ID Date Author Type Categoryup Subject Text Attachments
  15446   Wed Jul 1 18:03:04 2020 JonConfigurationVACUPS replacements​I looked into how the new UPS devices
suggested by Chub would communicate with
the vac interlocks. There are several possible
  
  15465   Thu Jul 9 18:00:35 2020 JonConfigurationVACUPS replacementsChub has placed the order for two new UPS
units (115V for TP2/3 and a 220V version
for TP1).
  
  15499   Thu Jul 23 15:58:24 2020 JonSummaryVACVacuum controls refurbishment planThis year we've struggled with vacuum
controls unreliability (e.g., spurious interlock
triggers) caused by decaying hardware. Here
  
  15500   Fri Jul 24 15:40:59 2020 JordanUpdateVACInstallation of two new UPS unitsI installed the Tripp
Lite SMX1000RT2U and Tripp
Lite Smart1000LCD at the
 UPS.jpg 
  15501   Mon Jul 27 15:48:36 2020 JonSummaryVACVacuum parts orderedTo carry out the next steps of the vac
refurbishment plan [ELOG
15499], I've ordered parts necessary
  
  15502   Tue Jul 28 12:22:40 2020 JonUpdateVACVac interlock test today 1:30 pmThis afternoon Jordan is going to carry
out a test of the V4 and V5 hardware
interlocks. To inform the interlock improvement
  
  15504   Tue Jul 28 14:11:14 2020 JonUpdateVACVac interlock test today 1:30 pmThis test has been completed. The IFO configuration
has been reverted to nominal.

For future reference: yes,
  
  15526   Fri Aug 14 10:10:56 2020 JonConfigurationVACVacuum repairs todayThe vac system is going down now for planned
repairs [ELOG
15499]. It will likely take most
  
  15527   Sat Aug 15 02:02:13 2020 JonConfigurationVACVacuum repairs todayVacuum work is completed. The TP2 and TP3
interlocks have been overhauled as proposed
in ELOG
  
  15528   Sat Aug 15 15:12:22 2020 JonConfigurationVACOverhaul of small turbo pump interlocksSummary

Yesterday I completed
the switchover of small turbo pump interlocks
as proposed in ELOG
 small_tp_signal_routing.pngsmall_tp_interlock_signal.pngsmall_tp_signal_routing.pngmedm_screen.png 
  15537   Mon Aug 24 08:13:56 2020 JonUpdateVACUPS installationI'm in the lab this morning to interface
the two new UPS units with the digital controls
system. Will be out by lunchtime. The disruptions
  
  15538   Mon Aug 24 11:25:07 2020 JonUpdateVACUPS installationI'm leaving the lab shortly. We're
not ready to switch over the vac equipment
to the new UPS units yet.
  
  15541   Wed Aug 26 15:48:31 2020 gautamUpdateVACControl screen left open on vacuum workstationI found that the control MEDM screen was
left open on the c1vac workstation. This
should be closed every time you leave the
  
  15556   Fri Sep 4 15:26:55 2020 JonUpdateVACVac system UPS installationThe vac controls are going down now to
pull and test software changes. Will advise
when the work is completed.
  
  15557   Fri Sep 4 21:12:51 2020 JonUpdateVACVac system UPS installationThe vac work is completed. All of the vacuum
equipment is now running on the new 120V
UPS, except for TP1. The 230V TP1 is still
  
  15558   Sat Sep 5 12:01:10 2020 JonUpdateVACVac system UPS installationSummary

Yesterday's UPS
switchover was mostly a success. The new
Tripp
 vac_medm.pngvac_medm.png 
  15577   Wed Sep 16 12:03:07 2020 JonUpdateVACReplacing pressure gaugesAssembled is the list of dead pressure
gauges. Their locations are also circled
in Attachment 1.
 vac_gauges.png 
  15582   Sat Sep 19 18:07:35 2020 KojiUpdateVACTP3 RP failureI came to the campus and Gautam notified
that he just had received the alert from
the vac watchdog.
 vac.pngScreen_Shot_2020-09-19_at_17.52.40.png 
  15586   Sat Sep 19 19:37:16 2020 not KojiUpdateVACTP3 RP failureDisconcerting because those tip seals were
just replaced [15417].
Maybe they were just defective, but if there
  
  15591   Mon Sep 21 15:57:08 2020 JordanUpdateVACTP3 Forepump Replacement and Vac resetI removed the forepump (Varian SH-110)
for TP3 today to see why it had failed over
the weekend. I tested it in the C&B lab
 Screenshot_from_2020-09-21_15-56-04.png20200921_145043.jpg20200921_145040.jpg 
  15599   Wed Sep 23 08:57:18 2020 gautamUpdateVACTP2 running HOTThe interlocks tripped at ~630am local
time. Jordan reported that TP2 was supposedly
running at 52 C (!).
 Screen_Shot_2020-09-23_at_8.55.39_AM.png 
  15600   Wed Sep 23 10:06:52 2020 KojiUpdateVACTP2 running HOTHere is the timeline. This suggests TP2
backing RP failure.

1st line: TP2 foreline pressure
 Screen_Shot_2020-09-23_at_10.00.43.png 
  15602   Wed Sep 23 15:06:54 2020 JordanUpdateVACTP2 Forepump Re-installI removed the forepump to TP2 this morning
after the vacuum failure, and tested in the
C&B lab. I pumped down on a small volume
 Screenshot_from_2020-09-23_15-15-43.pngMainVolPumpDown.png 
  15615   Tue Oct 6 14:35:16 2020 JordanUpdateVACSpare forepumpsI have placed 3 new in box, IDP 7 forepumps
along the x arm of the interferometer. These
are to be used as spares for both the 40m
  
  15668   Tue Nov 10 11:59:37 2020 gautamUpdateVACStuck RV2I've uploaded some more photos here.
I believe the problem is a worn out thread
where the main rotary handle attaches to
 damagedThread.001.jpegIFOstatus.pngP1a_leakTest.png 
  15681   Wed Nov 18 17:51:50 2020 gautamUpdateVACAgilent pressure gauge controller deliveredIt is stored along with the cables that
arrived a few weeks ago, awaiting the gauges
which are now expected next week sometime.
  
  15686   Mon Nov 23 16:33:10 2020 gautamUpdateVACMore vacuum deliveries Five Agilent pressure gauges were delivered
to the 40m. It is stored with the controller
and cables in the office area. This completes
  
  15692   Wed Dec 2 12:27:49 2020 JonUpdateVACReplacing pressure gaugesNow that the new Agilent full-range gauges
(FRGs) have been received, I'm putting
together an installation plan. Since
  
  15698   Thu Dec 3 10:33:00 2020 gautamUpdateVACTrippLite UPS deliveredThe latest greatest UPS has been delivered.
I will move it to near the vacuum rack in
its packaging for storage. It weighs >100lbs
 DFDD4F39-3F8A-439D-888D-7C0CE2E030CF.jpeg 
  15703   Thu Dec 3 14:53:58 2020 JonUpdateVACReplacing pressure gaugesUpdate to the gauge replacement plan (15692),
based on Jordan's walk-through today.
  
  15721   Wed Dec 9 20:14:49 2020 gautamUpdateVACUPS failureSummary:


The (120V) UPS at the vacuum
rack is faulty.
The drypump backing
 rackBeforenAfter.pdfIMG_0008.jpgIMG_0009.jpgvacStatus.png 
  15722   Thu Dec 10 11:07:24 2020 ChubUpdateVACUPS faultIs that a fault code that you can decipher
in the manual, or just a light telling you
nothing but your UPS is dead?
  
  15723   Thu Dec 10 11:17:50 2020 ChubUpdateVACUPS faultI can't find anything in the manual
that describes the nature of the FAULT message. 
In fact, it's not mentioned at all. 
  
  15724   Thu Dec 10 13:05:52 2020 JonUpdateVACUPS failureI've investigated the vacuum controls
failure that occurred last night. Here's
what I believe happened.
 TP2_time_history.pngslow_controls_monitors.png 
  15725   Thu Dec 10 14:29:26 2020 gautamUpdateVACUPS failureI don't buy this story - P2 only briefly
burped around GPStime 1291608000 which is
around 8pm local time, which is when I was
 vacDiag1.png 
  15748   Wed Jan 6 15:28:04 2021 gautamUpdateVACVac rack UPS batteries replaced[chub, gautam]

the replacement was done this afternoon.
The red "Replace Battery" indicator
  
  16047   Mon Apr 19 09:17:51 2021 JordanUpdateVACEmpty N2 TanksWhen I came into the lab this morning,
I noticed that both N2 tanks were empty.
I had swapped one on Friday (4-16-21) before
 N2_Pressure.PNG 
  16064   Wed Apr 21 12:56:00 2021 JordanUpdateVACEmpty N2 TanksInstalled T2 today, and leaked checked
the entire line. No issues found. It could
have been a bad valve on the tank itself.
 Screenshot_2021-04-21_12-53-26.png 
  16090   Wed Apr 28 11:31:40 2021 JonUpdateVACEmpty N2 TanksI checked out what happened on c1vac. There
are actually two independent monitoring codes
running:
  
  16305   Wed Sep 1 14:16:21 2021 JordanUpdateVACEmpty N2 TanksThe right N2 tank had a bad/loose valve
and did not fully open. This morning the
left tank was just about empty and the right
  
  16316   Wed Sep 8 18:00:01 2021 KojiUpdateVACcronjobs & N2 pressure alertIn the weekly meeting, Jordan pointed out
that we didn't receive the alert for
the low N2 pressure.
  
  16404   Thu Oct 14 18:30:23 2021 KojiSummaryVACFlange/Cable Stand ConfigurationFlange Configuration for BHD

We will need total 5 new
cable stands. So Qty.6 is the number to be
 40m_flange_layout_20211014.pdf 
  16410   Mon Oct 18 10:02:17 2021 KojiUpdateVACVent Started / Completed[Chub, Jordan, Anchal, Koji]

- Checked the main volume is isolated.
- TP1 and TP2 were made isolated from
 Screen_Shot_2021-10-18_at_14.52.34.pngScreenshot_2021-10-18_15-08-59.png 
  16412   Tue Oct 19 10:59:09 2021 KojiUpdateVACVent Started / Completed[Chub, Jordan, Yehonathan, Anchal,
Koji]

North door of the BS chamber opened
  
  16413   Tue Oct 19 11:30:39 2021 KojiUpdateVACHow to vent TP1I learned that TP1 was vented through the
RGA room in the past. This can be done by
opening VM2 and a manual valve ("needle
 Screen_Shot_2021-10-18_at_14.52.34.png 
  16418   Wed Oct 20 15:58:27 2021 KojiUpdateVACHow to vent TP1Probably the hard disk of c0rga is dead.
I'll follow up in this elog later today.

Looking at the log in /opt/rtcds/caltech/c1/scripts/RGA/logs
  
  16490   Mon Dec 6 14:26:52 2021 KojiUpdateVACPumping down the RGA sectionJordan reported that the RGA section needs
to be pumped down to allow the analyzer to
run at sufficiently low pressure (P<1e-4
  
  16493   Tue Dec 7 13:12:50 2021 KojiUpdateVACPumping down the RGA sectionSo that Jordan can run the RGA scan this
afternoon, I ran TP3 and started pumping
down the RGA section.
  
  16494   Wed Dec 8 10:14:43 2021 JordanUpdateVACPumping down the RGA sectionAfter an overnight pumpdown/RGA warm up,
I took a 100 amu scan of the RGA volume and
subsequent pumping line. Attached is a screenshot
 40m_RGAVolume_12_8_21.PNG40m_RGAVolume_Torr_12_8_21.PNG 
  16501   Fri Dec 10 19:22:01 2021 KojiUpdateVACPumping down the RGA sectionThe scan result was ~x10 higher than the
previously reported scan on 2020/9/15 (https://nodus.ligo.caltech.edu:8081/40m/15570),
  
ELOG V3.1.3-