40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 305 of 339  Not logged in ELOG logo
ID Date Author Type Categoryup Subject Text Attachments
  12518   Mon Sep 26 19:48:09 2016 LydiaUpdateSUSITMX stuck again, ITMY whitening issueThis afternoon around 2:45, ITMX started
ringing up at ~.9Hz for about a minute and
then
 44.png26.png 
  12519   Tue Sep 27 08:49:47 2016 SteveUpdateSUSseismic activity is upThe earth quake shook ITMX free for a 
short while.

 
 4.3mSaltonSee.pngITMXstuck.png 
  12520   Tue Sep 27 18:04:50 2016 LydiaUpdateSUSITMX slow channels down, ITMY diagonalization update[Teng, Lydia]

When we plugged in the back cables
yesterday on the whitening boxes after switching
 ITMY_diagsuccess.pdf 
  12523   Thu Sep 29 16:19:29 2016 LydiaUpdateSUSFree swing eigenmodes[Lydia, Teng]

Motivated by the strange pitch/yaw
coupling behavior we ran into while doing
 05.pngSUS_eigenmodes.png 
  12536   Thu Oct 6 15:42:51 2016 LydiaUpdateSUSOutput matrix diagonalizationSummary: At the 40m meeting yesterday,
Eric Q. gave the suggestion that we
accept the input matrix weirdness and adjust
  
  12540   Fri Oct 7 20:56:15 2016 KojiUpdateSUSOutput matrix diagonalizationI wanted to see what is the reason to have
such large coupling between pitch and yaw
motions.
 161007_P.pdf161007_Y.pdf 
  12541   Mon Oct 10 09:31:25 2016 SteveUpdateSUSPRM damping restoredLocal earth quake 3.7 mag  trips PRM

ETMY_UL glitch

What about the MC?
 3.7mLomaLinda.pngETMY_UL_glitch.pngMC_glitcing?.png 
  12546   Tue Oct 11 00:43:50 2016 ericqUpdateSUSPRM LR problematic againTonight, and during last week's locking,
we noticed something intermittently kicking
the PRM. I've determined that PRM's
 PRM_LR_Problem.pngsince_fix.pnglast10.png 
  12549   Tue Oct 11 10:15:04 2016 SteveUpdateSUSPRM LR problematic againIt started here

 
 PRMalfoiled.png 
  12550   Tue Oct 11 10:38:51 2016 SteveUpdateSUS wire standoffs update100 Sapphire prizms arrived.

 
 A028_-_20161010_170319.jpgA027_-_20161010_170202.jpg 
  12551   Tue Oct 11 13:30:49 2016 gautamUpdateSUSPRM LR problematic againPerhaps the problem is electrical? The
attached plot shows a downward trend for
the LR sensor output over the past 20 days
 PRM_LR_decay.png 
  12552   Wed Oct 12 13:34:28 2016 gautamUpdateSUSPRM LR problematic againLooks like what were PRM problems are now
seen in the SRM channels, while PRM itself
seems well behaved. This supports the hypothesis
 08.png 
  12553   Wed Oct 12 15:01:22 2016 steveUpdateSUSSOS ITM baffles plates are readyThe two 40 mm apeture baffles at the ends
were replaced by 50 mm one. ITM baffles with
50 mm apeture are baked ready for installation.
 baffle7x9_1.5.jpgbaffle_holder.jpgbaffle_top_view.jpg 
  12569   Wed Oct 19 08:28:11 2016 SteveUpdateSUSITMY_ULEverybody is happy, except ITMY_UL or satalite
box.

Gautam shows perfect form in the
 12hrs.pngvent79.jpg 
  12590   Tue Nov 1 09:03:08 2016 SteveUpdateSUSseismic activity is upSalton See is shaking again.

 
 seismicActivity.png 
  12598   Thu Nov 3 16:30:42 2016 LydiaConfigurationSUSETMX to coil matrix expanded[ericq, lydia]

Background: 

We believe the optimal OSEM damping
 01.png 
  12601   Mon Nov 7 08:00:11 2016 SteveUpdateSUSSRM -PRM sat. amp swapI just realized that Gautam set this test
up and turned damping off......He will explane
the details
 SRM.jpgSRM-UR_OK.png 
  12602   Mon Nov 7 16:05:55 2016 gautamUpdateSUSPRM Sat. Box. DebuggingShort summary of my Sat. Box. debugging
activities over the last few days. Recall
that the SRM Sat. Box has been plugged into
 D961289-B2.pdfPRMSatBoxtest.png 
  12606   Tue Nov 8 11:54:38 2016 gautamUpdateSUSPRM Sat. Box. looks to be fixedLooks like the PRM Sat. Box is now okay,
no evidence of the kind of glitchy behaviour
we are used to seeing in any of the 5 channels.
  
  12612   Sun Nov 13 23:42:43 2016 LydiaUpdateSUSETMX output matrix data I took data of the ETMX SUSPOS, SUSPIT
and SUSYAW channels while driving each of
the 4 face coils. I manually turned off all
  
  12643   Mon Nov 28 10:27:13 2016 gautamUpdateSUSITMY UL glitches are backI left the tester box plugged in from Thursday
night to Sunday afternoon, and in this period,
the glitches still appeared in (and only
 ITMY_satboxSpectra.pdfITMY_UL_testerBox.png 
  12650   Wed Nov 30 14:53:56 2016 SteveUpdateSUSnew sus wire stored in N2 filled dessicatorThe new SOS sus wire finally is stored
in a nitrogen filled dessicator. This was
recommended by Ca. Fine Wire to minimize
 dessicatorC.jpgwireN2c.jpg 
  12720   Sat Jan 14 22:39:30 2017 ranaSummarySUSITMY is drifting ?https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20170114/sus/susdrift/

ITMY is not like the others. Real
or just OSEM madness?
  
  12722   Mon Jan 16 18:54:01 2017 ranaUpdateSUSBS: whitening re-engagedFound that the BS whitening was off. Gautam
says that "it has always been that way"
and "there's nothing in the elog
 BSwhite.pdf 
  12725   Mon Jan 16 23:25:07 2017 gautamUpdateSUSMC1 SUS electronics investigation[rana,gautam]

Summary:


 20170116_231625.pngIMG_7175.JPGIMG_7174.JPG 
  12728   Tue Jan 17 21:29:52 2017 gautamUpdateSUSMC1 SUS electronics investigation 




  
  12731   Wed Jan 18 11:40:54 2017 gautamUpdateSUSMC1 SUS electronics investigationAfter the repair of the faulty LEMO cable,
I left MC1 with it's watchdog off overnight.
 MC1glitches.pngMC1coilGlitches.pngMC1CoilGlitchZoomed.png 
  12734   Wed Jan 18 14:23:47 2017 gautamUpdateSUSMC1 SUS electronics investigationAs part of the ongoing   
  12736   Wed Jan 18 18:44:53 2017 gautamUpdateSUSMC1 SUS electronics investigation


 MCsatBoxSwitch.png 
  12737   Thu Jan 19 08:25:12 2017 SteveUpdateSUSMC1 SUS electronics investigation


 MC1_MC3_ITMY_ETMX_sensors.pngsensors_UL.png 
  12739   Thu Jan 19 12:00:10 2017 gautamUpdateSUSMC1 SUS electronics investigationGoing through the last ~20 hours of data,
the MC1 sensor channels look glitch free
the entire period. However, there is a ~10min
 SatBoxSwitchGlitch.png 
  12741   Thu Jan 19 19:56:09 2017 ranaUpdateSUSMC1 SUS electronics investigationMight be. Or it might be in the satellite
box cabling. Hard to tell without a tester.
I recommend you squish the cables on there
  
  12742   Fri Jan 20 11:16:30 2017 gautamUpdateSUSMC1 SUS electronics investigationBoth suspensions have been relatively well
behaved for the best part of the last two
days, since I effected the Satellite Box
 MCglitchesReturn.png 
  12745   Mon Jan 23 10:24:01 2017 SteveUpdateSUSMC1 SUS electronics investigationTwo day plot of glitching suspentions:
MC3, ITMY and ETMX
 3glitchingSUS.png 
  12753   Wed Jan 25 10:46:58 2017 steveSummarySUSoplev laser summary updated                   
Oct.  5, 2015             
ETMY He/Ne
 oplev_sums.png 
  12760   Fri Jan 27 14:50:04 2017 SteveUpdateSUS wire standoffs updateThe 3 pieces of Sapphire
v-groove test cuts are back. They look
good. The suspension wire 0.0017" (
  
  12761   Fri Jan 27 15:36:17 2017 KojiUpdateSUS wire standoffs updateVery nice! I got excited.


Don't you ask Calum
and co to check the groove size with their
  
  12802   Mon Feb 6 10:05:28 2017 SteveUpdateSUSclamped cablesThe bottom 5 cable connections from Sat-Amp
to Whittering Filter at 1X5 were clamped
today.
 clamped.jpg 
  12811   Wed Feb 8 10:16:39 2017 steveUpdateSUSclipping ITMX oplevThe ITMX oplev beam is clipping. It will
be corected with locked arm

 
 ITMX_oplev_clipping.jpgITMX_clipping.jpg 
  12846   Thu Feb 23 09:32:20 2017 KojiUpdateSUS wire standoffs updateKyle took high quality images of 
the three sapphire prisms using the microscope
@Downs. He analyzed the images to see the
 Sapphire_prism_1(A015).pngSapphire_prism_2(A016).pngSapphire_prism_3(A014).png 
  12886   Tue Mar 14 10:40:30 2017 SteveUpdateSUSOSEM filters are inWe have 50 pieces in the clean cabinet.  filters.jpg 
  12889   Thu Mar 16 08:22:16 2017 SteveUpdateSUSETMX damping Finally I see what kicks the sus damping
off

 laser_power_glitch.png 
  12892   Fri Mar 17 15:30:39 2017 SteveSummarySUSoplev laser summary updated   
     March  17, 
2017        
 oplev_sums.png 
  12931   Fri Apr 7 13:46:23 2017 SteveUpdateSUSETMX enclosure feedthoughETMX enclosure feedtrouh cabeling corrected.  bad.jpggood.jpg 
  12933   Mon Apr 10 09:58:35 2017 SteveSummarySUSoplev laser summary updated 




  
  12941   Thu Apr 13 09:48:37 2017 SteveUpdateSUSITMY-UL and ETMX sensorsWhy ITMY UL can not see this earth quake?
SRM and PRM are misaligned. ETMX is still
not well.
 ITMY.pngITMY-UL.pngETMX?.png 
  12995   Wed May 17 08:19:59 2017 SteveUpdateSUS4.1M earthquake Sus dampings recovered. ETMY oplev needs
to be recentered.

 eq_4.1_SantaBarbara.png4.1m_Isla_Vista_CA.png 
  13030   Thu Jun 1 16:21:55 2017 SteveUpdateSUS wire standoffs updateRuby
wire standoff received from China. I
looked one of them with our small USB camera. 
  
  13039   Mon Jun 5 10:30:45 2017 SteveUpdateSUSruby wire standoff picturesAtm 1 & 5, showing the ruby R ~10 mm
as it is seated on Al SOS test mass

Atm. 2, 3 & 4  chipped
 A087_R.png.bmpA097_chipped_edges.png.bmpA099_cal_wire.png.bmpA101_cal_wire_43_micron.png.bmpAl_SOS_R39mm.jpg 
  13123   Mon Jul 17 16:22:01 2017 SteveUpdateSUSruby wire standoff picturesBluebean Optical Tech Limited of Shanghai
delivered 50 pieces red ruby
prisms
 contacting_ridge.bmpcontacting_ridge.png 
ELOG V3.1.3-