11863
|
Tue Dec 8 15:40:48 2015 |
Steve | Update | VAC | glitchy RGA scan at day 434 |
The noise floor of the Rga scan is glitching less today
|
Attachment 1: lessGlichingToday.png
|
|
11866
|
Wed Dec 9 11:25:55 2015 |
Steve | Update | VAC | normal RGA scan at day 435-436 |
Glitches are gone. Rga scan is good
|
Attachment 1: glichesGone.png
|
|
Attachment 2: d436.png
|
|
11960
|
Fri Jan 29 09:59:23 2016 |
Steve | Update | VAC | vacuum control upgrade |
Rana stated yesterday that there will be a vacuum control update in the close future. Witnesses : Rich, Chris and Dave
Can you give me this in writing?
|
Attachment 1: 1100days.png
|
|
11996
|
Wed Feb 17 09:05:07 2016 |
Steve | Update | VAC | forepump replaced |
TP3 drypump replaced after 10,344 hrs at 750 mTorr foreline pressure.
The foreline pressure is 13 mTorr after 8 hrs of running, TP3: 50K rpm, 0.14 Amp with all annuloses pumped.
The annulos pressures are 0.3 - 5 mtorr
************************
Frame builder just crashed again |
Attachment 1: jetStoreCrashedAgain.png
|
|
12014
|
Tue Mar 1 09:37:15 2016 |
steve | Update | VAC | vac rack UPS batteries replaced |
Amstron batteries replaced after 11 months with SP-12-5.5HR, 2 years warranty from replaceUPSbattery.com
Quote: |
Batteries replaced after 3.5 years with Amstron AP-1250F2, 8x 12V 6Ah
Quote: |
APC Smart -UPS 2200 model: SUA2200RM2U batteries were replaced by compatible RBC43, 8x 12V5A
|
|
Note: the replace battery LED did not go out ( well pasted 24 hrs ) till the self test bottom was hold down for 2-3 sec |
Attachment 1: vacRackBat.jpg
|
|
12086
|
Thu Apr 21 15:12:38 2016 |
Steve | Update | VAC | RGA is not working |
Quote: |
Steve pointed out that in the aftermath of the Nitrogen running out a couple of times last week, the RGA had shut itself off thinking that there was a leak and so it was not performing the scheduled scans once a day. So the data files from the scheduled scans were empty in the /opt/rtcds/caltech/c1/scripts/RGA/logs directory. The wiki page for getting it up and running again is up-to-date, but the script RGAset.py did not exist on the c0rga machine, which the RGA is communicating with via serial port. I copied over the script RGAset.py from rossa to c0rga and ran the script on that machine - but the error flags it returned were not all 0 (indicating some error according to the manual) - so I edited the script to send just the initialize command ('IN0') and commented out the other commands, after which I got error flags which were all 0. After this, I ran a manual scan using 'RGAlogger.py', and it appears that the RGA is now able to take scans again - I'm attaching a plot of the scan results. We've saved this scan as a reference to compare against after a few days.
|
Our last RGA scan is from February 14, 2016 We had a power outage on the 15th
Gautom has not succeded reseting it. The old c0rga computer looks dead. Q may resurrect it, if he can? |
12115
|
Wed May 11 16:39:01 2016 |
ericq | Update | VAC | c0rga alive, output wonky |
Quote: |
Our last RGA scan is from February 14, 2016 We had a power outage on the 15th
Gautom has not succeded reseting it. The old c0rga computer looks dead. Q may resurrect it, if he can?
|
The c0rga computer was off, I turned it on via front panel button. After running RGAset.py, RGAlogger.py seems to run. However, there are error messages in the output of the plotrgascan MATLAB script; evidiently there are some negative/bogus values in the output.
I'll look into it more tomorrow. |
12116
|
Thu May 12 14:29:58 2016 |
gautam | Update | VAC | RGA back up and running |
It looks like the hardware reset did the trick. Previously, I had just tried ssh-ing into c0rga and rebooting it. At the time, however, Steve and I noticed that the various LEDs on the RGA unit weren't on, as they are supposed to be in the nominal operating state. Today, Steve reported that all LEDs except the RS232 one were on today, so I just tried following the steps in this elog again, looks like things are back up and running. I'm attaching a plot of the scan generated using plotrgascan MATLAB script, it looks comparable to the plot in elog 11697, which if I remember right, was acceptable.
Unless there is some reason we want to keep this c0rga machine, I will recommission one of the spare Raspberry Pis lying around to interface with the RGA scanner when I get the time...
Quote: |
Quote: |
Our last RGA scan is from February 14, 2016 We had a power outage on the 15th
Gautom has not succeded reseting it. The old c0rga computer looks dead. Q may resurrect it, if he can?
|
The c0rga computer was off, I turned it on via front panel button. After running RGAset.py, RGAlogger.py seems to run. However, there are error messages in the output of the plotrgascan MATLAB script; evidiently there are some negative/bogus values in the output.
I'll look into it more tomorrow.
|
This is a cold scan. |
Attachment 1: RGAscan_12May2016.png
|
|
12117
|
Sun May 15 19:48:08 2016 |
Steve | Update | VAC | run out of N2 |
3-4 hrs ago we run out of nitrogen. We are back to Vacuum Normal
|
Attachment 1: noN2.png
|
|
12119
|
Tue May 17 14:46:51 2016 |
Steve | Update | VAC | RGA scan at day 595 |
We have good RGA scan now. There was no scan for 3 months. |
Attachment 1: RGAscan595d.png
|
|
Attachment 2: pd78-560Hz-d600.png
|
|
12174
|
Tue Jun 14 10:13:34 2016 |
Steve | Update | VAC | temporary N2 supply line |
The drill room floor will be retiled Thursday, June 16. Temporary nitrogen line set up will allow emptying the hole area.
Ifo room entry will be through control room.
|
Attachment 1: afterN2work.png
|
|
12195
|
Fri Jun 17 15:22:31 2016 |
gautam | Update | VAC | N2 supply line restored after retiling |
Quote: |
The drill room floor will be retiled Thursday, June 16. Temporary nitrogen line set up will allow emptying the hole area.
Ifo room entry will be through control room.
|
The retiling work has finished, Steve and I restored the N2 supply configuration to its normal state. The sequence of steps followed was:
- Went to the X end and closed the following valves, roughly in this order: VAEE, VAEV, VABS, VABSCI, VASV, VASE, V4, V1.
- Checked the RPM on the various turbo pump controllers to make sure they were in their nominal states
- Disconnect the electrical connections to V1, V4, V5 and VA6 - just to make sure some spurious signal doesn't unintentionally open any of these valves while we are mucking around with the N2 supply
- Close the valves on the N2 cylinders in the drill room. Disconnect the temporary nitrogen line (at this point, the N2 pressure to the IFO valves goes down from ~7-PSI to 0), reconnect the old supply chain, taking care that we aren't unintentionally loosening any of the Swagelock connections while unscrewing stuff
- Replaced one of the N2 cylinders that was running low.
- Reopen the cylinders, restore N2 pressure to IFO valves to ~70PSI.
- Do steps 1-3 in reverse: i.e. reconnect power to all valves, open them in the reverse order we closed them while monitoring the state of the various turbo pumps.
- Acknowledged the error message on the C0VAC medm screen
Note: the valve isolating the RGA automatically shutoff during this work, possibly because it detected a pressure above its threshold - after checking the appropriate pressure gauges, we reopened this valve as well.
The attached screenshot suggests that everything went as planned and that the vacuum system is back to normal...
|
Attachment 1: c0vac_06172016.png
|
|
12237
|
Fri Jul 1 09:20:44 2016 |
Steve | Update | VAC | vent 78 |
VENT OBJECTIVES:
1, Fix ETMX sus "jump issue"
2, First Contact clean the arms
3, Install new spare cold cathode and convectron gauges: InstruTech-Hornet
4, Install 50 mm apeture beam baffles
5, Check and clean optical quality viewport from inside
The following bullets were executed:
We are venting the 40m IFO
Quote: |
Steve has ordered some teflon parts to take the place of the metal parts in his acetone-soaking jig. They should arrive tomorrow.
So, we will be begin the venting process tomorrow. Doors to come off on Tuesday.
|
|
Attachment 1: vent78.png
|
|
Attachment 2: vent78at250T.png
|
|
12238
|
Fri Jul 1 15:48:10 2016 |
Steve | Update | VAC | vent 78 at 710 Torr |
I just disconnected the 6th instrument grade air cylinder from the vacuum envelope at 720 Torr. Now it will reach equilibrium through a filter as it sucks in lab air.
This is the sure way not to over pressurize the chamber.
|
Attachment 1: pd78at710T.png
|
|
12248
|
Wed Jul 6 08:34:00 2016 |
Steve | Update | VAC | RGA scan at day 639 |
The last RGA scan of this pumpdown 78
Pressure plot of 640 days long pd 78
CC1 cold cathode gauge was jump started with an accidental pressure glitch, that you can see on P1 plot
|
Attachment 1: RGAscan639d.png
|
|
Attachment 2: pd78d640.png
|
|
12250
|
Wed Jul 6 10:58:07 2016 |
Steve | Update | VAC | vent 78 valve configuration |
Vacuum Status: Chamber Open
All chamber annuloses are vented. Vac Monitor screen is not communicating with gauges. The valve position indicator are working.
RGA is pumped by Maglev through VM2
|
Attachment 1: CamberOpen.png
|
|
12293
|
Tue Jul 12 18:13:19 2016 |
Koji | Update | VAC | N2 bottle replaced |
Gautam, Koji
We replaced the right N2 bottle as it was empty. |
12314
|
Wed Jul 20 11:39:07 2016 |
Steve | Update | VAC | vent 78 day 20 |
RGA background scan
Quote: |
Vacuum Status: Chamber Open
All chamber annuloses are vented. Vac Monitor screen is not communicating with gauges. The valve position indicator are working.
RGA is pumped by Maglev through VM2
|
|
Attachment 1: vent78-d20.png
|
|
12337
|
Tue Jul 26 14:24:38 2016 |
Koji | Summary | VAC | Purge compressed air system at LHO |
I've visited the purge clean air system at LHO Yarm mid-station with John Worden.
The system is described C981637. There is a schematic in C981637-06-V (Vol.6).pdf although the schematic has some differences (or uncorrected mistakes).
This system is intended to provide positive pressure when a soft cover is attached to a chamber door. When the door is open, the purging does not help to keep the chamber clean because the flow is too slow. This protection has to be done with overhead HEPA filters (22x5000cfm). It may be possible that this purge air helps the tube not to allow dusts to come in. But before using this, the chambers and the tubes have to be cleaned, according to John.
- Here at the site, the purge air system is started up a day before the vent. This system is used for the vent air, the purge air, and turbo foreline filling.
- Air intake (attachment 1): At the site, the air is intaken from the VEA. We want to incorporate somewhat clean air instead of dirty, dusty, outside air.
- Initial filter (attachment 2): a high volume filter before the compressors.
- The compressors (attachment 3, 4) are 5x 6 horse power air compressor each goes up to 160 psi. They are turned on and off depending on the demand of the air. Which is turned on is revolved by the controller to equalize the compressor usage hours.
- The compressed air goes through the air cooler (heat exchanger) to remove the heat by the compressor work.
- This air goes through prefilters and accumulated in the air receiver (100psi) (attachment 5). This receiver tank has an automated vent valve for periodical water drainage at the bottom.
- The accumulated air is discharged to twin drier towers (attachment 6, blue). The tower is operated by the controller (attachment 7) alternately with a period of 4min (or 10min by setting). When one of the towers is working, a humid air comes from the bottom and the dry air is discharged from the top. A part of the dry air goes into the other tower from the top to the bottom and dries the tower. There is a vent at the bottom to discharge water periodically.
- The dried air goes through 4 types of filters. After the last filter, all of the plumbing should be made of stainless steel to keep cleanliness.
- The air goes to the pressure reducing regulator (attachment 8, gray). The final flow speed at the chamber side is 50cfm max, according to John.
- The lower pressure air goes through the final filter (attachment 8, blue). As the pressure is low, this filter is big in order to keep the volume of the air flow.
- The purge air is supplied to the chamber side with KF50 (attachment 9). There is a vent valve (attachment 10) for safety and also to run a dry air for at least a day before the use to clean up the supply line. The purge line is disconnected when no in use.
- The entire system (attachment 11) and size comparison (attachment 12).
|
Attachment 1: air_intake.jpg
|
|
Attachment 2: initial_filter.jpg
|
|
Attachment 3: compressors1.jpg
|
|
Attachment 4: compressors2.jpg
|
|
Attachment 5: air_receiver_dryer.jpg
|
|
Attachment 6: drier.jpg
|
|
Attachment 7: drier_controller2.jpg
|
|
Attachment 8: pressure_regulator_and_last_filter.jpg
|
|
Attachment 9: chamber_side_supply.jpg
|
|
Attachment 10: vent_valve_for_line_cleaning.jpg
|
|
Attachment 11: the_whole_system.jpg
|
|
Attachment 12: size_comparison.jpg
|
|
12338
|
Tue Jul 26 16:01:32 2016 |
Steve | Summary | VAC | Purge compressed air system |
Thanks for checking this out Koji
The builder in 1996 was Process System International, Inc ( Westborough, MA ) It does not exist any longer or I just could not find them. Flow diagramm at Atm1
Should I be keep looking for a company who could quote us for building a similar smaller unit with 10 - 15 cfm flowrate?
Note: my intension with the two mobile-overhead HEPA filter was the same as John Worden's " clean air overpressured tent " at chamber entrance.
Atm2, Our unit has 650 cfm, velocity 90 fpm at resistance 0.5" It may be enough to give a little overpressure if we seal it well to the chamber
We use to use them to minimize dirt getting inside the chanbers. |
Attachment 1: cleanAirSupl.pdf
|
|
Attachment 2: mobileHEPA.jpg
|
|
12339
|
Tue Jul 26 17:41:59 2016 |
Koji | Summary | VAC | Purge compressed air system |
We have no number for the CFM without calculation. We can't assume a random number like 10-15 |
12483
|
Mon Sep 12 17:54:24 2016 |
Steve | Update | VAC | pumpdown 79 |
We are pumping down. The annuloses are below 10 mTorr
The vacuum gauges are not communicating with the medm so there is no plot available.
The main volume pumping is stopped at P1 = 220 Torr ( New SuperBee 174 Torr ) for overnight.
note: SuperBee was reading 791 Torr at atm and it was not set to 760 ! |
12487
|
Tue Sep 13 16:16:28 2016 |
Chris | Omnistructure | VAC | vacuum interlock bypassed |
(Steve, Chris)
The pumpdown had stalled because of some ancient vacuum interlock code that prevented opening the valve V1 between the turbo pump and the main volume.
This interlock [0] compares the channels C1:Vac-P1_pressure and C1:Vac-PTP1_pressure, neither of which is functioning at the moment. The P1 channel apparently stopped reading sometime during the vent, and contained a value of ~700 torr, while the PTP1 channel contained 0. So the interlock code saw this huge apparent pressure difference and refused to move the valve.
To bypass this check, we used caput to enter a pressure of 0 for P1.
[0] /cvs/cds/caltech/state/from_luna/VacInterlock.st |
12488
|
Tue Sep 13 17:47:04 2016 |
Steve | Update | VAC | pumpdown 79 completed |
P1 IFO pressure is 1 mTorr, valve configuration: vacuum normal, annulosses are pumped, RGA is off, not pumped.
THANKS to Chris !
The shutters can be opened with high power.
No communication error message still exist.
I will reboot c1vac1 and c1vac2 to get gauge communication with medm tomorrow.
Quote: |
(Steve, Chris)
The pumpdown had stalled because of some ancient vacuum interlock code that prevented opening the valve V1 between the turbo pump and the main volume.
This interlock [0] compares the channels C1:Vac-P1_pressure and C1:Vac-PTP1_pressure, neither of which is functioning at the moment. The P1 channel apparently stopped reading sometime during the vent, and contained a value of ~700 torr, while the PTP1 channel contained 0. So the interlock code saw this huge apparent pressure difference and refused to move the valve.
To bypass this check, we used caput to enter a pressure of 0 for P1.
[0] /cvs/cds/caltech/state/from_luna/VacInterlock.st
|
|
12492
|
Wed Sep 14 09:44:09 2016 |
Steve | Update | VAC | pumpdown 79 completed |
The pumpdown started at 9-12-2016
The IFO is at 5e-5 Torr vacuum normal after 73 day at atm.
ITMX needed to be freed and ETMY-UL is still misbehaving occasionally.
New pirani and cold cathode gauges added at this vent. They were baked at 100 C for 6 hrs under vacuum.
Go to the Vac Rack to read IFO pressure from the gauge itself when Vac. Monitor is blank as it is now !
Quote: |
P1 IFO pressure is 1 mTorr, valve configuration: vacuum normal, annulosses are pumped, RGA is not pumped.
THANKS to Chris !
The shutters can be opened with high power.
No communication error message still exist.
I will reboot c1vac1 and c1vac2 to get gauge communication with medm tomorrow.
|
|
Attachment 1: pd79.png
|
|
Attachment 2: pd79d2.jpg
|
|
Attachment 3: CCcg.jpg
|
|
12496
|
Thu Sep 15 10:36:32 2016 |
Steve | Update | VAC | pumpdown 79 day 3 |
IFO pressure 3.7e-5 Torr at new cold cathode InstruTech - Hornet
New items in vacuum:
1, ETMX sus tower with new baked sus wire, EP30-2 epoxied magnets, same at different locations also........ .......... and 2 ruby wirestandoffs.
2, First Contact cleaned arm test masses only. This technic was a 1st time use in our vacuum system.
3, 50 mm ID green glass baffles at the ends
4, witness mirrors at ETMX and ITMY (old oplev mirrors) We observed a very dusty system: sides of optics, towers and tables were wiped. Hepa tents used at Y arm and BS
5, new pirani, cc gauge and 1.5" right angle valve
Quote: |
The pumpdown started at 9-12-2016
The IFO is at 5e-5 Torr vacuum normal after 73 day at atm.
ITMX needed to be freed and ETMY-UL is still misbehaving occasionally.
New pirani and cold cathode gauges added at this vent. They were baked at 100 C for 6 hrs under vacuum.
Go to the Vac Rack to read IFO pressure from the gauge itself when Vac. Monitor is blank as it is now !
Quote: |
P1 IFO pressure is 1 mTorr, valve configuration: vacuum normal, annulosses are pumped, RGA is not pumped.
THANKS to Chris !
The shutters can be opened with high power.
No communication error message still exist.
I will reboot c1vac1 and c1vac2 to get gauge communication with medm tomorrow.
|
|
|
Attachment 1: vent78for73d.png
|
|
12525
|
Fri Sep 30 10:37:57 2016 |
Steve | Update | VAC | vacuum VME machines rebooted |
[ Gautam and Steve ]
c1vac1 and c1vac2 were rebooted and the gauges are communicating now. V1, VA6, V5 and V4 were closed and disconnected to avoid unexpected valve switching. All went smoothly.
The new ITcc gauge is at 1e-5 Torr as CC1 This is the gauge that should be logged in slow channel.
TP2 fore line dry pump was replaced this morning after 382 day of operation.
TP3 dry pump is very noisy, but it's pressure still 47 mTorr
Quote: |
[ericq, Gautam, Steve]
Following roughly the same procedure as ELOG 11354, c1vac1 and c1vac2 were rebooted. The symptoms were identical to the situation in that ELOG; c1vac1 could be pinged and telneted to, but c1vac2 was totally unresponsive.
The only change in the linked procedure was that we did not shut down the maglev. Since I unwittingly had it running for days without V4 open while Steve was away, we now know that it can handle shorter periods of time than that...
Upon reboot, many channels were readable again, unfortunately the channels for TP2 and TP3 are still blank. We were able to return to "Vacuum normal state," but because of unknowned communication problems with VM1's interlock, we can't open VM1 for the RGA. Instead we opened VM2 to expose the RGA to the main IFO volumn, but this isn't part of the "Normal" state definite, so things currently read "Undefined state".
|
|
Attachment 1: rebooted.png
|
|
12530
|
Tue Oct 4 11:02:31 2016 |
Steve | Update | VAC | RGA is out of order |
The last good rga scan at vent 78 day 38
Quote: |
Quote: |
RGA background scan
Quote: |
Vacuum Status: Chamber Open
All chamber annuloses are vented. Vac Monitor screen is not communicating with gauges. The valve position indicator are working.
RGA is pumped by Maglev through VM2
|
|
|
|
Attachment 1: lastBGscan.png
|
|
12537
|
Fri Oct 7 10:29:57 2016 |
Steve | Update | VAC | RGA removed |
The RGA is removed for repaire. It's volume at atmophere and sealed.. P4 reading of 38 Torr is not correct.
|
Attachment 1: RGAremoved.png
|
|
12556
|
Thu Oct 13 00:23:54 2016 |
ericq | Update | VAC | Pre-vent prep |
I have completed the following non-Steve portions of the pre-vent checklist [wiki-40m.ligo.caltech.edu]
- Center all oplevs, transom QPDs (and IPPOS + IPANG if they are set up)
- Align the arm cavities for IR and align the green lasers to the arms.
- Update the SUS Driftmon values
- Reconcile all SDF differences
- Reduce input power to no more than 100mW (measured at the PSL shutter) by adjusting wave plate+PBS setup on the PSL table BEFORE the PMC. (Using the WP + PBS that already exist after the laser.)
- Replace 10% BS before MC REFL PD with Y1 mirror and lock MC at low power.
- Close shutter of PSL-IR and green shutters at the ends
All shutters are closed. Ready for Steve to check nuts and begin venting! |
12559
|
Thu Oct 13 16:48:34 2016 |
Steve | Update | VAC | vent 79 is completed |
IFO is at atmosphere. The MC can be locked in air now.
The doors will be coming off tomorrow 8am sharp.
Do we want to install the ITM baffles?
What about the found OSEM filters?
|
Attachment 1: vent_79.png
|
|
12574
|
Thu Oct 20 09:17:08 2016 |
Steve | Update | VAC | Pumpdown 80 has started |
Pumping again after 7 days at atmosphere.
BS, ITMY and OMC chambers were open only.
Checked: jam nuts, viewport covers and beam shutters.
Oplev servo turned off and medm screens shots taken.
New item in vacuum: green shade 14 glass beam block at IR-input [ from the PSL ] viewport to block green reflection-scatter.
Reminder: viewport is not AR coated for green!
|
Attachment 1: readyPd80.png
|
|
Attachment 2: beforePd80.png
|
|
Attachment 3: pd80at1h.png
|
|
Attachment 4: 7d_atm.png
|
|
12575
|
Thu Oct 20 15:59:52 2016 |
Steve | Update | VAC | Pumpdown 80 has completed |
IFO pressure 1.7E-4 Torr on new not logged cold cathode gauge. P1 <7E-4 Torr
Valve configuration: vac.normal with anunulossess closed off.
TP3 was turned off with a failing drypump. It will be replaced tomorrow. |
Attachment 1: 7hPd80.png
|
|
Attachment 2: 8hPd80.png
|
|
Attachment 3: 9hPd80.png
|
|
12577
|
Fri Oct 21 09:28:21 2016 |
Steve | Update | VAC | Vac Normal reached |
Dry pump of TP3 replaced after 9.5 months of operation.[ 45 mTorr d3 ]
The annulosses are pumped.
Valve configuration: vac normal, IFO pressure 4.5E-5 Torr [1.6E-5 Torr d3 ] on new ITcc gauge, RGA is not installed yet.
Note how fast the pressure is dropping when the vent is short.
Quote: |
IFO pressure 1.7E-4 Torr on new not logged cold cathode gauge. P1 <7E-4 Torr
Valve configuration: vac.normal with anunulossess closed off.
TP3 was turned off with a failing drypump. It will be replaced tomorrow.
|
All time stamps are blank on the MEDM screens. |
Attachment 1: VacNormal.png
|
|
12589
|
Mon Oct 31 16:20:50 2016 |
Steve | Update | VAC | RGA is reinstalled |
Quad rods and ionizer kit: consisting of repeller cage, anode grid, focus plate and filament were replaced....... under repair # RGA200/12 ECA 100416-12967
The electronic ECU is not connected. It is beeing pumped at IFO ITcc 9.7E-6 Torr vacNormal
Quote: |
The RGA is removed for repaire. It's volume at atmophere and sealed.. P4 reading of 38 Torr is not correct.
|
|
12633
|
Tue Nov 22 11:31:52 2016 |
Steve | Update | VAC | RGA is running again |
The Rga was turned on yesterday.
Quote: |
The RGA is removed for repaire. It's volume at atmophere and sealed.. P4 reading of 38 Torr is not correct.
|
|
Attachment 1: RGAisBack33d.png
|
|
Attachment 2: pd80-560Hz-d33.png
|
|
Attachment 3: afterRepair1d.png
|
|
12636
|
Wed Nov 23 11:32:13 2016 |
Steve | Update | VAC | TP3 drypump replaced again |
TP3 drypump replaced at 655 mTorr, no load, tp3 0.3A
This seal lasted only for 33 days at 123,840 hrs
The replacement is performing well: TP3 foreline pressure is 55 mTorr, no load, tp3 0.15A at 15 min [ 13.1 mTorr at d5 ]
Valve configuration: Vacuum Normal, ITcc 8.5E-6 Torr
Quote: |
Dry pump of TP3 replaced after 9.5 months of operation.[ 45 mTorr d3 ]
The annulosses are pumped.
Valve configuration: vac normal, IFO pressure 4.5E-5 Torr [1.6E-5 Torr d3 ] on new ITcc gauge, RGA is not installed yet.
Note how fast the pressure is dropping when the vent is short.
Quote: |
IFO pressure 1.7E-4 Torr on new not logged cold cathode gauge. P1 <7E-4 Torr
Valve configuration: vac.normal with anunulossess closed off.
TP3 was turned off with a failing drypump. It will be replaced tomorrow.
|
All time stamps are blank on the MEDM screens.
|
|
12642
|
Mon Nov 28 09:45:44 2016 |
Steve | Update | VAC | RGA scan at pumpdown day 40 |
The vacuum envelope ~1 C warmer today than 7 days ago. Bad peaks are coming down as normal.
|
Attachment 1: RGAscan_d40.png
|
|
Attachment 2: d2_&_d7scan.png
|
|
Attachment 3: RGAd2.png
|
|
12681
|
Thu Dec 22 09:37:20 2016 |
Steve | Update | VAC | RGA scan at day 63 |
Valve configuration: vacuum normal
RGA head temp: 43.5 C
Vac envelope temp: 23 C
|
Attachment 1: pd80-d63.png
|
|
Attachment 2: pd80-580Hz-d63.png
|
|
12698
|
Tue Jan 10 14:24:09 2017 |
Steve | Update | VAC | RGA scan at day 82 |
Valve configuration: vacuum normal
Vacuum envelope: 23C
Rga head: 44C
|
Attachment 1: pd80VNd82.png
|
|
12774
|
Tue Jan 31 14:14:29 2017 |
rana | Update | VAC | Pressure watch script |
I think this cron job is running on NODUS (our gateway) instead of our scripts machine:
*/1 * * * * /opt/rtcds/caltech/c1/scripts/Admin/n2Check.sh >> /opt/rtcds/caltech/c1/scripts/Admin/n2Check.log 2>&1
Quote: |
Based on Jenne's chiara disk usage monitoring script, I made a script that checks the N2 pressure, which will send an email to myself, Jenne, Rana, Koji, and Steve, should the pressure fall below 60psi. I also updated the chiara disk checking script to work on the new Nodus setup. I tested the two, only emailing myself, and they appear to work as expected.
The scripts are committed to the svn. Nodus' crontab now includes these two scripts, as well as the crontab backup script. (It occurs to me that the crontab backup script could be a little smarter, only backing it up if a change is made, but the archive is only a few MB, so it's probably not so important...)
|
moreover this script has a 90MB log file full of not finding its channel
I wish this script was in python instead of BASH and I wish it would run on megatron instead of nodus (why can't megatron send us email too?) and I wish that this log file would get wiped out once in awhile. Currently its been spitting out errors since at least a month ago:
Tue Jan 31 14:10:02 PST 2017 : N2 Pressure:
Channel connect timed out: 'C1:Vac-N2pres' not found.
(standard_in) 1: syntax error |
12853
|
Mon Feb 27 15:33:10 2017 |
Steve | Update | VAC | RGA scan at day 130 |
Valve configuration: vacuum normal
Vacuum envelope: 23.5 C
RGA head: 46.6 C
|
Attachment 1: rgascan@130d.png
|
|
13047
|
Wed Jun 7 11:32:56 2017 |
Steve | Update | VAC | smooth vac reboot |
Gautam and Steve,
The medm monitor & vac control screens were totally blank since ~ May 24, 2017 Experienced vacuum knowledge is required for this job.
IDENTIFY valve configuration:
How to confirm valve configuration when all vac mons are blank? Each valve has a manual-mechanical position indicator. Look at pressure readings and turbo pump controllers. VAC NORMAL configuration was confirmed based on these information.
Preparation: disconnect valves ( disconnect meaning: valve closes and stays paralized ) in this sequence VC2, VC1 power, VA6, V5, V4 & V1 power, at ifo pressure 7.3E-6 Torr-it ( it = InstruTech cold cathode gauge )
This gauge is independent from all other rack mounted instrumentation and it is still not logged.
Switching to this valve configuration with disconnected valves will insure NOT venting of the vacuum envelope by accidental glitching voltage drop or computer malfunction.
RESET v1Vac1 .........in 2-3 minutes........ ( v1Vac1 - 2 ) the vac control screen started reading pressures & position
Connected cables to valves (meaning: valve will open if it was open before it was disconnected and it will be control able from computer ) in the following order: V4, V1 power, V5, VA6, VC2 & VC1 power, at ifo 2E-5 Torr-it.....
....vac configuration is reading VAC NORMAL,
ifo 7.4E-6 Torr-it
We have to hook up the it-cold cathode gauge to be monitored - logged ! this should be the substitute for the out of order CC1 pressure gauge. |
Attachment 1: vac_reboot.png
|
|
13075
|
Tue Jun 20 16:28:23 2017 |
Steve | Update | VAC | RGA scan |
|
Attachment 1: RGAscan243d.png
|
|
Attachment 2: RGAscan.png
|
|
13132
|
Sun Jul 23 15:00:28 2017 |
Jamie | Omnistructure | VAC | strange sound around X arm vacuum pumps |
While walking down to the X end to reset c1iscex I heard what I would call a "rythmic squnching" sound coming from under the turbo pump. I would have said the sound was coming from a roughing pump, but none of them are on (as far as I can tell).
Steve maybe look into this?? |
13134
|
Mon Jul 24 09:55:29 2017 |
Steve | Omnistructure | VAC | controller failer |
Ifo pressure was 5.5 mTorr this Monday morning. The PSL shutter was still open. TP2 controller failed. Interlock closed V1, V4 and VM1
Turbo pump 2 is the fore pump of the Maglev. The pressure here was 3.9 Torr so The Maglev got warm ~38C but it was still rotating at 560 Hz normal with closed V1
Pressure plots are not available because of computer problems.
What I did:
Looked at pressures of Hornet and Super Bee Instru Tech. Inc
Closed all annuloses and VA6, disconnected V4 and VA6 and turned on external fan to cool Maglev
Opened V7 to pump the Maglev fore line with TP3
V1 opened manually when foreline pressure dropped to <2mTorr at P2 and the body temp of the Maglev cooled down to 25-27 C
VM1 opened at 1e-5 Torr
Valve configuration: vacuum normal with annuloses not pumped
Ifo pressure 8.5e-6 Torr -IT at 10am, P2 foreline pressure 64 mTorr, TP3 controller 0.17A 22C 50Krpm
note: all valves open manually, interlock can only close them
Quote: |
While walking down to the X end to reset c1iscex I heard what I would call a "rythmic squnching" sound coming from under the turbo pump. I would have said the sound was coming from a roughing pump, but none of them are on (as far as I can tell).
Steve maybe look into this??
|
PS: please call me next time you see the vacuum is not Vacuum Normal |
Attachment 1: TP2controllerFails.png
|
|
13142
|
Tue Jul 25 08:48:57 2017 |
Steve | Update | VAC | RGA scan at d278 |
The RGA did not shut down at the turbo pump controller failing.
Quote: |
Ifo pressure was 5.5 mTorr this morning. The PSL shutter was still open. TP2 controller failed. Interlock closed V1, V4 and VM1
Turbo pump 2 is the fore pump of the Maglev. The pressure here was 3.9 Torr so The Magelv got warm ~38C but it was still rotating at 560 Hz normal with closed V1
What I did:
Looked at pressures of Hornet and Super Bee Instru Tech. Inc
Closed all annuloses and VA6, disconnected V4 and VA6 and turned on external fan to cool Maglev
Opened V7 to pump the Maglev fore line with TP3
V1 opened manually when foreline pressure dropped to <2mTorr at P2 and the body temp of the Maglev cooled down to 25-27 C
VM1 opened at 1e-5 Torr
Valve configuration: vacuum normal with annuloses not pumped
Ifo pressure 8.5e-6 Torr -IT at 10am, P2 foreline pressure 64 mTorr, TP3 controller 0.17A 22C 50Krpm
note: all valves open manually, interlock can only close them
Quote: |
While walking down to the X end to reset c1iscex I heard what I would call a "rythmic squnching" sound coming from under the turbo pump. I would have said the sound was coming from a roughing pump, but none of them are on (as far as I can tell).
Steve maybe look into this??
|
PS: please call me next time you see the vacuum is not Vacuum Normal
|
|
Attachment 1: RGA278d.png
|
|
13143
|
Tue Jul 25 14:04:06 2017 |
Steve | Update | VAC | turbo controller installed and we are running at vac normal |
Gautam and Steve,
Spare Varian turbo-V 70 controller, Model 969-9505, sn 21612 was swapped in. It is running the turbo fine @ 50Krpm but it does not allow it's V4 valve to be opened............
It turns out that TP2 @ 75Krpm will allow V4 to open and close. This must be a software issue.
So Vacuum Normal is operational if TP2 is running 75,000 rpm
We want to run at 50,000 rpm on the long term.
Note: the RS232 Dsub connector on the back of this controller is mounted 180 degrees opposite than TP3 and old failed TP2 controller
PS: controller is shipping out for repair 7-28-2017
|
Attachment 1: TP2@75Krpm.png
|
|
13171
|
Tue Aug 8 17:04:26 2017 |
Steve | Update | VAC | unintended pump down |
IFO pressure 2 Torr, PSL shutter closed. I'm pumping down with 2 roughing pumps with ion pump gate valves open and annulosses at atm.
The vacuum envelope was vented to 17 Torr while I was replacing the USP battery stack. More about this later.....
Do not plan on using the interferrometer tonight. I will complete the pumpdown tomorrow morning.
|
Attachment 1: pumping_down.png
|
|
13172
|
Tue Aug 8 17:44:11 2017 |
Steve | Update | VAC | unintended pump down |
Pumpdown stopped for over night at ~ 1 Torr
The roughing line disconnected. Valves condition indicator "moving " means that it is closed and it's cable disconnected so it can not move.
The RGA is off and VM1 is stuck.
Quote: |
IFO pressure 2 Torr, PSL shutter closed. I'm pumping down with 2 roughing pumps with ion pump gate valves open and annulosses at atm.
The vacuum envelope was vented to 17 Torr while I was replacing the USP battery stack. More about this later.....
Do not plan on using the interferrometer tonight. I will complete the pumpdown tomorrow morning.
|
|
Attachment 1: stopped_pumping.png
|
|