40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 321 of 335  Not logged in ELOG logo
ID Date Author Type Categorydown Subject Text Attachments
  8812   Tue Jul 9 16:08:32 2013 JenneUpdateASCPOP QPD calibration attempt[Jenne, Alex] 
Calibration data for the POP QPD
has been taken, with the PRMI locked on sideband
  
  8840   Fri Jul 12 21:23:42 2013 JenneUpdateASCPOP QPD calibration attemptThese are the data, one plot for when the
vertical QPD position was changed, and one
for when the horizontal (yaw) QPD position
 ChangeVertMicrometer_July2013.pngChangeSideMicrometer_July2013.png 
  8851   Mon Jul 15 17:16:59 2013 JenneUpdateASCPOP QPD calibration attemptI tried to retake POP QPD calibration data
again today.  The MC was mostly fine,
but whenever the PRMI unlocked, both ITM
 ChangeSideMicrometer_July2013_calib2.pngChangeSideMicrometer_July2013_calib1.pngChangeVertMicrometer_July2013_calib1.png 
  8854   Tue Jul 16 01:17:21 2013 JenneUpdateASCPOP QPD calibration attempt[Rana, Jenne]
I took POP QPD calibration data with
a new method, on Rana's suggestion. 
  
  8897   Tue Jul 23 01:30:27 2013 JenneUpdateASCPOP QPD analysisI have some data for how much motion of
any PRMI-relevant optic affects the beam
seen by the POP QPD. 
 9x 
  8926   Fri Jul 26 13:39:30 2013 KojiUpdateASCPOP QPD analysisIt was not actually easy to see from the
entry what signal was taken in what condition
but from the shape of the spectra
  
  8977   Wed Aug 7 15:32:37 2013 KojiUpdateASCASS setting up accelerated (slightly)I moved bunch of ezcawrite from the ASS
Dither On script to a snapshot file.
This accelerated a half of the "up"
  
  8980   Wed Aug 7 19:16:20 2013 JenneUpdateASCASS setting up accelerated moreI have furthered Koji's work, and moved
the filter on/off state for all the filter
banks also to the burt snapshot.  
  
  8982   Wed Aug 7 22:18:43 2013 KojiUpdateASCASS updateWhile Gautam is working on the Xarm green
ASS...
The EPICS monitor points for the
 Screenshot-Untitled_Window.png 
  8997   Mon Aug 12 14:05:34 2013 JenneUpdateASCPRCL ASS software in place

    

       
            
  
  8998   Mon Aug 12 15:39:40 2013 JenneUpdateASCPRCL ASS screens in placeI have added the PRCL ASS to the main ASS
screen, and created the servo and lockin
screens.  The filters loaded are the
  
  8999   Mon Aug 12 17:30:03 2013 KojiUpdateASCPRCL ASS software in placeWhy POPDC???   
  9001   Mon Aug 12 23:13:14 2013 JenneUpdateASCPRCL ASS software in place

    

       
            
  
  9126   Thu Sep 12 01:06:09 2013 JenneUpdateASCSRCL ASS implementedI have modified the ASS model to also have
an ASS for SRCL.  The input options
are POPDC, POP110, AS110.  I suppose
  
  9207   Sun Oct 6 20:55:08 2013 ranaSummaryASCMC WFS Limits set based on 40 days of trendsMC3 watchdog gets tripped sometimes when
lock is lost. I noticed that there were no
limits set in the MC WFS drive. The attached
 mcwrfs_trend.png 
  9249   Thu Oct 17 13:26:13 2013 JenneUpdateASCPOP QPD realignedI locked the PRMI, and tried to turn on
the ASS, but this caused PRMI to lose lock. 
Since this is similar to what happened
  
  9460   Thu Dec 12 21:30:52 2013 JenneUpdateASCPRMI-relevant oplevs centeredThe ITM oplevs were pretty close to the
edge of their ranges, and none of the oplevs
have been centered in a while, so I centered
  
  9469   Fri Dec 13 19:33:56 2013 DenUpdateASCETM X,Y QPDs I have modified/compiled/installed/restarted
c1scx and c1scy models to include arm transmission
QPDs in angular controls.
  
  9558   Wed Jan 15 18:42:57 2014 JenneUpdateASCPOP ASC QPD offline for a few hours this afternoonI was in the lab, near the south end of
the ITMX oplev table, looking for something,
and I bumped the POP ASC QPD's power supply. 
  
  9619   Mon Feb 10 18:59:25 2014 JenneUpdateASCPRM ASC better, but not great yetI have turned off the 3.2Hz res gains in
the PRC ASC loops, since those seem to make
the loops unstable. 
  
  9620   Mon Feb 10 19:56:10 2014 ranaUpdateASCPRM ASC better, but not great yetIgnoring the OSEM damping loops, the oplev
servo loops make it so that the POP ASC loops
do not see a simple pendulum plant, but instead
  
  9769   Mon Mar 31 23:57:22 2014 KojiSummaryASCPRM ASC characterization / designA
series of measurements / calculations for
the PRM ASC characterization and servo design
 7x 
  9773   Tue Apr 1 22:03:44 2014 KojiSummaryASCNew PRM ASC is running[Koji Jenne]
New PRM ASC was implemented. [to
be cnt'd]
  
  9777   Wed Apr 2 19:50:12 2014 KojiSummaryASCNew PRM ASC is runningAs the designed ASC filters in this
entry had too little phase margins (~10deg),
I had to compromise the servo design.
  
  9779   Wed Apr 2 23:08:51 2014 KojiSummaryASCNew PRM ASC is runningThe new PRM ASC design

Pitch
FM1: zero at 0Hz, pole at 2000Hz,
 PRM_QPD.pdfQPD_spe.pdf 
  9930   Thu May 8 14:37:02 2014 JenneUpdateASCPOP ASC QPD powerI was thinking about POP today, and wanted
to know if there was something to be done
to allow us to use the PRCL ASC for at least
  
  10114   Mon Jun 30 22:06:55 2014 JenneUpdateASCIFO (mostly) aligned - ASS working[Koji, Jenne]
The Yarm ASS is now working (as is
the Xarm ASS).  Both of the TT's pitch
  
  10124   Wed Jul 2 16:18:32 2014 JenneUpdateASCIFO aligned, PRMI + 2 arms achievedAfter the meeting, I aligned the IFO to
the IR, and then I aligned the Ygreen to
the Yarm.  I then found the beatnotes
  
  10278   Sat Jul 26 14:45:33 2014 GabrieleMetaphysicsASCResponse of POP QPD Koji asked me to perform a simulation
of the response of POP QPD DC signal to mirror
motions, as a function of the CARM offset.
 pop_qpd_all.pngpop_qpd_carrier.pngpop_qpd_sb.pngfoldeddoublecavity.mistpop_qpd.m 
  10337   Wed Aug 6 10:45:56 2014 GabrieleUpdateASCPOP QPD signals In addition to the simulation described
in my
previous elog, I simulated the signal
 pop_qpd_2f.png 
  10338   Wed Aug 6 12:44:52 2014 KojiUpdateASCPOP QPD signalsThis is nice. Can we test this idea with
POP22 + a razor blade?
Just to take transfer functions in
  
  10398   Fri Aug 15 01:31:44 2014 JenneUpdateASCPOP QPD to ETMs model wiring completeI have added a few things to the ASS model,
and the ASC sub-block, so that we can send
POP QPD information down to the ETMs for
 ASC_newCARM.pngASSoverview_newCARM.pngASC_screen.png 
  10454   Thu Sep 4 18:30:13 2014 GabrieleSummaryASCOptimal Gouy phase for POP QPD Jenne asked me to simulate the signals
on POP QPD when moving different mirrors,
as a function of the Gouy phase where the
 pop_qpd_dc.pngfortymeters.mistfortymeters_pop_qpd.mistpop_qpd.m 
  10503   Fri Sep 12 15:10:09 2014 JenneUpdateASCMICH ASSDuring the Sim meeting today, I added parts
to the ASS model so that we can also dither
the BS and minimize the power at AS. 
  
  10662   Mon Nov 3 17:14:00 2014 ericqUpdateASCIMC to IFO angular motionSomething to note, as we have the IMC angular
controls under consideration:
Jenne has the DRMI locked right now.
 2014-10-03-MC2T_to_POPQPD.pdf 
  10663   Mon Nov 3 17:43:14 2014 KojiUpdateASCIMC to IFO angular motionI wonder if this is the coherence caused
by the beam itself, or caused by the same
ground motion.
  
  10680   Thu Nov 6 12:53:09 2014 diegoUpdateASCX arm restored[Diego, Koji]
X arm has been restored, after modifying
the two parameters mentioned in http://nodus.ligo.caltech.edu:8080/40m/10676
  
  10745   Tue Dec 2 01:27:22 2014 diegoUpdateASCASS Scripts for armsI updated the medm C1ASS page for the Arm
scripts:
ON : same as before
  
  10752   Thu Dec 4 00:26:07 2014 JenneUpdateASCPOP yaw razor blade installedWe would like the option of feeding
back the POP beam position fluctuations to
the PRM to help stabilize the PRC since we
 ASC_PRCLloop_POP22err.pdfPRC_YAW_QPDvs22_3Dec2014.pdf 
  10768   Tue Dec 9 03:34:52 2014 JenneUpdateASCPOP yaw razor tuningWith the re-do of the IFO alignment last
week, I think that the beam was no longer
about halfway on the POP22 razor blade. 
 PRC_YAW_QPDvs22_8Dec2014.pdf 
  10789   Fri Dec 12 04:33:49 2014 JenneUpdateASCASS retuned[Rana, Jenne]
We decided that tonight was the night
for ASS tuning. 
  
  10807   Wed Dec 17 01:51:44 2014 rana, jenneUpdateASCASS retunedDid a big reconfig to make the Y-arm work
again since it was bad again.

    Undid Koji's topology
 ASSconfig_141217_0205.png 
  10811   Wed Dec 17 18:14:36 2014 ericqUpdateASCTransmon QPD -> ASC servos ready for comissioning I have completed all of the model
modifications and medm screen updates to
allow for feedback from the  transmon QPD
 newASCscreen.png 
  10812   Wed Dec 17 19:04:12 2014 jenneUpdateASCASS retunedI made the Xarm follow the new (old) topology
of Length -> test masses, and Trans ->
input pointing.
  
  10813   Wed Dec 17 19:31:55 2014 KojiUpdateASCASS retunedI wonder what to do with the X arm.
The primary purpose of the
ASS is to align the arm (=transmission),
  
  10886   Mon Jan 12 18:11:25 2015 ericqUpdateASCTest Mass -> Transmon QPD TFs measuredWe want to have some angular control of
the arms during lock acquistion. 

In single arm lock, Diego
 Jan12_singleArmSensing.pdfJan12_singleArmSensing.xml.zip 
  10908   Thu Jan 15 18:57:41 2015 ericqUpdateASCTransmon QPD loops liveI've measured the sensing for each
of the arms, by using our calibrated oplevs,
in terms of QPD counts per micron. It is:
 loopDesign.pdf 
  10920   Mon Jan 19 18:27:16 2015 ericqUpdateASCQPD ASC saga continues. Herein,
I will try
to paint a more thorough picture of this
 oplevPlant.pdfloopDesignComparison.pdfJan16_newLoops_offOn.png 
  10921   Tue Jan 20 02:39:49 2015 ericqUpdateASCQPD ASC saga continues. Although the QPD loops are less than
ideal right now, I made changes to the ASC
model to trigger the QPD loops on and off
  
  10946   Tue Jan 27 21:33:39 2015 KojiUpdateASCASS retunedI checked the situation of ASS. I wanted
to know how much we are away from the maximum
transmittion.
  
ELOG V3.1.3-