16071
|
Thu Apr 22 08:50:21 2021 |
Anchal | Update | SUS | MC2 Suspension Optimization summary | Yes, during the AC balancing, POS column
was set to all 1. This table shows the final
values after all the steps. The first 3 columns |
|
16072
|
Thu Apr 22 12:17:23 2021 |
Anchal, Paco | Update | SUS | MC1 and MC3 Suspension Optimization Summary | |
|
16073
|
Thu Apr 22 14:22:39 2021 |
gautam | Update | SUS | Settings restored | The MC / WFS stability seemed off to me.
Trending some channels at random, I saw that
the MC3 PIT/YAW gains were restored mixed |
|
16074
|
Thu Apr 22 14:41:55 2021 |
Chub | Update | PSL | New HEPA speed control | When adjusting the blower speed, give the
blower at least 30 seconds to speed up or
slow down to the set speed. The flywheel |
|
16075
|
Thu Apr 22 14:49:08 2021 |
gautam | Update | Computer Scripts / Programs | rossa added to RTS authorized keys | This is to facilitate running of scripts
like the CDS reboot script, mx_stream restart,
etc, from rossa, without being pwd prompted |
|
16076
|
Thu Apr 22 15:15:26 2021 |
gautam | Update | PSL | PMC transmission | I was a bit surprised by these numbers
suggesting the PMC transmission is only 50-60%.
I went to the table today and confirmed that |
|
16077
|
Thu Apr 22 15:34:54 2021 |
Anchal | Update | PSL | PMC transmission | Koji mentioned that the mode of the laser
is different for lower diode currents. So
that might be the reason why we got less |
|
16078
|
Thu Apr 22 15:36:54 2021 |
Anchal | Update | SUS | Settings restored | The mix up was my fault I think. I restored
the channels manually instead of using burt
restore. Your message suggests that we can |
|
16079
|
Thu Apr 22 17:04:17 2021 |
gautam | Update | SUS | Settings restored | Indeed, you can make your own snapshot
by specifying the channels to snap in a .req
file. But what I meant was, we should confirm |
|
16080
|
Thu Apr 22 17:28:34 2021 |
Yehonathan | Update | PSL | Laser amplifier | According to the aLIGO
70W amplifier interlock concept the flow
rate of the chiller should be between 5 and |
  |
16081
|
Fri Apr 23 15:52:19 2021 |
Ian MacMillan | Update | CDS | 40m LSC simPlant model | I have attached the framework that I am
using for the full system. Plantframework.pdf
has the important aspects that I will be |
  |
16082
|
Fri Apr 23 18:00:02 2021 |
gautam | Update | PSL | HEPA speed lowered | I will upload some plots later - but in
summary, I set the HEPA speed to ~40%. I
used (i)IMC transmission RIN, (ii) Arm cavity |
|
16083
|
Fri Apr 23 19:26:58 2021 |
Koji | Update | PSL | HEPA speed lowered | I believe that there is an internal setting
for the minimum flow, so the flow is not
linear ("0%" is not zero), but |
|
16084
|
Sun Apr 25 21:21:02 2021 |
rana | Update | CDS | SUS simPlant model |
I suggest not naming this the LSC
model, since it has no LSC stuff.
Also
remove all the diagnostic stuff in the plant |
|
16086
|
Mon Apr 26 18:55:39 2021 |
Anchal, Paco | Update | SUS | MC2 F2A Filters Tested | Today we tested the F2A filters created
from the DC gain values listed in 16066.
Filters:
For |
  |
16087
|
Tue Apr 27 10:05:28 2021 |
Anchal, Paco | Update | SUS | MC1 and MC3 F2A Filters Tested | We extended the f2a filter implementation
and diagnostics as summarized in 16086
to MC1 and MC3. |
   |
16088
|
Tue Apr 27 15:15:17 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | The first version of the single filter
plant is below. Jon and I went through compiling
a model and running it on the docker (see |
 |
16089
|
Wed Apr 28 10:56:10 2021 |
Anchal, Paco | Update | SUS | IMC Filters diagnosed | Good
morning! |
7x |
16090
|
Wed Apr 28 11:31:40 2021 |
Jon | Update | VAC | Empty N2 Tanks | I checked out what happened on c1vac. There
are actually two independent monitoring codes
running: |
|
16091
|
Wed Apr 28 17:09:11 2021 |
Anchal | Update | SUS | Tuned Suspension Parameters uploaded for long term behavior monitoring | I have uploaded all the new settings mentioned
in 16066
and 16072. |
|
16092
|
Wed Apr 28 18:56:57 2021 |
Yehonathan, Jon | Update | CDS | Updated c1auxey wiring plan | We took a Supermicro from the lab (along
with a keyboard, a mouse, and a screen taken
from a table on the Y arm) and placed it |
|
16093
|
Thu Apr 29 10:51:35 2021 |
Jon | Update | CDS | I/O Chassis Assembly | Summary
Yesterday I unpacked
and installed the three 18-bit DAC cards
received from Hanford. I then repeated the |
|
16094
|
Thu Apr 29 10:52:56 2021 |
Anchal | Update | SUS | IMC Trans QPD and WFS loops step response test | In 16087
we mentioned that we were unable to do a
step response test for WFS loop to get an |
 |
16096
|
Thu Apr 29 13:41:40 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | To add the required library: put the .mdl
file that contains the library into the userapps/lib folder.
That will allow it to compile correctly |
|
16097
|
Thu Apr 29 15:11:33 2021 |
gautam | Update | CDS | RFM | The problem here was that the RFM errors
cropped up again - seems like it started
~4am today morning judging by TRX trends. |
 |
16098
|
Thu Apr 29 16:35:51 2021 |
Yehonathan | Update | CDS | Updated c1auxey wiring plan | I installed the EPICs base, asyn and modbus
modules according to Jon's
instructions. |
|
16099
|
Thu Apr 29 17:43:16 2021 |
Koji | Update | CDS | RFM | The other day I felt hot at the X end.
I wondered if the Xend A/C was off, but the
switch right next to the SP table was ON |
|
16100
|
Thu Apr 29 17:43:48 2021 |
Anchal | Update | CDS | F2A Filters double check | I double checked today and the F2A filters
in the output matrices of MC1, MC2 and MC3
in the POS column are ON. I do not get what |
|
16102
|
Thu Apr 29 18:53:33 2021 |
Anchal | Update | SUS | IMC Suspension Damping Gains Test | With the input matrix, coil ouput gains
and F2A filters loaded as in 16091,
I tested the suspension loops' step response |
  |
16103
|
Thu Apr 29 19:55:45 2021 |
Yehonathan | Update | CDS | Updated c1auxey wiring plan | We received a stock of DB9 male feed-through
connectors. That allowed me to complete the
remaining wiring on the c1auxey Acromag chassis. |
|
16105
|
Fri Apr 30 00:20:30 2021 |
gautam | Update | CDS | F2A Filters double check | The SDF system is supposed to help with
restoring the correct settings, complementary
to burt. My personal opinion is that there |
|
16106
|
Fri Apr 30 12:52:14 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | Now that the model is finally compiled
I need to make an medm screen for it and
put it in the c1sim:/home/controls/docker-cymac/userapps/medm/ directory. |
|
16107
|
Fri Apr 30 19:18:51 2021 |
Yehonathan, Jon | Update | CDS | Updated c1auxey wiring plan | We finished the installation procedure
on the c1auxey1 host machine. There were
some adjustments that had to be made for |
|
16108
|
Mon May 3 09:14:01 2021 |
Anchal, Paco | Update | LSC | IMC WFS noise contribution in arm cavity length noise | Lock
ARMs |
 |
16109
|
Mon May 3 13:35:12 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | When the cymac is started it gives me a
list of channels shown below.
|
|
16110
|
Mon May 3 16:24:14 2021 |
Anchal | Update | SUS | IMC Suspension Damping Gains Test Repeated with IMC unlocked | We repeated the same test with IMC unlocked.
We had found these gains when IMC was unlocked
and their characterization needs to be done |
  |
16111
|
Mon May 3 16:49:04 2021 |
Yehonathan | Update | BHD | SOS assembly | I found a "vice" in the cleanroom
(attachment 1). I used it to push dowel pins
into the last suspension block using some |
 |
16112
|
Mon May 3 17:28:58 2021 |
Anchal, Paco, Rana | Update | LSC | IMC WFS noise contribution in arm cavity length noise | Rana came and helped us figure us where
to inject the noise. Following are the characteristics
of the test we did: |
 |
16114
|
Mon May 3 20:36:46 2021 |
Yehonathan, Jon | Update | CDS | Updated c1auxey wiring plan | It seemed like the BIO channels were not
working, both the inputs and the outputs.
The inputs were working on the windows machine |
|
16116
|
Tue May 4 07:38:36 2021 |
Jon | Update | CDS | I/O Chassis Assembly | IOP models created
With all
the PCIe issues now resolved, yesterday I
proceeded to build an IOP model for |
|
16117
|
Tue May 4 11:43:09 2021 |
Anchal, Paco | Update | LSC | IMC WFS noise contribution in arm cavity length noise | We redid the WFS noise injection test and
have compiled some results on noise contribution
in arm cavity noise and IMC frequency noise |
  |
16118
|
Tue May 4 14:55:38 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | After a helpful meeting with Jon, we realized
that I have somehow corrupted the sitemap
file. So I am going to use the code |
|
16119
|
Tue May 4 19:14:43 2021 |
Yehonathan | Update | General | OSEMs from KAGRA | I put the box containing the untested OSEMs
from KAGRA near the south flow bench on the
floor. |
|
16120
|
Wed May 5 09:04:47 2021 |
Anchal | Update | SUS | New IMC Suspension Damping Gains uploaded for long term testing | We have uploaded the new damping gains
on all the suspensions of IMC. This completes
changing all the configuration to as mentioned |
|
16121
|
Wed May 5 13:05:07 2021 |
Chub | Update | General | chassis delivery from De Leone | Assembled chassis from De Leone placed
in the 40 Meter Lab, along the west wall
and under the display pedestal table. |
|
16122
|
Wed May 5 15:11:54 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | I added the IPC parts back to the plant
model so that should be done now. It looks
like this again here. |
|
16124
|
Thu May 6 16:13:24 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | When using mdl2adl I
was getting the error:
|
|
16126
|
Fri May 7 11:19:29 2021 |
Ian MacMillan | Update | CDS | SUS simPlant model | I copied c1scx.mdl
to the docker to attach to the plant using
the commands: |
|
16127
|
Fri May 7 11:54:02 2021 |
Anchal, Paco | Update | LSC | IMC WFS noise contribution in arm cavity length noise | We today measured the calibration factors
for XARM_OUT and YARM_OUT in nm/cts and replotted
our results from 16117 |
  |
16129
|
Mon May 10 18:19:12 2021 |
Anchal, Paco | Update | LSC | IMC WFS noise contribution in arm cavity length noise, Corrections | A few corrections to last analysis:
The first plot was not
IMC frequency noise but actually MC_F noise |
|