ID |
Date |
Author |
Type |
Category |
Subject |
12768
|
Sat Jan 28 01:25:51 2017 |
gautam | Update | IMC | 29.5 MHz modulation depth |
Some more details of our investigation:
- Here is a spectrum of the signal to the power combiner on the PSL table, measured on the output of the RF AM Stabilization box.

Perhaps these sidebands were the ones I observed while looking at the input to the WFS demod board.
- The signal looked like a clean sinusoid when viewed on an oscilloscope with input impedance set to 50ohms. There were no sharp features or glitches in the time we observed, except when the 29.5 MHz MEDM slider was increased beyond 5, as noted by Lydia.
We couldn't find a schematic for this RF AM Stabilization servo, so we are not sure what RF output power to the EOM we should expect. Schematic has since been found.
- I measured the power level at the input side (i.e. from the crystal) and found that it is ~12dBm, which seems reasonable (the front panel of the box housing the 29.5 MHz oscillator is labelled 13dBm). The schematic for the RF AM stabilization box says we should expect +10dBm at the input side, so all this points to a problem in the RF AM stabilization circuit...
- There is an attenuator dial on the front panel of the said RF AM stabilization servo that allows one to tune the power to the LO input of the WFS. Right now, it is set to approximately 7dB of attentuation, which corresponds to -12dBm at the WFS demod board input. I did a quick check to see if turning the dial changed the signal level at the LO input of the WFS board. The dial moves in clicks of 1dB, and the RF power at the LO input of the demod board increased/decreased by ~1dBm for each click the dial was rotated (I only explored the region 3dB-11dB of atttentuation). So it should be possible to increase the LO level to the WFS demod boards, is there any reason we shouldn't increase this to -8bBm (~0.25Vpp into 50ohms, which is around the level Koji verified the mixer to be working well at)?
- There were a couple of short ribbon cables which were just lying around on top of the cards in the eurocrate, Koji tells me that these were used as tester cables for checking the whitening filters and that they don't serve any purpose now. These have been removed.
- Added a button to IMC MEDM screen to allow easy access to the MEDM screen with slider to control the 29.5MHz modulation depth - though as mentioned in Lydia's elog, at the moment, this slider has no effect on the 29.5MHz power level to the EOM...

|
2801
|
Thu Apr 15 14:47:28 2010 |
steve | Update | Electronics | 25MHZ oscillation of HP4195A |
The 1979 vintage RF spectrum analyzer HP4195A sn2904J01587 shipped out for repair today to http://www.avalontest.com
It has a 25 MHZ oscillation when you go below 150 MHZ in your sweep....atm1 with the larger amplitude shows this 25 MHZ
Atm2 is displaying full sweep-sign scans from 1 to 500 MHZ.....here one can clearly see the three segment of the scan:
1, large amplitude 25 MHZ oscillation dominating the spectrum up to 150 MHZ
2, the mid section from 150 MHZ to 300 MHZ with medium size amplitude is normal
3, from 300 MHZ to 500 MHZ the amplitude is decreasing.......showing the disadvantage of using a 300 MHZ oscilloscope
|
16211
|
Thu Jun 17 22:19:12 2021 |
Koji | Update | Electronics | 25 HAM-A coil driver units delivered |
25 HAM-A coil driver units were fabricated by Todd and I've transported them to the 40m.
2 units we already have received earlier.
The last (1) unit has been completed, but Luis wants to use it for some A+ testing. So 1 more unit is coming. |
12660
|
Fri Dec 2 16:40:29 2016 |
gautam | Update | IMC | 24V fuse pulled out |
I've pulled out the 24V fuse block which supplies power to the AOM RF driver. The way things are set up on the PSL table, this same voltage source powers the RF amplifiers which amplify the green beatnote signals before sending them to the LSC rack. So I turned off the green beat PDs before pulling out the fuse. I then disconnected the input to the RF driver (it was plugged into a DS345 function generator on the PSL table) and terminated it with a 50 ohm terminator. I want to figure out a smart way of triggering the AOM drive and recording a ringdown on the scope, after which I will re-connect the RF driver to the DS345. The RF driver, as well as the green beat amplifiers and green beat PDs, remain unpowered for now... |
623
|
Wed Jul 2 13:56:10 2008 |
Rob, Yoichi, John | Update | Locking | 24.5 Hz resonance |
Work continues on trying to reduce the CARM offset using dc signals from PO_DC. Got up to arm powers of
~35 last night.
We found that progress was stymied by an oscillation around 24 Hz. This oscillation was clearly visible
in the intensity of the light at REFL, PO and TrX.
Initially we suspected that this oscillation was due to an instability in the CARM loop. We attempted to
solve the problem by tuning the crossover frequncy of the AO and MC_L paths and shaping the MC_L loop to
reduce the impact of the 24 Hz noise.
After some quick tests we found that the 24 Hz signal was present even when dc CARM was used. It appears
that the peak is in fact due to a SOS mechanical resonance. We currently suspect a roll mode.
We're going to check that PRC, MICH and DARM have filters to attenuate the 24 Hz line. We'll also look at the
SUS_POS bandstop filters to see where they are centred.
The ISS was behaving strangely again. Constantly saturated at 5dB of gain. Someone needs to look a this. |
6359
|
Mon Mar 5 20:31:33 2012 |
Koji | Update | LSC | 22/110MHz path for POP |
This a kind of self record...
We need an RF setup at POP to extract 22 and 110 MHz components separately.
I am planning to work on this in the daytime on Tuesday. |
6366
|
Tue Mar 6 22:23:04 2012 |
Koji | Update | LSC | 22/110MHz path for POP |
As par Kiwamu's request, RF filters for POP22 and POP110 were installed. They are not really nice. We need to replace it with more fancy electronics.
More to come later.
|
6374
|
Wed Mar 7 15:56:36 2012 |
Koji | Update | LSC | 22/110MHz path for POP |
The RF separator installed comprises of the Minicircuits filters cascaded as in the figure below.
This has one input and 4 output ports for 11, 22, 30-60, and 110MHz signal.
As seen in this entry #6167, we have 22 and 110MHz signals together with 11, 44, 66MHz signals.
They may be demodulated via a harmonic characteristic of the mixers. (Remeber mixers are not multipliers.)

Of course the big concern is the impedance matching for those signals as usual.
The 2nd attachment shows measured impedance of the circuits with all of the ports terminated.
From the complex impedance, we can calculate the reflection coefficient. The 44 and 110MHz
components look correctly matched while the others seems largely reflected.
This certainly is not a nice situation, as the reflection can make the amplifier next to the PD unhappy
(although the reflected power is tiny in our case).
In our case more eminent problem is that the amplitude of the 22MHz signal can vary depending on the cable length by
factor of 10 in amplitude. (c.f. VSWR on the 2nd attachment.)
The transmission to each port was measured. The separation of the signals looks good. But the attenuation of the
targetted signals (i.e. insertion losses) are qulitatively consistent with the impedance. Again these losses are depend
on the cable length.
|
8192
|
Wed Feb 27 20:50:41 2013 |
Manasa | Update | LSC | 22/110MHz path for POP |
[Yuta, Manasa]
Modified POP path.
1. Removed temporary POP DC and the BS 50 (elog)
2. Introduced a 95% BS after the POP steering mirrors (95% of the signal goes to PD10CF used for POP22 and 5% goes to POP camera)
3. Output of PD10CF goes to the LSC rack through POP110 heliax cable.
4. The PD output at the LSC rack goes through a DC block to separate DC from RF.

We could not find a power supply slot for the amplifiers on the LSC rack. We had to put a temporary power supply in contradiction to our 'no temporary power supply' policy.
|
8397
|
Tue Apr 2 23:14:02 2013 |
rana | Update | LSC | 22/110MHz path for POP |
Quote: |
We could not find a power supply slot for the amplifiers on the LSC rack. We had to put a temporary power supply in contradiction to our 'no temporary power supply' policy.
|
After 1 month, its hard to imagine that this could not have been fixed by putting in a proper fuse and fuse block. I will remove this tomorrow if I still find it this way in the bottom of the rack.
There are also 2 Sorensen switching supplies in the bottom of the LSC rack (with all of our sensitive demod boards). These should also be moved over to the old 'digital' LSC rack tomorrow for the post meeting lab cleanup.
Use fuse blocks with fuses with appropriate ampacity. |
2925
|
Wed May 12 23:31:17 2010 |
Alberto | Update | 40m Upgrading | 216 MHz resonance in the POY11 PD killed |
It turned out that the resonance at 216 MHz in the 11MHz PD that I showed in the elog entry 2902 was casued by an instability of the of the MAX4107 opamap' feedback loop.
As the datasheet of the opamp shows, the close-loop gain has a peak at about 200-250MHz, in presence of even small capacitive loads.
In my case, perhaps either the capacitance of the BNC cables plugged to the RF output of the PD box, or the shunt capacitance of the circuit parts after the opamap (traces and resistors) might have introduced capacitance at the output of the amplifier.
LISO had failed in predicting the resonance because it has only ideal transfer functions of the opamps. In particular the open-loop gain of the opamaps in the library is just a function with a simple pole.
At RF frequencies the output impedances of the opamp starts having a non-negligible inductance that interacts with the load capacitance, generating a typical LC-circuit resonance.
In cases like this, such effect can be mitigated by introducing an "isolating" resistor at the output of the opamp.
So I did that and modified the circuit as in this simplified schematic here:
The choice of 100 Ohm for the isolating resistor was mainly empirical. I started with 10, then 20 and 50 until I got a sufficient suppression of the resonance. Even just 10Ohm suppressed the resonance by several tens of dB.

In that way the gain of the loop didn't change. Before that, I was also able to kill the resonance by just increasing the loop gain from 10 to 17. But, I didn't want to increase the closed-loop gain.
One thing that I tried, on Koji's suggestion, was to try to connect the RF output of the PD box to an RF amplifier to see whether shielding the output from the cable capacitance would make the resonance disappear: It did not work. |
2926
|
Thu May 13 05:06:43 2010 |
rana | Update | 40m Upgrading | 216 MHz resonance in the POY11 PD killed |
This idea was tried before by Dale in the ~1998 generation of PDs. Its OK for damping a resonance, but it has the unfortunate consequence of hurting the dynamic range of the opamp. The 100 Ohm resistor reduces the signal that can be put out to the output without saturating the 4107.
I still recommend that you move the notch away from the input of the 4107. Look at how the double notch solution has been implemented in the WFS heads. |
13544
|
Fri Jan 12 20:35:34 2018 |
Udit Khandelwal | Summary | General | 2018/01/12 Summary |
- 40m Lab CAD
- Worked further on positioning vacuum tubes and chambers in the building.
- Next step would be to find some drawings for optical table positions and vibration isolation stack. Need help with this!
- Tip Tilt Suspension (D070172)
- Increased the length of side arms. The overall height of D070172 assembly matches that of D960001.
- The files are present in dropbox in [40mShare] > [40m_cad_models] > [TT - Tip Tilt Suspension]
|
12055
|
Wed Mar 30 16:40:24 2016 |
ericq | Update | LSC | 2016 vs 2010 |
I haven't found any data files for the DARM spectrum of the previous generation of 40m, but with some GIMP-fu, I have plotted Monday's spectrum (green) on top of one of the figures from Rob's thesis.

|
9730
|
Mon Mar 17 10:50:58 2014 |
steve | Update | safety | 2014 annual crane certification |
Quote: |
Quote: |
We had our annual safety inspection today. Our SOPs are outdated. The full list of needed correction will be posted tomorrow.
The most useful found was that the ITMX-ISCT ac power is coming from 1Y1 rack. This should actually go to 1Y2 LSC rack ?
Please test this so we do not create more ground loops.
|
Annual crane inspection is scheduled for 8-11am Monday, March 17, 2014
The control room Smart UPS has two red extension cords that has to be removed: Nodus and Linux1
|
KroneCrane Fred inspected and certified the 3 40m cranes for 2014. The vertex crane crane was load tested at fully extended position. |
4651
|
Fri May 6 10:20:00 2011 |
steve | Summary | SAFETY | 2011 safety audit |
The emphasis of this annual safety audit was on safe electrical housekeeping on March 3, 2011
Safety audit correction list for the electric shop:
1, install breaker panel door in room 101
2, install conduit- AC out let in the east arm for USB camera table and
south arm for maglev- external fan
3, replace AC cord to south end work bench and door alarm
4, trace breaker of 1Y4
Requested completion date: 3-28-2011 at estimated cost $1,500.
All recommendations for improvement were done by April 1, 2011
We thank the participants for making the 40m a safer place to work.
|
1620
|
Fri May 22 01:27:14 2009 |
pete | Update | SUS | 200 days of MC3 side |
Looks like something went nuts in late April. We have yet to try a hard reboot. |
547
|
Fri Jun 20 01:38:55 2008 |
rana | Update | PEM | 20 day Weather |
Yoichi showed me that its possible to make PNG images from PS using GS:
gs -sDEVICE=png16m -sOutputFile=foo.png bar.ps |
10899
|
Wed Jan 14 02:11:07 2015 |
rana | Summary | Treasure | 2-loop Algebra Loopology |
I show here the matrix formalism to calculate analytically the loop TF relationships for the IMC w/ both FSS actuators so that it would be easier to interperet the results.
The attached PDF shows the Mathematica notebook and the associated block diagram.
In the notebook, I have written the single hop connection gains into the K matrix. P is the optical plant, C is the Common electronic gain, F is the 'fast' NPRO PZT path, and M is the phase Modulator.
G is the closed loop gain matrix. The notation is similar to matlab SS systems; the first index is the row and the second index is the column. If you want to find the TF from node 2 to node 3, you would ask for G[[3,2]].
As examples, I've shown how to get the FAST gain TF that I recently made with the Koji filter box as well as the usual OLG measurement that we make from the MC servo board front panel. |
16464
|
Thu Nov 11 00:11:39 2021 |
Koji | Summary | SUS | 2" to 3" sleeve issue |
Yehonathan and Tega found that the new PR3 and SR3 delivered in 2020 is in fact 3/4" in thickness (!). Digging the past email threads, it seems that the spec was 10mm but the thickness was increased for better relieving the residual stress by the coatings.
There are a few issues.
1. Simply the mirror is too thick for the ring. It sticks out from the hole. And the mirror retainers (four plastic plates) are too far from the designed surface, which will make the plates tilted.
2. The front side of the mirror assembly is too heavy and the pitch adjustment is not possible with the balance mass.
Some possible solutions:
- How about making the recess deeper?
In principle this is possible, but the machining is tricky because the recess is not a simple round hole but has "pads" where the mirror sits. And the distance of the retainer to the thread is still far.
And the lead time might become long.
- How about making new holes on the ring to shift the clamp?
Yes it is possible. This will shift the mirror assembly by a few mm. Let's consider this.
- How about modifying the wire blocks?
Yes it is equivalent to shift the holes on the ring. Let's consider this too.
1. How to hold the mirror with the retainer plates
[Attachment 1] The expected distance between the retainer plate and the threaded hole is 13.4mm. We can insert a #4-40 x L0.5" stand off (McMaster-Carr 91197A150, SUS316) there. This will make the gap down to 0.7mm. With a washer, we can handle this gap with the plate. Note that we need to use vented & silver plated #4-40 screws to hold the plates.
[Attachment 2] How does this look like when the CoM is aligned with the wire plane? Oh, no... the lower two plates will interfere with the EQ stops and the EQ stop holders. We have to remove them. [Attachment 3]
We need to check with the suspension if the EQ stop screws may hit the protruded optics and can cause chipping/cracking.
2. Modifying the wire block
[Attachment 4] The 4x thru holes of the wire block were extended to be +/-0.1" slots. The slots are too long to form ovals and produce thin areas. With the nominal position of the balance mass, the clamp coordinates are y=1.016 (vertical) and z=-2.54mm (longitudinal).
==> The CoM is 0.19mm backside (magnet side) and 0.9134 mm lower from the wire clamping points. This looks mathematically doable, but the feasibility of the manufacturing is questionable.
[Attachment 5] Because the 0.1" shift of the CoM is large, we are able to make new #2-56 thread holes right next to the original ones. The clamp coordinates are y=1.016 (vertical) and z=-2.54mm (longitudinal).
==> The CoM is 0.188mm backside (magnet side) and 0.9136 mm lower from the wire clamping points. With the given parameters, the expected pitch resonant frequency is 0.756Hz
My Recommendation
- Modify the metal ring to shift the #2-56 threads by 0.1"
- The upper two retainer plates will have #4-40 x 0.5" stand off. Use vented Ag-coated #4-40 screws.
- The lower two are to be removed.
- Take care of the EQ stops.
- Of course, the best solution is to redesign the holder for 3/4" optics. Can we ask Protolab for rapid manufacturing???
|
Why did we need to place the mass forward to align the 1/4" thick optic?
We were supposed to adjust the CoM not to have too much adjustment. But we had to move the balance mass way too front for the proper alignment with a 1/4" thick optic. Why...?
This is because the ring was designed for a 3/8" thick optic... It does not make sense because the depth of the thread holes for the retainer plate was designed for 1/4" optics...
When the balance mass is located at the neutral position, the CoM coordinate is
x 0.0351mm (x+: left side at the front view)
y 0.0254mm (y+: vertical up)
z 0.4493mm (z+: towards back)
So, the CoM is way too behind. When the balance mass was stacked and the moved forward (center of the axis was moved forward by 0.27"), the CoM coordinate is (Attachment 6)
x 0.0351mm
y 0.0254mm
z 0.0011mm
This makes sens why we had to move the balance mass a lot for the adjustment. |
16296
|
Wed Aug 25 08:53:33 2021 |
Jordan | Update | SUS | 2" Adapter Ring for SOS Arrived 8/24/21 |
8 of the 2"->3" adapter rings (D2100377) arrived from RDL yesterday. I have not tested the threads but dimensional inspection on SN008 cleared. Parts look very good. The rest of the parts should be shipping out in the next week. |
16364
|
Wed Sep 29 09:36:26 2021 |
Jordan | Update | SUS | 2" Adapter Ring Parts for SOS Arrived 9/28/21 |
The remaining machined parts for the SOS adapter ring have arrived. I will inspect these today and get them ready for C&B. |
820
|
Mon Aug 11 00:58:31 2008 |
rana | Update | PEM | 2 years of temperature trend |
The PSL RMTEMP alarmed again because it says the room temperature is 19.5 C. Steve said in
an earlier log entry that this is a false alarm but he didn't say why he thought so...
I say that either the calibration of the RMTEMP channel has drifted, the setpoint of the HVAC
has shifted, or there's a drift in the RMTEMP channel. I don't know what electronics exactly
are used for this channel so not sure if its susceptible to so much drift.
However, since the Dust Monitor (count_temp) shows a similar temperature decline in the
last two years I am inclined to blame the HVAC system.
The attached plot shows 2 years of hour-mean trend. |
821
|
Mon Aug 11 09:39:29 2008 |
rana | Update | PEM | 2 years of temperature trend |
Steve and I went around and inspected and then adjusted the thermostats and humidostats.
All the thermostats were set at 70F in 2005 by Steve. We adjusted the ones on the arms up to 72F
and set the one on the wall west of the framebuilder up to 74F (this area was consistently colder
than all the others and so we're over-correcting intentionally). |
5783
|
Wed Nov 2 10:41:47 2011 |
steve | Update | General | 2 spare heliax at the LSC rack |
They were traced and labeled. One goes to 1X2 and the other to AS-ISCT. They are Andrew Heliax 1/4" od. made by CommScone, model number FSJ1-50A |
16778
|
Thu Apr 14 10:18:35 2022 |
Paco | Summary | BHD | 2 in oplev mirrors incompatible with LMR2V |
[Paco, JC]
We realized the 2 in oplev mirrors (Thorlabs BB2-E02) for ITMYOL, SRMOL, and BSOL, are 0.47 in thick, while the LMR2V fixed mount is 0.46 in deep, even without taking the retaining ring into account. After a brief exchange with Koji, and Ian, we decided to glue the mirrors onto the mounts using Torr Seal (a vac compatible epoxy). They are curating in the clean room and should be ready to install in about 2 hours. |
3622
|
Wed Sep 29 16:56:36 2010 |
yuta | Update | VAC | 2 doors opened |
(Steve, Koji, Joe, Kiwamu, Yuta)
Background:
The vent was started on Monday, and finished on Tuesday.
We were to open the doors on Tuesday, but we couldn't because the vertex crane got out of order.
Now the crane was fixed, and so we opened the doors today.
What we did:
We opened the north side of the BS chamber and the west side of the ITMX chamber.
Now, the light doors are put instead. |
2744
|
Wed Mar 31 16:55:05 2010 |
josephb | Update | Computers | 2 computers from Alex and Rolf brought to 40m |
I went over to Downs today and was able to secure two 8 core machines, along with mounting rails. These are very thin looking 1U chassis computers. I was told by Rolf the big black box computers might be done tomorrow afternoon. Alex also kept one of the 8 core machines since he needed to replace a hard drive on it, and also wanted to keep for some further testing, although he didn't specify how long.
I also put in a request with Alex and Rolf for the RCG system to produce code which includes memory location hooks for plant models automatically, along with a switch to flip from the real to simulated inputs/outputs.
|
9691
|
Wed Mar 5 11:33:10 2014 |
Koji | Summary | LSC | 2 arm ALS->LSC transition - road map |
Step by step description of transition from 2arm ALS to Common/Differential LSC for FPMI
- Step 0: Place the frequencies of the arm green beams at the opposite side of the carrier green.
- Step 1: Activate stablization loops for ALSX and ALSY simultaneously.
(Use LSC filter modules for the control. This still requires correct handling of the servo and filter module triggers)
- Step 2: Activate stablization loops for ALS Common and Differential by actuating ETMX and ETMY
- Step 2 (advanced): Activate stabilization loops for ALS Common by actuating MC2 and ALS Differential by ETMX and ETMY
- Step 3: Transition from ALS Common to 1/SQRT(TRX)+1/SQRT(TRY). Make sure that the calibration of TRX and TRY are matched.
The current understanding is that the offset for 1/SQRT(TRX)+1/SQRT(TRY) can't be provided at the servo filter. Figure out
what is the correct way to give the offsets to the TR signals.
- Step 4: Lock Michelson with AS55Q and then POP55Q (PD not available yet) or any other PD, while the arms are kept off-resonant using ALS.
- Step 5: Reduce the TR offsets. Transition to RF CARM signals obtained from POP55I or REFL11I in the digital land.
- Step 5 (advanced): Same as test6 but involve the analog common mode servo too.
- Step 6: Transition from ALS Differential to AS55Q
Independent test: One arm ALS (To be done everyday)
- ALS resonance scan
- Measurement of out-of-loop displacement (or frequency) stability
- Check openloop transer function
Independent test: Common Mode servo for one arm
- Reproduce Decmber CM servo result of transition from one arm ALS to CM servo
Insert 1/sqrt(TRY) servo in between?
- How can we realize smooth transition from ALS to POY11? |
6851
|
Fri Jun 22 02:21:57 2012 |
Jenne | Update | Green Locking | 2 arm ALS - Success!!!! |
[Yuta, Jenne]
We locked both arms using the ALS system simultaneously! Hooray!
Video of spectrum analyzer during lock acquisition of both beats is attached.
Jamie is super awesome, since he fixed us up a beatbox speedy-quick. Thanks Jamie!! 
Details:
1: Aligned PSL green optics
1.1: We added an amplifier of ~20dB after the X beat PD (more Xgreen power on the PSL table so the signal was ~3dB higher than Y, so required less amplification). The ~24dB amplifier is still in place after the Y beat PD. Both beat signals go to a splitter after their amplifiers. One side of each splitter goes to one of the channels on the beatbox. The other side of each splitter goes to a 3rd splitter, which we're using backwards to combine the 2 signals so we can see both peaks on the spectrum analyzer at the same time.
2: Found both beat notes
2.1: Y beat was easy since we knew the temps that have been working for the past several days
2.2: X beat was more tricky - the last time it was locked was the end of February (elog 6342)
2.2.1: We found it by adjusting the PSL laser temp nearly the full range - DC Adjust slider was at 8.8V or so (Y beat was found with the slider at ~1.1V tonight)
2.2.2: We then walked the beat around to get the PSL temp back to "normal" by moving the PSL temp, then compensating with the Xend laser temp, keeping the beatnote within the range of the spectrum analyzer.
2.2.3: Fine tuned the temps of all 3 lasers until we had 2 peaks on the analyzer at the same time!!
2.2.3.1: Yend - measured Temp=34.14 C, thermal Out of Slow servo=29820
2.2.3.2: Xend - displayed temp=39.33 C, thermal Out of Slow servo=5070
2.2.3.3: PSL - displayed temp=31.49 C, Slow actuator Adjust=1.100V
3: Locked both arms using ALS!!
3.1: We were a little concerned that the Xarm wasn't locking. We tried switching the cables on the beatbox so that we used the old channels for the Xarm, since the old channels had been working for Y. Eventually we discovered that the input of the filter module for ETMX's POS-ALS input was OFF, so we weren't really sending any signals to ETMX. We reverted the cabling to how it was this evening when Jamie reinstalled the beatbox.
3.1.1: We need to sort out our SUS screens - Not all buttons in medm-land link to the same versions of the SUS screens! It looks like the ALS screen was modified to point the ETMY button to a custom ETMY SUS screen which has the ALS path in the POS screen, along with LSC and SUSPOS. There is no such screen (that I have found) for ETMX. The regular IFO_ALIGN screen points to the generic SUS screens for both ETMY and ETMX, so we didn't know until Yuta searched around for the filter bank that the ALS input for ETMX was off. We just need to make sure that all of the screens reflect what's going on in the models.
3.2: See the video attached - it shows the beat peaks during locking!!! (how do I embed it? right now you have to download it)
3.2.1: First you will see both peaks moving around freely
3.2.2: Then X arm is locked briefly, then unlocked
3.2.3: Y arm is locked, steadily increasing gain
3.2.4: X arm is locked, so both arms locked simultaneously
3.2.5: Yuta clicked a button, accidentally unlocking the Xarm
4: The transmission of the X arm was not so great, and both of our green beams (although X green especially) were no longer nicely aligned with the cavities. Yuta tried to align the X arm to the X green, but it's bad enough that we really need to start over with the whole IFO alignment - we leave this until tomorrow. Since we didn't have any good IR transmission, we didn't bother to try to find and hold the Xarm on IR resonance using ALS, so we didn't measure a POX out of loop residual cavity motion spectrum. Again, tomorrow. |
6852
|
Fri Jun 22 03:37:42 2012 |
Koji | Update | Green Locking | 2 arm ALS - Success!!!! |
Are these correct?
1. It is a nice work.
2. This is not locking, but stabilization of the both arms by ALS.
3. We now have the phase trackers for both arms.
4. There is no coarse (i.e. short) delay line any more.
5. The splitters after the PDs are reducing the RF power to Beat-box.
Actually there are RF monitors on Beat-box for this purpose, but you did not notice them.
6. c1ioo channel list
https://wiki-40m.ligo.caltech.edu/CDS/C1IOO%20channel%20list
has to be updated.
7. Video can be uploaded to Youtube as Mike did at http://nodus.ligo.caltech.edu:8080/40m/6513 |
6853
|
Fri Jun 22 10:52:18 2012 |
yuta | Update | Green Locking | 2 arm ALS - Success!!!! |
Answers to questions from Koji.
Are these correct?
1. It is a nice work.
Correct, of course!
2. This is not locking, but stabilization of the both arms by ALS.
Correct.
3. We now have the phase trackers for both arms.
Correct.
4. There is no coarse (i.e. short) delay line any more.
Correct. No coarse, only fine delay line (30m) with the phase tracker.
5. The splitters after the PDs are reducing the RF power to Beat-box.
Actually there are RF monitors on Beat-box for this purpose, but you did not notice them.
Oh, yes. But distance between beatbox and spectrum analyzer in the control room is longer than distance between BBPD on PSL table and the spectrum analyzer. We were too lazy to do cabling, but maybe we should.
6. c1ioo channel list
https://wiki-40m.ligo.caltech.edu/CDS/C1IOO%20channel%20list
has to be updated.
Yes, we will.
7. Video can be uploaded to Youtube as Mike did at http://nodus.ligo.caltech.edu:8080/40m/6513
We didn't, but we can. |
3295
|
Mon Jul 26 20:30:35 2010 |
Jenne | Update | SUS | 2 Tip Tilts suspended and balanced |
[Koji, Jenne]
We were on Team Cleanroom, while Kiwamu and Alberto were on Team Chamber. Team Cleanroom suspended and balanced 2 Tip Tilts this afternoon.
One of the TTs that was suspended today is the one which was broken on Friday (see elog 3278). We resuspended it using the regular 0.0036" diameter wire (91um). We balanced it using the HeNe oplev, and then set it aside. This TT has serial number 2.
We noticed that, like the previous 2 TT suspensions (this one before it was broken, and the one actually installed in the BS chamber on Friday, which is #3), there seems to be a little bit of hysteresis in the pointing. The difference comes if we poke the top of the mirror holder and observe the place the reflected beam spot comes to rest at, and if we poke the bottom of the mirror holder. The beam spot stays a little higher when we poke the top vs. when we poke the bottom.
To combat this, we tried suspending our second TT of the day (the one that Kyung Ha and I had half finished) using thinner wire for the mirror holder. We used the 0.0017" diameter wire (43um) that is used for the SOSes. Unfortunately, it still seems like there is a similar hysteresis. The thin-wire TT has serial number 4.
While working on TT4, we recalled that we have to include rubber dampers for the vertical blade springs. Oooops! We used some of the leftover #4-40 screws with viton tips that Zach and Mott had made for Earthquake stops to damp the vertical resonance of the blades. We measured the Q factor by flicking the blades up or down. We changed the oplev setup to be a shadow sensor setup, and watched the ringdown of the vertical mode on the 'scope. We counted #cycles/time = frequency, and the t(1/2) time for the exponential ringdown to calculate the Q. For the shadow sensor, we positioned the QPD in line with the initial HeNe beam, and placed the edge of the mirror holder clamp partially in the beam, so the beam was partly occluded. When the mirror shook up and down, more or less of the beam was blocked, and we could see this power fluctuation on the 'scope.
Using the formula Q = pi f0 T1/2 / ln(2) = 4.53 f0 T1/2, where T1/2 is the the time it takes for the amplitude to decay by half, we measured a Q of 31 for the vertical mode with no damping, and a Q of 14 with damping. Koji confirmed the calculation and put it into wiki.
We need to go through the other TTs that have been assembled and give them their rubber dampers.
|
2574
|
Fri Feb 5 14:31:46 2010 |
Jenne | Update | SUS | 2 SOS towers assembled |
[Jenne, Kiwamu]
The 2 SOS towers for the ITMs have been assembled, and are on the flow bench in the cleanroom. Next up is to glue magnets, dumbells, guiderods and wire standoffs to the optics, then actually hang the mirrors.

|
4830
|
Fri Jun 17 00:17:26 2011 |
rana | Configuration | Electronics | 2 RFPDs sent to LLO |
Koji and I found 2 RFPD boxes to send to LLO. We've put them onto Steve's desk to be overnighted to Valera.
One of them is our old 21.5 MHz gold box RFPD from the FSS (which we don't use). The other one is a 2mm gold box one which was previously tuned for 66 MHz.
They shipped out on Friday |
1349
|
Tue Mar 3 11:39:50 2009 |
Osamu | DAQ | Computers | 2 PCs in Martian |
Kiwamu and I brought 2 SUPER MICRO PCs from Willson house into 40m.
Both PCs are hooked up into Martian network. One is named as bscteststand for BSC which has been set up by Cds people and another one is named kami1 for temporary use for CLIO which is a bland new, no operating installed PC. This bland new PC will be returned Cds or 40m once another new PC which we will order within several days arrives.
IP address for each machine is 131.215.113.83 and 131.215.113.84 respectively.
We have installed CentOS5.2 into the new PC. |
7385
|
Fri Sep 14 01:18:51 2012 |
rana | Update | COC | 2 Layout Changes |
After looking at the in-vacuum layout I think we should make two changes during the next vent:
1) Reduce the number of mirrors between the FI and its camera. We install a large silvered mirror in the vacuum flange which holds the Faraday cam (in the inside of the viewport). That points directly at the input to the Faraday. We get to remove all of the steering mirror junk on the IO stack.
2) Take the Faraday output (IFO REFL) out onto the little table holding the BS and PRM Oplevs. We then relocate all 4 of the REFL RFPDs as well as the REFL OSA and the REFL camera onto this table. This will reduce the path length from the FI REFL port to the diodes and reduce the beam clutter on the AS table. |
7398
|
Mon Sep 17 18:04:01 2012 |
Steve | Update | COC | 2 Layout Changes |
Quote: |
After looking at the in-vacuum layout I think we should make two changes during the next vent:
1) Reduce the number of mirrors between the FI and its camera. We install a large silvered mirror in the vacuum flange which holds the Faraday cam (in the inside of the viewport). That points directly at the input to the Faraday. We get to remove all of the steering mirror junk on the IO stack.
2) Take the Faraday output (IFO REFL) out onto the little table holding the BS and PRM Oplevs. We then relocate all 4 of the REFL RFPDs as well as the REFL OSA and the REFL camera onto this table. This will reduce the path length from the FI REFL port to the diodes and reduce the beam clutter on the AS table.
|
There is just so much room on this table. |
3960
|
Sat Nov 20 02:25:30 2010 |
yuta | Update | CDS | 2 LOCKINs for suspension models |
(Suresh, Koji, Yuta)
Background:
No AWG. No tdssine.
...... LOCKIN!
What we did:
1. Added 2 LOCKINs for c1sus model.
Currently, we cannot put cdsOsc in a subsystem.
So, we put LOCKINs just for BS for a test.
The signal going into LOCKIN can be anything. For now, we just put a matrix for selecting the signal and connected the input signals to the ground.
See the following page for the current simlink diagram of c1sus model.
https://nodus.ligo.caltech.edu:30889/FE/c1sus_slwebview_files/index.html
2. Edited MEDM screens. (see Attachment #1)
Result:
We succeeded in putting 2 LOCKINs and exciting BS.
During the update, we might destroyed things. For example, fb status is red in GDS screens.
We will wait for Joe to fix them.
Plan:
- Fix cdsOsc and put LOCKINs for all the other optics
- Come up with a good idea what to do with this LOCKIN. Remember, LOCKIN is not just a replacement for excitation points.
- Enhance an oscillator so that we can put a random noise |
3314
|
Wed Jul 28 18:24:57 2010 |
Jenne | Update | Green Locking | 2 Green Periscopes have mirrors, aligned |
[Koji, Jenne, Kiwamu]
This is to describe the work that went on in the Cleanroom today. Kiwamu's entry will detail the tidbits that happened in the chamber.
We engraved the periscope mounts with the mirror info for the mirrors which were placed in the periscope. We also engraved the barrels of the optics with their info, for posterity. Koji carefully put the mirrors into the periscopes. Since we have wedged optics, the goal was to have the front HR surface of the mirror parallel to the plane of the mount, and leave a bit of space behind one side of the optic (if we just pushed the optic fully in, the HR surface wouldn't be flat, and would send the beam off to the left or right somewhere). Once the mirrors were mounted in the periscopes, we checked the vertical levelness of the outcoming beam. For the first periscope (the one which has been installed on the BS table), the beam was deflected upward (2.5)/32 inches over 55inches. This corresponds to a 1.4mRad vertical deflection. The second periscope (which will eventually be installed on the OMC table) had a deflection of 1/32 over 55inches, or 0.6mRad. We did not check the side-to-side deflection for either of the periscopes.
We also engraved one more DLC mount with mirror info, and put a mirror into the mount. This is one of the optics that was placed onto the BS table today, which Kiwamu will describe.
We removed TT#3 from the BS chamber so that it could have rubber vertical dampers installed, and be characterized. For future reference, the #'s of the Tip Tilts refers to the serial number of the suspension block piece, which forms the top horizontal bar of the frame. |
329
|
Thu Feb 21 19:55:46 2008 |
rana | Update | Electronics | 2 BNC Cables, 1 Tee |
I'm not sure where Ward and Miller went to Analyzer school, but it was probably uncredited.
I turned it on and used 2 BNC cables and a T to hook up the source to the 2 inputs and measured the always-exciting TF of cable.
Score: HP Analyzer 1
Rob & John 0
I have left the analyzer on in this complicated configuration. RTFM boys.
Quote: | The HP 4195A network analyser may be broken, measurements below 150MHz are not reliable. Above 150MHz everything looks normal. This may be caused by a problem with its output (the one you'd use as an excitation) which is varying in amplitude in a strange way.
Analyzer |
|
7626
|
Thu Oct 25 21:02:34 2012 |
Den | Update | PEM | 1x7 dc power |
We now stop using bench DC power supplies for microphone preamp and PEM AA board. DC power is wired from 1x5 rack suppliers. I've installed a beam to mount fuse houses in the 1x7 as we did not have one.

|
6076
|
Tue Dec 6 02:57:44 2011 |
kiwamu | Update | Green Locking | 1st trial of handing off |
I succeeded in handing off the servo from that of the ALS to IR-PDH.
However the handing off was done by the coarse sensor instead of the fine sensor because I somehow kept failing to hand off the sensor from the coarse to the fine one.
The resultant rms in the IR-PDH signal was about a few 100 pm, which was fully dominated by the ADC noise of the coarse sensor.
Tomorrow I will try :
(1) Using the fine sensor.
(2) Noise budgeting with the fine sensor.
Here is the actual time series of the handing off.

(Upper left ): intracavity power.
As the offset was adjusted the power increased to ~ 0.8. Eventually the power becomes close to the nominal value of 1 after the handing off.
(Lower left) : Frequency of the beat-note.
After the engagement of the ALS servo, I was scanning the arm length and searching for the resonance by changing the error point of this signal.
(Lower right) : IR-PDH signal. |
8507
|
Mon Apr 29 18:53:03 2013 |
Jenne | Update | Electronics | 1pps timing fiber to OMC rack may be bent |
While helping Riju out this afternoon, I noticed that the timing fiber that goes to the OMC rack (near the AP table) was bent, and is now possibly kinked, after the installation of the fiber splitter box.
The fiber was hanging from the back of the rack, and had been strain relieved. However, the path that the fiber was taking is now occupied by the fiber splitter for the RF PD diagnostic stuff. So, the installation of the fiber splitter box put the old timing fiber under tension, causing the fiber to be bent at a little over 90 degrees, since it was pulled tightly against the corner of the splitter's front panel.
I adjusted the strain relief so that the fiber is loose again, although there is still a bit of a kink that you can feel. Things (for now) seem to be working, since the 1pps light on the front of the box at the top of the OMC rack is still blinking happily, indicating that the 1pps is still getting there.
We are not using most of the stuff in that rack right now, but if we have problems in the future, we should check out the fiber to make sure it is still good. |
593
|
Sun Jun 29 18:58:43 2008 |
rana | Summary | Computers | 1e20 is too big for AWG and/or IOVME |
While testing out my matlab/awgstream based McWFS diagnostic script I accidentally put a
huge excitation into C1:IOO-WFS1_PIT_EXC . This went to 1e20 and then caused
some SUS to trip and c1susvme2 to go red. I tried booting it via the normal procedures
but it wouldn't come back, even after 2 crate power cycles. I also tried booting AWG
via the vmeBusReset, but that didn't do it. Then I booted c1iovme from the telnet prompt
and then I could restart c1susvme2 successfully.
The reason the excitation was so large is that the following filter command is unstable:
[b,a] = butter(4,[0.02 30]/1024);
The low pass part is OK, but it looks like making such a low frequency digital filter
is not. Que lastima. On the bright side, the code now has some excitation amplitude
checking. |
2173
|
Tue Nov 3 12:47:01 2009 |
Koji | Configuration | CDS | 1Y9 Rack configuration update |
For the CDS upgrade preparation I put and moved those stuff at the rack 1Y9:
Placed 1Y9-12 ADC to DB44/37 Adapter LIGO D080397
Placed 1Y9-14 DAC to IDC Adapter LIGO D080303
Moved the ethernet switch from 1Y9-16 to 1Y9-24
Wiki has also been updated. |
948
|
Mon Sep 15 14:00:52 2008 |
josephb | Configuration | Computers | 1Y9 Hub and C1asc |
The 1Y9 switch is now using a labeled Cat6 cable in cable trays to connect to the main switch in the offices. In addition, the c1asc cable which had been coming out the door was fixed last Friday, and is now labeled, going out the top and connects to the hub in 1Y2.
Note: Do not connect new ethernet cable from switch to switch without disconnecting the old cable to the rest of the network - this tends to make the Ethernet network unhappy with white flashing alarms. |
14848
|
Fri Aug 16 16:40:04 2019 |
gautam | Update | CDS | 1Y3 work |
[chub, gautam]
Installation: The following equipment were installed in 1Y3, see Attachment #1:
- Supermicro server, which is the new c1iscaux machine, with IP Address 192.168.113.83.
- 6U Acromag chassis which contains all the ADCs, DACs and BIO units.
- 2 Sorensen DC power supplies to provide +24 V DC and +15 V DC to the Acromags.
- Fusable DIN rail power blocks were installed on the North side of the 1Y3 rack - I placed 2 banks of 5 connectors each for +15 V DC and +24 V DC.
Removal: The following equipment was removed from 1Y3:
- VME crates that were the old c1iscaux and c1iscaux2 machines.
- Spare VME crate that used to be c1susaux, which Chub and I brought over to 1Y3 in an attempt to revive the broken c1iscaux2.
- Approximately 30 twisted ribbon cables that were going to the cross connects. For now, we have not done a full cleanup and they are just piled along the east arm (see Attachment #2), beware if you are walking there!
Software:
- I connected the c1iscaux machine to the martian network.
- Then I edited the relevant files on chiara to free up the IP addresses previously used by c1iscaux (192.168.113.81) and c1iscaux2 (192.168.113.82), and re-assigned the IP address used for c1iscaux to be 192.168.113.83.
- I also changed the hostname of the c1iscaux machine (it was temporarily called c1iscaux3 to allow bench testing).
- I moved the old /cvs/cds/caltech/target/c1iscaux and /cvs/cds/caltech/target/c1iscaux2 directories to /cvs/cds/caltech/target/preAcromag_oldVME/c1iscaux and /cvs/cds/caltech/target/preAcromag_oldVME/c1iscaux2 respectively.
- I moved the temporarily named /cvs/cds/caltech/target/c1iscaux3 directory, from which I was running all the tests, to /cvs/cds/caltech/target/c1iscaux.
- I edited all references to c1iscaux3 in the systemd files so that we can run the approriate systemd services.
Next steps:
- We did not get around to running the DB37 cables between the Acromag chassis and the 1Y2 Eurocrates today - this operation itself took the whole day as we also needed to lay out some support struts etc on the rack to support the Sorensens and the Acromag chassis.
- Once the Acromags are connected to the Eurocrates, we have to run in-situ tests to make sure the appropriate functionality has been restored.
- We must have bumped something in the c1lsc expansion chassis - the CDS FE overview screen is reporting some errors (see Attachment #3). I will fix this.
- General tidiness, strain-relief etc.
Quote: |
I judge that we are good to go ahead with an install tomorrow.
|
|
4609
|
Tue May 3 10:59:31 2011 |
josephb | Update | CDS | 1Y2 binary output adapter board now powered |
I temporarily turned off the power to the 1Y2 rack this morning while wiring in the binary output adapter board power (+/- 15V) into the cross connects.
The board is now powered and we can proceed to testing if can actually control the LSC whitening filters. |
4717
|
Sat May 14 14:50:21 2011 |
Koji | Update | LSC | 1Y2 5V Blown Fuse found -> Fixed |
Incidentally, a blown fuse on 5V line at 1Y2 rack was found during the intallation of Sorensens.
The fuse (5A 125V) has been replaced and fixed.
When I plugged the fuse in, I heard some sound like relays were switched. Are there any relays in the LSC rack?
It was a 9th fuse from the top as seen in the picture. |