  1071   Thu Oct 23 11:26:25 2008 steveBureaucracyVACoptosigma forks for vacuum
Nine short forks from OptoSigma were cleaned, baked and measured by Bob T.
They passed the LIGO vac compatiblity test as Rana predicted.
Remember: Thorlab and Newfocus forks failed this qualification process.
  1082   Fri Oct 24 11:09:08 2008 steveUpdateSAFETYmore lexan plates under cameras
The MC2, MC3&1 and BSC-SUS cameras were repositioned somewhat in the
process of placing lexan disks under neat them.
MC1&3 will have to be readjusted.

Now all horizontal viewports are protected.
  1085   Fri Oct 24 15:05:13 2008 steveUpdateVACrga is out of order
The old Dycor RGA is out of order.
I'm getting ready to purchase an SRS instrument.
  1094   Mon Oct 27 11:23:10 2008 steveUpdatePhotosnew Olympus camera with IR vision
The IR blocker was removed from our new Olympus camera
SP 570UZ camera.
It has image stabilization, zoom 26-520 mm (20x optical)
and 10.7 Mpixel
Attachment 1: IRisin.JPG
IRisin.JPG
  1103   Fri Oct 31 08:52:19 2008 steveUpdateSUSETMY damping restored
ETMY watchdogs were tripped yesterday morning also. This is a problem that existed some months ago.(?)
Our seimo channels are inactive.
Attachment 1: etmy2d.jpg
etmy2d.jpg
  1107   Mon Nov 3 09:59:47 2008 steveUpdatePSLPSL HEPAs turned on
The psl enclosure HEPAs were tuned on.

Loose paper drawing was found on the psl inside shelf.
This can fall down into the beam and ignite a tragedy.

Thanks for the color coded correction. My spell checker is not reliable
  1116   Thu Nov 6 09:45:27 2008 steveUpdateMOPAhead temp hick-up vs power
The control room AC temp was lowered from 74F to 70F around Oct 10
This hold the head temp rock solid 18.45C for ~30 days as it shows on this 40 days plot.
We just had our first head temp hick-up

note: the laser chiller did not produce any water during this period
Attachment 1: htpr.jpg
htpr.jpg
  1117   Thu Nov 6 10:06:41 2008 steveUpdateLockingarms lock degradation
I have been locking the arms in the mornings lately.
The daily drift of LSC-TRX is ~ 15% and LSC-TRY ~5%
Attachment 1: arms.jpg
arms.jpg
  1120   Fri Nov 7 08:08:00 2008 steveUpdatePEMAC turned off in IFO room
All tree AC units in IFO room 104 switched off at 7:45am
I'm looking for the periodic thump signal in the Vertex area
noted in entry 1113 of PEM-ACC_MC1
  1121   Fri Nov 7 10:52:57 2008 steveUpdatePEMAC is back on
The 20 minutes and 6 sec thumps are not related to the 40m ac units
This period is precisely right on so it is not likely mechanical as a source.
The time and frequency domain of this signal at attachment 2&3

First I had the chilling water cut off and later I turned off the fans
as 5 hrs temp plot shows
Attachment 1: acf.jpg
acf.jpg
Attachment 2: acc_mc1.jpg
acc_mc1.jpg
Attachment 3: test.jpg
test.jpg
  1123   Fri Nov 7 16:05:55 2008 steveBureaucracySAFETYinsect killer sprayed at kitchen area !

Bob and I cleaned out the sink area and sprayed
Spectracide's BUG STOP insect killer solution on the shelfs and sink
table top area.

NO eating or coffee drinking till Monday

This is an effort to stop the ants coming.
  1127   Mon Nov 10 16:44:14 2008 steveUpdatePEMparticle counter at MC2
The particle counter was moved to MC2 table temporarily.
It is clamped to the table at the south west corner.
It is pumping for 20s and waiting for 2 minutes now.
  1129   Mon Nov 10 19:39:40 2008 steveUpdatePEMthump noise from particle counter
The particle counter is back on the IOOC location on a piece of FOAM
It needs this isolation, so when the pump is running, it's not shaking things.
The counter was counting for 6 sec and it was on holding for 20 mins.
Now I set the counter for 20 sec so it is easy to recognise it's signal and it holds for 2min only.
This will set the alarm handler in action.

Atm1: 40 mins plot
PEM-ACC_MC2_x,y,z up to 13 mins: pcounter at MC2 table, clamped, counting for 20s and holds for 2 mins
PEM-ACC_MC2_x,y,z from 13 to 26 mins: pcounter at MC2 table, not clamped, seated on 2" foam, counting 20s and holds for 2 mins
PEM-ACC_MC1_x,y,z from 26 to 40 mins: pcounter at MC1_IOOC location, not clamped, seated on 2" foam, counting 20s and holds for 2 mins

Rana won the bet



Quote:
I'll bet Steve a dollar that it is mechanical. The attached PDF compares all of the accelerometers from right now.

You can see that the RMS in MC2 is way bigger than MC1.

In the second PDF file you can see the time series. I had to play around a lot with DTT to get it to work. The DTT/Foton
combo on Allegra is not stable, so make sure your work early and often.

In the plots shown, I am bandpassing the time series from 600-700 Hz. I found that doing so allowed the burp in MC1 to remain
large and reduce the extraneous fuzz in MC2. As you can see there is no such noise in MC2.

So its a noise around 600-700 Hz that comes on quickly and then shuts off after several seconds. Its also very periodic in that
it comes on around every 20 minutes. Steve also tells me (although he refuses to put in the elog) that it started up around
August 20th (?). I feel like someone in the 40m lab ought to be able to guess what this is at this point.

Please convince Steve to elog his findings about when the noise started.

If one goes out there and stands next to it when the trend predicts its happening it becomes clear what it is.
Attachment 1: partcfm.jpg
partcfm.jpg
  1134   Fri Nov 14 11:33:19 2008 steveUpdateVACSRS-RGA installed
Old Dycor rga is removed and new SRS-RGA200 installed.
It is pumped down and ready for hooking up its RS-232 output for operation.
  1149   Thu Nov 20 09:37:39 2008 steveUpdatePSLMZ vs temp
This 12 days plot shows that it can hold lock if the daily temp variation is not more than 3/4 of a degree C
The MZ is happy if it's HV is above 100V
Attachment 1: mztemp.jpg
mztemp.jpg
  1197   Fri Dec 19 16:38:09 2008 steveUpdateLSCall optlevs centered
All optlevs were centered after full alignment.

Qpd sums are:
ETMX 12,229 counts
ITMX 9,932
ETMY 12,043
ITMY 4,362
BS 1,880
PRM 1,423
SRM 11,641
  1199   Sun Dec 21 13:00:06 2008 steveUpdateall down cond.vac and laser back on
There was a power outage sometimes early Saturday.

All things are down.

The Maglev was started and reached normal operating condition.
V1 was opened at P1 3.8 mTorr and cc1 is back to 3e-6 Torr now

The MOPA was turned on.
Ion pump HV on to FSS cavity.
  1219   Fri Jan 9 09:14:17 2009 steveUpdatePEMifo is recovered after eq
There is no obvious damage from last night earth quake.
All sus dampings were turned on, MC locked and the arms locked right on
Attachment 1: eq4.5sb.jpg
eq4.5sb.jpg
  1220   Fri Jan 9 16:52:18 2009 steveUpdateVACtp3 forline pump replaced
Alberto took 40m vacuum 101 class as we replaced the drypump at the annulus pump line.

He is still not authorized to use the monster 2 3/8" open end wrench that is 36" long.

The fore line pressure dropped to 20 mTorr from 1 Torr as the pumps were swapped.
Bob needs to be given credit for replacing the tip seal on this Varian SH-100 drypump
The ss-hose felt dry at the tp3 exhaust end but it was some what "teflon coated-placticky-
-almost oily" at intake end of the dry pump.
We'll have to replace this metal hose next time.

This is a reminder that the 40m vacuum operation is fully manual.
It requires two people to switch a vacuum valve and one of them has to be experienced.
  1226   Wed Jan 14 09:22:43 2009 steveHowToVACN2 supply pressure lowered for vac valves

Quote:
I've been replacing the N2 bottles recently.
I noticed that the consumption is too high. I had to replace them every two days.
Normally the interval is three or more days.
I suspect there is some leak in the line.

Strangely, it is always the left hand bottle which goes empty. The right hand bottle has been
there for more than a week at about 1000 psi.

We should check it when Steve is back.



All vac valves operated by N2
The two cylinders are located in entry room 103
There is an auto switch over valve between them to insure continuous supply.
The pressure regulator should be set 70 PSI on the gauge
This pressure we keep constant.
All vac valves will close in case of running out of N2 or losing ac power so
it is essential that one replaces empty N2 cylinder.

Simply disconnect large CGA580 fitting with a crescent wrench.
Swap in full cylinder from outside, reconnect fitting tightly and open cylinder valve.
Now you should be reading the full cylinder pressure.
Write each cylinder pressure and date-time on the board so one can see if there is a leak
Attachment 1: n280d.jpg
n280d.jpg
  1227   Wed Jan 14 10:59:06 2009 steveBureaucracyVACrga waiting to be "connected"
We have no RGA data since old Dycor passed away in the middle of Oct, 2008
New SRS-RGA200 was installed on the vac envelope on Nov 14, 2008
It is waiting to be software connected to the 40m logging
  1228   Wed Jan 14 15:53:32 2009 steveDAQPSLMC temperature sensor

Quote:
I added a channel for the temperature sensor on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder. The slow servo of the FSS had to get restarted, the reference cavity locked and so the PMC and MZ.


Where is this channel?
  1229   Thu Jan 15 09:19:32 2009 steveUpdateIOOMC locking
MC2 sus damping was found tripped at the morning the second time this week.

Damping was restored, ISS gain lowered to avoid saturation, MZ manually locked
and MC locking was back.
  1243   Thu Jan 22 08:34:10 2009 steveBureaucracySAFETYsafety training
Kakeru received 40m safety training yesterday.

He has no authorization to work alone in the high power beam path yet.
  1248   Fri Jan 23 10:00:21 2009 steveUpdatePSLPMC transmission is down
The PMC transmission is going down.
I have not relocked the PMC yet.
Attachment 1: pmc4d.jpg
pmc4d.jpg
  1273   Thu Feb 5 09:12:29 2009 steveBureaucracySAFETYsafety updates & fire alarm test

The fire alarm test was completed at 13:30 yesterday.

 

I updated the 40M Emergency Calling List by replacing Rob by Yoichy. The calling order: Vass,  Aso and Taylor.

Rana and Yoichy were added to the "Registered PSL Operator List"  and posted it in the lab. (not in the document, that should be up dated)

We are getting ready for the annual safety audit. It will be held next week  at 14:00 Friday, Feb 13, 2009

Please participate in the preparation by correcting it or just  tell me.

 

Rod Luna is organizing the pick up of the following old equipment:

HP laser jet  5000n, 3 of 19" 10 base-T network ports and  4 small hubs,

2 Sun monitors, 1 Viewsonic monitor, 7 keypads and

Hitachi scopes 2 of V-355, 2 of V-422, 1 of V-202 and 1 of V-6165

 

 

  1280   Fri Feb 6 14:49:31 2009 steveSummarySAFETYlaser inventory
40M Laser Inventory at Feb 05, 2009
 

1, Lightwave PA#102 @ 77,910 hrs 1064nm of 2.8W @ 27.65A
                   NPRO#206 @ 2.4A                    at PSL enclosure............"Big Boy" is waiting for to be retired but not now.

2, Lightwave  NPRO 1064nm of 700mW  #415   at AP table.......cavity length measurements of Alberto

3, CrystaLaser  IRCL-100-1064S, 1064nmS of 100mW  ,sn#IR81132 at east arm cabinet

4, CrystaLaser 1064nm of 180mW # -1274 flq at scattering setup.........flashlight quality

5, RF-PD tester 1064nm of 1.2mW @20mA at SP table

6, Lumix 800-1100nm of 500mW at east arm cabinet

7, JDS-Uniphase 633nmP of 4mW oplev sus laser at 5 places,
    plus four spares in east arm cabinet
 
 
The same information is posted at the 40M WIKI also
 
 
 
  1282   Fri Feb 6 16:23:54 2009 steveUpdateMOPAMOPAs of 7 years

MOPAs and their settings, powers of 7 years in the 40m

Attachment 1: 7ymopas.jpg
7ymopas.jpg
  1288   Tue Feb 10 10:48:48 2009 steveBureaucracySAFETYsafety glasses scanned
All 40m safety glasses are absorbent.
 
They were scanned by 180 mW  2mm spot size  of 1064 nm beam and  their  transmissions were measured
They showed no degradation.
 
UVEX 's  LOTG-YAG/CO2 glasses (fits over personal glasses) 11 pieces,
Glendale-XC by Bacou-Dalloz,  #1166218VLT68% , aerodynamic style, 7 pieces, (4 missing )
KG5- glasses , heavier- made out of glass, 3 pieces,
Thorlab's LG10, 1 piece
Old glasses by "Laser Glass" Nd-YAG, 2 pieces
Totaling 24
 
Atm.2  showing our muilty wavelength glasses of 1064 and 530 nm
Their transmittance for both wavelength were zero.
 
Note: if you have prescripton safety glasses I would like to scan it for you !
 
 
 

 

 

Attachment 1: sgscan1.png
sgscan1.png
Attachment 2: sgscan2.png
sgscan2.png
  1293   Wed Feb 11 11:39:17 2009 steveConfigurationGeneralvac envelope ground
The vacuum envelope was grounded from the IOO chamber to master ground north (under electric breaker panel L)
with 4GA  25ft red battery cable.

 

  1300   Fri Feb 13 08:38:03 2009 steveUpdateIOOMC2 damping restored
  1302   Fri Feb 13 16:30:49 2009 steveConfigurationGeneralstatus quo is disturbed

I have been getting ready for the annual safety inspection in the past 2-3 days.

Meaning cleaning up and disturbing the status quo on the floor  mostly under the optical tables and their surroundings.
For example: pd power supplies, He/Ne laser ps. and their positions.
BNC cables and ac power line positions can be different.
The new rule: no electronic equipment on the floor.
 
All electronic equipment were moved-placed into a plastic dish or tray.
  1319   Wed Feb 18 15:31:38 2009 steveBureaucracySAFETYemergency back-up lights checked
All 13 exit and  emergency back-up lights and their batteries were checked.
Two of  the batteries were replaced.
  1324   Thu Feb 19 11:51:56 2009 steveUpdateMOPAHTEMP variation is too much
The C1:PSL-MOPA_HTEMP variation is more than 0.5 C daily
Normally this temp stays well within 0.1 C
This 80 days plots shows that we have just entered this unstable region some days ago.
The control room temp set unchanged at 70 F, actual temp at ac-mon 69-70 with occasional peaks at 74 F
 
Water temp at chiller repeatedly around 20.6 C at 8 am
This should be rock solid at 20.00C +- 0.02C
 

 

Attachment 1: 80dhtemp.jpg
80dhtemp.jpg
  1332   Mon Feb 23 11:07:01 2009 steveBureaucracySAFETYKiwamu receives safety training

Osamu and Kiwamu received 40m safety training on Thursday, Feb 19, 2009

Kiwamu needs Osamu's close supervision at PSL enclosure and AP table
I hope they already read, understood and signed the laser SOP
  1347   Tue Mar 3 08:44:31 2009 steveUpdatePEMair cond. maintenance today
IFO room 104 air conditions will be shut down for maintenance today.
This should be finished by noon.
The temperature and particle count variation can be more than usual.
  1387   Wed Mar 11 16:41:22 2009 steveUpdateMOPAspare NPRO power

The spare M126N-1064-700,  sn 5519 of Dec 2006 rebuilt  NPRO's power output

 measured   750mW at DC2.06A with Ohpir meter.

Alberto's controller  unit 125/126-OPN-PS,  sn516m was disconnected from lenght measurment NPRO on the AP table.

5519 NPRO  was clamp to the optical table  without heatsink and it was on for 15 minutes.

  1411   Fri Mar 20 11:01:02 2009 steveUpdatePEMparticle counts are high

The outside particle counts for 0.5 micron are 3 million this morning at 9am. Low clouds, foggy condition with low inversion layer.

This makes the 40m lab 30-50K

I just turned on the HEPA filter at the PSL enclosure.

Please, leave it on high

 

Attachment 1: particles32d.jpg
particles32d.jpg
  1414   Fri Mar 20 15:54:29 2009 steveOmnistructureGeneral480V crane power switch on MEZ

CES Mezzanine is beeing rebuilt to accommodate our new neighbor: the 20ft high water slide...& .jacuzzi

All our ac power transformers are up there. Yesterday we labelled the power switch of 480VAC on the mezz

that we need to keep to run the 3 cranes in the lab.

  1420   Tue Mar 24 09:04:02 2009 steveUpdateSUS4.8 mag earthquake

SRM, ITMX, ETMX, ITMY and ETMY lost damping at 4:55am this morning from 4.8 magnitude earthquake.

Their damping were restored.

C1:SUS-ITMX_URSEN_OUTPUT swich was found in off position. It was turned on.

MZehnder  and MC were locked.

The WFS qpd spot needs recentering

  1427   Wed Mar 25 09:55:45 2009 steveUpdateIOOglitching sensors of MC

SUS-MC1_SENSOR_SIDE and SUS-MC2_SENSOR_UL are glitching

Yesterday's 4.8mag earthquake at Salton Sea is shown on Channel 1

Attachment 1: glitchesofMC.jpg
glitchesofMC.jpg
  1439   Sun Mar 29 13:44:27 2009 steveUpdateSUSETMY sus damping restored

ETMY sus damping was found to be tripped.

It was retored.

All fluorecent light were turned off. Please try to conserve some energy.

  1445   Mon Mar 30 15:51:27 2009 steveUpdateElectronicsHP4291A left the lab to be repaired

Eric Gustafson is handling the old HP4291A rehabilitation. Tarac picked both units up today.

March of 2008 Tucker Electronics failed to fix it's intermittent ~25MHz 0.5V oscillation at the swept sine output

See 40m-elog id:398 on 3-24-2008 by Rob Ward

 

  1447   Tue Mar 31 09:42:32 2009 steveUpdatePEMETMY sus damping restored again

The Caltech gasoline storage tank is being upgraded.

They are jack hammering and digging with bulldozer 50 yards south of  ETMY

  1448   Wed Apr 1 10:22:13 2009 steveUpdateVACRGA logging is working

Thanks to Joe B who made the SRS RGA working with linux

Last data file logged at 2008 Oct 24 with old Dycor unit

First data file logged at  2009 Feb 10 with SRS

 

Attachment 1: rga-090401.png
rga-090401.png
  1452   Fri Apr 3 10:01:50 2009 steveUpdateVACrgascan with temp plot

Rga scan of day 231 since pumpdown pd66-m-d231

m stands for maglev pumping speed, vacuum normal condition of valves,

cc4 cold cathode gauge at the rga location,

cc1 is real ifo pressure from the 24" tube at the pumpspool,

PEM-count temp: vac envelope temp at the top of IOO chamber

 

 

Attachment 1: pd66tempow.jpg
pd66tempow.jpg
Attachment 2: rga-090403scan.png
rga-090403scan.png
  1459   Wed Apr 8 09:36:20 2009 steveConfigurationVAC valve condition changed to BG to calibrate the RGA

Vacuum normal valve condition was changed to accommodate SRS-RGA calibration.

 

The annulus valves were closed to free TP3 to pump on the RGA
VM1 was closed to isolate the RGA from the IFO.
 
 
VM3 opened to TP3.
 
This condition is called back ground (BG) mode. It is where we can calibrate and see the RGA background without the IFO.
  1462   Thu Apr 9 11:27:19 2009 steveUpdateVACvac gauge reading problem

Cold cathode gauge CC4  is reading normal.

CC1 is glitching, it is probably dirty.

CC2 is fluctuating too much and it is cutting out for 6-7 minutes. It must be insulated by deposits and there is no emission current.

I think the same goes for P1

They will have to be replaced at the next vent

Attachment 1: vacgflsec.jpg
vacgflsec.jpg
Attachment 2: vacgflmin.jpg
vacgflmin.jpg
  1477   Mon Apr 13 08:59:57 2009 steveUpdatePSLmixers on order

Quote:

Quote:

I then changed the RF Output Adjust slider in increments of 0.5, and measured the peak-to-peak values on the scope. In the table and on the plots, I've taken into account the 12dB attenuation. i.e I actually measured 964mV, so 964mV*10^.6 = 3838mV.


3.8Vpp is about 16dBm.
The mixer for the PMC demodulator is level 23. So 16dBm is insufficient.
What is the level of the new mixer Steve ordered ? 13 ?



I ordered mixers level 13, 17 on Friday and level 23 now.
They should be here Tuesday

NOTE: level 23 power is illegal to use in the 40m lab
They get hot
  1486   Wed Apr 15 11:25:21 2009 steveConfigurationVAC RGA cal completed

Quote:

Vacuum normal valve condition was changed to accommodate SRS-RGA calibration.

 

The annulus valves were closed to free TP3 to pump on the RGA
VM1 was closed to isolate the RGA from the IFO.
 
 
VM3 opened to TP3.
 
This condition is called back ground (BG) mode. It is where we can calibrate and see the RGA background without the IFO.

 Vacuum valve configuration is back to VACUUM NORMAL  condition. RGA  calibration  completed.

RGA  scan attached is the backgroud of the rga with std cal leak open, sn 08581

Krypton at amu 84 and Argon at amu 40 are the cal signals.

 

Attachment 1: rga-090415-bg-d7-cald2.png
rga-090415-bg-d7-cald2.png