  15626   Wed Oct 14 17:03:55 2020 anchalSummaryALSALS noise budget update - Added whitening filter for ADC

Koji recommended that I can add whitening filters to suppress ADC noise easily. I added a filter before ADC in ALS loop with 4 zeros at 1.5 Hz and 4 poles at 100 Hz and added a reversed filter in the digital filter of ALS. This did not change the performance of the loop but significantly reduced the contribution of ADC noise above 1 Hz. One can see ALS_controls.yaml for the filter description. Please let me know if this does not make sense or there is something that I have overlooked.

Now, the dominant noise source is DFD noise below 100 Hz and green laser frequency noise above that. For DFD noise, I used data dating back to Kiwamu's paper. The noise contribution from DFD in the model is lower than the latest measured ALS noise budget post on elog. I'll look further into design details and noise of DFD.


Code, data, and schematics

  15751   Wed Jan 6 22:47:41 2021 gautamUpdateALSNoisy ALS

Summary:

I want to get back to locking the interferometer so I can test out the newly installed AS WFS. However, the ALS noise is far too high, at least the transition of arm length control from IR to ALS fails reliably with the same settings that worked so reliably previously. I worked on investigating it a bit today.

Timeline

In the latter half of last year, I was focused on the air-BHD setup, so I wasn't checking in on the ALS noise as regularly. 

  1. On Aug 17, the noise was fine.
  2. But on Oct 29, the noise is bad (and it continues to remain so, to the point where I cannot lock the interferometer). 
  3. Koji and Anchal confirmed nothing was touched while they were investigating the ALS system, also on Oct 29. The spectra attached in #15650 don't make any sense to me, the noise at 100 Hz cannot be <100mHz/rtHz. So, inconclusive.

Excess noise:

All tests are done with the arm cavity length locked to the PSL frequency using POX. Then, the EX laser is locked to the arm cavity length using the AUX PDH servo. The fluctuations in the beatnote between the two lasers is what is monitored as a diagnostic. See Attachment #1. The reference traces in the top pane are from a known good time. The large excess noise between ~80-200 Hz is what I'm concerned about.

A separate issue that can improve the noise is to track down the noise in the 20-80 Hz band - probably some IMC frequency noise issue.

Noise budget:

See Attachment #2

  • I am pretty confident the electronics after the beat mouth are not to blame - I injected a 50 MHz signal from a Marconi and adjusted the signal amplitude to mimic what we get from the beat mouth. The trace labelled "DFD electronics noise" is the noise in this config.
  • The unsuppressed AUX frequency noise was measured with an SR785 (converted to freqnecy noise units knowing the PDH horn-to-horn voltage and the cavity linewidth). I didn't confirm the sensing noise level (dark noise of the AUX PDH loop), but I figure that at 100 Hz (voltage noise of ~100 uV/rtHz on the SR785), we are above the sensing noise level, and so are truly measuring the in-loop frequency noise of the stabilized AUX laser. I also confirmed that the loop UGF was ~10 kHz and phase margin was ~60 degrees, which are nominal numbers.
  • The fact that the excess noise is only in the X arm channel means the PSL frequency is not to blame.

So what could it be? The only things I can think of are (i) the beat mouth photodiode (NF1611) or (ii) excess noise in the fiber carrying the light from EX to the PSL table (but only on this fiber, and not on the EY fiber). Both seem remote to me - I'll test the former by switching the EX and EY fiber inputs to the beat mouth, but apart from this, I'm out of ideas... 

To avoid this kind of issue, we should really have scripted locks of all the basic IFO configs and record the data to summary pages or something - maybe something to do once Guardian is installed, it'd be pretty hacky to do cleanly with shell scripts.

  15752   Thu Jan 7 19:16:11 2021 gautamUpdateALSNoisy ALS

I'm also wondering why the error monitors for the X and Y loops report such wildly different spectra for the suppressed frequency noise of the AUX laser relative to the cavity length, see Attachment #1. The y-axis should be approximately Hz/rtHz. In both cases, the servo's error point monitor is connected to the DAQ system via a G=10 SR560. With the SR785, I measure for EX a nice bucket-shaped spectrum, bottoming out at ~10 uV/rtHz around 40 Hz, see Attachment #2. The SR560 should have an input-referred noise much less than this at the G=10 setting. The ADC noise level is only ~5 uV/rtHz, and indeed, the EY spectrum shows the expected shape. So what's up with the EX error mon? Tried swapping out the SR560 for a different unit, no change. And both the SR560 noise, and the ADC noise, check out when everything is checked individually. So some kind of interaction once everything is connected together, but it's only present at EX...

Today, I tried switching the EX and EY fibers going into the beat mouth, but I preserved the channel mapping after the beat mouth by switching the electrical outputs as well (the goal was to make sure that the beat photodiodes weren't the issue here, I think the electronics are already exonerated since driving the channel with a Marconi doesn't produce these noisy features). The EX spectrum remains noisy. I've switched everything back to the nominal configuration now to avoid further confusion. So it would appeat that this is real frequency noise that gets added in the EX fiber somehow. What can I do to fix this? The source of coupling isn't at the PSL table, else the EY channel would also show similar features. Visually, nothing seems wrong to me at EX either. So the problem is somehow in the cable tray along which the 40m of fiber is routed? This is already inside some nice foam/tubing setup, what can be done to improve it? Still doesn't explain why it suddenly became noisy...

  15753   Thu Jan 7 20:07:27 2021 KojiUpdateALSNoisy ALS

How about resurrecting the PSL table green beat for the X arm to see if the non-fiber setup shows the same level of the freq noise (e.g. the PDH locking became super noisy due to misalignment etc).

  15754   Thu Jan 7 21:16:22 2021 gautamUpdateALSNoisy ALS

I thought about it, but wouldn't that show up at the AUX PDH error point? Or because the loop gain is so high there we wouldn't see a small excess? I suppose there could be some clipping on the Faraday or something like that. But the GTRX level and the green REFL DC level when locked are nominal.

Quote:

How about resurrecting the PSL table green beat for the X arm to see if the non-fiber setup shows the same level of the freq noise (e.g. the PDH locking became super noisy due to misalignment etc).

  15755   Thu Jan 7 23:25:19 2021 KojiUpdateALSNoisy ALS

If the sensing noise level of the end PDH degraded for some reason, it'd make the out of loop stability worse without making the end pdh error level degraded.
It's just speculation.

 

  15756   Fri Jan 8 20:01:11 2021 gautamUpdateALSNoisy ALS

I did this test today. The excess noise around 100 Hz doesn't show up in the green beat.

See Attachment #1. The setup was as usual:

  • X-Arm cavity length stabilized to PSL frequency using the POX locking loop.
  • EX laser frequency locked to the X-Arm cavity length using the AUX PDH loop.
  • The "BEATX" channel records frequency fluctuations in the beat sensed on the IR beat photodiode, while the "BEATY" channel records frequency fluctuations in the beat sensed on the Green beat photodiode.
  • Since the green beat frequency fluctuations are twice that of the IR beat frequency fluctuations, I scaled the former ASD by a factor of 0.5 so as to compare apples to apples.
  • At low frequencies, the green beat is noisier, but that channel doesn't show the excess noise at mid frequencies you see in the IR beat. So the AUX PDH sensing noise is not to blame I think.

So, this excess appears to truly be excess phase noise on the fiber (though I have no idea what the actual mechanicsm could be or what changed between Aug and Oct 2020 that could explain it. Maybe the HEPA?

For this work, I had to spend some time aligning the two green beams onto the beat photodiode. During this time, I shuttered the PSL, disabled feedback via the FSS servo, turned the HEPA high, and kept the EX green locked to the arm so as to have a somewhat stable beat signal I could maximize. Everything has been returned to nominal settings now (obviously, since I locked the arms to get the data).


You may ask, why do we care. In terms of RMS frequency noise, it would appear that this excess shouldn't matter. But in all my trials so far, I've been unable to transition control of the arm cavity lengths from POX/POY to ALS. I suppose we could try using the green beat, but that has excess low frequency noise (which was the whole point of the fiber coupled setup). 

Quote:

How about resurrecting the PSL table green beat for the X arm to see if the non-fiber setup shows the same level of the freq noise (e.g. the PDH locking became super noisy due to misalignment etc).

  15860   Wed Mar 3 23:23:58 2021 gautamUpdateALSArm cavity scan

I see no evidence of anything radically different from my PSL table optical characterization in the IMC transmitted beam, see Attachment #1. The lines are just a quick indicator of what's what and no sophisticated peak fitting has been done yet (so the apparent offset between the transmission peaks and some of the vertical lines are just artefacts of my rough calibration I believe). The modulation depths recovered from this scan are in good agreement with what I report in the linked elog, ~0.19 for f1 and ~0.24 for f2. On the bright side, the ALS just worked and didn't require any electronics fudgery from me. So the mystery continues.

  16161   Tue May 25 17:42:11 2021 Anchal, PacoSummaryALSALS Single Arm Noise Budget

Here is our first attempt at a single-arm noise budget for ALS.

Attachment 1 shows the loop diagram we used to calculate the contribution of different noises.

Attachment 2 shows the measured noise at C1:ALS-BEATX_PHASE_FINE_OUT_HZ when XARM was locked to the main laser and Xend Green laser was locked to XARM.

  • The brown curve shows the measured noise.
  • The black curve shows total estimated noise from various noise sources (some of these sources have not been plotted as their contribution falls off the plotting y-lims.)
  • The residual frequency noise of Xend green laser (AUX) is measured by measuring the PDH error monitor spectrum from C1:ALS-X_ERR_MON_OUT_DQ. This measurement was converted into units of V by multiplying it by 6.285e-4 V/cts. This factor was measured by sending a 43 Hz 100 mV sine wave at the readout point and measuring the output in the channel.
  • This error signal is referred to AUX_Freq input in the loop diagram (see attachment 1) and then injected from there.
  • All measurements were taken to Res_Disp port in the 'Out-of-Loop Beat Note' block (see attachment 1).
  • In this measurement, we did not DAC noise that gets added when ALS loop is closed.
  • We added ADC noise from Kiwamu's ALS paper after referring it to DFD input. DFD noise is also taken from Kiwamu's ALS paper data.

Inference:

  • Something is wrong above 200 Hz for the inclusion of AUX residual displacement noise. It is coming off as higher than the direct measured residual noise, so something is wrong with our loop diagram. But I'm not sure what.
  • There is a lot of unaccounted noise everywhere from 1 Hz to 200 Hz.
  • Rana said noise budget below 1 Hz is level 9 stuff while we are at level 2, so I'll just assume the excess noise below 1 Hz is level 9 stuff.
  • We did include seismic noise taken from 40m noise budget in 40m/pygwinc. But it seems to affect below the plotted ylims. I'm not sure if that is correct either.

Unrelated questions:

  • There is a slow servo feeding back to Green Laser's crystal temperature by integrating PZT out signal. This is OFF right now. Should we keep it on?
  • The green laser lock is very unreliable and it unlocks soon after any signal is being fed back to the ETMX position.
  • This means, keeping both IR and green light locked in XARM is hard and simultaneous oscillation does not last longer than 10s of seconds. Why is it like this?
  • We notice that multiple higher-order modes from the green laser reach the arm cavity. The HOMs are powerful enough that PDH locks to them as well and we toggle the shutter to come to TEM00 mode. These HOMs must be degrading the PDH error signal. Should we consider installing PMCs at end tables too?
  16164   Thu May 27 11:03:15 2021 Anchal, PacoSummaryALSALS Single Arm Noise Budget

Here's an updated X ARM ALS noise budget.

Things to remember:

  • Major mistake we were making earlier was that we were missing the step of clicking  'Set Phase UGF' before taking the measurement.
  • Click the clear phase history just before taking measure.
  • Make sure the IR beatnotes are less than 50 MHz (or the left half of HP8591E on water crate). The DFD is designed for this much beatnote frequency (from Gautum).
  • We took this measurement with old IMC settings.
  • We have saved a template file in users/Templates/ALS/ALS_outOfLoop_Ref_DQ.xml . This si same as ALS_outOfLoop_Ref.xml except we changed all channels to _DQ.

Conclusions:

  • Attachment 1 shows the updated noisebudget. The estimated and measured RMS noise are very close to eachother.
  • However, there is significant excess noise between 4 Hz and 200 Hz. We're still thinking on what could be the source of these.
  • From 200 Hz to about 3 kHz, the beatnote noise is dominated by AUX residual frequency noise. This can be verified with page 2 of Attachment 2 where coherence between AUX PDH Error signal and BEATX signal is high.
  • One mystery is how the measured beatnote noise is below the residual green laser noise above 3 kHz. Could this be just because the phase tracker can't measure noise above 3kHz?
  • We have used estimated open loop transfer function for AUX from poles/zeros for uPDH box used (this was done months ago by me when I was working on ALS noise budget from home). We should verify it with a fresh OLTF measurement of AUX PDH loop. That's next on our list.
  16168   Fri May 28 17:32:48 2021 AnchalSummaryALSSingle Arm Actuation Calibration with IR ALS Beat

I attempted a single arm actuation calibration using IR beatnote (in the directions of soCal idea for DARM calibration)


Measurement and Inferences:

  • I sent 4 excitation signals at C1:SUS-ITM_LSC_EXC wit 30cts at 31Hz, 200cts at 197Hz, 600cts at 619Hz and 1000cts at 1069 Hz.
  • These were sent simultaneously using compose function in python awg.
  • The XARM was locked to mai laser and alignment was optimized with ASS.
  • The Xend Green laser was locked to XARM and alignment was optimized.
    • Sidenote: GTRX is now normalized to give 1 at near maximum power.
    • Green lasers can be locked with script instead of toggling.
    • Script can be called from sitemap->ALS->! Toggle Shutters->Lock X Green
    • Script is present at scripts/ALS/lockGreen.py.
  • C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ was measured for 60s.
  • Also, measured C1:LSC-XARM_OUT_DQ and C1:SUS-ITMX_LSC_OUT_DQ.
  • Attachment 1 shows the measured beatnote spectrum with excitations on in units of m/rtHz.
  • It also shows resdiual displacement contribution PSD of (output referred) XARM_OUT and ITMX_LSC_OUT to the same point in the state space model.
    • Note: that XARM_OUT and ITMX_LSC_OUT (excitation signal) get coherently added in reality and hence the beatnote spectrum at each excitation frequency is lower than both of them.
    • The remaining task is to figure out how to calculate the calibration constant for ITMX actuation from this information.
    • I need more time to understand the mixture of XARM_OUT and ITMX_LSC_OUT in the XARM length node in control loop.
    • Beatnote signal tells us the actual motion of the arm length, not how much ITMX would have actuated if the arm was not locked.
  • Attachment 2 has the A,B,C,D matrices for the full state space model used. These were fed to python controls package to get transfer functions from one point to another in this MIMO.
    • Note, that here I used the calibration of XARM_OUT we measured earlies in 16127.
    • On second thought, maybe I should first send excitation in ETMX_LSC_EXC. Then, I can just measure ETMX_LSC_OUT which includes XARM_OUT due to the lock and use that to get calibration of ETMX actuation directly.

  16171   Tue Jun 1 16:55:32 2021 Anchal, PacoSummaryALSSingle Arm Actuation Calibration with IR ALS Beat

Rana suggested in today's meeting to put in a notch filter in the XARM IR PDH loop to avoid suppressing the excitation line. We tried this today first with just one notch at 1069 Hz and then with an additional notch at 619 Hz and sent two simultaneous excitations.


Measurement and Analysis:

  • We added notch filters with Q=10, depth=50dB, freq=619 Hz and 1069 Hz using foton in SUS-ETMX_LSC filter bank at FM10.
  • We sent excitation signals with amplitudes 600cts and 1000 cts for 619 Hz and 1069 Hz signals respectively.
  • We measured time series data of C1:SUS-ITMX_LSC_OUT_DQ and C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ for 60s.
  • Then, spectrum of both signals is measured with Hanning window using scipy.welch function with scaling set to  'spectrum', binwidth=1Hz.
  • The beatnote signal was converted into length units by multiplying it by 1064nm * 37.79m / c.
  • The ratio of the two spectrums at teh excitation frequency multiplies by excitation frequency squared gives us teh calibration constant in units of nm Hz^2/cts.
  • At 619 Hz, we got \frac{5.01}{f^2}nm/cts
  • At 1069 Hz, we got \frac{5.64}{f^2}nm/cts.
  • The calibration factor in use is from \frac{7.32}{f^2} nm/cts from 13984.
  • So, the calibration factor from this methos is about 23% smaller than measured using freeswinging MICH in 13984.
  • One possiblity is that our notch filter is not as effective in avoiding suppresion of excitation.
    • We tried increasing the notch filter depths to 100 dB but got the same result within 2%.
    • We tried changing the position of notch filters. We put them in POX filter banks. Again the result did not change more than 2%.
  • The open loop gain of green PDH at 619 Hz and 1069 Hz must be large enough for our assumption of green laser perfectly following length motion to be true. The UGF of green laser is near 11 kHz.
  • The discrepancy could be due to outdated freeswinging MICH measurement that was done 3 years ago. Maybe we should learn how to do the ITMX calibration using this method and compare our own two measurements.
  16192   Tue Jun 8 11:40:53 2021 Anchal, PacoSummaryALSSingle Arm Actuation Calibration with IR ALS Beat

We attempted to simulate "oscillator based realtime calibration noise monitoring" in offline analysis with python. This helped us in finding about a factor of sqrt(2) that we were missing earlier in 16171. we measured C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ when X-ARM was locked to main laser and Xend green laser was locked to XARM. An excitation signal of amplitude 600 was setn at 619 hz at C1:ITMX_LSC_EXC.

Signal analysis flow:

  • The C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ is calibrated to give value of beatntoe frequency in Hz. But we are interested in the fluctuations of this value at the excitation frequency. So the beatnote signal is first high passed with 50 hz cut-off. This value can be reduced a lot more in realtime system. We only took 60s of data and had to remove first 2 seconds for removing transients so we didn't reduce this cut-off further.
  • The I and Q demodulated beatntoe signal is combined to get a complex beatnote signal amplitude at excitation frequency.
  • This signal is divided by cts amplitude of excitation and multiplied by square of excitation frequency to get calibration factor for ITMX in units of nm/cts/Hz^2.
  • The noise spectrum of absolute value of  the calibration factor is plotted in attachment 1, along with its RMS. The calibration factor was detrended linearly so the the DC value was removed before taking the spectrum.
  • So Attachment 1 is the spectrum of noise in calibration factor when measured with this method. The shaded region is 15.865% - 84.135% percentile region around the solid median curves.

We got a value of \frac{7.3 \pm 3.9}{f^2}\, \frac{nm}{cts}.  The calibration factor in use is from \frac{7.32}{f^2} nm/cts from 13984.

Next steps could be to budget this noise while we setup some way of having this calibration factor generated in realitime using oscillators on a FE model. Calibrating actuation of a single optic in a single arm is easy, so this is a good test setup for getting a noise budget of this calibration method.

  16196   Wed Jun 9 18:29:13 2021 Anchal, PacoSummaryALSCheck for saturation in ITMX SOS Driver

We did a quick check to make sure there is no saturation in the C1:SUS-ITMX_LSC_EXC analog path. For this, we looked at the SOS driver output monitors from the 1X4 chassis near MC2 on a scope. We found that even with 600 x 10 = 6000 counts of our 619 Hz excitation these outputs in particular are not saturating (highest mon signal was UL coil with 5.2 Vpp). In comparison, the calibration trials we have done before had 600 counts of amplitude, so we can safely increase our oscillator strength by that much yes


Things that remain to be investigated -->

  • What is the actual saturation level?
  • Two-tone intermodulation?
  16242   Fri Jul 9 15:39:08 2021 AnchalSummaryALSSingle Arm Actuation Calibration with IR ALS Beat [Correction]

I did this analysis again by just doing demodulation go 5s time segments of the 60s excitation signal. The major difference is that I was not summing up the sine-cosine multiplied signals, so the error associated was a lot more. If I simply multpy the whole beatnote signal with digital LO created at excitation frequency, divide it up in 12 segments of 5 s each, sum them up individually, then take the mean and standard deviation, I get the answer as:
\frac{6.88 \pm 0.05}{f^2} nm/ctsas opposed to \frac{7.32 \pm 0.03}{f^2} nm/ctsthat was calculated using MICH signal earlier by gautum in 13984.

Attachment 1 shows the scatter plot for the complex calibration factors found for the 12 segments.

My aim in the previous post was however to get a time series of the complex calibration factor from which I can take a noise spectral density measurement of the calibration. I'll still look into how I can do that. I'll have to add a low pass filter to integrate the signal. Then the noise spectrum up to the low pass pole frequency would be available. But what would this noise spectrum really mean? I still have to think a bit about it. I'll put another post soon.

Quote:

We attempted to simulate "oscillator based realtime calibration noise monitoring" in offline analysis with python. This helped us in finding about a factor of sqrt(2) that we were missing earlier in 16171. we measured C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ when X-ARM was locked to main laser and Xend green laser was locked to XARM. An excitation signal of amplitude 600 was setn at 619 hz at C1:ITMX_LSC_EXC.

Signal analysis flow:

  • The C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ is calibrated to give value of beatntoe frequency in Hz. But we are interested in the fluctuations of this value at the excitation frequency. So the beatnote signal is first high passed with 50 hz cut-off. This value can be reduced a lot more in realtime system. We only took 60s of data and had to remove first 2 seconds for removing transients so we didn't reduce this cut-off further.
  • The I and Q demodulated beatntoe signal is combined to get a complex beatnote signal amplitude at excitation frequency.
  • This signal is divided by cts amplitude of excitation and multiplied by square of excitation frequency to get calibration factor for ITMX in units of nm/cts/Hz^2.
  • The noise spectrum of absolute value of  the calibration factor is plotted in attachment 1, along with its RMS. The calibration factor was detrended linearly so the the DC value was removed before taking the spectrum.
  • So Attachment 1 is the spectrum of noise in calibration factor when measured with this method. The shaded region is 15.865% - 84.135% percentile region around the solid median curves.

We got a value of \frac{7.3 \pm 3.9}{f^2}\, \frac{nm}{cts}.  The calibration factor in use is from \frac{7.32}{f^2} nm/cts from 13984.

Next steps could be to budget this noise while we setup some way of having this calibration factor generated in realitime using oscillators on a FE model. Calibrating actuation of a single optic in a single arm is easy, so this is a good test setup for getting a noise budget of this calibration method.

 

  16333   Wed Sep 15 23:38:32 2021 KojiUpdateALSALS ASX PZT HV was off -> restored

It was known that the Y end ALS PZTs are not working. But Anchal reported in the meeting that the X end PZTs are not working too.

We went down to the X arm in the afternoon and checked the status. The HV (KEPCO) was off from the mechanical switch. I don't know this KEPCO has the function to shutdown the switch at the power glitch or not.
But anyway the power switch was engaged. We also saw a large amount of misalignment of the X end green. The alignment was manually adjusted. Anchal was able to reach ~0.4 Green TRX, but no more. He claimed that it was ~0.8.

We tried to tweak the SHG temp from 36.4. We found that the TRX had the (local) maximum of ~0.48 at 37.1 degC. This is the new setpoint right now.

  16884   Wed Jun 1 11:56:28 2022 yutaUpdateALSShutter driver for GRY replaced

[JC, Yuta]

We replaced a shutter driver for GRY since it stopped working this morning.
We replaced it with a free driver which was sitting on the ITMY table.
The reverse polarity issue of C1:AUX-GREEN_Y_Shutter was fixed by switching one of the switches of the driver from N.O. to N.C.

Also, "Toggle" button was added to IFO_ALIGN.adl so that we can toggle shutters easily to find TEM00. It runs /home/controls/Git/40m/scripts/ALS/ShutterToggler.py.
 

Quote:
 

The green Y shutter now works but in reverese, meaning that sending 1 to C1:AUX-GREEN_Y_Shutter closes the shutter and vice versa. This needs to be fixed.

 

  16945   Fri Jun 24 17:16:59 2022 PacoUpdateALSXAUX cable in control room

[JC, Paco]

We took the long BNC cable that ran from ETMX to ETMY and ran it from ETMX into the control room instead. This way Cici and Deeksha can send small voltage signals to the AUX PZT and read back using the beatnote pickoff that is usually connected to the spectrum analyzer.

  16956   Tue Jun 28 16:59:35 2022 PacoSummaryALSALS beat allan deviation (XARM)

[Paco]

I took ~ 7 minutes of XALS beatnote data with the XAUX laser locked to the XARM cavity, and the XARM locked to PSL to develop an allan deviation estimator. The resulting timeseries for the channel C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ (decimated timeseries in Attachment #1) was turned into an allan variance using the "overlapped variable tau estimator":

\sigma_y^2(n\tau_0, N) = \frac{1}{2n^2\tau_0^2(N - 2n)} \sum_{i=0}^{N-2n-1} (x_{i+2n} - 2x_{i+n} + x_i)^2

Where x_k represents the k-th data point in the raw timeseries, and n\tau_0 are the variable integration intervals under which two point variances are computed (the allan variance is a special case of M-point variance, where M=2). Then, the allan deviation is just the square root of that. Attachment #2 shows the fractional deviation (normalized by the mean beat frequency ~ 3 MHz for this measurement) for 100 integration times spanning the full duration (~ 7 min = 420 s).

The code used for this lives in Git/40m/labutils/measuremens/ALS/


If this estimate is any good, wherever the fractional beatnote deviation reaches a minimum value can be used as a proxy for the longest averaging time that give a statistical increase in SNR. After this timescale, the frequency comparison is usually taken over by "environmental instabilities" which I don't think I can comment further on. In our particular estimate, the 100 second integration gives a fractional deviation of ~ 0.44 %, or absolute deviation of 12.925 kHz.

  16959   Tue Jun 28 18:53:16 2022 ranaSummaryALSALS beat allan deviation (XARM)

what's the reasoning behind using df/f_beat instead of df/f_laser ?

Quote:

[Paco]

I took ~ 7 minutes of XALS beatnote data with the XAUX laser locked to the XARM cavity, and the XARM locked to PSL to develop an allan deviation estimator.

 

  16962   Wed Jun 29 14:28:06 2022 PacoSummaryALSALS beat allan deviation (XARM)

I guess it didn't make sense since f_beat can be arbitrarily moved, but the beat is taken around the PSL freq ~ 281.73 THz. Attachment #1 shows the overlapping tau allan deviation for the exact same dataset but using the python package allantools, where this time I used the PSL freq as the base frequency. This time, I can see the minimum fractional deviation of 1.33e-13 happening at ~ 20 seconds.

Quote:

what's the reasoning behind using df/f_beat instead of df/f_laser ?


Another, more familiar interpretation

The allan variance is related to the beatnote spectral density as a mean-square integral (the deviation is then like the rms) with a sinc window.

\sigma^2_\nu = 2 \int_0^{\infty} S_\nu(f) \lvert \frac{\sin({\pi f \tau})}{\pi f \tau} \lvert ^2 df

  16965   Thu Jun 30 18:06:22 2022 PacoUpdateALSOptimum ALS recovery - part I

[Paco]

In the morning I took some time to align the AUX beams in the XEND table. Later in the afternoon, I did the same on the YEND table. I then locked the AUX beams to the arm cavities while they were stabilized using POX/POY and turned off the PSL hepa off temporarily (this should be turned on after today's work).

After checking the the temperature slider sign on the spectrum analyzer of the control room I took some out-of-loop measurements of both ALS beatnotes (Attachment #1) by running diaggui /users/Templates/ALS/ALS_outOfLoop_Ref_DQ.xml and by comparing them against their old references (red vs magenta and blue vs cyan); it seems that YAUX is not doing too bad, but XAUX has increased residual noise around and above 100 Hz; perhaps as a result of the ongoing ALS SURF loop investigations? It does look like the OLTF UGF has dropped by half from ~ 11 kHz to ~ 5.5 kHz.

Anyways let this be a reference measurement for current locking tasks, as well as for ongoing SURF projects.

  16994   Tue Jul 12 19:46:54 2022 PacoSummaryALSHow (not) to take NPRO PZT transfer function

[Paco, Deeksha, rana]

Quick elog for this evening:

  • Rana disabled MC servo .
  • Slow loop also got disengaged.
  • AUX PSL beatnote is best taken with *free running lasers* since their relative frequency fluctuations are lowest than when locked to cavities.
  • DFD may be better to get PZT transfer funcs, or get higher bandwidth phase meter.
  • Multi instrument to be done with updated moku
  • Deeksha will take care of updated moku
  17123   Wed Aug 31 12:57:07 2022 ranaSummaryALScontrol of ALS beat freq from command line -easy

The PZT sweeps that we've been making to characterize the ALS-X laser should probably be discarded - the DFD was not setup correctly for this during the past few months.

Since the DFD only had a peak-peak range of ~5 MHz, whenever the beat frequency drifts out of the linear range (~2-3 MHz), the data would have an arbitrary gain. Since the drift was actually more like 50 MHz, it meant that the different parts of a single sweep could have some arbitrary gain and sign !!! This is not a good way to measure things.

I used an ezcaservo to keep the beat frequency fixed. The attacehed screenshot shows the command line. We read back the unwrapped beat frequency from the phase tracker, and feedback on the PSL's NPRO temperature. During this the lasers were not locked to any cavities (shutters closed, but servos not disabled).

For the purposes of this measurement, I reduced the CAL factor in the phase tracker screen so that the reported FINE_PHASE_OUT is actually in kHz, rather than Hz on this plot. So the green plot is moving by 10's of MHz. When the servo is engaged, you can see the SLOWDC doing some action. We think the calibration of that channel is ~1 GHz/V, so 0.1 SLOWDC Volts should be ~100 MHz. I think there's a factor of 2 missing here, but its close.

As you can see in the top plot, even with the frequency stabilized by this slow feedback (-1000 to -600 seconds), the I & Q outputs are going through multiple cycles, and so they are unusable for even a non serious measurement.

The only way forward is to use less of a delay in the DFD: I think Anchal has been busily installing this shorter cable (hopefully, its ~3-5 m long so that the linear range is more. I think a 10 m cable is too long.), and the sweeps taken later today should be more useful.

  17131   Fri Sep 2 15:40:25 2022 AnchalSummaryALSDFD cable measurements

[Anchal, Yehonathan]

I laid down another temporary cable from Xend to 1Y2 (LSC rack) for also measuring the Q output of the DFD box. Then to get a quick measurement of these long cable delays, we used Moku:GO in oscillator mode, sent 100 ns pulses at a 100 kHz rate from one end, and measured the difference between reflected pulses to get an estimate of time delay. The other end of long cables was shorted and left open for 2 sets of measurements.

I-Mon Cable delay: (955+/- 6) ns / 2 = 477 +/- 3 ns

Q-Mon Cable delay: (535 +/- 6) ns / 2 = 267 +/- 3 ns

Note: We were underestimating the delay in I-Mon cable by about a factor of 2.

I also took the opportunity to take a delay time measurement of DFD delayline. Since both ends of cable were present locally, it made more sense to simply take a transfer function to get a clean delay measurement. This measurement resulted with value of 197.7 +/- 0.1 ns. See attached plot. Data and analysis here.

  677   Wed Jul 16 09:27:17 2008 steveUpdateALARMPSL-FSS_RMTEMP alarm is false
Morning alarm sound is good for people who does not drink coffee.
Our 40m alarm is on every morning.
Those whom are not here in the morning thinks that this beeping sound is inspirational.
Would someone change this sound into less punishing form, like mockingbird chirp....

The C1PSL_SETTINGS.adl (40mm PSL Settings ) indicating that
C1:PSL-FSS_INOFFSET (Input Offset Adjust ) should be 0.3 +-0.05 V (red warning tag )

Alarm Handler: 40M pointing to yellow grade warning of PSL-FSS_RMTEM
This is a false alarm.

Two years trend of these channels are here:
  1615   Thu May 21 12:58:32 2009 robConfigurationALARMPEM count-half disabled

I've disabled the alarm for PEM_count_half, using the mask in the 40m.alhConfig file.  We can't do anything about it, and it's just annoying.

  2549   Tue Jan 26 20:18:32 2010 ranaConfigurationALARMop540m: alarms and BLRMS and StripTool restored

I turned the StripTool and ALARMS and BLRMS back on on op540m. Looks like it has been rebooted 5 days ago and no one turned these back on. Also, there was a bunch of junk strewn around its keyboard which I restrained myself from throwing in the trash.

The BLRMS trends should be active now.

  11241   Thu Apr 23 23:07:23 2015 DugoliniFrogsALARMlaptops warning

Please!

Don't put laptops on the ISC Tables!

  11893   Sun Dec 20 23:23:54 2015 ericqUpdateALARMRats.

A small rat / large mouse just ran through the control room. Ugh.

  11901   Wed Dec 23 16:15:47 2015 ranaOmnistructureALARMfire alarm

Fire alarm went off several minutes ago. Talked to security and they said there was no fire. It beeped twice again just now. No one has been working on the IFO today.

  12216   Mon Jun 27 15:26:03 2016 SteveOmnistructureALARMfire alarm test

The fire alarm came on around 15:05  for about 2-3 minutes. We all  left the lab and counted heads.  I called Paul Mackel x2646 (cell 626/ 890- 3259) at Fire Protection Services. He said that this alarm test was planned and we should of got an email notice. Perhaps I missed that notes.

Quote:

Fire alarm went off several minutes ago. Talked to security and they said there was no fire. It beeped twice again just now. No one has been working on the IFO today.

 

  14294   Wed Nov 14 14:35:38 2018 SteveUpdateALARM emergency calling list for 40m Lab

It is posted at the 40m wiki with Gautam' help. Printed copies posted around doors also.

  14863   Fri Sep 6 16:38:24 2019 aaronUpdateALARMAlarm noise from smart-ups machine under workstation?

There was an alarm sound from the Smart-UPS 2200 sitting under the workstation. I see that the 'replace battery' light is red, and this elog tells me that these batteries are replaced every ~1-4 years; the last replacement was march 2016. Holding down the 'test' button for 2-3 seconds results in the alarm sound and does not clear the replace battery indicator.

  2053   Mon Oct 5 14:37:29 2009 AlbertoUpdateABSLAbsolute Length Meaasurement NPRO is on

In the revival of the experiement length measurement for the recycling cavities, I turned the auxiliary NPRO back on. The shutter is closed.

I also recollected all the equipment of the experiment after that during the summer it had been scattered around the lab to be used for other purposes (Joe and Zach's cameras and Stephanie and Koji's work with the new EOM).

  2209   Mon Nov 9 11:14:57 2009 AlbertoUpdateABSLStarted working on the PSL table

I'm working on the PSL table to set up the PLL setup for the AbsL experiment.

  2211   Mon Nov 9 13:17:07 2009 AlbertoConfigurationABSLNPRO on

I turned the auxialiary NPRO for the AbsL Experiment on. Its shutter stays closed.

  2229   Tue Nov 10 19:19:57 2009 AlbertoUpdateABSLRotated polarizer on the PSL table, along the MC input pick off beam

Aligning the beam for the PLL of the AbsL Experiement I rotated the polarizer along the path of the MC Input pick off beam (= the pick off coming from the MC periscope).

  2230   Tue Nov 10 19:21:53 2009 AlbertoUpdateABSLPLL Alignment

I've been trying to lock the PLL for the AbsL Experiment but I can't see the beat (between the auxiliary NPRO and the PSL).

I believe the alignment of the PLL is not good. The Farady Isolator is definitely not perfectly aligned (you can see it from the beam spot after it) but still it should be enough to see something at the PLL PD.

it's probably just that the two beams don't overlap well enough on the photodiode. I'll work on that later on.

I'm leaving the lab now. I left the auxiliary NPRO on but I closed its shutter.

All the flipping mirrors are down.

  2238   Wed Nov 11 15:04:52 2009 AlbertoUpdateABSLWorking on the AP table

I've opened the AP table and I'm working on it.

  2239   Wed Nov 11 16:18:57 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

I've opened the AP table and I'm working on it.

I re-aligned the Faraday on the AP table. I also aligned the beam to the periscope on the PSL and all the other optics along the beam path.  Now I have a nice NPRO beam at the PLL which overlaps with the PSL beam. The alignment has to be further improved because I see no beat yet.

I wonder if the all the tinkering on the PSL laser done recently to revive the power has changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat. So maybe the misalignment is the casue.

Not feeling very well right now. I need to go home for a while.

AP table closed at the moment.

NPRO shutter closed

  2240   Wed Nov 11 17:10:51 2009 JenneUpdateABSLWorking on the AP table

Quote:

Quote:

I've opened the AP table and I'm working on it.

I re-aligned the Faraday on the AP table. I also aligned the beam to the periscope on the PSL and all the other optics along the beam path.  Now I have a nice NPRO beam at the PLL which overlaps with the PSL beam. The alignment has to be further improved becasue I see no beat yet.

I wonder if the all the tinkering on the PSL laser done recently to revive the power have changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat. So maybe the misalignment is the casue.

Not feeling very well right now. I need to go home for a while.

AP table closed at the moment.

NPRO shutter closed

 We definitely changed the PSL NPRO temp while fiddling around, trying to increase the laser power.  I think it's noted in the elog both times that it's happened in the last few months (once when Rana, Koji and I worked on it, and then again when it was just Koji), but we opened up the side of the MOPA box so that we could get at (and change) the potentiometer which adjusts the NPRO temp.  So you may have to search around for a while.

  2241   Wed Nov 11 17:33:54 2009 KojiUpdateABSLWorking on the AP table

Yes it did.

For long time, the crystal temperature C1:PSL-126MOPA_LTMP was 43~46deg. Now it is 34deg. Try ~10deg lower temperature.

Quote:

I wonder if the all the tinkering on the PSL laser done recently to revive the power have changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat.

 

  2250   Thu Nov 12 10:45:36 2009 AlbertoUpdateABSLWorking on the AP table

I've opened the AP table and I'm working on it.

Also auxiliary NPRO turned on and mechanical shutter opened.

  2252   Thu Nov 12 11:34:38 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

Yes it did.

For long time, the crystal temperature C1:PSL-126MOPA_LTMP was 43~46deg. Now it is 34deg. Try ~10deg lower temperature.

Quote:

I wonder if the all the tinkering on the PSL laser done recently to revive the power have changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat.

 

 Beat found at 30MHz with auxiliary NPRO temperature of 37.19 degrees, vs. ~48 deg as it used to be.

The beat is small (-70dBm). PLL alignment has to be improved.

  2254   Thu Nov 12 12:51:45 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

I've opened the AP table and I'm working on it.

Also auxiliary NPRO turned on and mechanical shutter opened.

AP table and aux NPRO shutter just closed.

  2257   Thu Nov 12 16:53:59 2009 AlbertoUpdateABSLWorking on the AP table

Quote:

Quote:

Yes it did.

For long time, the crystal temperature C1:PSL-126MOPA_LTMP was 43~46deg. Now it is 34deg. Try ~10deg lower temperature.

Quote:

I wonder if the all the tinkering on the PSL laser done recently to revive the power have changed the PSL NPRO temperature and so its frequency. That could also explain why the beat doesn't show up at the same temperature of the NPRO as I used to operate it. Although I scanned the NPRO temperature +/- 2 deg and didn't see the beat.

 

 Beat found at 30MHz with auxiliary NPRO temperature of 37.19 degrees, vs. ~48 deg as it used to be.

The beat is small (-70dBm). PLL alignment has to be improved.

 PLL alignment improved. Beat amplitude = -10dBm. Good enough.

DC readouts at the PLL photodiode:

V_NPRO = -4.44V

V_PSL = -3.76V

The NPRO beam is attenuated by a N.D.=1 attenuator just before going to the photodiode.

Something strange happened at the last. Right before -10dBm, the amplitude of the beat was about -33dBm. Then I was checking the two interfering beams with the IR card and saw that they overlapped quite well. I then turned my head back to the spectrum analyzer and suddenly the beat was at -10dBm. Not only, but a bunch of new peaks had appeared on the spectrum. Either I inadvertently hit the PD moving it to a better position or something else happened.

Like if someone was making some other modulation on the beam or the modulation depth of the PSL's sidebands had gone up.

  2261   Thu Nov 12 18:10:27 2009 AlbertoUpdateABSLPLL Locked

I locked the PLL and made some first measuremtns of the spectrum of the error signal. I'll post them later.

I closed the shutter of the NPRO.

  2321   Tue Nov 24 14:33:22 2009 AlbertoUpdateABSLworking on the AP table

I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP table and PSL table.

  2324   Tue Nov 24 19:16:02 2009 AlbertoUpdateABSLworking on the AP table

Quote:

I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP table and PSL table.

 Closing the AP table and the NPRO shutter now.

ELOG V3.1.3-