40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 325 of 350  Not logged in ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
ID Date Authorup Type Category Subject
  1220   Fri Jan 9 16:52:18 2009 steveUpdateVACtp3 forline pump replaced
Alberto took 40m vacuum 101 class as we replaced the drypump at the annulus pump line.

He is still not authorized to use the monster 2 3/8" open end wrench that is 36" long.

The fore line pressure dropped to 20 mTorr from 1 Torr as the pumps were swapped.
Bob needs to be given credit for replacing the tip seal on this Varian SH-100 drypump
The ss-hose felt dry at the tp3 exhaust end but it was some what "teflon coated-placticky-
-almost oily" at intake end of the dry pump.
We'll have to replace this metal hose next time.

This is a reminder that the 40m vacuum operation is fully manual.
It requires two people to switch a vacuum valve and one of them has to be experienced.
  1226   Wed Jan 14 09:22:43 2009 steveHowToVACN2 supply pressure lowered for vac valves

Quote:
I've been replacing the N2 bottles recently.
I noticed that the consumption is too high. I had to replace them every two days.
Normally the interval is three or more days.
I suspect there is some leak in the line.

Strangely, it is always the left hand bottle which goes empty. The right hand bottle has been
there for more than a week at about 1000 psi.

We should check it when Steve is back.



All vac valves operated by N2
The two cylinders are located in entry room 103
There is an auto switch over valve between them to insure continuous supply.
The pressure regulator should be set 70 PSI on the gauge
This pressure we keep constant.
All vac valves will close in case of running out of N2 or losing ac power so
it is essential that one replaces empty N2 cylinder.

Simply disconnect large CGA580 fitting with a crescent wrench.
Swap in full cylinder from outside, reconnect fitting tightly and open cylinder valve.
Now you should be reading the full cylinder pressure.
Write each cylinder pressure and date-time on the board so one can see if there is a leak
Attachment 1: n280d.jpg
n280d.jpg
  1227   Wed Jan 14 10:59:06 2009 steveBureaucracyVACrga waiting to be "connected"
We have no RGA data since old Dycor passed away in the middle of Oct, 2008
New SRS-RGA200 was installed on the vac envelope on Nov 14, 2008
It is waiting to be software connected to the 40m logging
  1228   Wed Jan 14 15:53:32 2009 steveDAQPSLMC temperature sensor

Quote:
I added a channel for the temperature sensor on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder. The slow servo of the FSS had to get restarted, the reference cavity locked and so the PMC and MZ.


Where is this channel?
  1229   Thu Jan 15 09:19:32 2009 steveUpdateIOOMC locking
MC2 sus damping was found tripped at the morning the second time this week.

Damping was restored, ISS gain lowered to avoid saturation, MZ manually locked
and MC locking was back.
  1243   Thu Jan 22 08:34:10 2009 steveBureaucracySAFETYsafety training
Kakeru received 40m safety training yesterday.

He has no authorization to work alone in the high power beam path yet.
  1248   Fri Jan 23 10:00:21 2009 steveUpdatePSLPMC transmission is down
The PMC transmission is going down.
I have not relocked the PMC yet.
Attachment 1: pmc4d.jpg
pmc4d.jpg
  1273   Thu Feb 5 09:12:29 2009 steveBureaucracySAFETYsafety updates & fire alarm test

The fire alarm test was completed at 13:30 yesterday.

 

I updated the 40M Emergency Calling List by replacing Rob by Yoichy. The calling order: Vass,  Aso and Taylor.

Rana and Yoichy were added to the "Registered PSL Operator List"  and posted it in the lab. (not in the document, that should be up dated)

We are getting ready for the annual safety audit. It will be held next week  at 14:00 Friday, Feb 13, 2009

Please participate in the preparation by correcting it or just  tell me.

 

Rod Luna is organizing the pick up of the following old equipment:

HP laser jet  5000n, 3 of 19" 10 base-T network ports and  4 small hubs,

2 Sun monitors, 1 Viewsonic monitor, 7 keypads and

Hitachi scopes 2 of V-355, 2 of V-422, 1 of V-202 and 1 of V-6165

 

 

  1280   Fri Feb 6 14:49:31 2009 steveSummarySAFETYlaser inventory
40M Laser Inventory at Feb 05, 2009
 

1, Lightwave PA#102 @ 77,910 hrs 1064nm of 2.8W @ 27.65A
                   NPRO#206 @ 2.4A                    at PSL enclosure............"Big Boy" is waiting for to be retired but not now.

2, Lightwave  NPRO 1064nm of 700mW  #415   at AP table.......cavity length measurements of Alberto

3, CrystaLaser  IRCL-100-1064S, 1064nmS of 100mW  ,sn#IR81132 at east arm cabinet

4, CrystaLaser 1064nm of 180mW # -1274 flq at scattering setup.........flashlight quality

5, RF-PD tester 1064nm of 1.2mW @20mA at SP table

6, Lumix 800-1100nm of 500mW at east arm cabinet

7, JDS-Uniphase 633nmP of 4mW oplev sus laser at 5 places,
    plus four spares in east arm cabinet
 
 
The same information is posted at the 40M WIKI also
 
 
 
  1282   Fri Feb 6 16:23:54 2009 steveUpdateMOPAMOPAs of 7 years

MOPAs and their settings, powers of 7 years in the 40m

Attachment 1: 7ymopas.jpg
7ymopas.jpg
  1288   Tue Feb 10 10:48:48 2009 steveBureaucracySAFETYsafety glasses scanned
All 40m safety glasses are absorbent.
 
They were scanned by 180 mW  2mm spot size  of 1064 nm beam and  their  transmissions were measured
They showed no degradation.
 
UVEX 's  LOTG-YAG/CO2 glasses (fits over personal glasses) 11 pieces,
Glendale-XC by Bacou-Dalloz,  #1166218VLT68% , aerodynamic style, 7 pieces, (4 missing )
KG5- glasses , heavier- made out of glass, 3 pieces,
Thorlab's LG10, 1 piece
Old glasses by "Laser Glass" Nd-YAG, 2 pieces
Totaling 24
 
Atm.2  showing our muilty wavelength glasses of 1064 and 530 nm
Their transmittance for both wavelength were zero.
 
Note: if you have prescripton safety glasses I would like to scan it for you !
 
 
 

 

 

Attachment 1: sgscan1.png
sgscan1.png
Attachment 2: sgscan2.png
sgscan2.png
  1293   Wed Feb 11 11:39:17 2009 steveConfigurationGeneralvac envelope ground
The vacuum envelope was grounded from the IOO chamber to master ground north (under electric breaker panel L)
with 4GA  25ft red battery cable.

 

  1300   Fri Feb 13 08:38:03 2009 steveUpdateIOOMC2 damping restored
  1302   Fri Feb 13 16:30:49 2009 steveConfigurationGeneralstatus quo is disturbed

I have been getting ready for the annual safety inspection in the past 2-3 days.

Meaning cleaning up and disturbing the status quo on the floor  mostly under the optical tables and their surroundings.
For example: pd power supplies, He/Ne laser ps. and their positions.
BNC cables and ac power line positions can be different.
The new rule: no electronic equipment on the floor.
 
All electronic equipment were moved-placed into a plastic dish or tray.
  1319   Wed Feb 18 15:31:38 2009 steveBureaucracySAFETYemergency back-up lights checked
All 13 exit and  emergency back-up lights and their batteries were checked.
Two of  the batteries were replaced.
  1324   Thu Feb 19 11:51:56 2009 steveUpdateMOPAHTEMP variation is too much
The C1:PSL-MOPA_HTEMP variation is more than 0.5 C daily
Normally this temp stays well within 0.1 C
This 80 days plots shows that we have just entered this unstable region some days ago.
The control room temp set unchanged at 70 F, actual temp at ac-mon 69-70 with occasional peaks at 74 F
 
Water temp at chiller repeatedly around 20.6 C at 8 am
This should be rock solid at 20.00C +- 0.02C
 

 

Attachment 1: 80dhtemp.jpg
80dhtemp.jpg
  1332   Mon Feb 23 11:07:01 2009 steveBureaucracySAFETYKiwamu receives safety training

Osamu and Kiwamu received 40m safety training on Thursday, Feb 19, 2009

Kiwamu needs Osamu's close supervision at PSL enclosure and AP table
I hope they already read, understood and signed the laser SOP
  1347   Tue Mar 3 08:44:31 2009 steveUpdatePEMair cond. maintenance today
IFO room 104 air conditions will be shut down for maintenance today.
This should be finished by noon.
The temperature and particle count variation can be more than usual.
  1387   Wed Mar 11 16:41:22 2009 steveUpdateMOPAspare NPRO power

The spare M126N-1064-700,  sn 5519 of Dec 2006 rebuilt  NPRO's power output

 measured   750mW at DC2.06A with Ohpir meter.

Alberto's controller  unit 125/126-OPN-PS,  sn516m was disconnected from lenght measurment NPRO on the AP table.

5519 NPRO  was clamp to the optical table  without heatsink and it was on for 15 minutes.

  1411   Fri Mar 20 11:01:02 2009 steveUpdatePEMparticle counts are high

The outside particle counts for 0.5 micron are 3 million this morning at 9am. Low clouds, foggy condition with low inversion layer.

This makes the 40m lab 30-50K

I just turned on the HEPA filter at the PSL enclosure.

Please, leave it on high

 

Attachment 1: particles32d.jpg
particles32d.jpg
  1414   Fri Mar 20 15:54:29 2009 steveOmnistructureGeneral480V crane power switch on MEZ

CES Mezzanine is beeing rebuilt to accommodate our new neighbor: the 20ft high water slide...& .jacuzzi

All our ac power transformers are up there. Yesterday we labelled the power switch of 480VAC on the mezz

that we need to keep to run the 3 cranes in the lab.

  1420   Tue Mar 24 09:04:02 2009 steveUpdateSUS4.8 mag earthquake

SRM, ITMX, ETMX, ITMY and ETMY lost damping at 4:55am this morning from 4.8 magnitude earthquake.

Their damping were restored.

C1:SUS-ITMX_URSEN_OUTPUT swich was found in off position. It was turned on.

MZehnder  and MC were locked.

The WFS qpd spot needs recentering

  1427   Wed Mar 25 09:55:45 2009 steveUpdateIOOglitching sensors of MC

SUS-MC1_SENSOR_SIDE and SUS-MC2_SENSOR_UL are glitching

Yesterday's 4.8mag earthquake at Salton Sea is shown on Channel 1

Attachment 1: glitchesofMC.jpg
glitchesofMC.jpg
  1439   Sun Mar 29 13:44:27 2009 steveUpdateSUSETMY sus damping restored

ETMY sus damping was found to be tripped.

It was retored.

All fluorecent light were turned off. Please try to conserve some energy.

  1445   Mon Mar 30 15:51:27 2009 steveUpdateElectronicsHP4291A left the lab to be repaired

Eric Gustafson is handling the old HP4291A rehabilitation. Tarac picked both units up today.

March of 2008 Tucker Electronics failed to fix it's intermittent ~25MHz 0.5V oscillation at the swept sine output

See 40m-elog id:398 on 3-24-2008 by Rob Ward

 

  1447   Tue Mar 31 09:42:32 2009 steveUpdatePEMETMY sus damping restored again

The Caltech gasoline storage tank is being upgraded.

They are jack hammering and digging with bulldozer 50 yards south of  ETMY

  1448   Wed Apr 1 10:22:13 2009 steveUpdateVACRGA logging is working

Thanks to Joe B who made the SRS RGA working with linux

Last data file logged at 2008 Oct 24 with old Dycor unit

First data file logged at  2009 Feb 10 with SRS

 

Attachment 1: rga-090401.png
rga-090401.png
  1452   Fri Apr 3 10:01:50 2009 steveUpdateVACrgascan with temp plot

Rga scan of day 231 since pumpdown pd66-m-d231

m stands for maglev pumping speed, vacuum normal condition of valves,

cc4 cold cathode gauge at the rga location,

cc1 is real ifo pressure from the 24" tube at the pumpspool,

PEM-count temp: vac envelope temp at the top of IOO chamber

 

 

Attachment 1: pd66tempow.jpg
pd66tempow.jpg
Attachment 2: rga-090403scan.png
rga-090403scan.png
  1459   Wed Apr 8 09:36:20 2009 steveConfigurationVAC valve condition changed to BG to calibrate the RGA

Vacuum normal valve condition was changed to accommodate SRS-RGA calibration.

 

The annulus valves were closed to free TP3 to pump on the RGA
VM1 was closed to isolate the RGA from the IFO.
 
 
VM3 opened to TP3.
 
This condition is called back ground (BG) mode. It is where we can calibrate and see the RGA background without the IFO.
  1462   Thu Apr 9 11:27:19 2009 steveUpdateVACvac gauge reading problem

Cold cathode gauge CC4  is reading normal.

CC1 is glitching, it is probably dirty.

CC2 is fluctuating too much and it is cutting out for 6-7 minutes. It must be insulated by deposits and there is no emission current.

I think the same goes for P1

They will have to be replaced at the next vent

Attachment 1: vacgflsec.jpg
vacgflsec.jpg
Attachment 2: vacgflmin.jpg
vacgflmin.jpg
  1477   Mon Apr 13 08:59:57 2009 steveUpdatePSLmixers on order

Quote:

Quote:

I then changed the RF Output Adjust slider in increments of 0.5, and measured the peak-to-peak values on the scope. In the table and on the plots, I've taken into account the 12dB attenuation. i.e I actually measured 964mV, so 964mV*10^.6 = 3838mV.


3.8Vpp is about 16dBm.
The mixer for the PMC demodulator is level 23. So 16dBm is insufficient.
What is the level of the new mixer Steve ordered ? 13 ?



I ordered mixers level 13, 17 on Friday and level 23 now.
They should be here Tuesday

NOTE: level 23 power is illegal to use in the 40m lab
They get hot
  1486   Wed Apr 15 11:25:21 2009 steveConfigurationVAC RGA cal completed

Quote:

Vacuum normal valve condition was changed to accommodate SRS-RGA calibration.

 

The annulus valves were closed to free TP3 to pump on the RGA
VM1 was closed to isolate the RGA from the IFO.
 
 
VM3 opened to TP3.
 
This condition is called back ground (BG) mode. It is where we can calibrate and see the RGA background without the IFO.

 Vacuum valve configuration is back to VACUUM NORMAL  condition. RGA  calibration  completed.

RGA  scan attached is the backgroud of the rga with std cal leak open, sn 08581

Krypton at amu 84 and Argon at amu 40 are the cal signals.

 

Attachment 1: rga-090415-bg-d7-cald2.png
rga-090415-bg-d7-cald2.png
  1506   Tue Apr 21 18:18:27 2009 steveUpdateVACmaglev failed

Our Osaka TG360MB maglev failed with CSB error message. This means that the dry emergency landing bearing has to be replaced.

I will consalt with Osaka about the choice of replacing bearing or installing new spare  tomorrow.

Mean while V1 is closed and the vac envelope is not pumped.

Valve configuration: BG -background, pumping on the RGA-only

High voltage to IOO PZT steering mirrors and OMC are turned off.

PSL output shutter is closed and manual block is in place.

I will start cooling the CYO pump in the morning, so the IFO will be pumped by noon.

 

Outgassing plus leakrate after  10 hrs the pressure is 2.3 mTorr

This rate of rise is normal and it is safe to work with the ifo.

 

Attachment 1: nopumping10h.jpg
nopumping10h.jpg
  1507   Wed Apr 22 11:16:26 2009 steveConfigurationVACCryo pump is ON and the Maglev is dead

The CRYO pump cooled down and VC1 was opened. This valve configuration is Cryo-only


PSL output shutter opened at 4pm 

PZT HV power supplies turned on for OMC and IOO steering mirrors.

There positions were not corrected to strain gauge values.

 

Ben helped us to conclude that the FAILURE led indicator is working correctly and

has nothing to do with the one lose, dangling wire#258 in the side connects of the vac rack.

 

I reset the CSB switch inside the Maglev controller and tried to start accelerating.

It fails after 2-3 sec and failure led light comes on.

Now we can say the MAGLEV 360 is DEAD and the new OSAKA TG420M can be swapped in.

However it requires new interface with our epics based MEDM or better...?

 

 

Attachment 1: cryoOn.jpg
cryoOn.jpg
  1511   Thu Apr 23 16:38:33 2009 steveUpdateVACvac valve relay box is shorting

Ben and I found this vacuum valve relay box intermittently shorthing yesterday.

It effects V4, V5, VA6 and VM1........   Please do not touch this box under the beam pipe next to the vac rack!

The function of this box to send 120VAC to the vacuum valve to move.

Attachment 1: vacrel.png
vacrel.png
  1517   Fri Apr 24 16:02:31 2009 steveHowToVACcryo pump interlock rule

I tested the cryopump interlock today. It is touchy. I do not have full confidence in it.

I'm proposing that VC1 gate valve should be kept closed while nobody is working in the 40m lab.

How to open gate valve:

1, confirm temp of 12K on the gauge at the  bottom of the cryopump

2, if medm screen cryo reads OFF( meaning warm) hit reset will result reading ON (meaning cold 12K )

3, open VC1 gate valve if P1 is not higher than 20 mTorr

 

VC1 was closed at 18:25,

IFO condition: not pumped,

expected leak plus  out gassing should be less than 5 mTorr/day

The RGA is in bg-mode, annuloses are closed off

Attachment 1: cryo.png
cryo.png
  1524   Mon Apr 27 18:31:44 2009 steveConfigurationVACVC1 closed

CC1 5e-7 Torr,  VC1 closed at 18:25,  IFO is not pumped, RGA is in bg-mode

 

  1527   Tue Apr 28 09:27:32 2009 steveConfigurationVACcryopump deserves some credit

Congratulation Yoichi and Peter for full rf locking at night. Let's remember that the cryopump was shaking the hole vac envelope and ifo during this full lock.

Attachment 1: cryfl.jpg
cryfl.jpg
Attachment 2: seiscryofl.jpg
seiscryofl.jpg
  1532   Wed Apr 29 10:20:14 2009 steveHowToVACcryo pump interlock rule is waved

Quote:

I tested the cryopump interlock today. It is touchy. I do not have full confidence in it.

I'm proposing that VC1 gate valve should be kept closed while nobody is working in the 40m lab.

How to open gate valve:

1, confirm temp of 12K on the gauge at the  bottom of the cryopump

2, if medm screen cryo reads OFF( meaning warm) hit reset will result reading ON (meaning cold 12K )

3, open VC1 gate valve if P1 is not higher than 20 mTorr

 

VC1 was closed at 18:25,

IFO condition: not pumped,

expected leak plus  out gassing should be less than 5 mTorr/day

The RGA is in bg-mode, annuloses are closed off

 The Cryo pump is running reliably since April 22 hence there is no need to close VC1 repeatedly.

The photo switch interlock was put back onto the H2 vapor pressure gauge and it is working.

  1542   Mon May 4 10:38:52 2009 steveUpdateMOPAlaser power is dropped

As PSL-126MOPA_DTEC went up, the power out put went down yesterday

Attachment 1: dtecup.jpg
dtecup.jpg
  1547   Tue May 5 10:42:18 2009 steveUpdateMOPAlaser power is back

Quote:

As PSL-126MOPA_DTEC went up, the power out put went down yesterday

 The NPRO cooling water was clogged at the needle valve. The heat sink temp was around ~37C

The flow-regulator  needle valve position is locked with a nut and it is frozen. It is not adjustable. However Jeenne's tapping and pushing down on the plastic hardware cleared the way for the water flow.

We have to remember to replace this needle valve when the new NPRO will be swapped in. I checked on the heat sink temp this morning. It is ~18C

There is condensation on the south end of the NPRO body, I wish that the DTEC value would just a little higher like 0.5V

The wavelenght of the diode is temp dependent: 0.3 nm/C. The fine tuning of this diode is done by thermo-electric cooler ( TEC )

To keep the diode precisely tuned to the absorption of the laser gain material the diode temp is held constant using electronic feedback control.

This value is zero now.

 

Attachment 1: uncloged.jpg
uncloged.jpg
  1553   Thu May 7 10:28:20 2009 steveUpdateVACretrofitted maglev's needs

 

 Our spare Osaka maglev purchased in Oct 2005 turned out to be having a viton o-ring seal connection on the intake.

It was shipped back to San Jose for retrofitting it with 6" conflat flange ( CF ) This CF is using copper gasket so there will be no permeation of He when you leak check the IFO

 

The digital controller and cable are here. The controller needs to be interfaced with the interlocks and computer system; those have been in a neglected condition lately.

see elog #1505  Historically after every REBOOT of c1vac2 the readbacks works for 3-4 days only. Fixing of this was postponed many times in the past as low priority or lack of knowledgeable

enthusiast.

 

The maglev TG390MCAB wil be back on Tuesday, May 4, 2009.  The mourning of our fateful 360 will only end at the first levitation of the 390.

 

  1572   Sun May 10 13:41:17 2009 steveUpdateVACETMY damping restored, VC1 opened

ETMY damping restored.

Cryo  interlock closed VC1 ~2 days ago. P1 is 6.3 mTorr. Cryo temp 12K stable, reset photoswitch and opened VC1

  1573   Mon May 11 11:49:20 2009 steveUpdatePSLMOPA cooling water lines are backwards

Quote:
This is 8 days of 10-minute trend.

DTEC is just the feedback control signal required to keep the NPRO's pump diode at a constant temperature.
Its not the amplifier or the actual NPRO crystal's temperature readout.

There is no TEC for the amplifier. It looks like to me that by opening up the flow to the NPRO some more
we have reduced the flow to the amplifier (which is the one that needs it) and created these temperature
fluctuations.

What we need to do is choke down the needle valve and ream out the NPRO block.




I have measured the "input" line temp at the MOPA box 10 C and the "out" line 8 C

This must be corrected.

However look at the 80 days plot of operation where the head temp variation is nothing new
Attachment 1: htempvar80d.jpg
htempvar80d.jpg
  1586   Thu May 14 15:28:28 2009 steveSummarySUSApril 24 earthquake effect on MC2

Quote:

Quote:
The MC side problem could also be the side tramp unit problem. Set the tramp to 0 and see if that helps.


This started around April 23, around the time that TP1 failed and we switched to the cryopump, and also when there was a mag 4 earthquake in LA. My money's on the EQ. But I don't know how.



Only MC2 moved in this earth quake. Was the MC alignment touched up since than?
Have you guys swapped satellite amp of MC3 yet?
Attachment 1: eq042409.jpg
eq042409.jpg
  1604   Tue May 19 09:34:29 2009 steveConfigurationVACIFO is not pumped & CRYO is being regenerated
Morning Vacuum condition: IFO is not being pumped, P1 pressure is 1.8 mTorr and rising (see P1 pressure plot of 100 min ).

Overnight the RGA protection software interlock at closed the VM1 valve triggering on CC1 = 1e-5 torr.

This interlock blocked our attempt to hold the IFO operational pressure in the high 1e-5 Torr range with one small
"beer can" turbopump (Varian V70D drag-turbo pumping speed for N2 is ~60 l/s at 75KRPM).

I started CRYO regeneration using TP3. Pressure readout on the P3 gauge. This is after 30 days of CRYO operation.

V5 was closed for 60 sec to see the outgassing rate of the cryopump surfaces. It was good (but I am not going to elog
what 'good' actually means - instead I will write it in my paper logbook to prevent others from learning). I will now'
go start cooling down the cryo pump.

** translated into English by Rana
Attachment 1: cryoreg.jpg
cryoreg.jpg
Attachment 2: cryo30d.jpg
cryo30d.jpg
  1607   Tue May 19 15:57:07 2009 steveUpdateIOOMC2 damping restored after EQ

  Earthquake mag 4.0 at Lennox, Ca     trips MC2 watchdogs       http://quake.usgs.gov/recenteqs/Quakes/ci10411545.html

See 40m accelerometers as they see it.

Attachment 1: acc.jpg
acc.jpg
  1609   Tue May 19 16:18:45 2009 steveConfigurationVACIFO pressure is 4.2e-7 Torr with CRYO pump

Quote:
Morning Vacuum condition: IFO is not being pumped, P1 pressure is 1.8 mTorr and rising (see P1 pressure plot of 100 min ).

Overnight the RGA protection software interlock at closed the VM1 valve triggering on CC1 = 1e-5 torr.

This interlock blocked our attempt to hold the IFO operational pressure in the high 1e-5 Torr range with one small
"beer can" turbopump (Varian V70D drag-turbo pumping speed for N2 is ~60 l/s at 75KRPM).

I started CRYO regeneration using TP3. Pressure readout on the P3 gauge. This is after 30 days of CRYO operation.

V5 was closed for 60 sec to see the outgassing rate of the cryopump surfaces. It was good (but I am not going to elog
what 'good' actually means - instead I will write it in my paper logbook to prevent others from learning). I will now'
go start cooling down the cryo pump.

** translated into English by Rana


The Cryo cooled down to ~12K by noon.Photo switch was reset and VC1 was opened at 2pm

The VACMONITOR.adl screen is not working. Someone made some improvement on it last night.
  1612   Wed May 20 09:55:18 2009 steveUpdatePEMoplev servos turned off

All oplevs servos turned off to protect our suspentions from vibration due to drilling and pounding in CES high bay area.

This activity will be done from 10 am till 3 pm today.

 

Meanwhile our IFO-air conditions are turned off for maintenance.

Their performance of 6 months is shown on plot.

Attachment 1: ACpart6m.jpg
ACpart6m.jpg
  1613   Wed May 20 10:43:17 2009 steveOmnistructureEnvironmentaccelerometers sensitivity

Quote:

2009 May 18 03:39:36 UTC

Earthquake Details

Magnitude 5.0
Date-Time
  • Monday, May 18, 2009 at 03:39:36 UTC
  • Sunday, May 17, 2009 at 08:39:36 PM at epicenter
Location 33.940°N, 118.338°W
Depth 13.5 km (8.4 miles)
Region GREATER LOS ANGELES AREA, CALIFORNIA
Distances
  • 2 km (1 miles) E (91°) from Lennox, CA
  • 2 km (1 miles) SSE (159°) from Inglewood, CA
  • 3 km (2 miles) NNE (22°) from Hawthorne, CA
  • 7 km (4 miles) ENE (72°) from El Segundo, CA
  • 15 km (10 miles) SSW (213°) from Los Angeles Civic Center, CA
Location Uncertainty horizontal +/- 0.4 km (0.2 miles); depth +/- 0.9 km (0.6 miles)
Parameters Nph=139, Dmin=7 km, Rmss=0.42 sec, Gp= 40°,
M-type=local magnitude (ML), Version=C
Source
Event ID ci10410337

 Wilcoxon 731A seismic accelerometers and Guralp CMG-40T-old seismometer at magnitude 5 and 4 erthquakes

Attachment 1: eq5m4.jpg
eq5m4.jpg
ELOG V3.1.3-