40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 333 of 341  Not logged in ELOG logo
ID Date Author Type Category Subjectdown
  9576   Mon Jan 27 09:08:00 2014 SteveUpdateVAC4 days at atmosphere

 

 

  6929   Fri Jul 6 09:27:40 2012 steveUpdateVAC4 days at atm

 I'm looking for some movement indicators of the vent-pump down events.

 

  11005   Wed Feb 11 18:11:46 2015 KojiSummaryLSC3f modulation cancellation

33MHz sidebands can be elliminated by careful choice of the modulation depths and the relative phase between the modulation signals.
If this condition is realized, the REFL33 signals will have even more immunity to the arm cavity signals because the carrier signal will lose
its counterpart to produce the signal at 33MHz.

Formulation of double phase modulation

m1: modulation depth of the f1 modulation
m2: modulation depth of the f2 (=5xf1) modulation

The electric field of the beam after the EOM

E=E_0 \exp \left[ {\rm i} \Omega t + m_1 \cos \omega t +m_2 \cos 5 \omega t \right ]
\flushleft = {\it E}_0 e^{{\rm i} \Omega t} \\ \times \left[ J_0(m_1) + J_1(m_1) e^{{\rm i} \omega t}- J_1(m_1) e^{-{\rm i} \omega t} + J_2(m_1) e^{{\rm i} 2\omega t}+ J_2(m_1) e^{-{\rm i} 2\omega t} + J_3(m_1) e^{{\rm i} 3\omega t}- J_3(m_1) e^{-{\rm i} 3\omega t} + \cdots \right] \\ \times \left[ J_0(m_2) + J_1(m_2) e^{{\rm i} 5 \omega t}- J_1(m_2) e^{-{\rm i} 5 \omega t} + \cdots \right]
\flushleft = {\it E}_0 e^{{\rm i} \Omega t} \\ \times \left\{ \cdots + \left[ J_3(m_1) J_0(m_2) + J_2(m_1) J_1(m_2) \right] e^{{\rm i} 3 \omega t} - \left[ J_3(m_1) J_0(m_2) + J_2 (m_1) J_1(m_2) \right] e^{-{\rm i} 3 \omega t} + \cdots \right\}

Therefore what we want to realize is the following "extinction" condition
J_3(m_1) J_0(m_2) + J_2(m_1) J_1(m_2) = 0

We are in the small modulation regime. i.e. J0(m) = 1, J1(m) = m/2, J2(m) = m2/8, J3(m) = m3/48
Therefore we can simplify the above exitinction condition as

m_1 + 3 m_2 = 0

m2 < 0 means the start phase of the m2 modulation needs to be 180deg off from the phase of the m1 modulation.

E = E_0 \exp\left\{ {\rm i} [\Omega t + m_1 \cos \omega t + \frac{m_1}{3} \cos (5 \omega t + \pi)] \right \}

Field amplitude m1=0.3, m2=-0.1 m1=0.2, m2=0.2
Carrier 0.975 0.980
1st order sidebands 0.148 9.9e-2
2nd 1.1e-3 4.9e-3
3rd 3.5e-7 6.6e-4
4th 7.4e-3 9.9e-3
5th 4.9e-2 9.9e-2
6th 7.4e-3 9.9e-3
7th 5.6e-4 4.9e-4
8th 1.4e-5 4.1e-5
9th 1.9e-4 5.0e-4
10th 1.2e-3 4.9e-3
11th 1.9e-4 5.0e-4
12th 1.4e-5 2.5e-5
13th 4.7e-7 1.7e-6
14th 3.1e-6 1.7e-5
15th 2.0e-5 1.6e-4

 

  11019   Thu Feb 12 23:47:45 2015 KojiUpdateLSC3f modulation cancellation

- I built another beat setup on the PSL table at the South East side of the table.
- The main beam is not touched, no RF signal is touched, but recognize that I was present at the PSL table.
- The beat note is found. The 3rd order sideband was not seen so far.
- A PLL will be built tomorrow. The amplifier box Manasa made will be inspected tomorrow.

- One of the two beams from the picked-off beam from the main beam line was introduced to the beat setup.
(The other beam is used of for the beam pointing monitors)
There is another laser at that corner and the output from this beam is introduced into the beat setup.
The combined beam is introduced to PDA10CF (~150MHz BW).

- The matching of the beam there is poor. But without much effort I found the beat note.
  The PSL laser had 31.33 deg Xtal temp. When the beat was found, the aux laser had the Xtal temp of 40.88.

- I could observe the sidebands easily, with a narrower BW of the RF analizer I could see the sidebands up to the 2nd order.
  The 3rd order was not seen at all.

- The beat note had the amplitude of about -30dBm. One possibility is to amplify the signal. I wanted to use a spare channel
of the ALS/FOLL amplifier box. But it gave me rather attenuation than any amplification.
I'll look at the box tomorrow.

- Also the matching of two beams are not great. The PD also has clipping I guess. These will also be improved tomorrow

- Then the beat note will be locked at a certain frequency using PLL so that we can reduce the measurement BW more.

 

  11028   Sat Feb 14 00:48:13 2015 KojiUpdateLSC3f modulation cancellation

[SUCCESS] The 3f sideband cancellation seemed worked nicely.

- Beat effeciency improved: ~30% contrast (no need for amplification)

- PLL locked

- 3f modulation sideband was seen

- The attenuation of the 55MHz modulation and the delay time between the modulation source was adjusted to
have maximum reduction of the 3f sidebands as much as allowed in the setup. This adjustment has been done
at the frequency generation box at 1X2 rack.

- The measurement and receipe for the sideband cancellation come later.


- This means that I jiggled the modulation setup at 1X2 rack. Now the modulation setup was reverted to the original,
but just be careful to any change of the sensing behavior.

- The RF analyzer was returned to the control room.

- The HEPA speed was reduced from 100% (during the action on the table) to 40%.

  11029   Sat Feb 14 19:54:04 2015 KojiSummaryLSC3f modulation cancellation

Optical Setup

[Attachment 1]

Right before the PSL beam goes into the vacuum chamber, it goes through an AR-wedged plate.
This AR plate produces two beams. One of them is for the IO beam angle/position monitor.
And the other was usually dumped. I decided to use this beam.

A G&H mirror reflects the beam towards the edge of the table.
A 45deg HR mirror brings this beam to the beat set up at the south side of the table.
This beam is S-polarlized as it directly comes from the EOM.

[Attachment 2]

The beam from the PSL goes through a HWP and some matching lenses before the combining beam splitter (50% 45deg P).
The AUX laser beam is attenuated by a HWP and a PBS. The transmitted beam from the PBS is supposed
to have P-polarization. The beam alignment is usually done at the PSL beam side.

The combined beam is steered by a HR mirror and introduced to Thorlabs PDA10CF. As the PD has small diameter
of 0.5mm, the beam needed to be focused by a strong lens.

After careful adjustment of the beam mode matching, polarization, and alignment, the beatnote was ~1Vpp for 2.5Vdc.
In the end, I reduced the AUX laser power such that the beat amplitude went down to ~0.18Vpp (-11dBm at the PD,
-18dBm at the mixer, -27dBm at the spectrum analyzer) in order to minimize nonlinearity of the RF system and
in order that the spectrum analyzer didn't need input attenuation.

Electrical Setup

[Attachment 3]

The PD signal is mixed with a local oscillator signal at 95MHz, and then used to lock the PLL loop.
The PLL loop allows us to observe the peaks with more integration time, and thus with a better signal-to-noise ratio.

The signal from the PD output goes through a DC block, then 6dB attenuator. This attenuator is added to damp reflection
and distortion between the PD and the mixer. When the PLL is locked, the dominant signal is the one at 95MHz. Without this attenuator,
this strong 95MHz signal cause harmonic distortions like 190MHz. As a result, it causes series of spurious peaks at 190MHz +/- n* 11MHz.

10dB coupler is used to peep the PD signal without much disturbing the main line. Considering we have 6dB attanuator,
we can use this coupler output for the PLL and can use the main line for the RF monitor, next time.

The mixer takes the PD signal and the LO signal from Marconi. Marconi is set to have +7dBm output at 95MHz.
FOr the image rejection, SLP1.9 was used. The minicirsuit filters have high-Z at the stop band, we need a 50Ohm temrinator
between the mixer and the LPF.

The error signal from the LPF is fed to SR560 (G=+500, 1Hz 1st-order LPF). I still don't understand why I had to use a LPF
for the locking.
As the NPRO PZT is a frequency actuator, and the PLL is sensitive to the phase, we are supposed to use
a flat response for PLL locking. But it didn't work. Once we check the open loop TF of the system, it will become obvious (but I didn't).

The actuation signal is fed to the fast PZT input of the AUX NPRO laser.
 

  11031   Sat Feb 14 20:37:51 2015 KojiSummaryLSC3f modulation cancellation

Experimental results

- PD response [Attachment 1]

The AUX laser temperature was swept along with the note by Annalisa [http://nodus.ligo.caltech.edu:8080/40m/8369]
It is easier to observe the beat note by closing the PSL shutter as the MC locking yields more fluctuation of the PSL
laser freuqency at low frequency. Once I got the beat note and maximized it, I immediately noticed that the PD response
is not flat. For the next trial, we should use Newfocus 1611. For the measurement today, I decided to characterize the
response by sweeping the beat frequency and use the MAXHOLD function of the spectrum analyzer.

The measured and modelled response of the PD are shown in the attachment 1. It has non-intuitive shape.
Therefore the response is first modelled by two complex pole pair at 127.5MHz with Q of 1, and then the residual was
empirically fitted with 29th polynomial of f.

- Modulation profile of the nominal setting [Attachment 2]

Now the spectrum of the PD output was measured. This is a stiched data of the spectrum between 1~101MHz and 99~199MHz
that was almost simultaneously measured (i.e. Display 1 and Display 2). The IF bandwidth was 1kHz. The PD response correction
described above was applied.

It obviously had the peaks associated with our main modulations. In addition, there are more peaks seen.
The attachment 2 breaks down what is causing the peaks.

  • Carrier: The PLL LO frequency is 95MHz. Therefore the carrier is locked at 95MHz.
  • Modulation sidebands (11/55MHz series):
    Series of sidebands are seen at the both side of the carrier. Their frequency is 95MHz +/- n * fmod  (fmod = 11.066128MHz).
    Note that the sidebands for n>10 were above 200MHz, and n<-9 (indicated in gray) were folded at 0Hz.
    With this measurement BW, the following sidebands were buried in the noise floor.
    n = -8, -12, -13, and -14, n<= -16, and n>=+7
  • Modulation sidebands for IMC and PMC (29.5MHz and 35.5MHz):
    First order sidebands for the IMC and PMC modulations of sidebands are seen at the both side of the carrier.
    Their frequency is 95MHz +/- 29.5MHz or 33.5MHz. The PMC modulation sidebands are supposed to be blocked
    by the PMC. However, due to finite finesse of the PMC, small fraction of the PMC sidebands are transmitted.
    In deed, it is comparable to the modulation depth of the IMC one.
  • RF AM or RF EMI for the main modulation and the IMC modulationand:
    If there is residual RF AM in the PSL beam associated with the IMC and main modulations, it appears as the
    peaks at the modulation frequency and its harmonics. Also EM radiation couples into this measument RF system
    also appears at these frequencies. They are seen at n * fmod  (n=1,2,4,5) and 29.5MHz.
  • Reflection/distortion or leakage from mixer IF to RF:
    The IF port of the mixer naturally has 190MHz signal when the PLL is locked. If the isolation from the IF port to the RF port
    is not enough, this signal can appear in the RF monitor signal via an imperfection of the coupler or a reflection from the PD.
    Also, if the reflecrtion/distortion exist between the PD and the mixer RF input, it also cause the signal around 190MHz.
    It is seen at 190MHz +/- n* fmod. In the plot, the peak at n=0, -1 are visible. In fact these peak were secondarily dominant
    in the spectrum when there was no 6dB attenuation in the PD line. WIth the attenuator, they are well damped and don't disturb
    the main measurment.

From the measured peak height, we are able to estimate the modulation depths for 11MHz, 55MHz, IMC modulations, as well as
the relative phase of the 11MHz and 55MHz modulation. (It is not yet done).

- 3f modulation reduction [Attachment 3]

Now, the redcution of the 3f modulation was tried. The measured modulation levels for the 11MHz and 55MHz were almost the same.
The calculation predicts that the modulation for the 55MHz needs to be 1/3 of the 11MHz one. Therefore the attenuation of 9dB and 10dB
of the modulation attenuation knob at the frequency generation box were tried.

To give the variable delay time in the 55MHz line, EG&G ORTEC delay line unit was used. This allows us to change the delay time from
0ns to 63.5ns with the resolution of 0.5ns. The frequency of 55MHz yields a phase sensitivity of ~20deg/ns (360deg/18ns).
Therefore we can adjust the phase with the precision of 10deg over 1275deg.

The 3rd-order peak at 61.8MHz was observed with measurement span of 1kHz with very narrow BW like 30Hz(? not so sure). The delay
time was swept while measuring the peak height each time. For both the atteuation, the peak height clearly showed the repeatitive dependence
with the period of 18ns, and the 10dB case gave the better result. The difference between the best (1.24e-7 Vpk) and the worst (2.63e-6 Vpk)
was more than a factor of 20.
The 3rd-order peak in the above broadband spectrum measurement was 6.38e-6 Vpk. Considering the attenuation
of the 55MHz modulation by 10dB, we were at the exact unluck phase difference.
The improvement expected from the 3f reduction (in the 33MHz signal)
will be about 50, assuming there is no other coupling mechanism from CARM to REFL33.

I decided to declare the best setting is "10dB attenuation & 28ns delay".

- Resulting modulation profile [Attachment 4]

As a confirmation, the modulation profie was measured as done before the adjustment.
It is clear that the 3rd-order modulation was buried in the floor noise. 10dB attenuation of the 55MHz modulation yields corresponding reduction of the sidebands.
This will impact the signal quality for the 55MHz series error signals, particularly 165MHz ones. We should consider to install the Teledyne Cougar amplifier
next to the EOM so that we can increase the over all modulation depth.

  1354   Wed Mar 4 12:38:07 2009 AlbertoUpdateComputer Scripts / Programs3f locking simulations
I simulated the REFL signals demodulated at the differential frequencies of the sidebands (f2-f1), at their summed frequencies (f2+f1). I also simulated their combination as in the Double Demodulation.
 
I repeated the simulation for:
- Old (current) 40m
- 40m Upgrade
- AdvLIGO
 
I'm attaching the results to this elog entry.
 
The plots show how the signal varies exploring the two-dimensional space of the demodulation frequencies (differential and sum).

 Both the Upgrade and the Old40m's signals look anomalous since the zero-crossing point does not change with the demodulation phases.

I suspect there's is a problem with the optickle model of the 40m.

  5377   Sat Sep 10 14:55:28 2011 KeikoUpdateLSC3f demodulation board check

 To check the demodulation boards for REFL33 and REFL165, a long cable from ETMY (SUS-ETMY-SDCOIL-EXT monitor) is pulled to the rack on Y side.

(1) A filter just after the RF input and (2) transfer function from the RF input to the demodulated signal will be checked for the two 3f demod boards to confirm that they are appropriate for 33 and 165 MHz.

  5378   Sat Sep 10 16:10:42 2011 KeikoUpdateLSC3f demodulation board check

There is a LP filter just after the RF input of an demodulation board (its schematic can be found as D990511-00-C on DCC). I have checked if the 3f freq, 33MHz, can pass  this filter. The filter TF from the RF input to RF monitor (the filter is between the input and monitor) on REFL33 demo-board was measured as shown in Fig. 1. At 33MHz, the magnitude is still flat and OK, but the phase is quite steep. I am going to consider if it is ok for the PDH method or not.

REFL33-input-filter.png

 Fig. 1 Transfer function from the RF input to RF monitor on the REFL33 demodulation board. At 33MHz, a very steep phase is applied on the input signal.

Quote:

 To check the demodulation boards for REFL33 and REFL165, a long cable from ETMY (SUS-ETMY-SDCOIL-EXT monitor) is pulled to the rack on Y side.

(1) A filter just after the RF input and (2) transfer function from the RF input to the demodulated signal will be checked for the two 3f demod boards to confirm that they are appropriate for 33 and 165 MHz.

 

  5380   Sat Sep 10 18:57:52 2011 KeikoUpdateLSC3f demodulation board check

The phase delay due to the RF input filter on the demodulation board will not bother the resulting PDH signals.

I quickly calculated the below question (see the blue sentence in the quote below). I applied an arbitrary phase delay (theta) due to the filter I measured, on the detected RF signal by the photo detector. Then the filtered RF signal is multiplied by cos(omega_m) then filter the higher (2 omega_m) freqency as the usual mixing operation for the PDH signal. As a result, the I signal is delayed by cos(theta) and the Q signal is delayed by sin(theta). Therefore the resulting signals and its orthogonalitity is kept ok. From the sideband point of view, theta is applied on both upper and lower and seems to make the unbalance, however, as it is like a fixed phase offset on both SBs at the modulation frequency, the resulting signals is just multiplied by cos or sin theta for I and Q, respectively. It won't make any strange effect (it is difficult to explain by sentence not using equations!).

Quote:

There is a LP filter just after the RF input of an demodulation board (its schematic can be found as D990511-00-C on DCC). I have checked if the 3f freq, 33MHz, can pass  this filter. The filter TF from the RF input to RF monitor (the filter is between the input and monitor) on REFL33 demo-board was measured as shown in Fig. 1. At 33MHz, the magnitude is still flat and OK, but the phase is quite steep. I am going to consider if it is ok for the PDH method or not.

 Fig. 1 Transfer function from the RF input to RF monitor on the REFL33 demodulation board. At 33MHz, a very steep phase is applied on the input signal.

Quote:

 To check the demodulation boards for REFL33 and REFL165, a long cable from ETMY (SUS-ETMY-SDCOIL-EXT monitor) is pulled to the rack on Y side.

(1) A filter just after the RF input and (2) transfer function from the RF input to the demodulated signal will be checked for the two 3f demod boards to confirm that they are appropriate for 33 and 165 MHz.

 

 

  5385   Sun Sep 11 22:36:32 2011 KeikoUpdateLSC3f demodulation board check

Filters at the RF inputs of REFL33 and REFL165 demodulation boards were measured again. The filters will be totally fine for 33MHz and 165MHz.

Last time I forgot to calibrate the cable lengths, therefore the phase delay of the measurement included the cable lengths. This time the measurements were done for REFL33 and REFL165 demod board with calibration. As the cable lengths were calibrated, the shown plots (Fig.1 and Fig.2) do not include the phase delay dues to measurement cables. Please note that the x-axis is in linear. The phase delays of both boards seems to be not too steep (it will not affect anyway, as Kiwamu pointed out in his comment on the previous post). You can see that the two filters do not filter 33MHz and 165MHz component out.

REFL33.png

Fig.1 A response of a filter which is placed just after the RF input of the demodulation board for REFL33. X-axis is shown in linear (~50MHz).

REFL165.png

Fig.2 A response of a filter which is placed just after the RF input of the demodulation board for REFL165.

 

Quote:

There is a LP filter just after the RF input of an demodulation board (its schematic can be found as D990511-00-C on DCC). I have checked if the 3f freq, 33MHz, can pass  this filter. The filter TF from the RF input to RF monitor (the filter is between the input and monitor) on REFL33 demo-board was measured as shown in Fig. 1. At 33MHz, the magnitude is still flat and OK, but the phase is quite steep. I am going to consider if it is ok for the PDH method or not.

REFL33-input-filter.png

 Fig. 1 Transfer function from the RF input to RF monitor on the REFL33 demodulation board. At 33MHz, a very steep phase is applied on the input signal.

Quote:

 To check the demodulation boards for REFL33 and REFL165, a long cable from ETMY (SUS-ETMY-SDCOIL-EXT monitor) is pulled to the rack on Y side.

(1) A filter just after the RF input and (2) transfer function from the RF input to the demodulated signal will be checked for the two 3f demod boards to confirm that they are appropriate for 33 and 165 MHz.

 

 

  5386   Mon Sep 12 13:24:07 2011 KeikoUpdateLSC3f demodulation board check

I also quickly checked the orthogonality of the demodulation board for REFL33 and REFL165 using function generators and oscilloscope. I checked the frequencies at 1,10,100,1K,10KHz of the demodulated signals. They are fine and ready for 3f signal extraction.

  5387   Mon Sep 12 16:27:01 2011 KeikoUpdateLSC3f demodulation board check

Wait. I am checking the whitening filters of the 33 and 165 demodulation boards.

Also, LSC-REFL33-I-IN1(IN2, OUT) and LSC-REFL165-Q-IN1(IN2,OUT) channels may not be working??

 

Quote:

I also quickly checked the orthogonality of the demodulation board for REFL33 and REFL165 using function generators and oscilloscope. I checked the frequencies at 1,10,100,1K,10KHz of the demodulated signals. They are fine and ready for 3f signal extraction.

 

  5388   Mon Sep 12 18:40:35 2011 KeikoUpdateLSC3f demodulation board check

LSC-REFL33-I-IN1(IN2, OUT) and LSC-REFL165-Q-IN1(IN2,OUT) channels are back!

We disconnected and connected again the AA filters then the channels are fixed. Apparently the AA filters just before the digital world were somhow charged and not working... Thank you Kiwamu!

Quote:

Wait. I am checking the whitening filters of the 33 and 165 demodulation boards.

Also, LSC-REFL33-I-IN1(IN2, OUT) and LSC-REFL165-Q-IN1(IN2,OUT) channels may not be working??

 

  5394   Tue Sep 13 15:00:25 2011 KeikoUpdateLSC3f demodulation board check

Whitening filters for the REFL33 & 165 demodulated channels were measured and confirmed that they are working. They can be turned on and off by un-white filter switches on the MEDM screen because they are properly linked. The measured filter responses are showen below. (Sorry, apparentyl the thumbnails are not shown here. Please click the attachments.) 

WF33.pdf

WF165.pdf

Attachments: (top) Whitening filter for REFL33 demodulation board. (bottom) Whitening filter response for REFL 165 demodulation board.

  5399   Tue Sep 13 23:08:51 2011 KeikoUpdateLSC3f demodulation board check

Keiko, Jamie , Kiwamu

The I and Q orthogonalities of REFL33 and 165 demodulation board were measured by "orthogonality.py"  Python package scipy were addied on Pianosa to run this code. Please note that "orthogonality.py" can be run only on Pianosa.  

The results were:

REFL165

ABS = 1.070274, PHASE = -81.802479 [deg]

if you wanna change epics values according to this result, just copy and execute the following commands

ezcawrite C1:LSC-REFL165_Q_GAIN 0.934340 && ezcawrite C1:LSC-REFL165_PHASE_D -81.802479

- - - - - - - - - - - - - - - - - -

REFL33

ABS = 1.016008 PHASE = -89.618724 [deg]

if you wanna change epics values according to this result, just copy and execute the following commands

ezcawrite C1:LSC-REFL33_Q_GAIN 0.984244 && ezcawrite C1:LSC-REFL33_PHASE_D -89.618724


Fig.1 and 2  are the resulting plots for 33 and 165 MHz demod baoards, respectively.You should look at the 3Hz in x axis, as the demodulated signal frequency was set as 3 Hz.REFL33-modified.png

Fig. 1 REFL33 I and Q orthogonality at 3 Hz.

REFL165-modified.png

 

Fig. 2 REFL165 I and Q orthogonality at 3 Hz.

 

 

 

  5412   Thu Sep 15 01:06:20 2011 KeikoUpdateLSC3f demodulation board check

In addition to REFL 33 ans 165, I checked the orthogonality for the other existing three channels.

 

AS11

ABS = 1.025035  PHASE = -93.124929 [deg]

REFL11]

ABS = 0.920984  PHASE = -88.824691 [deg]

REFL55

ABS = 1.029985 , PHASE = -90.901123 [deg]

- - - - - - - - - - - - - - - - - -

 

The demodulated signal was set as 50 Hz (for example LO 11MHz and RF 11MHz+50Hz from function generators.) These AS11, REFL11, REL55, REFL33m REFL165 are the current available channels in terms of the connection to the data system from the demodulation board. I am going to estimate the error next.

Quote:

REFL165

ABS = 1.070274, PHASE = -81.802479 [deg]

- - - - - - - - - - - - - - - - - -

REFL33

ABS = 1.016008 PHASE = -89.618724 [deg]

 

  5440   Fri Sep 16 21:26:12 2011 KeikoUpdateLSC3f demodulation board check

The demodulation phases and gains for the all existing channels, AS11, REFL11,REFL55, REFL165, and REFL33, were adjusted by the command "ezcawrite" commands. 

Scripts are: 

REFL165 ezcawrite C1:LSC-REFL165_Q_GAIN 0.934340 && ezcawrite C1:LSC-REFL165_PHASE_D -81.802479

REFL33

ezcawrite C1:LSC-REFL33_Q_GAIN 0.984244 && ezcawrite C1:LSC-REFL33_PHASE_D -89.618

REFL11

ezcawrite C1:LSC-REFL11_Q_GAIN 1.173418 && ezcawrite C1:LSC-REFL11_PHASE_D -442.882697

AS11
ezcawrite C1:LSC-AS11_Q_GAIN 0.975576 && ezcawrite C1:LSC-AS11_PHASE_D -93.12492

AS55

ezcawrite C1:LSC-AS55_Q_GAIN 0.999164 && ezcawrite C1:LSC-AS55_PHASE_D -89.300986
  9976   Tue May 20 16:48:52 2014 ericqUpdateLSC3f Stability

So, I really should have done this as soon as Manasa measured the arm lengths... I've updated my MIST model with the real arm lengths, but still am using assumed identical losses of 75ppm on each mirror. (I've tried measuring the arm losses for real, but got numbers in the hundreds of ppms, so I need to reexamine things...) 

Here's a simulation of the fields in a perfectly locked PRC when CARM is swept (Normalized to input power = 1). 

CARMsweepPrcFields.pdf

More importantly, here's the latest simulation of MICH vs. PRCL demodulation angle separation in the 3F signals. It seems that we may be getting burned by using REFL33 for the PRC lock. REFL165, on the other hand looks much more robust. We should try this out. 

3fs.pdf

(Some of my previous simulations incorrectly implemented MICH excitations; I only moved the ITMS, not the ETMS along with them, so some other stuff slipped in... )

  9977   Tue May 20 22:42:28 2014 ericqUpdateLSC3f Stability

 Here's the angles of MICH and PRCL from the my earlier plot by themselves; this shows that the individual demod angles in REFL165 aren't changing much either. 

PRCangles.pdf

  5401   Wed Sep 14 01:19:20 2011 AnamariaConfigurationLSC3f PD Install in Progress

I have reconfigured the refl beam path on the AP table to include REFL33 and REFL165. Would be done if we hadn't prepared P BSs instead of S, which required some serious digging to find two others. And if someone hadn't stolen our two 3m SMA cables that Keiko and I made on our previous visit and I had left with the 3f PDs. I don't expect them to reappear but if they do, it would be grand.

Note: Refl beam from ifo looks a bit high, ~1cm on the lens 20'' from output port. Not sure what that means about ifo alignment change, I've left it as is. When we know we have a good alignment, we should be able to easily realign the beam path if necessary. If it remains the same, we might want to change the lens height.

Done:

1) REFL11 and REFL55 are now hooked up and aligned in a low power beam. (I set the power as low as I could by eye to not risk burning the PDs during alignment)

2) The required BSs and REFL33 and REFL165 are in place, powered.

3) I have set them in a configuration such that the beam is the same distance from the main beam, to adjust beam size easily for all 4.

4) Camera has been moved from main beam to behind a steering mirror, ND filters removed, centered on camera.

To Do:

1) Find one more longish SMA cable.

2) Align beam on REFL33 and REFL165.

3) Check beam size carefully. (I get a plateau on the scope, and I can "hide" the beam on the PD, but it could be better. The path has become longer by ~5-8inches.)

4) Adjust power.

5) Redo layout diagram, post in wiki.

  10698   Tue Nov 11 21:41:09 2014 KojiUpdateLSC3f DRMI sensing mat

Sensing matrix calculation using DTT + Matlab

Note: If the signal phase is, for example,  '47 deg', the phase rotation angle is -47deg in order to bring this signal to 'I' phase.

Note2: As I didn't have the DQ channels for the actuation, only the relative signs between the PDs are used to produce the radar chart.
This means that it may contain 180deg uncertainty for a particular actuator. But this does not change the independence (or degeneracy) of the signals.



=== Sensing Matrix Report ===
Test time: 2014-11-11 08:14:00
Starting GPS Time: 1099728855.0
 

== PRCL ==
Actuation frequency: 621.13 Hz
Suspension (PRM) response at the act. freq.: 5.0803e-14/f^2 m/cnt
Actuation amplitude: 20.3948 cnt/rtHz
Actuation displacement: 1.0361e-12 m/rtHz
 
C1:LSC-AS55_I_ERR_DQ 4.20e+10
C1:LSC-AS55_Q_ERR_DQ -1.91e+11
==> AS55: 1.95e+11 [m/cnt] -24.58 [deg]
C1:LSC-REFL11_I_ERR_DQ 3.17e+12
C1:LSC-REFL11_Q_ERR_DQ -8.04e+10
==> REFL11: 3.17e+12 [m/cnt] -18.20 [deg]
C1:LSC-REFL33_I_ERR_DQ 4.15e+11
C1:LSC-REFL33_Q_ERR_DQ 4.28e+10
==> REFL33: 4.17e+11 [m/cnt] -137.11 [deg]
C1:LSC-REFL55_I_ERR_DQ 1.90e+10
C1:LSC-REFL55_Q_ERR_DQ -9.91e+09
==> REFL55: 2.14e+10 [m/cnt] -58.58 [deg]
C1:LSC-REFL165_I_ERR_DQ -1.16e+11
C1:LSC-REFL165_Q_ERR_DQ -3.14e+10
==> REFL165: 1.20e+11 [m/cnt] 45.20 [deg]
 
 
== MICH ==
Actuation frequency: 675.13 Hz
Suspension (ITMX) response at the act. freq.: 1.0312e-14/f^2 m/cnt
Suspension (ITMY) response at the act. freq.: 1.0224e-14/f^2 m/cnt
Actuation amplitude: 974.2957 cnt/rtHz
Actuation displacement (ITMX+ITMY): 2.0007e-11 m/rtHz
 
C1:LSC-AS55_I_ERR_DQ 2.55e+12
C1:LSC-AS55_Q_ERR_DQ 4.51e+12
==> AS55: 5.18e+12 [m/cnt] 113.51 [deg]
C1:LSC-REFL11_I_ERR_DQ -4.84e+10
C1:LSC-REFL11_Q_ERR_DQ -4.07e+09
==> REFL11: 4.85e+10 [m/cnt] 168.06 [deg]
C1:LSC-REFL33_I_ERR_DQ 2.06e+10
C1:LSC-REFL33_Q_ERR_DQ -9.39e+09
==> REFL33: 2.26e+10 [m/cnt] -167.51 [deg]
C1:LSC-REFL55_I_ERR_DQ 2.52e+09
C1:LSC-REFL55_Q_ERR_DQ -1.02e+10
==> REFL55: 1.05e+10 [m/cnt] -107.09 [deg]
C1:LSC-REFL165_I_ERR_DQ -1.79e+10
C1:LSC-REFL165_Q_ERR_DQ -5.50e+10
==> REFL165: 5.79e+10 [m/cnt] 102.02 [deg]



== SRCL ==

Actuation frequency: 585.13 Hz
Suspension (SRM) response at the act. freq.: 5.5494e-14/f^2 m/cnt
Actuation amplitude: 1176.3066 cnt/rtHz
Actuation displacement: 6.5278e-11 m/rtHz
 
C1:LSC-AS55_I_ERR_DQ -9.90e+10
C1:LSC-AS55_Q_ERR_DQ -1.18e+11
==> AS55: 1.54e+11 [m/cnt] -76.89 [deg]
C1:LSC-REFL11_I_ERR_DQ 2.96e+08
C1:LSC-REFL11_Q_ERR_DQ 4.78e+08
==> REFL11: 5.62e+08 [m/cnt] 41.42 [deg]
C1:LSC-REFL33_I_ERR_DQ -2.93e+09
C1:LSC-REFL33_Q_ERR_DQ 1.23e+10
==> REFL33: 1.27e+10 [m/cnt] -39.63 [deg]
C1:LSC-REFL55_I_ERR_DQ 3.71e+09
C1:LSC-REFL55_Q_ERR_DQ 2.78e+09
==> REFL55: 4.63e+09 [m/cnt] 5.86 [deg]
C1:LSC-REFL165_I_ERR_DQ -1.80e+10
C1:LSC-REFL165_Q_ERR_DQ 2.68e+10
==> REFL165: 3.23e+10 [m/cnt] -26.02 [deg]
 


Demodulation phases of the day

    'C1:LSC-AS55_PHASE_R = -53'
    'C1:LSC-REFL11_PHASE_R = 16.75'
    'C1:LSC-REFL33_PHASE_R = 143'
    'C1:LSC-REFL55_PHASE_R = 31'
    'C1:LSC-REFL165_PHASE_R = 150'

  10701   Wed Nov 12 03:22:23 2014 JenneUpdateLSC3f DRMI sensing mat

Koji pointed out something to me that I think he had told me ages ago, and Rana alluded to last night:  Since I'm not tuning my "demod phase" for the sensing matrix lockins, unless I happened to get very lucky, I was throwing away most of the signal.  Lame.  

So, now the magnitude is sqrt(real^2 + imag^2), where real and imag here are the I and Q outputs of the lockin demodulator, after the 0.1Hz lowpass.  (I put in the low pass into all of the Q filter banks).  To keep the signs consistent, I did do a rough tuning of those angles, so that I can use the sign of the real part as the sign of my signal.  When I was PRMI locked, I set the phase for all things acutated by MICH to be 79deg, all things actuated by PRCL to be 20 deg, and when DRMI locked set all things SRCL to be 50deg. 

After doing this, the phases of my sensing matrix output matches Koji's careful analyses.  I don't know where the W/ct numbers I was using came from (I don't think I made them up out of the blue, but I didn't document where they're from, so I need to remeasure them).  Anyhow, for now I have 1's in the calibration screen for the W/ct calibration for all PDs, so my sensing matrices are coming out in cts/m, which is the same unit that Koji's analysis is in. (Plot for comparing to Koji is at end of entry).

While reducing the CARM offset, I left the sensing matrix lines on, and watched how they evolved.  The phases don't seem to change all that much, but the magnitudes start to decrease as I increase the arm power.

For this screenshot, the left plot is the phases of the sensing matrix elements (all the REFL signals, MICH and PRCL), and the right plot is the magnitudes of those same elements.  Also plotted is the TRX power, as a proxy for CARM offset.  The y-scale for the TRX trace is 0-15.  The y-scale for all the phases is -360 to +360.  The y-scale of the magnitude traces are each one decade, on a log scale.

SensMatVsPower_UpToArms10.png

Bonus plot, same situation, but the next lock held for 20 minutes at arm powers of 8.  We don't know why we lost lock (none of the loops were oscillating, that I could see in the lockloss plot).

PRMI_arms8_20minutes.png


Here are some individual sensing matrix plots, for a single lock stretch, at various times.  One thing that you can see in the striptool screenshots that I don't know yet how to deal with for the radar plots is the error bars when the phase flips around by 360 degrees.  Anyhow, the errors in the phases certainly aren't as big as the error boxes make them look.

PRMI just locked, CARM offset about 3nm, CARM and DARM on ALS comm and diff, arm powers below 1:

SensMatMeas_11Nov2014_PRMIarms_ArmPowSmall.png

PRMI still on REFL33 I&Q, CARM and DARM both on DC transmissions, arm powers about 4:

SensMatMeas_11Nov2014_PRMIarms_ArmPow3pt8.png

CARM offset reduced further, arm powers about 6:

SensMatMeas_11Nov2014_PRMIarms_ArmPow6.png

CARM offset reduced even more, arm powers about 7:

SensMatMeas_11Nov2014_PRMIarms_ArmPow7.png


For this plot for comparing with Koji's analysis, I had not yet put 1's in the calibration screen, so this is still in "W"/m, where "W" is meant to indicate that I don't really know the calibration at all.  What is good to see though is that the angles agree very well with Koji's analysis, even though he was analyzing data from yesterday, and this data was taken today.  This sensing matrix is DRMI-only (no arms), 1f locking.

SensMatMeas_11Nov2014_DRMI_fixedMags.png

Bonus plot, PRMI-only sensing matrix, with PRMI held using REFL 33 I&Q:

SensMatMeas_11Nov2014_PRMI_fixedMags.png

 

  10696   Tue Nov 11 03:48:46 2014 JenneUpdateLSC3f DRMI

I was able to lock the DRMI on 3f signals this evening, although the loops are not stable, and I can hear oscillations in the speakers.  I think the big key to making this work was the placement of the SHP-150 high pass filter at the REFL165 PD, and also the installation of Koji's 110 MHz notch filter just before the amplifier, which is before the demod board for REFL165.  These were done to prevent RF signal distortion.

DRMI 3f:   With DRMI locked on 1f (MICH gain = 1, PRCL gain = -0.05, SRCL gain = 2, MICH = 1*REFL55Q, PRCL = 0.1*REFL11I, SRCL = 1*REFL165I), I excited lines, and found the signs and values for 3f matrix elements.  I was using the same gains, but MICH = 0.5*REFL165Q, PRCL = 0.8*REFL33I and SRCL = -0.2*REFL165I.  Part of the problem is likely that I need to include off-diagonal elements in the input matrix to remove PRCL from the SRCL error signal. 

With the DRMI locked on 1f, I took a sensing matrix measurement.  I don't think we believe the W/ct of the photodiode calibration (we need to redo this), but otherwise the sensing matrix measurement should be accurate.  Since the calibration of the PDs isn't for sure, the relative magnitude for signals between PDs shouldn't be taken as gospel, but within a single PD they should be fine for comparison. 

As a side note, we weren't sure about the MICH -> ITMs balancing, so we checked during a MICH-only, and with the locking apparatus we are unable to measure a difference between 1's for both ITMs in the output matrix, and 1 for ITMX and 0.99 for ITMY.  So, we can't measure 1% misbalances in the actuator, but we think we're at least pretty close to driving pure MICH. 

We kind of expect that SRCL should only be present in the 55 and 165 PDs, although we see it strongly in all of the REFL PDs.  Also, PRCL and SRCL are not both lined up in the I-phase.  So, I invite other people to check what they think the sensing matrix looks like. 

  • The excitation lines (and matching notches) were on from 12:14am (
  • Nov 11 2014 08:14:00 UTC / GPS 1099728856) to 12:20am (
  •  
  • Nov 11 2014 08:20:00 UTC / GPS 
  • 1099729216) for 360sec. 
  • MICH was driven with 800 counts at 675.13 Hz, with +1*ITMY, -1*ITMX. 
  • PRCL was driven with 1000 counts at 621.13 Hz with the PRM. 
  • SRCL was driven with 800 counts at 585.13 Hz using the SRM. 

All 3 degrees of freedom have notches at all 3 of those frequencies in the FM10 of the filter banks (and they were all turned on).  During this time, DRMI was locked with 1f signals. 

DRMI sensing matrix:

 SensMatMeas_10Nov2014_DRMI.png

Earlier in the evening, I also took a PRMI sensing matrix, with the PRMI locked on REFL33 I&Q.  Watch out for the different placement of the plots - I couldn't measure AS55 in the DRMI case, since cdsutils.avg freaked out if I asked for more than 14 numbers (#PDs * #dofs) at a time.

SensMatMeas_10Nov2014_PRMI.png

Rana, Koji and I were staring at the DRMI sensing matrix for a little while, and we aren't sure why PRCL and SRCL aren't co-aligned, and why they aren't orthogonal to MICH.  Do we think it's possible to do something to digitally realign them?  Will the solution that we choose be valid for many CARM offsets, or do we have to change things every few steps (which we don't want to do)? 

Things to work on:

* Reanalyze DRMI sensing matrix data from 12:14-12:20am. 

* Make a simulated scan of higher order mode resonances in the arm cavities.  Is it possible that on one or both sides of the CARM resonance we are getting HOM resonances of the sidebands? 

* Figure out how to make DRMI 3f loops stable.

* Try CARM offset reduction with DRMI, and / or PRMI on REFL 165.

  10668   Wed Nov 5 01:58:54 2014 ericqUpdateLSC3F RFPD RF spectra

Given the checkout of the aLIGO BBPDs happening (aLOG link), wherein the PDs were acting funny, and Koji has made some measurements determining that intermodulation/nonlinearity of circuitry can corrupt 3F signals, I've made a similar measurement of the RF spectra of REFL165 when we're locked on DRMI using 1F signals. Maybe this could give us insight to our bad luck using REFL165...

In essence, I plugged the RF output of the PD into an AG4395, through a 10dB attenuator and downloaded the spectrum. I also did REFL33 as a possible comparison and because why not. The attached plots have the 10dB accounted for; the text files do not. 

REFL165 (Exposed PCB BBPD):

REFL165_DRMIspectrum.png

(What is all that crap between 8 and 9 fmod?)

REFL33 (Gold Box resonant RFPD):

REFL33_DRMIspectrum.png

  10669   Wed Nov 5 11:09:44 2014 KojiUpdateLSC3F RFPD RF spectra

If you look at the intermodulation at 14 (4+10) and 16 (6+10), 15 (5+10) would make any problem, thanks to the notch at 1f and 5f.

BUT, this absolute level of 165MHz is too tiny for the demodulator. From the level of the demodulated signal, I can say REFL165 has
too little SNR. We want to amplify it before the demodulator.

Can you measure this again with a directional coupler instead of the direct measurement with an attenuator?
The downstream has bunch of non-50Ohm components and may cause unknown effect on the tiny 165MHz signal.
We want to measure the spectrum as close situation as possible to the nominal configuration.

90MHz crap is the amplifier noise due to bad power bypassing or bad circuit shielding.

I have no comment on REFL33 as it has completely different amplification stages.

  10673   Wed Nov 5 22:25:42 2014 ericqUpdateLSC3F RFPD RF spectra

 

Now that I have followed the chain, the PD signal is indeed being amplified at the LSC rack. It goes into a ZFL-1000LN+ amplifier (~23dB gain at 165MHz and 15V supply), followed by a SHP-100 high pass filter, and then enters the RF IN of the demod board. 

I repeated the measurement in two spots.

First, I took a spectrum of the RF MON of the REFL165 demod board during DRMI lock; this was input-referred by adding 20dBm. 

Second, I inserted a ZFDC-10-5 coupler between the high pass and the RF input of the demod board. This was input-referred by adding 10dBm. 

REFL165_demod_DRMIspectrum.png

My calibration isn't perfect; the peaks above the high pass corner seem to be different by a consistent amount, but within a few dBm. 

Thus, it looks like the demod board is getting a little under -40dBm of 165MHz signal at its input. 

  10675   Thu Nov 6 01:58:55 2014 KojiUpdateLSC3F RFPD RF spectra

Where is the PD out spectrum measured with the coupler???

  10679   Thu Nov 6 11:49:58 2014 ericqUpdateLSC3F RFPD RF spectra

Quote:

Where is the PD out spectrum measured with the coupler???

 The "coupled" port of the coupler went to the AG4395 input, the output of the Highpass is connected to the "IN", and the "OUT" goes to the demod board. 

  10682   Thu Nov 6 14:41:49 2014 KoijUpdateLSC3F RFPD RF spectra

That's not what I'm asking.

Also additional cables are left connected to the signal path. I removed it.

  10683   Fri Nov 7 02:21:12 2014 ericqUpdateLSC3F RFPD RF spectra

 

After some enlightening conversation with Koji, we figured that the RF amplifier in the REFL165 chain is probably being saturated (the amp's 1dB compression is at +3dBm, has 23dB gain, and there are multiple lines above -20dBm coming out of the PD). I took a few more spectrum measurements to quantify the consequences, as well as a test with the highpass connected directly to the PD output, that should reduce the power into the amplifier. However, I am leaving everything hooked back up in its original state (and have removed all couplers and analyzers...)

I also took some DRMI sensing measurements. In the simple Michelson configuration, I took TFs of each ITMs motion to AS55Q to make sure the drives were well balanced. They were. Then, in the DRMI, I took swept sine TFs of PRCL, SRCL and differential ITM MICH motion to the Is and Qs of AS55 and all of the REFLs. I constrained the sweeps to 300Hz->2kHz; the loops have some amount of coupling so I wanted to stay out of their bandwidth. I also took a TF of the pure BS motion and BS-PRM MICH to the PDs. From these and future measurements, I hope to pursue better estimates of the sensing matrix elements of the DRMI DoFs, and perhaps the coefficients for compensating both SRCL and PRCL out of BS motion. 

I'm leaving analysis and interpretation for the daytime, and handing the IFO back to Diego...

  10685   Fri Nov 7 14:41:18 2014 ericqUpdateLSC3F RFPD RF spectra

Quote:

 After some enlightening conversation with Koji, we figured that the RF amplifier in the REFL165 chain is probably being saturated.

The measurements I took yesterday bear this out. However, even putting the high-pass directly on the PD output doesn't reduce the signal enough to avoid saturating the amplifier.

We need to think of the right way to get the 165MHz signal at large enough, but undistorted, amplitude to the demod board. 


 The current signal chain looks like:

AS Table                                  LSC RACK
[ PD ]----------------------------------->[ AMP ]------>[ 100MHzHPF ]----->[ DEMOD ]
      (1)                                        (2)                 (3)

I previously made measurements at (3). Let's ignore that. 

Last night, I took measurements with a directional coupler at points (1) and (2), to see the signal levels before and after the amplifier. I divided the spectrum at (2) by the nominal gain of the amplifier, 23.5dB; thus if everything was linear, the spectra would be very similar. This is not the case, and it is evident why. There are multiple signals stronger than -20dBm, and the amplifier has a 1dB compression point of +3dBm, so any one of these lines at 4x, 6x and 10x fMod is enough to saturate. 

 165_ampSaturation.png


I also made a measurement at point 4 in the following arrangement, in an attempt to reduce the signal amplitude incident on the amplifier.  

AS Table                                           LSC RACK
[ PD ]->[ 100MHzHPF ]----------------------------------->[ AMP ]--------->[ DEMOD ] 
                                                                (4) 

 Though the signals below 100MHz are attenuated as expected, the signal at 110MHz is still too large for the amplifier. 

165_HPatPD.png


Minicircuits' SHP-150 only has 13dB suppression at 110MHz, which would not be enough either. SHP-175 has 31dB suppression at 110MHz and 0.82dB at 160MHz, maybe this is what we want.

  10686   Fri Nov 7 16:15:53 2014 JenneUpdateLSC3F RFPD RF spectra

I have found an SHP-150, but no SHP-175's (also, several 200's, and a couple of 500's).

Why do you say the SHP-150 isn't enough?  The blue peak at 10*fmod in your plot looks like it's at -12 dBm.  -13 dB on top of that will leave that peak at -25 dBm.  That should be enough to keep us from saturation, right?  It's not a lot of headroom, but we can give it a twirl until a 175-er comes in.  

Koji also suggests putting in a 110 MHz notch, combined with either an SHP-150 or SHP-175, although we'll have to measure the combined TF to make sure the notch doesn't spoil the high pass's response too much.

Quote:

165_HPatPD.png


Minicircuits' SHP-150 only has 13dB suppression at 110MHz, which would not be enough either. SHP-175 has 31dB suppression at 110MHz and 0.82dB at 160MHz, maybe this is what we want.

 

  10689   Sat Nov 8 11:35:05 2014 ranaUpdateLSC3F RFPD RF spectra

 

 I think 'saturation' here is a misleading term to think about. In the RF amplifiers, there is always saturation. What we're trying to minimize is the amount of distorted waveforms appearing at 3f and 15f from the other large peaks. Usually for saturation we are worried about how much the big peak is getting distorted; not the case for us.

  10692   Mon Nov 10 18:11:57 2014 ericqUpdateLSC3F RFPD RF spectra

 Jenne and I measured the situation using a SHP-150 directly attached to the REFL165 RF output, and at first glance, the magnitude of the 165MHz signal seems to not be distorted by the amplifier. 

 165signals.pdf

We will soon investigate whether 165 signal quality has indeed improved. 

  11010   Thu Feb 12 03:43:54 2015 ericqUpdateLSC3F PRMI at zero ALS CARM

I have been able to recover the ability to sit at zero CARM offset while the PRMI is locked on RELF33 and CARM/DARM are on ALS, effectively indefinitely. However, I feel like the transmon QPDs are not behaving ideally, because the reported arm powers freqently go negative as the interferometer is "buzzing" through resonance, so I'm not sure how useful they'll be as normalizing signals for REFL11. I tried tweaking the DARM offset to help the buildup, since ALS is only roughly centered on zero for both CARM and DARM, but didn't have much luck.

Example:

Turning off the whitening on the QPD segments seems to make everything saturate, so some thinking with daytime brain is in order.


How I got there:

It turns out triggering is more important than the phase margin story I had been telling myself. Also, I lost a lot of time to needing demod angle change in REFL33. Maybe I somehow caused this when I was all up on the LSC rack today?

We have previously put TRX and TRY triggering elements into the PRCL and MICH rows, to guard against temporary POP22 dips, because if arm powers are greater than 1, power recylcing is happening, so we should keep the loops engaged. However, since TRX and TRY are going negative when we buzz back and forth through the resonsnace, the trigger row sums to a negative value, and the PRMI loops give up. 

Instead, we can used the fortuitously unwhitened POPDC, which can serve the same function, and does not have the tendancy to go negative. Once I enabled this, I was able to just sit there as the IFO angrily buzzed at me. 

Here are my PRMI settings

REFL33 - Rotation 140.2 Degrees, -89.794 measured diff

PRCL = 1 x REFL33 I; G = -0.03; Acquire FMs 4,5; Trigger FMs 2, 9; Limit: 15k ; Acutate 1 x PRM

MICH = 1 x REFL33 Q, G= 3.0, Acquire FMs 4,5,8; Trigger FM 2, 3; Limit: 30k; Actuate -0.2625 x PRM + 0.5 x BS

Triggers = 1 x POP22 I + 0.1 * POPDC, 50 up 5 down


Just for kicks, here's a video of the buzzing as experienced in the control room

  2126   Tue Oct 20 16:35:24 2009 robConfigurationLSC33MHz Mod depth

The 33MHz mod depth is now controlled by the OMC (C1:OMC-SPARE_DAC_CH_15).  The setting to give us the same modulation depth as before is 14000 (in the offset field).

  4151   Thu Jan 13 16:34:02 2011 josephbUpdateComputers32 bit matlab updated

There was a problem with running the webview report generator in matlab on Mafalada.  It complained of not having a spare report generator license to use, even though the report generator was working before and after on other machines such as Rosalba.  So I moved the old 32 bit matlab directory from /cvs/cds/caltech/apps/Linux/matlab to /cvs/cds/caltech/apps/Linux/matlab_old.  I installed the latest R2010b matlab from IMSS in /cvs/cds/caltech/apps/Linux/matlab and this seems to have made the cron job work on Mafalda now.

  9115   Fri Sep 6 09:27:10 2013 SteveUpdateVAC31 days after pumpdown

Quote:

 Valve configuration: Vacuum Normal

 

 

  1449   Wed Apr 1 15:47:48 2009 YoichiUpdateLocking3.8kHz peak looks like a real optical response of the interferometer
Yoichi, Peter

To see where the 3.8kHz peak comes from, we locked the interferometer with the CARM fed back only to ETM and increased the arm power to 4.
The CARM error signal was taken from the transmission DC (not PO_DC).
The attached plots show the CARM transfer functions taken in this state (called ETM lock in the legends) compared with the ones taken when the CARM is locked by the feedback to the laser frequency (called "Frequency lock").
The first attachment is the TFs from the CARM excitation (i.e. the ETMs were actuated) to the TR_DC and PO_DC signals.

The second attachment is the AO path loop TFs. This is basically the TF from the frequency actuator to the PO_DC error signal.
I injected a signal into the B-excitation channel of the common mode board (with SR785) and measured the TF from TP2B to TP2A of the board.
For the ETM lock case, the AO loop was not closed because I disabled the switch between TP2A and TP1B.

The observation here is that even with no feedback to the laser frequency, the 3.8kHz peak is still present.
This strongly suggests that the peak is a real optical response of the interferometer.

To realize the ETM lock with arm_power=4, I had to tweak the CM loop shape.
I wrote a script to do this (/cvs/cds/caltech/scripts/CM/ETM_CARM_PowerUp).
You can run this script after drstep_bang has finished.
  1450   Wed Apr 1 16:14:36 2009 YoichiUpdateLocking3.8kHz peak does not change with SRC offset
Yoichi, Peter

We suspected that maybe the 3.8kHz peak is the DARM RSE somehow coupled to the CARM.
So we added an offset to the SRC error signal to see if the peak moves by changing the offset.
It didn't (at least by changing the SRC offset by +/-1000).
(I had a nice plot showing this, but dtt corrupted the data when I saved it. So no plot attached.)

I also played with the PRC, DARM offsets which did not have any effect on the peak.
The only thing, I could find so far, having some effect on the peak is the arm power. As the arm power is increased, the peak height goes up and the frequency shifts slightly towards lower frequencies.
  1433   Thu Mar 26 04:27:26 2009 YoichiUpdateLocking3.8kHz peak as a function of the arm power
During the power ramp-up, I actuated CARM using ETMs and measured the transfer functions to the PO_DC at several arm powers.
The peak grows rapidly with the power. It also seems like the frequency shifts slightly as the power goes up, but not much.

Some sort of an RSE peak ? An offset in the PRC lock point ?
  580   Thu Jun 26 22:08:33 2008 JenneUpdateElectronics3.7MHz bandstop filter in MC Servo
The 3.7MHz elliptical bandstop filter that I made during my SURF summer is now installed in the MC servo loop to reduce the noise at 3.7MHz.

I have taken transfer functions with and without the filter between TP1A and TP2A, with EXCA at -20dBm, using the HP4195A Network Analyzer. I have also taken power spectra of TP1A with and without the filter, and time domain data with the filter of OUT2 on the MC Servo Board and Qmon on the Demod board just before the MC servo board. The filter is between Qmon and OUT2 in the loop.

The UGF and phase margin don't change noticeably with and without the filter, and the MC still locks nicely (after the minor fiasco this afternoon), so I think it's okay. The UGF is around 57kHz, with about 38 degrees phase margin.

1 July 2008: I redid the plots. Same info, but the traces with and without the filters are now on the same graph for easier readability.
  8526   Fri May 3 08:55:55 2013 SteveUpdatePEM3.2 M earthquake
  15929   Wed Mar 17 10:52:48 2021 JordanUpdateSUS3" Ring Adpater for SOS

I have added a .1", 45deg chamfer to the bottom of the adapter ring. This was added for a new placement of the eq stops, since the barrel screws are hard to access/adjust.

This also required a modification to the eq stop bracket, D960008-v2, with 1/4-20 screws angled at 45 deg to line up with the chamfer.

The issue I am running into is there needs to be a screw on the backside of the ring as well, otherwise the ring would fall backwards into the OSEMs in the event of an earthquake. The only two points of contact are these front two angled screws, a third is needed on the opposite side of the CoM for stability. This would require another bracket mounted at the back of the SOS tower, but there is very little open real estate because of the OSEMs.

 

Instead of this whole chamfer route, is it possible/easier to just swap the screws for the barrel eq stops? Instead of a socket head cap screw, a SS thumb screw such as this, will provide more torque when turning, and remove the need to use a hex wrench to turn.

 

  9549   Mon Jan 13 11:08:48 2014 SteveUpdatePSL3 good days of IOO pointing

 Three good days of IOO pointing: Friday, Sat and Sun    What was changed?  May be the clamping on Friday?

IOO vertical changes recovering as tempeture. IP is clipping at plastic enclosure of ETMY

 

NOTE: ANTS at the PSL optical table.  I will mop with chemicals tomorrow if we see more.

 

  17111   Mon Aug 29 15:15:46 2022 TegaUpdateComputers3 FEs from LLO got delivered today

[JC, Tega]

We got the 3 front-ends from LLO today. The contents of each box are:

  1. FE machine
  2. OSS adapter card for connecting to I/O chassis
  3. PCI riser cards (x2)
  4. Timing Card and cable
  5. Power cables, mounting brackets and accompanying screws
  17113   Tue Aug 30 15:21:27 2022 TegaUpdateComputers3 FEs from LHO got delivered today

[Tega, JC]

We received the remaining 3 front-ends from LHO today. They each have a timing card and an OSS host adapter card installed. We also receive 3 dolphin DX cards. As with the previous packages from LLO, each box contains a rack mounting kit for the supermicro machine.

  5970   Mon Nov 21 16:08:04 2011 kiwamuUpdateGreen Locking2nd trial of Y arm ALS noise budget : broad band noise gone

Quote from #5930

Right now the fluctuation of the green beat-note seems mostly covered by unknown noise which is relatively white.

The 2nd trial of the Y arm ALS noise budgeting :

(Removal of broad band noise)

  + The broad band noise decreased somewhat after I fixed a broken connection in the discriminator.
  + I took a look at the frequency discriminator setup and found one of the SMA-BNC adapter was broken.
     This adapter was attached to one of the outputs of the 4-way power splitter, which splits the signal into the coarse and find discriminator paths.
     And this broken adapter was in the coarse path, which actually I am not using for the noise budget.
     Depending on the stress acting on the adapter it was creating broadband noise, even in the fine path.
     So I threw it away and put another SMA-BNC adapter.
 
Here is a plot of the latest noise : high frequency noise is still unknown.

Yarm_ALS_2011Nov19.png

I will add the dark noise of the broad-band beat-note PD  and the MFD read out noise on the budget.

ELOG V3.1.3-