40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 323 of 326  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  8912   Tue Jul 23 20:41:40 2013 gautamConfigurationendtable upgradeFull range calibration and installation of PZT-mounted mirrors

 Given that the green beam is to be used as the reference during the vent, it was decided to first test the PZT mounted mirrors at the X-endtable rather than the Y-endtable as originally planned. Yesterday, I prepared a second PZT mounted mirror, completed the full range calibration, and with Manasa, installed the mirrors on the X-endtable as mentioned in this elog. The calibration constants have been determined to be (see attached plots for aproximate range of actuation):

M1-pitch: 0.1106 mrad/V

M1-yaw: 0.143 mrad/V

M2-pitch: 0.197 mrad/V

M2-yaw: 0.27 mrad/V


Second 2-inch mirror glued to tip-tilt and mounted:

  • The spot sizes on the steering mirrors at the X-end are fairly large, and so two 2-inch steering mirrors were required.
  • The mirrors already glued to the PZTs were a CVI 2-inch and a Laseroptik 1-inch mirror.
  • I prepared another Laseroptik 2-inch mirror (45 degree with HR and AR coatings for 532 nm) and glued it to a PZT mounted in a modified mount as before.
  • Another important point regarding mounting the PZTs: there are two perforated rings (see attached picture) that run around the PZT about 1cm below the surface on which the mirror is to be glued. The PZT has to be pushed in through the mount till these are clear of the mount, or the actuation will not be as desired. In the first CVI 2-inch mirror, this was not the the case, which probably explains the unexpectedly large pitch-yaw coupling that was observed during the calibration [Thanks Manasa for pointing this out]. 

Full range calibration of PZT:

Having prepared the two steering mirrors, I calibrated them for the full range of input voltages, to get a rough idea of whether the tilt varied linearly and also the range of actuation. 

Methodology:

  • The QPD setup described in my previous elogs was used for this calibration. 
  • The linear range of the QPD was gauged to be while the output voltage lay between -0.5V and 0.5V. The calibration constants are as determined during the QPD calibration, details of which are here.
  • In order to keep the spot always in the linear range of the QPD, I stared with an input signal of -10V or +10V (ie. one extreme), and moved both the X and Y micrometers on the translational stage till both these coordinates were at one end of the linear range (i.e -0.5V or 0.5V). I then increased the input voltage in steps of ~1V through the full range from -10V to +10V DC. The signal was applied using a SR function generator with the signal amplitude kept to 0, and a DC offset in the range -5V to 5V DC, which gave the desired input voltages to the PZT driver board (between -10V DC and 10V DC).
  • When the output of the QPD amp reached the end of the linear regime (i.e 0.5V or -0.5V), I moved the appropriate micrometer dial on the translational stage to take it to the other end of the linear range, before continuing with the measurements. The distance moved was noted. 
  • Both the X and Y coordinates were noted in order to investigate pitch-yaw coupling.

Analysis and remarks:

  • The results of the calibration are presented in the plots below. 
  • Though the measurement technique was crude (and maybe flawed because of a possible z-displacement while moving the translational stage), the calibration was meant to be rough, and I think the results obtained are satisfactory. 
  • Fitting the data linearly is only an approximation, as there is evidence of hysteresis. Also, PZTs appear to have some drift, though I have not been able to quantify this (I did observe that the output of the QPD amp shifted by an amount equal to ~0.05mm while I left the setup standing for an hour or so).  
  • The range of actuation seems to be different for the two PZTs, and also for each degree of freedom, though the measured data is consistent with the minimum range given in the datasheet (3.5 mrad for input voltages in the range -20V to 120V DC). 

 

PZT Calibration Plots

The circles are datapoints for the degree of freedom to which the input is applied, while the 'x's are for the other degree of freedom. Different colours correspond to data measured with the position of the translational stage at some value.

                                            M1 Pitch                                                                                             M1 Yaw

M1_Pitch_calib.pdf     M1_Yaw_calib.pdf

 

                                              M2 Pitch                                                                                        M2 Yaw 

M2_Pitch_calib.pdf     M2_Yaw_calib.pdf

 



Installation of the mirrors at the X-endtable:

The calibrated mirrors were taken to the X-endtable for installation. The steering mirrors in place were swapped out for the PZT mounted pair. Manasa managed (after considerable tweaking) to mode-match the green beam to the cavity with the new steering mirror configuration. In order to fine tune the alignment, Koji moved ITMx and ETMx in pitch and yaw so as to maximise green TRX. We then got an idea of which way the input pointing had to be moved in order to maximise the green transmission.

 

Attachment 5: PI_S330.20L.pdf
PI_S330.20L.pdf
  8930   Sun Jul 28 19:39:04 2013 AnnalisaUpdateendtable upgradePicture

 Yend table picture updated on the wiki page

  8932   Mon Jul 29 13:39:25 2013 gautamConfigurationendtable upgradePZT Driver Board-further changes

 

 

I have updated the schematic of the D980323 PZT driver boards to reflect the changes made. The following changes were made (highlighted in red on the schematic):

  • Gain of all four HV amplifier stages changed from ~15 to ~5 by swapping 158k resistors R43, R44, R69 and R70 for 51k resistors.
  • Electrolytic 10 uF capacitors C11, C12, C29 and C31 swapped for 470pF, 500V mica capacitors.
  • Fixed resistor in voltage divider (R35, R40, R59 and R64) replaced with 0 ohm resistors so as to be able to apply a bias of -10V to the HV amplifier
  • The DC-DC Series components, which I think were originally meant to provide the 100V DC voltage, have been removed.
  • The path between the point at which +100V DC is delivered and jumpers J3 and J6 has been shorted (bypassing R71 and R11 for J3, R73 and R12 for J6).
  • Tantalum capacitors C38 and C39 have been replaced with electrolytic capacitors (47 uF, 25V). One of the original tantalum capacitors had burned out when I tried installing the board in the eurocrate, shorting out -15V to ground. At Koji's suggestion, I made this switch. The AD797s do not seem to be oscillating after the switch.


I have also changed the routing of the 100V from the HV power supply onto the board, it is now done using an SMA T-connector and two short lengths of RG58 cable with SMA connectors crimped on.

The boards are functional (output swings between 0 and 100V as verified with a multimeter for input voltages in the range -10V to +10V applied using a function generator.

 



Revised schematics:

D980323-C-modified.pdf

D980323-C-modified-pg2.pdf

 

 

 

  8935   Mon Jul 29 21:57:45 2013 gautamConfigurationendtable upgradeHardware installed at 1X9

 The following hardware has been installed on rack 1X9;

  • KEPCO high voltage power supply (kept in a plastic box at the bottom of the rack, with the 3m SMA cable carrying 100V running along the inside side wall of the rack). The HV supply has not been connected to the driver board yet.
  • AI board D000186 installed in top eurocrate. The board does not seem to fit snugly into the slot, so I used a longish screw to bolt the front panel to the eurocrate.
  • PZT driver board D980323 installed in top eurocrate adjacent to the AI board.
  • Six 11m SMB-LEMO cables have been laid out from 1X9 to the endtable. I have connected these to the PZT driver board, but the other end (to the PZTs) is left unconnected for now. They have been routed through the top of the rack, and along the cable tray to the endtable. All the cables have been labelled at both ends. 


I have also verified that the AI board is functional in the eurocrate by using the LEMO monitoring points on the front panel.


The driver boards remain to be verified, but this cannot be done until we connect the HV supply to the board. 

 

 

  8942   Tue Jul 30 19:40:47 2013 gautamConfigurationendtable upgradeDAC-PZT Driver Board Output Signal Chain Tested

 

 [Alex, Gautam]

The signal chain from the DAC output to the output of the PZT driver board (including the HV supply) has been verified. 

I had installed the two boards in the eurocrate yesterday and laid out the cables from 1X9 to the endtable. The output of the AI board had been verified using the monitor port on the front panel, but the output from the PZT driver board was yet to be checked because I had not connected the HV supply yesterday.

When I tried this initially today, I was not getting the expected output from the monitor channels on the front panel of the PZT driver board, even though the board was verified to be working. Alex helped debug the problem, which was identified as the -15V supply voltage not making it onto the board.

I changed the slot the board was sitting in, and used a long screw to bolt the board to the crate. Both the AI board and the PZT driver board seem to be slightly odd-sized, and hence, will not work unless firmly pushed into the eurocrate and bolted down. This would be the first thing to check if a problem is detected with this system. 

In any case, I have bolted both boards to the eurocrate, and the output from the PZT driver board is as expected when I sent a 10Vp sine wave out from the DAC. I think the cables can now be hooked up to the PZTs once we are pumped down.

  8943   Tue Jul 30 19:44:05 2013 gautamConfigurationendtable upgradeSecond mirror glued to PZT and mounted

 

 I have glued a fourth mirror to a PZT (using superglue) and inserted it into a modified mount. This is to be used together with the 1-inch Laseroptik mirror I had glued a couple of weeks back at the Y-endtable. I will be calibrating both these mirrors tonight such that these are ready to put in as soon as we are pumped down.

The mirror was one of those removed from the X-endtable during the switch of the steering mirrors. It is a CVI 2-inch mirror, with HR and AR coatings for 532 nm. 

  8967   Mon Aug 5 18:48:44 2013 gautamConfigurationendtable upgradeFull range calibration of PZT mounted mirrors for Y-endtable

 I had prepared two more PZT mounted mirrors for the Y-end some time back. These are:

  • A 2-inch CVI mirror (45 degree, HR and AR for 532nm, was originally one of the steering mirrors at the X-endtable, and was removed while switching those out for the PZT mounted mirrrors).
  • A 1-inch Laseroptik mirror (45 degree, HR and AR for 532nm).

I used the same QPD set-up and the methodology described here to do a full-range calibration of these PZTs. Plots attached. The calibration constants have been determined to be:

CVI-pitch: 0.316 mrad/V

CVI-yaw:  0.4018 mrad/V

Laseroptik pitch: 0.2447 mrad/V

Laseroptik yaw:  0.2822 mrad/V

Remarks:

  • These PZTs, like their X-end counterparts, showed evidence of drift and hysteresis. We just have to deal with this.
  • One of the PZTs (the one on which the CVI mirror is mounted) is a used one. While testing it, I thought that its behaviour was a little anomalous, but the plots do not seem to suggest that anything is amiss.

Plots:

                                                        CVI YAW                                                                                                                         CVI PITCH

2-inch-CVI-Yawcalib.pdf      2-inch-CVI-Pitchcalib.pdf

                                                        Laseroptik YAW                                                                                                             Laseroptik PITCH

1-inch-Laseroptik-Yawcalib.pdf   1-inch-Laseroptik-Pitchcalib.pdf

 

  9451   Mon Dec 9 11:44:08 2013 SteveUpdateendtable upgradeETMY optable got new beam dump

Aluminum shield replaced by razor beam dump.

Attachment 1: ETMYoptBefore.jpg
ETMYoptBefore.jpg
Attachment 2: ETMYoptAfter.jpg
ETMYoptAfter.jpg
  9941   Mon May 12 14:42:25 2014 steveUpdateendtable upgradeoptical table enclosure wall proposal and table at ETMY

Quote:

 Sanwiched wall as shown: 1" clear acrylic, 2 layers of 0.004" thick "window tint", 1 layer of  0.007" thermashield  and  0.125" yellow acrylic

Visibility: 70 %,    Transmission of 1064 nm  2-3 % at 0-50 degrees incident,   power density  ~ 0.7 W/mm2

Max power 100 mW

        

More details about this east end  " acrylic + "   enclosure ( optical table cover ) can be found elog entry 6210, 7194 and  7106

 

Window tinted layer transmission plot is below.

 

We have a film which may meet your requirements and the values are shown below:

         

               Wavelength (nm)              Transmission                     Reflectance(front)            Reflectance (back)

 

                        1060                               .0772                               .604                                     .759

 

                        1070                               .0723                               .615                                     .772

 

These values are taken from the LBNL Optics 6 program and if you have access to that program, the NFRC ID for the film is 202.   If you do not have access to the program, I have a captured the graph which may be of some help.  I apologize for the appearance of the graph but someone at LBNL decided it would look better with a dark gray background – the yellow is the transmission curve, the blue is the reflectance (front) and the green is the reflectance (back).

 

 

The film is referred to as “Hilite 70” and has a 72% visible light transmission.  These results were obtained with the film mounted on 1/8” clear glass.

 

I

 

 

 

 

 

Saint-Gobain Performance Plastics

www.solargard.com

 

Please consider your environmental responsibility before printing this email.

 

 

 

Attachment 1: Hilite70.jpg
Hilite70.jpg
Attachment 2: ETMY-ISCT.jpg
ETMY-ISCT.jpg
Attachment 3: RoscoThermashield.pdf
RoscoThermashield.pdf
  11591   Fri Sep 11 10:56:47 2015 SteveUpdateendtable upgradeETMY optical table feedthrough

The ETMY enclosure feedthrough - north is installed. The sealing material is hard to work with.

The upper empty blocks will be replaced by something soft to make changing cables easy.

 

Attachment 1: ETMY-Nfeedt.jpg
ETMY-Nfeedt.jpg
  11600   Tue Sep 15 16:49:08 2015 SteveUpdateendtable upgradeETMY optical table feedthrough

ETMY optical table enclosure feedthrough- south is in. Now it is time to see how air tightness increases performance.

Quote:

The ETMY enclosure feedthrough - north is installed. The sealing material is hard to work with.

The upper empty blocks will be replaced by something soft to make changing cables easy.

 

 

Attachment 1: ETMYsFeedt.jpg
ETMYsFeedt.jpg
  11608   Thu Sep 17 02:22:53 2015 SteveUpdateendtable upgradeETMY optical table feedthrough

I doubt we'll see any effect until we carefully seal the holes. If there's 1 hole in your boat it still sinks.

Quote:

ETMY optical table enclosure feedthrough- south is in. Now it is time to see how air tightness increases performance.

 

  11897   Tue Dec 22 16:39:42 2015 SteveUpdateendtable upgradeETMY optical table enclosure

I think there should be a scientifically based aveluation of the ETMY enclosure so we can make the ETMX better.

Meanwhile I'm counting pieces to move on with the south end table cover.

 

Attachment 1: IMG_0028.JPG
IMG_0028.JPG
Attachment 2: IMG_0027.JPG
IMG_0027.JPG
Attachment 3: ETMY-ISCT_EISOL.jpg
ETMY-ISCT_EISOL.jpg
  12029   Thu Mar 10 16:29:32 2016 gautamUpdateendtable upgradeInventory check

I did a quick sweep of the lab to find out what hardware has already been acquired for the X-end table upgrade. The attached PDF is an inventory check in the spirit of this elog.

Some things we have to decide:

  • Are we okay with using the old green coloured faraday mount for the IR faraday? I have in hand a piece identical to the one used at the Y-end for the green faraday, that is red in colour, so I guess we can switch this out.
  • The way in which the doubling oven is currently mounted at the X-end is using some posts cobbled together. The Y-end looks to have a custom mount machined for it (see Attachment #2). Do we want to go ahead and get something like this done?
  • I suppose it is okay to reuse all the old optics (mirrors, lenses, harmonic separators) and PDs? It may be that we need to order som extra mirrors/lenses/posts (this will become clear once I do the layout)

I have not gotten around to planning the layout or doing drawings. I will try and first work through a mode-matching solution to make sure we have all the required lenses. It may be that we need some 1" or 2" mirrors as well. The beam from the lightwave NPRO is quite elliptical, but we have a number of cylindrical lenses in hand already if we decide we want to use these, so I guess we don't have to worry about this...

This is quite a preliminary list, and I will add/update over the coming days as I do more detailed planning, but have I missed out anything obvious?

Attachment 1: Inventory_check.pdf
Inventory_check.pdf
Attachment 2: Doubler_comparison.pdf
Doubler_comparison.pdf
  12030   Thu Mar 10 16:32:45 2016 ranaUpdateendtable upgradeInventory check

Its not a good idea to use green mounts with green lasers. Steve should be able to get another copy of the EY doubler mount made up if we really don't have another one sitting in the Manasa end table box which Koji mentioned.

  12033   Mon Mar 14 22:42:23 2016 gautamUpdateendtable upgradeInventory check
Quote:

Steve should be able to get another copy of the EY doubler mount made up if we really don't have another one sitting in the Manasa end table box which Koji mentioned.

I located the second doubler mount, it was sitting inside a cabinet along the Y-arm. So this will not have to be machined. The doubling oven mount is black in colour.

So as things stand now, the only thing that needs to be machined is a non-green mount for the IR faraday (IO-5-1064-HP) - is it possible to just coat the existing mount with a different color? I've got a drawing for this part ready, but it seems unnecessary to machine the whole thing from scratch when only the color is an issue. Steve was talking about dipping this in some sort of solution and taking the green off. But if this isn't possible, I'll send Steve the drawings tomorrow so that he can place the order with the machine shop...

I will work on the mode-matching calculations over the next couple of days to make sure we have all the mirrors and lenses we need.

 

  12051   Mon Mar 28 10:43:18 2016 SteveUpdateendtable upgradeETMX 4'x2' optical table pictures

 

 

Attachment 1: ETMX4x2Layout.jpg
ETMX4x2Layout.jpg
Attachment 2: ETMX_4x2.JPG
ETMX_4x2.JPG
Attachment 3: ETMXopt_4x2.JPG
ETMXopt_4x2.JPG
  12056   Wed Mar 30 17:38:52 2016 gautamUpdateendtable upgradeX end table proposed layout

Attachment 1: This is a photo of the current X end table optical layout with the beampaths of the various sub-systems overlaid. For the labels, see Attachment #2.

Attachment 2: This is a summary of all the optical components that are currently being used. I've noted some things we may want to change when we effect the swap. The important ones are:

  • Switch out all 1" and 2" optic mounts which are not of the Polaris type to the Polaris type. I have checked that we have sufficient numbers of these in hand.
  • Adjust the collimating lens of the fiber collimating telescope to get a better mode
  • Many of the labels are probably outdated, now would be a good time to update them
  • For the mode-matching of the AUX IR into the doubling crystal, a la mode suggests a better (i.e. less sensitive to lens position) solution is effected with L2 as a 100mm fl lens rather than 88.3mm. I did not change this during the laser swap in order to minimize the number of components changed. Since we are doing a wholesale change now, it may not be a bad idea to swap this out as well. I have checked that we have a suitable AR1064 coated lens.
  • Some optics probably need to be cleaned...
  • PZT mirror 2 has a new mount ready that is the "correct" height so we don't have to keep using makeshift stacked posts.
  • The plan as it stands is to use the green coloured mount for the IR faraday (IO-5-1064-HP).

Have I missed anything important?

Attachment #3: I've made a CAD drawing of the proposed new layout and have overlaid the beampath in an amateur way because I couldn't figure OptoCad out - I figure this will suffice for now. I have adopted elements from the current Y-end layout, but have used Anders' mode-matching solution (same lenses, same positions of optics) to make sure we have good Guoy phase separation between the two PZT steering mirrors. Some notes:

  • I've tried to palce the optics for the AUX IR into the doubler and subsequent steering of green into the arm cavity as per the mode matching solution. These should be pretty accurate, and the layout suggests we have some room to maneuver 
  • The Green REFL beampath is exaggerated but I think we have enough room to place Y16 appropriately and steer the reflected beam into the PDA36A
  • We need two more 1" 1064nm coated mirrors for the initial steering into the doubling oven, I have checked we have these in hand.
  • The IR pickoff into the fiber coupler may change somewhat once we change the mode and redo the mode-matching calculations. But again, I think we have sufficient room to implement a workable solution.
  • After accounting for the fact that the new endtable will be a little closer to the vacuum chamber, Y12 in the proposed layout will be ~10cm further away from ETMX than it is currently. But as discussed at the meeting today, the Rayleigh range of the green beam should be large enough here such that this shouldn't be a significant change.

Steve says the table is ready - so if we are happy with this layout, we can move forward...

Attachment 1: ETMX_3x2.JPG
ETMX_3x2.JPG
Attachment 2: layout_details_20160328.pdf
layout_details_20160328.pdf
Attachment 3: ETMX_proposed_layout.pdf
ETMX_proposed_layout.pdf
  12057   Thu Mar 31 09:38:41 2016 SteveUpdateendtable upgradeY end 4x3 existing layout

Beam colors: 1064 nm red, 514 nm green and 633 nm yellow.

There should be room for lens in front of the pd at red3 and a mirror for alignment in the new layout.

This picture may help you how to improve the new ETMX 4' x 3' optical layout.

 

Attachment 1: ETMY4x3layout.jpg
ETMY4x3layout.jpg
  12058   Thu Mar 31 19:49:31 2016 gautamUpdateendtable upgradeproposed layout v2

The major changes from the previous layout:

  1. I've depicted the Green reflected beam path more accurately - I approximately measured the angle of the rejected beam from the faraday from the Y-end setup. This looks like a workable solution, and is similar to what we have currently at the Y-end
  2. I've added some optics to monitor the DC power and RIN of the AUX laser
  3. I've added two lenses to the input path of the Oplev beam (the path is such that I think we can use the same lenses that are currently being used. 
  4. I've now drawn the beams in CAD so that is marginally neater.

To do:

  1. Post mode matching solutions for AUX laser to doubler and green beam to arm for this proposed layout (should be identical to what we have now, which at least according to the calculation is a good solution, but I will double check - I also need to quantify what the effect of the elliptical beam is)
  2. Check the Gouy phase of the transmitted IR beam at the QPD - we may need to change some lenses in this path. But I think the path as such is close enough (distance-wise) to what we have currently at the X end (after accounting for the fact that the new endtable edge will be closer to the ETM) so I don't expect this to be a show-stopper.

Does any part of this layout need a radical redesign? 

Attachment 1: ETMX_proposed_layout_v2.pdf
ETMX_proposed_layout_v2.pdf
  12060   Mon Apr 4 10:59:12 2016 gautamUpdateendtable upgradeproposed layout v3

I realized I had overlooked an important constraint in the layout, which is that the enclosure will have two supports that occupy some region of the table - these are denoted in blue in v3 of the layout (Attachment #1). I measured the dimensions for these from the existing Y-endtable. The main subsystem this has affected is the IR transmission monitors, but I've been able to move the photodiodes a little to accommodate this constraint.

I've also done the mode-matching calculations explicitly for the proposed new layout (Attachments #2 and #3, code in Attachment #4). While the layout was largely adopted from what Andres posted in this elog, I found that some of the parameters he used in his a la mode code were probably incorrect (e.g. distance between the 750mm lens and the ETM). More critically, I think the Gouy phase for the optimized solution in the same elog is more like 60 degrees. I found that I could get a (calculated) Gouy phase difference between the two PZT mirrors of ~81 degrees by changing the green path slightly, and making the two PZT mirrors Y7 and Y8 (instead of Y7 and Y11, for which the Gouy phase difference is more like 50 degrees). But this way the two steering mirrors are much closer to each other than they were before. Other misc. remarks about the mode matching calculations:

  • The beam diameter at the locations where the Faraday isolators should go is well below 5mm, the aperture size of the Faraday isolators
  • The calculated mode-matching efficiencies suggest that we don't need any cylindrical lenses though the mode from the NPRO is elliptical
  • Attachment #5 is a CAD drawing of the layout with all dimensions used for the mode-matching calculations included (although they are in inches)

These changes also necessitated minor changes to the transmitted IR beampath and the Oplev system, but these changes are minor. I've also switched the positions of the AUX IR power monitoring PD and the fiber coupler as suggested by Koji. The shutter has also been included. 

Attachment 1: ETMX_proposed_layout_v3.pdf
ETMX_proposed_layout_v3.pdf
Attachment 2: IR_modematch.pdf
IR_modematch.pdf
Attachment 3: Green_modematch.pdf
Green_modematch.pdf
Attachment 4: XendALaMode.zip
Attachment 5: ETMX_proposed_layout_dimensions.dwg
  12061   Mon Apr 4 15:04:14 2016 gautamUpdateendtable upgradeCOMPONENT REMOVAL


I'm planning to start removing components from the X endtable tomorrow morning at ~10AM - if anyone thinks I should hold off and do some further checks/planning, let me know before this so that I can do the needful.

  12063   Tue Apr 5 11:42:17 2016 gaericqutamUpdateendtable upgradeTABLE REMOVAL

There is currently no table at the X end!

We have moved the vast majority of the optics to a temporary storage breadbord, and moved the end table itself to the workbench at the end. 

Steve says Transportation is coming at 1PM to put the new table in.

  12065   Wed Apr 6 17:52:21 2016 gautamUpdateendtable upgradeFirst contact cleaning commenced

I've begun cleaning the optics that will eventually go back onto the newly installed X-endtable. We decided that First Contact was the way to go (as opposed to methanol drag wiping). Koji demonstrated the application of the (red) First Contact solution onto a 2" mirror - I then proceeded to work on the rest of the optics. We are broadly following the procedure in E1000079 - first one coat of First Contact solution is applied, then a small piece of PEEK is embedded by applying a second layer of solution over it (this will enable us to pull off the First Contact once we are ready - the plan is to do this after roughly placing the optic on the table. As of now, I've finished coating most of the optics that are part of the IR Transmon path - I will continue later in the evening.

The new endtable is almost ready for re-population. Steve just needs to shim the enclosure which will be done tomorrow morning. The game-plan as discussed at the meeting today is to first try and set up the IR Transmon path. This will allow us to verify that the endtable height is such that we can maintain a beam height of 4" everywhere on the table (I suspect we may have to compromise at some poing and do some fine adjustment of 1/4 to 1/2" somewhere though). It will also allow me to define the cavity axis relative to the table, which will be useful to place the green steering optics eventually. Doing this will be challenging though as right now, I can't see any of the arm flashes on the endtable using an IR card. Ideally, we want to somehow lock the X arm and then do the checks mentioned at the endtable, before beginning to put the endtable back together. 

  12066   Thu Apr 7 12:51:24 2016 gautamUpdateendtable upgradeBeam height differences

Steve has finished installing the enclosure on the new endtable. So Eric and I decided to try and lock the X arm and measure the beam height of the transmitted IR beam relative to the endtable. We initially thought of using POX DC as a the LSC trigger but this did not work as there was no significant change in it when the arm was flashing. Eric then tried misaligning the ITM and using AS110 as a trigger - this worked. We then recompiled the ASS model to take AS110 as an input, and ran the dither alignment. After doing so, I measured the beam height at two points on the new endtable.

Bottom line:

  • The beam is roughly level across the table (along the North-South direction, within the precision to which I could place the irides and measure the height). The table has also been levelled pretty well...
  • The beam height is ~4.7" across the endtable

So the beam is about 0.7" higher relative to the endtable than we'd like it to be. What do we do about this?

  • Is it even possible to raise the table by 0.7" so we can have a level beam everywhere? Are there some constraints related to how the enclosure is attached to the window?
  • Are we okay with tolerating a solution where we keep the beam level at 4", and use Y10 and Y11 (see layout in elog 12060) to raise the beam by 0.7", and then have slightly higher posts for the optics downstream of this point?

I've also placed two irides extending the cavity axis on the endtable. These should be helpful in aligning the green to the arm eventually.

  12067   Thu Apr 7 14:30:58 2016 SteveUpdateendtable upgradeETMX table height

 

Quote:

Steve has finished installing the enclosure on the new endtable. So Eric and I decided to try and lock the X arm and measure the beam height of the transmitted IR beam relative to the endtable. We initially thought of using POX DC as a the LSC trigger but this did not work as there was no significant change in it when the arm was flashing. Eric then tried misaligning the ITM and using AS110 as a trigger - this worked. We then recompiled the ASS model to take AS110 as an input, and ran the dither alignment. After doing so, I measured the beam height at two points on the new endtable.

Bottom line:

  • The beam is roughly level across the table (along the North-South direction, within the precision to which I could place the irides and measure the height). The table has also been levelled pretty well...
  • The beam height is ~4.7" across the endtable

So the beam is about 0.7" higher relative to the endtable than we'd like it to be. What do we do about this?

  • Is it even possible to raise the table by 0.7" so we can have a level beam everywhere? Are there some constraints related to how the enclosure is attached to the window?
  • Are we okay with tolerating a solution where we keep the beam level at 4", and use Y10 and Y11 (see layout in elog 12060) to raise the beam by 0.7", and then have slightly higher posts for the optics downstream of this point?

I've also placed two irides extending the cavity axis on the endtable. These should be helpful in aligning the green to the arm eventually.

The new TMC 4' x 3' x4" optical table and enclosure is installed - aligned- leveled.

Atm2,  Picture is taken ~42" from the window at 3.75 camera height. The leveled table height is wthin 1/4 at the center of the window.

I think this is close enough to move on with the installation of the optics.

We can raise the loaded table in the future if it is needed.

Atm4, Optical table height to floor 33" at the south west corner

Atm3, Enclosure top cover transmission at 1064 nm, 1mm beam size, power level 157 mW, 0 degree incident angle,   T 1.3% Metal shield is required above 100 mW hitting the wall of the enclosure!

Atm5, window to enclosure Kapton seal

Attachment 1: ETMXbrs.png
ETMXbrs.png
Attachment 2: ETMX.jpg
ETMX.jpg
Attachment 3: IMG_0618.JPG
IMG_0618.JPG
Attachment 4: ETMX.jpg
ETMX.jpg
Attachment 5: ETMXwc.jpg
ETMXwc.jpg
  12068   Fri Apr 8 09:13:20 2016 SteveUpdateendtable upgradeETMX-T beam height

X arm resonating after alignment, beam height on ETMX optical table ~4.75"

 

Attachment 1: ETMXalignedT.png
ETMXalignedT.png
Attachment 2: ETMXF.jpg
ETMXF.jpg
  12073   Wed Apr 13 00:56:07 2016 gautamUpdateendtable upgradeX endtable repopulation

Over the last couple of days, I've been working on restoring the optical layout on the X-endtable. Some notes about the status as of today:

Lightwave NPRO output power

The output power from the lightwave NPRO is about 210mW (as measured with the calorimeter). This is significantly lower than the value of ~300mW reported in this elog. It may be that the laser crystal temperature has changed compared to that measurement, but the "ADJ" parameter is at 0, both today and in that measurement. The laser has also been on for more than a day now, that should be sufficient time for the crystal to equilibriate to its final operating state? Is such a large change in output power possible just because of a change in laser crystal temperature? Or did the laser really lose ~1/3rd of its output power over the last two months?

Alignment into IR Faraday, and changes to the planned layout

I've set up the layout until steering the beam through the IR faraday. The input power into the IR Faraday is ~210mW. The output power is ~186mW, after optimizing the angle of the HWP. These numbers seem consistent with what I had reported in this elog (although this was for the Innolight NPRO). The alignment looks reasonably good to the eye as well.

I've made one change to the planned layout (latest version here). Y1 is now a 2" 99% reflective for S polarization beam splitter, instead of a 1" HR mirror. I made this change because we want some light from the NPRO to be transmitted through this optic to couple into the fiber eventually, for the IR beat. I measured the transmitted power to be ~1.5mW, which is around what we were coupling into the fiber before, and should suffice now. The Lightwave NPRO datasheet (page 4) suggests that the polarization of the output of the laser is S, and the measured power before and after this optic suggests that it is working as advertised. This means that HWP 1 also has to be moved downstream (to rotate the polarization so as to maximize transmission through the IR faraday). Space constraints meant that I could not mount HWP 1 on the baseplate+3/4" OD post assembly which is what we want where possible on the new table, so for this optic, I used a 1" OD post and a fork. There may be a couple of other optics in the final layout where space constraints dictate we compromise in this way.

I've also installed beam dumps for the rejected light from the Faraday. For now, these are the old beam dumps. They looked reasonably intact. I believe we have a bunch of new beam dumps on hand as well, so these can be swapped out if deemed necessary.

Cleaning of optics

All the optics are being cleaned using first contact before being installed on the table. 

As I found out the hard way, it is not a good idea to clean small optics like half-wave plates while in their mounts. The first contact tends to bond to the frame while drying, and doesn't come off cleanly. Koji helped me clean the offending pieces (he used tweezers to manually remove the residual first contact, and then some acetone to clean up any remaining residue). Subsequently, he re-cleaned these optics, again using first contact, but this time being careful not to extend all the way out to the edge of the optic. The idea is to cover as much area as possible with first contact, while staying clear of the edge. This approach worked reasonably well.


The next major step is to achieve optimal alignment into the doubler. I've placed the doubler on the table in it's approximate final position, I wanted to make sure the enclosure support wasn't in the way (it isn't). The cable from the oven won't run all the way to the Thorlabs temperature controller in it's usual place, we need to either extend the cable, or figure out a new place where we can keep the temperature controller.

  12074   Wed Apr 13 11:00:28 2016 SteveUpdateendtable upgradeX endtable repopulation

ETMX optical table is grounded to ETMX chamber through 1 Mohms

The doubling oven temp controller is installed to reach its cable.

 

  12075   Wed Apr 13 18:25:07 2016 gautamUpdateendtable upgradeLightwave health check

[Koji,gautam]

Lightwave NPRO information:

Model: 126-1064-700

Serial Number: 337

Manufactured: December 1998!!

Details of checks performed:

Koji tuned the parameters on the laser controller and we observed the following:

  1. Turning "ADJ" to +10 and the pumping current all the way up to the maximum (2.62A) allowed us to recover an output power of 300mW, at a laser crystal temperature of ~45degrees
  2. The output power increased almost monotonically as a function of the laser crystal temperature - why? We were able to see powers as high as 250mW (at ADJ=0) for the maximum crystal temperature of ~60 degrees.
  3. We checked that we could believe the readout of the power meter by measuring the power using the Scientech power meter - we saw ~270mW after the Faraday with this meter, accounting for ~10% loss through the Faraday, this corresponds to an output power of 300mW (all this was done at ADJ=+10, DC=2.62A). I suspect that the display is dodgy though, because changing the Diode Current from 2.52A to 2.62A increased the output power by almost 100mW, which seems hard to believe?
  4. The Lightwave NPRO does not have heat dissipation fins attached - could this be affecting the power output somehow? In any case, this has to be rectified. So if we decide to keep the Lightwave NPRO, the layout will still need minor changes to accommodate the heat fins. Steve, do we have these in hand?

Way forward

Ericq has begun the characterization of the repaired Innolight. We checked that it outputs 1W of power. We will now have to perform the following measurements:

  • Frequency noise using PLL
  • AM/PM response of the PZT
  • Laser power output as a function of diode current - this will be useful for diagnostic purposes in the future
  • AUX temperature vs PSL temperature at which beatnotes can be found
  • Waist measurement - the mode matching and optical layout upstream of the doubling oven at least will have to be modified significantly

All of these will have to be done before installing this laser at the endtable.

I believe the consensus as of now is to go ahead with carrying out the above measurements. Meanwhile, we will keep the Lightwave NPRO on and see if there is some miraculous improvement. So the decision as to whether to use the Innolight is deferred for a day or two.

  12079   Fri Apr 15 18:38:12 2016 gautamUpdateendtable upgradeLightwave health check - NO IMPROVEMENT

I re-measured the power levels today.

We have ~205mW out of the NPRO, and ~190mW after the Faraday. It doesn't look like the situation is going to improve dramatically. I'm going to work on a revised layout with the Innolight as soon as I've profiled the beam from it, and hopefully, by Monday, we can decide that we are going ahead with using the Innolight.

  12083   Tue Apr 19 18:37:29 2016 gautamUpdateendtable upgradeLaser swap + optical layout

Summary of work done over the last two days

  1. Lightwave NPRO + controller moved to PSL table
    • ​​The interlock is not connected to the controller
    • Controller is not powered
  2. Innolight NPRO + controller installed at endtable
    • ​​​​Interlock has been connected
    • For initial alignment purposes, I'm running it at an injection current of 1.000A (~50mW of IR out of the NPRO)
    • Temperature of crystal set to 31.66 degrees in anticipation of operation in the nominal state
  3. Laying out optics
    • ​​Given that the mode out of the NPRO is different from that from the Lightwave, the mode-matching had to be re-done
    • Attachment #1 shows the mode-matching solution being implemented
    • Current state - I've placed all the optics up to and including the doubling crystal + oven. Alignment through IR Faraday is pretty good, QWP+HWP angles optimized to maximize transmission through the Faraday (<10% loss). Oven has been hooked up to temperature controller, and is currently set to 36.3 degrees. Coarse alignment into doubling crystal done at lower power. Even with the low IR power, I am able to see some green. It remains to turn the injection current up and do the fine alignment + lens position tweaking to maximize the green power from the doubling crystal - with ~1W of power, assuming 2%/W SHG efficiency, we should be seeing 20 mW of green (which is probably way too much)

Immediate next steps:

  1. Some optimization to be done with regards to beam dumps for rejected beam from IR Faraday. Also double check to make sure that the reflected beam from L1 doesn't go back directly to the laser (at the moment it doesn't, is there a standard way to do this? I was trying to have the lens as close to normal incidence as possible, but I may not have been entirely successful which is why the reflected beam is not going straight back at the moment).
  2. Optimize mode-matching into the doubling crystal
  3. Once the desired green mode is obtained, continue with the rest of the layout
  4. Update CAD drawing to reflect new layout

 

Attachment 1: IR_modematch_19April2016_2.pdf
IR_modematch_19April2016_2.pdf
  12085   Thu Apr 21 14:25:52 2016 gautamUpdateendtable upgradeGreen light recovered

I've made progress on the new layout up to the doubling oven. After doing the coarse alignment with the diode current to the NPRO at ~1A, I turned it back up to the nominal 2A. I then rotated the HWP before the IR Faraday such that only ~470mW of IR power is going into the doubler (the rest is being dumped on razor beam dumps). After tuning the alignment of the IR into the doubling oven using the steering mirror + 4 axis translation stage on which the doubling oven is mounted, I get ~3.2mW of green after the harmonic separator and a HR mirror for green. The mode looks pretty good to the eye (see attachment #1), and the conversion efficiency is ~1.45%/W - which is somewhat less than the expected 2%/W but in the ballpark. It may be that some fine tweaking of the alignment + polarization while monitoring the green power can improve the situation a little bit (I think it may go up to ~4mW, which would be pretty close to 2%/W conversion efficiency). The harmonic separator also seems to be reflecting quite a bit of green light along with IR (see attachment #2) - so I'm not sure how much of a correction that introduces to the conversion efficiency. 

While doing the alignment, I noticed that some amount of IR light is actually transmitted through the HR mirrors. With ~500mW of incident light at ~45 degrees, this transmitted light amounts to ~2mW. Turns out that this is also polarization dependant (see attachment #3) - for S polarized light, as at the first two steering mirrors after the NPRO, there is no transmitted light, while for P-polarized light, which is what we want for the doubling crystal, the amount transmitted is ~0.5%. The point is, I think the measured levels are consistent with the CVI datasheet. We just have to take care find all these stray beams and dump them.

I will try and optimize the amount of green power we can get out of the doubler a little more (but anyway 3mW should still be plenty for ALS). Once I'm happy with that, I will proceed with laying out the optics for mode-matching the green to the arm.

Attachment 1: IMG_6567.JPG
IMG_6567.JPG
Attachment 2: IMG_6568.JPG
IMG_6568.JPG
Attachment 3: CVI_reflectivity.jpeg
CVI_reflectivity.jpeg
  12089   Tue Apr 26 15:22:35 2016 SteveUpdateendtable upgradeCleaning ETMX vacuum dirty window

Gautom is progressing with the layout nicely. The X-arm transmission window have not seen cleaning for decades. This should be the time to do it. Here is picture of dirtiness.

It is not that simple... How much effort should we put in it? The hole table with 1W inno laser plus... set up now about ~500 lbs  We can pull it off carefully, but it is not risk free.

We should look at our other signal port windows! Gautom's long reach able him to do the first contact cleaning without moving anything. It is great!
 

Attachment 1: ETMX-Tvp.jpg
ETMX-Tvp.jpg
Attachment 2: ETMX-TvpDetail.jpg
ETMX-TvpDetail.jpg
  12090   Tue Apr 26 23:19:42 2016 gautamUpdateendtable upgradeGreen aligned to arm - high order mode flashes seen

Attachment #1

Layout as of today. Most of the green path is done. The Green REFL PD + PZT mirrors have not been hooked up to their respective power sources yet (I wonder if it's okay to start laying cables through the feedthroughs on either end of the table already, or if we want to put whatever it is that makes it airtight eventually in first?). A rough power budget has been included (with no harmonic separator just before the window), though some optimization can be done once the table is completely repopulated.

Attachment #2

A zoomed-in version of the REFL path.

Some general notes:

  1. I've tried to use the custom 3/4" O.D. posts + baseplate arrangement wherever possible (only 1 steering mirror is on a 1" post clamped with a fork to the table because of space constraints). Where the baseplates could not be bolted onto the table directly, I've used Newport SS Dogs to do the job.
  2. I checked for continuity between the PZT outer case and the table top with a multimeter, and found none. So I chose to leave the Thorlabs baseplates in place. For the REFL PD, I've used an insulating baseplate given to me by Steve.
  3. I've used some custom length 3/4" O.D. posts to get the beam up to the right height (~4.75") just before sending the green beam in. The beam height is 4" elsewhere.
  4. I was playing around with positioning the harmonic separator immediately before the vacuum chamber window - I found that there is a substantial amount of green light that is reflected, though there doesn't seem to be any IR leaking through. The mirror was labelled Y1-1037-45P, which is a code for CVI mirrors, though I believe it is a LaserOptik product and that we have a couple of other such mirrors in the optics cabinet - though they are all 1". This document suggests that from the back side, there should be <0.1% reflection of green while on the front side it should be < 3%. I will have to hunt a little more for the specs, and measure the powers to see if they match the previously quoted numbers. In any case, I'll have to think of how to separate the (unwanted) reflected green and the transmitted IR from the cavity in the IR transmon path.
  5. There are some minor changes to the planned layout posted here - I will update these in due course once the Transmon path and Oplev have been set up.

I am closing the PSL shutter and the EX laser shutters for the night as I have applied a layer of first contact to the window for cleaning purposes, and we don't want any laser light incident on it. It may be that the window is so dirty that we may need multiple F.C. cleaning rounds, we will see how the window looks tomorrow...

 

Attachment 1: IMG_2219.JPG
IMG_2219.JPG
Attachment 2: IMG_2220.JPG
IMG_2220.JPG
  12094   Wed Apr 27 15:04:47 2016 SteveUpdateendtable upgradeCleaning ETMX vacuum dirty window

It looks very promising.

 

Attachment 1: 1cETMX-Tcmp.jpg
1cETMX-Tcmp.jpg
  12095   Thu Apr 28 00:41:08 2016 gautamUpdateendtable upgrademore progress - Transmon PD installed

The IR Transmon system is almost completely laid out, only the QPD remains to be installed. Some notes:

  1. The "problem" with excessive green power reflected from the harmonic separator has been resolved. It is just very sensitive to the angle of incidence. In the present configuration, there is ~10uW of green power reflected from either side, which shouldn't be too worrisome. But this light needs to be dumped. Given the tiny amount, I think a black glass + sticky tape solution is best suited, given the space constraints. This does not reach the Transmon PDs because there is a filter in the path that is transmissive to IR only. 
  2. I aligned the transmitted beam onto the Thorlabs PD, and reconnected the signal BNC cable (the existing cable wasn't long enough so I had to use a barrel connector and a short extension cable). I then reverted the LSC trigger for the X arm back to TRX DC and also recompiled c1ass to revert to TRX for the dither alignment. At the moment, both arms are stably locked, although the X arm transmission is saturated at ~0.7 after running the dither alignment. I'm not sure if this is just a normalization issue given the new beam path or if there is something else going on. Further investigations tomorrow.
  3. It remains to dump some of the unwanted green light from the addition of the harmonic separator...
  4. We may want to redesign some (or all) of the Transmon path - the lens currently in use seems to have been chosen arbitrarily. Moreover, it is quite stubbornly dirty, there are some markings which persist after repeated first contact cleaning...

I feel like once the above are resolved, the next step would be to PDH lock the green to the arm and see what sort of transmission we get on the PSL table. It may be the polarization or just alignment, but for some reason, the transmitted green light from the X arm is showing up at GTRY now (up to 0.5, which is the level we are used to when the Y arm has green locked!). So a rough plan of action:

  1. Install transmon QPD
  2. PDH lock green to X arm
  3. Fix the window situation - as Steve mentioned in an earlier elog, the F.C. cleaning seems to have worked well, but a little remains stuck on the window (though away from where any laser radiation is incident). This is resolved easily enough if we apply one more layer of F.C., but the bottle-neck right now is we are out of PEEK which is what we use to remove the F.C. once dried. Steve thinks a fresh stock should be here in the next couple of days...
  4. Once 3 is resolved, we can go ahead and install the Oplev.
  5. Which leaves the lst subsystem, coupling to the fiber and a power monitor for the NPRO. I have resolved to do both these using the 1% transmitted beam after the beamsplitter immediately after the NPRO rather than pick off at the harmonic separator after the doubling oven. I need to do the mode-matching calculation for coupling into the fiber and also adjust the collimating lens...
  6. Clean-up: make sure cables are tied down, strain-relieved and hooked up to whatever they are supposed to be hooked up to...
  12099   Fri Apr 29 00:55:46 2016 gautamUpdateendtable upgradegreen PDH locked to Xarm

Using the modulation frequency suggested here, I hooked up the PDH setup at the X-end and succeeded in locking the green to the X arm. I then rotated the HWP after the green Faraday to maximize TRX output, which after a cursory alignment optimization is ~0.2 (I believe we were used to seeing ~0.3 before the end laser went wonky). Obviously much optimization/characterization remains to be done. But for tonight, I am closing the PSL and EX laser shutters and applying first contact to the window once more courtesy more PEEK from Koji's lab in W Bridge. Once this is taken care of, I can install the Oplev tomorrow, and then set about optimizing various things in a systematic way.. MC autolocker has also been disabled...

Side note: for the IR Transmon QPD, we'd like a post that is ~0.75" taller given the difference in beam height from the arm cavity and on the endtable. I will put together a drawing for Steve tomorrow..

  12100   Fri Apr 29 16:05:23 2016 gautamUpdateendtable upgradeCleaning ETMX vacuum dirty window

After a second round of F.C. application, I think the window is clean enough and there are no residual F.C. pieces anywhere near the central parts of the window (indeed I think we got most of it off). So I am going to go ahead and install the Oplev. 

Quote:

It looks very promising.

 

 

Attachment 1: IMG_0755.JPG
IMG_0755.JPG
  12104   Mon May 2 19:14:18 2016 gautamUpdateendtable upgradeOptical layout almost complete

With Steve's help, I installed the Oplev earlier today. I adjusted the positions of the two lenses until I deemed the spot size on the QPD satisfactory by eye. As a quick check, I verified using the DTT template that the UGF is ~5Hz for both pitch and yaw. There is ~300uW of power incident on the QPD (out of ~2mW from the HeNe). In terms of ADC counts, this is ~13,000 counts which is about what we had prior to taking the endtable apart. There are a couple of spots from reflections off the black glass plate in the vacuum chamber, but in general, I think the overall setup is acceptable.

This completes the bulk of the optical layout. The only bits remaining are to couple the IR into the fiber and to install a power monitoring PD. Pictures to follow shortly. 

Now that the layout is complete, it remains to optimize various things. My immediate plan is to do the following:

  1. Maximize green transmission by tweaking alignment. I should also do a quick check using mirror specs to see that the measured transmitted green power compares favourably to what is expected.
  2. Check the green PDH loop transfer function at the X end - this will allow me to set the gain on the uPDH box systematically.
  3. Re-establish green beats, check noise performance.
  4. There are possibly multiple beam dumps that have to be installed. For now, I've made sure that no high power IR beams are incident on the enclosure. But there are a couple of red and green beams that have to be accounted for.

I will also need to upload the layout drawing to reflect the layout finally implemented.


Not directly related:

The ETMx oplev servo is now on. I then wanted to see if I could lock both arms to IR. I've managed to do this successfully - BUT I think there is something wrong with the X arm dither alignment servo. By manually tweaking the alignment sliders on the IFOalign MEDM screen, I can get the IR transmission up to ~0.95. But when I run the dither, it drives the transmission back down to ~0.6, where it plateaus. I will need to investigate further. 

 

GV Edit: There was some confusion while aligning the Oplev input beam as to how the wedge of the ETM is oriented. We believe the wedge is horizontal, but its orientation (i.e. thicker side on the right or left?) was still ambiguous. I've made a roughly-to-scale sketch (attachment #1) of what I think is the correct orientation - which turns out to be in the opposite sense of the schematic pinned up in the office area.. Does this make sense? Is there some schematic/drawing where the wedge orientation is explicitly indicated? My search of the elog/wiki did not yield any..

Attachment 1: ETMX_wedge.pdf
ETMX_wedge.pdf
  12105   Thu May 5 03:05:37 2016 gautamUpdateendtable upgradeALS status update

[ericQ, gautam]

Today we spent some time looking into the PDH situation at the X end. A summary of our findings.

  1. There is something that I don't understand with regards to the modulation signal being sent to the laser PZT via the sum+HPF pomona box - it used to be that with 2Vpp signal from the function generator, we got ~5mVpp signal at the PZT, which with the old specs resulted in a modulation of ~0.12rad. Now, however, I found that there was a need to place a 20dB attenuator after the splitter from the function generator in order to realize a modulation depth of ~0.25 (which is what we aim for, measured by locking to the TEM00 modes of the carrier and sidebands and comparing the ratio of powers). It could be that the PZT capacitance has changed dramatically after the repair. Nevertheless, I still cant reconcile the numbers. We measured the transfer function from the LO input of the pomona box to the output with the PZT connected, and figure there should be ~70dB of attentuation (with the 20dB additional attenuator in place). But this means 1Vpp*0.0003*70rad/V = 0.02rad which is an order of magnitude away from what the ratio of powers suggest. Maybe the measurement technique was not valid. In any case, this setup appears to work, and I'm also able to send +7dBm to the mixer which is what it wants (function generator output is 3Vpp).
  2. In addition to the above, I found that the demodulated error signal had a peak-to-peak of a few volts. But the PDH servo is designed to have tens of mV at the input. Hence, it was necessary to turn down the gain of the REFL PD to 10dB and add a 20dB attenuator between mixer output and servo input.
  3. While Johannes and I were investigating this earlier in the afternoon, we found that the waveform going to the laser PZT was weirdly distorted (still kind of sinusoidal in shape, but more rounded, I will put up a picture shortly). This may not be the biggest problem, but perhaps there is a better way to pipe the LO signal to the PZT and mixer than what is currently done.
  4. We then looked at loop transfer function and spectrum of the control signal. Plots to follow. They look okay.
  5. I measured the green power coming onto the PSL table. It is ~400uW. After optimizing alignment, the green transmission is ~0.4 according to whatever old normalization we are using.
  6. We then recovered the X green beatnote and looked at the ALS noise spectrum. Beatnote amplitude at the beat PD is ~ -27dBm. The coherence in the region of a few hundred Hz suggests that some improvements can be made to the PDH situation (the gain of the PDH servo is maxed out at the X end at the moment...). But the bottom line is this is probably good enough to get back to locking...
Attachment 1: ALS_noiseSpec_5May2016_2.pdf
ALS_noiseSpec_5May2016_2.pdf
Attachment 2: Coherence_5May2016.pdf
Coherence_5May2016.pdf
Attachment 3: image.jpeg
image.jpeg
  12108   Thu May 5 14:05:01 2016 ranaUpdateendtable upgradeALS status update

All seems very fishy. Its not good to put attenuators and filters in nilly-willy.

  1. Once the post-PD bandpass has been designed and constructed, you should be able to use whatever PD gain setting gives you the best SNR. There's no need to use more PD gain than necessary; it just reduces the PD bandwidth. What is the input referred current noise of the PD at the different gain settings?
  2. The open loop mixer output *should* be very large. It should be reduced to mV only when the loop is closed.
  3. The better way to estimate the modulation depth is to lock the arm on red as usual and then scan the EX laser and look at the green transmission. The FSR is 3.7 MHz, so the SBs should show up well in a narrow scan around the carrier.
  4. I guess its going to be tough to impedance match the splitter box to the NPRO PZT, since its impedance is all over the place at 200-300 kHz, but you could put a 50 Ohm in-line terminator in there somewhere?
  5. The Bode plot seems to indicate that we could easily get a 10 kHz UGF and then switch on a Boost. Is the remote Boost switch disabled or always ON? I am suspicious of the plot and think that the coarse trace is probably missing some sharp resonances which will sneakily bite you.
  12109   Thu May 5 21:28:44 2016 gautamUpdateendtable upgradeInnolight PZT capacitance

I suggested in an earlier elog that after the repair of the NPRO, the PZT capacitance may have changed dramatically. This seems unlikely - I measured the PZT capacitance with the BK Precision LCR meter and found it to be 2.62 nF, which is in excellent agreement with the numbers from elogs 3640 and 4354 - but this makes me wonder how the old setup ever worked. If the PZT capacitance were indeed that value, then for the Pomona box design in elog 4354, and assuming the PM at ~216kHz which was the old modulation frequency was ~30rad/V as suggested by the data in this elog, we would have had a modulation depth of 0.75 if the Function Generator were set to output a Signal at 2Vpp (2Vpp * 0.5 * 0.05 * 30rad/V = 1.5rad pp)! Am I missing something here?

Instead of using an attenuator, we could instead change the capacitor in the pomona box from 47pF mica to 5pF mica to realize a modulation depth of ~0.2 at the new modulation frequency of 231.25 kHz. In any case, as elog 4354 suggests, the phase introduced by this high-pass filter is non-zero at the modulation frequency, so we may also want to install an all-pass filter which will allow us to control the demodulation phase. This should be easy enough to implement with an Op27 and passive components we have in hand...

 

  12122   Thu May 19 16:29:20 2016 SteveUpdateendtable upgradeOptical layout almost complete

 

 

Attachment 1: ETMX_4x3_closed.jpg
ETMX_4x3_closed.jpg
Attachment 2: sealedETMXenclosure.jpg
sealedETMXenclosure.jpg
  12526   Fri Sep 30 19:53:07 2016 gautamUpdateendtable upgradeX end IR pickoff fiber coupled

[johannes, gautam]

Today we re-installed the fiber coupler on the X-endtable to couple some of the PSL light into a fiber that runs to the PSL table, where it is combined with a similar PSL pickoff to make an IR beat between the EX AUX laser and the PSL. The main motivation behind this was to make the process of finding the green beatnote easier. We used JAMMT (just another mode matching tool) to calculate a two lens solution to couple the light into the collimator - we use a +200mm and -200mm lens, I will upload a more detailed mode matching calculation + plot + picture soon. We wanted to have a beam waist of 350um at the collimator, a number calculated using the following formula from the Thorlabs website:

d =4\lambda (\frac{f}{\pi*MFD})

where d is the diameter of the output beam from the collimator, f is the collimating lens focal length and MFD is 6.6um for the fiber we use.

There is ~26mW of IR light coming through the BS after the EX AUX - after playing around with the 6 axis stage that the coupler is mounted on, Johannes got the IR transmission to the PSL table up to ~11.7mW. The mode matching efficiency of 45% is certainly not stellar, but we were more curious to find a beat and possibly measure the X arm loss so we decided to accept this for now - we could probably improve this by moving the lenses around. We then attenuated the input beam to the fiber by means of an ND filter such that the light incident on the coupler is now ~1.3mW, and the light arriving at the PSL table from the EX laser is ~550uW. Along with the PSL light, after the various couplers, we have ~500uW of light going to the IR beat PD - well below its 2mW threshold.

The IR beat was easily found with the frequency counter setup. However, there was no evidence of a green beat. So we went to the PSL table and did the near-field-far-field alignment onto the beat PD. After doing this, we were able to see a beat - but the amplitude was puny (~-60dBm, we are more used to seeing ~-20dBm on the network analyzer in the control room). Perhaps this can be improved by tweaking the alignment onto the PD while monitoring the RF output with an oscilloscope.

Moreover, the green PDH problems with the X end persist - even though the arm readily locks to a TEM00 mode, it frequently spontaneously drops lock. I twiddled around with the gain on the uPDH box while looking at the error signal while locked on a oscilloscope, but was unable to mitigate the situation. Perhaps the loop shape needs to be measured and that should tell us if the gain is too low or high. But ALS is getting closer to the nominal state...

Johannes is running his loss measurement script on the X arm - but this should be done by ~10pm tonight.

 

  12532   Wed Oct 5 16:28:10 2016 gautamUpdateendtable upgradeEX laser power monitor PD installed

I installed a 10% BS to pick off some of the light going to the IR fiber, and have added a Thorlabs PDA55 PD to the EX table setup. The idea is to be able to monitor the power output of the EX NPRO over long time scales, and also to serve as an additional diagnostic tool for when ALS gets glitchy etc. There is about 0.4mW of IR power incident on the PD (as measured with the Ophir power meter), which translates to ~2500 ADC counts (~1.67V as measured with an Oscilloscope set to high impedance directly at the PD output). The output of the PD is presently going to Ch5 of the same board that receives the OL QPD voltages (which corresponds to ADC channel 28). Previously, I had borrowed the power and signal cables from the High-Gain Transmon PD to monitor this channel, but today I have laid out independent cabling and also restored the Transmon PD to its nominal state.

On the CDS side of things, I edited C1SCX to route the signal from ADC Ch28 to the ALS block. I also edited the ALS_END library part to have an additional input for the power monitor, to keep the naming conventions consistent. I have added a gain in the filter module to calibrate the readout into mW using these numbers. The channel is called C1:ALS-X_POWER_OUT, and is DQed for long-term trending purposes.

The main ALS screen is a bit cluttered so I have added this channel to the ALS overview MEDM screen for now..

  438   Tue Apr 22 22:19:02 2008 robMetaphysicslorejiggling sliders

In the interests of tacit communication of scientific knowledge, I here reveal a nugget of knowledge which may or may not prove useful to new LIGOites: sometimes when front-end machines are rebooted, the hardware they control can wind up in a state which is not accurately represented by the EPICS values you may see. This can be easily rectified by momentarily changing the EPICS settings in question. For reference, this came up tonight in the context of the whitening gain sliders for the TransMon QPDs.
  1104   Sun Nov 2 20:21:58 2008 ranaConfigurationloreHP 5550dtn (Grazia) set up on allegra
I set up printing to grazia from allegra. The CUPS interface was not as straightforward as Tobin had made it seem in the Wiki. I had to type in the IP address and port number by hand.

The steps (AFAIR):
1) Goto http://localhost:631/
2) Click on "Add Printer"
3) Choose HP JetDirect
4) Use the correct address (socket://131.215.115.220:9100)
5) Choose HP and the 5550 postscript driver as the options
6) Try to only print useful stuff and not kill too many trees.
  1217   Thu Jan 8 16:49:37 2009 ranaConfigurationloreHP 5550dtn (Grazia) set up on allegra

Quote:
I set up printing to grazia from allegra. The CUPS interface was not as straightforward as Tobin had made it seem in the Wiki. I had to type in the IP address and port number by hand.

The steps (AFAIR):
1) Goto http://localhost:631/
2) Click on "Add Printer"
3) Choose HP JetDirect
4) Use the correct address (socket://131.215.115.220:9100)
5) Choose HP and the 5550 postscript driver as the options
6) Try to only print useful stuff and not kill too many trees.


It ought to be root to do that.
  1579   Wed May 13 02:53:12 2009 robSummaryloreChannel Hopping: That ancient enemy (MC problems)

We were stymied tonight by a problem which began late this afternoon.  The MC would periodically go angularly unstable, breaking lock and tripping the MC2 watchdogs.  Suspicion fell naturally upon McWFS.

Eventually I traced the problem to the MC3 SIDE damping, which appeared to not work--it wouldn't actually damp, and the Vmon values did not correspond to the SDSEN outputs.  Suspicion fell on the coil driver.

Looking at the LEMO monitors on the MC3 coil driver, with the damping engaged, showed clear bit resolution at the 100mV level, indicating a digital/DAC problem.  Rebooting c1sosvme, which acquires all the OSEM sensor signals and actually does the side damping, resolved the issue. 

ELOG V3.1.3-