct 31 08:55:19 2011 steveUpdateVACpressure plot at day 53


   
        
            
 pd71-m-d53.png 
  5768   Mon Oct 31 09:42:12 2011 steveUpdateIOOPMC lockedThe PMC HV drops off more offen lately.  pmcHV.png 
  5771   Mon Oct 31 15:58:25 2011 steveUpdateSUSthinking of black glass baffles for 40mShade 12 black glass-green lines: 1/8"
thick,  6" wide  x 7 "
high  mounted on angle bracket below. 
 03260901.PDF03260902.PDF 
  5782   Wed Nov 2 10:04:05 2011 steveUpdateSUSMC2 chamber anchoring tightened


   
        
            
  
  5783   Wed Nov 2 10:41:47 2011 steveUpdateGeneral2 spare heliax at the LSC rackThey were traced and labeled. One goes
to 1X2 and the other to AS-ISCT. They are
Andrew Heliax 1/4" od. made by CommScone, 
  
  5826   Mon Nov 7 08:08:24 2011 steveUpdateIOOPMC lockedThe PMC acted like it was sleeping. The
HV slider was dead. The MC locked instantly
as the PMC had transmission.
 pmcHV.png 
  5827   Mon Nov 7 08:25:15 2011 steveUpdateGeneralno electricity on Nov. 20The
Nov.19 interruption was rescheduled to be
on Nov.20,2011
  
  5831   Mon Nov 7 14:58:17 2011 steveUpdateGeneralChamber illuminator remote switches installed

    

       
            
  
  5844   Wed Nov 9 07:57:39 2011 steveUpdateSUSETMX oplev is down and sus damping restoredThe ETMX oplev returning beam is well centered
on the qpd. We lost this signal 2 days ago.
I will check on the qpd. 
 etmxtripped.pngetmxOpl.png 
  5887   Mon Nov 14 15:46:14 2011 steveUpdateVACTP2's forepump changedThe foreline pressure of TP2 was 1.4 Torr
this morning. This drypump worked well for
ten months.
 drypump.png 
  5897   Tue Nov 15 16:16:34 2011 steveUpdatePSLenclosure interlocks are working on all sliding doorsBen and Sam came over to fix one of 
the east side  sliding  door sensor
that had to be moved from the inside 
  
  5898   Tue Nov 15 16:25:38 2011 steveUpdatePSLIOO angle qpd centeredThis moring I centered the IOO Angle QPD.
The IOO Pos QPD was not centered. The existing
layout does not allow the beam centering
  
  5907   Wed Nov 16 10:11:20 2011 steveUpdatePSLIOO angle & pos qpd centered


   
        
            
 iooqpds.png 
  5909   Wed Nov 16 10:25:57 2011 steveUpdateSUSPRM damping restoredThe  PRM lost damping about a day
ago. It was restored.
  
  5912   Wed Nov 16 14:34:18 2011 steveUpdatePSLIOO beam moves in pitchC1:IOO-QPD_ANG_VERT beam movement 
in 1 degree C temp change in 3 hrs
 
 iooqpds.png 
  5927   Thu Nov 17 15:19:06 2011 steveUpdateSUSTi spring plunger to hold OSEM is not affortableOur existing 300 series SS plungers from
McMastercar #8476A43 are silver plated as
Atm2 shows.
 20111116111042405.pdfP1080216.JPG 
  5950   Fri Nov 18 16:37:14 2011 steveUpdateVACpreparing for ac power interruptionThe vacuum is ready for no AC power for
1 hr on Sunday morning at 10am
 
 planednopower.png 
  5962   Sun Nov 20 13:51:54 2011 steveUpdateVACvacuum is back up after ac interruptionState condition: Vac Normal,         
CC1 ~8e-6 Torr
The IFO pressure peaked at 
 notpumping.png 
  6025   Mon Nov 28 15:43:36 2011 steveUpdateRF SystemEOM temp zoomed


   
        
            
 eomtempmon.png 
  6036   Tue Nov 29 15:25:29 2011 steveUpdateRF SystemEOM temp stabilization performance It is not obvious what is working.
 
 2days.png 
  6058   Thu Dec 1 11:25:10 2011 steveUpdatePEM40m infrastructure holds up well in strong wind condition Santa Anna wind speed was locked
around 60 kmph last night on campus. The
strongest in 30 years.  The lab hold
 santaannawind.pngwind.png 
  6120   Wed Dec 14 14:40:53 2011 steveUpdateGreen Lockingdelay line bnc cable specsThe existingly used used Pasternack Enterprices
RG58 C/U cable lenght ~ 140 ft and the specs
are here at Atm1
 rg58cu.jpgRG58-P.pdf 