Entry  Thu Jun 21 09:01:18 2012, steve, Update, STACIS, the resurrection of STACIS -surf IMG_1364.JPG
There is a small wood cabinet under the south end flow bench, labeled STACIS.
Unit is complete with extension cards and cables.
Entry  Thu Jun 21 09:10:08 2012, steve, Update, PEM, flow bench must be running all times 
The south end flow bench HEPA filter should be run all times. You can turn it off for a measurement or two but remember we are storing clean optics there.
The zero count bench will reach  room particle count  ~ 10,000 in one minute.
Entry  Thu Jun 21 15:00:55 2012, steve, Update, SAFETY, SURFs 2012 safety training surfs2012.JPG
Masha, Eric, Yaakov, Liz and Sasha received 40m specific basic safety training.
Entry  Sat Jun 23 18:44:15 2012, steve, Update, General, power surge has no effect on the lab 
I was notified by CIT Utilities that there was a power surge or short power outage this after noon.
Lab conditions are normal:  c1ioo is down.  The south arm AC was off......I turned it back on.
    Reply  Mon Jun 25 08:21:40 2012, steve, Update, General, AC power disturbance on Sat powerdisturbed.png


    
        
            Quote:
        
        
            

           I was notified by CIT Utilities that there was a power surge or short power outage this after noon.
    Reply  Fri Jun 29 11:02:00 2012, steve, Update, VAC, the vent is at 500 torr sussum.pngpzt.png


    
        
            Quote:
        
    


 Steve, Yuta and Jamie
    Reply  Fri Jun 29 13:11:31 2012, steve, Update, VAC, 40m IFO at atm vent06292012.png40m@ATM.png
 The 4 hrs vent plot at 3 torr/min rate.
Nitrogen was used from 1e-6 torr to 35 torr  at intake pressure 14 PSI
The rest was filled with 5 cylinders of Instrument Grade Air at intake pressure 14 PSI
    Reply  Sat Jun 30 18:31:38 2012, steve, Update, IOO, input telescope beam clipping on Faraday 
  We could set up a simple pick  off after the Faraday  and bring it  out the north window of IOO chamber. No monitor needed,
just take the cover off when you want to see it.
Most people have no idea how to get the MC through the F
    Reply  Tue Jul 3 17:23:50 2012, steve, Update, VAC, pumpdown completed pd72.pngpd72vacnormal.pngpd72at9hrs.png
Vacuum Normal State is reached in 9 hours.  CC1 =  2e-5 Torr
Aux dry pump #3 is still running.  The RGA is not pumped yet.
Entry  Fri Jul 6 08:38:04 2012, steve, Update, Computers, time out timingblanks.png
Why do we have these timing blanks?
    Reply  Fri Jul 6 09:27:40 2012, steve, Update, VAC, 4 days at atm vent_moves.png
 I'm looking for some movement indicators of the vent-pump down events.
 
    Reply  Fri Jul 6 09:52:35 2012, steve, Update, VAC, pump down hick up d3wFireworks.pngpdhickup.png
 I can not understand what really happened here with CC1 gauge or there was really a pressure glitch.
 
attachment 1,  pump down day 3 with 4th of July fireworks in the lab
Entry  Wed Jul 11 11:18:21 2012, steve, Update, PEM, more seismic noise next week 
 
   The fabricators of the big flume in the CES lab have begun  testing the sediment feed system which is the noisiest
component and  plan to test off and on during the day for the next week.
   Please let me  know if you detect the noise or have any
issues.
 
 
Brian Fuller
phone: 626-395-2465
Entry  Thu Jul 12 16:27:04 2012, steve, Update, VAC, drypump changed 
TP2  dry_pump was changed at intake pressure 982 mTorr at 84,889 hrs  This seal hold up for one year.
The rebuilt  pump seal performing well at 28 mTorr
Entry  Fri Jul 27 14:28:36 2012, steve, Update, VAC, cold cathode gauges cc_gauges.png
Our cold cathode 423 gauges are 10 years old. They get insulated by age and show no ionization
current.  We should replace them at the next vent. I'm buying 4 at $317ea
 
Entry  Mon Jul 30 09:00:18 2012, steve, Update, PSL, PMC is still not fixed pmcA.pngpmcB.png
The PMC was locking right the way, but it's transmission would not go up.  Finally I get it back up by moving the "sticky" DC Gain slider
up and down a few times.
    Reply  Mon Jul 30 15:49:16 2012, steve, Update, VAC, cc1 switched cc1changedfVtH.png


    
        
            Quote:
        
        
            

           Our cold cathode 423 gauges are 10 years old. They get insulated by age and show no
    Reply  Tue Jul 31 08:01:22 2012, steve, Update, VAC, cc1 switched cc1h1d.png


    
        
            Quote:
        
        
            

           
            
                
           
Entry  Wed Aug 1 16:38:46 2012, steve, Update, SUS, spare SRMU 03 optic moved IMG_1464.JPGIMG_1465.JPGIMG_1466.JPG
SOS optics prepared to be hanged are moved from the South Flow Bench to S15 Clean Cabinet.
SRMU 03 (1-25-2010) specification summery E080460-05-D,  older vintage SRM 01 and PRM 02 (need more specification)
There was one  NOT MARKED SOS with two broken magnets on its face. This is labeled ???
    Reply  Mon Aug 6 07:59:33 2012, steve, Update, SUS, PRM oplevs servo is still bad PRMeq.png


    
        
            Quote:
        
        
            

           
            
                
           
Entry  Mon Aug 6 09:46:31 2012, steve, Update, IOO, poweroutage turns laser off 

. Power outage turned off the PSL Innolight laser on Sunday afternoon.  It  was turned back on and 
locked happily right on. The green lasers were not effected.
    Reply  Mon Aug 6 10:53:32 2012, steve, Update, VAC, status: vauum normal CCgauges.png
 
 The power outage did not have any effect on the vacuum system.  I had to open VM1 to the RGA because of flaky CC1 cold cathode gauge
fluctuations.
Entry  Mon Aug 6 16:48:43 2012, steve, Update, SUS, SRM oplev is clipping 
The SRM oplev beam is clipping.
Entry  Tue Aug 7 16:11:23 2012, steve, Update, SUS, optical table box wall proposal IMG_1479.JPGIMG_1478.JPG
 Sanwiched wall as shown: 1" clear acrylic, 2 layers of 0.004" thick "window tint", 1 layer of  0.007" thermashield
and  0.125" yellow acrylic
Visibility: 70 %,    Transmission of 1064 nm  2-3 % at 0-50 degrees incident  power density 0.7 W/mm2
Entry  Sun Aug 12 01:21:26 2012, steve, Update, SAFETY, exit door left unlocked 
Caltech security called me at 1am Sunday.  Control room emergency exit door was
found unlocked!
It is your responsibility to lock door if you unlocked it !
Entry  Mon Aug 13 15:31:09 2012, steve, Update, General, larger optical tables at the ends ETMYtable.jpgETMY4X3.jpg
I'm proposing larger optical tables at the ends to avoid the existing overcrowding. This
would allow the initial pointing and optical level beams to set up correctly.
The existing table is 4 x 2 would be replaced by 4' x 3'   We would lose only ~3" space  toward exist door.
    Reply  Wed Aug 15 16:01:47 2012, steve, Update, General, larger optical tables at the ends ? bottleneck.jpg
The drawing of the 4' x 2'  table cover can be seen at entry  #6190 The new proposed
wall #7106  The yellow acrylic would be ~ 0.25" thick and it will be the inside. It
is not shown on the drawing.
    Reply  Wed Aug 22 11:41:20 2012, steve, Update, General, grout plate for optical table at the ends IMG_1561.JPG08231201.PDF
Reinforced concrete grout plate for existing carbon steel stands at the ends.
 
    Reply  Mon Oct 1 17:03:01 2012, steve, Update, SUS, ETMY oplev relayed IMG_1673oplevETMY.jpg


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Fri Oct 12 16:59:26 2012, steve, Update, SUS, TT PITCH damping 


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Mon Oct 15 08:08:33 2012, steve, Update, SUS, PRM damping restored 


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Thu Nov 1 08:48:42 2012, steve, Update, Alignment, progress, then setback - trend of BS table shift alignmentstrugle2d.pngalignmentstrugle2da.pngbefpd09252012.png


    
        
            Quote:
        
        
            

             But these jumps in the OSEMs are all at the level of 10-20 microns. Seems like that wouldn't be enough to account for anything;
    Reply  Fri Dec 14 09:05:04 2012, steve, Update, PEM, TT after eq 6.1 eq6.1.png
 
 SRM, ITMY and MC3 moved some what, but how are the TT? It may worth looking at them before we vent.
    Reply  Wed Mar 6 11:33:09 2013, steve, Update, SAFETY, safety audit 2013 IMG_0010.JPG
 
                                         
       Recommended correction list:
    Reply  Wed Mar 13 16:04:24 2013, steve, Update, SAFETY, laser safety glasses checked safetyglasseschecked.jpg
 All safety glasses were cleaned in soapy water by Bob. I measured their transmission at 1064 nm, 150 mW,  beam diameter 1.5 mm  They
are in working order, no transmission.
 Inventory: 
    Reply  Tue Mar 19 10:30:00 2013, steve, Update, SAFETY, crane inspection 2013 sippingOIL.jpg03191301.PDF
 
Professional crane inspector: Fred Goodbar found two small leaks at the Vertex trolley as he was conducting the annual inspection of the 40m cranes. 
Otherwise the cranes are in safe condition.
    Reply  Fri May 31 10:25:17 2013, steve, Update, safety, first aid kits refilled refilledFAidK.jpg


    
        
            Quote:
        
        
            

            
    Reply  Thu Jun 6 16:33:19 2013, steve, Update, safety, updated laser inventory ELSOconnected.jpg
 Updated laser inventory and operator list. They are posted in the 40m wiki and entry doors of the 40m IFO room.
Let me know if this list needs correction.
Lightwave  M126N-1064-700 NPRO sn 337 in the PSL enclosure got connected to local emergency shut off switch. This is a LIGO operational safety
    Reply  Fri Aug 30 08:08:29 2013, steve, Update, safety, fire horns-flashers tested OK 
 
 We had fire alarm tests  and evacuation drills at 1:30pm yesterday. All flashers and horns are functioning unbearably loud and bright
including clean assembly room.
    Reply  Wed Feb 19 08:07:01 2014, steve, Update, safety, annual laser safety glasses check  AA2014checked.jpg2014check.jpg
 
 All 40m laser safety glasses are cleaned and measured this morning.  Bring
your own safety glasses if you have to enter the 40m IFO room.
    Reply  Tue Feb 25 16:54:57 2014, steve, Update, safety, safety audit 2014 
 
 We had our annual safety inspection today.  Our SOPs are outdated. The full list of needed
correction will be posted tomorrow.
    Reply  Thu Feb 27 13:11:13 2014, steve, Update, safety, safety audit correction noDaisyChaining.jpgZbThumbnail.info


    
        
            Quote:
        
        
            

            
Entry  Thu Mar 6 11:43:17 2014, steve, Update, General, Guralp control unit GuralpCU.jpgBAH.jpg
CMG-40T handheld controller unit is missing its power supply. In order to zero the instrument one has to apply plus and minus DC voltage.
The wiring on this 10 pin Amphenol PT02E-12-10P is shown.
    Reply  Fri Mar 7 16:56:17 2014, steve, Update, IOO, IOO qpds centered IOOqpdCentered.png


    
        
            Quote:
        
        
            

           
            
                
           
    Reply  Mon Mar 10 09:28:48 2014, steve, Update, IOO, IOO pointing 2 days IOOpointing2d.pngmorningseconds.png
 Morning seconds without adjustment.
 
    Reply  Thu Mar 13 16:05:48 2014, steve, Update, safety, safety audit 2014 T1400198-v1-CIT-LIGO_Labs_Annual_Safety_Audit_2014.pdf


    
        
            Quote:
        
        
            

            
    Reply  Mon Mar 17 10:50:58 2014, steve, Update, safety, 2014 annual crane certification 2014craneCert.jpg


    
        
            Quote:
        
        
            

           
            
                
           
Entry  Fri Mar 21 14:59:44 2014, steve, Update, PEM, controling dust IMG_0044.JPG
Please take a look at the table top with the flashlight before
removing it. If it is dusty, wipe it down with dry lint free cloth in the box.
There is one box with flash light and wiper at AP, ETMY & ETMX  optical tables.
    Reply  Mon Mar 24 22:13:37 2014, steve, Update, General, Power Failure poweroutage.png


    
        
            Quote:
        
        
            

           I'm checking the status from home.
    Reply  Tue Mar 25 14:52:57 2014, steve, Update, VAC, vacuum is recovered 17hrsNoPumping.png
 Out gassing plus leak rate   0.15  mTorr / hour
 The pressure rose to 2.5 mTorr in 17 hours
 V1 was opened at 1:56pm