ID Date Author Type Category Subject
  491   Thu May 22 11:21:45 2008 steveBureaucracySAFETYearly surf sudent
Caltech under grad Eric Mintun received the 40m safety training.
Now he has to read and sign SOP of laser and ifo.
He'll be working with GigE cameras with Joe
  490   Wed May 21 15:21:33 2008 robUpdateComputer Scripts / Programsautolockers and cron

I added hourly cron jobs to op340m to ensure that

MC autolocker
FSS Slow Servo
PSL watch

are running. I've also edited the wiki procedure to reflect the fact that these no longer need to be restarted by hand.
  489   Tue May 20 18:33:01 2008 Andrey, JohnConfigurationIOOMode Cleaner is locked again

It was noticed by Mr.Adhikari earlier today that the MC became unlocked at about 11AM.

There is no clear understanding what caused the problem.

Trying to restore the modecleaner locking, we noticed with John that the beam was not centered on the wavesensors (WFS1. WFS2 on the screen "C1IOO_LockMC.adl"). We decided to adjust the beam position moving slightly the bias sliders for pitch and yaw degrees of freedom for MC1.
This allowed to make the MC locked.

Old positions for the MC1 sliders: Pitch = 2.9934, Yaw = -0.6168;
New positions --------//---------: Pitch = 3.0604, Yaw = -0.7258.

At the same time, FSS for PSL is still showing the values in the range 0.720 - 0.750 which is lower than the usual values. The indicator for FSS value is yellow when it is below 0.750.
  488   Tue May 20 09:28:42 2008 steveBureaucracySAFETYsafety traning for 40m
Tara Chalernsongsak, new gradstudent for K. Libbrecht was introduced to the basics of the 40m operations.
  487   Mon May 19 11:49:57 2008 steveUpdateSUSetmy oplev laser replaced
The 9 years old Uniphase HeNe was replaced by a new JDSU 1103P
Power output ~3.5 mW, reflected light back on qpd ~140 microW, 12,800 counts
Andrey will get the ~2.5 mm od spot on the qpd smaller by replacing
the f1000 lens
  486   Sun May 18 18:59:15 2008 ranaConfigurationComputerscron and hosts
I added rosalba to the hosts file for the control room machines (131.215.113.103).

I also removed the updateddb cron from our op440m crontab because it was running at 5 PM
even though I had set it to run at 5:57 AM. If it still runs then, it must be because of
another crontab.
  485   Sun May 18 18:44:48 2008 ranaSummarySUSOptical Lever SUM Trend - 80 days
I used the OL-Trend.xml dataviewer template to make this plot. Looks like the ETMY optical lever
slowly degraded over the last few months and then finally died 10 days ago. Would someone please
replace this laser and tune the lens position to minimize the spot size on the quad?
Attachment 1: e.pdf
e.pdf
  484   Sun May 18 18:14:30 2008 ranaSummaryComputer Scripts / Programsautlockers and cron
Today I noticed that the MC was unlocked, the MC autolocker was off, the PSL autolocker was off,
and the PSL Slow loop was off.

Reading through .log files and our elog it seems like things have been this way since Tuesday
when Andrey went around rebooting computers to bring things back.

And it looks like the ETMY optical lever is dead.

I restarted the PSL & MC scripts on op340m. I also locked and aligned the X arm to verify that
not everything is broken. The Y Arm is unalignable until we replace the HeNe.
  483   Fri May 16 17:27:55 2008 AndreyOmnistructureGeneralToilets are broken, do not use them !!!

Both toilets in 40-meter were constantly flushing, the leaking water was on the floor inside of the restrooms, so

BOTH RESTROOMS ARE CLOSED TILL MONDAY


I have heard the constant loud sound of flushing water, opened the door, and was unpleasantly surprised because all the floor was under the layer of water and the toilets were constantly flushing. I called security at X5000, a plumber came in and told that a team of plumbers needs to repair the flushing system after the weekend. The plumber today just shut off the flushing water, wiped off the floor and told not to use the restrooms in the weekend. We should expect a team of plumbers on Monday.

Sinks are working, so you can wash your hands.
  482   Fri May 16 14:38:50 2008 josephbSummaryCamerasTwo cameras setup
I've changed the pickoff setup from yesterday for the GigE cameras to include a 33% beam splitter (first one I could find). The reflection is going to the GC650 (CCD camera) while the transimission is going to the GC750 CMOS camera. This means the CMOS camera has roughly twice the light incident as the GC650 and should be kept in mind in all comparisons. The distances from the beam splitter are approximately the same both cameras, but some more accurate positioning might be useful.

Its very easy to get the GC650 camera into a bad state where you need to go out and cycle the power (simply unplug and re-plug in the power supply either at the camera or outlet). If the ListCamera program doesn't see it, this is probably necessary.

Andrey added at 6.30PM: Actually the 650 camera keeps crashing constantly. Every time I attempt to capture an image, the camera fails.
  481   Thu May 15 16:24:18 2008 josephbSummaryCameras 
The GC750 camera is currently looking at a very small pickoff of the PSL output (transmission of a Y1-1037-45-S mirror). The plan is to take images tomorrow with it and the GC650 from the same spot and do comparisons.

For those interested, the camera can be run with two codes, from mafalda. Use ssh -X mafalda to login, to allow the live stream to work with the SampleViewer code. The codes can be found in:

/cvs/cds/caltech/target/Prosilica/40mCode/Snap

and

/cvs/cds/caltech/target/Prosilica/bin-pc/x86/SampleViewer

Type Snap --help for a list of options for that program. Click the circle looking thing in SampleViewer to start the live stream. Note only 1 of the two programs can be running at a time, and the only way to change settings (such as exposure length) is with Snap at the moment.
  480   Thu May 15 14:39:33 2008 CarynSummaryPEMfiltering mode cleaner with mic
Tried filtering for mode cleaner data(C1:IOO-MC_L) using a siso-firwiener filter and microphone data(C1:PEM-AS_MIC) for noise input. The noise reduction in mode cleaner data using the microphone-filter is comparable to the noise reduction when an accelerometer(C1:PEM-ACC_MC1_X) filter is used. See attached graphs.
Attachment 1: MC_L_with_PEM-AS_MIC_filter.pdf
MC_L_with_PEM-AS_MIC_filter.pdf
Attachment 2: MC_L_with_PEM-ACC_MC1_X_filter.pdf
MC_L_with_PEM-ACC_MC1_X_filter.pdf
  479   Thu May 15 12:05:49 2008 josephbConfigurationPSLPath to PSL Position QPD
The 50/50 beamsplitter that was being used as the last turning mirror to the PSL Position QPD has been replaced with a Y1-1037-45-S plate. This turning mirror was also moved 4" farther along the beam path, so as to produce as small (few microwatts) transmission through the plate. The lensing optics were also shifted so as to maintain a focused beam on the photodiode. Lastly, the rotating ND filter was increased from 1.5 to 2.0 to reduce the incident power on the photodiode, since twice the power is now reaching it.

The small beam on transmission will be used by the digital cameras as a test beam.
  478   Thu May 15 10:40:21 2008 steveHowToGeneralLisa Goggin, PhD
Lisa Goggin successfully defended her thesis on May, 13 2008

"A Search For Gravitational Waves from Perturbed Black Hole Ringdowns in Ligo Data"

She started out as a surf student in the 40m.

Congratulation!
Attachment 1: lisa.JPG
lisa.JPG
  477   Wed May 14 14:05:40 2008 AndreyUpdateComputersComputer Linux-2, MEDM screen "Watchdogs"

Computer "Linux-2", MEDM screen "C1SUS_Watchdogs.adl": there is no indication for ETMY watchdogs, everything is white. There is information on that screen "C1SUS_Watchdogs.adl" about all other systems (MC, ETMX,...), but something is wrong with indicators for ETMY on that particular control computer.
  476   Wed May 14 13:14:19 2008 AndreySummaryComputersReflective Memory Network is restored

Reflective Memory Network is restored, all watchdogs and oplevs are returned to the "enabled" state.

In order to revive the computers, several things were done.

1) Following Mr. Adhikari's elog entry #353, I walked around the interferometer room, and switched off the power keys in all crates with computers whose names are contained in the MEDM Reflective Memory screen, including the rack with the framebuilder. By the way, it was nontrivial to find the switch in the 1Y4 crate that would shut off/on processors "c1susvme1" and "c1susvme2": the switch turned out to be located at the rear side of the crate, and it is not a key but it is a button.

2) I was trying to follow wiki-40 computer restart procedures, but every time that I was trying to run "startup.cmd" screen from the corresponding target subdirectory, I got the error message "Device or resource busy".
By the way, one more thing was learned: if you firstly open in terminal burtgooey, select the snap file, then reboot the processor, and then will try to burt-restore it, you will get the message "Status Not OK". In order to really burt-restore the processor which was recently rebooted, you need to close the terminal with burtgooey and open burtgooey in a new terminal window which should be opened after rebooting the processor.

Feeling that my activities according to wiki-40 procedures do not revive computers, I invited Alex Ivanov.

3) Alex tried to touch the memory card in "c1iovme" in rack 1Y2, because once before this card failed causing network problems, but this did not help.

4) We shutted off and restarted again (pressing the power-switching button) the black Linux machine "c1dcuepics" (located in the very bottom below the framebuilder). Alex says that this machine is responsible for all EPICS. It was not restarted for 182 days, and probably some process there went wrong.

After restarting this machine "c1dcuepics" we were able to follow wiki-40 procedures for restarting all other computers (whose names are on the MEDM RFM network). We ran correcponding "startup.cmd" files and burt-restored them without error messages.

Now all the computers work and communicate in a proper way.

Mr. Joseph Betzwiezer was helping me with all these activities (we decided that it is more important that cameras for now), thanks to him. But our joint skills turned out to be insufficient, so Alex Ivanov's contribution was the most important.
  475   Tue May 13 10:38:28 2008 steveUpdateComputersrfm network is down
The RFM network went down yesterday around 5pm
Only c1susvme1 is alive but it's timing is off.
Andrey is bringing the network up.

Andrey wants to make an addition first that our situation is very much similar to that described by Rana in his elog entry # 353 (March 03). All of the rectangular boxes are red, except for SUS1-c1susvme1 and AWG (only these two rectangles are green).
  474   Tue May 13 10:15:52 2008 steveUpdateSUSrestored damping of BS & PRM
I think our janitor was cleaning too heavy handedly.
The BS and PRM sus damping were lost.
They were restored.
  473   Fri May 9 10:15:36 2008 josephbUpdateComputersNodus has moved
Steve and myself moved Nodus from under the table in the control room, to just above the Rana computer in the control room rack.
  472   Fri May 9 08:40:24 2008 steveUpdateSUSETMY sus damping restored
ETMY lost damping at 19:10 last night.
There was no seismic event than.
Sus damping was restored this morning.
  471   Thu May 8 16:40:36 2008 josephbConfigurationCamerasGige Camera currently on PSL table
Andrey and myself were working on the PSL table today, using a pickoff of a pickoff of the main beam (adding a microscope slide to pickoff ~4% of the original pickoff) to the GC750 GigeCam.

At the time we left, we scanned the area with a beam scan and didn't see any new stray beams, and nothing in any useful beam paths should have changed. We also strung a Cat 6 cable from the control room switch out to the PSL table in the cable trays, and then above the PSL table.

Currently, its not as well aligned as it could be, and also requires a very low exposure setting, of -E 50 or so to avoid saturation.
  470   Thu May 8 02:06:13 2008 ranaSummaryCOCThermal Lensing in the ITMs and BS may be a problem
The iLIGO interferometers start to see thermal lensing effects with ~2W into the MC, a recycling
gain of ~50, and a beam waist on the ITMs of ~3.5 cm.

At the 40m, the laser power into the MC is 1/2 as much, the recycling gain is 4-5x less, but the
beam on the ITM has a 3 mm waist. So the power in the ITM bulk is 10x less but the power density
is 100x more
. Seems like the induced lens in the ITM bulk might be larger and that if there's
significant absorption on the ITM face (remember our Finesse is 4-5x higher) the beam size in the
arm cavity may also change enough to measure.

Someone (like Andrey) should calculate how much the beam sizes change with absorbed power.
  469   Thu May 8 01:50:25 2008 ranaSummaryASCArm Cavity HOM Resonances
Nothing new, but I calculated the frequencies of the first 22 higher order transverse modes and thought I might as well list them here.

To do this I took formula (23) from page 762 of Siegmans book and put it into this form:
         f_fsr
dfmn =   ----- * (m+n) * acos(sqrt(g1*g2))
           pi

and then calculated them from m+n = 1..22 (22 is not a magic number).

I also used the 'mod' function of matlab to calculate the frequency mod FSR so that we would know how far away
from a cavity resonance it is. I took as parameters: Larm = 38.55 m, Ritm = 1e6 m, Retm = 57.1 m. Kirk measured
the arm length some time ago; we need to measure the arm g-factor...maybe we'll put Tobin on this when he comes
by for a visit.

1.1936 (TEM01, TEM10)
2.3871
3.5807
0.8859 (TEM22, TEM13, TEM31)
2.0795
3.2730
0.5782
1.7718
2.9654
0.2706 (TEM55, ...)
1.4641
2.6577
3.8512
1.1564
2.3500
3.5436
0.8488
2.0423
3.2359
0.5411
1.7347
2.9282
  468   Thu May 8 01:07:24 2008 ranaSummaryLSCFrequency Noise test: MC Trans Backscatter
There is a wandering hump in the MC_F spectrum. It can move around on the time
scale of seconds between 40 and 200 Hz. It has an amplitude ~5-50x above the background spectrum. This seems new; I don't remember it
from a year ago. It is there in the IFO unlocked as well as the IFO locked as well as the locked + CM mode.

Tapping the AS table and/or the PSL table enclosures produces a broadband increase in the MC_F spectrum but doesn't
selectively effect the hump.

We thought it might be backscatter from the MC TRANS path and so we stuck in one of Steve's cool black glass V's into
this space. No effect. We should design a black glass V dump which we can replicate in large quantities for us and for
the sites. Something like the one on the LSC PDs, but with a 1 sq. inch opening area and a 2 inch depth.


We have also done this on the MC2 - TRANS beam before. No noise reduced there either.

The noise hump is appearing in MC_F but not in CARM_IN1 (after the CM handoff) so it seems like the MC has enough gain
to squash it. This also exonerates the MC REFL path since anything there would not be effected by the MC servo gain and
so would be visible in CARM.

My best guess is that there is something really, really scattery on the PSL table. But for now it looks like this is not a
big factor in the locking
issues.
  467   Wed May 7 15:25:41 2008 robConfigurationLSCAP33 -> POX33

Quote:

I am in the process of switching the POX166 and AP33 photodetectors, so that they become POX33 and AP166. The IFO_CONFIGURE buttons won't work until I finish.


Done. We're now in the 40m CDD configuration.
  466   Tue May 6 17:28:39 2008 robConfigurationLSCAP33 -> POX33

I am in the process of switching the POX166 and AP33 photodetectors, so that they become POX33 and AP166. The IFO_CONFIGURE buttons won't work until I finish.
  465   Mon May 5 15:53:14 2008 steveUpdateVACdrypump replaced & readbacks are out
The small turbo pump of the annulus system is tp3
It's drypump was replaced at 1.4 Torr after 8 months of operation.
The rebuilding cost of this SH100 drypump is getting ridiculously high $1349.
I'll look for a replacement.

Joseph rebooted c1vac2 card on Apr 24 entry#441
This restarted the readbacks of the turbos and ion pumps for a while.
That was nice. The readbacks are not working again

  464   Mon May 5 11:04:30 2008 robOmnistructureComputersNetwork setup

Mafalda was not connected to the network, and so our DMF-based seisBLRMS has not been running for ~1 week. I traced this to a broken ethernet cable connecting mafalda to the network switch in the rack next to the B&W printer. This cable has a broken connector at the switch side, which means it can't stay connected if there's any tension. It needs to be replaced.
  463   Thu May 1 12:46:02 2008 josephbConfigurationComputersNodus gateway is up
The computer Nodus is now acting as a gateway machine between the GC network and the martian network in the 40m. It has the same passwords as the rana gateway machine.

Its name on the GC side is nodus (ip: 131.215.115.52) and on the martian side is nodus113 (ip: 131.215.113.200). Will need to update the hosts file on the control room machines so you can just use the name nodus113 rather than the full ip.

Software is still being added to the computer, and it will remain in parallel with the rana gateway machine until everything has been working properly for a week or so.
  462   Thu May 1 08:31:51 2008 steveUpdateSUSearthquake trips etmy & mc1
Earthquake at Lake Isabel magnitude 4.4 at 1am today shakes the 40m
ETMY and MC1 watchdogs tripped.
Sus damping restored.
  461   Wed Apr 30 20:48:58 2008 AndreySummaryPEMNew Weather Channels

I created the new channels for the weather station, all letters are capital ones. They are of the form "C1 : PEM-WS_PARAMETER" where "PARAMETER" is temperature, pressure, wind,... characteristics (names are self-obvious).

These new weather channels are indicated on the "Weather Checklist" MEDM screen. Also, units of pressure were changed from Pascal to torr and mbars.

The new weather channels are also visible in Dataviewer. I updated the template, and as an example of Dataviewer data I attach the following 5-hour trends of weather parameters from 3.30PM to 8.30PM on April 30th.
Attachment 1: April30-5hours.png
April30-5hours.png
  460   Tue Apr 29 21:30:49 2008 AndreyUpdatePEMIn the process of renaming channels for Weather Station

I startted renaming channels for the weather station, and I will continue this tomorrow, on Wednesday.

I have restarted 'c1pem1' several times and reconfigured "C0DCU1" on the framebuilder MEDM screen.

Framebuilder now does not work.
  459   Tue Apr 29 21:09:12 2008 ranaDAQCDSFE Filters
These are new FE filters for downsampling and upsampling. We will be going from native hardware sampling rates of 64k down to 32k, 16k, and 2k.

The attached plot shows these filters. They are 3dB ripple, 40 dB stopband, 4th order elliptic filters in which I have moved the zeros around
into good places (e.g. to the Nyquist frequency).

I'm also attaching the .txt file containg the filter coefficients and the design strings. The filters are called x2, x4, and x32, for the
D2, D4, and D32 downsampling, respectively.
Attachment 1: fefilters.jpg
fefilters.jpg
Attachment 2: fefilters.txt
# FILTERS FOR ONLINE SYSTEM
#
# Computer generated file: DO NOT EDIT
#
# MODULES ULYAW
#
################################################################################
### ULYAW                                                                    ###
################################################################################
# SAMPLING ULYAW 65536
... 28 more lines ...
  458   Mon Apr 28 23:44:33 2008 AndreyUpdateComputer Scripts / ProgramsWeather.db

I was trying to figure out how to modify the file "Weather.db" so that the atm.pressure would be recalculated from Pa to bar before appearing in the EPICS screen, but so far I did not succeed. I restarted processor "c1pem1" several times. I will continue this tomorrow, and also I will modify the nmaes of the weather channels.
  457   Sun Apr 27 22:57:15 2008 ajwDAQComputersbr40m?

Quote:

The testpoint manager (which runs on fb40m) crashed this afternoon. Upon re-starting it, I found there was a rogue dtt process on op440m and also a daqd daemon running on br40m. One or both of these caused the tpman to crash. br40m is the frame broadcaster, which is never used here as we don't run DMT. I killed the daqd process there.

The way to find if there is a rogue process is to watch the output to the console from the tpman when you start it:

Allocate new TP handle 56 by 131.215.113.203
Allocate new TP handle 57 by 131.215.113.203
Allocate new TP handle 58 by 131.215.113.203
Allocate new TP handle 59 by 131.215.113.203
Allocate new TP handle 60 by 131.215.113.203
Allocate new TP handle 61 by 131.215.113.203
Allocate new TP handle 62 by 131.215.113.203
Allocate new TP handle 63 by 131.215.113.203
Allocate new TP handle 64 by 131.215.113.203
Allocate new TP handle 65 by 131.215.113.203
Allocate new TP handle 66 by 131.215.113.203
Allocate new TP handle 67 by 131.215.113.203
Allocate new TP handle 68 by 131.215.113.203


If you see something like this, with a new TP handle being allocated every few seconds, you need to log in to the corresponding host and kill whatever process has run away.


I *think* Alex is responsible for the daqd daemon running on br40m (he set up some new stuff recently, a data concentrator and broadcaster); I'll make sure he sees this post.
  456   Sun Apr 27 18:11:58 2008 robDAQComputersbr40m?

The testpoint manager (which runs on fb40m) crashed this afternoon. Upon re-starting it, I found there was a rogue dtt process on op440m and also a daqd daemon running on br40m. One or both of these caused the tpman to crash. br40m is the frame broadcaster, which is never used here as we don't run DMT. I killed the daqd process there.

The way to find if there is a rogue process is to watch the output to the console from the tpman when you start it:

Allocate new TP handle 56 by 131.215.113.203
Allocate new TP handle 57 by 131.215.113.203
Allocate new TP handle 58 by 131.215.113.203
Allocate new TP handle 59 by 131.215.113.203
Allocate new TP handle 60 by 131.215.113.203
Allocate new TP handle 61 by 131.215.113.203
Allocate new TP handle 62 by 131.215.113.203
Allocate new TP handle 63 by 131.215.113.203
Allocate new TP handle 64 by 131.215.113.203
Allocate new TP handle 65 by 131.215.113.203
Allocate new TP handle 66 by 131.215.113.203
Allocate new TP handle 67 by 131.215.113.203
Allocate new TP handle 68 by 131.215.113.203


If you see something like this, with a new TP handle being allocated every few seconds, you need to log in to the corresponding host and kill whatever process has run away.
  455   Sun Apr 27 05:09:30 2008 ranaConfigurationIOOMC WFS Whitening turned on
I hardwired on the MC WFS whitening filters.

The MAX333A switches which choose between whitening and bypass on that board were in the bypass position
because the Xycom220 connections are not there. So the control switch gets +15V but there is no pull
down to set it to the whitened mode.

The least invasive (easiest) change I could do was to tie all of those inputs to ground. This pulls a few mA
through the pull-down resistors but is otherwise innocuous. All of these control lines come in on the A-row
of the P1 connector, so I was able to solder a single wire across all of them to ground them all.

The WFS2 board had a blown electrolytic capacitor on the -15 V line and so there was probably some extra noise
getting in that way. I couldn't find any extra SMD to replace it so I cut the legs off of a 22 uF polarized
tantalum and stuck it in there. Its even close to being the same color. I checked out the other caps, and they were all
close to 68 uF as spec'd. This one had luckily blown open and so didn't suck down the Sorensen and destroy everything.

Plugged everything back in switched the WFS servos back on. Looks good. Took before and after spectra.

In the plot:

GREEN: Open loop dark noise before changes
RED: Open loop bright (MC locked but MCWFS off)
BLUE: Closed loop, MC locked

BLACK: Dark noise after whitening
ORANGE:Closed loop after whitening

The cursor is at 16.25 Hz, the SOS bounce mode.

The I ran the new setMCWFSgains script which uses pzgain to set the UGFs of the 4 loops to 4.01 Hz.
We have in the past had problems with high WFS gains causing instabilities with the CM servo around 10-30 Hz. If this happens we should
just lower the gain by a factor of ~5.
Attachment 1: mcnoise.png
mcnoise.png
  454   Sun Apr 27 02:11:11 2008 ranaConfigurationIOOMC WFS Notes
As noted in the elog from Friday, the WFS has been bad ever since someone switched on the digital whitening filters (FM1 & FM2)
in the MC WFS I&Q filter banks.

On Friday evening, John, Alan, and I went to the rack and verified that although the drawing shows a hookup for the whitening
filters, there is actually no such thing and so we can't have the whitening. So the anti-whitening turns on two lag filters
(2 poles at 4 Hz) and without the hardware this makes the servos unstable by adding 90 deg of phase lag at 4 Hz.

There are still several problems in this system:
- AD797 is used after the mixer. This is an unreliable, noisy part. We need to change this out
  with some OP27s so that this becomes reliable and has a more reasonable noise figure.

- Hard wire the whitening filters ON. We never want these to be off. Then we can turn on the
  anti-whitening. This will give us a factor of 100 better noise without filtering.

- The AD602 on the front of the whitening board has a 100 Ohm internal impedance and the 
  resistor between the demod board and the AD602 is 909 Ohms. This results in dividing the
  signal by 10.

- The signal at the ADC is ~100 cts peak-peak. The full ADC range is, of course, 65000 cts. So
  we could use a lot more gain. The mean quadrant signals are also ~100 cts so we could easily
  up the analog DC gain by a factor of 30 on top of the whitening filter increase.

- The AD602 at the input and the AD620 on the output are both variable gain stages but because
  of our lack of control are set to ambiguous gain levels. We should set the AD602 on the input
  to its max gain of 30 dB. With the -20 dB from the x10 voltage division, this will give us
  an overall gain of 3 for the puny demod signals.

  453   Sat Apr 26 11:21:15 2008 ajwOmnistructureComputersbackup of /cvs/cds restarted
The backup of /cvs/cds (which runs as a cron job on fb40m; see /cvs/cds/caltech/scripts/backup/000README.txt) 
has been down since fb40m was rebooted on March 3.
I was unable to start it because of conflicting ssh keys in /home/controls/.ssh .
With help from Dan Kozak, we got it to work with both sets of keys
( id_rsa, which allows one to ssh between computers in our 113 network without typing a password,
 and backup2PB which allows the cron job to push the backup files to the archive in Powell-Booth).

It still goes down every time one reboots fb40m, and I don't have a solution.
A simple solution is for the script to send an email whenever it can't connect via ssh keys
(requiring a restart of ssh-agent with a passphrase), but email doesn't seem to work on fb40m.
I'll see if I can get help on how to have sendmail run on fb40m.
  452   Sat Apr 26 01:45:38 2008 AndreySummaryPEMWeather Station enhancement
Two more things concerning weather monitoring have been done during this week.

1) A Dataviewer template was created, so that it allows to see "real-time" information from weather channels immediately, without adding many channels "manually".

If one wants to use this template,
open Dataviewer -> "File" -> "Restore Settings", /cvs/cds/caltech/users/Templates/Dataviewer_Templates/Weather.xml.

2) I wrote a couple of Matlab scripts that allow to read data (minute trends) from the Dataviewer channels over some time in the past, save the received data in mat-files, and plot those minute-trends. Thus, one can get plots that are very much similar to what one can see in Dataviewer. These two Matlab files are located in the directory
"/cvs/cds/caltech/users/weather_station". File "WeatherReading.m" allows reading from the weather channels (paths to mDV directory must be configured before using my script), file "WeatherTrends.m" allows plotting of those minute trends.

Unfortunately, hardware problems arise very often if we want to read for a somewhat long time in the past, so until now I have not succeeded in getting trends for more than 20 minutes. As an example, see the attached png-file with the 20-minutes trends of data from Thursday evening.

3) So far I did not have success in learning how to recalculate pressure from Pascals to mbars in EPICS (although I tried google-search).

4) I am making every effort in recent weeks not to put any personal or non-scientific information into elog, but this message could be important for all of us, so I cannot resist:
a shark in the Pacific Ocean has killed a swimmer near San-Diego (I saw this in russian news and then made a quick google-search).
http://latimesblogs.latimes.com/lanow/2008/04/this-just-in-fa.html
Attachment 1: Matlab_Weather_Trends.png
Matlab_Weather_Trends.png
  451   Fri Apr 25 20:53:02 2008 ranaConfigurationIOOMC WFS with more gain
Quick update: we found that the reason for the MC WFS instability was that the digital anti-whitening was one but not the analog whitening.

We turned off the digital filters and were able to increase the gain by a factor of ~30. It is left like this, but if it hampers IFO locking then best to just turn it back down to an overall gain of 0.1 or 0.05.
  450   Fri Apr 25 15:44:21 2008 steveUpdatePSLscattering measurments
In pursuit of a low back scattering, high power beam dump we looked at materials such:
polished copper, polished aluminum, diamond cut aluminum, variety of polished & heat treated stainless steel and shades of black glasses.
Black glass is ideal at low power. Superpolished SS 304 #8 is the only material that measures close to bg
We are still looking for a conductive low back scattering material that would be ideal for a good high power beam trap.

Black glass-shade 12, ss304 superpolished #8 and white paper-B98-24lbs back scattering were compared at 1064 nm

Atm 1: data plot numbers from front panel of SR830 display,
sensitivity: 1x1 mV for bg and ss
1x1 V for wp

Atm 2: drawing of measurement set up
Atm 3: SR830 lock.amp settings
Atm 4: view from steering mirror
Atm 5: view from black glass trap
Atm 6: white paper
Attachment 1: scatmeas20080425.xls
Attachment 2: scattring_set_up_20080425.ppt
Attachment 3: sr830settings.pdf
sr830settings.pdf
Attachment 4: viewfmirror.pdf
viewfmirror.pdf
Attachment 5: viewfbgtrap.pdf
viewfbgtrap.pdf
Attachment 6: wpnmount.pdf
wpnmount.pdf
  449   Fri Apr 25 13:53:11 2008 josephbSummaryComputersNetwork setup
This is the promised more in detail summary from Andrey's log ID 444.

What we did was go around to each hub, one at a time, unplug the network connection, and figure out which light on which hub went out. We then, went back to the control room, confirmed that we were still able to talk to the devices connected to the hub, and if not, rebooted them. This process was repeated for each hub.

As it stands, the hubs located at the ends of arms (in racks 1X4 and 1Y9) are connected to the really old 24 port 10 Base T hub located in 1Y7. In addition, the 5 port SMC hub is plugged into the 8 port SMC switch in 1Y5 (which actually has enough ports to simply move all the connections over to it, so I'm not sure why there are two...).

All other hubs/switches are connected back to the control room 24 port switch.

Attached is a simple diagram of the network connections for the 40m lab.
Attachment 1: 40m_network_90.pdf
40m_network_90.pdf
  448   Fri Apr 25 13:20:04 2008 AndreyUpdatePEMMicrophone test
In response to Rana's request, I tested the microphone (if it is alive or not) by clapping my hands and speaking aloud nearby.

The microphone is alive, see the attached "Full Data" for 5 minutes from Dataviewer.
Attachment 1: Microphone.png
Microphone.png
  447   Fri Apr 25 11:33:40 2008 AndreyConfigurationComputersComputer controlling vaccum equipment

Old computer (located in the south-end end of the interferometer room) that was almost unable to fulfill his duties of controlling vacuum equipment has been replaced to "Linux-3". MEDM runs on "Linux-3".

We checked later that day together with Steve Vass that vacuum equipment (like vacuum valves) can be really controlled from the MEDM-screen 'VacControl.adl'.

Unused flat LCD monitor, keyboard and mouse (parts of the former LINUX-3 computer) were put on the second shelf of the computer rack in the computer room near the HP printer.
  446   Thu Apr 24 23:50:10 2008 ranaUpdateGeneralSyringes in George the Freezer
There are some packets of syringes in the freezer which are labeled as belonging to an S. Waldman.
Thu Apr 24 23:48:55 2008

Be careful of them, don't give them out to the undergrads, and just generally leave them alone. I
will consult with the proper authorities about it.
  445   Thu Apr 24 23:27:48 2008 ranaUpdatePEMacoustic noise in MC_F
I looked at the coherence between the Microphone in the PSL (PEM-AS_MIC) and the MC_F channel.

We want to use a microphone to do Wiener/Adaptive noise cancellation on the MC and so we need to
have a coherence of more than ~0.1 in order for that to have any useful effect.

The attached plot shows the spectrum and coherence with and without the HEPA turned up. As you can
see, the HEPA noise is just barely noticeable in this microphone. Mad

We will need to get something with at least 20 dB more sensitivity.:P
  444   Thu Apr 24 22:06:47 2008 AndreySummaryComputersEthernet Cables and Hubs
Today in the morning (between 8.30AM and noon) Joe and I were working on understanding which ethernet cables connect "processors controlling the work of equipment in the interferometer room" and "Internet hub in the computer room".

Firstly, we took off several times the blue ethernet cables from the router located near ETMX in the morning. We were trying to understand which port in the hub is responsible for the interaction with that processor.

Secondly, we were working on reviving the connection with the computer controlling vacuum in the interferometer.

Later in the middle of the day (around 2PM) Joe continued some work with ethernet cables without me. We plan on continuing the cable work on Friday morning. A better and more detailed elog will appear then.
  443   Thu Apr 24 15:57:53 2008 steveConfigurationSAFETYSafety at AP-ISCT
I measured the output power of the psl after the mechanical shutter.

It was 1.1 W with Ophir power meter, than unlocked the MC and measured
the power at the MC-REFL Beam Dump at the AP-ISCT 0.9 W
Power on MC-REFL photodiode 92 mW

High power metal beam shields were installed around the beam path of
MC-REFL between AP-Viewport and MC-REFL Beam Dump.
Placed HIGH POWER LASER BEAM PATH warning signs on table frame and top
covers.

Last week I placed a small monitor on the top of the OOC that
monitors the resonant spot of MC2. Please keep an eye on this monitor
when working on the AP-ISCT

AP table should NOT be left uncovered. One experienced laser operator
has to be present if the top is removed and IR-viewer scan required.
We need your full cooperation to keep this lab safe.
Attachment 1: P1020197.JPG
P1020197.JPG
Attachment 2: mcrefl3.JPG
mcrefl3.JPG
  442   Thu Apr 24 14:10:26 2008 robUpdateLockinglocking work
Rob, Johnnie

We made some progress on locking last night (Wed night), namely that we were able to handoff (briefly) the CARM-MCL path the REFL-DC error signal. We tried this because we suspect that the reason the PO-DC is not a good CARM error signal is because at low powers, the dc light level in the recycling cavity is dominated by the +f2 RF sideband. Thus, REFL-DC should work a bit better at low powers, which it did. It wasn't super stable, though, so this will require a bit of work to make the transition reliable & stable. The next things to work on include setting the AO path gain properly and possibly going to higher arm powers before handing off (thus increasing the discriminant).

Another thing we found is that the alignment scripts are not working in an ideal fashion. Running the alignment scripts for the two arms (XARM & YARM) leaves the Michelson badly misaligned, making it impossible to get good DRM alignment. This will have to be fixed.