ID |
Date |
Author |
Type |
Category |
Subject |
10665
|
Tue Nov 4 10:40:46 2014 |
steve | Update | PEM | lab temperatures and particle counts |
|
Attachment 1: PEM100d.png
|
|
670
|
Tue Jul 15 09:47:09 2008 |
steve | Update | PEM | lab temps and particles |
All air condtion units were serviced last Friday.
AC filters are trying to control our particle counts but they have no capacity to match bad Pasadena conditions.
IFO room filters at CES were really clean.
Air make up filters inside and outside were dirty.
They showed the construction effect.
Control room and clean assembly units needed all filters replaced.
Note: the PSL-FSS_RCTEMP droped o.1C when enclouser HEPAs were turned back on
The RC temp controller should be better than that! |
Attachment 1: temps24d.jpg
|
|
11241
|
Thu Apr 23 23:07:23 2015 |
Dugolini | Frogs | ALARM | laptops warning |
Please!

Don't put laptops on the ISC Tables! |
1840
|
Thu Aug 6 09:05:29 2009 |
steve | Update | VAC | large O-rings of vacuum envelope |
The 40m-IFO vacuum envelope doors are sealed with dual viton O-rings and they are pumped through the annulos lines.
This allows easy access into the chambers. The compression of the o-rings are controlled by the o-ring grooves.
The OOC (output optic chamber)'s west side door has no such groove and it is sealed by just one single O-ring.
We have to protect this O-ring from total compression by 3 shims as shown below.
There were control shims in place before and they disappeared.
Let's remember that these shims are essential to keep our vacuum system in good condition.
|
Attachment 1: vacsor1.png
|
|
Attachment 2: vacsor2.png
|
|
11217
|
Tue Apr 14 11:19:52 2015 |
Steve | Update | PEM | large chamber has arrived |
It is here.
Quote: |
The 40m fenced area will start storing this large ~ 8000 lbs chamber on April 14. The asphalt will be cut, jack hammered the next 2-3 days in order to lay concrete.
Their schedule is from 8 to 5 starting tomorrow. We are asking them to work from 6 to 3pm
ETMX is about 12-15 ft away
|
|
Attachment 1: ETMXfriend.png
|
|
Attachment 2: 8000lbs.png
|
|
4989
|
Tue Jul 19 10:54:14 2011 |
steve | Update | General | large sensor card can not be found |
Please return sensor card to laser log box so others can use it. We have only one larger fluorescent sensor card. |
8311
|
Tue Mar 19 17:00:14 2013 |
Steve | Update | VAC | large window quote of Cascade |
Optical quality 6.5" OD window specs and quotes for vacuum view ports and optical table enclosure are on the 40m wiki_ Aux Optics page.
We have just received a very good quote #2 from Cascade Optical. |
7160
|
Mon Aug 13 15:31:09 2012 |
steve | Update | General | larger optical tables at the ends |
I'm proposing larger optical tables at the ends to avoid the existing overcrowding. This would allow the initial pointing and optical level beams to set up correctly.
The existing table is 4 x 2 would be replaced by 4' x 3' We would lose only ~3" space toward exist door.
I'm working on the new ACRYLIC TABLE COVER for each end that will cost around $4k ea. The new cover should fit the larger table.
Let me know what you think. |
Attachment 1: ETMYtable.jpg
|
|
Attachment 2: ETMY4X3.jpg
|
|
7163
|
Mon Aug 13 18:00:30 2012 |
jamie | Update | General | larger optical tables at the ends |
Quote: |
I'm proposing larger optical tables at the ends to avoid the existing overcrowding. This would allow the initial pointing and optical level beams to set up correctly.
The existing table is 4 x 2 would be replaced by 4' x 3' We would lose only ~3" space toward exist door.
I'm working on the new ACRYLIC TABLE COVER for each end that will cost around $4k ea. The new cover should fit the larger table.
Let me know what you think.
|
I'm not sure I see the motivation. The tables are a little tight, but not that much. If the issue is the incidence angle of the IP and OPLEV beams, then can't we solve that just by moving the table closer to the viewport?
The overcrowding alone doesn't seem bad enough to justify replacing the tables. |
7194
|
Wed Aug 15 16:01:47 2012 |
steve | Update | General | larger optical tables at the ends ? |
The drawing of the 4' x 2' table cover can be seen at entry #6190 The new proposed wall #7106 The yellow acrylic would be ~ 0.25" thick and it will be the inside. It is not shown on the drawing.
Question remaining: should get a larger table 4' x 3' as outlined by red lines and make new cover to fit this
The oplev beam path needs larger incident angle to get in and out of the chamber: REMOVE BOTTLENECK for easy traffic
Moving the existing table closer to ETMY chamber - as Jamie suggested- would help but there is no room for this solution.
The larger table solve this issue and leave more room for initial pointing, arm transmitted and future experiments.
Other benefits: no tube to make between table and chamber. It is easier to make the the larger box air tight.
The new isolation box with feed through, cover, seals will cost $4-5K ea
|
Attachment 1: bottleneck.jpg
|
|
9291
|
Fri Oct 25 10:45:16 2013 |
Steve | Update | PSL | laser drift monitor set up |
Quote: |
Quote: |
I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?
|
The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.
I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.
|
The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown. |
Attachment 1: laserDriftMon.jpg
|
|
Attachment 2: PMCT_120d.png
|
|
Attachment 3: PMCT_1000d.png
|
|
9292
|
Fri Oct 25 19:56:58 2013 |
rana | Update | PSL | laser drift monitor set up |
I went to re-align the beam into the PMC just now. I also tapped all the components between the laser and the PMC; nothing seems suspicious or loose.
The only problem was that someone (probably Steve or Valera) had closed down the iris just downstream of the AOM to ~1-2 mm diameter. This is much too tight! Don't leave irises closed down after aligning. An iris is not to b used as a beam dump. Getting it within a factor of 5-10 of the beam size will certainly make extra noise from clipping/scattering. After opening the iris, the reflected beam onto the PMC REFL camera is notably changed.
Not sure if this will have any effect on our worsening transmission drift, but let's see over the weekend.
I took pictures of this clipping as well as the beam position on Steve's new Retro Position Sensor, but I can't find the cable for the Olympus 570UZ. Steve, please buy a couple more USB data cables of this particular kind so that we don't have to hunt so much if one of the cryo (?) people borrows a cable.
Attachment shows PMC power levels before and after alignment. After alignment, you can see spikes from where I was tapping the mounts in the beamline. We ought to replace the U-100 mount ahead of the AOM with a Polanski
EDIT: Cryo team returns cable - receives punishments. Picture added. |
Attachment 1: PMC-IRSISSS.png
|
|
Attachment 2: PA250052.JPG
|
|
Attachment 3: PA280044.JPG
|
|
9547
|
Fri Jan 10 15:33:02 2014 |
Steve | Update | PSL | laser drift monitor set up idea |
this locationQuote: |
Quote: |
Quote: |
I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?
|
The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.
I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.
|
The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown.
|
IOO pointing is drifting in pitch. I'd like to use a QPD instead of the paper target to see if the Innolite output is stable. The idea is to move temporarily IOO-QPD_POS to this location |
Attachment 1: 2daysDrift.png
|
|
9552
|
Tue Jan 14 10:12:12 2014 |
Steve | Update | PSL | laser drift monitor set up idea |
Quote: |
this locationQuote: |
Quote: |
Quote: |
I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?
|
The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.
I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.
|
The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown.
|
IOO pointing is drifting in pitch. I'd like to use a QPD instead of the paper target to see if the Innolite output is stable. The idea is to move temporarily IOO-QPD_POS to this location
|
I do like to move IOO-QPD_POS temporarily to see that the feedback has anything to do with with the pointing. |
Attachment 1: bad4thday.png
|
|
6395
|
Fri Mar 9 16:00:46 2012 |
steve | Update | Green Locking | laser emergency shut down switch replaced at the south end |
Over-sized local laser emergency switch was held by large C clamp at the south end. This was replaced by a smaller one and it is mounted with magnets.
The Innolight laser was turned off, while the interlock was wired. |
3108
|
Wed Jun 23 17:48:16 2010 |
steve | Update | MOPA | laser head temp |
The laser chiller temp is fluctuating and the power output is decreasing. See 120 days plot.
Yesterday I removed ~300cc water from the overflowing chiller tank. |
Attachment 1: htemp120d.jpg
|
|
325
|
Wed Feb 20 11:34:17 2008 |
steve | Update | PSL | laser head temp is up |
MOPA head temp is running at 20.3C now
Nomally it is at 18.5C |
Attachment 1: htempup.jpg
|
|
12
|
Wed Oct 24 08:58:09 2007 |
steve | Other | PSL | laser headtemp is up |
C1:PSL-126MOPA_HTEMP is 19.3C
Half of the chiller's air intake was covered by loose paper |
Attachment 1: htempup.jpg
|
|
3033
|
Wed Jun 2 07:54:55 2010 |
steve | Update | MOPA | laser headtemp is up |
Is the cooling line clogged? The chiller temp is 21C See 1 and 20 days plots |
Attachment 1: htemp.jpg
|
|
Attachment 2: htemp20d.jpg
|
|
3035
|
Wed Jun 2 11:28:31 2010 |
Koji | Update | MOPA | laser headtemp is up |
Last night we stopped the air conditioning. It made HDTEMP increase.
Later we restored them and the temperature slowly recovered. I don't know why the recovery was so slow.
Quote: |
Is the cooling line clogged? The chiller temp is 21C See 1 and 20 days plots
|
|
6245
|
Fri Feb 3 14:47:51 2012 |
steve | Update | PEM | laser interlock drawing |
Rough draft of updated interlock drawing by Ben is here.
|
1280
|
Fri Feb 6 14:49:31 2009 |
steve | Summary | SAFETY | laser inventory |
40M Laser Inventory at Feb 05, 2009
1, Lightwave PA#102 @ 77,910 hrs 1064nm of 2.8W @ 27.65A
NPRO#206 @ 2.4A at PSL enclosure............"Big Boy" is waiting for to be retired but not now.
2, Lightwave NPRO 1064nm of 700mW #415 at AP table.......cavity length measurements of Alberto
3, CrystaLaser IRCL-100-1064S, 1064nmS of 100mW ,sn#IR81132 at east arm cabinet
4, CrystaLaser 1064nm of 180mW # -1274 flq at scattering setup.........flashlight quality
5, RF-PD tester 1064nm of 1.2mW @20mA at SP table
6, Lumix 800-1100nm of 500mW at east arm cabinet
7, JDS-Uniphase 633nmP of 4mW oplev sus laser at 5 places,
plus four spares in east arm cabinet
The same information is posted at the 40M WIKI also
|
3514
|
Thu Sep 2 16:41:32 2010 |
steve | Configuration | SAFETY | laser is ON: safety glasses required! |
I hooked up interlock to the Innolight 2W 1064 nm in the enclosure. The manual shutter is closed on this unit.
SAFETY GLASSES REQUIRED ! |
6242
|
Wed Feb 1 17:00:57 2012 |
steve | Update | IOO | laser is back ON |
Quote: |
The 2W PSL laser is turned off. The danger laser lights are not illuminated at the entry doors because of malfunctioning electronic circuit!!!
Laser safety glasses are still required! Other lasers are in operation!
|
BEN fixed the interlock. The laser is turned ON. Thanks for all, Rich and Sam who came over to help. Atm1
All emergency shut- off switches, lights and door indicators are working at this moment. More about this tomorrow.
Atm2, PSL enclosure interlock jungle without REAL schematic drawing.....at this point.... We all agreed it is easier to redo the hole thing than find the problem
Atm3, Emergency shut off switches and illuminated signs from entry doors to AC on-off box ( Use this switches in emergency ONLY, otherwise leave alone , even it is labeled obsolete !)
Summery: I still do not really know what was wrong.
|
Attachment 1: P1080525.JPG
|
|
Attachment 2: P1080514.JPG
|
|
Attachment 3: P1080518.JPG
|
|
7172
|
Tue Aug 14 08:43:42 2012 |
Steve | Update | IOO | laser off and on |
The janitor accidentally hit the laser emergency kill switch at room 103 entry door. It did shut down the PSL laser. The laser was turned back on. |
Attachment 1: 1day.png
|
|
365
|
Fri Mar 7 19:04:39 2008 |
steve | Omnistructure | PSL | laser pointer |
Green laser pointer was found in my desk.
I blamed Rana for not returning it to me after a conference talk.
It is surprisingly bright still.
I will bring sweets for Wednesday meeting. |
712
|
Tue Jul 22 09:24:17 2008 |
steve | Update | PSL | laser power |
Laser power reality of 120 days |
Attachment 1: power120d.jpg
|
|
1547
|
Tue May 5 10:42:18 2009 |
steve | Update | MOPA | laser power is back |
Quote: |
As PSL-126MOPA_DTEC went up, the power out put went down yesterday
|
The NPRO cooling water was clogged at the needle valve. The heat sink temp was around ~37C
The flow-regulator needle valve position is locked with a nut and it is frozen. It is not adjustable. However Jeenne's tapping and pushing down on the plastic hardware cleared the way for the water flow.
We have to remember to replace this needle valve when the new NPRO will be swapped in. I checked on the heat sink temp this morning. It is ~18C
There is condensation on the south end of the NPRO body, I wish that the DTEC value would just a little higher like 0.5V
The wavelenght of the diode is temp dependent: 0.3 nm/C. The fine tuning of this diode is done by thermo-electric cooler ( TEC )
To keep the diode precisely tuned to the absorption of the laser gain material the diode temp is held constant using electronic feedback control.
This value is zero now.
|
Attachment 1: uncloged.jpg
|
|
2142
|
Mon Oct 26 15:40:01 2009 |
steve | Update | PSL | laser power is down |
The laser power is down 5-6% |
Attachment 1: laserpowerdown.jpg
|
|
2147
|
Mon Oct 26 23:14:08 2009 |
Koji | Update | PSL | laser power is down |
I adjusted the steerings to the PMC and gained 7%. Now the MC_TRANS 7.0 has been recovered.
Actually I need another 7% to get MC_TRANS 7.5.
But I couldn't find how I can recover 126MOPA-AMPMON to 2.8ish.
Quote: |
The laser power is down 5-6%
|
|
Attachment 1: PSL091026.png
|
|
1542
|
Mon May 4 10:38:52 2009 |
steve | Update | MOPA | laser power is dropped |
As PSL-126MOPA_DTEC went up, the power out put went down yesterday |
Attachment 1: dtecup.jpg
|
|
1543
|
Mon May 4 16:49:56 2009 |
Alberto | Update | MOPA | laser power is dropped |
Quote: |
As PSL-126MOPA_DTEC went up, the power out put went down yesterday
|
Alberto, Jenne, Rob, Steve,
later on in the afternoon, we realized that the power from the MOPA was not recovering and we decided to hack the chiller's pipe that cools the box.
Without unlocking the safety nut on the water valve inside the box, Jenne performed some Voodoo and twisted a bit the screw that opens it with a screw driver. All the sudden some devilish bubbling was heard coming from the pipes.
The exorcism must have freed some Sumerian ghost stuck in our MOPA's chilling pipes (we have strong reasons to believe it might have looked like this) because then the NPRO's radiator started getting cooler.
I also jiggled a bit with the valve while I was trying to unlock the safety nut, but I stopped when I noticed that the nut was stuck to the plastic support it is mounted on.
We're now watching the MOPA power's monitor to see if eventually all the tinkering succeeded.
[From Jenne: When we first opened up the MOPA box, the NPRO's cooling fins were HOT. This is a clear sign of something badbadbad. They should be COLD to the touch (cooler than room temp). After jiggling the needle valve, and hearing the water-rushing sounds, the NPRO radiator fins started getting cooler. After ~10min or so, they were once again cool to the touch. Good news. It was a little worrisome however that just after our needle-valve machinations, the DTEC was going down (good), but the HTEMP started to rise again (bad). It wasn't until after Alberto's tinkering that the HTEMP actually started to go down, and the power started to go up. This is probably a lot to do with the fact that these temperature things have a fairly long time constant.
Also, when we first went out to check on things, there was a lot more condensation on the water tubes/connections than I have seen before. On the outside of the MOPA box, at the metal connectors where the water pipes are connected to the box, there was actually a little puddle, ~1cm diameter, of water. Steve didn't seem concerned, and we dried it off. It's probably just more humid than usual today, but it might be something to check up on later.] |
3202
|
Tue Jul 13 10:02:30 2010 |
steve | Update | MOPA | laser power is dropping slowly |
I have just removed an other 400 cc of water from the chiller. I have been doing this since the HTEMP started fluctuating.
The Neslab bath temp is 20.7C, control room temp 71F
|
Attachment 1: power100d.jpg
|
|
335
|
Fri Feb 22 14:45:06 2008 |
steve | Update | MOPA | laser power levels |
At the beginning of this 1000 days plot shows the laser that was running at 22C head temp
and it was send to LLO
The laser from LHO PA#102 with NPRO#206 were installed at Nov. 29, 2005 @ 49,943 hrs
Now,almost 20,000 hrs later we have 50% less PSL-126MOPA_AMPMON power |
Attachment 1: lpower1000d.jpg
|
|
865
|
Thu Aug 21 10:24:20 2008 |
steve | Configuration | SAFETY | laser safe mode condition |
The MOPA and PSL shutters are closed.
Manual beam blocks are in place.
Enclosure interlock is enabled.
No other high power laser is in operation.
We are in laser safe of operation for visiting students from Japan
NO safety glasses required
|
6347
|
Fri Mar 2 16:05:52 2012 |
Den | Update | SAFETY | laser safety |
Today I've attended the laser safety seminar. |
12921
|
Fri Mar 31 10:16:07 2017 |
Steve | Update | safety | laser safety glasses annual inspection |
Laser safety glasses cleaned in " Dawn Ultra " mild soap - water solution and measured for 1064 nm transmission at 150 mW
|
Attachment 1: march2017.jpg
|
|
Attachment 2: 2017march.jpg
|
|
8287
|
Wed Mar 13 16:04:24 2013 |
steve | Update | SAFETY | laser safety glasses checked |
All safety glasses were cleaned in soapy water by Bob. I measured their transmission at 1064 nm, 150 mW, beam diameter 1.5 mm They are in working order, no transmission.
Inventory:
10 pieces of KG-5, fit over, from Laser Safety
4 pieces of KG-5, std size, from Drever Lab, best visibility
1 piece of KG-5 coated for visible, std size, from Kentek
15 pieces of green-plastic LOTG-YAG, fit over, from UVEX
7 pieces of green-plastic B-D+S 137, std areo fit, from Sperian
3 pieces of green-plastic, old Thorlab, fit over
2 pieces of green-plastic, fit over, from Laservision
8 pieces of braun- plastic, fit over, for green & IR protection, from UVEX & Thorlabs |
Attachment 1: safetyglasseschecked.jpg
|
|
203
|
Wed Dec 19 16:40:12 2007 |
steve | Update | SAFETY | laser safety glasses measured |
I measured the coarse transission at 1064nm of the 40m safety glasses today.
12 pieces of UVEX # LOTG-YAG/CO2 light green, all plastic construction, ADSORBANT
3 pieces of 6KG5, Scott colored filter glass type,
individual prescription glasses: alan, bob, ben, jay and steve
7 pieces of dual waveleght glasses
These glasses showed 0.00mW transmission out of 170mW Crysta Laser 1064 |
6239
|
Tue Jan 31 08:44:10 2012 |
steve | Update | IOO | laser shuts down |
The 2W Innilight shutdown shut when I opened side door for safety scan. This was not a repeatable by opening -closing side doors later on. Turned laser on, locked PMC and MC locked instantly. The MC was not locked this moring and it seemed that the MC2 spot was still some high order mode
like yesterday. MC lock was lost when the janitor bumped something around the MC. |
Attachment 1: laseroff.png
|
|
6240
|
Tue Jan 31 14:58:30 2012 |
kiwamu | Update | IOO | laser shuts down |
[Steve/ Kiwamu]
We found that the laser had completely shut off for ~ 4 hours even with all the PSL doors closed.
We are guessing it is related to the interlock system and Steve is working on it to fix it.
Quote from #6239 |
The 2W Innilight shutdown shut when I opened side door for safety scan. This was not a repeatable by opening -closing side doors later on. Turned laser on, locked PMC and MC locked instantly. The MC was not locked this moring and it seemed that the MC2 spot was still some high order mode
like yesterday. MC lock was lost when the janitor bumped something around the MC.
|
|
4614
|
Tue May 3 15:48:26 2011 |
Larisa Thorne | Update | Electronics | laser temperature control LPF, final version! |
This is a continuation of this
The low pass filter is finally acceptable, and its Bode graph is below (on a ~3Hz frequency span that shows the cutoff frequency is at 0.1Hz) |
Attachment 1: LPF100k.2.jpg
|
|
9814
|
Tue Apr 15 13:24:42 2014 |
Steve | Update | PSL | laser turned on |
The 2W Innolight was off for 4 hours. |
Attachment 1: laserOff4hrs.png
|
|
Attachment 2: laserShutsDown.png
|
|
11373
|
Wed Jun 24 08:05:43 2015 |
Steve | Update | PSL | laser turned on |
The laser went off around 11am yesterday. It was turned on |
Attachment 1: laserWasOff.png
|
|
7853
|
Tue Dec 18 16:37:40 2012 |
Steve | Update | VAC | last RGA scan before vent |
|
Attachment 1: pd74m36d.png
|
|
Attachment 2: 36d_vacuum_normal.png
|
|
1009
|
Tue Sep 30 13:43:43 2008 |
rob | Update | Locking | last night |
Steady progress again in locking again last night. Initial acquisition of DRMI+2ARMs was working well.
Short DOF handoff, CARM->MCL, AO on PO_DC, and power ramping all worked repeatedly, in the cm_step script.
This takes us to the point where the common mode servo is handed off to an RF signal and the CARM offset
is reduced to zero. This last step didn't work, but it should just require some tweaking of the gains
during the handoff. |
1011
|
Wed Oct 1 00:24:54 2008 |
rana | Update | Locking | last night |
I had mistakenly left the MC boost off during my FAST investigations. The script is now restored.
The ISS is still saturating with gains higher than -5 dB. We need to request a PeterK / Stefan consult in the morning.
Also found the MZ gain down at -10 dB around midnight - need an alarm on that value. |
1024
|
Fri Oct 3 15:57:05 2008 |
rob | Update | Locking | last night, again |
Last night was basically a repeat of the night before--marginally better locking with the DRMI resonating the +f2
sideband. Several stable locks were achieved, and several control handoffs to DDM signals worked, but never from
lock to lock--that is, a given DD handoff strategy would only work once. This really needs to work smoothly before
more progress can be made.
Also, a 24Hz mode got rung up in one/several of the suspensions--this can also impede the stability of locks. |
6703
|
Tue May 29 15:29:16 2012 |
Jamie | Update | Computers | latest pynds installed on all new control room machines |
The DASWG lscsoft package repositories have a lot of useful analysis software. It is all maintained for Debian "sqeeze", but it's mostly installable without modification on Ubuntu 10.04 "lucid" (which is based on Debian squeeze). Basically the only thing that needs to access the lscsoft repositories is to add the following repository file:
controls@rossa:~ 0$ cat /etc/apt/sources.list.d/lscsoft.list
deb http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze contrib
deb-src http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze contrib
deb http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze-proposed contrib
deb-src http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze-proposed contrib
controls@rossa:~ 0$
A simple "apt-get update" then makes all the lscsoft packages available.
lscsoft includes the nds2 client packages (nds2-client-lib) and pynds (python-pynds). Unfortunately the python-pynds debian squeeze package currently depends on libboost-python1.42, which is not available in Ubuntu lucid. Fortunately, pynds itself does not require the latest version and can use what's in lucid. I therefore rebuilt the pynds package on one of the control room machines:
$ apt-get install dpkg-dev devscripts debhelper # these are packages needed to build a debian/ubuntu package
$ apt-get source python-pynds # this downloads the source of the package, and prepares it for a package build
$ cd python-pynds-0.7
$ debuild -uc -us # this actually builds the package
$ ls -al ../python-pynds_0.7-lscsoft1+squeeze1_amd64.deb
-rw-r--r-- 1 controls controls 69210 2012-05-29 11:57 python-pynds_0.7-lscsoft1+squeeze1_amd64.deb
I then copied the package into a common place:
/ligo/apps/debs/python-pynds_0.7-lscsoft1+squeeze1_amd64.deb
I then installed it on all the control room machines as such:
$ sudo apt-get install libboost-python1.40.0 nds2-client-lib python-numpy # these are the dependencies of python-pynds
$ sudo dpkg -i /ligo/apps/debs/python-pynds_0.7-lscsoft1+squeeze1_amd64.deb
I did this on all the control room machines.
It looks like the next version of pynds won't require us to jump through these extra hoops and should "just work". |
3024
|
Tue Jun 1 11:47:14 2010 |
steve | Update | PEM | lead balls on concrete |
Quote: |
Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.
There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.
You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.
|
The tiles were cut out in 1.5" ID circle to insure that the 7/16" OD lead balls would not touch the tiles on Wednesday, May 26, 2010
Granite surface plate specifications: grade B, 18" x 24" x 3" , 139 lbs
These balls and granite plate were removed by Rana in entry log #3018 at 5-31-2010 |
Attachment 1: P1060269.JPG
|
|