13296
|
Tue Sep 5 17:52:06 2017 |
Kira | Update | PEM | temp sensor update |
to get the sensors to read the same values they have to be in direct thermal contact with the metal block - there can't be any adapter board in-between
for the 2nd attempt, I also recommend encasing it in a metal block rather than just one side. You can drill some 7-10 mm diameter holes in an aluminum or copper block. Then put the sensors in there and plug it up with some thermal paste. |
13303
|
Fri Sep 8 10:22:30 2017 |
Steve | Update | PEM | temp sensor update |
The weight of SS can with copper liner is 12.2 kg
Is 1 Amp for the heating jacket going to be enough? We should have some headroom.
Quote: |
to get the sensors to read the same values they have to be in direct thermal contact with the metal block - there can't be any adapter board in-between
for the 2nd attempt, I also recommend encasing it in a metal block rather than just one side. You can drill some 7-10 mm diameter holes in an aluminum or copper block. Then put the sensors in there and plug it up with some thermal paste.
|
|
13311
|
Tue Sep 12 11:44:16 2017 |
Kira | Update | PEM | temp sensor update |
Got it to work. A cable was broken and the AD586 also broke at the same time so it took a while to find the problem. I had to create a makeshift cable out of three parts so once I replace it for an actual cable, it will be good to go for a test.
Quote: |
Today, I stuck on the sensors to a metal block using a flag, rubber bands, and some thermal paste (1st attachment). I then wrapped the whole thing in about 4 layers of insulation and a lot of tape (2nd attachment). The only things leading out of the box were the three connections to the sensors and a thermometer. I then connected the wires to their respective places on the board of the sensor. To get the readings out we would need to use an ADC. Gautam and I checked to make sure the ADC we have inside the lab goes from -10V to 10V so that it would be able to measure the 3V value the sensor typically measures. We then tried to connect all three sensors to a DC source simultaneously, but unfortunately one of them seems to have disconnected somewhere during the process, as it only showed 1.2V instead of 3V. I plan to fix this tomorrow morning so that we can hopefully set this up soon.
Quote: |
I took off the AD590 and attached it to two long wires leading out from the board. This will allow us to attach the sensor to a metal block and not have to stick the whole board to it. I have also completed three identical copies of this and it's pretty much ready to be tested. According to Craig and Andrew's elog here, the sensor is very noisy and they added in a low pass filter to fix that, so that's something to consider for the final version of the circuit. I'll test what I have so far and see how that goes. We still need to figure out how to get readings from the sensors.
To attach the sensor to the metal block, I'll use some thermal paste and fasteners. I'll also put a thermometer on the block to record the actual temperature. I'll then wrap it in some insulation we have in the lab and have only some wires leading out of it to make measurements. I'll leave this setup overnight and record the outputs for about a full day. The fluctuations between the sensors will then indicate the noise of each individual sensor.
|
|
|
1161
|
Mon Nov 24 19:15:16 2008 |
rana, alberto, john | Configuration | Environment | temperature |
The PSL Room Temperature was alarming because it had gone above 23 C. This set off an unfortunate chain of events:
We found that the PSL HEPA was set low (20%). This is a fine setting for when no one is working in there but it
does raise the temperature since there are heat sources inside the blue box.
We tried to change the office area temperature to compensate and also the westmost sensor inside the lab area by 2 deg F.
The office area one was problematic - there was so much dust in it that the gas valve nipple was clogged. So we've
now blown it all clean with a compressed air can. We're now tuning the calibration screw to make our new
digital sensor agree with the setpoint on the controller.
Expect wild temperature swings of the office area for a couple days while Alberto and I tune the servo. |
1163
|
Tue Nov 25 19:29:15 2008 |
rana, alberto, john | Configuration | Environment | temperature |
Quote: | The PSL Room Temperature was alarming because it had gone above 23 C. This set off an unfortunate chain of events:
We found that the PSL HEPA was set low (20%). This is a fine setting for when no one is working in there but it
does raise the temperature since there are heat sources inside the blue box.
We tried to change the office area temperature to compensate and also the westmost sensor inside the lab area by 2 deg F.
The office area one was problematic - there was so much dust in it that the gas valve nipple was clogged. So we've
now blown it all clean with a compressed air can. We're now tuning the calibration screw to make our new
digital sensor agree with the setpoint on the controller.
Expect wild temperature swings of the office area for a couple days while Alberto and I tune the servo. |
This morning Bob found 92F in the office area and in the control room of the lab. He turned down the thermostat and when I got in at about 9 I found 65. After a few hours of adjustment of the thermostat's calibration, I could stabilize the room temperature to about 72F. I also turned down the thermostats inside the lab of a couple of degrees F. |
1164
|
Thu Nov 27 22:56:42 2008 |
rana | Configuration | Environment | temperature |
8-) |
Attachment 1: mc.png
|
|
13611
|
Tue Feb 6 16:58:19 2018 |
Kira | Update | PEM | temperature measurements |
I decided to plot the temperatures measured over two days for the sensor inside the can and inside the lab just to see if there was any significant difference between the two, and obtained the following plot. This shows that there is a difference in measurements of a few 0.01 C. The insulated seismometer can didn't change temperature as much as the lab did, which is as expected. I'll work on properly calibrating the sensors sometime in the future so that we can use the sensor that's just in the lab as an accurate thermometer. |
Attachment 1: temps.png
|
|
1331
|
Sun Feb 22 23:43:07 2009 |
caryn | Summary | General | temperature sensor |
Comparing PSL-FSS-RMTEMP and PEM-MC1-TEMPS
So, to compare temp channels, I made a plot of PSL-FSS_RMTEMP and PEM-MC1_TEMPS(the test temp sensor channel after converting from cts to degC). This plot begins about 2 months ago t_initial=911805130. The temperature channels look kinda similar but MC1-TEMPS (the temp sensor clamped to MC1,3 chamber) is consistently higher in temperature than FSS_RMTEMP. See compare_temperature_channels.png.
MC1-TEMPS isn't exactly consistent with FSS-RMTEMP. I attached a few plots where I've zoomed in on a few hours or a few days. See compare_temperature_channels_zoom1.pdf & compare_temperature_channels_zoom2.pdf
Change the room temperature, see what happens to the chamber temperature
A while ago, somebody was fiddling around with the room temperature. See compare_temperature_channels_zoom4.pdf. This is a plot of PEM-MC1_TEMPS and PSL-FSS_RMTEMP at t0=911805130. You can see the chamber heating up and cooling down in happy-capacitory-fashion. Although, the PSL-FSS_RMTEMP and the PEM-MC1_TEMPS don't really line up so well. Maybe, the air in the location of the MC1,3 chamber is just warmer than the air in the PSL or maybe there's an offset in my calibration equation.
Calibration equation for PEM-MC1-TEMPS
For the calibration (cts to degC) I used the following equation based on the data-sheet for the LM34 and some measurements of the circuit:
TEMPERATURE[degC]=5/9*(((-CTS/16384/451.9/1.04094)-(.0499*10^-3))/(20*10^-6)-35);
How does the chamber temperature compare with the air temperature?
It looks like the chamber may be warmer than the air around it sometimes.
I wanted to check the temperature of the air and compare it with the temperature the sensor had been measuring. So, at t=918855087 gps, I took the temp sensor off of the mc1-mc3 chamber and let it hang freely, close to the chamber but not touching anything. See compare_temperature_chamber_air.png. MC1_TEMPS increases in temperature when I am handling the temp-sensor and then cools down to below the chamber temperature, close to FSS_RMTEMP, indicating the air temperature was less than the chamber temperature.
When, I reattached temp sensor to the chamber at t=919011131 gps, the the temperature of the chamber was again higher than the temperature of the air. See compare_temperature_air2chamber.pdf.
Also, as one might expect, when the temp-sensor is clamped to the chamber, the temperature varies less, & when it's detached from the chamber, the temperature varies more. See compare_temperature_air_1day.pdf & compare_temperature_chamber_1day.pdf.
New temp-sensor power supply vs old temp-sensor power supply
The new temp-sensor is less noisy and seems to work OK. It's not completely consistent with PSL-FSS_RMTEMP, but neither was the old temp-sensor. And even the air just outside the chamber isn't the same temperature as the chamber. So, the channels shouldn't line up perfectly anyways.
I unplugged the 'old' temp-sensor power supply for a few hours and plugged in the 'new' one, which doesn't have a box but has some capacitors and and 2 more voltage regulators. The MC1_TEMPS channel became less noisy. See noisetime.png & noisefreq.pdf. For that time, the minute trend shows that with the old temp-sensor power supply the temp sensor varies +/-30cts and with the new power supply, it is more like +/-5cts (and Volt/16,384cts * 1degF/10mV --> apprx +/-0.03degF). So, it's less noisy.
I kept the new temp-sensor power supply plugged in for about 8 hours, checking if new temp sensor power supply worked ok. Compared it with PSL-FSS_RMTEMP after applying an approximate calibration equation. See ver2_mc1_rmtemp_8hr_appxcal.png.
Just for kicks
Measuring time constant of temp sensor when detached from chamber. At 918858981, I heated up the temp sensor on of the mc1-mc3 chamber with my hand. Took hand off sensor at 918859253 and let it cool down to the room temperature. See temperature_sensor_tau.pdf. |
Attachment 1: compare_temperature_channels.png
|
|
Attachment 2: compare_temperature_channels_zoom1.pdf
|
|
Attachment 3: compare_temperature_channels_zoom2.pdf
|
|
Attachment 4: compare_temperature_channels_zoom4.pdf
|
|
Attachment 5: compare_temperature_chamber_air.png
|
|
Attachment 6: compare_temperature_air2chamber.pdf
|
|
Attachment 7: compare_temperature_air_1day.pdf
|
|
Attachment 8: compare_temperature_chamber_1day.pdf
|
|
Attachment 9: noisetime.pdf
|
|
Attachment 10: noisefreq.pdf
|
|
Attachment 11: ver2_mc1_rmtemp_8hr_appxcal.pdf
|
|
Attachment 12: temperature_sensor_tau.pdf
|
|
13183
|
Thu Aug 10 14:13:23 2017 |
Kira | Summary | PEM | temperature sensor |
Goal is to build a temperature sensor accurate to 1 mK. Schematic is shown below. This does not take into account the DC gain that occurs.
Parts that would be used for this: LM317 regulator, AD592 temperature transducer, OP amp (low input noise and high impedance), 100K (or maybe 10k) resistor. This is what is currently proposed, but the exact parts we use could be changed to better fit the sensor. The resistor and the OP amp will be decided depending on the output of the AD592.
Once this is built, I would like to create a few copies of it and put them into an insulated container and measure the output from each one. This would allow us to calculate the temperature noise of the circuit, as we can take out the variations due to temperature changes inside the container by comparing the outputs.
I can also model the noise in the circuit to see how much noise there is before building it. There are three terms to the noise that we have, and we need to decide which one dominates at low frequencies.
Our final goal is to create an additional circuit that could cancel out the DC gain. I have attached an additional schematic proposed by Rana that would help with this issue. I will leave this second half for when the first part works. |
Attachment 1: IMG_20170810_121637~2.jpg
|
|
Attachment 2: IMG_20170810_134422~2.jpg
|
|
13188
|
Thu Aug 10 21:22:06 2017 |
rana | Summary | PEM | temperature sensor |
- Should we use the AD590 or the AD592? They're sort of the same, but have slightly different packages and specs.
- Given the sensitivity of the AD590 to power supply drift, I would recommend using a precision voltage reference like the AD581 or AD587. Take a look at the datasheets and order a few different varieties so we can see what works best for us. I believe that the voltage regulators have too much drift to use for precision temperature control.
- The AD590 datasheet has a few example circuits showing how we can subtract off the offset which comes the 1 uA/K coefficient of the AD590 (i.e. we would have 295 uA at room temperature, trying to stabilize to +/- 0.1 K)
- For the first prototypes its fine to use some solderable protoboard assembly, but we will eventually have to figure out how to package this thing and interface it with our Acromag slow controls system.
also, I've attached some temperature noise spectra from the LISA group at the AEI in Hannover. It will be interesting to see if we get the same results. |
Attachment 1: tempsensnoise2.pdf
|
|
Attachment 2: tempnoise_final.pdf
|
|
13190
|
Fri Aug 11 10:27:49 2017 |
Kira | Update | PEM | temperature sensor |
Since there seems to be little difference between AD590 and AD592, I guess we could just go with the AD590. The temperature noise spectrum in the first graph are for the AD590, so if we want to reproduce those results, we should use AD590.
For the AD581/AD587, we could go with a few varieties that have the least output voltage drift, although I am not sure what precision we will need. So maybe we could try AD587U and AD581L. We could also try AD587K and AD581K and see if those work as well.
We will also need to calibrate the sensor, as it takes an input of 5V, but the AD581/AD587 provides 10V, which will give about a 1 degree error according to the datasheet. It does state that this is only a calibration error, so it shouldn't be too much of an issue.
I will figure out the packaging once I construct the sensor and verify that it works. Maybe we could use a box similar to the existing sensor, but it depends on the size of the finished circuit. |
13191
|
Fri Aug 11 10:48:39 2017 |
Kira | Update | PEM | temperature sensor |
Quick update: we actually have AD587KRZ and AD592, so we could start by using that and seeing how it works. |
13193
|
Fri Aug 11 11:56:02 2017 |
rana | Update | PEM | temperature sensor |
Might as well order several of a few different varieties today. Its good to have some extra in stock; we don't always want to have to wait days for parts to show up. If you give Steve a list of parts to buy he can order them today or Monday.
There should also be some precision 5V sources (e.g. AD586) that you can try. |
13194
|
Fri Aug 11 12:27:25 2017 |
Kira | Update | PEM | temperature sensor |
Used AD592CNZ and AD586 (5V output) to create a circuit that works and is responsive to temperature changes. At room temp, using ~1K resistor, it showed ~0.3V across it, as expected. The voltage went up when we heated it with a heating gun. Next step will be to add in an OP amp and design some experiments to check to see how accurate it is. Thanks to Gautam for helping me with it!
I have attached the working circuit and a close up of the connections. |
Attachment 1: IMG_20170811_121608.jpg
|
|
Attachment 2: IMG_20170811_121619.jpg
|
|
13202
|
Mon Aug 14 09:49:18 2017 |
Kira | Update | PEM | temperature sensor |
Decided to try adding in an OP amp just to see if it would work. Added LT1012 and a 100k resistor to the circuit (I originally wanted to do AD743 as it seems to be the best choice according to Zach's elog here, but it said that they are very precious so I went with LT1012 for testing purposes). When heating it with a heating gun, the output voltage went down by a few 0.01V. The maximum voltage was 0.686V. Similar thing happened when I switched to a 10k resistor, where the maximum was 0.705V and it also went down by a few 0.01V upon heating.
I've attached a few pictures showing the circuit. |
Attachment 1: IMG_20170814_092452.jpg
|
|
Attachment 2: IMG_20170814_092513.jpg
|
|
13203
|
Mon Aug 14 12:52:33 2017 |
Kira | Update | PEM | temperature sensor |
I didn't realize that the LT1012 needed an additional input to function. I added in +15V and -15V to pins 7 and 4, respectively and placed a 10k resistor and the numbers make more sense now. The voltage showed a negative value, but it became more negative as I heated it up (it's negative due to how a transimpedance amplifier works).
I have attached the new setup and the value it shows (~-3V). It became more negative by about 0.4V, which translates to about a 40K increase in temperature, which makes sense.
In addition, I have attached an updated sketch of the circuit. I will need to do more testing to determine how accurate this is. The next step would be to calculate how much noise there is currently and figure out how to remove this circuit from the breadboard and use a PCB or something like that for final testing in an insulated container.
The reason I chose AD743 initially for the OP amp is because at low frequencies (which is what we are working with), a FET amp such as AD743 will have a low current noise at high impedance, which is what we have in this case. While a FET amp has high voltage noise compared to other OP amps, the current noise becomes more important at high impedance, so it will work better. According to Zach's graphs, the AD743 is best at high impedances, followed by LT1012.
Quote: |
Decided to try adding in an OP amp just to see if it would work. Added LT1012 and a 100k resistor to the circuit (I originally wanted to do AD743 as it seems to be the best choice according to Zach's elog here, but it said that they are very precious so I went with LT1012 for testing purposes). When heating it with a heating gun, the output voltage went down by a few 0.01V. The maximum voltage was 0.686V. Similar thing happened when I switched to a 10k resistor, where the maximum was 0.705V and it also went down by a few 0.01V upon heating.
I've attached a few pictures showing the circuit.
|
|
Attachment 1: IMG_20170814_121131.jpg
|
|
Attachment 2: IMG_20170814_121139.jpg
|
|
Attachment 3: IMG_20170814_121758~2.jpg
|
|
13210
|
Tue Aug 15 13:32:38 2017 |
Kira | Update | PEM | temperature sensor |
Tested to make sure that even when only the AD586 was heated that there was no change in the reading. I did so by placing the AD586 away from the rest of the circuit and blowing hot air only on it. There was, in fact, no change.
Quote: |
I didn't realize that the LT1012 needed an additional input to function. I added in +15V and -15V to pins 7 and 4, respectively and placed a 10k resistor and the numbers make more sense now. The voltage showed a negative value, but it became more negative as I heated it up (it's negative due to how a transimpedance amplifier works).
I have attached the new setup and the value it shows (~-3V). It became more negative by about 0.4V, which translates to about a 40K increase in temperature, which makes sense.
In addition, I have attached an updated sketch of the circuit. I will need to do more testing to determine how accurate this is. The next step would be to calculate how much noise there is currently and figure out how to remove this circuit from the breadboard and use a PCB or something like that for final testing in an insulated container.
The reason I chose AD743 initially for the OP amp is because at low frequencies (which is what we are working with), a FET amp such as AD743 will have a low current noise at high impedance, which is what we have in this case. While a FET amp has high voltage noise compared to other OP amps, the current noise becomes more important at high impedance, so it will work better. According to Zach's graphs, the AD743 is best at high impedances, followed by LT1012.
Quote: |
Decided to try adding in an OP amp just to see if it would work. Added LT1012 and a 100k resistor to the circuit (I originally wanted to do AD743 as it seems to be the best choice according to Zach's elog here, but it said that they are very precious so I went with LT1012 for testing purposes). When heating it with a heating gun, the output voltage went down by a few 0.01V. The maximum voltage was 0.686V. Similar thing happened when I switched to a 10k resistor, where the maximum was 0.705V and it also went down by a few 0.01V upon heating.
I've attached a few pictures showing the circuit.
|
|
|
1102
|
Thu Oct 30 20:39:47 2008 |
caryn | Configuration | PEM | temperature sensor |
We attached the temperature sensor box to the MC1/MC3 chamber with a C-clamp. We connected the temp sensor to a 2nd box with a short BNC. Bob set up a power cable coming from the X-end towards the MC1/MC3 chamber(Thanks, Bob!) We soldered the end of Bob's power cable to a plug and attached it to the 2nd box (The power supply enters through the 2nd box). A ~20ft BNC cable connects the output signal of the 2nd box to the tall thing by the PSL where all the signals go labeled 1Y2. Once we had everything connected, we put in the fuses for the power supply. So, now the temperature sensor is receiving power. We checked that the power supply was working (we measured +15.08V and -14.95V, and we wanted 15V and -15V so it's OK for now). Tomorrow we will modify C1IOOF.INI file and reboot the frame builder.
About sensor-
There is an LM34 (looks like a transistor) glued w/ epoxy and thermal paste to the inside of a Pomona box ~1"x"1.5"x2". The lid to the box is covered with a 1-2mm thick piece of copper and a little thermal paste is sandwiched between the Pomona lid and the copper piece. A C-clamp attaches the copper piece to the chamber. A BNC is connected to one side of the box (the side with less copper)
About power supply box-
There is a power regulator and an op-amp inside a Pomona box ~2.5"x4"x2". The power regulator is attached to the center of lid of the pomona box with a screw and washer. There's a power plug on the front of the box
Left:+15V:red wire
Center:GND:white wire
Right:-15V:black wire
There are 2 BNC connections on the sides of the box. The left BNC connection is for the output signal and the right BNC connection is for the temperature sensor (if the power plug is coming out of the box towards you).
Sensor location-
Chamber which contains MC1/MC3. On the door facing towards the Y-end. On the bottom-left side. Behind the door. Attached with a C-clamp.
Power supply box location-
Chamber which contains MC1/MC3. On some metal leg thing near the floor facing towards the Y-end. Attached with a zip-tie
Power supply-
Coming from the X-end from a tall thing with all the fuses labeled 1X1
Fuse 160:+15V:red wire
Fuse 171:GND:white wire
Fuse 172:-15V:black wire
Signal-
Going towards the PSL to the tall thing labeled 1Y1 on the rack labeled SN208
ICS-110B
J12 (which we believe corresponds to 50-51 and channel number 13650)
Temperature sensor is connected to J12 with a ~20ft BNC attached to a BNC2LEMO connector we found lying around |
6731
|
Thu May 31 16:19:07 2012 |
yuta | Update | Green Locking | temperature setting for PSL doubling crystal |
I fixed the temperature control of the oven for the PSL doubling crystal.
The PID settings were not good, and also, TC200 was beging DETUNED. So, I activated TUNE function and adjusted PID settings.
I'm not sure what the DETUNE function is for. The manual can be found here;
http://www.thorlabs.com/thorproduct.cfm?partnumber=TC200
Current settings for Thorlabs TC200 are (Red ones are what I changed from the previous setting);
parameters |
Xend |
Yend |
PSL |
TEMP SET (deg C) |
37.5 |
35.7 |
36.9 |
P |
250 |
250 |
250 |
I |
60 |
60 |
200 (was 117) |
D |
25 |
25 |
40 (was 19) |
(DE)TUNE on? |
TUNE |
TUNE |
TUNE (was DETUNE) |
TMAX (deg C) |
200 |
200 |
170 |
PMAX (Watts) |
18 |
18 |
18 |
temperature sensor |
PTC100 |
PTC100 |
PTC100 |
|
11131
|
Wed Mar 11 13:54:18 2015 |
Steve | Update | General | temporaly storage in the east arm |
Green glass for aLIGO OMC shield is temporarly stored in the inside of the Y-arm. |
Attachment 1: greenG.jpg
|
|
Attachment 2: EastArm.jpg
|
|
12174
|
Tue Jun 14 10:13:34 2016 |
Steve | Update | VAC | temporary N2 supply line |
The drill room floor will be retiled Thursday, June 16. Temporary nitrogen line set up will allow emptying the hole area.
Ifo room entry will be through control room.
|
Attachment 1: afterN2work.png
|
|
4113
|
Wed Jan 5 16:11:17 2011 |
kiwamu | Summary | IOO | temporary PZT connection |

This is a connection diagram for the input PZTs (i.e. PZT1 and PZT2).
As drawn in the diagram, the signals don't go through the anti-imaging filter D000186 in the current configuration. |
2
|
Thu Oct 18 14:52:35 2007 |
rana | Routine | ASC | test |
test
X-(:P;(:)) |
12763
|
Fri Jan 27 17:49:41 2017 |
jamie | Update | CDS | test of new daqd code on fb1 |
Just FYI I'm running a test of updated daqd code on fb1.
fb1 has it's own fiber to the daq network switch, so nothing had to be modified to do this test. This *should* not affect anything in the rest of the system, but as we all know these are famous last words.... If something is going haywire, and you can't get in touch with me and can't figure what else to do, you can just log on to fb1 and shut it down. It's not writing any data to any of the network filesystems.
The daqd code under test is from the latest advLigoRTS 3.2.1 tag, which has daqd stability fixes that will hopefully address the problems we were seeing last time I tried this upgrade. We'll see...
I'm going to let it run over the weekend, and will check in periodically. |
12765
|
Fri Jan 27 20:52:36 2017 |
gautam | Update | CDS | test of new daqd code on fb1 |
I'm not sure if this is related, but since today morning, I've noticed that the data concentrator errors have returned. Looking at daqd.log, there is a 1 second timing mismatch error that is being generated. Usually, manually running ntpdate on the front ends fixes this problem, but it did not work today. |
Attachment 1: DCerrors.png
|
|
12769
|
Sat Jan 28 12:05:57 2017 |
jamie | Update | CDS | test of new daqd code on fb1 |
Quote: |
I'm not sure if this is related, but since today morning, I've noticed that the data concentrator errors have returned. Looking at daqd.log, there is a 1 second timing mismatch error that is being generated. Usually, manually running ntpdate on the front ends fixes this problem, but it did not work today.
|
If this problem started before ~4pm on Friday then it's probably unrelated, since I didn't start any of these tests until after that. If unexplained problem persist then we can try shutting of the fb1 daqd and see if that helps. |
3625
|
Thu Sep 30 11:07:20 2010 |
josephb, alex | Update | CDS | test points starting to work |
The centos 5.5 compiled gds code is currently living on rosalba in the /opt/app directory (this is local to Rosalba only). It has not been fully compiled properly yet. It is still missing ezcaread/write/ and so forth. Once we have a fully working code, we'll propagate it to the correct directories on linux1.
So to have a working dtt session with the new front ends, log into rosalba, go to opt/apps/, and source gds-env.bash in /opt/apps (you need to be in bash for this to work, Alex has not made a tcsh environment script yet). This will let get testpoints and be able to make transfer function measurements, for example
Also, to build the latest awgtpman, got to fb, go to /opt/rtcds/caltech/c1/core/advLigoRTS/src/gds, and type make. This has been done and mentioned just as reference.
The awgtpman along with the front end models should startup automatically on reboot of c1sus (courtesy of the /etc/rc.local file). |
2306
|
Fri Nov 20 11:14:22 2009 |
josephb, alex | Configuration | Computers | test points working on megatron and we may have filters with switch outputs built in |
Alex tooked at the channel definitions (can be seen in tpchn_C1.par), and noticed the rmid was 0.
However, we had set in testpoint.par the tst system to C-node1 instead of C-node0. The final number inf that and the rmid need to be equal. We have changed this, and the test points appear to be working now.
However, the confusing part is in the tst model, the gds_node_id is set to 1. Apparently, the model starts counting at 1, while the code starts counting at 0, so when you edit the testpoint.par file by hand, you have to subtract one from whatever you set in the model.
In other news, Alex pointed me at a CDS_PARTS.mdl, filters, "IIR FM with controls". Its a light green module with 2 inputs and 2 outputs. While the 2nd set of input and outputs look like they connect to ground, they should be iterpreted by the RCG to do the right thing (although Alex wasn't positive it works, it worth trying it and seeing if the 2nd output corresponds to a usable filter on/off switch to connect to the binary I/O to control analog DW. However, I'm not sure it has the sophistication to wait for a zero crossing or anything like that - at the moment, it just looks like a simple on/off switch based on what filters are on/off. |
13684
|
Thu Mar 15 17:33:56 2018 |
Kira | Update | PEM | test setup |
I have attached the setup I completed today. The metal box contains the heater circuit and the board for the temperature sensor is right above it. This is basically the same setup as before, but I've just packaged everything up neater. I expect to be able to perform the test tomorrow and begin implementing PID control. I still need a DAC input for the heater circuit and the temperature sensor is having some issues as well. |
Attachment 1: IMG_20180315_172512.jpg
|
|
13691
|
Tue Mar 20 16:56:01 2018 |
Kira | Update | PEM | test setup |
The MOSFET was getting pretty hot, so I switched it out to a larger heat sink and it's fine now. I then used a function generator in place of the DAC to provide ~3.5V. I got the current in the circuit to 1.7A, which is as expected, since we have 24V input, the heater resistance is 12.5ohm and the resistor we are using is 1ohm, so 24V/(12.5+1)ohm = 1.7A. The temperature inside the can rose about 5 degrees in half an hour. The only issue now is the voltage regulators and OP amp inside the box get hot, though it doesn't seem to be dangerous. I switched the function generator input to a DAC and Gautam set it to 1.5V. If it works, then we'll leave this on overnight and work on the PID control tomorrow. I've attached images of the current heater circuit box when it is open and the new heat sink for the MOSFET.
gautam: we also tried incorporating the EPICS channels from the Acromag into the RTCDS so that we can implement PID control by using Foton. I tried doing this using the "EpicsIn" and "EpicsOut" blocks from CDS_PARTS. While the model recompiled smoothly, I saw no signals in the filter module i had connected in series with the EpicsIn block. So I just reverted c1pem to its original state and recompiled the model. Guess we will stick to python script PID reading EPICS channels to implement the PID servo. |
Attachment 1: IMG_20180320_154516.jpg
|
|
Attachment 2: IMG_20180320_145957.jpg
|
|
13692
|
Tue Mar 20 19:48:10 2018 |
gautam | Update | PEM | test setup |
according to the temp sensor readout, which was ~-3.35V which corresponds to ~335K, the temperature of the can is now 60 deg C. This is a bit warm for my liking so i'm turning the heater current down to 0 now by writing 0 to C1:PEM-SEIS_EX_TEMP_CTRL |
13700
|
Fri Mar 23 12:00:20 2018 |
rana | Update | PEM | test setup |
we don't ever want to use our 16 kHz real time system for such low frequency action; its main purpose is for real-time controls, whereas we are OK with multiple seconds of delay in a thermal loop. The Python PID script is sufficient and highly reliable (after years of testing). |
13701
|
Fri Mar 23 12:45:08 2018 |
Kira | Update | PEM | test setup |
I fit the data that we got from the test. The time constant for the cooling came out to be about 4.5 hours. The error is quite large and we should add a low pass filter to the temperature sensor eventually in order to minimize the noise of the measurements. |
Attachment 1: seis_fit.png
|
|
13859
|
Thu May 17 15:38:19 2018 |
Kira | Update | PEM | test setup with seismometer |
I've moved my setup to the actual seismometer. I attached the temperature sensor to the seismometer (attachment 1) with duct tape, though this is temporary. I will be monitoring the temperature fluctuations of the seismometer for a whole day then take the can off and repeat the test. The can isn't clamped down so the insulation isn't perfect, so I'd expect to see some noticeable fluctuations even with the can on. I've also labeled the long cable for the temperatuse sensor readout (attachments 2 and 3). There will also be an out of loop sensor added in later, but for this test since I am not running the loop it doesn't matter which sensor I monitor. Attachment 4 is a picture of the current setup. |
Attachment 1: IMG_20180517_144420.jpg
|
|
Attachment 2: IMG_20180517_145754.jpg
|
|
Attachment 3: IMG_20180517_151956.jpg
|
|
Attachment 4: IMG_20180517_145121.jpg
|
|
13864
|
Fri May 18 14:33:34 2018 |
Kira | Update | PEM | test setup with seismometer |
Here is the result of my test. I think I'll leave the can on over the weekend because there's a long period of time where the seismometer heated up by 0.8 degrees so I can't fully see the fluctuations over a full 24 hour period. |
Attachment 1: seis_temp_can_on.png
|
|
13871
|
Mon May 21 10:15:35 2018 |
gautam | Update | PEM | test setup with seismometer |
I guess it's fine for now while we are still finalizing the setup at EX, but we should eventually line up the seismometer axes with the IFO axes. Is there a photo of the orientation of the seismometer pre heater can tests? If not, probably good to make some sort of markings on the granite slab / seismometer to allow easy lining up of these axes... |
13872
|
Mon May 21 14:17:28 2018 |
Kira | Update | PEM | test setup with seismometer |
I have attached the graph for the seismometer temperature fluctuations over 3 days. As we can see, there is a noticeable fluctuation in daily temperature as well as a difference between days in the maximum and minimum temperatures. I will repeat this test but take the can off to see if there's any difference between having the can on or off. |
Attachment 1: seis-temp.png
|
|
13877
|
Tue May 22 14:49:03 2018 |
Kira | Update | PEM | test setup with seismometer |
It appears that one of the wires was disconnected overnight or this morning so I wasn't able to gather data over a full 24 hour period. Perhaps someone accidentally kicked it. I placed some cones in that area so hopefully the wires won't be in the way as much and I can get the data tomorrow. From the data I do have it seems that the seismometer is at a colder temperature when the can is not on, though it is difficult to see by how many degrees the temperature fluctuates. I've included the data from 5 days back to see the comparison. |
Attachment 1: seis-temp-2.png
|
|
13882
|
Wed May 23 14:50:33 2018 |
Kira | Update | PEM | test setup with seismometer |
This time the test went without issue. The first attachment is the data for the past 24 hours and the second attachment is the full data over 6 days. The average temperature fluctuations (from highest point to lowest point) for the can on was 0.43 C and for the can off it came out to 0.55 C. In addition the seismometer with the can off is about 1 C cooler than with the can on. I'd like to leave the can off until the end of the week so we can get a comparable data set for both the can on and off. Eventually I'll need to figure out a way to clamp the can down to the block in order to get better insulation and hopefully get even smaller temperature fluctuations. |
Attachment 1: seis-temp-3.png
|
|
Attachment 2: seis-temp-full-2.png
|
|
3604
|
Fri Sep 24 00:56:35 2010 |
koji, tara | Update | Electronics | testing TTFSS |
We found that a transistor was broken from yesterday spark too. We partially fixed TTFSS, and it should be enough for testing purpose.
From yesterday test, we found that the RF amplifier for LO signal was broken. There was no spare at the electronic shop at Downs,
so we shorted the circuit for now. Another part which was broken too was a transistor, Q3 PZT2222A, on D0901846.
It was removed and two connections, which are for Q3's 1 and 3 legs, are shorted. Now the voltages out from the regulators are back to normal.
We are checking a MAX333A switch, U6A on D0901894. it seems that the voltage that controls the switch disappears.
There might be a bad connection somewhere. This will be investigated next. |
3607
|
Fri Sep 24 23:47:10 2010 |
koji, tara | Update | Electronics | testing TTFSS |
Q3, a PZT2222A transistor, on D0901846 is replaced by a GE-82. However, the board is still not fully function.
Since Q3, PZT2222A, was broken, I went to Wilson house and got some SP3904's for replacement. But somehow, I broke it during
installation, and did not notice it, and resumed the test. When I got to test 8 on the list, the TTFSS did not work as specified.
Koji checked and found out that -15V, Nref, Vref voltages output did not work correctly. So the SP3904 I installed was removed
and replaced with another SP3904 by Koji, and Vref is working.
Q4 transistor is broken as well and it was replaced by GE 82.
Q1 might be broken too since -15V out is not working.
I'll go to Wilson house to get more transistors next week.
After the broken parts have been replaced, I have to make sure that I separate the power supply board from the rest of the circuit and
check if all V outputs are working, then reconnect the board and check if the current input is reasonable before resume the test.
I hope the wrong input voltage problem today wouldn't damage anything else.
|
3608
|
Sat Sep 25 19:01:13 2010 |
Koji | Update | Electronics | testing TTFSS |
How much current do you need for each voltages?
GE-82 was the only PNP transister I could find in the lab. It's too old but we just like to confirm any other components are still functioning.
Similarly, we can confirm the functionality of the other components by skipping those current boost transisters,
if we don't need more than 30mA.
|
5324
|
Tue Aug 30 11:42:29 2011 |
jamie | Update | CDS | testpoint.par file found to be completely empty |
The testpoint.par file, located at /opt/rtcds/caltech/c1/target/gds/param/testpoint.par, which tells GDS processes where to find the various awgtpman processes, was completely empty. The file was there but was just 0 bytes. Apparently the awgtpman processes themselves also consult this file when starting, which means that none of the awgtpman processes would start.
This file is manipulated in the "install-daq-%" target in the RCG Makefile, ultimately being written with output from the src/epics/util/updateTestpointPar.pl script, which creates a stanza for each front-end model. Rebuilding and installing all of the models properly regenerated this file.
I have no idea what would cause this file to get truncated, but apparently this is not the first time: elog #3999. I'm submitting a bug report with CDS.
|
6844
|
Thu Jun 21 09:01:18 2012 |
steve | Update | STACIS | the resurrection of STACIS -surf |
There is a small wood cabinet under the south end flow bench, labeled STACIS.
Unit is complete with extension cards and cables. |
Attachment 1: IMG_1364.JPG
|
|
3697
|
Tue Oct 12 15:51:18 2010 |
yuta | Bureaucracy | SAFETY | the 40m squad received safety training from Peter |
Yuta, Joonho, and Suresh received the Basic Laser Safety Training from Peter King today.
Now, we got homework. |
659
|
Fri Jul 11 09:29:02 2008 |
steve | Update | PSL | the PMC still hangs up |
Morning alarms:
PMC high voltage is railing, it's auto locker is frosen
I have to move DC Output Adjust slider by hand and it locks immediately
PSL_ISS is saturating. Saturation goes away when PMC is locked.
Laser chiller water is overflowing again: removed 450 cc water |
Attachment 1: pmc4d.jpg
|
|
3569
|
Mon Sep 13 21:52:53 2010 |
kiwamu | Update | Green Locking | the X end laser is ON |
I turned ON the laser at the X end station, which had been OFF for several weeks because of the crane business.
Now the green beam hits the ITMX and I got a reflection back to the end table.
This green beam will be a nice reference when we install the green periscope in the chamber.
If it's necessary, feel free to correct the alignment of the green beam during my absence. |
12453
|
Tue Aug 30 09:51:37 2016 |
Steve | Update | SUS | the chamber HEPA tents are back |
How to minimize particles entering the vacuum envelope.
Just the way it was in August 2011 vent and before.
The portable HEPAs were set up at ETMY and ITMY with CP STAT 100 curtains.
The 40m particles on the floor at ITMY 3000-5000 counts of 0.5 micron cf / min and 0.3 micron size particles are 55,000 - 65,000 counts cf / min
At this condition the MET One Counter #3 on the floor inside the tent goes to zero count of 0.5 micron and 20-40 counts cf / min for 0.3 micron when the tent is slightly overpressured.
|
Attachment 1: ETMYtentHEPA.jpg
|
|
Attachment 2: vented_for_60_days.png
|
|
12458
|
Wed Aug 31 23:01:02 2016 |
gautam | Update | SUS | the chamber HEPA tents are back |
Some more numbers we found while working in/around the chamber today:
Particle counts
Area |
0.3 micron count (per cfm) |
0.5 micron count (per cfm) |
Vacuum chamber (vented) ITMY |
~5000 |
~600 |
HEPA enclosure on ITMY door |
<100 |
0 |
Lab |
~100,000 |
~6000 |
These numbers were measured using our particle counter, which has a pump rate of 0.1 cfm, so the numbers above are 10x the numbers shown on the instrument after a measurement to account for this.
Essentially, the chamber is pretty dirty. Peeling the F.C with hard to reach optics like the ITM installed in place is not really feasible, and after peeling the F.C, we are looking at a best case of an additional 1-2 weeks in air to align the IFO, during which the optic is apparently exposed to quite a lot of particulates. In fact, with the high intensity flashlight left on, I actually saw some flecks of dust occassionally floating around inside the chamber while I was working on the optic. But this is just something we have to accept I guess. |
4092
|
Thu Dec 23 08:54:32 2010 |
Suresh | Update | elog | the delhi univ syndrome |
Sorry folks! I couldnt get to the elog and didnt know that the elog was crashing every time I tried to access it.
But have found other means to access it and the elog is safe for now! |