ID Date Authorup Type Category Subject Text Attachments
  2762   Sun Apr 4 00:21:42 2010 rana, kojiSummaryElectronicsCheckout of EG&G (PARC) preamp model #113, s/n 49135We tested out the functionality of the
EG&G 113 preamp that I found in one of
the cabinets. This is one of the ancestors
 IMG_0628.JPG 
  3903   Fri Nov 12 00:42:11 2010 rana, kojiUpdateIOOMC alignmentWe decided to ignore the computer script
outputs for the beam positions and use instead
the eyeball method to get the beam into the
  
  4253   Fri Feb 4 23:39:56 2011 rana, kojiUpdateLSCmixer based FD set up for noise testWe set up the mixer based FD to check
out its noise performance.
It is being acquired as C1:GCV-XARM_FINE_OUT_DAQ.
  
  4254   Sat Feb 5 23:03:04 2011 rana, kojiSummaryElectronicsAnalog Frequency Discriminator: splitter + mixer + long cableThis diagram shows the setup  mixer.pdfUntitled.pngfsm.png 
  6198   Sat Jan 14 00:50:08 2012 rana, kojiConfigurationIOOTowards coating thermal noise measurement with RefCav / MC beatKoji asked aloud tonight if we could   
  9324   Thu Oct 31 21:22:00 2013 rana, kojiSummaryIOOmodulation beat note in MC servoI hooked up the 4395 to the MC servo board
test out (TP2A) and looked at the spectrum
using our new SPAG4395.py script. We noticed
 13031.png13031_200.png13031_200b.png 
  9760   Fri Mar 28 22:10:00 2014 rana, kojiUpdateSUSrecovery from * EQ Southeast of LA around 45 minutes
ago. Callum and I felt it.
 Screenshot.png 
  10832   Mon Dec 22 21:53:08 2014 rana, kojiUpdateIOOSeven transfer functionsToday we were looking at the MC TFs and
pulled out the FSS box to measure it. We
took photos and removed a capacitor with
 MC_OLG.pdf 
  10833   Tue Dec 23 01:55:35 2014 rana, kojiUpdateIOOSeven transfer functionsSome TFs of the TTFSS box  MC_FSS_TF.pdf 
  10841   Tue Dec 23 20:50:39 2014 rana, kojiUpdateIOOSeven transfer functionsToday we decided to continue to modify
the TTFSS board.
The modified schematic can be found
 MC_OLTF.pdfMC_OLTF2.pdfmatlab.zip 
  11343   Tue Jun 2 21:22:07 2015 rana, kojiConfigurationIOOAOM inserted in beam and alignedWe
spent an hour today to put the AOM back in
  
  11361   Mon Jun 15 22:36:40 2015 rana, kojiUpdateGreen LockingBeatBox Assay: not looking goodBecause the ALS beatbox schematic is out-of-date
and misleading, we pulled the box to photograph
the current implementation and figure out
  
  11363   Fri Jun 19 01:24:26 2015 rana, kojiUpdateGreen LockingBeatBox Assay: not looking goodWe had decided a few days ago, to bypass
the IF part of the BeatBox board and put
some of the Haixing Maglev generic filter
  
  12639   Wed Nov 23 17:48:16 2016 rana, kojiUpdateIOOHow bad is the McWFS?Medium.


Previous elog entries on
this:
  
  14730   Fri Jul 5 23:28:52 2019 rana, kruthiSummarySUSETMX unstuck by shaking the stackWe unstuck ETMX by shaking the stack. Most
effective was to apply large periodic human
sized force to the north STACIS mounts.
  
  8741   Tue Jun 25 00:28:52 2013 rana, manasaUpdateLSCArm Cavity scan with X-ALS after ALS servo upgrade[Rana, Manasa]
ALS noise suppressed to 
 6x 
  3173   Wed Jul 7 22:52:38 2010 rana, nancyConfigurationIOObad length control offset for the MCRana found out that a connection was bad
in the shown place, due to which the MEDM
screen was showing bad offset for length
 P7070251.jpgScreenshot.png 
  9089   Fri Aug 30 01:01:28 2013 rana, nicSummaryComputer Scripts / ProgramsaLIGO Noise Budget code installed and runningChris Wipf has been developing a new Noise
Budget code that allows us to use our existing
Simulink models to handle all of the noise
 NBplots.pngScreen_Shot_2013-08-30_at_1.00.02_AM.png 
  1888   Tue Aug 11 23:55:04 2009 rana, richSummaryOMCQuantum Efficiency and Dark Current measurements of eLIGO PhotodiodesRich Abbott, Rana
Summary: We found that the 3mm
InGaAs photodiodes from eGTRAN which are
 IMG_0135.png 
  2095   Thu Oct 15 02:38:10 2009 rana, robUpdateOMCDark Port Mode Scan using the OMCBottom trace is proportional to the OMC
PZT voltage - top trace is the transmitted
light through the OMC. Interferometer is
 OMC-ModeScan_091015.png 
  2101   Fri Oct 16 03:16:50 2009 rana, robSummaryLSCfunny timing setup on the LSCWhile measuring the Piezo Jena noise
tonight we noticed that the LSC timing is
setup strangely. 
 1X3_1.JPG 
  2118   Mon Oct 19 14:48:15 2009 rana, robSummaryElectronicspiezo jena measuring boxAttached is the schematic of the Piezo Jena
driver measuring box made in a Pomona box:
 PA160153.JPGPA160151.JPG 
  2311   Mon Nov 23 00:46:09 2009 rana, robUpdatePSLISS RIN: Its too high by 10xThis plot shows the RIN as measured
by the ISS. Its ~2 x 10^-7, whereas its supposed
to be more like 3 x 10^-8.
 Untitled.png 
  2125   Tue Oct 20 11:38:10 2009 rana, rolfUpdateAdaptive Filteringextra delay and noise in PEM -> ASS/OAF systemAn email from Rolf about the delay in
the 110Bs: 
  
  3052   Sun Jun 6 08:08:05 2010 rana, sanjitSummaryElectronicsCapacitor Bridge TestTo get a feel for the Capacitive Bridge
problems, we setup a simple bridge using
fixed (1 nF) caps on a breadboard. We used
 CapacitanceBridge.png 
  1601   Mon May 18 19:44:52 2009 rana, steveConfigurationVACCryo Pump turned off and valved off: 1 beer can onlyI was seeing some excess noise in the ETMY
oplev yaw and so we turned off the cryo and
restarted c1vac2 to get the turbo pump channels
  
  2598   Fri Feb 12 14:19:28 2010 rana, steveHowToloreInternational FaxSteve showed me how to send an international
fax today:

    Load paper.
   
  
  3284   Sat Jul 24 13:13:41 2010 rana, steve, albertoUpdateGeneralInitial Crane Inspection reveals flaws: wiring and oilThe guy from KroneCrane (sp?) came today
and started the crane inspection on the X
End Crane. There were issues with our crane
  
  3573   Wed Sep 15 01:27:52 2010 rana, steve, valeraUpdatePSLFSS cables connected- connected the TTFSS cables (FSS fast
goes directly to NPRO PZT for now)
- measured the reference cavity 21.5
  
  4281   Mon Feb 14 00:39:21 2011 rana, sureshUpdateElectronicsVCO Frequency Noise Measurement with the MFDWe hooked up the VCO Driver output to the
MFD. We adjusted the levels with attenuators
to match up to the Level 7 mixer that's being
  
  3640   Fri Oct 1 21:34:14 2010 rana, taraUpdatePSLHigh Voltage Driver added to TTFSS -> NPRO

    
        
 NPROTF.png2010_10_01.png 
  3641   Mon Oct 4 06:47:46 2010 rana, taraUpdatePSLHigh Voltage Driver added to TTFSS -> NPROInside the FSS box, the FAST path has a
~10 Hz pole made up from the 15k resistor
and the 1 uF cap before the output connector.
  
  3542   Wed Sep 8 00:01:07 2010 rana, valeraUpdatePSLPMC updateWe ran the cables for the PMC: The RF cable
for the 35.5 MHz drive was cheap and so we
swapped the 29.5 MHz cable for it.
  
  4792   Mon Jun 6 23:56:16 2011 rana, valeraConfigurationSUSETMX/ETMY OSEM whitening We measured the OSEM PD whitening
transfer function of the ETMX OSEM UL whitening
stage (
  
  12573   Wed Oct 19 18:32:25 2016 rana, yinziUpdatePSLRefCav thermal control: heater is deadWe wanted to re-activate the Heater for
the reference cavity today to use it as a
testbed for PID autotuning and the new heater
  
  1314   Mon Feb 16 22:58:51 2009 rana, yoichiConfigurationSUSHysteresis in SUS from MisalignmentsWE wondered if there was some hysteresis in
the SUS alignments. When we leave the optics
misaligned for a
 a.png 
  1425   Wed Mar 25 01:37:35 2009 rana, yoichiSummaryIOONo Reference Cavity RequiredWe were wondering if we need to have a reference
cavity. One possible reason to have one is
to reduce the free running
  
  1484   Wed Apr 15 02:20:46 2009 rana, yoichiUpdateDMFDMF now working copyWe found that DMF/ was not an SVN working
copy, so I wiped out the SVN version, imported
the on-disk copy, moved it to DMFold/ and
  
  1563   Fri May 8 04:46:01 2009 rana, yoichiSummaryoplevsBS/PRM/SRM table bad!We went to center the oplevs because they
were far off and found that (as usual) the
numbers changed
 a.png 
  3438   Wed Aug 18 20:54:23 2010 rana, yoichiConfigurationEnvironmentChiller's chiller turned offWE programmed the MOPA chiller's AC unit
to turn off at 6PM each day and automatically
turn on at 7 AM each morning. Its now very
  
  7169   Tue Aug 14 04:32:49 2012 rana, yoichiUpdateLockingPOX signal sometimes looks very funny The alignment was way off.
We moved the PZT, the BS, and the x arm to
get it to lock. Along the way we noticed
 ITMX_1028974969.bmpETMXF_1028975007.bmpETMXT_1028975105.bmpAS_1028975166.bmp 
  10015   Mon Jun 9 22:26:44 2014 rana, zachUpdateCDSSLOW controls recovery All of the SLOW computers were
in limbo since the fileserver/nameserver
change, but me and Zach brought them back.
  
  554   Mon Jun 23 19:48:28 2008 rana,albertoSummaryIOOStochMon trends (80 days)Here's a StochMon plot showing the RFAM after
the MC. Remember that in these units, 2V
means no RFAM
 e.png 
  710   Mon Jul 21 19:55:16 2008 rana,jenneConfigurationIOONoise in MC_F[COLOR=darkgreen]Jenne put the MC board on
extender today - its still that way but everything
is probably connected (check AO).[/COLOR]
 SCRN0024.GIF 
  892   Wed Aug 27 13:55:43 2008 rana,jenneUpdatePSLPMC Servo BoardBoard is back in. PMC is locked.

Nominal gain is now 15 dB with brick. We
  
  895   Fri Aug 29 02:40:43 2008 rana,jenneUpdatePSLPMC Servo Board[quote]Board is back in. PMC is locked.
[/quote]
 pmclp-07.pdfpmclp_40m_080824.mpmclp.m 
  1208   Tue Dec 30 18:51:18 2008 rana,yoichiConfigurationElectronicsIlluminator Power Supply reset[COLOR=blue][B]We noticed that none of the
illuminators were working.[/B][/COLOR]
  
  13278   Thu Aug 31 00:19:35 2017 rana[^r]UpdatePSLIMC/FSS FAST gainnominal changed from 22 to 23 dB to minimize
PC drive RMS

  
  12937   Mon Apr 10 16:00:26 2017 rebeccaUpdateCamerasPylon installation warningWhen trying to install an older version
of Pylon packaged for Debian that Joe B.
had sent, it gave the warning that the package
  
  12956   Fri Apr 28 18:01:56 2017 rebeccaUpdateCamerasAttempting to Load Camera ClientUsing /ligo/apps/linux-x86-64/camera/bin/camera_client.py
-c  /opt/rtcds/caltech/c1/scripts/GigE/SnapPy/L1-CAM-MC1.ini,
the Python script was able to run without
  