2418
|
Tue Dec 15 05:29:31 2009 |
Alberto | Update | General | Arm Cavity Poles measured again after cleaning the optics last week |
The Y arm cavity pole moved down by 130 Hz, whereas the X arm moved by only 34 Hz. I wonder if that is because Kiwamu spent much more time on cleaning ITMY than on any other optic. |
2420
|
Tue Dec 15 21:39:34 2009 |
Alberto | Update | ABSL | brief summary of this afternoon's measurements |
I took measurements of the open loop gain of the AbsL PLL with the old Universal PDH Box.
I Also measured the filter shape of both the new and the old PDH box.
I'm going to plot the results in a nice form tomorrow morning.
For who's interested, the PLL UGF was at 10KHz.
I can't lock the PLL with the new PDH box. Measuring its filter's shape, as suggested by Koji, I found out that it differs from the old one. That despite the fact that the two boxes should share the same circuit schematic. O,r at least, that is what it looks like from the schematics in the DCC.
I need to understand whether that is intentional and, if that was the case, what kind of use Rich Abbott designed it for.
Tomorrow I'm going to post in the elog the filter's transfer functions too.
Before leaving the lab I closed the auxiliary laser's mechanical shutter. |
2421
|
Wed Dec 16 11:21:20 2009 |
Alberto | Update | ABSL | Universal PDH Box Servo Filters |
Yesterday I measured the shape of the servo filter of both the old and the new Universal PDH boxes.
Here they are compared.

The way the filter's transfer function has been measured is by a swept sine between the "SERVO INPUT" and the "PIEZO DRIVE OUTPUT" connection on the box front panel. The spectrum analyzer used for the measurement is the SR785 and the source amplitude is set at 0.1V.
The two transfer functions are clearly different. In particular the old one looks like a simple integrator, whereas the new one already includes some sort of boost.
That probably explains why the new one is unable to lock the PLL. Indeed what the PLL needs, at least to acquire lock, is an 1/f filter.
I thought the two boxes were almost identical, at least in the filter shapes. Also the two schematics available in the DCC coincide. |
Attachment 1: NewandOlfFilterTF.png
|
|
2422
|
Wed Dec 16 11:46:25 2009 |
Alberto | Update | ABSL | Absl PLL Open Loop Gain |
Yesterday I measured the Open Loop Gain of the PLL in the absolute length experiment. The servo I used was that of the old Universal PDH box.
The OLG looks like this:

The UGF is at 10 KHz. |
2427
|
Thu Dec 17 09:30:08 2009 |
Alberto | HowTo | Computers | Nodus sluggish |
The elog has been quite slow for the last two days. The cause is nodus, that has been slowing down the access to it.
I looked at the list of the running processes on nodus by typing the command prstat, which is the equivalent for Solaris of the Linux "top". I didn't see any particular process that might be absorbing too many resources.
I remember Rana fixing the same problem in the past but couldn't find his elog entry about that. Maybe we should just restart nodus, unless someone has some other suggestion. |
2428
|
Thu Dec 17 17:13:50 2009 |
Alberto | Omnistructure | Environment | STACIS stuff |
One of the electronics benches is currently occupied by the STACIS equipment.
We need that table If no one is working on the STACIS anymore, it should be removed from there. |
2429
|
Thu Dec 17 19:03:14 2009 |
Alberto | HowTo | Computers | Nodus sluggish |
Quote: |
The elog has been quite slow for the last two days. The cause is nodus, that has been slowing down the access to it.
I looked at the list of the running processes on nodus by typing the command prstat, which is the equivalent for Solaris of the Linux "top". I didn't see any particular process that might be absorbing too many resources.
I remember Rana fixing the same problem in the past but couldn't find his elog entry about that. Maybe we should just restart nodus, unless someone has some other suggestion.
|
Problem solved. Nodus and the elog are running fine. It's just that the elog takes some time to make a preview of complex pdf attachments, like those in Kiwamu's entry 2405. |
2459
|
Mon Dec 28 15:19:19 2009 |
Alberto | Update | Computers | Burtrestored to Dec 26 at 20:00 |
Since it wasn't sure whether all the front-ends had been restored after the bootfest, I burtrestored everything to Dec 26 at 20:00.
Always keep in mind that to burtrestore c1dcuepics, the snapshot file has to be modified by hand by moving the last quote up to the line before the last. |
2460
|
Mon Dec 28 15:34:14 2009 |
Alberto | Update | ABSL | Working on the AP table |
I opened the auxiliary laser's shutter.
I'm currently working on the AP table. |
2461
|
Mon Dec 28 18:35:27 2009 |
Alberto | Update | ABSL | Working on the AP table |
Quote: |
I opened the auxiliary laser's shutter.
I'm currently working on the AP table.
|
I finished working on the table.
I closed the AUX NPRO's shutter. |
2467
|
Wed Dec 30 10:58:48 2009 |
Alberto | Update | General | All watchdogs tripped this morning |
This morning I found all the watchdogs had tripped during the night.
I restored them all.
I can't damp ITMX. I noticed that its driving matrix is all 1s and -1s as the the right values had been lost in some previous burtrestoring. |
2476
|
Tue Jan 5 09:18:38 2010 |
Alberto | Omnistructure | Electronics | Universal PDH Box Stored in the RF Cabinet |
FYI: I stored the Universal PDH boxes in the RF cabiner in the Y arm. |
2479
|
Tue Jan 5 13:23:45 2010 |
Alberto | Omnistructure | Environment | turning page |
In the lab there are lots of old posters with outdated autocad drawings, or printouts with schematics of old electronics hanging on the walls.
Can we get rid of those and start giving the lab a fresh and modern look? |
2485
|
Fri Jan 8 10:03:04 2010 |
Alberto | Omnistructure | LSC | SPOB shutter was closed |
This morning I found that there was no light on the SPOB PD. I went looking at the photodetector and I found that the shutter in front of it was closed.
I switched the shutter driver from n.c. to n.o. which had the effect of opening it.
I guess we inadvertently closed the shutter with Rana when last week we were tinkering with the ITMY camera. |
2491
|
Sat Jan 9 09:47:03 2010 |
Alberto | Update | LSC | Problems trying to lock the arms |
This morning I've been having problems in trying to lock the X arm.
The X arm's filter FM6 in the LSC screen starts blinking as it was halfloaded. Then the transmitted power drops from 1 to ~0.5 and eventually the arm loses lock.
To me it looked like a computer related issue. So I decide to reboot C1ISCEX by powercycling it.
That doesn't seem to have solved the problem. The X arm can get locked but TRX slowly moves between 0.2 and 1. |
2492
|
Sat Jan 9 11:07:30 2010 |
Alberto | Update | LSC | Problems trying to lock the arms |
Quote: |
This morning I've been having problems in trying to lock the X arm.
The X arm's filter FM6 in the LSC screen starts blinking as it was halfloaded. Then the transmitted power drops from 1 to ~0.5 and eventually the arm loses lock.
To me it looked like a computer related issue. So I decide to reboot C1ISCEX by powercycling it.
That doesn't seem to have solved the problem. The X arm can get locked but TRX slowly moves between 0.2 and 1.
|
The X arm is now locked with TRX stable at ~1.
I think earlier on today I was having problems with running the alignment scripts from op540. Now I'm controlling the IFO from Rosalba and I can easily and stably lock all degrees of freedom.
I needed the X arm to be locked to align the auxiliary beam of the AbsL experiment to the IFO. To further stabilize TRX I increased the loop gain from 1 to 1.5.
Now the auxiliary beam is well aligned to the IFO and the beat is going through the PRC. I'm finally ready to scan the recycling cavity.
I also changed the gain of the PRC loop from -0.1 to -0.5. |
2493
|
Sat Jan 9 15:02:01 2010 |
Alberto | Update | ABSL | PRC scanning |
I scanned the PRC in the frequency range of 30-60 MHz, untill the PLL lost lock. But everything is working fine.
The PRC remained lock for all time, with SPOB at ~1000.
I'm leaving the lab now, planning to come back tomorrow.
I turned the flipping mirrors down and closed the mechanical shutter of the auxiliary NPRO. |
2496
|
Sun Jan 10 16:05:51 2010 |
Alberto | Omnistructure | Environment | Lab Thermostats Temperature Lowered by 1 deg F |
Rana noticed that recently the temperature inside the lab has been a little bit too high. That might be causing some 'unease' to the computers with the result of making them crash more often.
Today I lowered the temperature of the three thermostats that we have inside the lab by one degree:
Y arm thermostat: from 71 to 70 F
X arm thermostat: from 70 to 69 F
Aisle thermostat: from 72 to 71 F.
For the next hours I'll be paying attention to the temperature inside the lab to make sure that it doesn't go out of control and that the environment gets too cold. |
2500
|
Mon Jan 11 09:18:44 2010 |
Alberto | Update | General | Measurement running |
I'm working on the AbsL experiment. A measurement which involved the PRC locked is running at the moment.
Please make sure of not interfering with the interferometer until it is done. Thank you. |
2501
|
Mon Jan 11 10:01:06 2010 |
Alberto | Omnistructure | Environment | Lab Thermostats Temperature Lowered by 1 deg F |
Quote: |
Rana noticed that recently the temperature inside the lab has been a little bit too high. That might be causing some 'unease' to the computers with the result of making them crash more often.
Today I lowered the temperature of the three thermostats that we have inside the lab by one degree:
Y arm thermostat: from 71 to 70 F
X arm thermostat: from 70 to 69 F
Aisle thermostat: from 72 to 71 F.
For the next hours I'll be paying attention to the temperature inside the lab to make sure that it doesn't go out of control and that the environment gets too cold.
|
Today the lab is perceptibly cooler.
The temperature around the corner is 73 F. |
2502
|
Mon Jan 11 11:06:53 2010 |
Alberto | Update | ABSL | Measurement running |
Quote: |
I'm working on the AbsL experiment. A measurement which involved the PRC locked is running at the moment.
Please make sure of not interfering with the interferometer until it is done. Thank you.
|
I'm done for the moement.
I realized that I need to take into account somehow the DC power from the photodiode. By now the measurement of the transmitted beat's power is affected by the total power circulating inside of the PRC and thus it depends on the cavity alignment.
I closed the laser shutter and turned down the flipping mirrors.
I'm planning to restart measuring by 2.30pm today. Till then the interferometer is available. |
2504
|
Mon Jan 11 16:59:14 2010 |
Alberto | Update | ABSL | Interferometer Busy |
I'm currently running a measurement on the PRC.
Please don't interfere with the IFO until it is done. Talk with Alberto if you've been intending to work inside the lab.
Thank you. |
2505
|
Mon Jan 11 19:36:13 2010 |
Alberto | Update | ABSL | Measurement running |
Leaving for dinner. Back in ~1hr.
I left a measurement running. Please don't interfere with it till I'm back. Thanks. |
2508
|
Tue Jan 12 09:37:05 2010 |
Alberto | Update | ABSL | IFO available |
I finished measuring the AbsL for this morning. The IFO is again available.
Please don't mess up with the interferometer though. I'll be back in a couple of ours |
2512
|
Wed Jan 13 12:01:06 2010 |
Alberto | Update | Computers | c1dcuepics, c1lsc rebooted this morning |
Since last night the alignemtn scripts couldn't work.
c1lsc wasn't working properly because attempts to lock the X arm would try to control ETMY and attempts of locking the Y arm, wouldn't actuate any optics.
Also, another sign of a malfunctioning c1lsc was that one of the LSC filter modules, FM6, couldn't get loaded properly. It looked like only half loaded on the LSC MEDM screen.
On the other hand, plotting the trend of the last month, c1lsc's CPU didn't look more loaded that usual.
Rebooting and restarting C1lsc wasn't enough and I also had to reboot c1dcuepics a couple of times beforse getting things back to work. |
2513
|
Wed Jan 13 12:03:00 2010 |
Alberto | Update | ABSL | Measurement now running. Please be careful |
At the moment I'm running a measurement on the PRC and I'm planning to leave it going for the time we'll be at the 40m meeting.
Please be careful in the lab. Thank you. |
2519
|
Sun Jan 17 19:18:55 2010 |
Alberto | Update | Computers | Boot fest to restart the computer and c1iscey not responding. |
Thi afternoon I found that the RFM network in trouble. The frontends sync counters had railed to 16384 counts and some of the computers were not responding. I went for a bootfest, but before I rebooted c1dcu epics. I did it twice. Eventually it worked and I could get the frontends back to green.
Although trying to burtrestore to snapshots taken at any time after last wednesday till today would make the RFM crash again. Weird.
Also, c1iscey seems in a coma and doesn't want to come back. Power cycling it didn't work. I don't know how to be more persuasive with it. |
2521
|
Mon Jan 18 18:34:01 2010 |
Alberto | Update | ABSL | Measurement in progress |
I started a long measurement of the PRC's transmissivity. I'm leaving the lab and I'm going to be back at about 8 tonight. Please do not disturb the interferometer. it is important that the MC and the PRC stay locked all the time. |
2522
|
Mon Jan 18 20:58:40 2010 |
Alberto | Update | ABSL | Measurement in progress |
Quote: |
I started a long measurement of the PRC's transmissivity. I'm leaving the lab and I'm going to be back at about 8 tonight. Please do not disturb the interferometer. it is important that the MC and the PRC stay locked all the time.
|
That measurement is finished. I'm now going to start another one that will take another hour or so. I'm leaving it running for the night. If you want to work on the IFO, it should be definitely done by 11pm. |
2531
|
Tue Jan 19 12:54:39 2010 |
Alberto | Update | ABSL | Measurement in progress |
A measurement will be running for the next hour. Please be careful. |
2532
|
Tue Jan 19 16:21:18 2010 |
Alberto | Update | ABSL | Watchdogs not working and then fixed |
This afternoon the watchdogs stopped working: they didn't trip when the suspension positions crossed the threshold values.
I rebooted c1susaux (aka c1dscl1epics0 in the 1Y5 rack), which is the computer that runs the watchdog processes.
The reboot fixed the problem. |
2534
|
Wed Jan 20 20:11:56 2010 |
Alberto | Update | ABSL | Some preliminary results from measuring PRC's transmissivity for an amplitude modulated beam |
Here I'm posting a plot showing a set of measurements that I made in the past few days to determine the absolute length of the PRC cavity.
As in my other AbsL measurements, I inject an auxiliary laser beam into the cavity and look at the transmission. In the PRC case, the beam is injected through the dark port and I look at a pick-off of the REFL beam. The aux laser is phase locked to the PSL beam and I control the differential frequency between the two. The PSL and the aux beam interfere and beat at their differential frequency.
The attached plot shows the transmitted power as a function of the beat frequency.
Fitting the data with the model would let me determine the cavity length.
By now I can estimate the length of the PRC at about 2.257m, but it's still a rather approximate value.
I can't provide accurate error bars yet. I need to optimize the measurements to get a more precise value.
I will go more through the details of the measurement technique and of the fitting function as soon as I have more definitive results.
The data points shown here were taken at different times and not always in optimal alignment condition of the PRC.
To get a good fit of the data I should have fewer frequency segments, taken in a shorter period of time, and in which the power circulating inside of the cavity (ie SPOB) fluctuates as little as possible.
For what regards the time needed for a measurements, I already significantly sped up the measurements (i.e. optimizing the scanning and acquisition GPIB scripts, and fixing a couple of problems with the PDH box used in the PLL), and finally now I can scan several tens of MHz in few minutes.
About the frequency segments, so far they have been determined by two factors
1) Tthe frequency generator in the PLL: the Marconi works as a continuous wave generator only in limited ranges. Switching from one to another brakes the wave in a way that causes the PLL to lose lock.
2) Getting below 18 MHz a series of other beats appear on the PLL photodiode and make the PLL lose lock.
For the first problem, I'm thinking of using two Marconis and to mix their signals. I would keep one at 300MHz and I would scan the other from 300MHz to 500MHz. In fat, in that frequency range the Marconi has not discontinuity.
To try to avoid the other beats at low frequency, I'm not entirely sure about what to do yet.
To be continued... |
Attachment 1: 2010-01-19_PRCtransmissivityVsModel.png
|
|
2541
|
Fri Jan 22 02:54:06 2010 |
Alberto | Update | ABSL | Overnight measurement |
I'm leaving a measurement running overnight. It should be done in about one hour.
Tomorrow morning, If you need to use the interferometer, and you don't want to have the auxiliary beam going onto the dark port, you can turn down the flipping mirror and close the NPRO's mechanical shutter. |
2543
|
Fri Jan 22 14:40:49 2010 |
Alberto | Update | ABSL | Overnight measurement |
Quote: |
I'm leaving a measurement running overnight. It should be done in about one hour.
Tomorrow morning, If you need to use the interferometer, and you don't want to have the auxiliary beam going onto the dark port, you can turn down the flipping mirror and close the NPRO's mechanical shutter.
|
This is what I measured last night:

This is not a fit. It's just a comparison of the model with the data. |
2545
|
Mon Jan 25 16:30:37 2010 |
Alberto | Update | ABSL | 166 MHz sideband turned off |
I turned off the modulation at 166MHZ becasue I don't need it if I'm only locking the PRC.
It was introducing extra amplitude modulations of the beam which interfered with the AbsL's PLL photodiode.
I'm going to turn it back on later on. |
2546
|
Mon Jan 25 16:46:33 2010 |
Alberto | Update | ABSL | 166 MHz sideband turned off |
Quote: |
I turned off the modulation at 166MHZ becasue I don't need it if I'm only locking the PRC.
It was introducing extra amplitude modulations of the beam which interfered with the AbsL's PLL photodiode.
I'm going to turn it back on later on.
|
I turned back on the 166MHz modulation just a bit. I set the slider at 4.156.
When it was totally off the MZ seemd quite unhappy. |
2550
|
Wed Jan 27 11:02:30 2010 |
Alberto | Update | ABSL | PRC Cavity Length |
I fitted the data from scanning the PRC by changing the beat frequency of the auxiliary laser beam with the PSL beam.
The data points that I've taken so far over the entire frequency range (0-300 MHz) are not continuous. For several reasons the PLL was unable to maintain lock for such a large range and I had to break it into smaller segments. The measurements to acquire them stretched over a too long period of time during which the status of the PRC changed.
Because of that, before I get a continuous set of data points (perhaps normalized by the circulating power inside of the cavity), I restricted the fit to a 55MHz range around 100MHz. I obtained the following numbers for the fit parameters:
Length PRC = 2.169 +/- 0.007 m
Schnupp Asymmetry: 0.471+/- 0.006 m
The fit is shown in the attached plot:
When I fit over the entire set of data I get this:
Length PRC = 2.224 +/- 0.005 m
Schnupp Asymmetry: 0.457+/- 0.005 m
The results are different. Evidently I have to improve the measurement. I'm working on it.
For posterity:
The function I used to fit the transmitted beat power vs. frequency is the following:
E_trans = - t_prm * r_itm * exp(1i*2*wb*l_prc/c) .* sin(wb*l_/c) ./ ( 1 + r_prm * r_itm * exp(1i*2*wb*l_prc/c) .* cos(wb*l_/c)
Where wb is the angular frequency of the beat, l_prc and l_ are the length of the PRC and the Schnupp asymmetry, respectively; r_itm, t_itm, r_prm, t_prm are reflectances and transmittances of PRM and ITM; c is the speed of light.
|
2551
|
Thu Jan 28 09:14:51 2010 |
Alberto | Configuration | Computers | c1iscey, c1iscex, c1lsc, c1asc rebooted |
This morning the LSC scripts wheren't running properly. I had to reboot c1iscey, c1iscex, c1lsc, c1asc .
I burtrestored to Monday January 25 at 12:00. |
2552
|
Thu Jan 28 09:17:32 2010 |
Alberto | Update | LSC | 166 Modulation turned off |
I temporarily turned off the 166 modulation. |
2553
|
Fri Jan 29 12:06:58 2010 |
Alberto | Update | ABSL | Measurement running today at lunch time |
I just started a measuremtn that will be running for the next hour or so. Please be careful with the interferometer. |
2554
|
Fri Jan 29 13:14:49 2010 |
Alberto | Update | ABSL | Measurement running today at lunch time |
Quote: |
I just started a measuremtn that will be running for the next hour or so. Please be careful with the interferometer.
|
Done. IFO available |
2560
|
Tue Feb 2 15:30:03 2010 |
Alberto | Frogs | Treasure | Wild Oats |
FYI. Sitting on the top shelf of George I found an opened jar of raspberry jam and an opened jar of creamy peanut butter. Both are branded Wild Oats Market.
Wikipedia:
"Wild Oats Markets was an operator of natural foods stores and farmers markets in North America... Whole Foods officially completed their buyout of Wild Oats on August 27, 2007 [...]" |
2562
|
Tue Feb 2 18:15:47 2010 |
Alberto | Update | elog | Elog restarted it |
Zach made me notice that the elog had crashed earlier on this afternoon.
I just restarted it with the restarting script.
Instructions on how to run the last one are now in the wiki page. Look on the "How To" section, under "How to restart the elog". |
2567
|
Wed Feb 3 10:46:12 2010 |
Alberto | Update | elog | elog restarted |
Again, this morning Zach told me that the elog had crashed while he was trying to post an entry.
I just restarted it. |
2576
|
Mon Feb 8 14:13:03 2010 |
Alberto | Update | ABSL | PLL Characterization |
Lately I've been trying to improve the PLL for the AbsL experiment so that it could handle larger frequency steps and thus speed up the cavity scan.
The maximum frequency step that the PLL could handle withouth losing lock is given by the DC gain of the PLL. This is the product of the mixer's gain factor K [rad/V ], of the laser's calibration C [Hz/V] and of the PLL filter DC gain F(0).
I measured these quantities: K=0.226 V/rad; C=8.3e6 Hz/V and F(0)=28.7dB=21.5. The max frequency step should be Delta_f_max = 6.4MHz.
Although in reality the PLL can't handle more than a 10 KHz step. There's probably some other effect that I'm not.
I'm attaching here plots of the PLL Open Loop Gain, of the PLL filter and of a spectra of the error point measured in different circumstances.
I don't have much time to explain here how I took all those measurements. After I fix the problem, I'm going to go go through those details in an elog entry.
Does anyone have any suggestion about what, in principle, might be limiting the frequency step?
I already made sure that both cables going to the mixer (the cable with the beat signal coming from the photodiode and the cable with the LO signal coming from the Marconi) had the same length. Although ideally, for phase locking, I would still need 90 degrees of phase shift between the mixing signals, over the entire frequency range for which I do the cavity scan. By now the 90 degrees are not guaranteed.
Also, I have a boost that adds another 20 dB at DC to the PLL's filter. Although it doesn't change anything. In fact, as said above calculating the frequency step, the PLL should be able to handle 100KHz steps, as I would want the PLL to do. |
Attachment 1: 2010-02-08_Old_PDH_Box_Filter_TF_gain_knob_0_Boost_OFF.png
|
|
Attachment 2: 2010-02-08_PLL_OLG_gain_knob_0_Boost_OFF.png
|
|
Attachment 3: 2010-02-08_PLL_Noise_Budget.png
|
|
2577
|
Mon Feb 8 14:56:17 2010 |
Alberto | Update | ABSL | Suddenly a much better alignment of PRC |
I just aligned PRM and locked PRC and I noticed that SPOB is much higehr than it used to be. It's now about 1800, vs 1200 than it used to be last week.
Isn't anyone related to that? If so, may I please know how he/she did it? |
2579
|
Mon Feb 8 15:41:51 2010 |
Alberto | Update | ABSL | Suddenly a much better alignment of PRC |
Quote: |
Quote: |
I just aligned PRM and locked PRC and I noticed that SPOB is much higehr than it used to be. It's now about 1800, vs 1200 than it used to be last week.
Isn't anyone related to that? If so, may I please know how he/she did it?
|
oops, my bad. I cranked the 33MHz modulation depth and forgot to put it back. The slider should go back to around 3.
|
I was actually hoping that the alignment got better. |
2581
|
Tue Feb 9 09:07:06 2010 |
Alberto | Update | ABSL | PLL Characterization |
Quote: |
Lately I've been trying to improve the PLL for the AbsL experiment so that it could handle larger frequency steps and thus speed up the cavity scan.
The maximum frequency step that the PLL could handle withouth losing lock is given by the DC gain of the PLL. This is the product of the mixer's gain factor K [rad/V ], of the laser's calibration C [Hz/V] and of the PLL filter DC gain F(0).
I measured these quantities: K=0.226 V/rad; C=8.3e6 Hz/V and F(0)=28.7dB=21.5. The max frequency step should be Delta_f_max = 6.4MHz.
Although in reality the PLL can't handle more than a 10 KHz step. There's probably some other effect that I'm not.
I'm attaching here plots of the PLL Open Loop Gain, of the PLL filter and of a spectra of the error point measured in different circumstances.
I don't have much time to explain here how I took all those measurements. After I fix the problem, I'm going to go go through those details in an elog entry.
Does anyone have any suggestion about what, in principle, might be limiting the frequency step?
I already made sure that both cables going to the mixer (the cable with the beat signal coming from the photodiode and the cable with the LO signal coming from the Marconi) had the same length. Although ideally, for phase locking, I would still need 90 degrees of phase shift between the mixing signals, over the entire frequency range for which I do the cavity scan. By now the 90 degrees are not guaranteed.
Also, I have a boost that adds another 20 dB at DC to the PLL's filter. Although it doesn't change anything. In fact, as said above calculating the frequency step, the PLL should be able to handle 100KHz steps, as I would want the PLL to do.
|
I might have found the problem with the PLL that was preventing me from scanning the frequencies by 100KHz steps. A dumb flimsy soldering in the circuit was making the PLL unstable.
After I fixed that problem and also after writing a cleverer data acquisition script in Python, I was able to scan continuosly the range 10-200MHz in about 20min (versus the almost 1.5-2 hrs that I could do previously). I'm attaching the results to this entry.
The 'smears' on the right side of the resonance at ~33MHz, are due to the PSL's sideband. I think I know how to fix that.
As you can see, the problem is that the model for the cavity transmission still does not match very well the data. As a result, the error on the cavity length is too big (~> 10 cm - I'd like to have 1mm).
Anyway, that was only my first attempt of scanning. I'm going to repeat the measurement today too and see if I can come out better. If not, than I have to rethink the model I've been using to fit. |
Attachment 1: 2010-02-08_PRCtransmissivity_EntireFreqRange_VsFit.png
|
|
2582
|
Tue Feb 9 10:10:58 2010 |
Alberto | Update | ABSL | back to analog |
I want to try to do the measurement with the network analyzer used as local oscillator, instead of the Marconis that I'm using now. Tha could give me better noise rejection. It would also give me information about the phase.
Also I wouldn't dislike abandoning the GPIB interfaces to acquire data. |
2604
|
Tue Feb 16 09:51:22 2010 |
Alberto | Update | Green Locking | take some optics away from the ETM end table |
Quote: |
In the last two days Steve and I took some optics away from the both ETM end table.
This is because we need an enough space to set up the green locking stuff into the end table, and also need to know how much space is available.
Optics we took away are : Alberto's RF stuff, fiber stuff and some optics obviously not in used.
The picture taken after the removing is attached. Attachment1:ETMX, Attachment2:ETMY
And the pictures taken before the removing are on the wiki, so you can check how they are changed.
http://lhocds.ligo-wa.caltech.edu:8000/40m/Optical_Tables
|
The PD Kiwamu removed from the Y table was TRY, which we still need.
My bad if he took that. By mistake I told him that was the one I installed on the table for the length measurement and we didn't need it anymore.
I'm going to ask Kiwamu if he can kindly put it back. |