ID |
Date |
Author |
Type |
Category |
Subject |
721
|
Wed Jul 23 10:49:37 2008 |
Max | Update | Computer Scripts / Programs | Weekly Progress Report | This week I installed the magnetometer. The channels seem to be reading correctly. I'm back to working on noise budget and have added the MICH and will soon add the PRC source. The various source-specific scripts still need to be adjusted and the transfer functions remeasured since they do not match in any reasonable manner the SRD Rana put out in the e-log yesterday. |
720
|
Wed Jul 23 10:47:05 2008 |
Sharon | Update | | Weekly update | This week I spent some time with Alex who updated the adaptive code to save the filter's coeffs all the time, stop when I open the loop, and reload the latest coeffs. when I start it again.
The point was to minimize the adaptation rate. Unfortunately, seems it is making some filters go wild, so it is not in use yet.
After taking some more measurements with the adaptive filter running, I have noticed a new peak in the signal around 22-23 Hz. My first assumption was that this is caused due to internal resonance of MC1 (which is driven when the adaptive code is running, and not when it's not). Therefore, I drove MC1 without the adaptive filter looking for the same peak... which wasn't there.
This sent me back to the adaptive code... Seems there is a matrix in the simulink file of the adaptive filter which doesn't have an MEDM screen. I am now working on making this screen. Once I am done with that, and make sure there is correlation between simulink and MEDM, I'll keep on chasing the peak in the code. |
719
|
Wed Jul 23 01:42:26 2008 |
rana | Configuration | PSL | FSS RFPD: Examined, "repaired", and re-installed | Rob said that there might be something wrong with the FSS RFPD since the loop gain is so low.
Next time we should just use the Jenne laser on it in-situ and compare with our reference.
We had a 24.5 MHz LSC PD which Rob got from Sam. Sam got it from Rai. I gave it to Rai in Livingston
because it seemed suspicious. Seems fine now. This black box PD had a lower overall response than
the goldbox one we already had. The 2001-2005 era diodes which we got from the Canadian Perkin-Elmer
all had high capacitance and so that's not a surprise.
So the goldbox one was not broken totally.
I found that the offset came from a cracked capacitor. C25 was a yellow thru-hole ceramic 0.1 uF.
Its a surface mount board...don't know why this was like this but there's also no reason it should
have cracked unless it was soldered on with too much heat. I replaced it with a 0.47 uF ceramic
surface mount. Also R24 was a 20 Ohm resistor and L3 was not stuffed?? Removed R24 and put a 1 uH
inductor into L3. This is there so that the input to the MAX4107 is AC coupled.
However, the DC signal that Rob saw was actually because of the cracked C25. It had shorted and was
making a 25 mV signal at the input to the MAX4107 which has a gain of 10. This was producing ~165 mVdc
into a 50 Ohm load and so it could have saturated most mixers. The FSS board, however, has an overly
monstrous level 21 (I think) mixer and so this should not have been an issue. Maybe.
I was able to lock with the 24.5 MHz black box PD but it was not too hard to repair the gold box one
so I did. I tuned it so that the notch is truly at 43 MHz (2x the FSS 21.5 MHz modulation) but because
someone has done this using a hacky cap in parallel with the main PD, I am unable to get the resonant
peak to line up at 21.5 MHz. Its at 23 MHz instead. This loses us ~2 dB in signal. Since the frequency
is so low, we can increase the gain in the MAX4107 by another factor of 3 or so in the future.
So the PD is not our problem. Still worth verifying that the cable is good -- its around 10 miles long!!
And loops around in there with a bunch of other cables. We have an electronic phase shifter so this seems
totally misguided.
The other bad problem is that the mode matching is pretty horrible. Something like 1/3 of the carrier
power doesn't go into the cavity.
FSS TODO:
1) Check cable between RFPD and FSS box for quality. Replace with a good short cable.
2) Using a directional coupler, look at the RFPD output in lock on a scope with 50 Ohm term.
I suspect its a lot of harmonics because we're overmodulating to compensate for the bad
mode matching.
3) Purchase translation stages for the FSS mode matching lenses. Same model as the PMC lenses.
Fix the mode matching.
4) Get the shop to build us up some more bases for the RFPDs on the PSL such as we have for the LSC.
Right now they're on some cheesy Delrin pedestals. Too soft...
5) Dump the beam reflected off the FSS RFPD with a little piece of black glass or a razor dump.
Anodized aluminum is no good and wiggles too much.
The attached PDF shows photos of the old and new style PDs. One page 3 there's a wire that I soldered on
as a handle so that we can remove the RF can (occasionally people claim that soldering to the lid screws
up the magnetic shielding magic of the lid. use this as a litmus test of their electronics know-how; its
a tin can - not an orgone box). Pages 4 & 5 are the circuit before I soldered, page 6 the cap after I
tried to remove it, page 7 is the circuit after I put in the new cap, and page 8 is the schematic with
the mark up of the changes. |
718
|
Tue Jul 22 22:25:31 2008 |
rana | Update | LSC | Looptickle for existing 40m | John and I have adapted the Stefan-Looptickle model of the 40m upgrade to have the parameters of the old one.
(old one = what we have had for the last 4 years).
Its in the /cvs/cds/caltech/iscmodeling directory on the CDS computers but you can also check it out from the
MIT CVS repo; its part of the whole shebang.
It makes the attached theoretical NB. Feel free to modify it. |
717
|
Tue Jul 22 22:11:58 2008 |
Yoichi | Update | LSC | X-arm g factor measurement | Alberto, Yoichi
We measured the g factor of the X-arm by slightly shifting the 166MHz sideband frequency:
We first locked the X-arm to TEM00 mode. Then misaligned the ETMX in yaw a little bit until the transmitted power is a half of the normal value.
In this way, we can expect that TEM01 mode will be resonated in the arm if a sideband with a suitable frequency is introduced.
To add such a sideband, we used the 166MHz EOM. According to John's calculation (ELog entry 690), the TEM01 mode of the 166MHz upper sideband is only about 100kHz away from the resonance. So by changing the 166MHz modulation frequency, we should be able to see the 166MHz upper sideband resonating in the X-arm.
We used the 166MHz PD at the AS to find the resonance.
When we modulated the 166MHz RF frequency by +/- 100kHz, we could see spikes in the AS166_I signal.
Then we fine tuned the RF frequency slowly by hand to find the exact resonant frequency. At that time, the X-arm PDH servo was oscillating at ~480Hz.
So the resonance was determined by maximizing this signal in the AS166_I.
The 166MHz signal was originally at 165.977195 MHz. I found the resonance around 165.985MHz. It is surprisingly close to the original modulation frequency (only 7.3kHz away). This number yields the g factor of 0.626 and the transverse mode interval of 0.285*FSR. I used the arm length of 38.5750m in this calculation. Because of the 480Hz oscillation, it was difficult to precisely determine the resonant frequency. We will try this again tomorrow after mitigating the oscillation.
Although the resonance of the 166MHz upper sideband is located at a lower frequency in John's prediction, we found a resonance at a higher frequency.
This can be interpreted as the discrepancy between the actual g-factor and the designed g-factor.
To confirm what we saw was really an arm cavity resonance, we will try to do the same thing with the arm cavities all mis-aligned.
(We expect no signal in this configuration.)
Appendix: the expected signal from AS166 port when the 166MHz upper sideband passes by a resonance of the arm cavity.
Since the carrier is resonating in the cavity and kept there by the PDH feedback using 33MHz sideband, its phase is virtually fixed at the AS166 port. The lower sideband's phase also does not change much because it is off resonance. The upper sideband get a large phase change when approaching to the resonance. This effectively rotates the modulation angle of the 166MHz sidebands, which was orthogonal to the carrier when off resonance (i.e. phase modulation), to create 166MHz amplitude modulation. Because the sideband axis is rotated, the signal should appear both in I and Q phases. |
716
|
Tue Jul 22 16:50:09 2008 |
steve | Metaphysics | Environment | professorial clean up of work bench | Atm1: is showing the spiritual satisfaction after work bench clean up by the professor himself.
Atm2: some items are still waiting to be placed back to their location |
715
|
Tue Jul 22 13:16:09 2008 |
John, Rob | Update | PSL | FSS open loop transfer function |
Quote: | With the common gain slider maxed out the unity gain frequency is 58kHz.
The reference cavity refl diode appears to be okay. RF OUT/ TEST IN transfer function was normal.
There is a ~220mV offset in the RF out. We removed this using a coupler - no change. We also checked the
diode->FSS cable.
Tomorrow I'll take a closer look at the board. |
Should note that the UGF of 58kHz was measured with the test cable (from RFPD to board), so the demod phase was presumably sub-optimal. |
714
|
Tue Jul 22 13:15:14 2008 |
rob | Update | PSL | Note from R. Abbott re: the PMC |
Quote: | an email from Rich:Your PZT is broken.
R |
Quelle surprise
 |
713
|
Tue Jul 22 11:55:22 2008 |
rana | Update | PSL | Note from R. Abbott re: the PMC | an email from Rich:Your PZT is broken.
R |
712
|
Tue Jul 22 09:24:17 2008 |
steve | Update | PSL | laser power | Laser power reality of 120 days |
711
|
Tue Jul 22 03:03:22 2008 |
John, Rob | Update | PSL | FSS open loop transfer function | With the common gain slider maxed out the unity gain frequency is 58kHz.
The reference cavity refl diode appears to be okay. RF OUT/ TEST IN transfer function was normal.
There is a ~220mV offset in the RF out. We removed this using a coupler - no change. We also checked the
diode->FSS cable.
Tomorrow I'll take a closer look at the board. |
710
|
Mon Jul 21 19:55:16 2008 |
rana,jenne | Configuration | IOO | Noise in MC_F | Jenne put the MC board on extender today - its still that way but everything is probably connected (check AO).
We measured the TFs of the DAQ section for MC_F because of how everything looked wrong in the plots Jenne
put in the log earlier. Everything we measured today seemed to jive with the schematic. We also looked up
the original traveler for this board which Betsy filled in years ago: it also is in spec for the DAQ filtering.
So then I looked at the power spectrum of the output signal to the VCO. It had lots of HFC (high frequency crap).
I adjusted the parameters of the FSS (common gain, fast gain, RF phase ) and lowered the MC common gain. This
produced a global minimum in that 4D parameter space.
I think that basically, the FSS gain is too low even with the common gain slider maxed. Having the fast gain up
at 19 dB like I had left it was bad - even though it minimizes the PC control signal, it produces a lot of HFC
up around 100 kHz in MC_F. After John (finally) gets around to measuring the FSS loop we can figure out how to
better tune this. The MC gain then has to be tuned so as to best minimize the HFC given the new FSS gain; there's
basically no coupling from the MC gain to the FSS loop shape so its always best to tune the FSS first. 
The RF phase of the FSS was a mystery - I have no idea why it should do anything and I have never heard of this
and I don't know why I tried it today. But...changing it by ~0.6-0.7 slider units reduced the HFC by another factor
of ~3. Somebody should put this slider into units of degrees.8-)
Here's a table of the changes. Please make these the new nominals:you asked for: diff 2008/07/21,13:00 2008/07/22,2:44:16 utc '.*FSS.*|.*MC.*'
LIGO controls: differences, 2008 07/21 13:00:00 utc vs. 2008 07/22 02:44:16 utc
__Epics_Channel_Name______ __Description__________ __value1____ __value2____
C1:IOO-MC_REFL_GAIN 22.000000 19.000000
C1:IOO-MC_REFL_OFFSET 0.818380 0.818100
C1:PSL-FSS_MGAIN 10.000000 30.000000
C1:PSL-FSS_PHCON 2.073170 1.413170
The attached plot shows the "SERVO" TNC output of the board; this is supposed to be the same as the voltage going to the
VCO box. So its V/Hz transfer function is flat above 40 Hz. Tomorrow Jenne will post more data and remove the extender
board.
Since I only used an SR785, I only saw noise up to 100 kHz. Its key to use an RF spectrum analyzer when checking out
the FSS and the MC systems. |
709
|
Mon Jul 21 19:48:57 2008 |
Sharon | Update | | how tp restart C1ASS | How to restart C1ASS:
1. reboot
2. as root: caltech/target/c1ass:> ./startass
3. no need for root: burtgooey
that's it... |
708
|
Mon Jul 21 15:52:22 2008 |
steve | Bureaucracy | SAFETY | fire alarms test | The fire alarm test and evacuation drill was successfully completed at 13:45 Wednesday, July 16, 2008
Everybody was on time for the 40m meeting. |
707
|
Mon Jul 21 14:26:11 2008 |
Max | Summary | PEM | Added Channels | The following channels have been added.
Channel Name DAQ port
C1 : PEM-MAG_BSC_X 27
C1 : PEM-MAG_BSC_Y 28
C1 : PEM-MAG_BSC_Z 29
Jenne and I ran the wires from near the beam splitter chamber (as described in a previous elog) to the rack Y7 and plugged the labeled BNC's into ports 27-29. The computer was c0dcu1. John then restarted the frame builder and Alberto and I restarted the front end of c0dcu1 as per the wiki's instructions. The channels seem to be working. - Max. |
706
|
Mon Jul 21 11:54:00 2008 |
Jenne | Update | General | MC Servo Board | I pulled the MC Servo Board again, to check the components that are on the board, and compare them with the schematics. The filters that I'm interested in on the Fast Path haven't been changed. The high pass filters on the Fast Path have been changed.
Component Schematic Actual
--------- --------- ------
C140 10u open
C144 10u open
C149 open a gray Cap. value unknown
C141 10u open
C145 10u open
R97 1.58K 0
R99 open 1130
R103 open 1130
R100 open 0
R104 100 1130
R98 1.58K open
R109 367 365
Board is back in, and MC locks. |
705
|
Mon Jul 21 10:00:42 2008 |
steve | Update | PEM | ifo room temp | Five years of temp trend with 5 C fluctuation. The campus chilled water supply was upgraded ~ 1 year ago.
I requested our thermostats to be calibrated. |
704
|
Mon Jul 21 09:52:05 2008 |
Sharon | Update | | Adaptive code changes | Thanks to Alex, we now save the coefficients of the adaptive filter every cycle. When we choose ENABLE: OFF on the MEDM screen, suppressing the signal to the MC1, we stop and save the last coefficients. When enabling it again, it starts from the last coefficients saved. I will take some measurements today to check how this contributes to the adaptation rate. If you change the number of taps or the number of AUX channels, the coefficients are again set to zero. |
703
|
Sat Jul 19 19:41:56 2008 |
Yoichi | AoG | PSL | The author of the entry 702 is Yoichi not Rob | I made a mistake. |
702
|
Sat Jul 19 19:39:44 2008 |
rob | Update | PSL | PMC PZT investigation |
Quote: |
Quote: | The 64k is the series resistor after the PA85, not the modeled resistance of the PZT itself. |
Yes. What I meant was that because the measured voltage across the PZT was a half of the open voltage of the HV amp, the DC impedance of the PZT is expected to be similar to the output impedance of the HV amp. Of course, I don't think the DC impedance of a normal PZT should be such low.
I'm puzzled by the discrepancy between this expected DC impedance and the directly measured impedance by the Fluke DVM (1.5M Ohm).
One possibility is that the PZT leaks current only when a high voltage is applied. |
701
|
Fri Jul 18 23:24:24 2008 |
rob | Update | PSL | PMC PZT investigation |
Quote: | I measured the HV coming to the PMC PZT by plugging it off from the PZT and hooking it up to a DVM.
The reading of DVM is pretty much consistent with the reading on EPICS. I got 287V on the DVM when the EPICS says 290V.
Then I used a T to monitor the same voltage while it is connected to the PZT. I attached a plot of the actual voltage measured by the DVM vs the EPICS reading.
It shows a hysteresis.
Also the actual voltage drops by more than a half when the PZT is connected. The output impedance of the HV amp is 64k (according to the schematic). If I believe this number, the impedance of the PZT should also be 64k. The current flowing the PZT is 1.6mA at 200V EPICS reading.
The impedance of the PZT directly measured by the DVM is 1.5M ohm, which is significantly different from the value expected above. I will check the actual output impedance of the HV amp later.
The capacitance of the PZT measured by the DVM is 300nF. I don't know if I can believe the DVM's ability to measure C.
I noticed that when a high voltage is applied, the actual voltage across the PZT shows a decay.
The second plot shows the step response of the actual voltage.
The voltage coming to the PZT was T-ed and reduced by a factor of 30 using a high impedance voltage divider to be recorded by an ADC.
The PMCTRANSPD channel is temporarily used to monitor this signal.
After the voltage applied to the PZT was increased abruptly (to ~230V), the actual voltage starts to exponentially decrease.
When the HV was reduced to ~30V, the actual voltage goes up. This behavior explains the weird exponential motion of the PZT feedback signal when the PMC is locked.
The cause of the actual voltage drop is not understood yet.
From the above measurements, we can almost certainly conclude that the problem of the PMC is in the PZT, not in the HV amp nor the read back. |
I'd believe the Fluke's measurement of capacitance. Here's some info from PK about the PZT:
Quote: |
But the PMC ones were something like
0.750 in. thick x 0.287 in. thick. 2 microns per 200 V displacement,
resonant frequency greater than 65 kHz. Typical capacitance is around 0.66
uF.
|
If the PZT capacitance has dropped by a factor of two, that seems like a bad sign. I don't know what to expect for a resistance value of the PZT, but I wouldn't be surprised if it's non-Ohmic. The 64k is the series resistor after the PA85, not the modeled resistance of the PZT itself. |
700
|
Fri Jul 18 19:43:55 2008 |
Yoichi | DAQ | Computers | PSL fast channels cannot be read by dataviewer | At this moment only the PSL fast channels have trouble.
Rob restarted fb40m, c1IOVME, but no effect. |
699
|
Fri Jul 18 19:41:09 2008 |
Yoichi | Update | PSL | PMC PZT investigation | I measured the HV coming to the PMC PZT by plugging it off from the PZT and hooking it up to a DVM.
The reading of DVM is pretty much consistent with the reading on EPICS. I got 287V on the DVM when the EPICS says 290V.
Then I used a T to monitor the same voltage while it is connected to the PZT. I attached a plot of the actual voltage measured by the DVM vs the EPICS reading.
It shows a hysteresis.
Also the actual voltage drops by more than a half when the PZT is connected. The output impedance of the HV amp is 64k (according to the schematic). If I believe this number, the impedance of the PZT should also be 64k. The current flowing the PZT is 1.6mA at 200V EPICS reading.
The impedance of the PZT directly measured by the DVM is 1.5M ohm, which is significantly different from the value expected above. I will check the actual output impedance of the HV amp later.
The capacitance of the PZT measured by the DVM is 300nF. I don't know if I can believe the DVM's ability to measure C.
I noticed that when a high voltage is applied, the actual voltage across the PZT shows a decay.
The second plot shows the step response of the actual voltage.
The voltage coming to the PZT was T-ed and reduced by a factor of 30 using a high impedance voltage divider to be recorded by an ADC.
The PMCTRANSPD channel is temporarily used to monitor this signal.
After the voltage applied to the PZT was increased abruptly (to ~230V), the actual voltage starts to exponentially decrease.
When the HV was reduced to ~30V, the actual voltage goes up. This behavior explains the weird exponential motion of the PZT feedback signal when the PMC is locked.
The cause of the actual voltage drop is not understood yet.
From the above measurements, we can almost certainly conclude that the problem of the PMC is in the PZT, not in the HV amp nor the read back. |
698
|
Fri Jul 18 19:30:20 2008 |
Masha | Update | Auxiliary locking | moving from 40m | I will be working in the basement of Bridge probably starting next week; I moved the NPRO laser and some of the optics from my mach zehnder setup on the SP table to Bridge. Thanks for your help! |
697
|
Fri Jul 18 19:15:15 2008 |
Jenne | Update | IOO | Checking out the MC Servo Board |
Quote: | [Jenne, Max]
I haven't checked to ensure that the MC still locks, because Yoichi is doing stuff on the PSL table, but I didn't change anything on the board, and hooked all the cables back where they were, so hopefully it's all okay. |
I put the PMC back and the MC now locks. |
696
|
Fri Jul 18 17:12:35 2008 |
Jenne | Update | IOO | Checking out the MC Servo Board | [Jenne, Max]
One of the things that Rana thinks that might be causing my MC_F calibration to be off is that the MC Servo Board's filters don't match those on the schematics. Max and I pulled the MC servo board today to check resistor and capacitor values. Alberto needed the Mode Cleaner, so we put the board back before finishing checking values. We will probably pull the board again next week to finish checking the values.
I haven't checked to ensure that the MC still locks, because Yoichi is doing stuff on the PSL table, but I didn't change anything on the board, and hooked all the cables back where they were, so hopefully it's all okay. |
695
|
Fri Jul 18 17:06:20 2008 |
Jenne | Update | Computers | Computers down for most of the day, but back up now | [Sharon, Alex, Rob, Alberto, Jenne]
Sharon and I have been having trouble with the C1ASS computer the past couple of days. She has been corresponding with Alex, who has been rebooting the computers for us. At some point this afternoon, as a result of this work, or other stuff (I'm not totally sure which) about half of the computers' status lights on the MEDM screen were red. Alberto and Sharon spoke to Alex, who then fixed all of them except C1ASC. Alberto and I couldn't telnet into C1ASC to follow the restart procedures on the Wiki, so Rob helped us hook up a monitor and keyboard to the computer and restart it the old fashioned way.
It seems like C1ASC has some confusion as to what its IP address is, or some other computer is now using C1ASC's IP address.
As of now, all the computers are back up. |
694
|
Fri Jul 18 16:57:37 2008 |
Jenne | Update | IOO | Calibrated MC_F | I have calibrated MC_F. The conversion factor is 137.49 MHz/count.
The calibration data taken is attached, along with a calibrated power spectrum.
On the data plot, the x axis is volts from the C1:IOO-MC_FAST_MON channel, with the calibration between FAST_MON and MC_F = -788.18 volts/count.
The linear term of the fit line = -0.085MHz/volt. Error bars are +/- 1 in the last digit of what the spectrum analyzer gave me for frequency (+/- 0.01MHz).
The net conversion factor is then (-788.18)*(-0.085)*(2) = 137.49 MHz/count. The factor of 2 is because the light passes through the AOM twice.
On the power spectrum,
REF0 and REF1 = MC unlocked, HEPAs on, MC Refl gain = 22
REF2 and REF3 = MC locked, HEPAs on, MC Refl gain = 22
REF4 and REF5 = MC locked, HEPAs on, MC Refl gain = 19
REF6 and REF7 = MC locked, HEPAs off, MC Refl gain = 19 |
693
|
Fri Jul 18 12:24:15 2008 |
josephb, Eric | Configuration | Cameras | Changed Lenses on GC750 at ETMX | We removed the giant TV zoom lens and replaced it with a much smaller fixed zoom lens. Currently it views the entire optic. We have another (also small) zoom lens which focuses much better on the spot itself. With how far back the camera is currently placed, neither of these fixed zoom lenses can touch or hit the view port or the chamber while still attached to camera and mount, even using all of the mount's motion range. So this should be less of a safety issue.
Ideally, we'd like to get some images of the full optic (including osems and so forth) with the X-arm locked, and then use the higher zoom lens while still locked, to get images we can use to calibrate the x and y length scales. |
692
|
Thu Jul 17 20:13:34 2008 |
Yoichi | Update | PSL | PMC alignment/mode matching effort | I'm working to improve the mode matching of PMC.
Because I noticed that the beam was hitting the aperture of the EOM for PMC, I moved the EOM a little bit to maximize the transmission.
This did not change the alignment to the reference cavity but changed the alignment of the PMC a lot.
So I adjusted it back.
The alignment of the PMC can be easily optimized but the Hermite 02 mode still remains. This means the mode matching is bad.
Moving the lenses by a small amount (a few mm) did not change the height of 02 mode.
I'm planning to move the lenses by a large amount tomorrow. But it will destroy the alignment to the PMC.
So I installed two irises in the beam path after the lenses to remember the alignment roughly.
Right now the PMC transmission is slightly worse than before because the lens positions are not good. |
691
|
Thu Jul 17 16:39:58 2008 |
Max Jones | Update | DAQ | Magnetometer Installed | Today I installed the magnetometer near the beam splitter chamber. It is located on the BSC chamber at head height on the inner part of the interferometer (meaning I had to crawl under the arms to install it). I don't think I disturbed anything during installation but I think that's it's probably prudent to tell everyone that I was back there just in case. I plan to run 3 BNC cables (one for each axis) from the magnetometer to the DAQ input either tonight or tomorrow. Suggestions are appreciated. - Max. |
690
|
Thu Jul 17 13:08:37 2008 |
John | Summary | LSC | HOM resonances in the arms | On Tuesday night we attempted to lock the full DRFPMI in the optical spring configuration with the +f2 sideband resonant in the SRC.
Despite having no problems locking on the +f2 in a DRM we couldn't lock the full IFO.
There was some discussion about whether a +f2 higher order mode resonance in the arms could cause this problem.
I calculated the positions of the first six higher order modes for the carrier and all sidebands (using Siegman p 762 (23) with a plus sign).
Plot attached. Colors indicate different frequency components, numbers are the mode index (m+n). Thick lines are fundamental modes of
the sidebands. Heights of HOM indicators have been scaled by 1/(m+n)^2.
It appears that the first order transverse mode of the +166 is indeed partially resonant. We might try to tweak the sideband frequencies a
little to see if this helps us. It would probably be prudent to measure the MC length first.
Numbers used:
L = 38.5750; %average of Alberto's recent measurements elog #556
Retm = 57.375;
f166 = 165.977195e6;
f33 = 33.195439e6; |
689
|
Thu Jul 17 12:15:21 2008 |
Eric | Update | PSL | Swept PMC PZT voltage range | I unlocked the PMC and swept over C1:PSL-PMC_RAMP's full range a couple of times this morning. The PMC should now be relocked and returned
to normal. |
688
|
Thu Jul 17 08:30:15 2008 |
steve | Update | PSL | PMC relocked manually | The PMC pzt HV and the servo gain adj. are railing at max this morning
Why is it on the decreasing side of FSS_RTTEMP slope? |
687
|
Thu Jul 17 00:59:18 2008 |
Jenne | Summary | General | Funny signal coming out of VCO | While working on calibrating the MC_F signal, Rana and I noticed a funny signal coming out of the VCO. We expect the output to be a nice sine wave at about 80MHz. What we see is the 80MHz signal plus higher harmonics. The reason behind the craziness is to be determined. For now, here's what the signal looks like, in both time and frequency domains.
The first plot is a regular screen capture of a 'scope. The second is the output of the SR spectrum analyzer, as seen on a 'scope screen. The leftmost tall peak is the 80MHz peak, and the others are the harmonics. |
686
|
Wed Jul 16 22:29:05 2008 |
Masha | Update | Auxiliary locking | long measurement |
Quote: | I'm taking a measurement on the SR785 spectrum analyzer at low frequencies, so I'm going to leave it by the symmetric port table for a while. Please don't move it! |
all done thanks. |
685
|
Wed Jul 16 17:51:58 2008 |
Masha | Update | Auxiliary locking | long measurement | I'm taking a measurement on the SR785 spectrum analyzer at low frequencies, so I'm going to leave it by the symmetric port table for a while. Please don't move it! |
684
|
Wed Jul 16 17:36:51 2008 |
John | DAQ | PSL | FSS input offset | I changed the nominal FSS input offset to 0 from 0.3. Tolerance remains unchanged at +/-0.05. |
683
|
Wed Jul 16 16:59:07 2008 |
Alberto | Update | General | Aligment | I think the two beams are aligned again - they both pass the Faraday, they match at the irises and all along the optical path on the AP table. Although the NPRO beam does not show up at the AS port. |
682
|
Wed Jul 16 16:28:14 2008 |
josephb | Configuration | Computers | Fixed IP address on Switch | Realized today that the change I made back on June 30th to the switch was to the wrong switch. I had disabled the DHCP setting and mislabeled the switch in the control room (which seems to not have affected anything).
I've turned DHCP back on and labeled it correctly using the Netgear "Smartwizard discovery" program. |
681
|
Wed Jul 16 15:59:04 2008 |
josephb, Eric | Configuration | Cameras | PMC trans camera path | In order to reduce saturation, we placed a Y1 plate (spare from the SP table) in transmission just before the GC650 camera looking at the PMC transmision. The reflection (most of the light) was dumped to a convient razor blade dump. We also removed the 0.3 and 0.5 ND filters and placed them in the 24 hour loan ND filter box.
Good exposure values to view are now around 3000 for that camera. |
680
|
Wed Jul 16 11:26:47 2008 |
Max Jones | Update | | This Week | Baffles.
I got a battery for the magnetometer today which is slightly too large (~2 mm) in one dimension. Not sure what I'm going to do.
I'm attempting to calibrate the magnetometer but I'm having a hard time calibrating the axis that I cannot simply put through a coil parallel to the coils length. I have attempted to use the end fields of the solenoid but the measurements from the magnetometer are significantly different from the theoretical calculations.
I would appreciate suggestions. - Max. |
679
|
Wed Jul 16 11:00:15 2008 |
Masha | Update | Auxiliary locking | improving ADC input for the mach zehnder setup and completely unrelated happenings | For most of last week, the SURFs + Jenne were helping Mike and Ken with "stray light control for
Enhanced LIGO", i.e. cleaning and baking many many baffles which will catch scattered light in the
interferometer.
Otherwise, the two channels of the Mach Zehnder which will be used to measure fibre noise were
balanced, which should reduce the effect of laser amplitude noise in phase noise detection. I have
set up two digital channels to collect time series data from the two photodiodes and took some
preliminary noise measurements. I will be using Matlab to combine the signals as to directly measure
the phase noise, and I wrote some Matlab code to speed up this process: loading the files,
manipulating time series data, and converting into frequency domain. Currently I am building a
filter that will attenuate the signal at frequencies below 1Hz and amplify at higher frequencies in
order to whiten the spectra and reduce ADC noise. |
678
|
Wed Jul 16 10:50:55 2008 |
Eric | Summary | Cameras | Weekly Summary | Finished unwrapping, cleaning, baking, wrapping, wrapping again, packing, and shipping the baffles.
Attempted to set up the Snap software so that it could talk directly to EPICS channels. This is not currently working due to a series of very strange bugs in compiling and linking the channel access libraries. Alex Ivanov directed Joe and me to a script and makefile that are similar to what we're trying to do and it may solve our problem, but at the moment this still doesn't work. We're currently using a workaround that involves making unix system calls to ezca command line tools, but this is too hacky to leave in the final program.
Attempted to fit Josh's PZT voltage vs power plot of the OMC (from about a year ago) to lorentzians in order to try to develop fitting tools for more recent data. This isn't working, due to systematic error distorting the shapes of the peaks. Good fits can be obtained by cutting the number of points to a very small number around the peak of resonance, but this leads to such a small percentage of the peak being used that I don't trust these results either. (In the graph (shows the very top of the tallest peak): blue is Josh's original data, green is a fit to this peak using the top 66% of the peak and arbitrary, equal values for the error on each point, red is Josh's data averaged over bins of size 0.005, teal is a fit to these bins where the error on each point is the standard deviation of each bin, and magenta is a fit to these same bins, except cropped to the top ~10% of the peak, x-axis is voltage, y-axis is transmission power). Rana suggested that I take my own sweeps of the PMC using scripts that are already written: I'm currently figuring out where these scripts are and how to use them without accidently breaking something.
We've begun running the Snap software for long periods of time to see how stable it is. Currently, its only problem appears to be that it memory leaks somewhat: it was up 78% memory usage after a little over an hour. It doesn't put much strain on the computer, using only ~20% CPU. Stress put on the network from the constant transfer of images from the camera to the computer is not yet known. |
677
|
Wed Jul 16 09:27:17 2008 |
steve | Update | ALARM | PSL-FSS_RMTEMP alarm is false | Morning alarm sound is good for people who does not drink coffee.
Our 40m alarm is on every morning.
Those whom are not here in the morning thinks that this beeping sound is inspirational.
Would someone change this sound into less punishing form, like mockingbird chirp....
The C1PSL_SETTINGS.adl (40mm PSL Settings ) indicating that
C1:PSL-FSS_INOFFSET (Input Offset Adjust ) should be 0.3 +-0.05 V (red warning tag )
Alarm Handler: 40M pointing to yellow grade warning of PSL-FSS_RMTEM
This is a false alarm.
Two years trend of these channels are here: |
676
|
Tue Jul 15 19:15:57 2008 |
rana | Summary | SUS | ETMX Dewhitening characterization | Since the boys found that the ETM dewhitening transient was kicking the IFO out of lock we
decided to investigate.
First, we wrote a script to diagnose and then tune the DC gain of the dewhitening filters'
digital compensation filter (a.k.a. FM9 or SimDW). It is in the scripts/SUS/ directory
and is called dwgaintuner. It puts in an offset on each coil's DAC channel and
then reads back the Vmon on the coil driver with the DWF on and off. It reports the ratio of these
voltages which you can then type into the FM9/SimDW filter's gain field. We learned that the
difference between the analog DWF path and the bypass path was ~3% (which is consistent with
what you expect from the use of 1% resistors). We need to repeat this for all of the rest of
the suspended optics except for MC1 and MC3.
This Vmon method is better than what's used at the sites so we will export this new technology.
The attached plot shows some switching transients with only the local damping on:
BLUE: Output of filter bank during an FM9 turn off. This is the transient which goes to the DAC.
The transients are mostly of the same magnitude as this.
RED: This is the input of the filter module during another such transient.
GREEN: Tried another switch; this time I filtered the time series in DTT by typing the SimDW into
the Triggered Time Series filter field. This should be simulating what comes out of the
output of the DW board - to convert to volts multiply by 15/32768.
PURPLE: Same kind of filtering as the GREEN, but with also a double 30 Hz highpass to remove the low
frequency damping control signals. You can see that the total transient is only ~5 counts
or ~1 mV at the coil driver output. This is comparable to the relative offset in the bypass
and filter paths.
|
675
|
Tue Jul 15 12:44:08 2008 |
John | Summary | Locking | DRFPMI with DC readout | Rob, John
Last night, despite suspect alignment, we were again able to reduce the CARM offset to zero using
the RF signal.We were also able to transfer to dc readout taking calibrated spectra in both states.
DC readout shows a marked improvement over RF above ~1kHz but introduces some noise around 100Hz.
Broadband sensitivity appears to be more than ten times worse than previously. The calibration
being used remains to be confirmed.
Engaging the ETMY dewhitening caused lock to be lost. We'll check this today. The OMC alignment loops
may also need some attention.
We looked at REFL_166 as a candidate for CARM, at present POX still looks better.
The DARM filters were modified to reduce excess noise around 3Hz. Updating filter coefficients does
not cause loss of lock. |
674
|
Tue Jul 15 12:23:22 2008 |
Jenne | Update | General | MC2 Watchdog tripped | Alberto, Jenne
Mode Cleaner was unlocked. We checked, and found that MC2's watchdog was tripped. It didn't look like anything bad was going on, so we turned the optic back on, and tried to relock the MC. It looks like the Mode Cleaner is now locked, but the lock bit on the LockMC screen is still red. I don't know what's up. |
673
|
Tue Jul 15 11:47:56 2008 |
Jenne | DAQ | PEM | Accelerometer channels in ASS Adapt MEDM screen | Jenne, Sharon
We have traced which accelerometers correspond to which channels in the C1ASS_TOP MEDM screen.
Accelerometer Channel
------------- --------------------------
MC1-X C1:ASS-TOP_PEM_2_ADAPT_IN1
MC1-Y C1:ASS-TOP_PEM_3_ADAPT_IN1
MC1-Z C1:ASS-TOP_PEM_4_ADAPT_IN1
MC2-X C1:ASS-TOP_PEM_5_ADAPT_IN1
MC2-Y C1:ASS-TOP_PEM_6_ADAPT_IN1
MC2-Z C1:ASS-TOP_PEM_7_ADAPT_IN1
SEISMOMETER C1:ASS-TOP_PEM_1_ADAPT_IN1 |
672
|
Tue Jul 15 10:24:57 2008 |
steve | Update | PSL | PMC temp & pzt voltage | The PMC pzt HV was happy with no HEPA temp stability.
Can we thermally insulate the pmc ? |
|
ELOG V3.1.3- |