ID |
Date |
Author |
Type |
Category |
Subject |
1220
|
Fri Jan 9 16:52:18 2009 |
steve | Update | VAC | tp3 forline pump replaced |
Alberto took 40m vacuum 101 class as we replaced the drypump at the annulus pump line.
He is still not authorized to use the monster 2 3/8" open end wrench that is 36" long.
The fore line pressure dropped to 20 mTorr from 1 Torr as the pumps were swapped.
Bob needs to be given credit for replacing the tip seal on this Varian SH-100 drypump
The ss-hose felt dry at the tp3 exhaust end but it was some what "teflon coated-placticky-
-almost oily" at intake end of the dry pump.
We'll have to replace this metal hose next time.
This is a reminder that the 40m vacuum operation is fully manual.
It requires two people to switch a vacuum valve and one of them has to be experienced. |
1226
|
Wed Jan 14 09:22:43 2009 |
steve | HowTo | VAC | N2 supply pressure lowered for vac valves |
Quote: | I've been replacing the N2 bottles recently.
I noticed that the consumption is too high. I had to replace them every two days.
Normally the interval is three or more days.
I suspect there is some leak in the line.
Strangely, it is always the left hand bottle which goes empty. The right hand bottle has been
there for more than a week at about 1000 psi.
We should check it when Steve is back. |
All vac valves operated by N2
The two cylinders are located in entry room 103
There is an auto switch over valve between them to insure continuous supply.
The pressure regulator should be set 70 PSI on the gauge
This pressure we keep constant.
All vac valves will close in case of running out of N2 or losing ac power so
it is essential that one replaces empty N2 cylinder.
Simply disconnect large CGA580 fitting with a crescent wrench.
Swap in full cylinder from outside, reconnect fitting tightly and open cylinder valve.
Now you should be reading the full cylinder pressure.
Write each cylinder pressure and date-time on the board so one can see if there is a leak |
Attachment 1: n280d.jpg
|
|
1227
|
Wed Jan 14 10:59:06 2009 |
steve | Bureaucracy | VAC | rga waiting to be "connected" |
We have no RGA data since old Dycor passed away in the middle of Oct, 2008
New SRS-RGA200 was installed on the vac envelope on Nov 14, 2008
It is waiting to be software connected to the 40m logging |
1228
|
Wed Jan 14 15:53:32 2009 |
steve | DAQ | PSL | MC temperature sensor |
Quote: | I added a channel for the temperature sensor on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder. The slow servo of the FSS had to get restarted, the reference cavity locked and so the PMC and MZ. |
Where is this channel? |
1229
|
Thu Jan 15 09:19:32 2009 |
steve | Update | IOO | MC locking |
MC2 sus damping was found tripped at the morning the second time this week.
Damping was restored, ISS gain lowered to avoid saturation, MZ manually locked
and MC locking was back. |
1243
|
Thu Jan 22 08:34:10 2009 |
steve | Bureaucracy | SAFETY | safety training |
Kakeru received 40m safety training yesterday.
He has no authorization to work alone in the high power beam path yet. |
1248
|
Fri Jan 23 10:00:21 2009 |
steve | Update | PSL | PMC transmission is down |
The PMC transmission is going down.
I have not relocked the PMC yet. |
Attachment 1: pmc4d.jpg
|
|
1273
|
Thu Feb 5 09:12:29 2009 |
steve | Bureaucracy | SAFETY | safety updates & fire alarm test |
The fire alarm test was completed at 13:30 yesterday.
I updated the 40M Emergency Calling List by replacing Rob by Yoichy. The calling order: Vass, Aso and Taylor.
Rana and Yoichy were added to the "Registered PSL Operator List" and posted it in the lab. (not in the document, that should be up dated)
We are getting ready for the annual safety audit. It will be held next week at 14:00 Friday, Feb 13, 2009
Please participate in the preparation by correcting it or just tell me.
Rod Luna is organizing the pick up of the following old equipment:
HP laser jet 5000n, 3 of 19" 10 base-T network ports and 4 small hubs,
2 Sun monitors, 1 Viewsonic monitor, 7 keypads and
Hitachi scopes 2 of V-355, 2 of V-422, 1 of V-202 and 1 of V-6165
|
1280
|
Fri Feb 6 14:49:31 2009 |
steve | Summary | SAFETY | laser inventory |
40M Laser Inventory at Feb 05, 2009
1, Lightwave PA#102 @ 77,910 hrs 1064nm of 2.8W @ 27.65A
NPRO#206 @ 2.4A at PSL enclosure............"Big Boy" is waiting for to be retired but not now.
2, Lightwave NPRO 1064nm of 700mW #415 at AP table.......cavity length measurements of Alberto
3, CrystaLaser IRCL-100-1064S, 1064nmS of 100mW ,sn#IR81132 at east arm cabinet
4, CrystaLaser 1064nm of 180mW # -1274 flq at scattering setup.........flashlight quality
5, RF-PD tester 1064nm of 1.2mW @20mA at SP table
6, Lumix 800-1100nm of 500mW at east arm cabinet
7, JDS-Uniphase 633nmP of 4mW oplev sus laser at 5 places,
plus four spares in east arm cabinet
The same information is posted at the 40M WIKI also
|
1282
|
Fri Feb 6 16:23:54 2009 |
steve | Update | MOPA | MOPAs of 7 years |
MOPAs and their settings, powers of 7 years in the 40m |
Attachment 1: 7ymopas.jpg
|
|
1288
|
Tue Feb 10 10:48:48 2009 |
steve | Bureaucracy | SAFETY | safety glasses scanned |
All 40m safety glasses are absorbent.
They were scanned by 180 mW 2mm spot size of 1064 nm beam and their transmissions were measured
They showed no degradation.
UVEX 's LOTG-YAG/CO2 glasses (fits over personal glasses) 11 pieces,
Glendale-XC by Bacou-Dalloz, #1166218VLT68% , aerodynamic style, 7 pieces, (4 missing )
KG5- glasses , heavier- made out of glass, 3 pieces,
Thorlab's LG10, 1 piece
Old glasses by "Laser Glass" Nd-YAG, 2 pieces
Totaling 24
Atm.2 showing our muilty wavelength glasses of 1064 and 530 nm
Their transmittance for both wavelength were zero.
Note: if you have prescripton safety glasses I would like to scan it for you !
|
Attachment 1: sgscan1.png
|
|
Attachment 2: sgscan2.png
|
|
1293
|
Wed Feb 11 11:39:17 2009 |
steve | Configuration | General | vac envelope ground |
The vacuum envelope was grounded from the IOO chamber to master ground north (under electric breaker panel L)
with 4GA 25ft red battery cable.
|
1300
|
Fri Feb 13 08:38:03 2009 |
steve | Update | IOO | MC2 damping restored |
|
1302
|
Fri Feb 13 16:30:49 2009 |
steve | Configuration | General | status quo is disturbed |
I have been getting ready for the annual safety inspection in the past 2-3 days.
Meaning cleaning up and disturbing the status quo on the floor mostly under the optical tables and their surroundings.
For example: pd power supplies, He/Ne laser ps. and their positions.
BNC cables and ac power line positions can be different.
The new rule: no electronic equipment on the floor.
All electronic equipment were moved-placed into a plastic dish or tray. |
1319
|
Wed Feb 18 15:31:38 2009 |
steve | Bureaucracy | SAFETY | emergency back-up lights checked |
All 13 exit and emergency back-up lights and their batteries were checked.
Two of the batteries were replaced. |
1324
|
Thu Feb 19 11:51:56 2009 |
steve | Update | MOPA | HTEMP variation is too much |
The C1:PSL-MOPA_HTEMP variation is more than 0.5 C daily
Normally this temp stays well within 0.1 C
This 80 days plots shows that we have just entered this unstable region some days ago.
The control room temp set unchanged at 70 F, actual temp at ac-mon 69-70 with occasional peaks at 74 F
Water temp at chiller repeatedly around 20.6 C at 8 am
This should be rock solid at 20.00C +- 0.02C
|
Attachment 1: 80dhtemp.jpg
|
|
1332
|
Mon Feb 23 11:07:01 2009 |
steve | Bureaucracy | SAFETY | Kiwamu receives safety training |
Osamu and Kiwamu received 40m safety training on Thursday, Feb 19, 2009
Kiwamu needs Osamu's close supervision at PSL enclosure and AP table
I hope they already read, understood and signed the laser SOP |
1347
|
Tue Mar 3 08:44:31 2009 |
steve | Update | PEM | air cond. maintenance today |
IFO room 104 air conditions will be shut down for maintenance today.
This should be finished by noon.
The temperature and particle count variation can be more than usual. |
1387
|
Wed Mar 11 16:41:22 2009 |
steve | Update | MOPA | spare NPRO power |
The spare M126N-1064-700, sn 5519 of Dec 2006 rebuilt NPRO's power output
measured 750mW at DC2.06A with Ohpir meter.
Alberto's controller unit 125/126-OPN-PS, sn516m was disconnected from lenght measurment NPRO on the AP table.
5519 NPRO was clamp to the optical table without heatsink and it was on for 15 minutes. |
1411
|
Fri Mar 20 11:01:02 2009 |
steve | Update | PEM | particle counts are high |
The outside particle counts for 0.5 micron are 3 million this morning at 9am. Low clouds, foggy condition with low inversion layer.
This makes the 40m lab 30-50K
I just turned on the HEPA filter at the PSL enclosure.
Please, leave it on high
|
Attachment 1: particles32d.jpg
|
|
1414
|
Fri Mar 20 15:54:29 2009 |
steve | Omnistructure | General | 480V crane power switch on MEZ |
CES Mezzanine is beeing rebuilt to accommodate our new neighbor: the 20ft high water slide...& .jacuzzi
All our ac power transformers are up there. Yesterday we labelled the power switch of 480VAC on the mezz
that we need to keep to run the 3 cranes in the lab. |
1420
|
Tue Mar 24 09:04:02 2009 |
steve | Update | SUS | 4.8 mag earthquake |
SRM, ITMX, ETMX, ITMY and ETMY lost damping at 4:55am this morning from 4.8 magnitude earthquake.
Their damping were restored.
C1:SUS-ITMX_URSEN_OUTPUT swich was found in off position. It was turned on.
MZehnder and MC were locked.
The WFS qpd spot needs recentering |
1427
|
Wed Mar 25 09:55:45 2009 |
steve | Update | IOO | glitching sensors of MC |
SUS-MC1_SENSOR_SIDE and SUS-MC2_SENSOR_UL are glitching
Yesterday's 4.8mag earthquake at Salton Sea is shown on Channel 1 |
Attachment 1: glitchesofMC.jpg
|
|
1439
|
Sun Mar 29 13:44:27 2009 |
steve | Update | SUS | ETMY sus damping restored |
ETMY sus damping was found to be tripped.
It was retored.
All fluorecent light were turned off. Please try to conserve some energy. |
1445
|
Mon Mar 30 15:51:27 2009 |
steve | Update | Electronics | HP4291A left the lab to be repaired |
Eric Gustafson is handling the old HP4291A rehabilitation. Tarac picked both units up today.
March of 2008 Tucker Electronics failed to fix it's intermittent ~25MHz 0.5V oscillation at the swept sine output
See 40m-elog id:398 on 3-24-2008 by Rob Ward
|
1447
|
Tue Mar 31 09:42:32 2009 |
steve | Update | PEM | ETMY sus damping restored again |
The Caltech gasoline storage tank is being upgraded.
They are jack hammering and digging with bulldozer 50 yards south of ETMY |
1448
|
Wed Apr 1 10:22:13 2009 |
steve | Update | VAC | RGA logging is working |
Thanks to Joe B who made the SRS RGA working with linux
Last data file logged at 2008 Oct 24 with old Dycor unit
First data file logged at 2009 Feb 10 with SRS
|
Attachment 1: rga-090401.png
|
|
1452
|
Fri Apr 3 10:01:50 2009 |
steve | Update | VAC | rgascan with temp plot |
Rga scan of day 231 since pumpdown pd66-m-d231
m stands for maglev pumping speed, vacuum normal condition of valves,
cc4 cold cathode gauge at the rga location,
cc1 is real ifo pressure from the 24" tube at the pumpspool,
PEM-count temp: vac envelope temp at the top of IOO chamber
|
Attachment 1: pd66tempow.jpg
|
|
Attachment 2: rga-090403scan.png
|
|
1459
|
Wed Apr 8 09:36:20 2009 |
steve | Configuration | VAC | valve condition changed to BG to calibrate the RGA |
Vacuum normal valve condition was changed to accommodate SRS-RGA calibration.
The annulus valves were closed to free TP3 to pump on the RGA
VM1 was closed to isolate the RGA from the IFO.
VM3 opened to TP3.
This condition is called back ground (BG) mode. It is where we can calibrate and see the RGA background without the IFO. |
1462
|
Thu Apr 9 11:27:19 2009 |
steve | Update | VAC | vac gauge reading problem |
Cold cathode gauge CC4 is reading normal.
CC1 is glitching, it is probably dirty.
CC2 is fluctuating too much and it is cutting out for 6-7 minutes. It must be insulated by deposits and there is no emission current.
I think the same goes for P1
They will have to be replaced at the next vent |
Attachment 1: vacgflsec.jpg
|
|
Attachment 2: vacgflmin.jpg
|
|
1477
|
Mon Apr 13 08:59:57 2009 |
steve | Update | PSL | mixers on order |
Quote: |
Quote: |
I then changed the RF Output Adjust slider in increments of 0.5, and measured the peak-to-peak values on the scope. In the table and on the plots, I've taken into account the 12dB attenuation. i.e I actually measured 964mV, so 964mV*10^.6 = 3838mV.
|
3.8Vpp is about 16dBm.
The mixer for the PMC demodulator is level 23. So 16dBm is insufficient.
What is the level of the new mixer Steve ordered ? 13 ? |
I ordered mixers level 13, 17 on Friday and level 23 now.
They should be here Tuesday
NOTE: level 23 power is illegal to use in the 40m lab
They get hot |
1486
|
Wed Apr 15 11:25:21 2009 |
steve | Configuration | VAC | RGA cal completed |
Quote: |
Vacuum normal valve condition was changed to accommodate SRS-RGA calibration.
The annulus valves were closed to free TP3 to pump on the RGA
VM1 was closed to isolate the RGA from the IFO.
VM3 opened to TP3.
This condition is called back ground (BG) mode. It is where we can calibrate and see the RGA background without the IFO.
|
Vacuum valve configuration is back to VACUUM NORMAL condition. RGA calibration completed.
RGA scan attached is the backgroud of the rga with std cal leak open, sn 08581
Krypton at amu 84 and Argon at amu 40 are the cal signals.
|
Attachment 1: rga-090415-bg-d7-cald2.png
|
|
1506
|
Tue Apr 21 18:18:27 2009 |
steve | Update | VAC | maglev failed |
Our Osaka TG360MB maglev failed with CSB error message. This means that the dry emergency landing bearing has to be replaced.
I will consalt with Osaka about the choice of replacing bearing or installing new spare tomorrow.
Mean while V1 is closed and the vac envelope is not pumped.
Valve configuration: BG -background, pumping on the RGA-only
High voltage to IOO PZT steering mirrors and OMC are turned off.
PSL output shutter is closed and manual block is in place.
I will start cooling the CYO pump in the morning, so the IFO will be pumped by noon.
Outgassing plus leakrate after 10 hrs the pressure is 2.3 mTorr
This rate of rise is normal and it is safe to work with the ifo.
|
Attachment 1: nopumping10h.jpg
|
|
1507
|
Wed Apr 22 11:16:26 2009 |
steve | Configuration | VAC | Cryo pump is ON and the Maglev is dead |
The CRYO pump cooled down and VC1 was opened. This valve configuration is Cryo-only
PSL output shutter opened at 4pm
PZT HV power supplies turned on for OMC and IOO steering mirrors.
There positions were not corrected to strain gauge values.
Ben helped us to conclude that the FAILURE led indicator is working correctly and
has nothing to do with the one lose, dangling wire#258 in the side connects of the vac rack.
I reset the CSB switch inside the Maglev controller and tried to start accelerating.
It fails after 2-3 sec and failure led light comes on.
Now we can say the MAGLEV 360 is DEAD and the new OSAKA TG420M can be swapped in.
However it requires new interface with our epics based MEDM or better...?
|
Attachment 1: cryoOn.jpg
|
|
1511
|
Thu Apr 23 16:38:33 2009 |
steve | Update | VAC | vac valve relay box is shorting |
Ben and I found this vacuum valve relay box intermittently shorthing yesterday.
It effects V4, V5, VA6 and VM1........ Please do not touch this box under the beam pipe next to the vac rack!
The function of this box to send 120VAC to the vacuum valve to move. |
Attachment 1: vacrel.png
|
|
1517
|
Fri Apr 24 16:02:31 2009 |
steve | HowTo | VAC | cryo pump interlock rule |
I tested the cryopump interlock today. It is touchy. I do not have full confidence in it.
I'm proposing that VC1 gate valve should be kept closed while nobody is working in the 40m lab.
How to open gate valve:
1, confirm temp of 12K on the gauge at the bottom of the cryopump
2, if medm screen cryo reads OFF( meaning warm) hit reset will result reading ON (meaning cold 12K )
3, open VC1 gate valve if P1 is not higher than 20 mTorr
VC1 was closed at 18:25,
IFO condition: not pumped,
expected leak plus out gassing should be less than 5 mTorr/day
The RGA is in bg-mode, annuloses are closed off |
Attachment 1: cryo.png
|
|
1524
|
Mon Apr 27 18:31:44 2009 |
steve | Configuration | VAC | VC1 closed |
CC1 5e-7 Torr, VC1 closed at 18:25, IFO is not pumped, RGA is in bg-mode
|
1527
|
Tue Apr 28 09:27:32 2009 |
steve | Configuration | VAC | cryopump deserves some credit |
Congratulation Yoichi and Peter for full rf locking at night. Let's remember that the cryopump was shaking the hole vac envelope and ifo during this full lock. |
Attachment 1: cryfl.jpg
|
|
Attachment 2: seiscryofl.jpg
|
|
1532
|
Wed Apr 29 10:20:14 2009 |
steve | HowTo | VAC | cryo pump interlock rule is waved |
Quote: |
I tested the cryopump interlock today. It is touchy. I do not have full confidence in it.
I'm proposing that VC1 gate valve should be kept closed while nobody is working in the 40m lab.
How to open gate valve:
1, confirm temp of 12K on the gauge at the bottom of the cryopump
2, if medm screen cryo reads OFF( meaning warm) hit reset will result reading ON (meaning cold 12K )
3, open VC1 gate valve if P1 is not higher than 20 mTorr
VC1 was closed at 18:25,
IFO condition: not pumped,
expected leak plus out gassing should be less than 5 mTorr/day
The RGA is in bg-mode, annuloses are closed off
|
The Cryo pump is running reliably since April 22 hence there is no need to close VC1 repeatedly.
The photo switch interlock was put back onto the H2 vapor pressure gauge and it is working. |
1542
|
Mon May 4 10:38:52 2009 |
steve | Update | MOPA | laser power is dropped |
As PSL-126MOPA_DTEC went up, the power out put went down yesterday |
Attachment 1: dtecup.jpg
|
|
1547
|
Tue May 5 10:42:18 2009 |
steve | Update | MOPA | laser power is back |
Quote: |
As PSL-126MOPA_DTEC went up, the power out put went down yesterday
|
The NPRO cooling water was clogged at the needle valve. The heat sink temp was around ~37C
The flow-regulator needle valve position is locked with a nut and it is frozen. It is not adjustable. However Jeenne's tapping and pushing down on the plastic hardware cleared the way for the water flow.
We have to remember to replace this needle valve when the new NPRO will be swapped in. I checked on the heat sink temp this morning. It is ~18C
There is condensation on the south end of the NPRO body, I wish that the DTEC value would just a little higher like 0.5V
The wavelenght of the diode is temp dependent: 0.3 nm/C. The fine tuning of this diode is done by thermo-electric cooler ( TEC )
To keep the diode precisely tuned to the absorption of the laser gain material the diode temp is held constant using electronic feedback control.
This value is zero now.
|
Attachment 1: uncloged.jpg
|
|
1553
|
Thu May 7 10:28:20 2009 |
steve | Update | VAC | retrofitted maglev's needs |
Our spare Osaka maglev purchased in Oct 2005 turned out to be having a viton o-ring seal connection on the intake.
It was shipped back to San Jose for retrofitting it with 6" conflat flange ( CF ) This CF is using copper gasket so there will be no permeation of He when you leak check the IFO
The digital controller and cable are here. The controller needs to be interfaced with the interlocks and computer system; those have been in a neglected condition lately.
see elog #1505 Historically after every REBOOT of c1vac2 the readbacks works for 3-4 days only. Fixing of this was postponed many times in the past as low priority or lack of knowledgeable
enthusiast.
The maglev TG390MCAB wil be back on Tuesday, May 4, 2009. The mourning of our fateful 360 will only end at the first levitation of the 390.
|
1572
|
Sun May 10 13:41:17 2009 |
steve | Update | VAC | ETMY damping restored, VC1 opened |
ETMY damping restored.
Cryo interlock closed VC1 ~2 days ago. P1 is 6.3 mTorr. Cryo temp 12K stable, reset photoswitch and opened VC1 |
1573
|
Mon May 11 11:49:20 2009 |
steve | Update | PSL | MOPA cooling water lines are backwards |
Quote: | This is 8 days of 10-minute trend.
DTEC is just the feedback control signal required to keep the NPRO's pump diode at a constant temperature.
Its not the amplifier or the actual NPRO crystal's temperature readout.
There is no TEC for the amplifier. It looks like to me that by opening up the flow to the NPRO some more
we have reduced the flow to the amplifier (which is the one that needs it) and created these temperature
fluctuations.
What we need to do is choke down the needle valve and ream out the NPRO block. |
I have measured the "input" line temp at the MOPA box 10 C and the "out" line 8 C
This must be corrected.
However look at the 80 days plot of operation where the head temp variation is nothing new |
Attachment 1: htempvar80d.jpg
|
|
1586
|
Thu May 14 15:28:28 2009 |
steve | Summary | SUS | April 24 earthquake effect on MC2 |
Quote: |
Quote: | The MC side problem could also be the side tramp unit problem. Set the tramp to 0 and see if that helps. |
This started around April 23, around the time that TP1 failed and we switched to the cryopump, and also when there was a mag 4 earthquake in LA. My money's on the EQ. But I don't know how. |
Only MC2 moved in this earth quake. Was the MC alignment touched up since than?
Have you guys swapped satellite amp of MC3 yet? |
Attachment 1: eq042409.jpg
|
|
1604
|
Tue May 19 09:34:29 2009 |
steve | Configuration | VAC | IFO is not pumped & CRYO is being regenerated |
Morning Vacuum condition: IFO is not being pumped, P1 pressure is 1.8 mTorr and rising (see P1 pressure plot of 100 min ).
Overnight the RGA protection software interlock at closed the VM1 valve triggering on CC1 = 1e-5 torr.
This interlock blocked our attempt to hold the IFO operational pressure in the high 1e-5 Torr range with one small
"beer can" turbopump (Varian V70D drag-turbo pumping speed for N2 is ~60 l/s at 75KRPM).
I started CRYO regeneration using TP3. Pressure readout on the P3 gauge. This is after 30 days of CRYO operation.
V5 was closed for 60 sec to see the outgassing rate of the cryopump surfaces. It was good (but I am not going to elog
what 'good' actually means - instead I will write it in my paper logbook to prevent others from learning). I will now'
go start cooling down the cryo pump.
** translated into English by Rana |
Attachment 1: cryoreg.jpg
|
|
Attachment 2: cryo30d.jpg
|
|
1607
|
Tue May 19 15:57:07 2009 |
steve | Update | IOO | MC2 damping restored after EQ |
Earthquake mag 4.0 at Lennox, Ca trips MC2 watchdogs http://quake.usgs.gov/recenteqs/Quakes/ci10411545.html
See 40m accelerometers as they see it. |
Attachment 1: acc.jpg
|
|
1609
|
Tue May 19 16:18:45 2009 |
steve | Configuration | VAC | IFO pressure is 4.2e-7 Torr with CRYO pump |
Quote: | Morning Vacuum condition: IFO is not being pumped, P1 pressure is 1.8 mTorr and rising (see P1 pressure plot of 100 min ).
Overnight the RGA protection software interlock at closed the VM1 valve triggering on CC1 = 1e-5 torr.
This interlock blocked our attempt to hold the IFO operational pressure in the high 1e-5 Torr range with one small
"beer can" turbopump (Varian V70D drag-turbo pumping speed for N2 is ~60 l/s at 75KRPM).
I started CRYO regeneration using TP3. Pressure readout on the P3 gauge. This is after 30 days of CRYO operation.
V5 was closed for 60 sec to see the outgassing rate of the cryopump surfaces. It was good (but I am not going to elog
what 'good' actually means - instead I will write it in my paper logbook to prevent others from learning). I will now'
go start cooling down the cryo pump.
** translated into English by Rana |
The Cryo cooled down to ~12K by noon.Photo switch was reset and VC1 was opened at 2pm
The VACMONITOR.adl screen is not working. Someone made some improvement on it last night. |
1612
|
Wed May 20 09:55:18 2009 |
steve | Update | PEM | oplev servos turned off |
All oplevs servos turned off to protect our suspentions from vibration due to drilling and pounding in CES high bay area.
This activity will be done from 10 am till 3 pm today.
Meanwhile our IFO-air conditions are turned off for maintenance.
Their performance of 6 months is shown on plot. |
Attachment 1: ACpart6m.jpg
|
|
1613
|
Wed May 20 10:43:17 2009 |
steve | Omnistructure | Environment | accelerometers sensitivity |
Quote: |
2009 May 18 03:39:36 UTC
Earthquake Details
Magnitude |
5.0 |
Date-Time |
- Monday, May 18, 2009 at 03:39:36 UTC
- Sunday, May 17, 2009 at 08:39:36 PM at epicenter
|
Location |
33.940°N, 118.338°W |
Depth |
13.5 km (8.4 miles) |
Region |
GREATER LOS ANGELES AREA, CALIFORNIA |
Distances |
- 2 km (1 miles) E (91°) from Lennox, CA
- 2 km (1 miles) SSE (159°) from Inglewood, CA
- 3 km (2 miles) NNE (22°) from Hawthorne, CA
- 7 km (4 miles) ENE (72°) from El Segundo, CA
- 15 km (10 miles) SSW (213°) from Los Angeles Civic Center, CA
|
Location Uncertainty |
horizontal +/- 0.4 km (0.2 miles); depth +/- 0.9 km (0.6 miles) |
Parameters |
Nph=139, Dmin=7 km, Rmss=0.42 sec, Gp= 40°,
M-type=local magnitude (ML), Version=C |
Source |
|
Event ID |
ci10410337 |
|
Wilcoxon 731A seismic accelerometers and Guralp CMG-40T-old seismometer at magnitude 5 and 4 erthquakes |
Attachment 1: eq5m4.jpg
|
|