  14525   Tue Apr 9 00:16:22 2019 ranaUpdateALSEX Green PDH error monitor calibratedG=10 or
G=100?
  
  14526   Tue Apr 9 00:18:19 2019 gautamUpdateALSEX Green PDH error monitor calibratedwrong assumption - i checked the gain just
now, it is G=10, and is running in the "low-noise"
mode, so can only drive 4V. fixed elog, filter.
  
  14527   Tue Apr 9 18:44:00 2019 gautamUpdateALSEX Green PDH discriminant measurementI decided to use the more direct method,
of disconnecting feedback to the EX laser
PZT, and then looking at the cavity flashes. 
 cavityFlashes.pdfmodelPDH.pdf 
  14528   Tue Apr 9 19:07:12 2019 gautamUpdateALSIR ALS noise budgetUpdated the noise budget to include the
unsuppressed frequency noise from the EX
laser. It does not explain the noise
 ALS_noiseBudget.pdfALS_noiseBudget.pdf 
  14533   Thu Apr 11 01:10:05 2019 gautamUpdateALSLarge 2kHz peak (and harmonics) in ALS XThese weren't present last week. The
peaks are present in the EX PDH error monitor
signal, and so are presumably connected with
 EX_PDH_2kNoise.pdf 
  14548   Wed Apr 17 00:50:17 2019 gautamUpdateALSLarge 2kHz peak (and harmonics) in ALS X no moreI looked into this issue today. Initially,
my thinking was that I'd somehow caused
clipping in the beampath somewhere which
 EX_PDHnoise.pdf 
  14549   Wed Apr 17 11:01:49 2019 gautamUpdateALSLarge 2kHz peak (and harmonics) in ALS X no moreEX green stayed locked to XARM length overnight
without a problem. The spectrogram doesn't
show any alarming time varying features around
 EX_PDH_specGram.pdf 
  14643   Wed May 29 18:13:25 2019 gautamUpdateALSFiber beam-splitters are now PMTo maintain PM fibers all the way through
to the photodiode, I had ordered some PM
  
  14645   Fri May 31 15:55:16 2019 gautamUpdateALSPSL + X beat restoredCoupling into the fast axis
of the fiber:

The PM couplers I bought require
  
  14740   Tue Jul 9 18:42:15 2019 gautamUpdateALSEX green doubling oven temperature controller power was disconnectedThere was no green light even though the
EX NPRO was on. I checked the doubling oven
temperature controller and found that its
  
  14800   Mon Jul 22 23:53:16 2019 gautamUpdateALSIR ALS locking attemptSummary:

My goal tonight was to lock the
PSL frequency to be resonant in the
 ALS_X_outOfLoopnoise.pdfALS_X_inLoopnoise.pdfCARM_loopShape.pdf 
  14802   Wed Jul 24 00:22:24 2019 gautamUpdateALSPSL frequency locked to XARM length using ALSSummary:

I succeeded in locking the PSL
frequency to the XARM cavity length, with
 ALS_X_noise_POX.pdfBR_comparison.pdfALS_CARM_OLG.pdf 
  14811   Thu Jul 25 12:25:56 2019 gautamUpdateALSIR ALSX noiseSummary:


There are some broad peaks
in the ALS out-of-loop noise, centered at
 ALS_broadPeaks.pdf 
  14823   Fri Aug 2 11:37:38 2019 gautamUpdateALSEY IR ALS AssaySummary:

I'd like to confirm that the
IR ALS scheme will work for locking. The
 IMG_7780.JPGEy_MM_20190802.pdf 
  14835   Tue Aug 6 23:09:20 2019 gautamUpdateALSEY table work
Removed power monitoring PD (It
was off anyways)
Installed Steering
mirror and collimator in K6XS mount (fast
  
  14838   Fri Aug 9 16:37:39 2019 gautamUpdateALSMore EY table workSummary:


220 uW / 600 uW (~36 %
mode-matching) of IR light coupled into fiber
 IMG_7791.JPGfiberCleaning.pdf 
  14846   Thu Aug 15 18:54:54 2019 gautamUpdateALSALS sensing noise due to IMCSummary:

I came aross an interesting
suggestion by Yutaro that KAGRA's
 ALSsensingNoise.pdf 
  14847   Fri Aug 16 04:24:03 2019 ranaUpdateALSALS sensing noise due to IMCWhat about just use high gain feedback
to MC2 below 20 Hz for the IMC lock? That
would reduce the excess if this theory is
  
  14918   Mon Sep 30 18:20:26 2019 gautamUpdateALSALS OOL noise - a first lookAttachment #1 shows a
first look at the IR ALS noise after my re-coupling
 ALS_OOL_20190930.pdf 
  14938   Fri Oct 4 00:32:24 2019 gautamUpdateALSMore locking updatesSummary:

I managed to achieve a few transitions
of control of the XARM length using the ALS
 ALS_singleArm.pngALS_OOL_20191003.pdf 
  14943   Sat Oct 5 21:26:34 2019 gautamUpdateALSY-end green alignment tweakedSummary:

I improved the alignment of the
green beam into the Y arm cavity.
 ALSY_alignment.pngALSY_OOL.pdf 
  14944   Sun Oct 6 15:23:27 2019 gautamUpdateALSArm control using error signals achievedSummary:

I
 ALSY_20191006.pdftransitionIRALS.pngarms_ALS.pdf 
  14979   Fri Oct 18 20:21:33 2019 shrutiUpdateALSAM measurement attempt at X end[Shruti, Rana]

- At the X end, we set up the network
analyzer to begin measurement of the AM transfer
  
  14981   Mon Oct 21 12:25:46 2019 gautamUpdateALSDFD electronics checkoutSummary:

There are no unexpected red-flags
in the performance of the DFD electronics.
 ALSnoiseAnomaly.pdfDFDcalib.pdfspectra.pdfPTgains.pdf 
  14984   Tue Oct 22 15:32:15 2019 gautamUpdateALSEX uPDH electronics checkoutSummary:

The EX PDH setup had what I thought
was insufficient phase and gain margins.
 uPDH_X_OLTFs.pdf 
  14986   Wed Oct 23 10:23:26 2019 ranaUpdateALSEX uPDH electronics checkoutDuring our EX AM/PM setups, I don't
think we bumped the PDH gain knob (and I
hope that the knob was locked). Possible
  
  14987   Wed Oct 23 11:11:01 2019 gautamUpdateALSEX uPDH electronics checkoutThe closest thing I can think of is here.

  
  14993   Fri Oct 25 01:04:49 2019 gautamUpdateALSALS electronics chain was saturating[Koji, gautam]

Summary:

We think we got to the bottom of
  
  14995   Mon Oct 28 23:20:11 2019 gautamUpdateALSALS power budget 


IR
  
  15020   Thu Nov 7 17:46:10 2019 shrutiUpdateALSAM measurement at X endSome details:

- There
was a SR560+SR785 (not connected for measurement)
placed near the X end which I moved; it is
 AMTF20191107.png 
  15058   Mon Dec 2 00:27:20 2019 gautamUpdateALSGreen ALS resurrectionAttachment
#1 -
 ALSnoiseComparison.pdfALS_TR_RIN.pdfGreemAM.pdf 
  15059   Mon Dec 2 18:20:29 2019 gautamUpdateALSEY uPDH post mixer LPFAs part of characterization, I wanted to
calibrate the EY uPDH error point monitor
into units of Hz. So I thought I'd measure
 EY_uPDH_LPF.pdfEY_uPDH_OLTF.pdf 
  15060   Mon Dec 2 20:18:18 2019 ranaUpdateALSEY uPDH post mixer LPF         
         filter
Q seems too high, 
  
  15063   Tue Dec 3 00:10:15 2019 KojiUpdateALSEY uPDH post mixer LPFI got confused. Why don't we see that
too-high-Q pole in the OLTF? 
  
  15064   Tue Dec 3 00:51:25 2019 gautamUpdateALSEY uPDH post mixer LPFI'm not sure - maybe it was measurement
error on my part, I will double check. Moreover,
the EX
  
  15066   Tue Dec 3 18:15:42 2019 gautamUpdateALSEY uPDH post mixer LPFRana and I discussed this alogrythym a
bit today - here are some bullet points,
I'll work on preparing a notebook. We
  
  15068   Tue Dec 3 21:28:24 2019 gautamUpdateALSEY uPDH post mixer LPFHere are some loop transfer functions.
I basically followed the decomposition of
the end PDH loop as was done in the multi-color
 uPDH.pdf 
  15130   Fri Jan 17 18:02:21 2020 gautamUpdateALSGain blocks packaged and characterizedSummary:


The ZHL-1010+
gain blocks acquired from MiniCircuits
 photos.pdfgain.pdfnoise.pdfmeasSchem.pdfzhl1010Data.zip 
  15153   Fri Jan 24 17:14:01 2020 gautamUpdateALSGain blocks installedJordan will write up the detailed elog
but in summary,


Former +24V Sorensen in
  
  15157   Sun Jan 26 14:40:55 2020 gautamUpdateALSALS OOL noiseIn preparation for resuming IFO locking
activities, I measured the ALS noise with
the arm lengths locked to IR, AUX laser frequencies
 ALSnoise_20200126.pdf 
  15206   Tue Feb 11 16:39:00 2020 shrutiUpdateALSAM/PMThe results of the AM/PM measurements:


Attachment 1: Traces
of 9 AM TFs overlaid on top of each other,
 AM.pdfPM.pdfRatio_all.pdfRatios_FM_PM.pdfRatio_all_new.pdf 
  15209   Thu Feb 13 01:47:39 2020 gautamUpdateALSFast ALS - delay line prepA few years ago, Koji and I setup a delay
line phase shifter, which can be used to
impart a (switchable) delay to a signal path.
 delaySlider.pngdelayLineLosses.pdfdelayLineCal.pdf 
  15211   Thu Feb 13 21:30:55 2020 shrutiUpdateALSALS OOL noise with arms locked[Meenakshi, Gautam, Shruti]

Summary:

- We initially aligned the arm
 20200213_ALS.pdf 
  15212   Fri Feb 14 00:53:50 2020 gautamUpdateALSFast ALS - more setupIn the process of setting up some cabling
at 1Y2, I must've bumped a cable to the
c1lsc expansion chassis. Anyways, the c1lsc
 CMSlow_ALSnoise.pdfOLTFmeas.pdf 
  15213   Fri Feb 14 14:02:13 2020 shrutiUpdateALSALS OOL noise with arms locked[Meenakshi, Shruti]

Even though we were not able to
lock the the IR beat (by enabling LSC) during
 819472_10.pdf225642_10.pdf693500_10.pdf 
  15214   Fri Feb 14 14:52:41 2020 gautamUpdateALSALS OOL noise with arms lockedUnlikely, the alignment was probably just
not good. I restored the alignment and now
the arms can be locked to IR frequency.
  
  15216   Tue Feb 18 18:14:59 2020 shrutiUpdateALSALS OOL noise with arms lockedWe proceeded with the trying to measure
the ALS out-of-loop noise of the X arm when
the X arm cavity is green-locked using different
 ALSNoise.pdf 
  15217   Wed Feb 19 22:20:22 2020 ranaUpdateALSALS OOL noise with arms lockedCould you please put physical units on
the Y-axis and also put labels in the legend
which give a physical description of what
  
  15218   Fri Feb 21 10:59:08 2020 shrutiUpdateALSPDH error signals?   IMG_0082.jpgIMG_0083.jpgIMG_0084.jpgIMG_0085.jpg 
  15219   Fri Feb 21 13:02:53 2020 KojiUpdateALSPDH error signals?Check out this elog:
ELOG
4354
  